KLINGER US201 Ultrasonic Heatmeter Instruction Manual

June 3, 2024
KLINGER

US201 Ultrasonic Heatmeter

Ultrasonic Heatmeter Operation Manual
Ultrasonic Heatmeter Type US201 Operation Manual
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

Product Overview

1.1. Introduction

Klinger US201 is a wall-mount, clamp- on type ultrasonic flow meter which uses the transfer time technology. Clamp on type ultrasonic flow meter is easy to install and no need to cut off the pipe that saves you lots of troubles and cost. At the same time US201 has our unique calculate software to ensure the high accuracy and low velocity response.
US201 ultrasonic flow meter widely application in oil industry, water treatment, pure water, chemical and etc.
US201 could add the RTD model and temperature sensor become an energy meter to monitoring the energy use, help to save the energy.
1.2. Features of US201

Comparing With other traditional flow meter or ultrasonic flow meter, it has distinctive features such as high precision, high reliability, high capability and low cost, the Flow meter features other advantages:
1. With ARM COMA chip, low power consumption, high reliability, anti-jamming and outstanding benefits.
2. User-friendly menu designed. Parameters of pipe range, pipe material, pipe wall thickness, output signals, etc can be conveniently entered via the windows. British and Metric measurement units are available.
3. Daily, monthly and yearly totalized flow: Totalized flow for the last 64 days and months as well as for the last 6 years are may be viewed. With the SD Card, 512 files can be stored; the time interval can be within 1 second.
4. Parallel operation of positive, negative and net flow totalizer with scale factor and 7 digit display. Internally configured batch controller makes batch control convenient.
The flow meter ensures the higher resolution and wider measuring range by the 0.04ns high resolution, high linearity and high stability time measuring circuit and 32 bits digits processing program.
1.3. Theory of Operation

When the ultrasonic signal is transmitted through the flowing liquid, there will be a difference between the upstream and downstream transit time ( travel time or time of flight ), which is proportional to flow velocity,

V = MD ´ DT sin 2q Tup · Tdown
Remarks:
V Medium Velocity M Ultrasonic frequency of reflection D Pipe Diameter The angle between the ultrasonic signal and the flow Tup Transit time in the forward direction Tdown Transit time in the reverse direction T = Tup ­ Tdown

Downstream Transducer

Tdown

Flow

Tup

D

Upstream Transducer

4
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

1.4. Specifications

Performance specifications

Flow Rage

0 ft/s ~ ± 40 ft/s ( 0 m/s ~ ± 12 m/s ).

Accuracy

±1% of measured value.

Repeatability

0.2%.

Pipe Size

1 ~ 48 ( 25 mm ~ 1200 mm ).

Function Specifications

Output

Analog output: 0/4 ~ 20 mA, ( max load 750 ) Pulse output: 0 ~ 9999 Hz, OCT ( min. and max. frequency is adjustable ); Relay output: max. Frequency 1Hz (1A 125VAC or 2A 30VDC ).

Communication Interface

RS232 & RS485.

SD Card ( Option )

Max record: 512 days. Record time interval: 1 ~ 3600 s.

Power Supply

10 ~ 36 VDC.

Keypad

16 light tactile keys.

Display

240*128 lattice, backlit LCD.

Temperature

Transmitter: -40°F ~ 140 °F (- 40 ~ 60 ). Transducer: -40 °F ~ 176 °F ( – 40 ~ 80 , standard ).

Humidity

Up to 0 ~ 99% RH, non – condensing.

Physical specifications

Transmitter

PC/ABS, IP65.

Transducer

Encapsulated design. Standard / Maximum cable length: 30 ft / 1000 ft ( 9m / 305 m ).

Weight

162328cm, 3.2kg

5
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

2. Connection

2.1. Wire Connecting
2.1.1. Power supply option
Customers should pay special attention to specify the desired power supply when wiring. Factory standard power supply is 10~36VDC/1A max. To ensure the transmitter can work normally, please pay attention to the followings when wiring: Ensure that power connections are made in accordance with the specifications shown on the transmitter.
2.1.2. Transmitter Wiring
Once the electronics enclosure has been installed, the flow meter wiring can be connected. Open the case, you will find the Power board wiring ports, from left to right, are as follows; Connect to power supplier, Relay Output, OCT Output, Transducer wiring, 4-20mA Output, RS232 Output, RS485 Output, Analog Input. For double-shielded transducer cable: “-” on the Blue wire, “+” on the Brown wire and “shield” on the Black shield wire. Refer to the below form for specific connection:

Sign DC+ DC-
RL OUT+ RL OUTOCT OUT+ OCT OUTGND UP+ UPGND DN+ DNI OUT+ I OUT-

Description DC Power DC1036V + DC Power DC1036V Grounding
Relay Output
OCT Output
Upstream sensor Grounding Black Upstream sensor + Brown Upstream sensor – Blue Downstream sensor Grounding Black Downstream sensor + Brown Downstream sensor – Blue
40~20mA Output

6
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
AI1 AI2 GND TX HX GND A B IN1+ IN1GND IN2+ IN2GND

Analog Signal Input ( Only Energy Meter )
RS232 Output
RS485 Output Temperature sensor water inline + Temperature sensor water inline Temperature sensor water inline grounding Temperature sensor water outline + Temperature sensor water outline Temperature sensor water outline grounding

Warning Wire when it is power-off. Reliable grounding must be taken for the instrument before installation and use. Use either AC or DC power supply. Do not connect them both at the same time.
2.2. Powering on
As soon as the flow meter is switched on, the system will run automatically according to the last input parameters. If the installation is accomplished when system is switched on, gain adjustment can be monitored in Window M04. After code “R” are displayed on the upper right corner of the screen, the system will activate the normal measurement condition automatically. It is indicated by code “R” on the upper left corner of the screen. If it is the first time to use or install on a new site, the customer need to input the new installation site parameters. Any parameters which are set by user will be saved permanently until they are changed by the user. When the user modifies the parameters and removes the transducers, the meter will recalculate automatically, and operate normally with the parameters. The flow meter can always complete all tasks at the same time. The tasks (Including measurement, output, etc) will be carried out as usual, no matter in which display window.

7
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
2.3. Keypad Functions
Numbers “0~9” and “.” Input Numbers or Menu Code ” “Backspace or delete characters to the left or back to the previous menu. “”Return to the last menu or open the next menu. Acts as “+” and “-” are used to enter numbers. ” Menu”Select a menu. Press this key first, then input two menu numbers to display the selected menu.
SD card data memory is an optional.
2.4. Keypad Operation
The flow meter adopts the window software design to consolidate or subdivide all of the parameters entered, the instrument setup and measurement result displays into independent windows. The operator can input parameters, modify settings or display measurement results by “visiting” a specific menu window. Each window serial number, or so-called window ID code, has a defined meaning. For example, Window M10 indicates the parameter input for pipe outside diameter, while Window M14 indicates the mounting spacing between the transducers, etc. (Refer ­ Windows Display Explanations). The keypad shortcut to visit a specific window is to press the ” Menu” key at any time, then input the 2-digit window ID code. For example, to input or check the pipe outside diameter, just press the ” Menu” “1” “0” keys for window ID code 10. Use “”and “” to switch. Another method to visit a particular window is to press “”and “” to scroll the screen. You can check the corresponding parameters by visiting the Data Type Windows. If you want to modify the parameters, press ” Enter ” first, input the digits then press ” Enter ” again to confirm.
Attention Generally, press Enter key first if operator wants to enter “modify” status. If the “modify” is still not possible even after pressing the Enter key, it means that system is locked by a password. To “Unlock” it, select “Unlock” in Window M54 and enter the original password.
8
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

3.Quick start

3.1. Basic settings
For example, let us you have a pipe of 200mm outer diameter4mm pipe thickness, measuring medium is water, Pipe Material is PVC with no Liner, These parameters should be operated as follows:

Step1. Pipe Size Settings Find M10; enter the pipe size, the outer diameter of the pipe and the pipe thickness. Press the ” Enter” to confirm.

