HOVERTECH HOVERMATT Air Transfer System User Manual
- June 3, 2024
- HOVERTECH
Table of Contents
- HOVERTECH HOVERMATT Air Transfer System
- Intended Use and Precautions
- HT-Air® Keypad Functions
- ATTACHING TO BEDFRAME
- LITHOTOMY POSITION
- Product Specifications/Required Accessories
- Using the HoverMatt® Air Transfer System in the Operating Room
- Cleaning and Preventive Maintenance
- AIR SUPPLY CLEANING AND MAINTENANCE
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
HOVERTECH HOVERMATT Air Transfer System
Symbol Reference
DECLARATION OF CONFORMITY TO MEDICAL DEVICE DIRECTIVE
Intended Use and Precautions
INTENDED USE
The HoverMatt® Air Transfer System is used to assist caregivers with patient transfers, positioning (including boosting and turning), and proning. The Hover Tech Air Supply inflates the Hover Matt to cushion and cradle the patient, while air simultaneously escapes from the holes on the underside, reducing the force needed to move the patient by 80-90%.
INDICATIONS
- Patients unable to assist in their own lateral transfer
- Patients whose weight or girth poses a potential health risk for the caregivers responsible for repositioning or laterally transferring said patients
CONTRAINDICATIONS
- Patients who are experiencing thoracic, cervical or lumbar fractures that are deemed unstable should not use the HoverMatt unless a clinical decision has been made by your facility.
INTENDED CARE SETTINGS
- Hospitals, long term or extended care facilities
PRECAUTIONS – HOVERMATT
- Caregivers must verify that all caster brakes have been engaged prior to transfer.
- Use a minimum of two caregivers during air-assisted lateral patient transfers.
- For in-bed air-assisted positioning tasks, more than one caregiver may need to be used. Side rails must be raised with one caregiver.
- For air-assisted pruning, see HoverTech’s training video on www.HoverMatt.com.
- Never leave patient unattended on an inflated device.
- Use this product only for its intended purpose as described in this manual.
- Only use attachments and/or accessories that are authorized by Hover Tech International.
- When transferring to a low air loss bed, set the bed mattress air flow to the highest level for a firm transfer surface.
- Never attempt to move a patient on an uninflated Hover Matt.
- WARNING: In the OR – To prevent the patient from slipping, always deflate the Hover Matt and secure the patient and Hover Matt to the OR table prior to moving the table into an angled position.
PRECAUTIONS – AIR SUPPLY
- Not for use in the presence of flammable anesthetics or in a hyperbaric chamber or oxygen tent.
- Route the power cord in a manner to ensure freedom from hazard.
- Avoid blocking the air intakes of the air supply.
- When using the Hover Matt in the MRI environment, a 25 ft. specialty MRI hose is required (available for purchase).
- CAUTION: Avoid electric shock. Do not open air supply.
- WARNING: Reference product specific user manuals for operating instructions.
Part Identification – HoverMatt® Air Transfer Mattress
Part Identification – HT-Air® Air Supply
HT-Air® Keypad Functions
ADJUSTABLE: For use with HoverTech air-assisted positioning devices. There are
four different settings. Each press of the button increases the air pressure
and rate of inflation. The Green Flashing LED will indicate the inflation speed
by the number of flashes (i.e. two flashes equals the second inflation speed).
All of the settings in the ADJUSTABLE range are substantially lower than the
HoverMatt and HoverJack settings. The ADJUSTABLE function is not to be used
for transferring.
The ADJUSTABLE setting is a safety feature that can be used to ensure the
patient is centered on HoverTech air-assisted devices and to gradually
accustom a patient who is timid or in pain to both the sound and functionality
of the inflated devices.
STANDBY: Used to stop inflation/air flow (Amber LED indicates STANDBY
mode).
HOVERMATT 28/34: For use with 28″ & 34″ HoverMatts and HoverSlings.
HOVERMATT 39/50 & HOVERJACK: For use with 39″ & 50″ HoverMatts and
HoverSlings and 32″ & 39″ HoverJacks.
Instructions for Use – HoverMatt® Air Transfer System
- Patient should preferably be in a supine position.
- Place HoverMatt underneath patient using a log-rolling technique and secure patient safety straps loosely.
- Plug the HoverTech Air Supply power cord into an electrical outlet.
