humimeter FS1 Grain Moisture Meter Instruction Manual
- September 6, 2024
- humimeter
Table of Contents
FS1 Grain Moisture Meter
Product Information
Specifications:
- Product: Moisture Meter
- Model: humimeter FS1 & FS1.1
- Version: 3.4
Product Usage Instructions
Main Unit:
The main unit consists of the following components:
- Auxiliary plate for filling
- Measuring chamber
- Display
- Keypad
Rear of the Main Unit:
The rear of the main unit features the battery compartment for
easy access.
The Display:
The display provides information on:
- Calibration curve
- Moisture content in %
- Display symbols
- Currently applied offset
- Temperature display
The Menus:
The device has three main menus:
- Product selection menu
- Offset menu
- Main menu
Product Selection Menu:
- Setting the offset
- Display illumination / device on/off
- For changing the calibration curve
Main Menu:
The main menu includes various menu items for device settings
and operations.
Frequently Asked Questions (FAQ)
Q: How do I change the calibration curve on the humimeter FS1
& FS1.1?
A: To change the calibration curve, access the product selection
menu and select the option for changing the calibration curve.
Q: What is the proper way to set the offset on the moisture
meter?
A: To set the offset correctly, navigate to the offset menu and
follow the on-screen instructions to adjust the offset value as
needed.
Moisture meter
Operating Manual humimeter FS1 & FS1.1
Grain moisture meter
78,0°F | 6,16% | 456kg/m3 | -27,3td | 0,64aw | 51,9%r.H. | 14,8%abs | 100,4g/m2 | 09m/s | 4,90Ugl | 1
Version 3.4
Always read this operating manual before using the humimeter FS1 & FS1.1!
humimeter FS1 & FS1.1 Operating Manual
Your humimeter FS1 & FS1.1 at a glance
The main unit 1 2 3
4
No. Name 1 Auxiliary plate for filling 2 Measuring chamber 3 Display 4 Keypad
Page 2
Rear of the main unit 1
No. Name 1 Battery compartment
Page 3
humimeter FS1 & FS1.1 Operating Manual
The display
1
5
4
2
3
No. Name
1 Calibration curve 2 Moisture content in % (“6.1 How moisture
content is defined”) 3 Display symbols 4 Currently applied offset 5
Temperature display
The display symbols
Symbol Name Enter Up Down Back Enter numbers Enter letters Continue / go right Left
Symbol Name
Yes
No
Change input Level OK
Change menu
On/off button, display light Setting the offset
Page 4
The menus
The device has three different menus: product selection menu, offset menu and
main menu
Product selection menu
1
2
3
No. Name 1 Setting the offset 2 Display illumination / device on/off
3 For changing the calibration curve
Main menu
The main menu comprises the following menu items:
· Options: Language, Unlock, °C/°F, BL On Time, Auto Off Time,
Materialcalibration, Password, Reset
· Status
Accessing the main menu: “4.3 Accessing the main menu” page 14
Page 5
humimeter FS1 & FS1.1 Operating Manual
Table of contents
Your humimeter FS1 & FS1.1 at a glance ……………………………… 2
The main unit ……………………………………………………………………………………………. 2
Rear of the main unit ………………………………………………………………………………….. 3
The display ………………………………………………………………………………………………… 4
The display symbols ……………………………………………………………………………………. 4
The menus ………………………………………………………………………………………………… 5
Introduction ………………………………………………………….. 9
1.1
Information about this operating manual ………………………………………… 9
1.2
Limitation of liability …………………………………………………………………….. 9
1.3
Symbols used in this manual ……………………………………………………….. 10
1.4
Customer service ………………………………………………………………………. 10
For your safety …………………………………………………….. 11
2.1
Proper use ……………………………………………………………………………….. 11
2.2
Improper use ……………………………………………………………………………. 11
2.3
User qualifications ……………………………………………………………………… 11
2.4
General safety information ………………………………………………………….. 12
2.5
Warranty …………………………………………………………………………………. 12
On receipt of your device …………………………………….. 12
3.1
Taking the device out of its packaging …………………………………………… 12
3.2
Making sure that all of the components have been included …………….. 13
3.2.1 FS1’s scope of supply …………………………………………………………………. 13
3.2.2 FS1.1’s scope of supply ………………………………………………………………. 13
3.3
Inserting batteries ……………………………………………………………………… 13
Using the device – Basics ………………………………………. 14
4.1
Switching the device on ……………………………………………………………… 14
4.2
Automatic Adjustment ……………………………………………………………….. 14
4.3
Accessing the main menu …………………………………………………………… 14
Page 6
4.4 4.5 4.6
5.
5.1 5.2 5.3 5.3.1
6.
6.1
7. 8.
8.1 8.2 8.3 8.4 8.5 8.5.1 8.5.2 8.6 8.7 8.8
9.
9.1 9.2 9.3
10. 11.
11.1
Selecting the calibration curve ……………………………………………………… 15 Taking a measurement
……………………………………………………………….. 15 Switching the device off
……………………………………………………………… 15
The measuring process …………………………………………. 16
Preparing a measurement …………………………………………………………… 16 Taking a measurement
……………………………………………………………….. 17 Offset function – Device adaptation
………………………………………………. 19 Setting the offset ………………………………………………………………………. 19
Calibration curves ………………………………………………… 20
How moisture content is defined …………………………………………………. 21
Checking the device’s status …………………………………. 22 Configuring the device
…………………………………………. 23
Selecting a language ………………………………………………………………….. 23 Activating options
……………………………………………………………………… 23 Deactivating options
………………………………………………………………….. 24 Selecting °C/°F …………………………………………………………………………..
24 Reducing the device’s power consumption …………………………………….. 25 Configuring the
display illumination time ………………………………………. 25 Configuring automatic switch-off
………………………………………………… 25 Configuring the material calibration function
…………………………………. 25 Changing the password ……………………………………………………………… 26 Resetting
the device to its factory settings ……………………………………… 27
Cleaning and maintenance …………………………………… 27
Changing batteries ……………………………………………………………………. 27 Care instructions
……………………………………………………………………….. 28 Cleaning the device
…………………………………………………………………… 28
Faults ………………………………………………………………….. 29 Storage and disposal …………………………………………….
30
Storing the device ……………………………………………………………………… 30
Page 7
humimeter FS1 & FS1.1 Operating Manual
11.2
Disposing of the device ………………………………………………………………. 30
Device information ………………………………………………. 31
12.1
CE declaration of conformit ………………………………………………………… 31
12.2
Technical data …………………………………………………………………………… 35
Page 8
1. Introduction
1.1 Information about this operating manual
This operating manual is designed to enable you to use the humimeter FS1 &
FS1.1 safely and effectively. It is part of the device, has to be stored
nearby and must be easily accessible to users at all times. All users are
required to carefully read and make sure that they have understood this
operating manual before using the humimeter FS1 & FS1.1. All of the safety and
operating instructions detailed in this manual have to be observed to ensure
the safety of the device.
1.2 Limitation of liability
All of the information and instructions provided in this operating manual have
been compiled on the basis of the current standards and regulations, the state
of the art, and the extensive expertise and experience of Schaller Messtechnik
GmbH. Schaller Messtechnik GmbH does not accept any liability for damage
associated with the following, which also voids the warranty:
· Non-observance of this operating manual · Improper use · Inadequately
qualified users · Unauthorised modifications · Technical changes · Use of
unapproved spare parts This fast measuring procedure can be affected by a
range of different factors. For this reason, we recommend periodically
checking the device’s measurements with a standardised oven-drying method. We,
as the manufacturer, do not accept any liability for any incorrect
measurements and associated consequential damage.
Page 9
humimeter FS1 & FS1.1 Operating Manual
1.3 Symbols used in this manual
All of the safety information provided in this manual is shown with a
corresponding symbol.
