PCE-TG 75 Thickness Gauge User Manual
- June 9, 2024
- PCE
Table of Contents
PCE-TG 75 Thickness Gauge User Manual
Introduction
Safety notes
Please read this manual carefully and completely before you use the device for
the first time. The device may only be used by qualified personnel and
repaired by PCE Instruments personnel.
Damage or injuries caused by non-observance of the manual are excluded from
our liability and not covered by our warranty.
- The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
- The instrument may only be used if the environmental conditions (temperature, relative humidity, …) are within the ranges stated in the technical specifications. Do not expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
- Do not expose the device to shocks or strong vibrations.
- The case should only be opened by qualified PCE Instruments personnel.
- Never use the instrument when your hands are wet.
- You must not make any technical changes to the device.
- The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner, no abrasives or solvents.
- The device must only be used with accessories from PCE Instruments or equivalent.
- Before each use, inspect the case for visible damage. If any damage is visible, do not use the device.
- Do not use the instrument in explosive atmospheres.
- The measurement range as stated in the specifications must not be exceeded under any circumstances.
- Non-observance of the safety notes can cause damage to the device and injuries to the user.
We do not assume liability for printing errors or any other mistakes in this
manual.
We expressly point to our general guarantee terms which can be found in our
general terms of business.
If you have any questions please contact PCE Instruments. The contact details
can be found at the end of this manual.
Specifications
Technical specifications
Specification | Value |
---|---|
Model | PCE-TG 75 |
Measurement range | 1.00 … 225.0 mm |
Operating memory | 500 readings |
Probe selection | no |
Operating frequency | 5 MHz |
Accuracy | ±(0.5 % H*+0.05) mm |
Sound velocity setting range | 1000 … 9999 m/s |
Operating temperature | 0 … 40 °C / 32 … 104 °F |
Power supply | 3 x 1.5 V AA battery |
Dimensions | 163 x 82 x 38 mm / 6.42 x 3.23 x 1.50“ |
Weight | 320 g / 0.70 lbs |
Delivery contents
Model | Delivery contents |
---|---|
**** PCE-TG 75 | 1 x thickness gauge PCE-TG 75 |
1 x coupling gel TT-GEL
3 x 1.5 V AA battery
1 x probe PCE-TG 5M10d
1 x user manual
**** PCE-TG 150| 1 x thickness gauge PCE-TG 150
1 x coupling gel TT-GEL
3 x 1.5 V AA battery
1 x probe PCE-TG 5M10d
1 x probe PCE-TG 5M6d 1 x user manual
PCE-TG 150 2.5F| 1 x thickness gauge PCE-TG 150
1 x coupling gel TT-GEL
3 x 1.5 V AA battery
1 x probe PCE-TG 2.5M
1 x user manual
PCE-TG 150 HT| 1 x thickness gauge PCE-TG 150
1 x coupling gel TT-GEL
3 x 1.5 V AA battery
1 x probe PCE-TG HT
1 x user manual
Models
Model | Included probe |
---|---|
PCE-TG 75 | PCE-TG 5M10d |
PCE-TG 150 | PCE-TG 5M10d PCE-TG 5M6d |
PCE-TG 150 2.5F | PCE-TG 2.5M |
PCE-TG 150 HT | PCE-TG HT |
Probes
Probe | Specifications |
---|---|
PCE-TG 5M10d | Application |
radius >10 mm
Frequency| 5 MHz
Diameter| 10 mm / 0.29“
Temperature range| -10 … 60 °C / 14 … 140 °F
PCE-TG 5M6d| ****
Application
| Standard probe for flat materials, pipes with a radius >6 mm
Frequency| 5 MHz
Diameter| 6 mm / 0.24“
Temperature range| -10 … 60 °C / 14 … 140 °F
PCE-TG 2.5M| Application| Thickness measurement of cast materials
Frequency| 2.5 MHz
Diameter| 12 mm / 0.47“
Temperature range| -10 … 60 °C / 14 … 140 °F
PCE-TG HT| Application| Measurement of the material thickness of hot
components
Frequency| 5 MHz
Diameter| 10 mm / 0.29“
Temperature range| -10 … 350 °C / 14 … 662 °F
Optional accessories
Item number | Description |
---|---|
TT-GEL | Ultrasonic coupling gel |
K-GEL | Coupling gel for temperatures of up to 350 °C / 662 °F (100 ml) |
