HUMAN CARE Roomer S Overhead Lift with Hanger Bar User Guide

June 8, 2024
HUMAN CARE

HUMAN CARE Roomer S Overhead Lift with Hanger Bar

Important! This quick reference guide does not replace the user manual which can be downloaded at www.humancaregroup.com. Read all instructions before use!

Components

  1. Secondary Panel
  2. Lift Belts
  3. Direction Levers
  4. Hanger Bar
  5. Emergency Stop
  6. Hand Control (IPX7)

Hand Control Functions

  1. Up Button
  2. Down Button
  3. USB-C and Charge Port

Secondary Panel Functions

Secondary Panel Functions

Flashing Green
On/Ready for Use Full Battery
Flashing Yellow
On/Ready for Use Charge Battery Warning
Flashing Red
On/Ready for Use Low Battery
Constant Green
Lift in Use Lift Loaded
Constant Red
Safety Catch Engaged Troubleshooting!
Constant Yellow
Lift Charging Functions Blocked
Flashing Green/Yellow
Service Warning Perform Inspection
Flashing Red/Yellow
Service Passed Battery Low

Charging the Lift

Flashing Red – Charge lift immediately! If the indicator light is not on, check the following:

  1. Connect charger plug to handset.
  2. Connect plug to charger and insert plug into wall to charge. The charger must only be connected to an earthed socket.

Note: If the emergency stop is activated, the lift will not charge!

Important! The lift can only be used with a Human Care charger and must be disconnected to use the lift.

Models: 55300H-(Country Code)
(AU-Australia, GB-Great Britan, EU-Europe, JP-Japan, NA-North America)
Human Care HC AB. All Rights Reserved.

Room to Room Transfer Feature

Room to room transfers Activate desired belt by turning directional lever. If lever faces towards the leg hook, the belt nearest to the leg hook is active. If it faces towards the hanger bar, the belt closest to the bar is active.

**** Important! Do not turn the lever unless both belts are connected to their respective rail rollers. The lever must always point to the connected belt.

Using the lift

Emergency Stop Activation
Press down on red button to activate.
Emergency De-Activation
Turn red button clockwise

Important! Emergency lowering must only be used in emergency situations. There will be a 3 second delay.

To activate, press “Down” button (4) and secondary panel emergency lowering button (5) at the same time

Suspending the lift

Place the lift on a suitable surface and feed out belt by gently pulling out belt while at the same time pressing the “Down” button on either the hand control or  secondary panel. Hook lift belt onto distance strap so thatit is held firmly in position (easiest to hook from above at an angle as shown). Ensure the lift belt is firmly in place before use.

Slings

Attach head sling loops to carry bar first, then leg straps to leg hooks. Ensure the straps are secured beneath the safety pin.

Proper lifting techniques

Do not use the telescoping bar to move the patient and the lift. Do not hold onto the hooks while performing a lift. Once the patient is comfortably secured in sling, raise lift, hold sling and position/move patient.

General Care

Charge lift every night.
Wipe down with slightly dampened drying cloth
Disinfect with DuPont TM Virkon®.
Never use strong detergents on lift.
Do not immerse in d water.
Activate emergency stop when not in use.
Activate emergency stop when not in use. Important!
Use only authorized methods to transport! Lift must always be stored or moved in horizontal position. Rest on cart to move or use approved carry strap1
1Caution! When using strap, extra caution must be taken not to wrinkle or bend the lift belt.

Inspection

Annual Inspection – The lift must be inspected at least once per year by a qualified technician certified by Human Care. Monthly
Inspection – Check lift belts for damage or wear. Feed entire length to ensure proper functioning.
Daily Inspection – Always inspect the lift for visible signs of damage before use. Ensure the lift is correctly mounted on the distance strap or rail roller. Always check slings and belts for signs of wear before use. Do not use if wear evident.

Troubleshooting

If the indicator light is not on, check the following:
– Mains switch is on
– Batteries do not require charging
– Emergency stop button is not pushed in (deactivate by turning clockwise)

If the indicator light is flashing, check the following:
– Check the hand control button functions up / down when depressed (replace if necessary)
– When pressing up, the lift is not already in its maximum upper position
– When pressing down, the lift is not already in its maximum lower position

If the indicator light is constant red:
– The lift motor is overheated – allow to cool
– The lift is overloaded
– lower using emergency lowering button  If the indicator light is constant yellow:
– Charging in progress

If the lift belt does not feed out properly:
– Pull the belt upwards at the same time as you depress the Down button on the hand control / lift panel
If the indicator changes to flashing red:
– The batteries need charging, patient may be lifted a few more times

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals