PROCESS SENSING PST-UM-2010-EN-02 Minox i Oxygen Transmitter User Manual

June 9, 2024
PROCESS SENSING

Minox i Oxygen Transmitter
User Manual PST-UM-2010-EN-02 User Manual

Before using your Minox i

Safety information
Please read this manual, ensuring that you fully understand the content before attempting to install, maintain or use the Minox i. Important safety information is highlighted  throughout this document as follows:
The warning symbol indicates instructions that must be followed to avoid minor, serious or even fatal injury to personnel.
The caution symbol indicates instructions that must be followed to avoid damage to equipment (hardware and/or software) or the occurrence of a system failure.
NOTE highlights an essential operating procedure, condition, or statement.

Abbreviations

AC Alternating current
DC Direct current
oC Degrees celcius
oF Degrees fahrenheit
EC Electrochemical
g Grams
GND Ground
kg Kilogram
LED Light emitting diode
LPM Liters per minute
mA Milliampere
oz Ounces
PC Personal computer
PCB Printed circuit board
PLC Programmable logic controller
PPM Parts per million
SCFH Standard cubic feet per hour
S-S Solid-state
SS Stainless steel

This page is intentionally blank.
Introduction to Minox i
This manual is applicable to Minox i oxygen transmitter with electrochemical (EC) or solid-state (S-S) sensors; and KF-40 flange, flow-through or 2” tri- clamp connections.

Overview

ATEX, IECEx and UKCA certified, the Minox i is an intrinsically safe, 2-wire, loop-powered transmitter, with a 4…20mA output. It is designed with proven sensor technology to accurately measure O₂ in a variety of gases, in the most demanding applications and hazardous environments. Minox i is available with the three process connections shown below.
NOTE: You must provide suitable mating flange and clamp for KF-40 and 2” tri-clamp connections.

Figure 1. Three process connections for Minox i

Minox i can be used with a Programmable Logic Controller (PLC) or suitable equipment capable of providing a 4…20 mA loop-power interface. For operation in hazardous  Ex-rated areas, the interface must be provided via a suitable isolation barrier device.
The Minox i oxygen transmitter must only be operated in the way set out in this user manual.
Always verify electrical interface circuit compatibility before connection.

1.2 Structure
The Minox i oxygen transmitter consists of an oxygen sensor and inbuilt microprocessor. They are encased in stainless steel, with a process connection on the removable  housing and an electrical signal connector on the main housing.
The sensor is replaceable and can be accessed via the removable housing, as shown in Figure 2:

Figure 2. Structure of Minox i with EC sensor (KF-40 flange pictured)

Sensors

1.3.1 Electrochemical % (EC%)
The Minox i EC% oxygen sensor is a galvanic fuel cell capable of superior performance, accuracy, and stability. Its operational life is typically up to 18 months in ambient air.  The sensor is designed to be unaffected by the presence of other background gases.
The Minox i is required to be mounted in a vertical orientation with the electrical signal connector at the top. For gases with a composition of hydrocarbons and/or CO2, the  solid-state sensor should be used when % level O2 measurements are required.

1.3.2 Electrochemical ppm (EC ppm)
The Minox i can be specified with a choice of EC ppm oxygen sensors. Both are galvanic fuel cells with different chemistry. With both EC ppm sensors, the Minox i must be  mounted vertically with the electrical signal connector at the top.
The OC-60 sensor is the ppm version of the EC% and is suitable for inert, H2 and He gases. The OC-69 sensor is compatible with gases that have a hydrocarbon composition or CO2 element.

1.3.3 Solid-state (S-S)
The Minox i S-S sensor is not gas-dependent or specific, and offers a robust measurement capability on process gas streams or in containment environments. It can operate with or without sample flow, making it ideal for in-situ monitoring, and it is not orientation-specific. With the tri-clamp and KF-40 process connections, the Minox i can be mounted on a T-piece of pipeline or on the vertical panel of a containment system without concern. The transmitter provides a linear output with a sensitivity that is constant throughout its operating range.

Applications

Typically, the intrinsically safe Minox i oxygen transmitter can be used in the following applications:

  • Inerting
  • Glove box purge and containment solutions
  • Sieving and powder transfer systems (PTS)
  • Additive manufacturing
  • Hydrogen generation
  • Nitrogen generation
  • Biogas and biomethane
  • Pharmaceutical.

Installation

Before installing your Minox i oxygen transmitter, ensure that its specifications are suitable for the process or pipeline where it will be installed.
See “2.2.1 KF-40 flange and 2” tri-clamp process connections with dimensions” on page 5 or “2.2.2 Flow-through process connection with dimensions” on page 6 for more  guidance.
This product is not approved for use in applications where the sample gas is greater than 21 % volume O.

