The LS137 UV Energy Meter is a device that can measure UV energy, UV intensity, and temperature simultaneously. It is designed for UV energy, UV intensity, and temperature detection in various equipment such as UV curing machines, UV dryers, mobile phone UV coating machines, exposure machines, printing machines, and UV LED light sources.

  • Parameters: The parameters of the LS137 UV Energy Meter can be found on page 2 of the user manual.
  • Spectral Response Curve: The spectral response curve of the LS137 UV Energy Meter can be found in the user manual.

Characteristics

  1. Ultra-small size design with a thickness of 7.6mm and a width of 44mm.
  2. High-precision and quick-response temperature detector to measure the actual temperature in the UV curing machine.
  3. Built-in timer to accurately record and display the UV curing time.
  4. Metal shell with high temperature resistance, allowing the meter to operate in an environment of up to 100°C for a long time.
  5. Data storage feature ensures that stored data is not lost when the power is off, and the last test data will be displayed automatically after power on.
  6. Provides two standard data options: Linshang standard and Chinese national standard.

Product Usage Instructions

Parameter Setting

In OFF mode, follow these steps to enter the parameter setting mode:

  1. Trigger mode: Auto/Manual
    • Short press the SELECT button to select Auto/Manual.
    • Select Auto for automatic trigger mode, where the measurement starts when the power value is greater than the selected triggervalue.
    • Select Manual to manually start and end a measurement by short pressing the POWER button.
  2. Smooth : 50HZ/60HZ
    • If the UV light source uses AC power supply, the frequency of the AC power will affect the accuracy of the power measurement.
    • Short press the SELECT button to select the power frequency of the light source (select 50HZ for DC power supply).
    • Press the POWER button to confirm the setting and enter the STOP interface.
  3. Standard Selection
    • Short press the POWER button to select the standard.
    • LS-STD (Enterprise standard): The meter is calibrated according to the enterprise standard.
    • JJG 879-2015 (Chinese national standard): The meter is calibrated according to the Chinese national standard.
    • Press the POWER button to confirm the setting and enter the STOP interface.
  4. Factory default settings
    • Trigger mode: Manual
    • Smooth: 50HZ
    • Standard Selection: LS-STD (Enterprise standard)

Product Introduction

UV energy meter can measure UV energy, UV intensity and temperature at the same time. It is suitable for UV energy, UV intensity and temperature detection of UV curing machines, UV dryers, mobile phone  UV coating machines, exposure machines, printing machi nes and other equipment. This meter is suitablefor measuring the UV intensity and UV energy of UV LED light sources.

Parameters

  1. Spectral range: 3 40 nm 4 2 0nm , calibrated at 395nm
  2. Irradiance measuring range: 0 40 000mW/cm 2
  3. Irradiance resolution: 1mW/cm 2
  4. Energy measuring range: 0 999999mJ/cm 2
  5. Energy measuring accuracy: ±10%, ±5% (typical)
  6. Sampling speed: 2048 times/second
  7. Recording period : 32 minutes
  8. Temperature measuring range : 55°C +125°C
  9. Power supply: Rechargeable lithium batteries 3.7V@400mAh
  10. Display: 128*64 Dot matrix OLED
  11. Dimension: Diameter 102mm * thickness 7.6 mm
  12. Weight: 136 g

Spectral response curve

Characteristics

  1. A dopt ultra small size design, the thickness is 7.6mm and the width is 44mm.
  2. High precision and quick response temperature detector, dynamically measuring the actual temperature in the UV curing machine.
  3. Built in timer, the meter can accurately record and display the UV curing time.
  4. Metal shell, high temperature resistant design, the meter can run in an environment of 100 for a long time.
  5. The stored data will not be lost when the power is off and the last test data will be displayed automatically after power on.
  6. The meter provide two standard data, users can use Linshang standard and Chinese national standard.

Operation

Parame ter setting
In OFF mode, long press the “POWER” button for 3s and enter the parameter setting mode:

  • A. Trigger mode:
    Auto/Manual Short press the “SELECT” button and select Auto/Manual Select Auto, the automatic trigger mode, when the power value is greater than the selected trigger value, the measurement is automatically started. Select Manual, short press the “POWER” button start a measurement, and end a measurement. Short press the “POWER” button to confirm and start another setting. Choose “Auto”, the mete r enters the trigger power value setting interface; Select Manual, the meter enters the smooth setting interface.

    • Note:  For the recording time is only 32 minutes, if the production line is very long and need a long time to reach the UV Lamp position, the “AUTO” mode must be
  • B. Trigger power: 0.1 5.0mW/cm² can be set
    Short press the “SELECT” button to set the trigger power value and you can long press the button for quick setting. Short press the “POWER” button to confirm the setting and the meter e nters the smooth interface
    C. Smooth:
    50HZ/60HZ If the UV light source uses an AC power supply, the frequency of the AC power will affect the accuracy of the power measurement. Short press the “SELECT” button to select the power frequency of the light source (s elect 50HZ for DC power supply).