**** PCE-TG 5M10d | Standard probe for PCE TG 75 and PCE-TG 150 series, 5 MHz, |
∅10 mm / 0.29“, 1 … 225 mm (PCE-TG 75), 1 … 300 mm (PCE-TG 150), -10 … 60 °C /
14 … 140 °F
PCE-TG 5M6d| Miniature probe for PCE-TG 150 series, 5 MHz, ∅6 mm / 0.24“, 1 …
50 mm, -10 … 60 °C / 14 … 140 °F
PCE-TG 2.5M| 2.5 MHz probe for PCE-TG 150 series, ∅12 mm / 0.47“, 1.2 … 300
mm, -10 … 60 °C / 14 … 140 °F
PCE-TG HT| High-temperature probe for PCE-TG 150 series, ∅10 mm / 0.29“, 1.2 …
300 mm, -10 … 350 °C / 14 … 662 °F
PCE-BAG M| Universal bag for meters: inside size M
(190 x 120 x 70 mm / 7.48 x 4.72 x 2.76”) black/blue
System description
The PCE-TG 75 is a thickness gauge that measures the speed of ultrasonic waves and can thus determine the thickness of materials accurately and non- destructively. The meter can also be used to detect corrosion damage to piping and pressure vessels of production equipment. It can also be used in manufacturing, metal processing and cargo inspection. The meter is suitable for the thickness measurement of any homogeneous material that transmits ultrasonic waves at a constant rate and reflects them off the back of the material.
Applications
This meter can be used to measure the thickness of any good ultrasonic conductor with parallel top and bottom, such as metal, plastics, ceramics and glass. Materials such as aluminum, copper, gold, resin, etc. are not suitable for this measuring instrument!
Device
**No.** | **Key** | **Description** | **Functions** |
---|---|---|---|
**** 1. | **** Probe socket | ||
**** 2. | **** Up | **** Navigate up, increase value | |
**** 3. | **** Back | **** Cancel, back, exit | |
**** 4. | **** OK | **** Confirm | |
**** 5. | **** Probe head | ||
**** 6. | **** Down | **** Navigate down, decrease value | |
--- | --- | --- | --- |
**** 7. | Calibration block | ||
**** 8. | **** On/off & menu | Press and release to turn on the meter |
Press and hold to turn off the meter Press and release to open the menu
9.| | Change| Change between different materials
10.| | Display|
11.| | **** Battery compartment|
Interfaces
- Probe socket
Display (in measuring condition)
- Title bar
- Battery level
- Sound velocity
- Coupling icon
- Selected memory method
- Selected probe
- Selected file
- Reading
- Unit
- Information about the measurement
Getting started
Power supply
Three 1.5 V AA batteries are required for power supply. The battery compartment is located on the back of the instrument and is secured with two screws. Before replacing the batteries, switch off the device. Loosen the screws, lift off the cover, insert the batteries as marked and close the battery compartment again by tightening the screws.
First use
Before switching on the instrument, insert the probe into the probe socket of the instrument. To start the meter, press the on/off & menu key until the display shows the name of the meter. The interface for measuring the sound velocity will then open. The thickness of the material is determined by measuring the sound velocity. The measurement can be started here.
Menu
The menu of the meter can be reached by pressing the menu key after startup.
With the arrow keys , you can navigate through the menu. You can confirm your
selection with the OK key. To navigate one step back, press the back key .
Measure
In this menu item, make the measurement settings:
- d. Material to be measured (sound velocity)
- e. Unit of measurement m/s (unit)
- f. Resolution
Probe (PCE-TG 150 only)
In this menu item, select the probe you want to use for the measurement.
Limit
In this menu item, make the limit value settings:
- c. Set the upper and lower limit value for the measurement
- d. Enable or disable the alarm for exceeded limit values
Memory
Measurement processes can be saved and started under this menu item. If you
activate “Auto save”, all measuring processes are saved automatically.
- d. Select a directory in which the measurement is to be saved.