ATTENTION
It is essential to observe this warning. Non-compliance can lead to damage to
property or equipment.
i Information
This symbol indicates important information that enables users to use the
device more efficiently and cost-effectively.
1.4 Customer service
For technical advice, please contact our customer service department at
Schaller Messtechnik GmbH Max-Schaller-Straße 99 A – 8181 St.Ruprecht an der
Raab Telephone: +43 (0)3178 28899 Fax: +43 (0)3178 28899 – 901 E-mail:
info@humimeter.com Internet: www.humimeter.com © Schaller Messtechnik GmbH
2024
Page 10
2. For your safety
The device complies with the following European directives: · Restriction of
Hazardous Substances in Electrical and Electronic Equipment (RoHS) ·
Electromagnetic compatibility (EMC)
The device corresponds to state-of-the-art technology. However, it is still
associated with a number of residual hazards. These hazards can be avoided
through strict observance of our safety information.
2.1 Proper use
· Easy to use device for quickly measuring the moisture content of grain. ·
The device must only be used for taking measurements on the products defined
in the following sections of this manual (see “6. Calibration curves”es”).
2.2 Improper use
· The device is not suitable for measuring mouldy material. · The device is
not waterproof and must be protected from water and fine dust.
2.3 User qualifications
The device must only be operated by people who can be expected to reliably
take the measurements. The device must not be operated by people whose
reaction times may be slowed due to, e.g. the use of drugs, alcohol or
medication. All persons using this device must have read, understood and
follow the instructions provided in the operating manual.
Page 11
humimeter FS1 & FS1.1 Operating Manual
2.4 General safety information
The following safety information has to be observed at all times to avoid
damage to objects and injury to people:
· Remove the batteries if the device isn’t used for a prolonged period of time
(4 weeks).
· In case of damages or loose parts on the device, remove the batteries and
contact Schaller Messtechnik GmbH or your dealer.
All of the device’s technical features have been inspected and tested before
delivery. Every device has a serial number. Do not remove the tag with the
serial number.
2.5 Warranty
The warranty does not apply to: · Damage resulting from non-observance of the
operating manual · Damage resulting from third-party interventions · Products
that have been used improperly or modified without authorisation · Products
with missing or damaged warranty seals · Damage resulting from force majeure,
natural disasters, etc. · Damage from improper cleaning · Batteries older than
six months
3. On receipt of your device
3.1 Taking the device out of its packaging
· Take the device out of its packaging. · Next, make sure that it is not
damaged and that no parts are missing.
Page 12
3.2 Making sure that all of the components have been included
Make sure that all of the components have been included by checking the
package contents against the following list:
3.2.1 FS1’s scope of supply
· humimeter FS1 · 4 pieces of AA Alkaline batteries · Digital scale 150 grams
· Measuring cup 0.25 liter · Plastic case · Operating manual
3.2.2 FS1.1’s scope of supply
· humimeter FS1.1 · 4 pieces of AA Alkaline batteries · Digital scale 150
grams · Measuring cup 0.25 liter · Operating manual
3.3 Inserting batteries
1
1. Take hold of the device with one hand, press your thumb onto the engraved area of the battery compartment (1) and drag downwards (2) (figure 1).
2. Insert the batteries with negative and positive terminals matching those indicated on the battery compartment. 2 Press down the batteries so that they lay flat on the bottom of the housing (figure 2).
» As soon as all batteries have been inserted, the device switches on automatically.
3. Push the battery cover onto the housing until it clicks into
place (figure 3).
3
Page 13
humimeter FS1 & FS1.1 Operating Manual
4. Using the device – Basics
4.1 Switching the device on
· Press the button for 3 seconds. 4
» The display will then show the status indicator (figure 4).
» After inserting the batteries, the device switches on automatically.
4.2 Automatic Adjustment
» The display will show the message Adjust?
5
(figure 5).
1. Make sure that the measuring chamber of the device is empty and place the device on a level table.
2. Confirm by pressing .
6
» The display will now appear as shown in figure 6.
» The bar will run upwards. During this period, the device must remain on the table without external influence,
» which only takes a couple of seconds to complete.
» Once completed, the device will show the measuring window (see “Product selection menu” page 5).
4.3 Accessing the main menu
To do so: The devices is switched off.
· Switch on the device (see “4.1 Switching the device on”). · While switching
on, keep both the and buttons pressed. · The display will then show the main
menu.
Page 14
4.4 Selecting the calibration curve
To do so: The device has to be in the product selection menu (figure 7).
For an overview of the different calibration curves and the criteria for selecting them, please refer to “6. Calibration curves”es”.
1. Press the or button to move from one
7
calibration curve to the next Or
1. Press the or button for 3 seconds to open the calibration curve overview (figure 8).
2. Use the arrow keys to move from one calibration 8 curve to the next
3. and keep any of them pressed to scroll through the types.
4. Confirm your selection by pressing .
» The calibration curve you selected will now be shown at the top of the display.
4.5 Taking a measurement
· For information on how to take a measurement, see section “5. The measuring
process”.
4.6 Switching the device off
To do so: The device has to be in the product selection menu. It is not
possible to switch off the device when it is in the main menu.
· Press the button for 3 seconds.
Page 15
humimeter FS1 & FS1.1 Operating Manual
5. The measuring process
5.1 Preparing a measurement
To do so: The device has to have nearly the same temperature than the product being measured. It is recommended to let your humimeter device adjust to the surrounding temperature of the material being measured for at least 30 minutes.
1. Place the empty and clean measuring cup (0.25
liter) on the switched-off scale (figure 9). Then
9
switch on the scale.
» The scale must display 0.0 g with the empty measuring cup on it. The measuring cup must not be weighed.
2. Check whether the measuring chamber of the device is empty. When the
device is switched on, there must not be any material in it. 10
» Empty the instrument and clean the measuring chamber if necessary (see “9.3
Cleaning the device”).
3. Switch on the device (see “4.1 Switching the device
on”).
11
4. Perform the automatic adjustment (see “4.2 Automatic Adjustment”).
5. Select the required calibration curve (see “6.
Calibration curves”es”) by pressing the or calibration curve”rve”).
button (see “4.4 Selecting the
Page 16
5.2 Taking a measurement
To do so: The device has to have nearly the same tem- 12 perature than the
product being measured.
1. Fill the measuring cup with the filling quantity displayed in the
calibration curve name (+/- 1.0 g) (figure 12) (see “6. Calibration
curves”es”).
2. Now slowly and evenly fill the measuring chamber of the device with the
material being measured (figure 13).
» For the filling, no funnel or similar device may be 13 used.
3. The device will now display the moisture content (figure 14).
» The displayed value flashes when the moisture
content exceeds the measuring range of the
selected calibration curve (figure 15). A flashing
14
value signals a decreasing accuracy of the measure-
ment. The measuring range is dependent on the
calibration curve (see “6. Calibration curves”es”).
» It is possible to apply an offset to the displayed measurement value (see “5.3 Offset function – De- 15 vice adaptation”).
4. Empty the device and make sure that there are no residues in the measuring
chamber.
» Clean the measuring chamber if necessary (see “9.3 Cleaning the device”).
i Information – Measuring accuracy
Benefit from this rapid and non-destructive measuring procedure and take
several moisture readings of the same sample material.
Page 17
humimeter FS1 & FS1.1 Operating Manual
i Information – Incorrect readings
Always make sure to select the correct calibration curve and the correct
filling quantity for the material you are measuring. This prevents taking
incorrect readings (see “10. Faults”).
Page 18
5.3 Offset function – Device adaptation
By changing the offset, the displayed measurement values can be adapted to
other norms/standards. The displayed measuring value is corrected by the
entered offset.
Example: An offset of 1.5 % applied to a measurement value of 10.0 % results
in a displayed measurement value of 11.5 %. An offset of – 0.5 % applied to a
measurement value of 10.0 % results in a displayed measurement value of 9.5 %.
5.3.1 Setting the offset
To do so: The device has to be switched on and be in the product selection
menu.
1. Select the required calibration curve (see “6. Cali- 16 bration
curves”es”) by pressing the or button (see “4.4 Selecting the calibration
curve”rve”).
2. Press for 3 seconds.
» The display will now show the material calibration 17 menu for the selected
calibration curve (figure 17).
» The offset is part of the material calibration menu.
3. Select Offset. To do so, press or and confirm by pressing .
4. The data you have inputted can be overwritten at any time.
5. Inputting numbers:
Press and hold
to quickly scroll to the
18
required number and either press it for 3 seconds
or press to confirm the selected number
(figure 18).
» Setting a negative offset is also possible! To do so, insert a minus sign
before the first digit.
» Take care of the position of the comma to prevent setting an offset that is
too high!
6. Moving forward: To move forward, press .
Page 19
humimeter FS1 & FS1.1 Operating Manual
7. Moving back: Press to switch to another input level. To move back, press .
8. Confirm the offset by pressing .
» The offset has been saved.
9. Press to leave the material calibration menu. 19
10. The set offset will now be applied to the selected calibration curve and
shown in the display (figure 19).
» The displayed measurement value now deviates from the standard calibration!
6. Calibration curves
50g Corn 8 % – 30 %
60g Rye 8 % – 23 %
60g Triticale 8 % – 23 %
60g Wheat 8 % – 23 %
50g Barley 8 % – 23 %
30g Oats 8 % – 23 %
55g Rape 4 % – 15 %
60g Soybeans 6 % – 18 %
Page 20
60g Horse Beans 8 % – 18 %
35g Sunflower 5 % – 18 %
65g Rice peeled 8 % – 18 %
40g Rice unpeeled 5 % – 22 %
50g Buckwheat 8 % – 18 %
60g Mustard Seeds 5 % – 18 %
60g Sorghum Millet 8 % – 18 %
30g Coffee roasted 1 % – 10 %
50g Raw Coffee peeled
5 % – 18 %
Reference
! Only for testing the moisture meter !
6.1 How moisture content is defined
The device measures and shows a material’s moisture content. The moisture
content readings it displays are calculated in relation to the material’s
overall mass:
Mn: Mt: %WG:
Mass of the sample with average moisture content Mass of the sample with zero moisture content Moisture content (in accordance with the corresponding product norms)
Page 21
humimeter FS1 & FS1.1 Operating Manual
7. Checking the device’s status
1. Access the main menu (see “4.3 Accessing the main menu”). 2. Select
Status. To do so, press or and confirm by pressing .
» The display will then show the status indicator humimeter. » The display
will show the following information:
1 2 3
No. Name 1 Serial number 2 Software version 3 Battery status 3. Confirm by
pressing . 4. Press to leave the main menu.
Page 22
8. Configuring the device
8.1 Selecting a language
1. Access the main menu (see “4.3 Accessing the main menu”).
2. Select Options. To do so, press or and confirm by pressing .
3. Select Language. To do so, press or and confirm by pressing .
4. Navigate to the required language. To do so, press or and confirm by
pressing .
» The settings have been saved.
5. Press 6. Press
to leave the Options menu. to leave the main menu.
8.2 Activating options
To do so: Some of the options must be deactivated. 1. Access the main menu
(see “4.3 Accessing the main menu”).
2. Select Options. To do so, press or and confirm by pressing .
3. Select Unlock. To do so, press or and confirm by pressing .
» The display will now appear as shown in figure 20. 20
» On delivery, the four-digit password is the device’s serial number.
4. Inputting numbers:
Press and hold
to quickly scroll to the
required number and either press it for 3 seconds 21 or press to confirm the selected number
(figure 21).
5. Moving back:
Press to switch to another input level. To move back, press .
Page 23
humimeter FS1 & FS1.1 Operating Manual
6. Confirm the four-digit password by pressing .
» The settings have been saved.
» The °C/°F, BL On Time, Auto Off Time, Materialcalibration, Password, Reset
options are now activated.
7. Press 8. Press
to leave the Options menu. to leave the main menu.
8.3 Deactivating options
Once the device has been switched restarted, the °C/°F, BL On Time, Auto Off
Time, Materialcalibration, Password, Reset options will be deactivated again.
8.4 Selecting °C/°F
To do so: All of the options must be activated (see “8.2 Activating options”).
- Access the main menu (see “4.3 Accessing the main menu”).
2. Select Options. To do so, press or and confirm by pressing .
3. Select °C/°F. To do so, press or and confirm by pressing . 4. Navigate to
the required temperature scale, i.e. Celsius (°C) or Fahrenheit (°F). To
do so, press or and confirm by pressing . » The setting has been saved.
5. Press 6. Press
to leave the Options menu. to leave the main menu.
Page 24
8.5 Reducing the device’s power consumption
8.5.1 Configuring the display illumination time To do so: All of the options must be activated (see “8.2 Activating options”). 1. Access the main menu (see “4.3 Accessing the main menu”).
2. Select Options. To do so, press or and confirm by pressing .
3. Select BL On Time. To do so, press or and confirm by pressing . 4. Select
the required display illumination period (30 seconds, 2 minutes, 5 minutes,
10 minutes). To do so, press or and confirm by pressing . » The setting has
been saved.
5. Press 6. Press
to leave the Options menu. to leave the main menu.
8.5.2 Configuring automatic switch-off To do so: All of the options must be activated (see “8.2 Activating options”). 1. Access the main menu (see “4.3 Accessing the main menu”).
2. Select Options. To do so, press or and confirm by pressing .
3. Select Auto Off Time. To do so, press or and confirm by pressing . 4.
Select the period of time you want the device to stay switched on (3 minutes,
5
minutes, 10 minutes). To do so, press or and confirm by pressing . » The
setting has been saved.
5. Press 6. Press
to leave the Options menu. to leave the main menu.
8.6 Configuring the material calibration function
The type calibration function is described in a separate operating manual.
Page 25
humimeter FS1 & FS1.1 Operating Manual
8.7 Changing the password
To do so: All of the options must be activated (see “8.2 Activating options”).
- Access the main menu (see “4.3 Accessing the main menu”).
2. Select Options. To do so, press or and confirm by pressing .
3. Select Password. To do so, press or and confirm by pressing .
» The display will show the current password.
4. Overwrite the current password. To do so, press and hold
to quickly
scroll to the required number and either press it for 3 seconds or press to
confirm the selected number.
Moving back: Press to switch to another input level. To move back, press .
5. Confirm the new four-digit password by pressing .
» The setting has been saved.
6. Press 7. Press
to leave the Options menu. to leave the main menu.
Page 26
8.8 Resetting the device to its factory settings
To do so: All of the options must be activated (see “8.2 Activating options”).
- Access the main menu (see “4.3 Accessing the main menu”).
2. Select Options. To do so, press or and confirm by pressing .
3. Select Reset. To do so, press or and confirm by pressing .
» The display will then show the message Reset?
22
(figure 22).
4. Confirm by pressing .
» The device will now be reset to its factory settings. All of your personal
settings will be lost. 23
» The display will show the status indicator humimeter (figure 23).
9. Cleaning and maintenance
Regularly cleaning and maintaining the device will ensure that it will have a
long service life and stay in good condition.
9.1 Changing batteries
The device constantly monitors the charge level of the
24
batteries. The current battery status is shown on the
status screen.
If the battery´s charge is very low, the battery symbol will
be shown with an exclamation mark. In that case, the
batteries must be changed immediately (figure 25).
For changing the batteries, see section “3.3 Inserting
25
batteries”.
As the device’s user, you are responsible by law for properly disposing of all used batteries, which must not be disposed of as domestic waste (Battery Directive).
Page 27
humimeter FS1 & FS1.1 Operating Manual
9.2 Care instructions
· Do not leave the device out in the rain. The device is not waterproof. · Do
not expose the device to extreme temperatures. · Protect the device from
strong mechanical shocks and loads. · Remove the batteries after the
harvesting season.
9.3 Cleaning the device ATTENTION
Do not clean with fluids Water or cleaning fluid getting inside the device can
destroy the device. Only clean with dry materials. Plastic housing · Clean the
plastic housing with a dry cloth. Measuring chamber · Clean the measuring
chamber with a soft brush.
Page 28
10. Faults
If the measures listed below fail to remedy any faults or if the device has
faults not listed here, please contact Schaller Messtechnik GmbH.
Fault Measuring error
Cause
Remedy
The temperature of the material being measured is too low or high. I.e. the
material’s temperature is lower than 0 °C or higher than +40 °C. Temperature
discrepancy between device and material being measured
Wrong calibration curve
Wrong filling quantity
Mouldy or rain wet material
Stored and fermented corn from whole grain silage Frozen material or material
mixed with snow Contaminated material
The temperature of the material being measured has to be between 0 °C and +40
°C.
Let the temperature adjust to the material being measured (permitted
difference of max. 3 °C).
Check whether you have selected the right calibration curve (product) before
taking a reading (see “6. Calibration curves”es”).
Exactly fill in the filling quantity displayed in the calibration curve name
(+/- 0.5 gram).
The accuracy decreases significantly.
May lead to a too high displayed measuring value.
The accuracy decreases significantly.
Highly contaminated material such as long ears of barley or foreign material
can strongly influence the measuring result.
Page 29
humimeter FS1 & FS1.1 Operating Manual
Fault
Cause Entered offset
Remedy
An entered offset leads to deviations on the displayed measured value. If the
deviation does not match your reference procedure, enter an offset
corresponding to the difference or set the offset to 0.0 (see “5.3 Offset
function – Device adaptation”) to restore the factory characteristic curve.
11. Storage and disposal
11.1 Storing the device
The device must be stored as follows:
· Do not store outdoors. · Store in a dry and dust-free place. · Protect the
device from sunlight. · Avoid mechanical shocks/loads. · Remove the batteries
if the device isn’t used for a period of 4 weeks or longer · Storage
temperature: -20 °C to +60 °C
11.2 Disposing of the device
Devices marked with this symbol are subject to Directive 2012/19/ EU of the
European Parliament and of the Council of 4 July 2012 on waste electrical and
electronic equipment (WEEE). If the device is being operated outside the
European Union, the national regulations on the disposal of such devices that
apply in the country of use must be observed.
Electronic devices must not be disposed of as domestic waste.
The device must be disposed of appropriately using appropriate collection
systems.
Page 30
12. Device information
12.1 CE declaration of conformit
___ KONFORMITÄTSERKLÄRUNG DECLARATION OF CONFORMITY
Name/ Adresse des Herstellers: Name/ address of manufacturer:
Schaller Messtechnik GmbH Max-Schaller-Straße 99 A 8181 St. Ruprecht
Produktbezeichnung: Product designation:
Typenbezeichnung: Type designation:
Produktbeschreibung:
Product description
humimeter
FS1 ; FS1.1 ; FS2 ; FS3 ; FS4 ; FS4.1 ; FS4.2 ; BP1
Messgerät zur Bestimmung des Wassergehalts in Lebensmitteln Measuring
instrument for determining the water content in foodstuffs
Das bezeichnete Produkt erfüllt die Bestimmungen der Richtlinien: The designated product is in conformity with the European directives:
EMV – Richtlinie 2014/30/EC RoHS – Richtlinie 2011/65/EG
EMC Directive 2014/30/EU RoHS-Directive 2011/65/EU
Die Übereinstimmung des bezeichneten Produktes mit den Bestimmungen der
Richtlinien wird durch die vollständige Einhaltung folgender Normen
nachgewiesen:
Full compliance with the standards listed below proves the conformity of the
designated product with the provisions of the above-mentioned EC Directives:
EN 613261:2013
EN IEC 63000:2019-05 ersetzt / replaced EN 50581:2012
Elektrische Mess-, Steuer-, Regel- und Laborgeräte – EMV-Anforderungen
Electrical equipment for measurement, control, and laboratory use EMC
requirements
Technische Dokumentation zur Beurteilung von Elektro- und Elektronikgeräten
hinsichtlich der Beschränkung gefährliche Stoffe. Technical documentation for
the assessment of electrical and electronic products with respect to the
restriction of hazardous substances.
Page 31
humimeter FS1 & FS1.1 Operating Manual
Für das angeführte Produkt ist eine vollständige Dokumentation mit
Betriebsanleitung in Originalfassung vorhanden.
For the mentioned product a complete documentation with manual of instruction
in original version is available.
Bei Änderungen, die nicht vom Hersteller spezifiziert sind, verliert diese
Konformitätserklärung die Gültigkeit.
In case of any changes not agreed upon with the manufacturer, this declaration
of conformity loses its validity.
St. Ruprecht a.d. Raab, 31.07.2022
…………………………………………………..
Bernhard Maunz Rechtsverbindliche Unterschrift des Ausstellers Legal binding
signature of the issuer
Page 32
___ DECLARATION OF CONFORMITY
Name/ address of manufacturer:
Schaller Messtechnik GmbH Max-Schaller-Straße 99 A 8181 St. Ruprecht
Product designation: Type designation: Product description:
humimeter
FS1 ; FS1.1 ; FS2 ; FS3 ; FS4 ; FS4.1 ; FS4.2 ; BP1
Measuring instrument for determining the water content in foodstuffs
The designated product is in conformity with the following directives:
· Electromagnetic Compatibility Regulations 2016 Great Britain
· RoHS-Directive 2011/65/EU Directive on the restriction of the use of certain
hazardous substances in electrical and electronic equipment
Full compliance with the standards listed below proves the conformity of the designated product with the provisions of the above-mentioned Directives:
EN 613261:2013
EN IEC 63000:2019-05 replaced EN 50581:2012
Electrical equipment for measurement, control, and laboratory use EMC
requirements
Technical documentation for the assessment of electrical and electronic
products with respect to the restriction of hazardous substances.
Page 33
humimeter FS1 & FS1.1 Operating Manual
For the mentioned product, a complete documentation with manual of instruction
in original version is available.
In case of any changes not agreed upon with the manufacturer, this declaration
of conformity loses its validity.
St. Ruprecht a.d. Raab, 31.07.2022
…………………………………………………..
Bernhard Maunz Legal binding signature of the issuer
Page 34
12.2 Technical data
Display resolution
Measuring range (depending on calibration curve) Calibration accuracy to
reference material Operating temperature Storage temperature Temperature
compensation Power supply Current consumption Menu languages
Display Device dimensions Device weight Case dimensions (FS1’s scope of
supply) Weight of device + case Device IP rating
0.1 % moisture content, 0.5 °C/°F temperature
(FS1) 5 % to 30 % moisture content (FS1.1) 5 % to 25 % moisture content
(FS1) +/- 0.7 % (FS1.1) +/- 0.9 % 0 °C to +40 °C -20 °C to +60 °C Automatic 4
x 1.5 Volt AA Alkaline batteries 60 mA (incl. display illumination) English,
German, French, Italian, Spanish, Portuguese, Czech, Polish, Hungarian,
Russian, International 128 x 64 illuminated matrix display 155 x 75 x 90 mm
360 g
270 x 220 x 80 mm
850 g IP 40
Page 35
Schaller Messtechnik develops, produces and sells professional moisture meters
and turnkey solutions.
Schaller Messtechnik GmbH Max-Schaller-Straße 99, A – 8181 St. Ruprecht an der
Raab Tel +43 (0)3178 – 28899 , Fax +43 (0)3178 – 28899 – 901
info@humimeter.com, www.humimeter.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>