icare TONOVET Pro Premium Tonometer User Guide

September 26, 2024
iCare

QUICK GUIDE
NO ANESTHESIA
NO CALIBRATION

TONOVET Pro Premium Tonometer

TURNING THE TONOMETER ON AND LOADING THE PROBE

  1. Press the Select or Measure button to turn the tonometer on. Illustrations of these alternative ways of starting. Remember to use the wrist strap. Pressing the Select button Pressing the Measure button
  2. Open the probe tube by removing the cap and insert the probe into the probe base as shown in the image.
  3. When the Default measure text appears on the display press the Measure button to start Default measurement or press left to select Quick Measure mode.
  4. Press the Measure button to start the measurement.

Changing the setting for different species.
To change the setting for different species press the select button to access the menu. Press Navigation buttons and select SETTING by pressing the Select button. Use the Navigation buttons to toggle the settings. To select a species, press the Select button. To start measuring, use the Navigation buttons to return to MEASURE.

MEASUREMENT

  1. Bring the tonometer near the patient’s eye.
    NOTE! Avoid excessive restraining, as it may alter the IOP. Patient’s head should be held as lightly as possible; be careful not to put pressure on the neck or the eye ball. If a collar is worn, make sure it is not too tight or remove it for measurement.

  2. To perform a successful measurement, the distance from the tip of the probe to the patient’s cornea (see pictures) should be about 4-8 mm (about 5/32 – 5/16 in.). Bring the tonometer in front of the patient’s eye with the probe pointing to the center of the cornea. The probe base light must be green (indicating correct measuring position.)

  3. Default measure. The probe will contact the eye during measurements: press and hold the Measure button to collect 6 measurements in Series mode or press the button briefly 6 times to collect 1 measurement at a time in Single mode. The grey segments of the circle on the display screen will turn green one by one.

  4. Quick measure. The probe will contact the eye during measurements: press and hold the Measure button to collect 2-3 measurements in Series mode or press the button briefly 2-3 times to collect 1 measurement at a time in Single mode. The grey segments of the circle on the display screen will turn magenta one by one.

DISPLAY DURING MEASUREMENT

DEFAULT MEASURE

|
---|---
Before first measurement| After the second measurement
|
After the sixth measurement| After the sixth measurement but with acceptable variation*
|
After the sixth measurement but with too much variation| Repeat the measurement

QUICK MEASURE

|
---|---
Before first measurement| After the second measurement
|
After the second or third measurement| After the third measurement but with acceptable variation*

  • Measurement should be repeated if the result is over 25 mmHg
    ** Measurement should be repeated (NOTE in very high IOPs, the fluctuation may be higgher than normal)


ONE SHORT BEEP

Indicates a correct single measurement.
DOUBLE BEEP
Indicates an incorrect single measurement. (See error messages from instruction manual).
LONG BEEP
Indicates that six succesful measurements are done and IOP result is available on the display.

PARTS OF THE TONOMETER

  1. Measuring support
  2. Probe base
  3. Probe base collar
  4. Adjustment wheel for measuring support
  5. Display screen
  6. User interface
  7. Navigation buttons
  8. Select button
  9. Measure button

REPLACING OR CLEANING THE PROBE BASE

During the use some dirt may collect in the probe base, affecting the probe movement. Replace the probe base every 12 months. Clean the probe base every 3 months. The cleanest probe base is a new probe base.

  1. Turn power off from tonometer.
  2. Unscrew the probe base collar and remove the probe base from the tonometer.
  3. Insert the probe base into the cleaning container filled with 70-100% isopropyl alcohol and let soak for 5-30 minutes.
  4. Remove and dry the probe base by blowing some clean canned or compressed air into it.
  5. Reinsert the completely dry probe base and screw back the probe base collar. Alternatively, insert new probe base.

Please refer to the instruction manual for more detailed instructions of changing and cleaning the probe base.

Visit the TONOVET YouTube Channel for IOP measuring and probe base cleaning videos.
youtube.com/@tonovettonometer4327

Refer to the instruction manual
for additional information.