- e. View saved data
- f. Delete the data of the selected file
Analyze
Under this menu item, you can display the saved data in the form of a graph.
- a. You can view basic information such as the data number, the maximum value, the minimum value and the average value
- b. You can enlarge the diagrammed
- c. Use the cursor to view a certain data value
Settings
Make further settings in the meter:
- a. Language
- b. Automatic power off time
- c. Key tone
- d. Backlight brightness
- e. Colour
- f. Software version number
- g. Factory reset
Submenu
Operation
-
a. Enter the submenu:
Press the back key in the measuring interface to open the submenu. -
b. Select menu item:
-
Press the up and down keys to change between the menu items
-
c. Opening the selected menu item:
Press OK to enter the selected menu item -
d. Leave menu item:
Press the back key to leave the menu item
Functions of the submenu
-
e. Save:
Save current readings to the selected file -
f. Clear:
Delete readings and measuring information -
g. Calibration (Cal):
See calibration instructions for details -
h. Sound velocity calibration (V.Cal):
See chapter 8.2 Sound velocity calibration (V. Cal) for details.
Measurement
Apply the TT-Gel evenly to the surface to be measured. Place the probe onto the surface covered with the coupling agent.
Clean surface
Before the measurement, the surface of the object to be tested should be
cleaned of dust, dirt and rust. Coatings such as paint must also be removed.
Reducing the roughness
Rough surfaces can cause measurement errors or missing readings. Before the
measurement, the material surface should be made as smooth as possible by
grinding, polishing or using a high-viscosity coupling agent.
Rough surfaces due to heavy machines
Regular fine grooves which occur, for example, during production with lathe
machines and planers can also cause measurement errors. The corrective measure
is the same as in chapter 7.2. In addition, a better result can be obtained by
adjusting the angle using the silver marking on the center of the probe head
and the fine grooves of the material to be tested (orthogonal or parallel to
the marking).
Measurement of a round surface
To measure a round surface such as a pipe or an oil barrel, it is important to
set the angle between the silver line of the probe and the axis of the
material to be tested. In short, the mark in the sensor head must be kept
parallel or perpendicular to the axis of the material
under test. Slowly move the probe head perpendicular to the axis across the
material under test and the values on the display will change regularly. The
lowest reading displayed is the minimum thickness of the material being
measured.
The direction in which the probe is guided depends on the curvature of the
material. For large diameter pipes, the mark in the probe head should be
perpendicular to the pipe axis; for smaller diameter pipes, the mark in the
probe head can be parallel or perpendicular to the material axis. The lowest
reading is saved as the measured value.
Measurement of composite material
When measuring composite shapes (such as pipe elbows), the method described in
chapter 7.4 can be used. However, the measurement must be carried out twice to
obtain two readings. The silver line of the probe must be vertical or parallel
to the axis. The lower value is used as the measured thickness.
Non-parallel surfaces
To obtain a satisfactory result, the surface of the material to be tested must
be parallel or coaxial to the surface of the sensor, otherwise measurement
errors will occur or no reading will be obtained.
Effects of the material temperature
The material thickness and the transmission speed of ultrasonic waves are
influenced by the temperature. To achieve high measurement accuracy, measure
samples of the same material under the same temperature conditions. In case of
steel, high temperatures cause strong measurement errors (measured values are
lower than the actual data).
Damping materials
Materials with fibers, pores and coarse particles cause strong scattering and
energy attenuation of the ultrasonic waves, resulting in abnormal readings or
even missing readings (usually the abnormal reading is lower than the actual
thickness). In this case, the material is not suitable for testing with this
thickness gauge.
Reference examples
A material of known thickness or sound velocity is used to calibrate the
meter. At least one reference sample is required to calibrate the instrument.
The ultrasonic thickness gauge is equipped with a calibration block with a
thickness of 4.0 mm / 0.16”. Refer to the calibration instructions in chapter
8 of this manual for the calibration method.
Calibration (Cal)
Calibration instructions
- a. Press the back key when you are in the measuring interface, select the menu item Calibration and confirm this selection with the OK key.
- b. The display will look as shown in the image to the right.
- c. During the measurement, the sound velocity is automatically set to 5900 m/s.
- d. Press the probe onto the calibration block to calibrate the meter.
