RECI PROF VacuScope Easy Milking Machine Test Device User Manual
- June 3, 2024
- RECI PROF
Table of Contents
PROF VacuScope Easy Milking Machine Test Device
**User Manual
VacuScope easy
Milking Machine Test Device
Version V1.1
April 2022**
VacuScope Easy Milking Machine Test Device
NOTES ON THE INSTRUCTIONS MANUAL
In the interest of proper functionality and your own safety, please
familiarize yourself with the contents of this operating manual before use. It
contains important information that will enable you to use the device
properly.
The manufacturer reserves the right to make technical changes in the interest
of further development.
These instructions are part of the scope of delivery.
SAFETY INSTRUCTIONS
For your personal safety and to protect your warranty claims, please read the
operating manual and follow the safety instructions before starting up the
equipment.
Only persons who have been instructed in the intended use of the VacuScope
easy and who are familiar with the operating instructions are authorized to
commission the device. Knowledge of the accident prevention regulations and
the generally recognized safety rules is required to perform measurements.
The manufacturer is not liable for damage caused by improper use or
unauthorized tampering, especially with the electronic assemblies.
DESCRIPTION
3.1 INTRODUCTION
VacuScope easy is a handy device for checking pulsators and measuring the
vacuum curve in milking systems with outstanding product features:
- Visualization of the vacuum courses by a graphic display during testing
- Clear operation with fixed and adjustable keys
- Several interface connections for devices and sensors
- Large memory for storing measurement data and settings
Based on a wide range of programmes the device is used in development,
maintenance and servicing of milking units.
3.2 TECHNICAL DATA
Graphic display | LCD-Modul with 128 x 128 Pixel and and LED-backlight |
---|---|
Number of keys | 10 |
Measuring range | Vacuum 0…60 kPa |
Measuring accuracy | 0,6 kPa; 0,3 kPa (typ.) |
Measuring solution | 0,1 kPa |
Overload vac. sensors | max. 200 kPa |
Interfaces | RS 232 for printer, computer and air flow meter 0-5V-interface for |
external temperature sensor and pressure sensor
Data memory| 200 series in programme Pulsation 200 series in programme
Fluctuation
Power supply| 4 x AA, NiCd, NiMH
Protection| IP 65
Operating temperature| +5 bis +40 °C
Storage temperature| -20 bis +60 °C
Dimensioning| 195 x 100 x 50 mm
Weight| 420 g (w.o. batteries)
Tests| EN 61000-6-3, EN 61000-6-1
External Sensors (optional)
| Measuring range| Measuring accuracy| Measuring solution
---|---|---|---
Vacuum sensor/
Pressure sensor| 0…-50,0 kPa 0… 20,0 kPa| <0,6 kPa, <0,3 (typ)
<0,6 kPa, <0,3 (typ)| 0,1 kPa
0,1 kPa
Temp. sensor| -20,0….140,0 °C| 1,5%| 0,1 K
AFM 6001| 50-600 1/min at 50 kPa| 50…100I/min 5I/min 100…600I/min 5%| 1 1/min
AFM 30001| 100-3000I/min at 50 kPa| 5%| 1 1/min
FIRST START
4.1 PREPARATION
The battery compartment is on the back of the device. The instrument operates
with 4 1,5V- batteries (AA) or 4 accumulators (NiMH, NiCd).
Attention ! Check the polarity when inserting the batteries. Incorrect
polarity can destroy the electronics of the unit.
Attention ! VacuScope easy is designed for the measuring of atmospheric
pressure. Avoid measuring of other media.
This could falsify the measurement results or destroy the electronics. Clean
and dry hoses are required to ensure proper functioning of the pressure
sensors.
Attention ! Before commissioning, all hose connections must be checked
for leaks.
4.2 SWITCH DEVICE ON
The VacuScope easy is started by pressing the button.
While starting the instrument, the measuring systems zero calibration is
checked.
This defines the predominant air-pressure as zero level of the device.
****Achtung !
Measuring failures may appear when the device is started under pressure,
vacuum or pulsation
OPERATING ELEMENTS
5.1 KEY BOARD
The keys of the key board have the following meaning:
| Switch on the device
---|---
| Turn on / turn off the backlight of the display
| Moving in the menu up, setting parameters
| Moving in the menu down, setting parameters
| Close programme, cancel action
| Start programme, confirm of action
| Function key, function is displayed on screen
The function keys have the following meaning:
| System setting (language, user, calibration values)
---|---
| Delete Data
| Print out Data
| Transfer Data to computer
| Display measured data
| Save settings or measurement results
| Ruler, graphic tool
| Next readings (data sheet)
| Back to the first readings (first data sheet)
5.2 SCREEN
The instrument VacuScope easy has a backlit screen with a resolution of 128 x
128 pixels.
In the upper area of the menu windows, the current programme and the battery
charge status are displayed.
In the middle area are the menu items with the setting options for the
respective programmes.
The available functions are assigned to the function keys and displayed in the
lower part of the menu window.
5.3 INTERFACES
The VacuScope easy measuring device has a sensor interface (0-5V) for
connecting a temperature sensor or an external vacuum/pressure sensor.
The RS232 interface is used both to connect the air flow meter and to connect
the instrument to the computer.
The sensors, the air flow meter and the PC software are not included in the
scope of delivery of the meter and must be ordered separately.
Attention!
To ensure proper functioning of the external devices and sensors, screw them
tightly together with the meter.
PROGRAMMES
6.1 OPERATION
In the menu navigation, the programme items or the associated parameters are
selected with the and buttons (the current item is displayed inversely).
