arroyo instruments 274 TECMount Water Cooled Mount User Manual
- June 3, 2024
- arroyo instruments
Table of Contents
arroyo instruments 274 TECMount Water Cooled Mount
Introduction
Thank you for choosing the 274 TECMount from Arroyo Instruments. The 274
TECMount is designed for high performance and long-term use. The 274 TECMount
integrates a high-power Peltier cooler for precise control and substantial
heating and cooling capacity for your powerful devices. The standard 274
TECMount has an operating range of -25°C to +85°C. Operation below the dew
point requires a nitrogen-purged environment. The 274 TECMount requires water
cooling for proper operation. For maximum performance, a chiller rated for at
least 400 Watts or greater is recommended, but smaller chillers can be used
for lower-power applications. The 5400-15-28 TECSource temperature controller
is a good match for the mount delivering maximum performance and complete
control across the 274 TECMount operating range.
The 274 TECMount also offers all the features you would expect from a modern
diode fixture, including:
- Hard nickel over the copper thermal plate.
- Designed to be quickly integrated with Arroyo’s TECSource instruments.
- Industry-standard D-sub connectors and pin-outs allow for quick integration into existing laser applications.
- External sensor input for easy injection of on-board device sensor as sensor feedback.
Installation and Use
Connecting to the TEC Controller
First, connect the 274 TECMount to your temperature controller. Make sure the
temperature controller’s current limit is set to a maximum value of no more
than 16 Amps. Where possible, we recommend the use of Arroyo Instruments TEC
cables. If you are using an Arroyo Instruments 5300 Series TECSource
temperature controller, as 1262B TECSource cable is the recommend interconnect
between the mount and controller. For the 5400 Series controllers, use the
1264-DB15 cable.
NOTE
Arroyo Instruments offers TEC cables designed to connect directly between
our TECSource products. If you use your own cables, ensure the connections are
properly made between the instrument and the mount, and that proper grounding
techniques are used. The pin-out of the connectors can be found later in this
document.
Connecting the water supply
The 274 comes with factory-installed brass barbs suitable for 3/8” I.D. hoses,
but they can be replaced with any 1/8” NPT male fitting. It is recommended
that when using the barbs, pipe clamps be used to secure the hose to the 274.
NOTE
Earth Grounding Considerations If your application requires that the cold
plate be earth grounded, you should directly wire the cold plate to earth-
ground.
Mounting Plate
The 274 TECMount has a breadboard configuration with M3 holes on 10mm centers.
Optional blank adapter plates are available in several sizes (3.2”, 4.2”, and
5.2” square) to allow for custom mounting patterns or larger mounting
surfaces. See the mechanical information below.
Connector Pin-Outs
Sensor Polarity and 4-Wire Connections
While the thermistor and RTD inputs are not polarized, when using a 4-wire RTD
connection from the mount, it is important to properly connect the polarity of
the sense wires to the sensor. Pins 7 and 14 should be one polarity (+) and
pins 8 and 15 should be the opposite polarity (–). If polarities are not
matched, the instrument will indicate a sensor error.
EXT Input and SENSOR Switch
The 274 features a SENSOR switch to quickly switch between internal and
external temperature sensors. With the SENSOR switch in the INT position, the
internal thermistor embedded in the cold plate is used to provide feedback for
the temperature controller. With the SENSOR switch in the EXT position, the
EXT+ and EXT- inputs on the Phoenix connector are connected to pins 7 & 14
Sensor+) and 8 & 15 (Sensor–). The 274 comes with one Phoenix 2-pin plug.
Additional plugs can be ordered online at
www.mouser.com or
www.digikey.com, Phoenix part number 1803578.
Technical Specifications
274 TECMount
COLD PLATE|
Mounting Holes| M3
TEMPERATURE CONTROL|
Standard Model (+85 ° C)|
Temperature Range (°C)| -25 to +85, non-condensing
Sensor Type| 10kΩ Thermistor
High Temp Model (+150 ° C)|
Temperature Range (°C)| -25 to +150, non-condensing
Sensor Type| 100Ω Platinum RTD
TE Module (at 25°C) 1| Imax = 16A
| Vmax = 33V
CONNECTORS / INTERFACES|
Temperature Controller| DB-15, male
External Sensor| Phoenix 2-Pin (plug p/n 1803578)
Coolant Inlet/Outlet| 1/8” Female NPT with pre-installed
| brass hose barbs (3/8” I.D. hose)
GENERAL|
Recommended Coolant| Distilled Water
Max Water Pressure (psi)| 80
Max Housing Temperature (°C)| 60
Size (H x W x D) [in(mm)]| 2.5 (63.5) x 4.8
(121.9) x 3.2 (81.3)
Weight (lbs [kg])| 2.8 [1.27]
Wetted Materials
For purposes of evaluating galvanic corrosion, the following materials will make contact with the coolant:
- Housing: Anodized AL 6061-T6
- Fittings: UNS C36000 Brass
- O-ring: EPDM
When using pure water as a coolant, distilled water is typically recommended unless your application dictates a different fluid.
Pressure Drop vs Flow Graph
The graph represents typical performance and will vary by flow rate and other conditions.
Configuring the Temperature Controller
When using an Arroyo Instruments temperature controller, the easiest method
for configuring the controller to operate with the mount is to change the
Mount setting in the menu by selecting 274 or 274-150, depending This will
change the sensor settings and current limit to be appropriate for this mount.
If you will be using a non-Arroyo controller, make sure to adjust the limits
and sensor settings appropriately to ensure proper and safe operation of the
mount.
Using the Thermistor on Standard Versions
The 274 LaserMount is equipped with a 10kΩ negative temperature coefficient
(NTC) thermistor, specifically, the BetaTHERM 10K3A1. A thermistor works by
translating temperature into resistance, with resistance decreasing as
temperature increases (hence the ‘negative coefficient’). Below is the
response curve of the thermistor:
As can be seen, by the graph, the resistance of the thermistor drops very quickly. In the typical control range (0°C to 40°C), typical 10K thermistors offer good sensitivity to changes in temperature, and this is the range in which most 10K thermistors are typically used. 10K thermistors can be used at much higher temperatures but will suffer poorer temperature stability performance because of the lower sensitivity. All Arroyo temperature controllers support operation using a 10μA or 100μA thermistor bias, which limits the upper control range to 450kΩ or 45kΩ, respectively. To minimize noise and maximize stability, you should select the highest current while still allowing you full operation across your required temperature range. The typical setting is 100μA, but your application will determine the actual needs.
The Steinhart-Hart Equation
As can be seen from the temperature versus resistance graph above, the
resistance varies inversely with temperature in a non-linear fashion. This
relationship can be accurately modeled by polynomial equations, one such being
the Steinhart-Hart equation: The coefficients for the BetaTHERM 10K3A1
thermistor are:
- A = 1.12924×10-3
- B = 2.34108×10-4
- C = 0.87755×10-7
These are the default coefficients for Arroyo Instruments temperature controllers but can be changed in the Sensor menu, or by selecting the appropriate 274 mounts from the Mount menu setting.
Using the RTD on 150°C Versions
The 274 TECMount can optionally be configured for up to 150°C operations. To support this high-temperature operation, an RTD sensor with a 0.00385 Ω / Ω / °C sensitivity is used. Like thermistors, RTDs also function by converting temperature into resistance, but unlike thermistors, RTDs increase in resistance as temperature increases. RTDs are also fairly linear devices, meaning they can be used across a much broader temperature control range. You can identify a 150°C configured 274 by its part number: a “-150” will be added to the end, for example, 274-150. According to IEC751, the resistance/temperature relationship is determined using one of two equations, dependent on the temperature or resistance value being measured. For resistances above the R0 value (resistance at 0°C, typically 100Ω, as is the case with the RTD used in the 274) of the RTD, the following equation is used:
- R = R + AT + BT
Below R0, an additional term is added to the equation: In both of these equations, R0 is the resistance of the RTD at 0°C, and A, B, and C are the coefficients as defined by IEC751, through regression analysis, or by using the Callendar-van Dusen method. Not all Arroyo Instruments temperature controllers support RTD operation. Check with the factory for the recommended controller. In most cases, the 5400-15-28 is the recommended controller for the 274, although lessor-powerful controllers may be used, depending on the thermal load capacity required. For the Arroyo Instruments controllers that support RTD sensors, the default coefficients are not correct for this mount. They must be changed to use the0.00385 Ω / Ω / °C curve, which has the following coefficients:
- A = 3.9080×10-3
- B = -0.58019×10-6
- C = -4.2735×10-12 R0 = 100
These coefficients can be changed in the Sensor menu, or by selecting the appropriate 274 mounts from the Mount menu setting.
2-Wire versus 4-Wire Measurements
One concern in using RTDs is their relatively low resistance (typically 100Ω
at 0°C), and small Ω/°C. Because of these two factors, the resistance of the
cable used to connect to the sensor can become a significant error in the
sensor measurement. Most Arroyo Instruments controllers offer two RTD
measurement modes: a conventional two-wire measurement mode, which is subject
to this error, and a four-wire measurement mode that uses separate sensor and
source lines to remotely sense the actual resistance of the RTD and eliminate
the cable or connector resistances. When using 4-wire measurement mode, you
must select ‘RTD (4-wire)’ as the sensor type, and then connect the Sensor+
and Remote Sensor+ at one side of the RTD, and Sensor– and Remote Sensor– to
the other side of the RTD. Make these connections as close to the sensor as
possible.
The drawings below illustrate how 2-wire and 4-wire connections work. Note
that 4-wire measurements require all four wires to be brought through the
cable to the mount. The 1262 TECSource cable carries this connection through
to the mount, but the 1260B cable does not.
Operating at High Temperatures
The 274-150 (150°C-capable version) has additional requirements that should be considered when operating in the upper-temperature range:
- The voltage requirements of the TEC increase significantly when operating at higher temperatures, so much so that the standard 5305 or 6300 Series controllers will be voltage limited when controlling the mount. To gain the maximum performance of the 274-150, the recommended controller the 5400-15-28 has a 15A / 28V output. Contact the factory for more details.
- If the mount is heating when at high temperatures, if possible, raise the water temperature. By lowering the temperature difference between the water temperature and the plate temperature, the TEC will not work as hard, provide better plate uniformity as well as not require as much power to operate the TEC.
Mechanical Drawings
274 TECMount User’s Manual · Page 15
Optional Adapter Plates
274 TECMount User’s Manual · Page 17
Warranty
Arroyo Instruments warrants this product to be free from defects in material
and workmanship under normal use and service for a period of one (1) year from
the date of shipment. It does not apply when the product has been misused,
altered or damaged by an accident or abnormal conditions of operation. If
found to be defective during the warranty period, the product will either be
repaired or replaced at Arroyo Instruments’s option.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR
PURPOSE. ARROYO INSTRUMENTS SHALL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, OR
CONSEQUENTIAL DAMAGES RESULTING FROM THE PURCHASE OR USE OF ITS PRODUCTS.
Service and Support
The 274 contains no user-serviceable parts. For service and support,
contact your local distributor or Arroyo Instruments.
- Telephone: +1 805-543-1302
- Facsimile: +1 805-543-1303
- Email: support@arroyoinstruments.com
- Web: http://www.arroyoinstruments.com
- Address: 1201 Prospect Street
- San Luis Obispo, CA 93401
- USA
References
- Arroyo Instruments - Laser Diodes and Temperature Controllers
- DigiKey - Electronic Components Distributor
- DigiKey - Electronic Components Distributor
- Electronic Components Distributor - Mouser Electronics Europe
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>