OD200.0mm

Step2. Pipe Material Use “” to switch to select the material of the pipe.
Press the ” Enter” to confirm.

SizeM.

Step 3. Water Temperature Find M12, select the temperature of the water, temperature should be 0-80 deg. C.
Press the ” Enter” to confirm. Note: Room temperature is 25 deg.C

M12Medium*R

WTMP

20(C)

9
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
Step4. Transducer Type Find M13, select the transducer type, here we select the 1. Clamp-On-D, our standard clamp on type transducer. Press the ” Enter” to confirm. .
Step 5. Transducer Mounting Methods Use “” to switch to select transducers mounting method, here we select 0. V type, directly method. Press the ” Enter” to confirm.
Step 6. Installation Spacing Find M14, accurately install the transducer according to the displayed transducer mounting spacing and the selected mounting method.
Step 7. Display Measurement Results Menu 01 will display flow rate. (Subject to the real measurement. )
3.2. Measurement Site Selection
The installation of this ultrasonic flow meter is the simplest one of all kinds of flow meters. Only one suitable measuring site needed, plug the transducers on the pipe and then start the measurement. When selecting a measurement site, it is important to select an area where the fluid flow profile is fully developed to guarantee a highly accurate measurement. Use the following guidelines to select a proper installation site: l Choose a section of pipe that is always full of liquid, such as a vertical pipe with flow in the upward direction
or a full horizontal pipe.
10
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
l Ensure enough straight pipe length at least equal to the figure shown below for the upstream and downstream transducers installation. Try to avoid Ensure enough straight pipe length at least equal to the figure shown below for the upstream and downstream transducers installation.
l On the horizontal pipe, the transducer should be mounted on the 9 and 3 of the pipe, avoiding the position of 6 and 12, in case of the signal attenuation caused by pipe at the bottom sediment or bubblecavitation on the pipe.
l Ensure that the measuring site temperature is under the transducer temperature limits. l Consider the inside condition of the pipe carefully. If possible, select a section of pipe where the inside is free
of excessive corrosion or scaling. l Choose a section of sound conducting pipe.
11
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
12
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

4.Transducer Installation

4.1. Transducer Installation
Please make sure the pipe surfaces where the transducers are to be mounted are all clean. Including the rust, scale or loose paint to have a smooth surface. Choose the section and don’t forget apply the coupling compound. Apply the coupling compound down the center of the face of each transducer as well as on the pipe surface, ensure there are no air bubbles between the transducers and the pipe wall, and then attach the transducers to the pipe with the straps provided and tighten them securely.
Note: The two transducers should be mounted at the pipe’s centerline on horizontal pipes.
Make sure that the transducer mounting direction is parallel with the flow.
During the installation, there should be no air bubbles or particles between the transducer and the pipe wall. On horizontal pipes, the transducers should be mounted in the 3 o’clock and 9 o’clock positions of the pipe section in order to avoid any air bubbles inside the top portion of the pipe. (Refer to Transducer Mounting). If the transducers cannot be mounted horizontally symmetrically due to limitation of the local installation conditions, it may be necessary to mount the transducers at a location where there is a guaranteed full pipe condition (the pipe is always full of liquid).
4.1.1. Transducer Spacing

The spacing between the ENDS of the two transducers is considered as the standard transducer spacing (Refer to MENU14). After entering the required parameters, Check the data displayed in Window M14 and adjusted the transducers spacing according to the data displayed in Windows M14.
1.1.2 Transducer Mounting Methods

There are two mounting method, you could use depend on the measuring environment. V method and Z method (Reflect method and Direct method). V method is easy to installation and fit for mostly ultrasonic environment but Z method has stronger signal and works better in the complicated measuring environment.
4.1.3. V Method

The V method is considered as the standard method. It is convenient to use, but still requires proper installation of the transducers, contact on the pipe at the pipe’s centerline and equal spacing on either side of the centerline.

Side view

Section

Top View
13
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

4.1.4. Z Method

The signal transmitted in a Z method installation has less attenuation than a signal transmitted with the V method
When the pipes are too large, there are some suspended solid in the fluid, or the scaling and liner are too thick. This is because the Z method utilizes a directly transmitted (rather than reflected) signaling which transverses the liquid only once. The Z method is able to measure on pipe diameters ranging from 100mm to 5000mm (4 inch to 200 inch) approximately. Therefore, we recommend the Z method for pipe diameters over 300mm (12 inch).

Side view

Section

Top View

4.2. Transducer Mounting Inspection
Check to see if the transducer is installed properly and if there is an accurate and strong enough ultrasonic signal
14
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
to ensure proper operation and high reliability of the transducer. It can be confirmed by checking the detected signal strength, total transit time, delta time as well as transit time ratio. The “mounting” condition directly influences the flow value accuracy and system long-time running reliability. In most instances, only apply a wide band of sonic coupling compound lengthwise on the face of the transducer and stick it to the outside pipe wall to get good measurement results. However, the following inspections still need to be carried out in order to ensure the high reliability of the measurement and long-term operation of the instrument.
4.2.1. Signal Strength
Signal strength (displayed in Window M04) indicates a detected strength of the signal both from upstream and downstream directions. The relevant signal strength is indicated by numbers from 00.0 ~ 99.9. 00.0 represents no signal detected while 99.9 represent maximum signal strength. Normally, the stronger the signal strength detected, the longer the operation of the instrument reliably, as well as the more stable the measurement value obtained. Adjust the transducer to the best position and check to ensure that enough sonic coupling compounds is applied adequately during installation in order to obtain the maximum signal strength. System normally requires signal strength over 75.0, which is detected from both upstream and downstream directions. If the signal strength detected is too low, the transducer installation position and the transducer mounting spacing should be re-adjusted and the pipe should be re-inspected. If necessary, change the mounting method to be Z method.
4.2.2. Signal Quality (Q value)
Q value is short for Signal Quality (displayed in Window M04). It indicates the level of the signal detected. Q value is indicated by numbers from 00 ~ 99. 00 represents the minimum signal detected while 99 represent the maximum. Normally, the transducer position should be adjusted repeatedly and coupling compound application should be checked frequently until the signal quality detected is as strong as possible.
4.2.3. Total Time and Delta Time
“Total Time and Delta Time”, which displays in Window M04, indicates the condition of the installation. The measurement calculations in the Flow meter are based upon these two parameters. Therefore, when “Delta Time” fluctuates widely, the flow and velocities fluctuate accordingly, this means that the signal quality detected is too poor. It may be the resulted of poor pipe- installation conditions, inadequate transducer installation or incorrect parameter input.
Generally, “Delta Time” fluctuation should be less than ±20%. Only when the pipe diameter is too small or
velocity is too low can the fluctuation be wider.
4.2.4. Transit Time Ratio
Transit Time Ratio indicates if the transducer mounting spacing is accurate. The normal transit time ratio should be 100+/-3 if the installation is proper. Check it in Window M04.
Attention
If the transit time ratio is over 100± 3%, it is necessary to check:
(1) If the parameters (pipe outside diameter, wall thickness, pipe material, liner, etc.) have been entered correctly,
(2) If the transducer mounting spacing is accordance with the display in Window M14, (3) If the transducer is mounted at the pipe’s centerline on the same diameter, (4) If the scale is too thick or the pipe mounting is distorted in shape, etc.
15
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
4.2.5. Warnings
(1) Pipe parameters entered must be accurate; otherwise the Flow meter will not work properly. (2) During the installation, apply enough coupling compounds in order to stick the transducers onto the
pipe wall. While checking the signal strength and Q value, move the transducers slowly around the mounting site until the strongest signal and maximum Q value can be obtained. Make sure that the larger the pipe diameter, the more the transducers should be moved. (3) Check to be sure the mounting spacing is accordance with the display in Window M14 and the transducer is mounted at the pipe’s centerline on the same diameter. (4) Pay special attention to those pipes that formed by steel rolls (pipe with seams), since such pipe is always irregular. If the signal strength is always displayed as 0.00, that means there is no signal detected. Thus, it is necessary to check that the parameters (including all the pipe parameters) have been entered accurately. Check to be sure the transducer mounting method has been selected properly, the pipe is not worn-out, and the liner is not too thick. Make sure there is indeed fluid in the pipe or the transducer is not too close to a valve or elbow, and there are not too many air bubbles in the fluid, etc. With the exception of these reasons, if there is still no signal detected, the measurement site has to be changed. (5) Make sure that the Flow meter is able to run properly with high reliability. The stronger the signal strength displayed, the higher the Q value reached. The longer the Flow meter runs accurately, the higher the reliability of the flow rates displayed. If there is interference from ambient electromagnetic waves or the signal detected is too poor, the flow value displayed is not reliable; consequently, the capability for reliable operation is reduced. (6) After the installation is complete, power on the instrument and check the result accordingly.
16
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
5.Operating Instructions
5.1. System Normal Identification
If the letter “*R” displays on the screen, it indicates system normal. If the letter “D” is displayed, it indicates that system is adjusting the signal gain prior to the measurement. Also, it means system normal. Only when the adjustment takes too long without stopping, can system be identified as abnormal. Letter “E” indicates no signal is being detected. Check the transducer wiring connections are correct, the transducers are installed firmly, etc. For further information, please refer to “Error Diagnosis”.
5.2. Low Flow Cutoff Value
The data in M21 is Low Flow Cutoff Value. If the flow rate falls below the low flow cutoff value, the flow indication is driven to zero. This function can prevent the flow meter from displaying flow as “0”after a pump was shut down, but there is still liquid movement in the pipe, which will result in cumulative error. Generally, 0.03m/s is recommended to enter as the low flow cutoff point. The low flow cutoff value has no relation to the measurement results once the velocity increases over the low flow cutoff value.
5.3. Zero Settings
Once zero flow occurs, a zero point may indicate on each measuring instrument, but the displayed measuring value is not equal to “0”, this value indicates “Zero”. To any measuring instrument, the smaller the “Zero” is, the better the quality is. Conversely, if the Zero is too big, that indicates the quality of the instrument is poor. If the zero set point is not at true zero flow, a measurement difference may occur. The smaller the physical measurement capacity is, the larger the measurement difference from the zero point will exist. Only when zero point reduced to a definite degree, as compared with the physical measurement capacity, can the measuring difference from zero point be ignored. For an ultrasonic Flow meter, the measurement error from zero point cannot be ignored under low flow conditions. It is necessary to perform a static zero set calibration to improve low flow measurement accuracy. Cutoff Zero In Window M22- Cutoff- 1.Yes, window will show the “success” and back to M01 when you cut off the zero point successfully. Performing Set Zero In Window M22- Reset
5.4. Scale Factor
Scale factor refers to the ratio between “actual value” and “reading value”. For example, when the measurement is 2.00, and it is indicated as 1.98 on the instrument, the scale factor reading is 2/1.98. This means that the best scale factor constant is 1. However, it is difficult to keep the scale factor as “1” on the instrument especially in batch productions. The difference is called “consistency”. During operation, there still exists possible difference in pipe parameters, etc. The “scale factor” may be necessary when used on different pipes. Thus, scale factor calibration is specially designed for calibrating the differences that result from application on different pipes. The scale factor entered must be one that results from actual flow calibration. The scale factor can be input in Window M26.
17
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
5.5. System Lock
System lock is intended to prevent operation error due to tampering by unauthorized personnel. M54 is for system lock, unlock it by using the selected password only. If “lock” is displayed on the screen, then enter the correct password. Keep the password in mind or recorded in a safe place, otherwise the instrument cannot be used.
5.6. 4 ~ 20mA Current Loop Output
With a current loop output exceeding an accuracy of 0.1%, the flow meter is programmable and configurable with outputs such as 4 ~ 20mA or 0 ~ 20mA selected in Menu 32. For details, please refer to Menu 32 in “Window Display Explanations”. In Window M32- Range- LowL, enter a 4mA flow value. Enter the 20mA flow value in Window M32Range-UpperL. For example, if the flow range in a specific pipe is 0 ~ 1000m3/h, enter 0 in Window M32 and 1000 in Window M32. If the flow ranges from -1000 ~ 0 ~ 2000m3/h, configure the 20 ~ 4 ~ 20mA output by selecting in Window M32 when flow direction is not an issue. Enter 1000 in Window M32 LowL and 2000 in Window M32 UpperL. When flow direction is an issue, module 0 ~ 4 ~ 20mA is available. When the flow direction displays as negative, the current output is in range of 0 ~ 4mA, whereas the 4 ~ 20mA is for the positive direction. The output module options are displayed in Window M32. Calibrating and testing the current loop is performed in Window M32-Check. Complete the steps as follows: Use “” and “” to switch. “check 4mA”, “check 8mA”, “check 16mA”, “check 20mA” readings, connect an ammeter to test the current loop output and calculate the difference. Calibrate the 4-20mA is in M62.
5.7. Frequency Output
The flow meter is provided with a frequency output transmitter function. The high or low frequency output displayed indicates the high or low flow rate reading. The user can reset the frequency output as well as flow rate as the user’s actual requirements. For example: if a pipe flow range is 0 ~ 5000m3/h, the relative frequency output required is 100 ~ 1000Hz, and the configuration is as follows: In Window M33-Range-LowerL (lower limit frequency output flow value), input 0; In Window M33-Range -UpperL (upper limit frequency output flow value), input 5000; In Window M33-Mode-Frange( frequency range), input 1001000; In Window M33-Mode-Option, select “a.Flow Rate”; Typical OCT Output wiring diagram as below:
18
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
5.8. Totalizer Pulse Output
Each time the flow meter reaches a unit flow, it may generate a totalizer pulse output to a remote counter. The totalizer pulse output can be transmitted through OCT or a relay. Therefore, it is necessary to configure OCT and the relay accordingly. (Please refer to Window M33 and M34). For example, if it is necessary to transmit the positive totalizer pulse through a relay, and each pulse represents a flow of 10m3, the configuration is as follows: In Window M41-Unit, select the totalizer flow unit “m3”; In Window M41-MULT, select the scale factor “e. x10”; In Window M34-Option, select “g. POS Total “;
Attention Make sure to select an appropriate totalizer pulse. If the totalizer pulse is too big, the output cycle will be too long; if the totalizer is too small, the relay will operate too faster, you may shorten the life of the relay, as well as skip some pulses. The totalizer is recommended to transmit within the range of 1 ~ 3 pulse per second.
5.9. Alarm Programming
The on-off output alarm is generated through OCT or transmission to an external circuit by opening or closing a relay. The on-off output signal is activated under the following conditions:
(1) Signal not detected; (2) Poor signal detected; (3) The flow meter is not ready for normal measurement; (4) The flow is in the reverse direction (back flow). (5) The analog outputs exceed span by 120%. (6) The frequency output exceeds span by 120%. (7) The flow rate exceeds the ranges configured (Configure the flow ranges using the software alarm system.
There are two software alarms: Alarm#1 and Alarm #2. Example 1: When flow rate exceeds 300 ~ 1000 m3/h, in order to program the relay output alarm, Complete the steps as follows:
(1) In Menu 35, Alarm1 LowL 300; (2) In Menu 35, Alarm1 Upper 1000; (3) In Menu 34, Relay Setting-Option-d.Alarm1
5.10.4-20mA Analog Output Calibration
Attention Each flow meter has been calibrated strictly before leaving factory. It is unnecessary to carry out this step except when the current value (detected while calibrating the current loop) displayed in Window M32 is not identical with the actual output current value.
The hardware detect window must be activated prior to calibration the Analog Output. The procedure is as follows: Menu 62 is for 4-20mA calibration, if you need enter the pass word to enter. With no effect to next power on, this window will close automatically as soon as the power is turned off. Use “” and “” to switch. Calibrate the current loop 4mA output. Use an ammeter to measure the output current of current loop and adjust the displayed numbers at the same time. Watch the ammeter until it reads 4.00.
19
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
Stop at this point, the 4mA has been calibrated. Use “” and “” to switch. Calibrate the current loop 20mA output. The method is the same as 4mA calibration. The results are automatically saved in EEPROM and won’t lose when power off.
5.11.SD Card Operation
5.1.1. Specifications
Data collection interval: any interval settings from 1 to 3600 seconds are OK according to the requirement. Data content: date and time, flow rate, flow velocity, total flow, positive totalizer, negative totalizer. Data storage format:
a=2017-11-16,16:21:12 b=+2.652471E+00 m3/h c=+9.380460E-02 m/s d=+3.520580E+02 m3 e=+3.520580E+02 m3 f=+0.000000E+00 m3 g=+0.000000E+00 GJ/h h=+0.000000E+00 GJ i=+0.000000E+00 GJ j=+0.000000E+00C k=+0.000000E+00CFile system format: FAT16. File type: plain text file (.TXT). File number: maximum 512pcs. It can save 120 bytes of data each time. If it is set to save once in per 5 seconds, the capacity of storing file in 24 hours is 1203600/5242073600byte2.1Mbyte, therefore, 1Gbyte SD card can store for days: 1024/2.1= 487.6487 days. When the capacity of the SD card is full, the new data will override the earliest files automatically.
5.1.2. Install or Remove the SD card while the meter is powered on
If the operator desires to insert the SD card with power on, please remove the power with power off. The following operation is to be used:
Attention: Do not remove the SD card from the reader while actively working with the data. Data should be saved and stored in a separate location on the PC, and then processed form that file location. Processing the data directly from the SD card file location on the PC could result in losing or destroying data if the SD card is removed while still being processed.
20
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
5.1.3. ESN
We provide the flow meter with a unique electronic serial number to identify each flow meter for the convenience of the manufacturer and customers. The ESN, instrument types and versions are able to view in Window M50.
ATTENTION
Other Operating Refer to “6.2 Window Display Explanations”.
21
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

6. Windows Display Explanations
6.1. Windows Display Codes

M0X M1X M2X M3X

Easy Introduction Display Value and Condition R- System Normal E – Signal Not Detected *D- Adjusting Gain
Installation Setting
Calibration Setting
Input and Output Settings

A class of the menu M00 Flow Totalizer M01 Flow Rate M02 Hot M03 Cool M04 Status M10 Pipe Settings M11 Lining Settings M12 Liquid Settings M13 Transducer Settings M14 Installation Space M20 Damping M21 Low Flow Cut off Value M22 Zero Point Settings M23 Totalizer M24 Temperature M25 Power -off COMP M26 K Factor M27 Correction M28 SQA M30 Serial Port Parameter M31 AI Settings M32 CL Settings M33 OCT Settings

22
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

M3X M4X M5X M6X

Input and Output Settings Flow Unit Opinions System Settings Others

M34 Relay Settings M35 Alarm Value Settings M36 Ration M37 Micro SD Settings (option ) M40 Toggle Units M41 Flow Units M42 Energy Units M43 Temperature Units M50 Serial Number M51 Time and Date M52 Key Ton M53 Language M54 System Lock M55 System Reset M60 Date Totalizer M61 Running Timer M62 CL Adjust M63 RTD Adjust M64 AI Adjust

NOTEThe other menu features are retained by manufacturers.

23
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

6.2. Display Explanations
M00
Flow Total Display Net Totalizer. Display Positive totalizer. Display Negative totalizer.

Use

and

to switch.

M01
Flow Rate
Display the Flow Rate and Flow Total
Display the Velocity.

  • Flow Rate and the Flow Total switch every 6 seconds, Use the ENTER to stop the switch.

24
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

M02
Heat Rate Display the Heat Total. Display the Heat Rate and the Inlet Water Temp and Outlet Water Temp. NOTE: Instrument needs energy capacity.

x 0.001 (E-3)

x 0.01(E-2)

x 0.1(E-1)

x 1(E+0)

x 10(E+1)

x 100(E+2)

x 1000(E+3)

x 10000(E+4)

  • Heat Rate and the Heat Total switch every 6 seconds, Use the ENTER to stop the switch.

Cool Rate
Display the Cool Total.
Display the Cool Rate and the Inlet Water Temp and Outlet Water Temp.
NOTE: Instrument needs energy capacity.

x 0.001 (E-3)

x 0.01(E-2)

x 0.1(E-1)

x 1(E+0)

x 10(E+1)

x 100(E+2)

x 1000(E+3)

x 10000(E+4)

  • Cool Rate and the Cool Total switch every 6 seconds, Use the ENTER to stop the switch.

25
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
M04
Status Display the Signal strength, the Upstream signal strength and Downstream signal strength. Signal quality Q is indicated by 00 ~ 99. Therefore, 00 indicates the poorest signal while 99 indicates the best signal. Normally, signal quality Q value should be better than 60
Display the measured fluid sound velocity. Normally this value should be approximately equal to the entered value in Window M12. If the difference is too large, it probably results from an incorrect value entered in Window M12 or improper installation of the transducers. Display the ratio between the actual measured transmit time and the calculated transmit time according to customer’s requirement. Normally the ratio should be 100±3. If the difference is too large, the user should check that the parameters are entered correctly, especially the sound velocity of the fluid and the installation of the transducers. This data is of no use before the system is ready. Display the measured ultrasonic average time (unit: us) and delta time of the upstream and downstream (unit: ns) time. The velocity calculation in the flow meter is based on the two readings. The delta time is the best indication that the instrument is running steadily. Normally the fluctuation in the ratio of the delta time should be lower than 20%. If it is not, it is necessary to check if the transducers are installed properly or if the parameters have been entered correctly.
26
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
M10
Pipe settings Enter the pipe outer diameter; the pipe outer diameter must range from 25mm to 1200mm. Note: Enter Either pipe outer diameter or pipe outer perimeter Enter the pipe wall thickness. Pipe wall thickness is necessary.
Enter pipe material. The following options are available: 0. PVC 1. CS Carbon Steel 2. SSP Stainless Steel Pipe 3. CIP Cast Iron Pipe 4. DIP Ductile Cast Iron Pipe 5. Copper 6. Alu. Aluminum pipe 7. ACP Asbestos Cement Pipe 8. FPG Fiberglass Pipe 9. Other It is possible to enter other materials, which are not included in previous eight items. Once item 9 is selected, the relevant pipe sound velocity must be entered.
M11
Lining Enter liner thickness. Refer to item 9 “Other”; it is possible to enter other materials, which are not included in previous eight items. Once item 9 is selected, the relevant pipe sound velocity must be entered. Use and to switch.
27
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
M11
Lining Enter liner thickness. Select the Liner Material. The following options are available: 0. No liner 1. Tar Epoxy 2. Rubber 3. Mortar 4. PP Polypropylene 5. Polystryol 6. PS Polystyrene 7. Polyester 8. PE Polyethylene 9. Ebonite 10. Teflon 11. Other Item 11 “Other” is available to enter other materials that are not included in previous ten items. Once the “Other” is selected, the relevant liner sound velocity must be entered.
M12
Medium Select the temperature of water. Temperature should be 0-80 deg. C. Press “Enter” to confirm. Note: Room temperature is 25 deg.C
28
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
M13
Transducer Select transducer type The following options are available: 0. Clamp-On C 1. Clamp-On D 2. Clamp-On X 3. Plus-In 4. Plus-In X Select transducer Mounting Methods Three mounting methods are available: 0. V Reflect method 1. Z Direct method 2. N
29
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
M14
Installation space This value is calculated by the flow meter The operator must mount the transducer according to the transducer spacing displayed (ensure that the transducer spacing is measured precisely during installation). The system will display the data automatically after the pipe parameter had been entered.
M20
Damping The damping factor ranges from 1 ~ 999 seconds.1 indicates no damping; 999 indicates the maximum damping. The damping function will stabilize the flow display. Usually a damping factor of 3 to 10 is recommended in applications.
M21
Low Vel. Cut off Low Flow Cut off is used to make the system display as “0” value at lower and smaller flows to avoid any invalid totalizing. For example, if the cutoff value is set as 0.03, system will take all the measured flow velocity values from – 0.03 to + 0.03 as “0”. Generally, 0.03 is recommended in most applications.
30
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
M22
Zero Settings When fluid is in the static state, the displayed value is called “Zero Point”. When “Zero Point’ is not at zero in the flow meter, the difference is going to be added into the actual flow values and measurement differences will occur in the flow meter. Set zero must be carried out after the transducers are installed and the flow inside the pipe is in the absolute static state (no liquid movement in the pipe). Thus, the “Zero Point” resulting from different pipe mounting locations and parameters can be eliminated. The measuring accuracy at low flow is enhanced by doing this and flow offset is eliminated.
Select “YES”; reset “Zero Point” which was set by the user.
This method is not commonly used. It is only suitable for experienced operators to set zero under conditions when it is not preferable to use other methods. Enter the value manually to add to the measured value to obtain the actual value. For example: Actual measured value 240 m3/H Value Deviation 10 m3/H Flow meter Display 250 m3/H Normally, set the value as “0”. Use “” and “” to switch.
31
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
M23
Totalizer Select the totalizer type 0. POS Positive Totalizer 1. NEG Negative Totalizer 2. NET Select energy type 0. Heat 1. Cool Select “ON”/”OFF” to switch the totalizer.
Select the flow totalizer value you want Reset 0. POS Positive Totalizer 1. NEG 2. NET Negative Totalizer 3. All Select the energy totalizer value you want Reset 0. Heat 1. Cool 2. All
M24
Temperature Select Heat Input Options: 0. RTD 1. AI Use “” and “” to switch
32
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
Temperature Sensitivity Setting When the delta temperature is less than the sensitivity set, energy will not be accumulated. Set the adjustable temperature range of 0 ~ 20. The factory default setting is 0.2 . Select Specific Heat Options: 0. CJ128 SHC 1. Other Use “” and “” to switch.
M25
Power Down Correction Switch With the function of power down automation correction switch, the flow lost in an offline session can be estimated and automatically adjusted. The estimate is based on the average value, which is obtained from flow rate before going offline and flow measured after going online the next time, multiplied times the time period that the meter was offline. Select “ON” to use this function, select “OFF” to cancel this function.
M26
K Factor The K factor is used to modify the measurement results. The user can enter a numerical value (other than “1”) according to the actual calibration results.
33
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
M27
Correction K-Array Sectional Correction ON: Open the Sectional Correction Function; OFF: Close the Sectional Correction Function
Delay correction Engineer menu, suggest customer use the factory setting.
TPC Transducers power control Engineer menu, suggest customer use the factory setting. 0. Auto 1. Low 2. High
34
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
M28
SQA Statistic Analysis
M30
RS232/RS485 Serial Port Setting . 2400 None . 4800 None . 9600 None . 19200 None . 38400 None . 56000 None You can setting the order as following : a. 1-0: 3-2 b. 0-1:2-3 c. 3-2:1-0 d. 2-3:0-1
35
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
M31
AI Setting Display analog input AI1 analog value.
Display analog input AI2 analog value.
M32
CL Setting Current Loop Mode Options
Select the CL Range value Set the CL output value according to the flow value at 4mA or 0MA. Set the CL output value according to the flow value at 20mA.
36
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
4-20mA check opinions a. Check 4mA b. Check 8mA c. Check 12mA d. Check 20mA
M33
OCT Setting The following signal options are available: a. Flow Rate b. POS Total c. NEG Total d. NET Total e. Energy Rate f. Heat Total g. Cool Total h. Rationing i. Uart CTRL
Select the OCT Range value.
OCT check opinions a. Check 500 b. Check 1000 c. Check 3000 d. Check 5000
37
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
M34
Relay Setting The following signal options are available: a. No Signal b. *E c. Reverse d. Alarm1 e. Alarm2 f. Ration g. POS Total h. NEG Total i. NET Total j. Not Using
M35
Alarm Setting Enter the Lower alarm value, any of the measured flow, which is lower than the low value, will activate the alarm in the OCT hardware or relay output signal.
Enter the Upper alarm value, any of the measured flow, which is higher than the high value, will activate the alarm in the OCT hardware or relay output signal.
38
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
M36
Ration Following is the Ration opinions: a. Key CTRL b. AI1 CTRL c. AI2 CTRL d. Uart CTRL
M37
Micro SD Following is the opinions for the record. a. No Energy b. All Input the data collection time interval in this menu. Time is in seconds. The interval can be selected in the range of 1 ~ 3600 seconds. NOTE: this is an option.

M40
Toggle Unit Select the measurement unit as follows: a. Metric b. British

M41

Flow Unit

The following flow rate units are available:

0. Cubic Meters (m3)

1. Liters

(1)

2. USA Gallons (GAL)

3. Imperial Gallons (Imp gal)

4. Million Gallons (mg)

5. Cubic Feet

(cf)

6. USA Barrels (US bbl)

7. Imperial Barrels (Imp bbl)

8. Oil Barrels

(Oil bbl)

The following time units are available:

/Day

/Hour

/Min

/Sec

39
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

Factory default is Cubic Meters/hour.

a. x 0.001 (E-3)

b. x 0.01(E-2)

c. x 0.1(E-1)

d. x 1(E+0)

e. x 10(E+1) g. x 1000(E+3)

f. x 100(E+2)
h. x 10000(E+4)

M42

Energy Unit The following Energy units are available:

0. Giga Joule (GJ) 1. Kilocalorie (Kc)

2. MBtu

3. KJ

4. Btu

5. KWh

6. MWh

7. TH

a. x 0.001 (E-3) c. x 0.1(E-1) e. x 10(E+1) g.x 1000(E+3)

b. x 0.01(E-2) d. x 1(E+0) f. x 100(E+2) h.x 10000(E+4)

M43
Temperature Unit a. b. Use “” and “” to switch.

40
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
M50
Serial Number Display electronic serial number (S/N) of the instrument. This S/N is the only one assigned to each flow meter ready to leave the factory. The factory uses it for files setup and for management by the user. SVN is the software version
M51
Time and Data Date and time modifications are made in this menu.
M52
Key Tone Use this menu to “ON”/ “OFF” the key tone.
M53
Language Setting the language of the flow meter 0.English 1.Chinese
41
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
M54
System Lock Lock the instrument. Once the system is locked, any modification to the system is prohibited, but the parameter is readable. Entering your designated password correctly can be the only way to “Unlock”. The password is composed of 6 numbers. (Please contact the representative or manufacturer as soon as possible when the password is lost.)
M55
System Reset Select 1. Reset to make the instrument back to factory.
M60
Date Totalizer The following options are available:
0. Day 1. Month 2. Year In this window, it is possible to review the historical flow data net totalizer for any day for the last 31 days, any month for last 12 months and any year for last 6 years.
M61
Running Time With this function, it is possible to view the total Working days since the flow meter left the factory.
42
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
M62
CL Adjust This menu is for the 4-20mA calibration; enter the pass word to adjust.
M63
RTD Adjust This menu is for the RTD calibration; enter the pass word to adjust.
M64
RTD Adjust This menu is for the AI calibration; enter the pass word to adjust.
Warning The entire Menu which related to the temperature, cooling, heating, energy. Only display when it’s an energy meter. Flow meter doesn’t have the function.
43
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

7. Error Diagnoses

The ultrasonic flow meter has advanced self-diagnostics functions and displays any errors in the upper right corner of the LCD via definite codes in a date/time order. Some errors can be detected during normal operation. Undetectable errors caused by unskilled operation, incorrect settings and unsuitable measurement conditions can be displayed accordingly during work. This function helps the user detect the errors and find causes quickly; thus, problems can be solved in a timely manner according to the solutions listed in the following table.
If a problem still exists, please contact the factory or the factory’s local representative for assistance.

7.1. Table 1. Error Codes and Solutions (during operation)

Codes

The upper right corner of the screen

*R System Normal

Signal Not Detected

*E

Adjusting Gain *D

Causes

Solutions

  • System normal.

  • Signal not detected.

  • Spacing is not correct between the transducers or not enough coupling compound applied to face of transducers.

  • Transducers installed improperly.

  • Scale is too thick.

  • New pipe liner.

  • Attach transducer to the pipe and tighten it securely. Apply a plenty of coupling compound on transducer and pipe wall.

  • Remove any rust, scale, or loose paint from the pipe surface. Clean it with a file.

  • Check the initial parameter settings.

  • Remove the scale or change the scaled pipe section. Normally, it is possible to change a measurement location. The instrument may run properly at a new site with less scale.

  • Wait until liners solidified and saturated.

  • Adjusting gain for normal measurement.

44
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

7.2. Frequently Asked Questions and Answers
Question: New pipe, high quality material, and all installation requirements met: why still no signal detected?
Answer: Check pipe parameter settings, installation method and wiring connections. Confirm if the coupling compound is applied adequately, the pipe is full of liquid, transducer spacing agrees with the screen readings and the transducers are installed in the right direction.

Question: Old pipe with heavy scale inside, no signal or poor signal detected: how can it be resolved?

Answer:

Check if the pipe is full of fluid. Try the Z method for transducer installation (If the pipe is too close to a wall, or it is necessary to install the transducers on a vertical or inclined pipe with flow upwards instead of on a horizontal pipe).
Carefully select a good pipe section and fully clean it, apply a wide band of coupling compound on each transducer face (bottom) and install the transducer properly.
Slowly and slightly move each transducer with respect to each other around the installation point until the maximum signal is detected. Be careful that the new installation location is free of scale inside the pipe and that the pipe is concentric (not distorted) so that the sound waves do not bounce outside of the proposed area.
For pipe with thick scale inside or outside, try to clean the scale off, if it is accessible from the inside. (Note: Sometimes this method might not work and sound wave transmission is not possible because of the a layer of scale between the transducers and pipe inside wall).

Question: Why is the CL output abnormal?
Answer: Check to see if the desired current output mode is set in Window M32-Mode. Check to see if the maximum and minimum current values are set properly in Windows M32-Range. Re-calibrate CL and verify it in Window M32-Check.

Question: Why is the flow rate still displayed as zero while there is fluid obviously inside the pipe and a symbol of “R” displayed on the screen?
Answer: Check to see if “Set Zero” was carried out with fluid flowing inside the pipe (Refer to Window M22). If it is confirmed, recover the factory default in Window M22-Reset.

45
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

8. Appendix 3 ­ Serial Interface Network Use and Communications Protocol

8.1. Overview

The flow meter has perfect communication protocol. It can also be connected to a RS-485 Modbus .
Two basic schemes can be chosen for networking, i.e. the analog current output method only using the flow meter or the RS232 communication method via serial port directly from the flow meter. This method is suitable to replace dated instruments in old monitoring networks. The later method is used in new monitoring network systems. It has advantages include low hardware investment and reliable system operation.
When the serial port communications method is directly used to implement a monitoring network system, the address identification code of the flow meter is used as a network address code. Expanded command set with [W] is used as communication protocol.
RS-232 (Cable length 0 ~ 15m ) or RS-485 ( cable length 0 ~ 1000m ) can be directly used for data transmission links for a short distance. Current loop can be used in medium or long distance transmission.
When the flow meter is used in a network environment, various operations can be performed by a host device, except for programming of the address identification code, which needs to be done via the flow meter keyboard.
The command answer mode is used in data transmission, i.e. the host device issues commands and the flow meter answers correspondingly.
Common/special flow / thermal data monitoring system developed by our company can be used for flow data collection. Based on characteristics of the flow meter, the system makes full use of software and hardware designs with flow meter features. The system is simple, clear, economical, and reliable in operation.
Attention
In the communication protocol used functions, RS232 and RS485 serial communications cannot be used at the same time.
8.2. Serial port definitions

Flow meter – RS232: TXD send RXD receive GND ground
PC: PIN 1 empty PIN 2 RXD send

PIN 3 TXD send PIN 4 ground PIN 5 ground PIN 6 empty PIN 7 empty PIN 8 empty PIN9empty

46
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
8.3. Direct connection via RS232 to the host device
See the below list of flowmeter serial port definitions.

8.4. Communications protocol and the use

The flow meter supports these three communication protocols: FUJI Protocol, MODBUS-C Protocol, MODBUS-I Protocol.
8.4.1. HL Protocol

The communication protocol format used by the ultrasonic flow meter is an expanded set of the HL FLV series flow meter protocol. The host device requests the flow meter to answer by sending a “command”. The baud rate of asynchronous communication (Primary station: computer system; Secondary station: ultrasonic flow meter) is generally 9600BPS. A single byte data format (10 bits): one start bit, one stop bit and 8 data bits. Check bit: none.
A data character string is used to express basic commands and a carriage return (ENTER) is used to express the end of a command. The characteristic is that the string of data is flexible. The order applies to both RS232 and RS485. Frequently used commands are as follows:
Communications commands

Command

Description

Data format

RFR(cr)(lf)

Return instantaneous flow

±d.ddddddE±dd(cr) Note1

RVV(cr)(lf)

Return instantaneous velocity

±d.ddddddE±dd(cr)

RT+(cr)(lf)

Return positive accumulative flow

±ddddddd.dE±d(cr) Note 2

RT-(cr)(lf)

Return negative accumulative flow

±ddddddd.d±d(cr)

47
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

RTN(cr)(lf)

Return net accumulative flow

±ddddddd.d±d(cr)

RTH(cr)(lf)

Return net accumulative energy(hot)

±ddddddd.d±d(cr)

RT-(cr)(lf)

Return net accumulative energy(cold)

±ddddddd.d±d(cr)

RER(cr)(lf)

Return instantaneous energy value

±d.ddddddE±dd(cr)

RA1(cr)(lf)

Return analog input value of AI1 (Temperature, Pressure, etc.)

±d.ddddddE±dd(cr)

RA2(cr)(lf)

Return analog input value of AI2 (Temperature, Pressure, etc.)

±d.ddddddE±dd(cr)

RID(cr)(lf)

Return Net address of the instrument

ddddd(cr) 5 bits in length

RSS(cr)(lf)

Return signal intensity

UP:dd.dDN:dd.dQ=dd(cr)

REC(cr)(lf)

Return current error code

R/D/*E Note 3

RRS(cr)(lf)

Return Relay Status

ON/OFF(cr)

RDT(cr)(lf)

Current date and time

yy-mm-ddhh:mm:ss(cr)

RSN(cr)(lf)

Return serial number

dddddddt(cr) Note 4

SFQdddd.d(cr)(lf) OCT setting

dddd.d(cr) Successful setting will back to “OK”

SCLdd.d(cr)(lf)

Current setting

dd.d(cr) Successful setting will back to “OK”

SRS(cr)(lf)

Start quantitative control

OK(cr) Successful setting will back to “OK”

P

Prefix of return command with check

Note 5

W

Networking command prefix of numeric string address

Note 6

Note:
1.(cr)expresses carriage return. Its ASCII value is 0DH. (lf) expresses line feed. Its ASCII value is 0AH.
2.”d” expresses 0-9 number. 0 value is expressed as +0.000000E+00.
3.”d” expresses 0-9 numbers. There is no decimal point in integral part before “E”.
4. dddddddd means the serial number of the instrument, t means the model of the instrument.
5. The character P can be added before every basic command. It means that the transferred data has CRC verification. The method of verification is to add all of the data back to the data, which is cumulative and binary, and its low 8-bit binary data is taken.
E.g. The return information of the RT(cr)(lf) is :+1234567E+0m3(cr)(lf), (the relative binary system data is 2BH31H32H33H34H35H36H37H45H2BH30H6DH33H20H0DH

48

KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

0AH) The sum of all of its return data is =2BH+31H+32H+33H+34H+35H+ 36H+37H+45H+2BH+30H+6DH+33H+20H=2F7, The low 8-bit data of its binary is F7.
Therefore, the data of the order PRT (cr) (lf) is called + 1234567E + 0m3!F7 (cr) (lf), “!”For delimiters, the preceding is the character of the summation, followed by a check code of 1 byte.
6.Usage of prefix W: W+ numeric string address code +basic command. Value range of the numeric string is 0 ~ 255, except 13 (0DH carriage return ), 10 ( 0AH line feed ). If the instantaneous velocity of No. 123 flow meter is to be accessed, the command W123DV (cr)(lf) can be issued. The corresponding binary code is 57H, 31H, 32H, 33H, 44H, 56H, 0DH, 0AH, only the same instrument with the same address of the Internet address and command will send back the data.
7.W and P commands can be used in combination, for example, W123PRT +, which means that the instrument that reads the network address is the cumulative value of the instrument with 123, and its return data has eight accumulations and checksums.”s” expresses ON or OFF or UD. For example, “TR:ON, RL:ON” expresses that the OCT and relay are in an actuated status; “TR:UDRL:UD” expresses that the OCT and relay are not actuated.
8.4.2. MODBUS-I Communication Protocol

This MODBUS-I Protocol uses RTU transmission mode. The Verification Code uses CRC-16-IBM (polynomial is X16+X15+X2+1, shield character is 0xA001) which is gained by the cyclic redundancy algorithm method. MODBUS-I RTU mode uses hexadecimals to transmit data.
1. MODBUS-I Protocol Function Code and Format The flow meter protocol supports the following two-function codes of the MODBUS:

Function Code

Performance data

0x03

Read register

0x06 2. MODBUS Protocol function code 0x03 usage The host sends out the read register information frame format:

Write single register

Slave Address 1 byte

Operation Function Code
1 byte

First Address Register 2 bytes

Register Number 2 bytes

Verify Code 2 bytes

0x01 ~ 0xF7

0x03

0x0000 ~ 0xFFFF

0x0000 ~ 0x7D

CRC (Verify )

The slave returns the data frame format:

Slave Address

Read Operation Function Code

Number of Data Bytes

1 byte

1 byte

1 byte

0x01 ~ 0xF7

0x03

2xN*

N* = Data register number 3. MODBUS Protocol function code 0x06 usage

Data Bytes Nx2 byte Nx2 ( Data )

Verify Code 2 bytes
CRC ( Verify )
49

KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

The host sends a command to write a single register information frame format ( function code 0x06 ):

Slave Address 1 byte

Operation Function Code
1 byte

Register Address 2 bytes

Register Data 2 bytes

Verify Code 2 bytes

0x01 ~ 0xF7

0x06

0x0000 ~ 0xFFFF

0x0000 ~ 0xFFFF CRC ( Verify )

The slave returns the data frame format ( function code 0x06 ):

Slave Address 1 byte

Operation Function Code
1 byte

Register Address 2 bytes

Register Data 2 bytes

Verify Code 2 bytes

0x01 ~ 0xF7

0x06

0x0000 ~ 0xFFFF

0x0000 ~ 0xFFFF CRC ( Verify )

The range of flow meter addresses 1 to 247 (Hexadecimal: 0x01 ~ 0xF7 ), and can be checked in the Menu 46. For example, decimal number “11” displayed on Menu 46 means the address of the flow meter in the MODBUS protocol is 0x0B.
The CRC Verify Code adopts CRC-16-IBM (polynomial is X16+X15+X2+1, shield character is 0xA001) which is gained by the cyclic redundancy algorithm method. Low byte of the verify code is at the beginning while the high byte is at the end.

For example, to read the address 1 (0x01) in the RTU mode, if the instantaneous flow rate uses hour as a unit m3/h, namely reads 40005 and 40006 registers data, the read command is as follows:

0x01

0x03

0x00 0x04

0x00 0x02

0x85 0xCA

Flow meter Address Function Code First Address Register Register Numbers CRC Verify Code Flow meter returned data is (assuming the current flow=1.234567m3/h)

0x01

0x03

0x04

0x06 0x51 0x3F 0x9E0x3B 0x32

Flow meter Address Function Code Data Bytes Data1.2345678 CRC Verify Code

The four bytes 3F 9E 06 51 is in the IEEE754 format single precision floating point form of 1.2345678.

Pay attention to the data storage order of the above example. Using C language to explain the data, pointers can be used directly to input the required data in the corresponding variable address, the low byte will be put at the beginning, such as the above example 1.2345678 m/s, 3F 9E 06 51 data stored in order as 51 06 9E 3F.

For example, it converts the address 1 (0x01) to 20x02under the RTU mode, so to write the data of flowmeter 44100 register as 0x02, the write command is as follows:

0x01

0x06

0x10 0x03

0x00 0x02 0xFC 0xCB

Flow meter Address Function Code Register Address Register Number CRC Verify Code

Flow meter returned data is:

0x01

0x06

0x10 0x03

0x00 0x02 0xFC 0xCB

Flow meter Address Function Code Register Address Register Number CRC Verify Code

4. Error Check

The flow meter only returns one error code 0x02 which means data first address in error.

50
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

For example, to read address 1 (0x01) of the flow meter 40002 register data in the RTU mode, the flow meter considers it to be invalid data, and sends the following command:

0x01

0x03

0x00 0x01

0x00 0x01 0xD5 0xCA

Flow meter Address Function Code Register Address Register Number CRC Verify Code

Flow meter returned error code is:

0x01

0x83

0x02

0xC0 0xF1

Flow meter Address Error Code Error Extended Code CRC Verify Code

5. MODBUS Register Address List The flow meter MODBUS Register has a read register and a write single register. a) Read Register Address List (use 0x03 function code to read)

PDU Address

Register

Read

$0000

40001 Flow/s – low word

$0001

40002 Flow/s – high word

$0002

40003 Flow/m – low word

$0003

40004 Flow/m- high word

$0004

40005 Flow/h – low word

$0005

40006 Flow/h – high word

$0006

40007 Velocity ­ low word

$0007

40008 Velocity ­ high word

$0008

40009 Positive total ­ low word

$0009

40010 Positive total ­ high word

$000A

40011 Positive total ­ exponent

$000B

40012 Negative total ­ low word

$000C

40013 Negative total ­ high word

$000D

40014 Negative total ­ exponent

Write

Type

No. registers*

32 bits real

2

32 bits real

2

32 bits real

2

32 bits real

2

32 bits int.

2

16 bits int.

1

32 bits int.

2

16 bits int.

1

51
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

$000E $000F $0010 $0011 $0012 $0013 $0014 $0015 $0016 $0017 $0018 $0019 $001A $001B $001C
$001D

40015 Net total ­ low word

40016 Net total ­ high word

40017 Net total ­ exponent

40018 Energy flow ­ low word

40019 Energy flow ­ high word

40020 Energy total(hot) ­low word

40021 Energy total(hot) ­high word

40022 Energy total(hot) ­ exponent

40023 Energy total(cold ) ­high word

40024 Energy total(cold ) ­ exponent

40025 40026 40027 40028 40029

Energy total(cold ) ­ exponent Up signal int ­ low word Up signal int ­ high word Down signal int ­ low word Down signal int ­ high word

40030 Quality

$001E

40031 Error code ­ char 1

$003B $003C $003D $003E $003F

40060 Flow velocity unit ­char 1,2 40061 Flow velocity unit ­char 3,4 40062 Flow rate unit ­char 1,2 40063 Flow rate unit ­char 3,4 40064 Flow total unit ­ char 1,2

32 bits int.

2

16 bits int.

1

32 bits int.

2

32 bits real 16 bits int. 32 bits real 16 bits int. 32 bits real

2

1

2

1

2

0 ~ 99.9

32 bits real 16 bits int.
String String

2

0 ~ 99.9

1

0 ~ 99

Refer to “Error

1

Analysis” for detailed codes

meanings.

2

Only m/s right now

String

2

String

1

Note 1

52
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

$0040 $0041 $0042 $0043 $0044 $0045 $0046

40065 Energy rate unit ­ char1,2 40066 Energy rate unit ­ char 3,4 40067 Energy total unit ­ char 1,2 40068 Instrument address ­low word 40069 Instrument address ­high word 40070 Serial number ­ char 1,2 40071 Serial number ­ char 3,4

String

2

String

1

32 bits real

2

String

4

Note 2

$0047 $0048

40072 Serial number ­ char 5,6 40073 Serial number ­ char 7,8

String

4

$0049 $004a $004b $004c

40074 Analog Input AI1 Value- low word 40075 Analog Input AI1 Value- high word

32 bits real

40076 Analog Input AI2 Value- low word 40077 Analog Input AI2 Value- high word

32 bits real

2
Returned temperature value with RTD option
2

$004d $004e

40078 4-20mA Value- low word 40079 4-20mA Value- high word

32 bits real

2

Unit: mA

b) Single Write Register Address List ( use 0x06 performance code to write )

PDU Address Register

Description

Read/W rite

Type

No. registers*

$1003

44100

Flow meter address ( 1 – 255 )

R/W

16 bits int.

1

Communication Baud Rate 0 =

$1004

44101

2400,1 = 4800, 2 = 9600, 3 =

R/W

16 bits int.

1

19200, 4 = 38400,5 = 56000

Notes: 1.

The following flow rate units are available: 0. “m3” Cubic Meter 1. “l” Liters 2. “ga” Gallons

53

KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual
3. “ig” Imperial Gallons 4. “mg” Million Gallons 5. “cf” Cubic Feet 6. “ba” US Barrels 7. “ib” Imperial Barrels 8. “ob” Oil Barrels 2. The following energy units are available 0. “GJ” Giga Joule 1. “Kc” Kilocalorie 2. “MB” MBtu 3. “KJ” Kilojoule 4. “Bt” Btu 5. “Ts” US Tonnes 6. “Tn” US Tons 7. “kw” Kwh 3. 16 bits int–short integer, 32 bits int ­ long integer, 32 bits real–floating point number, String–alphabetic string
54
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

9. Appendix 6 – Flow Application Data

9.1. Sound Velocity for Various Materials Commonly Used

Pipe Material Steel ABS Aluminum Brass Cast iron Bronze Fiber glass-epoxy Glass Polyethylene PVC

Sound Velocity (m/s) 3206 2286 3048 2270 2460 2270 3430 3276 1950 2540

Liner Material Teflon Titanium Cement Bitumen Porcelain enamel Glass Plastic Polyethylene PTFE Rubber

Sound Velocity 1225 3150 4190 2540 2540 5970 2280 1600 1450 1600

55
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

9.2. Sound Velocity in Water (1 atm) at different temperatures

t()

v(m/s)

t()

v(m/s)

t()

0

1402.3

34

1517.7

68

1

1407.3

35

1519.7

69

2

1412.2

36

1521.7

70

3

1416.9

37

1523.5

71

4

1421.6

38

1525.3

72

5

1426.1

39

1527.1

73

6

1430.5

40

1528.8

74

7

1434.8

41

1530.4

75

8

1439.1

42

1532.0

76

9

1443.2

43

1533.5

77

10

1447.2

44

1534.9

78

11

1451.1

45

1536.3

79

12

1454.9

46

1537.7

80

13

1458.7

47

1538.9

81

14

1462.3

48

1540.2

82

15

1465.8

49

1541.3

83

16

1469.3

50

1542.5

84

17

1472.7

51

1543.5

85

18

1476.0

52

1544.6

86

19

1479.1

53

1545.5

87

20

1482.3

54

1546.4

88

21

1485.3

55

1547.3

89

22

1488.2

56

1548.1

90

23

1491.1

57

1548.9

91

24

1493.9

58

1549.6

92

25

1496.6

59

1550.3

93

26

1499.2

60

1550.9

94

27

1501.8

61

1551.5

95

28

1504.3

62

1552.0

96

29

1506.7

63

1552.5

97

30

1509.0

64

1553.0

98

31

1511.3

65

1553.4

99

32

1513.5

66

1553.7

33

1515.7

67

1554.0

Refer to the sound velocity of other fluids and materials, please contact the factory.

v(m/s) 1554.3 1554.5 1554.7 1554.9 1555.0 1555.0 1555.1 1555.1 1555.0 1554.9 1554.8 1554.6 1554.4 1554.2 1553.9 1553.6 1553.2 1552.8 1552.4 1552.0 1551.5 1551.0 1550.4 1549.8 1549.2 1548.5 1547.5 1547.1 1546.3 1545.6 1544.7 1543.9

56
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

Ultrasonic Heatmeter Operation Manual

10Selection Code

Tr a n s m i t t e r
Ultrasonic Flowmeter (Energy version) Wall mounted transmitter Flow Range: ±0.03m/s … ±5m/s Accuracy : ±1% measured value Repeatability: 0.2% measured value Display: 240*128 LCD backlit Power supply: 10-36VDC@1A max Transmitter housing: IP65, ABS (Temperature: -20grC…50grC) Output: Frequency/pulse output 0-10KHz, Relay, 4-20mA CommuniCation: RS232, Modbus Protocol

Code

Type Transmitter

1

Pulse/frequency, Relay, RS232/RS485, 4-20mA

2 Code

Pulse/frequency, Relay, RS232/RS485, 4-20mA / RTD Type Transmitter

CD01

Clamp-on, IP68. Temperature: -40grC … +80grC

C1U XXX 030 XXX

Clamp-on, IP68. Temperature: –40grC … +130grC Sensor cable length Standard length 9m Max. length 274m

Code

Temperature sensor

PT1000

Pt1000 temperature sensor

Code

Option

AC

AC power, 90 … 245 VAC

SD

SD card (datalogger)

Sample:
Ultrasonic Flowmeter (Energy version) w. Pulse, Relay, RS485, 4-20mA output, 2xPt1000 sensors, max 80grC, 9 m cable

Code: US201-2-CD01-030-PT1000

57
KLINGER Denmark A/S | Nyager 12-14 | DK-2605 Brøndby | +45 43 64 66 11 | salesinstrumentation@klinger.dk | www.klinger.dk

KLINGER Danmark A/S Nyager 12-14
DK-2605 Broendby Denmark
Phone +45 4364 6611
www.klinger.dk

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals