PROTEUS SDI-12 Water Quality Multiprobe Instruction Manual
- September 9, 2024
- Proteus
Table of Contents
PROTEUS SDI-12 Water Quality Multiprobe
FAQs
Q: How do I change the MODBUS Baud Rate?
A: Use the MODBUS Control Register Mapping to change the Baud Rate. The available indices are 0 to 4 corresponding to different rates.
Q: What is the default address for the MIB?
A: The default address for the MIB is 1. If unknown, the MIB will respond to MODBUS address 0.
Introduction
In March of 2020, Proteus introduced integrated SDI-12 and RS-485 MODBUS data
outputs for the Proteus water-quality multiprobe product line. This optional
feature replaces the external converter units previously required for those
output formats.
The new feature is called “Multi-Protocol Interface Board”, or MIB. One
adapter cable provides MODBUS output, and a different adapter cable provides
SDI-12 output. The MIB is normally built into the Proteus at the time of the
unit’s manufacture; it cannot be seen from outside the instrument and does not
alter the Proteus’ size or appearance. The photo below shows a Proteus,
Underwater Cable, and at the top of the photo, the short SDI-12 Adapter Cable.
The SDI-12 master is attached to the three bare wires on one end of the
Adapter Cable. The MODBUS Adapter Cable works the same way. If you wish to
retrofit a Proteus with the MIB option, please contact us. If you wish to
build your own MODBUS or SDI-12 converter cable, or convert your Underwater
Cable into a converter cable, use the wiring diagrams in Appendices One and
Two.
Note that Proteus Data Cables (the short cables used for tasks such as calibration) will communicate with a PC or laptop, but do not support MODBUS or SDI-12 operation.
MIB-equipped Proteus can communicate with a PC and other RS-232 devices as usual (this is the “transparent” mode). Most MIB-equipped Proteus’ can be operated with USB power as usual. However, some of the larger Proteus’ and/ or longer cables (>20m cable or P35/P40 with >250mA consumption – Use Battery life Calculator to check) may require the USB Converter’s 12V adapter to communicate with that Proteus. The photo at right shows a power adaptor that provides 12 volts to the USB Adapter that connects an Underwater Cable or Data Cable to the USB port on a PC or laptop. You can use a 12-volt battery as well.
Note also that early Proteus Data Cables will work only in transparent mode; the Modbus and SDI-12 features work only with an Underwater Cable.
MODUS Operation
-
a) How It Works
For MODBUS communication, simply connect the MODBUS Adapter Cable to the nine- pin connector on the Underwater Cable attached to your MIB-equipped Proteus. The Adapter Cable gives you the wires you need to connect the Proteus to a MODBUS device and is wired to inform the Proteus that you wish to communicate in the MODBUS format instead of the usual RS-232.
Note that the MODBUS interface can use either half-duplex differential RS-485 or full-duplex RS-232 (separate transmit and receive). -
b) The MODBUS Technical Details
An MIB-equipped Proteus uses MODBUS protocol over RS-485 or RS-232 to read the parameters processed by the unit. The upstream communication functions either as a full-duplex RS-232 standard interface, or as a half-duplex, RS-485 standard interface. The data format is 8-bits with no parity, one stop bit. The baud rate is 19,200. The MODBUS interface provides measurement values, beginning at holding register 40001 (see Table 1), for all enabled Proteus parameters, with each measurement value occupying 2 MODBUS registers. Values are formatted in IEEE-754 32-bit floating-point representation.
Table 1: Mapping MODBUS Parameter Measurement Values
MODBUS Holding Register| Bus Address| Read Value| Format
40001| 0| Parameter 1 MSW| IEEE32
40002| 1| Parameter 1 MSW|
40003| 2| Parameter 2 MSW| IEEE32
40004| 3| Parameter 2 MSW|
..| ..| |
40035| 34| Parameter 18 MSW| IEEE32
40036| 35| Parameter 18 MSW|
In RS-485 operation, two communication lines to the MIB are used for the differential lines Data+ and Data- (see Appendix 1). In RS-232 operation, the Data+ line is connected to the Proteus Rx line, and the Data- line is connected to the Proteus Tx line. A negative voltage on the Proteus Rx signals the MIB that the Tx line from an RS-232 host is connected so that the MIB operates in MODBUS/RS-232 format; otherwise, RS-485 format is assumed. The data format is 8 bits with no parity and one stop bit. The MIB normally operates at 19,200 baud. If you wish to change that rate, please see Table 2.
