Universal Medical Antimicrobial Digital Platform Scale with Extended
Handrail
Product Specifications
- Model: UM3001KL
- Capacity: Up to 1000 lb / 454 kg
- Minimum Calibration Weight: 200 lb / 100 kg
- Calibration Increments: 10 lb or 10 kg
Product Usage Instructions
Accessing the Calibration Menu
Press Help to access the Help menu options and navigate to Calibrate.
Initiating Calibration
Follow the on-screen prompts to enter the password and proceed with the calibration procedure.
Preparing for Calibration
Ensure the scale is on a level surface with no weight on the platform before proceeding.
Zeroing the Scale
Allow the scale to calculate the zero value without any interference. Do not touch the scale during this process.
Setting Calibration Weight
Enter the calibration weight within the range of 200 lb / 100 kg to 1000 lb / 454 kg in increments of 10 lb or 10 kg. Press Enter once the calibration weight is set.
Placing Calibration Weights
Position the calibration weights on the platform as instructed and press Enter to proceed.
Completing Calibration:
The display will show Calibrating during the process. If successful, it will indicate Calibration Complete with a calibration count. If unsuccessful, follow on-screen instructions to correct the calibration weight.
Frequently Asked Questions
Q: Do I need to calibrate the scale before use?
A: The scale comes factory calibrated and does not require calibration prior to use. However, users should adhere to their institution’s calibration policies.
Q: What is the minimum and maximum calibration weight for this scale?
A: The minimum calibration weight is 200 lb / 100 kg, and the maximum calibration weight is 1000 lb / 454 kg.
CALIBRATION
This scale has been factory calibrated, and does not require calibration prior
to use. There are no requirements for calibration in the field; users should
adhere to their institution’s calibration policies.
Calibration of your scale can be performed using calibrated kilogram (kg) or
pound (lb) weight standards. Minimum calibration is 200 lb / 100 kg. Maximum
calibration weight is 1000 lb / 454 kg. The required calibration standard will
be the Unit of Measure (UOM) for which the scale is configured. If Everlock®
is activated the scale can only be calibrated in the locked unit of measure.
Note: Only weights certified and traceable to national standards should be used for calibration.
Warning: Always ensure that the scale is returned to the correct UOM after scale calibration.
-
Press to access the Help menu options.
-
While in the Help menu, press to select “Calibrate”. Press to advance to the next step.
-
The display will prompt for entry of the password. Press the following keys in this sequence to proceed with the calibration procedure.
-
The display will show a warning. If prepared and qualified to perform the calibration procedure, press to advance to the next step.
-
Follow the screen prompt and ensure the scale is on a level surface with no weight on the platform.
Press to advance to the next step. -
The display will show “Zeroing Scale” as it calculates the zero value. Do not touch the scale during the zeroing process.
-
When the zeroing process is complete the display will prompt for entry of the calibration weight. Minimum calibration is 200 lb / 100 kg. Maximum calibration weight is 1000 lb / 454 kg. Calibration weight can be adjusted in 10 lb or 10 kg increments. Press when the calibration weight is set.
-
The display will prompt for the calibration weights to be placed on the platform. Press after calibration weights are placed on the platform.
-
The display will show “Calibrating” during the calibration process.
-
If calibration is successful, the display will show “Calibration Complete” with a 4 digit calibration count.
-
If calibration is unsuccessful, the display will show an error message that the entered calibration weight and the certified weights do not match. The display will then revert back to the set calibration weight screen for entry of calibration weight.