uv medico UV222 Filtered Far UV C Instruction Manual

June 3, 2024
uv medico

uv medico logouv medico logo 1 Filtered Far UV C
Instruction Manualuv medico UV222 Filtered Far UV
C

(Filtered Far UV-C)
Technical & Installation Manual

CAUTION UV LIGHT
THE PRODUCT AS DESCRIBED HEREIN CONTAINS COMPONENTS THAT EMIT FILTERED FAR UV LIGHT.
DO NOT USE THE PRODUCT BEFORE CAREFULLY READING THIS MANUAL.
ALWAYS KEEP THIS MANUAL WITH THE PRODUCT AND CONSULT IT REGULARLY.

RISK GROUP 3
WARNING: UV emitted from this product. Avoid eye and skin exposure to unshielded product.
CAUTION: Possibly hazardous optical radiation emitted from this product. Do not stare at operating lamp.

USING THIS MANUAL ­ GENERAL

This manual (hereinafter referred to as the “Manual”) is intended to be used as work of reference for professional, well trained and authorised users to be able to safely install and use the product mentioned on the cover of this document (hereinafter referred to as the “Product”). The observation of the Manual does not dispense the user from the observance of the local safety instructions and regulations other than provided by the Manual. Such safety instructions and regulations must be observed at all time.
For information about specific adjustments, maintenance or repair jobs which are not dealt with in this manual, please contact your Product supplier or UV Medico A/S.
If you suspect that the Product is damaged or detect a potential or actual danger when using the Product, immediately discontinue use of the Product and contact your supplier of the Product.
1.1. BASIC SAFETY INSTRUCTIONS

To minimise the risk of injury, the following basic safety precautions should be observed with each use of the Product:
Do not remove the sticker on the front of the unit until the UV222 is commissioned by a qualified installer.
Do not put any object in close front of the surface of the UV222. Do not put plastic or glass in front of the UV222. It does not transmit UV. Make sure the UV222 has an unobstructed path towards the area to be disinfected.
Do not use flammable substances near the lamp module. Do not use flammable or combustible gases or liquids, such as oxygen, hydrogen, or chemicals, near a lamp module. Failure to obey may result in fire or explosion.
If Front glass is broken, disconnect UV222 immediately and and contact an authorised installer. Do not turn on any UV222 with a broken front glass.
Do not disassemble or modify the lamp module. Disassembly or modification may result in fire, electric shock, or injury.
Turn off the electrical power during lamp module maintenance.
Do not remove any safety labelling or parts.
Do not touch the lamp module during operation. Otherwise, it may result in fire, electric shock, or injury.
Ground the device or metal parts of the housing that are close to the lamp module. Otherwise, it may result in fire, electric shock, or injury.
For those with photosensitivity, take measures similar to those for daily sunlight ultraviolet rays. Due to the absorption characteristics of ultraviolet rays in the skin, it is considered that the effect of 222nm light sensitivity on people exposed to ultraviolet rays will be less than when exposed to sunlight. For those with photosensitivity, take measures similar to those for daily sunlight ultraviolet rays, such as long sleeves and hats.
Do not use the unit for purposes other than disinfection.
The unit is designed for disinfection. Using it for any other purpose may result in unexpected accidents.

DISCLAIMERS

UV Medico A/S with registered office at Søren Frichs Vej 40E, 8230 Åbyhøj – Denmark, or any of its affiliates, does not accept any liability for material damage, including damage to the Product or personal injury caused by non- observance by the user of the safety instructions in the Manual or the applicable laws. As such UV Medico A/S shall among others not be liable for any damage or injury if the Product is used contrary to or for purposes other than its intended use, is installed, used, maintained or repaired negligently or contrary to the instructions mentioned in the Manual or on the Product; also in the case the Product is modified, and when safety devices and features are disregarded or overridden.