2.1 Unpack the Minox i
NOTE:
The sensor is packed in a sealed package that has been purged to remove O2 from the packed environment. It should not be removed from the package until the  transmitter has been installed at the measurement point and you are ready to perform measurements.
Your Minox i oxygen transmitter pack is comprised of the following equipment:

  1. Minox i oxygen transmitter
  2. Electrochemical or solid-state oxygen sensor
  3. A MagTip calibration tool
  4. Certificate of calibration
  5. Cable fitted with M12 connector.

Leaving a sensor in ambient air conditions will reduce sensor life.

2.2 Mount the Minox i
The Minox i transmitter requires a two-wire industry-standard 4…20 mA loop- powered connection, which you must be provide. It must be installed in the measurement point before the sensor is removed from its sealed package.
For flow-through connection, continue to “2.2.2 Flow-through process connection with dimensions” on page 6.

2.2.1 KF-40 flange and 2” tri-clamp process connections with dimensions
A suitable mating flange and seal/centering ringwill be required to install the KF-40 flange or triclamp Minox i. These are not included in the pack.

Figure 4. Technical drawing of KF-40 flange and 2” tri-clamp process connections

  1. Place the seal/centering ring on the pipe where the Minox i will be installed.
  2. Put the process connection end of the Minox i into the pipe.
  3. Place your flange clamp or tri-clamp over the seal/centering ring and the Minox i where they meet the pipe.
  4. Fasten the flange or tri-clamp until finger-tight.
  5. Continue to “2.2.3 Electrical connection” on page 7.

2.2.2 Flow-through process connection with dimensions
The flow-through Minox i is designed for measurement on an extracted sample gas. The sample gas should be at atmospheric pressure +/- 100 mbar (xxx psig), and a flow-rate of 250…500 ml (0.5…1.0 SCFH) is required.
If the sample gas is at a pressure regulator, a flow-meter must be placed before the transmitter. The transmitter must be able to vent the gas to atmosphere or an atmospheric vent.
If the sample gas is at sub-atmospheric pressure, the sample must be drawn through the flowthrough block. Ntron can supply systems to aid this installation upon request. Figure 5. Technical drawing of flow-through process connection

  1. Align the flow-through Minox i 1/8” NPT with your pipe fitting.
  2. Fasten screws to adapt to pipe fitting.

2.2.3 Electrical connection

Figure 6. Typical connection cable wiring

A suitable galvanic isolation barrier is required if the Minox i is located in an Ex area.
Consult the EC-type examination certificate Baseefa19atex0200 for Parameter Matching.
The Minox i oxygen transmitter operates using a 2-wire industry-standard 4…20 mA loop-powered supply which you must provide.
Once the Minox i is connected to your process, it is ready to be connected to your power supply.
To do this:

  1. Align the pins in the Minox i electrical signal connector with the M12 mating connector on the cable.
    The cable connector can then be gently pushed into place.

  2. Rotate the fastener clockwise until finger tight. Your Minox i oxygen transmitter is now connected. A pulsing red LED will indicate it is operating correctly.

2.2.4 Install your sensor
The sensor supplied should only be removed from the sealed package when the transmitter is installed and ready to make a measurement.
Installing the sensor is straightforward, but should be done quickly to limit exposure to ambient O2 levels, this is particularly important for the EC ppm sensors. Prolonged  exposure to air will shorten the operational life of the sensor.
As a guideline, you should install the sensor into the Minox i, and the mount the transmitter in the measurement environment in 15-30 seconds. To install the sensor, please  see “4.2 Replace the sensor” on page 12.

Calibration

NOTE: Calibration should be carried out by an appropriately trained or suitably qualified person.
The oxygen sensor output will decrease over time, and periodic calibration will be required. This is a non-intrusive procedure, achieved by using the MagTip one-touch  calibration tool included with your Minox i.
Calibration cycles for the Minox i depend on the application, the applied gases, and the sensor type. They can be carried out in-situ or after safely removing the oxygen  transmitter from your process in a bench test.
NOTE: For optimal calibration, the atmosphere surrounding the sensor must be ambient air (20.9 % volume O2) with no contaminants present. This can be achieved using  an oxygen analyzer or by flushing the system with clean, oil- free compressed air.
The Minox i ppm must be calibrated in a controlled environment. Calibrating the Minox i % in ambient air will significantly reduce sensor life.
When the MagTip tool is applied, the indicator LED will flash six times in rapid succession. This confirms calibration is in progress.
During normal operation, the indicator LED will flash once every two seconds. This slower flash confirms Minox i is powered up and operational.

