APERA LabSen 241-180 pH Electrode User Manual

June 3, 2024
APERA

LabSen® 241-180 pH Electrode
User Manual

LabSen 241-180 pH Electrode

Apera Instruments LabSen® pH electrodes are built with proprietary sensor technologies and key components from Switzerland. LabSen 241-180 Micro pH electrode is designed for testing the pH level of micro samples (50L) in small and deep containers such as NMR tubes.

Features

  • Proprietary LabSen® micro cylindrical sensor tip (180mm), designed for micro-volume pH measurement in slim and deep containers.
  • Blue gel inner solution – does not flow and will not cause air bubbles.
  • Long life reference system with better stability and longer service life.
  • The low membrane resistance increases response speed and measurement stability.

Technical Data Range

Range 0-14 pH Electrolyte 3M KCL
Temperature Range 32 to 212°F (0 to 100°C) Soaking Solution 3M KCL
Housing Material Lead-free glass Micro Sample Volume 50 pL
Reference Long Life Electrode Dimension Electrode (012×60) mm Measuring Tip

(cD3x 180) mm
Junction| Ceramic| Connector| BNC
Membrane Type| S| Cable| c1)3x1m
Membrane
Resistance (©25°C)| <450 MO| |

Usage and Maintenance

  1. The measuring tip of electrode is only 3mm. Please handle with care in case of breakage.
  2. Unplug the blue rubber plug before using to maintain the pressure of the reference solution, keeping consistent flow rate of reference solution and stable potentials of junction. After using, plug it back.
  3. Before measuring, unscrew the bottle cap, pull out the electrode and rinse it with deionized water. After using, put the electrode back into the storage bottle and screw tight the cap. When not in use, the electrode should be soaked in the storage bottle containing 3M KCL soaking solution to keep the glass membrane and junction in a healthy condition. Clean the bottle and replace the soaking solution if it gets contaminated. The electrode should never be stored in pure water such as deionized or distilled water.
  4. Perform at least a two-point calibration before measuring after connecting the new electrode to your pH meter.
  5. The reference solution will run low as you use the electrode. Whenever the solution level falls to 1/2 height of the electrode, add 3M KCL solution to the refilling hole using a syringe or pipette.
  6. After using, put the electrode back into the storage bottle, twist on the bottle cap, and plug in the refilling hole.
  7. The connector of the electrode should be kept clean and dry. If contaminated, please clean it with medical cotton and isopropyl alcohol and blow-dry it to prevent short circuit of the electrode or slow response of the electrode.
  8. pH electrodes don’t last forever. Every pH electrode will eventually age and fail even if you don’t use it that often. The typical service life of Apera pH electrodes is 12-24 months depending on the frequency of usage and how well you keep it clean and properly stored. We recommend replacing your electrode at least every 12-18 months to ensure the best accuracy.

Limited Warranty

We warrant this electrode to be free from defects in material and workmanship and agrees to repair or replace free of charge, at option of APERA INSTRUMENTS (Europe) GmbH, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS for a period of six months. Warranty period is the time limit to provide free service for the products purchased by customers, not the service life of the tester or electrodes.
This limited warranty does not cover any damages due to:
I. transportation;
II. storage;
III. improper use;
IV. failure to follow the product instructions or to perform any preventive maintenance;
V. modifications;
VI. combination or use with any products, materials, processes, systems or other matter not provided or authorized in writing by us;
VII. unauthorized repair;
VIII. normal wear and tear; or
IX. external causes such as accidents, abuse, or other actions or events beyond our reasonable control.

APERA INSTRUMENTS (Europe) GmbH
Wilhelm-Muthmann-Str. 18
42329 Wuppertal Germany
Tel.: +49 202 51988998
Email: [email protected]
www.aperainst.de

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

APERA User Manuals

Related Manuals