Table 2: MODBUS Baud Rate Indicies
Index| BaudRate
0| 9600
1| 19200 (default)
2| 38400
3| 57600
4| 115200
The MIB provides store/update read-only registers and read/write settings for communicating with common SCADA systems, PLC interfaces, or other data collection platforms. A built-in MODBUS map provides aggregated sensor readings and other equipment information. The MODBUS interface provides measurement values, beginning at holding register 40001 (see Table 3), for all enabled Proteus parameters, with each measurement value occupying two MODBUS registers. Values are formatted in IEEE-754 32-bit floating-point representation. The MIB’s address is register-programmable (the default value is 1). The MIB will always respond to MODBUS address 0 if you don’t know the real address.
Table 3: MODBUS Control Register Mapping
MODBUS Holding Register| Bus Address| Read/Write Value| Format
40201| 200| Baud Rate – Upstream| Fixed at 19, 200 baud
40202| 201| MODBUS Device Address| Integer 1-250 Default=1
40203| 202| Baud Rate- Downstream| Integer index, 0-4
40204| 203| SDI-12 Address| Integer 0-9, A-Z, a-z
40205| 204| Power Switch Delay| Integer 0-60
40206| 205| Proteus Wipe Interval| Integer 0-1440 (minutes)
40207| 206| Proteus Wipe Freeze Time| Integer 0-60 (seconds)
C) MIB Commands for MODBUS
A MIB-equipped Proteus can be connected to a host PC or laptop to send
commands directly to the Proteus CPU, as well as special commands (see Table
4) to the MIB itself. This mode of communication – using the Proteus’ normal
RS-232 output and not MODBUS – is called the “transparent mode”. When a
terminal emulator, such as TeraTerm or Hyperterminal, is used to talk to the
Proteus in this transparent mode, the MIB recognizes and responds to certain
ASCII commands to allow the pro-gramming/verifying of some parameters, as
shown below. The format of these MIB command is “$ccxxx
- ‘$’ indicates a MIB command
- cc is a two-character MIB command identifier
- xxx is a parameter values specific to the command
Table 4: Special MIB Commands
Command| Description| Parameters| Response
$AMxxx
Address
| xxx; 001 to 250| OK
$AM?
$WPxxxx
default= 0
| OK
$WP?
$WFxx
$WF?
$FV?
d) MODBUS Automatic Wiper Operation
Some Proteus models include a sensor-cleaning wiper built into the turbidity sensor. The wiper clears debris, foulants, and bubbles from the sensors’ active faces when the Proteus is first powered-up, and when a WIPE command is sent to the Proteus. If your Proteus is continuously powered during MODBUS operation, you may wish to periodically initiate wipe cycles using MIB commands (see Table 4). The Wipe Interval is the number of minutes between wipe cycles. Note that setting the Wipe Interval to 0 disables automatic wiping. Some parameter values are invalid during normal wiper cycles because of the movement of the wiper. When the cycle ends, the data resume their real- time format. But if your MODBUS controller may create an alarm on account of the invalid data during the wipe cycle, you can use MIB WIPE commands (see Table 4) to “freeze” all sensor data while the wiper is cycling. That means that all data coming from the Proteus during the wiper cycle is the same data sent in the last data transmission before the wipe cycle started, i.e. the readings stay the same during the wiper cycle. This programmable freeze time sets the number of seconds (default is 15 seconds) that the data are frozen after the Proteus is given a WIPE command. The data resume their real-time format after that number of seconds has passed.
