SENVA SENSORS 153-0048-0A Gas Sensors User Guide
- July 18, 2024
- SENVA SENSORS
Table of Contents
SENVA SENSORS 153-0048-0A Gas Sensors
Product Information
Specifications
- Model: Gas Sensor XYZ
- Power Source: 2 AAA batteries
- Detection Range: 0-100 ppm
- Calibration Modes: Baseline and Span
- Indicator Lights: Green, Orange, Red
Product Usage Instructions
Hold down the left + right buttons for 5 seconds. The buzzer will begin to sound and the LED will blink Green 1 time every 2 seconds to indicate successful entry into calibration mode.
- Hold down the left button (sensor 1) or right button (sensor 2) for 5 seconds.
- The buzzer will begin to sound and the LED will blink orange: 1 time for sensor 1, or 2 times for sensor 2, every 2 seconds.
- IMPORTANT: Calibration requests take place instantly. Ensure the sensor is in a stable environment without gas for at least 15 minutes before applying a baseline calibration.
- Hold down the left button for 5 seconds to activate baseline calibration.
- The buzzer will sound and the LED will blink 3 times to indicate the status of the request. Red LED if failed, Green LED if succeeded.
- IMPORTANT: Check the provided table to ensure the correct gas is being used.
- Apply gas until the output stabilizes, which should take 60-90 seconds.
- Hold down the right button for 5 seconds to activate span calibration.
- The buzzer will sound and the LED will blink 3 times to indicate the status of the request. Red LED if failed, Green LED if succeeded.
NOTE: If calibration is done incorrectly, go back through steps 1 and 2, then hold down the left + right buttons for 10 seconds to reset to factory calibration. The buzzer will sound and the LED will blink 3 times to indicate the status of the request. Red LED if failed, Green LED if succeeded.
FAQ
- Q: How often should I perform calibration on the Gas Sensor XYZ?
- A: It is recommended to perform calibration at least once every month for accurate results.
- Q: What should I do if the LED blinks red during calibration?
- A: If the LED blinks red, it indicates a failed calibration. Please retry the calibration process ensuring the correct steps are followed.
Applies To: Senva TGR Series
Description: Recommendations for field calibration of TG sensors.
EQUIPMENT
- Contact Senva (866-660-8864) to buy calibration gas kits and shrouds which will contain all necessary equipment for each sensor.
- If using gas not provided by Senva, please ensure your calibration gas is not expired that it meets the below requirements for concentration (PPM or %) and you are using a 0.5 LPM regulator. It is also recommended to calibrate using Senva’s gas shrouds to ensure proper gas allocation.
FIELD CALIBRATION PROCEDURE
- LED indicator
- Buzzer
- Sensor Element 1
- Sensor Element 2
- Capacitive touch button
- Capacitive touch button
– NFC Antena
Enter Calibration
- Hold down the left + right buttons for 5 seconds. The buzzer will begin to sound and led will blink Green 1 time every 2 seconds to indicate successful entry into calibration mode.
Choose Sensor
- Hold down the left button (sensor 1) or right button (sensor 2) for 5 seconds.
- The buzzer will begin to sound and the LED will blink orange: 1 time for sensor 1, or 2 times for sensor 2, every 2 seconds.
Apply baseline calibration
- IMPORTANT: Calibration requests take place instantly. When calibrating the baseline, ensure the sensor is in a good stable environment without gas for at least 15 minutes before applying a baseline calibration.
- Hold down the left button for 5 seconds to activate baseline calibration.
- The buzzer will sound and the LED will blink 3 times to indicate the status of the request. Red LED if failed, Green LED if succeeded)
Apply Span Calibration
- IMPORTANT: Check the table above to ensure the correct gas is being used.
- Apply gas until output is stabilized, should take 60-90 seconds.
- Hold down the right button for 5 seconds to activate span calibration.
- The buzzer will sound and the LED will blink 3 times to indicate the status of the request. Red LED if failed, Green LED if succeeded)
NOTE: if calibration is done incorrectly go back through steps 1 and 2
then hold down the left + right buttons for 10 seconds to reset to factory
calibration.
The buzzer will sound and the LED will blink 3 times to indicate the status of
the request. Red LED if failed, Green LED if succeeded)