- Insert the hose nozzle into either of two hose entries at foot end of HoverMatt and snap into place.
- Ensure that transfer surfaces are as close as possible and lock all wheels.
- If possible, transfer from a higher surface to a lower surface.
- Turn on HoverTech Air Supply.
- Push HoverMatt at an angle, either headfirst or feetfirst. Once half-way across, opposite caregiver should grasp closest handles and pull to desired location.
- Ensure patient is centered on receiving equipment prior to deflation.
- Turn off air supply and employ the bed/stretcher rails. Unbuckle patient safety straps.
NOTE: When using the 50” HoverMatt, two air supplies may be used for inflation.
ATTACHING TO BEDFRAME
- Remove connecting straps from pockets and loosely attach to solid points on the bed frame to allow the SPU Link to move with the patient.
- Before lateral transfers and positioning, disconnect connecting straps from bed frame and stow in corresponding storage pockets.
LATERAL TRANSFER
- Patient should preferably be in a supine position.
- Place HoverMatt underneath patient using a log-rolling technique and secure patient safety straps loosely.
- Plug the HoverTech Air Supply power cord into an electrical outlet.
- Insert the hose nozzle into either of two hose entries at foot end of HoverMatt and snap into place.
- Ensure that transfer surfaces are as close as possible and lock all wheels.
- If possible, transfer from a higher surface to a lower surface.
- Turn on HoverTech Air Supply.
- Push HoverMatt at an angle, either headfirst or feetfirst. Once half-way across, opposite caregiver should grasp closest handles and pull to desired location.
- Ensure patient is centered on receiving equipment prior to deflation.
- Turn off air supply and employ the bed/stretcher rails. Unfasten patient safety straps.
- Remove connecting straps from pockets and loosely attach to solid points on bed frame.
LITHOTOMY POSITION
- Separate the legs into two individual sections by disconnecting the snaps.
- Position each section on the table with the patient’s legs.
LATERAL TRANSFER
- Ensure all snaps located at the center leg and foot sections are connected.
- Patient should preferably be in supine position.
- Place the HoverMatt underneath the patient using a log-rolling technique and secure patient safety strap loosely.
- Plug HoverTech Air Supply power cord into an electrical outlet.
- Insert hose nozzle into either of the two hose entries located at the headend of the Reusable Split-Leg Matt, or at the footed of the Single-Patient Use Split-Leg Matt, and snap into place.
- Ensure that transfer surfaces are as close as possible and lock all wheels.
- If possible, transfer from a higher surface to a lower surface.
- Turn on HoverTech Air Supply.
- Push HoverMatt at an angle, either headfirst or feetfirst. Once half-way across, opposite caregiver should grasp closest handles and pull to desired location.
- Ensure patient is centered on receiving equipment prior to deflation.
- Turn off HoverTech Air Supply and employ bed/stretcher rails. Unbuckle patient safety strap.
- When the Split-Leg Matt is deflated, position each leg section as appropriate.
Instructions for Use – HoverMatt® Half-Matt
- Patient should preferably be in a supine position.
- Place HoverMatt underneath patient using a log-rolling technique and secure patient safety strap loosely.
- Plug the HoverTech Air Supply power cord into an electrical outlet.
- Insert hose nozzle into either of two hose entries at foot end of Hover-Matt and snap into place.
- Ensure that transfer surfaces are as close as possible and lock all wheels.
- If possible, transfer from a higher surface to a lower surface.
- Turn on HoverTech Air Supply.
- Push HoverMatt at an angle, either headfirst or feetfirst. Once half-way across, opposite caregiver should grasp closest handles and pull to desired location. Ensure caregiver at footed guides patient’s feet during the transfer.
- Ensure patient is centered on receiving equipment prior to deflation.
- Turn off HoverTech Air Supply and employ the bed/stretcher rails. Unbuckle patient safety strap.
PRECAUTION: USE A MINIMUM OF THREE CAREGIVERS DURING AIR-ASSISTED LATERAL PATIENT TRANSFERS WHEN USING THE HOVERMATT HALF-MATT.
Product Specifications/Required Accessories
HOVERMATT® AIR TRANSFER MATTRESS (REUSABLE)
Material:| Heat-Sealed: Nylon twill
Double-Coated: Nylon twill with silica polyurethane coating on patient side
---|---
Construction:| RF-Welded
Width:| 34″ (86 cm), 39″ (99 cm), 50″ (127 cm)
Length:| 78″ (198 cm) Half-Matt: 45″ (114 cm)
Heat-Sealed Construction
- Model #: HM28HS – 28” W x 78” L
- Model #: HM34HS – 34″ W x 78″ L
- Model #: HM39HS – 39″ W x 78″ L
- Model #: HM50HS – 50″ W x 78″ L
Double-Coated Construction
- Model #: HM28DC – 28” W x 78” L
- Model #: HM34DC – 34″ W x 78″ L
- Model #: HM39DC – 39″ W x 78″ L
- Model #: HM50DC – 50″ W x 78″ L
- WEIGHT LIMIT 1200 LBS/ 544KG
HoverMatt Half-Matt
- Model #: HM-Mini34HS – 34″ W x 45″ L
- Double-Coated Construction
- Model #: HM-Mini34DC – 34″ W x 45″ L
- WEIGHT LIMIT 600 LBS/ 272 KG
REQUIRED ACCESSORY:
- Model #: HTAIR1200 (North American Version) – 120V~, 60Hz, 10A
- Model #: HTAIR2300 (European Version) – 230V~, 50 Hz, 6A
- Model #: HTAIR1000 (Japanese Version) – 100V~, 50/60 Hz, 12.5A
- Model #: HTAIR2356 (Korean Version) – 230V~, 50/60 Hz, 6A
- Model #: AIR200G (800 W) – 120V~, 60Hz, 10A
- Model #: AIR400G (1100 W) – 120V~, 60Hz, 10A
LATEX FREE
Material: | Top: Non-woven polypropylene fiber Bottom: Nylon twill |
---|---|
Construction: | Sewn |
Width: | 34″ (86 cm), 39″ (99 cm), 50″ (127 cm) |
Length: | Varies by product |
Half-Matt: 45″ (114 cm)
HoverMatt Single-Patient Use
- Model #: HM34SPU-B – 34″ W x 78″ L (10 per box)*
- Model #: HM39SPU-B – 39″ W x 78″ L (10 per box)*
- Model #: HM50SPU-B – 50″ W x 78″ L (5 per box)*
- Model #: HM50SPU-B-1Matt – 50″ W x 78″ L (1 Unit)*
HoverMatt SPU Split-Leg Matt
- Model #: HM34SPU-SPLIT-B – 34″ W x 70″ L (10 per box)*
HoverMatt SPU Link
- Model #: HM34SPU-LNK-B – 34″ W x 78″ L (10 per box)*
- Model #: HM39SPU-LNK-B – 39″ W x 78″ L (10 per box)*
- Model #: HM50SPU-LNK-B – 50″ W x 78″ L (5 per box)*
- Model #: HM50SPU-LNK-B-1Matt – 50” W x 78” L (1 Unit)*
- WEIGHT LIMIT 1200 LBS/ 544 KG
HoverMatt SPU Half-Matt
- Model #: HM34SPU-HLF-B – 34″ W x 45″ L (10 per box)*
- Model #: HM39SPU-HLF-B – 39″ W x 45″ L (10 per box)*
- WEIGHT LIMIT 600 LBS/ 272 KG
- *Breathable model
- REQUIRED ACCESSORY:
- Model #: HTAIR1200 (North American Version) – 120V~, 60Hz, 10A
- Model #: HTAIR2300 (European Version) – 230V~, 50 Hz, 6A
- Model #: HTAIR1000 (Japanese Version) – 100V~, 50/60 Hz, 12.5A
- Model #: HTAIR2356 (Korean Version) – 230V~, 50/60 Hz, 6A
- Model #: AIR200G (800 W) – 120V~, 60Hz, 10A
- Model #: AIR400G (1100 W) – 120V~, 60Hz, 10A
LATEX FREE
Using the HoverMatt® Air Transfer System in the Operating Room
OPTION 1
Place the HoverMatt on the Pre-Op stretcher or bed prior to patient arrival.
Have patient ambulate onto bed/stretcher or use a HoverMatt to perform a
lateral transfer. Once in the OR, ensure the OR table is secured and locked to
the floor, then transfer the patient onto the OR table. Have a caregiver at
headend of OR table ensure the patient is centered before deflating the
HoverMatt. Position the patient as needed for surgery. Tuck the edges of the
HoverMatt under the OR table pad, and ensure table rails are accessible. For
supine surgeries, follow your facility’s patient positioning protocol. After
the case, release the edges of the HoverMatt from under the OR table. Buckle
patient safety straps loosely. Partially inflate the HoverMatt using the
ADJUST-ABLE setting, have headend caregiver ensure the patient is centered,
then fully inflate using the appropriate high speed setting. Transfer the
patient to the stretcher or bed.
OPTION 2
Prior to patient arrival, place the HoverMatt on the OR table and tuck the
edges under the OR table pad. Ensure table rails are accessible. Transfer the
patient onto the table, and proceed as described in Option 1.
TRENDELENBURG POSITION
If Trendelenburg or Reverse Trendelenburg is required, an appropriate anti-
slide device that secures to the frame of the OR table must be used. For
Reverse Trendelenburg, a device that clamps to the OR table frame, such as a
footplate, should be used. If the surgery also includes a tilt side-to-side
(sailplaning), the patient must be safely secured to accommodate this position
prior to starting surgery.
Cleaning and Preventive Maintenance
In between patient uses, the HoverMatt should be wiped down with a cleaning
solution used by your hospital for medical equipment disinfec-tion. A 10:1
bleach solution (10 parts water: one part bleach) or disinfec-tant wipes can
also be used. It is important to follow the cleaning solution manufacturer’s
instructions for use, including dwell time and saturation.
NOTE: Cleaning with bleach solution may discolor fabric.
If a reusable HoverMatt becomes badly soiled, it should be laundered in a
washing machine with a 160° F (71° C) maximum water temperature. A 10:1 bleach
solution may be used (10 parts water: one part bleach) during the wash cycle.
The HoverMatt should be air dried if possible. Air drying can be expedited by
using the air supply to circulate air through the inside of the HoverMatt. If
using a dryer, the temperature setting should be set on the coolest setting.
Drying temperature should never exceed 115° F (46° C). The backing of the
nylon is polyurethane and will begin to deteriorate after repeated high
temperature drying. The Double-Coated HoverMatt should not be put in the
dryer.
To help keep the HoverMatt clean, HoverTech International recommends the use
of the HoverCover™ Disposable Absorbent Cover or their Disposable Sheets.
Whatever the patient is lying on to keep the hospital bed clean may also be
placed on top of the HoverMatt.
The Single-Patient Use HoverMatt is not intended to be laundered.
AIR SUPPLY CLEANING AND MAINTENANCE
See air supply manual for reference.
NOTE: CHECK YOUR LOCAL/STATE/FEDERAL/INTERNATIONAL GUIDELINES BEFORE
DISPOSAL.
PREVENTIVE MAINTENANCE
Prior to use, a visual inspection should be performed on the HoverMatt to
ensure that there is no visible damage that would render the HoverMatt
unusable. The HoverMatt should have all of its patient safety straps and
handles (reference the manual for all appropriate parts). There should be no
tears or holes that would prevent the HoverMatt from inflating. If any damage
is found that would cause the system not to function as intended, the
HoverMatt should be removed from use and returned to HoverTech International
for repair (Single-Patient Use HoverMatts should be discarded).
INFECTION CONTROL
HoverTech International offers superior infection control with our heat-sealed
reusable HoverMatt.
This unique construction eliminates the needle holes of a sewn mattress which
can be potential bacterial entry ways. Additionally, the heat-sealed, Double-
Coated HoverMatt offers a stain and fluid proof surface for easy cleaning.
The Single-Patient Use HoverMatt is also available to eliminate the
possibility of cross-contamination and the need for laundering.
If the HoverMatt is used for an isolation patient, the hospital should employ
the same protocols/procedures it utilizes for the bed mattress and/or for
linens in that patient room.
Returns and Repairs
All products being returned to HoverTech International (HTI) must have
a Returned Goods Authorization (RGA) number issued by the company. Please call
800-471-2776 and ask for a member of
the RGA Team who will issue you an RGA number. Any product returned without an
RGA number will cause a delay in the repair time.
Returned products should be sent to:
HoverTech International
Attn: RGA # ___
4482 Innovation Way
Allentown, PA 18109
4482 Innovation Way Allentown, PA 18109
800.471.2776
Fax 610.694.9601
www.HoverMatt.com
Info@HoverMatt.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>