SAFETY INSTRUCTIONS AND WARNINGS

3.1. User manual
Every person using the Product must read, familiarise and understand the contents of the Manual before any use of the Product and must strictly observe the instructions set out herein. If you are an employee or are acting of a person, your employer/that person must instruct you in accordance with the Manual and ensure that all instructions set out herein are complied with. Please do not use the Product if that is not the case.
3.2. Intended use and installation
The Product is intended exclusively for air and surface disinfection. Using the Product for other purposes is contrary to its intended use and is done at own risk.
To ensure safe use of the Product, its installation, including all its safety features and devices, and its maintenance and repair must be carried out correctly by a qualified installer. All of the Product safety features and devices must be mounted and may not be removed but for the purpose of maintenance and repair.
The Product is not intended to be used by minors. Any use of the Product by minors, regardless of whether or not under the supervision of an adult, is at own risk.
3.3. Safety features of the Product
Each Product is equipped with the following safety features:

  • Motion sensor: To detect motion.
  • Lidar sensor: To measure distance to the occupancy of the room.
  • Multicolour LED to signal the operation and state of the unit.

It is strictly forbidden to override these safety devices or to use the Product if the safety features are not fully present or disfunction.
The safety features should be regularly checked for their proper functioning, and if required, be immediately repaired or replaced.
The pictograms, warnings and instructions attached to the Product are part of its safety features. They must not be covered or removed and must be present and legible during the entire use of the Product.
Immediately replace or repair missing, damaged or illegible pictograms, warnings and instructions.
3.4. Modification
No modification of the product is allowed. All warranty and liability are void if any modifications are done to the UV222.

GENERAL CHARACTERISTICS AND TECHNICAL SPECIFICATIONS

The Product is designed to emit Far UV-C radiation with a wavelength of 222 nm, in order to eliminate viruses, bacteria, fungi and other pathogens, from air and surfaces.
The Far UV-C lamps are designed for 5000 hours of efficient operation. Far UV-C lamps used longer than this, may no longer emit the required level of ultraviolet radiation to ensure the desired germicidal efficacy on harmful microorganisms.
Technical data

Light Source Kr-Cl Excimer lamp
Wavelength 222 nm
UV output power 130 mW (typical)
Input voltage 85-305V AC 50/60HZ
Max power consumption 17 W
Weight 1.9 KG
Dimensions 303.45 x 120 x 75.25
Power lead 3 x 0.75mm² -18AWG
Operating temperature 0 to +50°C

CONSTRUCTION

INSTALLATION OF THE PRODUCT

Perform the following steps for the installation of the Product.
Outside every room where Far UV-C light is used, there must be information signs in clear line of sight.
Installation must be done according to the local regulations by a qualified installer.
6.1. Choose the appropriate location
Choosing the appropriate location is important for the proper functioning and safe use of the Product.
A dialux simulation can help you to choose the right position for disinfection.
Install the Product in a protected area where the temperature does not fall below 0° C and the humidity level is low. The Product functions ideally in a temperature range from 0°C to 50° C.
Select a Product mounting location that is close to surface/area that you want to disinfect.
6.2. Unboxing and inspection
Read and save all notices, warnings and safety instructions received with the Product.
Open the box and carefully take the unit out of the box. Take the unit out of the plastic bag. Do not remove the protection label on the front of the UV222. The protection label is intended to avoid unexpected UV radiation. Make a visual inspection of the unit.
Do not power on the unit until installed by a qualified installer!
CAUTION: Do not touch the glass of the lamp without gloves. Damage to the lamp may result. Oil from fingerprints will permanently etch the glass of the lamp and weaken the structure.
6.3. UV222 fixture installation
The unit must be attached to a surface by means of at least 4 mounting pins, preferably near electrical wires with grounding. The mounting pins and surface must be strong enough to hold 4 times the weight of the unit.
Take out the 2x DIN 912 M4 on the bottom sides to take off the installation plate of the unit.
The installation plate can be now installed to the surface with 4 mm screw or M4 bolts:
6.4. Wiring the Product to power source
Far UV-C lamps must be operated continuously from a separate, dedicated, 85-305 V 50/60Hz power source.

Standard wiring:| Phase/ Hot +
Neutral/Cold
­ Earth| (brown)(black)
(blue) (white)
(green/yellow) (green)
---|---|---

How to wire the UV222

  1. Take of the installation plate (foot plate) A 2) Unscrew 3x din965 with T15 screwdriver B
  2. Take of the service plate
  3. Unscrew 2x ph1 (cable blockage)
  4. Unscrew 2x ph1 (cable connecter)
  5. Unscrew nut for earth connection
  6. Rewire the connecter correctly
  7. Put cable trough service plate and screw back the cable blockage
  8. Close the backplate with 3x din965
  9. Put the unit back on its installation plate (foot)

6.5. Understanding the unit’s operation:
The maximum permissible UV-C exposure shall not exceed the ACGIH TLV and NIOSH REL of 22,0 mJ/cm² for an 8 h day, 40 h work week exposure to UV radiation at 222 nm. (UV Medico A/S expects positive changes into the ACGIH that will result in a higher exposure than allowed at this moment. If this changes, UV Medico A/S will update the software of the unit to align with the new threshold of exposure).
To achieve the correct radiation within these limits, following information will be programmed into the UV222 before the start-up of this installation.
6.6. Setting up the UV222:
The installer will need a Microsoft Windows computer with the UVMEDICO application installed. You can download the application on
www.uvmedico.com/download-centre. Download the latest firmware to update the UV222. To launch the app, go to the installed folder and open the file:
UV Medico Configuration App.exe
The app includes two different set up modes: simple and advanced. For the simple mode, follow the instructions that will appear on the screen. For the advanced mode, see below:
There is a USB-C connection D on the side of the unit to connect to a computer with a USB cable.
Connect your computer to the UV222 unit and it will show up automatically on the application. When the app is launched, following steps needs to be taken:

  1. If the UV222 was already installed, you can read out the setting of the unit by pressing “Get” all settings will be shown in the application. A

  2. Check if software version is the latest version. If not, press “Update SW” button B and confirm the pop up by pressing “Yes”. C
    Select the correct software version D and press “Open” to upload the software. When finished installing, the software should be up to date.

  3. Set the correct hour by pressing “Set Device Time”. A The date and hour are now the same as your computer.

  4. Define the timing of the use (duty cycle) of the room where the UV222 is installed: “On time”: Time from which the room will be occupied by people. B “Off time”: Time from which the room will not be occupied. C

  5. Add the pre and post cleaning cycle. D The unit will/can light up before and after the duty cycle so you can ensure a “disinfected room” when people enter or leave.D

  6. Add the “Maximum Occupant Time”: how long the room is occupied during the duty cycle. E

  7. Add 3 distances:
    a. “Distance to eye height”: Closest distance of the unit to the eye of a human (even with people walking around). A
    b. “Safety distance”: Closest distance people are allowed to get to the lamp before it turns itself off. The unit will turn itself back on after 10 sec. when/if the safety distance is regained. B
    c. “Primary surface distance”: Distance from the unit to the main surface to disinfect. C
    All measurements are in centimetres.

  8. Change the intensity of the LED signal to the desired output. D

  9. Save setup into template by pressing “Save Setup”. E Name the template. This can be recalled whenever you have similar settings.