    • 50HZ: 50HZ AC.
    • 60HZ: 60HZ AC.
      Short press the “POWER” button to co nfirm the setting and enter the “STOP” interface.
  • D. Standard Selection
    Short press the “POWER” button to select the standard. LS
    STD ( Enterprise standard The meter is calibrated according to Enterprise standard JJG 879 2015 (Chinese national standard): The meter is calibrated according to Chinese national standard. Short press the “POWER” button to confirm the setting, the meter enters the “STOP” interface e.
    E. Factory default settings:

    • Trigger mode: Manual
    • Smooth: 50HZ
    • Standard Selection: LS STD ( Enterprise standard)

ON/OFF

  1. In POWER OFF state, short press “POWER” button to power on the meter. After turning on the meter, the meter displays the calibration coefficient, version number, serial number , and then enters the last measurement interface (STOP interface). As shown below: In the “STOP” interface, long press the “POWER” button to power off the meter.
  2. In the “STOP” interface, the meter will automatically power off in 3 minutes without any operation.
  3. In the automatic trigger measurement “Ready” state, the longest waiting time is 50 minutes. If the measurement cannot be triggered with in 50 minutes, it will automatically turn off.

Measurement

In the measurement mode, three operation states exist:

  • READY: Ready state; in the auto trigger mode, this means the meter is waiting for triggering to begin a measurement.
  • RUN: Measuring state; th is means the meter is measuring data.
  • STOP: Measurement is over.
  • In measurement interface:
  • M:Manual measurement mode
  • A:Auto measurement mode
    • Manual measurement mode: in the STOP interface. Short press the “POWER” button and then short press the “SELECT” button to confirm the new measurement, clear the historical data (display ” “———-“), automatically enter the measurement interface (RUN state) a fter 1s. The interface dynamically displays energy E, real time power value P, real time temperature value T, measurement time. Short press the “POWER” button or after 32 minutes, the measurement is over and enter the STOP interface. The interface displays the measurement result energy E, maximum power value P MAX , maximum temperature value T MAX and measurement time.Linshang-LS137-UV-Energy-Meter-fig-1 \(4\)
    • Automatic mode: After pressing the button to confirm the new measurement, the meter enter the READY state to clear the historical data and wa it for the trigger condition (Trigger power) to be met, the energy display ” “——–” and flashing. When the power value is greater than the set trigger power, the trigger condition is met, enter the measurement interface (RUN state). The interface dynamically displays energy E, real time power value P, real time temperature value T and measurement time. When the recording time reaches 32 minutes or the power value is less than the set trigger power, the measurement is automatically ended and enters the STOP in terface. The interface displays the measurement result energy E, maximum power value P MAX , maximum temperature value T MAX and measurement time. In the stop status, long press the “SELECT” key to clear the current test data.

Charging

This meter has a buil t in lithium rechargeable battery. Please charge it in time when the battery symbol on the screen is empty during use. After plugging in the charger, the energy bar in the battery symbol increases, indicating that it is charging. Charging can be done in th e following ways:

  1. Use an USB data cable, plug it directly into the USB port of the computer.
  2. Use the matching 5V switching power supply.
  3. The mobile power supply for smart phones.

Measurement and Notes

  1. The sensor must face the UV light source.
  2. When not in use, please long press the “POWER” button turn off the meter.
  3. Avoid contact with corrosive materials and keep away from high humidity.
  4. Please put it in the specialized package after power off and keep properly.
  5. The suggested calibrating period is one yea r and our company has the standard light source and provides calibration service. (The previous calibration time “Calibration: year/month/day” will be displayed on the boot screen.)
  6. For the UV sensor is very sensitive to humidity, the storage environment i s very important. For a long time storage, please be sure to keep the meter in dry environment.

Heat Shield

Equipped with heat shield to avoid damage to the screen caused by high power UV lamps irradiating the instrument at close range.

No Heat Shield Linshang-LS137-UV-Energy-Meter-fig-1
\(6\)

Using Heat Shield Linshang-LS137-UV-Energy-Meter-fig-1
\(7\)

Packing details

No. Description Quantity Unit
1 UV Energy Meter 1 pcs
2 USB data cable 1 pcs
3 DC 5V Power adapter 4 pcs
4 User Manual 1 pcs
5 Certificate/Warranty card 1 pcs
6 Plastic Case 1 pcs
7 Heat Shield 1 pcs

Service

  1. The meter has one year warranty. If the meter works abnormally, please send the whole meter to the company for maintenance.
  2. Provide users with spare parts and lifelong maintenance services.
  3. Provide the users with the meter i inspection service for free.
  4. Free technical support for the long term.