Sound velocity calibration (V. Cal)
- a. Enter the submenu and select the item Sound velocity calibration (V.Cal).
- b. The display will look as shown in the image to the right.
- c. As shown in the image, press the up and down keys to set the value.
- d. When the value has been adapted, press OK to complete the calibration.
Maintenance
Cleaning
Alcohol and thinner can damage the housing, especially the display.
Therefore, only clean the meter with a damp cloth.
Troubleshooting
Very thin material
With any ultrasonic thickness gauge, measurement errors occur when the
thickness of the material being tested is smaller than the lower limit of the
probe. When measuring ultra-thin materials, an error called “double
refraction” sometimes occurs, where the measured value is twice the actual
thickness. Another error is called “pulse envelope and loop jump” where the
measured value is higher than the actual thickness. To avoid such errors,
repeat the control measurement of critically thin material.
Rust stains and corrosion
Rust stains and corrosion pits on the surface to be measured will cause the
readings to change irregularly or not at all in extreme cases; in addition,
small rust stains are sometimes difficult to detect. If pits (blowholes) are
found or suspected, be careful when measuring this area and place the silver
line of the probe at different angles for several tests.
Incorrect material identification
If you have calibrated the instrument with one material and use it to test
another material, an incorrect result will be displayed. Make sure to select
the correct sound velocity. If the sound velocity in the current test has some
deviation from the calibration, errors may also occur. Make an adjustment when
using it.
Probe abrasion
The surface of the probe is made of acrylic resin. After prolonged use of the
meter, the roughness of the probe surface increases, resulting in a decrease
in accuracy. If the user can detect measurement errors caused by this, the
probe surface can be polished with a little bit of sandpaper or 500 grit
grindstone to smooth the surface and ensure parallelism. If the result is
still incorrect, you may need to replace the probe.
Laminated material, composite material
It is not possible to measure uncoupled laminated materials because ultrasonic
waves cannot penetrate the uncoupled space. Since ultrasonic waves cannot be
transmitted at a constant velocity in composite materials, measuring
instruments that measure thickness by ultrasonic reflection are not suitable
for measuring laminated and composite materials.
Effects of oxide layers on surfaces
Some metals can form a dense oxide layer on the surface, for example aluminum.
This oxide layer is firmly bonded to the substrate without an obvious boundary
layer. However, the transmission speed of ultrasonic waves is different in
these two substances, which causes errors. Different oxide layers cause
different errors that the user should be aware of. You can use a piece of the
same material the thickness of which you have checked with a gauge to
calibrate the meter.
Use and selection of the coupling agent
Coupling agent is used to transmit high frequency ultrasonic waves between the
probe and the material being tested. The wrong coupling gel or improper use
can lead to errors or failed measurements. It should be used in appropriate
amounts and applied evenly.
The coupling agent is usually applied to the surface of the material under
test. If the temperature is too high, the coupling agent is applied to the
bottom of the probe.
Only PCE Instruments coupling gel should be used.
Warranty
You can read our warranty terms in our General Business Terms which you can
find here:
https://www.pce-instruments.com/english/terms.
Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the
European Parliament applies. Due to the contained pollutants, batteries must
not be disposed of as household waste.
They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back.
We either re-use them or give them to a recycling company which disposes of
the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in
accordance with your local waste regulations.
If you have any questions, please contact PCE Instruments.
PCE Instruments contact information
User manuals in various languages (français, italiano, español, português, nederlands, türk, polski, русский, 中文) can be found by using our product search on: www.pce-instruments.com
United Kingdom
PCE Instruments UK Ltd
Unit 11 Southpoint Business Park
Ensign Way, Southampton
Hampshire
United Kingdom, SO31 4RF
Tel: +44 (0) 2380 98703 0
Fax: +44 (0) 2380 98703 9
info@pce-instruments.co.uk
www.pce-instruments.com/english
References
- France.fr : Explore France and its wonders - Explore France
- iberica.es
- Make an offer on the domain instruments.co.uk - Domains.co.uk
- Computer Instruments | Home
- Discover Italy: Official Tourism Website - Italia.it
- PCE Teknik Cihazlar Paz. Tic. Ltd.Şti. | PCE Instruments
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>