  10. Press “Set” to upload the settings into the UV222. A

Once this information is saved and programmed into the unit, it is ready to test/start up. The unit will calculate the radiation automatically to stay within the limits of the ACGIH like the example below:
Time If during the duty cycle a person comes closer to the unit than the programmed distance from the eye, UV222 will automatically stop, and it will resume the activity 10 sec. after this security distance is cleared.
If movement in the room is detected during the pre or post disinfection cycle, the unit will stop working, and will resume its activity after no more movement is detected.
If the information programmed into the UV222 when installed is different than the use/ occupancy of the room, UV Medico A/S cannot be held liable for any injuries or misuse of these units.
If the room changes of use, occupancy or any change that has an influence on the risks of exposure, the installer needs to be contacted and unit needs to be reprogrammed.
11) Testing the Product. Right click on the photo of the UV222 B and a pop up of the test mode will show up. Here you can quickly light up the UV222 and test the signal LED.
When you press “On” C at 222 LightSource section, it will turn on the UV222 with no safety! This is only to test the source, (the box “Enable” will be automatically ticked)
When you go to LED Mode you have 3 possibilies to choose:
“Normal Mode” Work as UV222 is programmed
“Led Off” Turns the LED off
“Led RGBW Off” LED test cycle
When you untick the “Enable” box it goes back to normal configuration/use.

6.7. Understanding the function of the signal LED

LED ACTION
No light No power
Red continuous Error: no feedback from UV222 light source
Cyan continuous Pre disinfection / after disinfection light source on
Magenta continuous Light source is on
Blue continuous In-cycle UV off
Green continuous Out of cycle
Flashing red Object to close
Flashing cyan Pre disinfection / after disinfection per seen movement

6.8. Replacements
The UV222 has an internal battery and fuse that can be replaced if broken/empty.
Take the unit off the installation plate and unscrew the 3x din965. Take off the cover and you will see the fuse (T250V 500mA) and battery (3V CR2032). Replace only with correct fuse and battery to prevent damage.

PRODUCT FAILURE

FAILURE TO DO
Unit does not light up Check incoming power.

Check cable damage.
Replace fuse.
Check if UV222 is correctly installed.
Unit flickers| Check if UV222 is correctly installed.
If not, contact your local distributor.
White flash| Power cycle the unit.
Does not hold settings| Replace battery.

If the failure is unsolved or recurring, please return the unit to an official UV Medico dealer/ service centre for repair or replacement (see point 10. REPAIRS).

CLEANING AND MAINTENANCE

Clean the lamps in time.
Before performing any maintenance always unplug the Product and disconnect the power supply.
Clean the external surfaces with a soft cloth and non-aggressive cleaning agents only.
CAUTION: Turn off and lock out ALL unit power before performing any cleaning. Do not wash the Product with water under pressure or by immersion.

REPLACEMENT OF UV-C LAMP

UV Medico A/S recommends replacing the Far UV lamp when LED signals red-blue- green (loop).
Contact your supplier/installer replacing the UV lamp. No user-serviceable parts are in the unit.

REPAIRS

If the Product is damaged, malfunctions, or for other specific inquiries, please contact your authorised UV Medico distributor.

Warranty

2-year warranty is given from invoice date regardless on the hours the unit is used.
UV222TM and its logo are registered trademarks of UV Medico A/S. All rights reserved.
No part of this manual may be copied or published by means of printing, photocopying, microfilm or otherwise without the prior written consent of the manufacturer. This restriction also applies to the corresponding drawings.
The information given in this manual has been collected for the general convenience of our customers. It has been based on general data pertaining to construction material properties and working methods known to us at the time of issue of the manual and is therefore subject at any time to change or amendment, and the right to change or amend is hereby expressly reserved. The instructions in this manual only serve as a guideline for installation, use, maintenance and repair of the Product mentioned on the cover page of this document. This manual is to be used for the standard model of the Product of the type given on the cover page. Thus, the manufacturer cannot be held responsible for any damage resulting from the application of this manual to any other product.
This manual has been written with great care. However, the manufacturer cannot be held responsible, either for any errors occurring in this manual or their consequences.
NOTES:

UV Medico A/S
Søren Frichs
Vej 40E 8230 Åbyhøj Denmark
+45 20 90 71 30
info@uvmedico.com
www.uvmedico.com
Version 02/2021