Using the MIB for SDI 12 Communication
a) How It Works
- For SDI-12 communication, simply connect the SDI-12 Adapter Cable to the nine-pin connector on the Data Cable or Underwater Cable attached to your MIB-equipped Proteus. The Adapter Cable gives you the wires you need to connect the Proteus to an SDI-12 device, and is wired to inform the Proteus that you wish to communicate in the SDI- 12 format instead of the usual RS-232 format (i.e. transparent mode). Appendix One shows the wire assignments by color.
b) The SDI-12 Technical Details
- The host computer-to-Proteus communication adheres to the requirements of the SDI-12 Support Group, Version 1.3. Table 5 summarizes the implemented SDI-12 commands. If you are not familiar with the SDI-12 Protocol, the SDI-12 Support Group website (www.sdi-12.org) provides more detail.
Table 5: MIB SDI-12 Commands (a = SDI-12 address)
a!| Empty Command|
aA!| Change address|
aC!| Request a concurrent measurement| returns up to 20 values
aM!| Request a measurement| returns up to 9 values
aM1!| Request an additional measurement| returns up to 9 additional values
aM2!| Request an additional measurement| returns up to 2 additional values
aCC!| Request a concurrent measurement with CRC|
aMC!| Request a measurement with CRC|
aDn!| Read measurement results data| n=0..2
aI!| Request device identification string|
c) Special MIB Commands for SDI-12
A MIB-equipped Proteus can be connected to a host PC or laptop to send
commands directly to the Proteus CPU, as MIB-equipped Proteus can be connected
to a host PC or laptop to send commands directly to the Proteus CPU, as well
as special commands to the MIB itself. Using the Proteus’ normal RS-232 output
and not SDI-12 output is called the “transparent mode”. When a terminal
emulator, such as TeraTerm or Hyperterminal, is used to talk to the Proteus in
this transparent mode, the MIB recognizes and responds to certain ASCII
commands (see Table 6) to allow the programming/verifying some parameters, as
shown below. The format of these MIB command is “$ccxxx
- ‘$’ indicates an MIB command
- cc is a two-character MIB command identifier
- xxx is a parameter value specific to the command
Table 6: MIB Transparent-Mode Commands
Command| Description| Parameter(s)| Response
$ASx
Address
| x= 0-9, A-Z, a-z; default= 0| OK
$AS?
$PDxx
$PD?
$FV?
Table 7 shows example SDI-12 commands and responses for a Proteus for which 10 parameters have been selected for SDI-12 monitoring.
Table 7: Sample SDI-12 Commands and Responses for a Proteus with 10 Parameters Selected
Command| Response
0!| 0
0I!| 013 PROTEUS 711SN10162469
0V!| 00000
0M!| 00169
0D0!| 0+0+408.6999+4938.999+489.3999
0D1!| 0+4494.399+132.6000+3651.699+131.2000
0D2!| 0+2269.900
0M1!| 00031
0D0!| 0+11.70000
0C!| 000310
0D0!| 0+0+1.800000+2.100000+489.6999
0D1!| 0+4523.299+133.1000+3591.099+132.2000
0D2!| 0+2243.600+11.72000
0MC!| 00039
0D0!| 0+0+1.900000+2.100000+488.999AD
0D1!| 0+4538.699+133.0000+3557.699+132.4000@Zy
0D2!| 0+2224.000NWS
0MC1!| 00031
0D0!| 0+11.68000BS_
0CC!| 000310
0D0!| 0+0+1.900000+2.000000+489.0999EHG
0D1!| 0+4546.699+133.100.3540.199+132.6000O]X
0D2!| 0+2214.500+11.70000CSh
Appendix 1 – MODBUS and SDI-12 Adapter Cable Wiring Assignments
Appendix Two – Making Your Own MODBUS and SDI-12 Adapter Cables
CONTACT INFORMATION
- Proteus Instruments Ltd, Canalside, Harris Business Park, Hanbury Road, Stoke Prior, Bromsgrove, B60 4DJ,
- United Kingdom www.proteus-instruments.com