The selected program item is started with the button.
An input window opens for setting parameters (e.g. memory space).
The parameter can be changed with the keys and . To accept the set value press
the key to discard it press
To exit the menu and return to the higher-level menu press the button
6.2 MAIN MENU
After switching on the device and zeroing the measuring system, the display
shows the main menu. The programmes and functions in the
main menu:
PULSATION| Testing of clusters according to milking system standards ISO 5705
and ISO 6690
---|---
FLUCTUATION| Measurement and display of vacuum curves with freely selectable
measurement settings
SENSORS| Measurement with the external sensors and devices, numerical display
of the measured values
EXIT| Turn off the instrument
| System settings (language and user data), service programmes
6.3 SETTINGS AND SERVICE
In this programme, the operator can make settings and service activities.
Settings and functions in the SYSTEM
programme:
ENGL,DEUT,FRANCEITALI,ESPAN,PYCK | Selection of language |
---|
NAME1
NAME2| Setting of user data (e.g. name, company, location)
This data will be used in the documentation
CALIBRIERUNG| Checking the sensors with suitable measuring devices
CALIBRIERUNG| Configuration of the system, adjustment of the sensors This
programme is password protected and only accessible to trained personnel.
6.4 PROGRAMME PULSATION
6.4.1 MENU PULSATION
For the qualitative assessment of milking machines and systems, operators,
service personnel and control bodies may carry out an audit in accordance with
ISO 5707/6690. Settings and functions in the PULSATION programme:
MEMORY-NO.| Selection of the storage location for the pulsation measurement,
after saving, this value increases (+1) Setting range: 1…200
---|---
PULSATOR-NO| Selection of the pulsator number or the milking station in the
milking facility, after saving, this value increases (+1) Setting range: 1…100
START| Start of the PULSATION measurement
| Delete of stored pulsation measurements
| Print out of stored pulsation measurements
| Data transfer of stored pulsation measurements to the PC
| Display of stored pulsation measurements
6.4.2 MEASUREMENT OF PULSATION
After starting the measurement, the pulsation is analyzed according to the
specifications of ISO 6690.
After completion of the pulse analysis, the measurement results are displayed
graphically and numerically in the 1st window.
The determined pulse phase values are displayed in the 2nd window.
The graphics and all measured values can be saved or printed out from both the
first and second window.
Deviations from the specifications of ISO 5707:2010 are determined and
displayed flashing.
The target values for the pulse analysis according to ISO 5707:2010.
Limping [%] | < 5,0 % |
---|---|
Phase B [%] | >=30,0 % |
Phase D [%] | > =150ms |
The measuring window is closed with .
6.4.3 DATA MEMORY OF PROGRAMME PULSATION
One or more saved measurements can be sent to the PC for documentation or to
an external log printer. The first and last memory location to be transfered
must be entered beforehand.
PC software and printer are not included in the scope of delivery of the
VacuScope easy measuring device.
One or more saved measurements can be deleted in the PULSATION menu. The first
and last memory location to be deleted must be entered beforehand.
Stored data can also be shown on the display in the PULSATION menu. The set
memory location can be called up with the function key .
Use the keys and to call up further measurements.
6.5 PROGRAMME FLUCTUATION
6.5.1 MENU FLUCTUATION
The FLUCTUATION programme is available to the user for the graphic analysis
of pulsation and vacuum curves: Settings and functions in the FLUCTUATION
programme:
VACUUM CH3 | Selection of an external vacuum sensor |
---|---|
SCAN [MS] | Selection of the time interval between two measuring points |
MEMORY-NO. | Selection of the storage location for the fluctuation |
measurement, after saving, this value increases (+1)
Setting range: 1…200
START| Start of the fluctuation measurement
| Delete of stored fluctuation measurements
| Print out of stored fluctuation measurements
| Data transfer of stored fluctuation measurements to the PC
| Display of stored fluctuation measurements
6.5.2. MEASUREMENT OF FLUCTUATION
In contrast to the PULSATION programme, the measuring parameters can be freely
selected in the FLUCTUATION programme. Fluctuating vacuum conditions in the
milking system can be displayed and analyzed here. At the end of the vacuum
curve, the pulse frequency is determined on channel 1 and the maximum and
minimum on all connected measuring channels.
The measurements can be stored in the measurement memory.
The measuring window is closed with .
6.5.3 DATA MEMORY OF PROGRAMME FLUCTUATION
Stored data can also be shown on the display in the FLUCTUATION menu. The set
memory location can be called up with the function key . In this window, the
graph with its measured values can be printed out and also deleted. The RULER function can be used ti display a selected vacuum
level as a comparison.
Further measurements can be called up with the keys and from the data memory
of the FLUCTUATION programme.
6.6 PROGRAMME SENSORS
The following programmes are available for the static measurement of physical
measurands using external sensors and measuring devices:
Measuring programmes in the SENSORS programme:
VACUUM CH1 | Measuring a static vacuum value on channel 1 |
---|---|
VACUUM CH1-CH2 | Measuring a vacuum difference between channel 1 and 2 |
VACUUM CH3 | Measuring a static vacuum value with the external vacuum sensor on |
channel 3
PRESSURE CH3| Measuring a static pressure value with the external pressure
sensor on channel 3
TEMPERATURE| Measuring a temperature value with the external temperature
sensor
AIRFLOW| Measuring of air flow in milking installations with the measuring
devices AFM600 and AFM3000
Sensors and air flow meters are not included in the scope of delivery of the VacuScope easy measuring device.