3.1 In-situ
3.1.1 Minox i ppm/%
Please see above before proceeding.
1. Apply the calibration gas (20.9 % volume O2) for a minimum of 15 minutes.
The pressure flow should replicate that applied to the Minox i during normal operation.
This will ensure calibration is accurate, and avoid over-pressurizing the oxygen sensor.

  1. Allow one to two minutes for the output to stabilize (i.e reach its ambient air concentration value and remain at the value). An independent portable oxygen analyzer can  be used to check this.

  2. Observe the O2 concentration on your oxygen analyzer display. It should read 20.9 % volume O2.

  3. Place the MagTip on the Minox i main housing, adjacent to the red indicator LED, as shown in Figure 8 below.
    The indicator LED will begin to flash rapidly.

  4. When calibration is complete, the indicator LED will resume flashing at a steady pulse. For the standard 0…25 % O2 range, the 4…20 mA output will be automatically  adjusted to 17.376 mA.
    This value, or the reading 20.9%, must be observed to confirm calibration is complete.
    When calibration is complete, the Minox i LED will resume flashing at a steady pulse.

 Air calibration

The following method of air calibration should only be performed on a Minox i %.

  1. Unfasten and remove the Minox i from your process, and leave it in an ambient air environment. Ensure the environment has normal oxygen concentration, an independent portable oxygen analyzer can be used to check this.

  2. Observe the O2 concentration on your oxygen analyzer display. When the concentration has stabilized it should read 20.9 % volume O2.
    This usually takes between five and ten minutes.

  3. Place the MagTip on the Minox i main housing, adjacent to the indicator LED, as shown in Figure 8. The red indicator LED will begin to flash rapidly.

  4. When calibration is complete, the indicator LED will resume flashing at a steady pulse.

  5. Re-install your Minox i following the instructions “2.2.1 KF-40 flange and 2” tri-clamp process connections with dimensions” on page 5 or “2.2.2 Flow-through process connection with dimensions” on page 6, depending on your process connection.

Maintenance

4.1 General
To maintain performance, the sensor in Minox i will require replacement. Please check the Technical Specifications for frequency. During sensor replacement, it is recommended that light cleaning of the external housing is carried out, as well as the checks outlined in this section. Never use chemical cleaning agents or high pressure water or steam to clean the equipment. Do not submerge in water.

4.2 Replace the sensor
The Minox i oxygen sensor is replaceable. When it reaches the end of serviceable life, calibration can no longer be performed and the sensor must be replaced.
A regular program of calibration will mitigate against sudden sensor failure. It is advisable to establish a program of preventative maintenance to ensure process downtime is  kept to a minimum or avoided.
The sensor supplied should only be removed from the sealed package when the transmitter is installed and ready to make a measurement.
Installing the sensor is straightforward, but should be done quickly to limit exposure to ambient O2 levels, this is particularly important for the EC ppm sensors. Prolonged  exposure to air will shorten the operational life of the sensor.
As a guideline, you should install the sensor into the Minox i, and the mount the transmitter in the measurement environment in 15-30 seconds.
To replace your sensor:

  1. Unscrew the removable housing from the main housing of Minox i by turning it clockwise.
    The sensor will be located in the removable housing, as shown.

  2. Gently tip the sensor out of the removable housing into your hand or onto a soft surface.

Figure 10. Oxygen sensor placement in removable housing

3. Place your new EC or S-S oxygen sensor into the removable housing, ensuring the sensor’s brass rings are facing out of the removable housing as shown in Figure 10.
4. Screw the removable housing back onto the main housing, by turning it counter-clockwise.
NOTE: The brass slip rings on the sensor must face into the main housing and out of the removable housing as shown above.

4.3 Check your Minox i
The following checks will ensure your Minox i oxygen transmitter is operating as its optimum.

  • Check for any damage to the connecting cable
  • Depending on process conditions and location, check the gas entry orifice at the face-end of the Minox i.

This is to ensure there is no build-up of particulate matter, and that moisture is not being retained at the orifice.
Typically, the Minox i will require removal from the process to carry out checks, cleaning and sen-sor replacement.

4.4 Operational spares
The only replaceable part for Minox i is the sensor. Sensor model details can be found in ” Appendix A – Technical Specifications” on page 14.
The replacement sensor must be the same as the installed sensor, for example, it is not possible to replace an OC-90 EC sensor with an OC-60 EC sensor.
NOTE: The sensors have a shelf life. Please see ” Appendix A – Technical Specifications” on page 14 for details.

Appendices

Appendix A – Technical Specifications
Sensor

| Solid-state| Electrochemical (%)| Electrochemical (ppm)| Electrochemical (ppm)
---|---|---|---|---
Model| OC-92| OC-90| OC-60| OC-69
Operating temperature| -20 °C…+50 °C
(-4 °F…+122 °F)| 5 °C…+45 °C
(+31 °F…+113 °F)| +5 °C…+45 °C
(+31 °F…+113 °F)| -10 °C…+45 °C
(+14 °F…+113 °F)
Calibration interval| Up to 1 year| Up to 6 months| Up to 3 months
Operational life (dependent on O2 and moisture concentration)| Up to 2 years| Up to 18 months| Up to 12 months
Shelf life (in original packaging)| Up to 1 years| | Up to 6 months
Calibration interval| Up to 1 year| Up to 6 months| Up to 3 months
Gas-wetted materials| Acetal, Stycast 2651 mmstainless steel, PTFE, FR4 and gold| HDPVDF, ABS, Teflon, FR4 and gold| HDPVDF, stainless steel, Teflon, FR4 and gold

Performance

Measuring range 0…25% O2 0…10, 0…100, 0…1000 ppmV O2
Output resolution 0.01 % O2 1 ppmV O2
LDL (sensitivity) 0.05 % O2 (500 ppm) 1 ppmV O2
Accuracy (intrinsic error) +/-2 % of reading @ calibrated temperature and

pressure
Linearity| +/- 2 % of reading
Response time (T90)| < 15 seconds @ 25 °C (77 °F) within selected range
Flow rate| 250…500 ml/min (0.5…1.0 SCFH)
Humidity| 0…98 %rh non-condensing
Operating pressure| Nominally atmospheric (+/-100 mBar)

Electrical input/output

Output signal 4…20 mA
Power supply 24 V DC +/- 10 %
Maximum power consumption 50 mA @ 24 V DC
Electrical interface M12 connector (1.5 metre cable supplied)

Mechanical

Housing material 303 and 316 stainless steel
Gas-wetted materials excluding sensor 316 stainless steel
O-ring material Nitrile
Ingress protection IP66 (NEMA4)
Weight 800 g (28 oz)

Appendix B – Hazardous area certification

Area Certification details Standards
Europe & Japan (Pending) IECEx / ATEX / UKCA

IECEx BAS 19.0013 / Baseefa 19ATEX0020 / BAS22UKEX0052
****  II 1 GD
Ex ia IIC T4 Ga (-20 °C≤Ta≤+55 °C)
Ex ia IIC T200 135°C Da (-20 °C≤Ta≤+55 °C)
Ui=28 V Ii=93 mA Pi=0.65 W Ci=12nF Li=705uH| EN IEC 60079-0:2018
EN 60079-11:2012
IEC 60079-0 Ed. 7 2017
IEC 60079-11 Ed. 6 2011
North America/Canada (Pending)| Class I Division 1 Group A-D T4
Class II Division 1 Groups E-G T135°C
Class I Zone 0 AEx ia IIC T4 Ga / Ex ia IIC T4 Ga
Class 1 Zone 20 AEx ia IIC T135°C Da / Ex ia IIC T135°C Da
Ta = -20°C≤Ta≤+55°C
UI/Vmax=28 V II/Imax=93 mA Pi=0.65 W Ci=12nF Li=705uH| CSA C22.2 No. 60079-0:19
CSA C22.2 No. 6007911:14
UL 60079-0:2019
UL 60079-11:2014
CSA C22.2 No. 61010-1-12
(Reference only)

Appendix C- Quality, recycling and warranty information
The PST Oxygen group of companies – AII, Ntron and SST – comply with applicable national and international standards and directives.
Full information can be found on our website at https://www.processsensing.com/en-us/compliance
The compliance site contains information on the following directives:

  • ATEX (equipment for explosive atmosphere)
  • REACH (Registration, Evaluation, Authorization, and Restriction of Chemicals)
  • RoHS (Restriction of Hazardous Substances in electrical and electronic equipment
  • ETL (The ETL mark is proof of product compliance to North American safety standards
  • WEEE (Waste Electrical and Electronic Equipment recycling
  • Recycling policy
  • Warranty and returns.

This page is intentionally blank. Minox i Oxygen Transmitter
For PST contact information, visit
ProcessSensing.com

This document is the property of Process Sensing Technologies and may not be copied or otherwise reproduced, communicated to third parties, nor stored in any Data  Processing System without the express written authorization of Process Sensing Technologies.
©2022 Process Sensing Technologies

Documents / Resources

| PROCESS SENSING PST-UM-2010-EN-02 Minox i Oxygen Transmitter [pdf] User Manual
PST-UM-2010-EN-02 Minox i Oxygen Transmitter, PST-UM-2010-EN-02, Minox i Oxygen Transmitter
---|---

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals