Perkin Elmer T2 Toxin Fast User Guide

August 12, 2024
Perkin Elmer

Perkin Elmer T2 Toxin Fast

Product Specifications

  • Product Name: Toxin Fast T2
  • Quantitative Kit for Corn Flour and Corn
  • Model: MZLFQTT2
  • Target: Soft Wheat
  • Detection: T2 and HT2 toxins
  • Quantification Range for Corn Flour and Corn: 40-300 ppb
  • Quantification Range for Soft Wheat: 10-260 ppb

Product Usage Instructions

Sample Preparation

  1. Collect and grind a representative sample. Ideal particle size: 0.8 mm (20 mesh).

Sample Extraction

  1. Filter the sample. Ensure the mixture pH is > 6.5.

Reading Results

  1. Insert the strip and wait for a 5-minute reaction time.
  2. Before reading the result, scan the QR code corresponding to the matrix under examination included in the kit (white sheet).
  3. Remove the absorbent pad with scissors. Read the strip using the portable reader MZREAD-F.

FAQ

Frequently Asked Questions

  • What should I do if the pH of the sample mixture is below 6.5?
    • If the pH is below 6.5, adjust it using the provided pH corrector Cat# ACC005 before proceeding with the test.
  • How long does it take to complete the test?
    • The test typically takes around 15 minutes to complete, including sample preparation, extraction, and reading of results.
  • Can I reuse any of the components in the kit?
    • No, for accurate results and to avoid contamination, it is recommended to use the components provided in the kit for each test.

“`