BUREAU VERITAS TKR-373MPPLUS Wireless Microphones Instruction Manual

July 6, 2024
BUREAU VERITAS

BUREAU VERITAS TKR-373MPPLUS Wireless Microphones

BUREAU-VERITAS-TKR-373MPPLUS-Wireless-Microphones-
PROIDUCT

Specifications

  • Applied Standard: FCC Part 15, Subpart C 15.236
  • FCC Part Section(s): 15.236(d), 15.236(f)(2), 15.236(g), 15.247(f)(3), 15.207, 15.203
  • Maximum Radiated Power: 50 mW EIRP
  • Occupied Bandwidth: < 200 kHz
  • Frequency Stability: PASS
  • Radiated Spurious Emission: < FCC 15.207 limits
  • Antenna Requirement: PASS

Product Usage Instructions

General Information
Ensure that the product is used in accordance with FCC Part 15 regulations as specified in the manual.

Test Results
Refer to the test results section to verify compliance with the specified standards.

Decision Rules for Statement of Conformity
Follow the QUA-52 Decision Rule for conformity assessment, including re- testing procedures if needed.

FAQ:
Q: What should I do if the test results are unclear?
A: If the original test results are subject to retesting and the judgement is unclear, follow the re-test procedure outlined in the manual.

RF Test Report

BUREAU-VERITAS-TKR-373MPPLUS-Wireless-Microphones-
\(2\) BV CPS ADT Korea Ltd.
This report is for your exclusive use. Any copying or replication of this report to or for any other person or entity, or use of our name or trademark, is permitted only with our prior written permission. This report sets forth our findings solely with respect to the test samples identified herein. The results set forth in this report are not indicative or representative of the quality or characteristics of the lot from which a test sample was taken or any similar or identical product unless specifically and expressly noted. Our report includes all of the tests requested by you and the results thereof based upon the information that you provided to us. You have 60 days from date of issuance of this report to notify us of any material error or omission caused by our negligence, provided, however, that such notice shall be in writing and shall specifically address the issue you wish to raise. A failure to raise such issue within the prescribed time shall constitute your unqualified acceptance of the completeness of this report, the tests conducted and the correctness of the report contents. Unless specific mention, the uncertainty of measurement has been ovnlicitly taken into account to derlare the comnliance or non-comnliance to the snecification

RELEASE CONTROL RECORD

REPORT NO. REASON FOR CHANGE DATE ISSUED
FCCCBNW-WAY-P24040288 Original release 2024.06.17
FCCCBNW-WAY-P24040288R1 Update the date of test 2024.06.28

Summary of Test Results

The EUT has been tested according to the following specifications

Applied Standard : FCC Part 15, Subpart C 15.236

FCC Part Section(s)| Test Description| Limit| Test Condition| Test Result| Reference

15.236(d)

|

Maximum Radiated Power

| (1) In the bands allocated and assigned for broadcast television and in the 600 MHz service band: 50 mW EIRP|

Conduted

|

PASS

|

Section 3.2

15.236(f)(2)| Occupied Bandwidth| < 200 kHz| PASS| Section 3.3
15.236(g)| Necessary Bandwidth| ETSI EN 300 422-1 V1.4.2

(2011-08) § 8.3.1.2

|

PASS

|

Section 3.4

15.247(f)(3)| Frequency Stability| < ±0.005%| PASS| Section 3.5

15.236(g)

| Radiated Spurious Emission| ETSI EN 300 422-1 V1.4.2

(2011-08) § 8.4.3

|

Radiated

|

PASS

|

Section 3.6

15.207

| AC Conducted Emissions

(150 kHz – 30 MHz)

|

< FCC 15.207 limits

| AC Line Conduted|

NANote3)

|

Section 3.7

15.203| Antenna Requirement| FCC 15.203| –| PASS| Section 3.1

NOTES

  1. The general test methods used to test on this devices are ANSI C63.10 / ETSI EN 300 422-1.
  2.  Determining compliance based on the results of the compliance measurement, not taking into  account measurement instrumentation uncertainty.
  3. Devices which only employ battery power for operation.

Decision Rules for Statement of Conformity

BUREAU-VERITAS-TKR-373MPPLUS-Wireless-Microphones-
\(3\) QUA-52 Decision Rule(QA Document) was applied.

  1. Step 1) : Reference Check, Daily Check, Peripheral device Check
  2. Step 2) : Re-test Procedure (Repeat the test maximum 3 times, Different Test Engineer)
  3. If the original test results are subject to retesting and the judgement is unclear, the retest is carried out.
  4. If the result of the first retest is the same as the initial test, the judgement is made based on the value.
  5. If the result of the first retest differ from the results of the initial test, the second re-test is carried out.
  6.  After completion of the second retest, the average of the three test results is determined as the final result. However, if the deviation of the three test values is more than 5 % of the reference value, the technical manager should review the reproducibility of the test from the beginning.

Measurement Uncertainty
Where relevant, the following measurement uncertainty levels have been estimated for tests performed on the EUT as specified in CISPR 16-4-2

Measurement Items| Frequency Range| Expanded Uncertainty U = k Uc ( k = 2)
---|---|---

Radiated Spurious Emissions

| 9 kHz – 30 MHz| 2.00
30 MHz – 1 GHz| 4.22
1 GHz – 18 GHz| 5.40
18 GHz – 26.5 GHz| 5.08
Measurement Items| Expanded Uncertainty U = k Uc ( k = 2)

Conducted

| Maximum Output Power| 1.20
Spurious Emissions| 1.36

This uncertainty represents an expanded uncertainty expressed at approximately the 95 % confidence level using a coverage factor of k = 2.

General Information

General Description of EUT

Equipment Class Wireless Microphone (DWM)
Product name MIC Karaoke
FCC ID TO8-TKR-373MPPLUS
Model TKR-372MP PRO PLUS
Additional model name
Power Supply DC 3.0 V (Battery 1.5 V x 2)
Modulation Type FM
Operating Frequency 502.0 MHz to 521.5 MHz
Output Power 9.64 dBm
Antenna Type Helical Antenna

Antenna Gain

| 502.0 MHz : -0.611 dBi

511.1 MHz : -1.811 dBi

521.5 MHz : -1.044 dBi

H/W Version| Ver 1.0
S/W Version| Ver 12

  1. NOTE 1: For a more detailed features description, please refer to the manufacturer’s specifications or the user’s  manual.
  2. NOTE 2: For the test results, the EUT had been tested with all conditions. But only the worst case was shown in

Tested sample and Tested companion device information

Type Model Note
Test sample (Conducted) TKR-373MP PLUS S/N: GY2004-00010
Test sample (Radiated) TKR-373MP PLUS S/N: GY2004-00013

Description of Test Mode
The EUT has been tested with all modes of operating conditions to determine the worst case emission characteristics.

Test Mode Tested Frequency (MHz)
TM 1 Wireless Microphone

Tested Frequency Channel Information


Channel

| Tested Frequency (MHz)
---|---
RF1 Frequency| RF2 Frequency
1| 502.0| 512.4
2| 503.3| 513.7
3| 504.6| 515.0
4| 505.9| 516.3
5| 507.2| 517.6
6| 508.5| 518.9
7| 509.8| 520.2
8| 511.1| 521.5

INSTRUMENT CALIBRATION
The measuring equipment, which was utilized in performing the tests documented herein, has been calibrated in accordance with the manufacturer’s recommendations for utilizing calibration equipments, which is traceable to recognized national standards.

EUT CONFIGURATION
The EUT configuration for testing is installed on RF field strength measurement to meet the Commissions requirement and operating in a manner that intends to maximize its emission characteristics in a continuous normal application.

General Description of Applied Standards
Generally the tests were performed according to the specifications of the standard, it must comply with the requirements of the following standards.

  • FCC CFR 47 Part 15, Subpart C (§15.236)
  • KDB 206256 D01 Wireless Microphones v02r01
  • ANSI C63.10-2013
  • All test items in this test report have been performed and recorded as per the above standards.

Test Equipment
Test Equipment is traceable to the National Institute of Standards and Technology (NIST). Measurement antenna used during testing were calibrated in accordance to the requirements of ANSI C63.5-2017.

Equipment Name| Manufacturer| Model No.| Serial No.| Next Cal.

Date

---|---|---|---|---
Signal Analyzer| Keysight Technologies| N9030B| MY57142476| 2024-11-22
Signal Analyzer| Keysight Technologies| N9020B| MY62150135| 2025-05-22
Signal Analyzer| R&S| FSV30| 103017| 2024-11-22
Power Sensor| R&S| NRP-Z211| 102377| 2024-11-22
Power Meter| R&S| NRX| 103577| 2024-11-22
Function Generator| Agilent| 33220A| MY44024913| 2025-05-22
Temperature & Humidity Chamber| Espec| PL-2J| 15015910| 2025-05-21
DC Power Supply| Agilent| 6674A| MY41003717| 2024-11-22
DC Power Supply| Keysight| U8002A| MY59020015| 2024-11-22
True-RMS Digital Multimeter| Fluke| 177| 43240434| 2025-05-21
Trilog Antenna (with 6 dB ATT.)| Schwarzbeck| VULB 9163| 01100| 2025-02-08
Horn Antenna| TA Engineering| TA-DRG118| TAEDRG2022 0407| 2024-07-25
Horn Antenna| TA Engineering| TAE1018-24-27A| 01| 2024-12-06
Horn Antenna| Q-PAR| QWH-DL-2-18-S- HG-R| 27679| 2024-07-27
Horn Antenna| TA Engineering| WR42-23A| 220412| 2024-07-26
Horn Antenna| TA Engineering| WR28-20A| 220412| 2024-07-26
Amplifier| Mini Circuits| ZFL-1000H+| 2232002150| 2025-05-25
Amplifier| B&Z Technologies| BZRU-01001800- 232054-302323| 27457-27462| 2025-05-25
Band Reject Filter| MICRO-TRONICS| BRM50702| G574| 2025-05-22
High Pass Filter| Wt Microwave| WT-A1699-HS| WT190313-6- 5| 2024-11-22
Attenuator| Mini-Circuits| VAT-10W2+| 1622| 2024-11-22
RSE Chamber| Daihan Sheild| –| 22-03-001| –
Signal Generator| R&S| SMB100A| MY41006053| 2025-05-21
Humidity Barometer TEMP Meter| LUTRON| MHB-382SD| AJ.38459| 2024-11-22

Test Results

Antenna Requirement
Except from §15.203 of the FCC Rules/Regulations:

An intentional radiator antenna shall be designed to ensure that no antenna other than that furnished by the responsible party can be used with the device. The use of a permanently attached antenna or of an antenna that uses a unique coupling to the intentional radiator shall be considered sufficient to comply with the provisions of the section.

  • The antenna(s) of the EUT are Permanently attached.
  • There are no provisions for connection to an external antenna.

Result
The EUT complies with the requirement of §15.203

Maximum Radiated Power

Regulation
§15.236(d) :

  1. In the bands allocated and assigned for broadcast television and in the 600 MHz service band: 50 mW EIRP

Test Procedure
Method AVGPM-G is a measurement using a gated RF average power meter.
Alternatively, measurements may be performed using a wideband gated RF power meter provided that the gate parameters are adjusted such that the power is measured only when the EUT is transmitting at its maximum power control level. Because the measurement is made only during the ON time of the transmitter, no duty cycle correction factor is required.

Test Setup

BUREAU-VERITAS-TKR-373MPPLUS-Wireless-Microphones-
\(4\)

Test Result
[Test Data of Maximum Radiated Power] Limit: 50 mW EIRP


Test Mode

| Tested Frequency [MHz]| Peak Output Power [dBm]| Antenna Gain [dBi]| ****

EIRP [dBm]

| ****

EIRP [mW]

---|---|---|---|---|---



TM 1

| 502.0| 9.530| -0.611| 8.92| 7.80
511.1| 9.600| -1.811| 7.79| 6.01
521.5| 9.640| -1.044| 8.60| 7.24

Note: EIRP = Peak Output Power + Antenna Gain

Occupied Bandwidth
Regulation
§15.236(f)(2) : One or more adjacent 25 kHz segments within the assignable frequencies may be combined to form a channel whose maximum bandwidth shall not exceed 200 kHz. The operating bandwidth shall not exceed 200 kHz.

Test Procedure
The occupied bandwidth is the frequency bandwidth such that, below its lower and above its upper frequency limits, the mean powers are each equal to 0.5% of the total mean power of the given emission. The following procedure shall be used for measuring 99% power bandwidth:

  • The instrument center frequency is set to the nominal EUT channel center frequency. The frequency span for the spectrum analyzer shall be between 1.5 times and 5.0 times the OBW.
  •  The nominal IF filter bandwidth (3 dB RBW) shall be in the range of 1% to 5% of the OBW, and VBW shall be approximately three times the RBW, unless otherwise specified by the applicable requirement.
  • Set the reference level of the instrument as required, keeping the signal from exceeding the maximum input mixer level for linear operation. In general, the peak of the spectral envelope shall be more than [10 log (OBW/RBW)] below the reference level. Specific guidance is given in 4.1.5.2.
  •  Step a) through step c) might require iteration to adjust within the specified range.
  • Video averaging is not permitted. Where practical, a sample detection and s26ingle sweep mode shall be used. Otherwise, peak detection and max hold mode (until the trace stabilizes) shall be used.
  •  Use the 99% power bandwidth function of the instrument (if available) and report the measured bandwidth.
  • If the instrument does not have a 99% power bandwidth function, then the trace data points are recovered and directly summed in linear power terms. The recovered amplitude data points, beginning at the lowest frequency, are placed in a running sum until 0.5% of the total is reached; that frequency is recorded as the lower frequency. The process is repeated until 99.5% of the total is reached; that frequency is recorded as the upper frequency. The 99% power bandwidth is the difference between these two frequencies.
  • The occupied bandwidth shall be reported by providing plot(s) of the measuring instrument display; the plot axes and the scale units per division shall be clearly labeled. Tabular data may be reported in addition to the plot(s).

Test Setup

BUREAU-VERITAS-TKR-373MPPLUS-Wireless-Microphones-
\(5\)

Test Result
[Test Data of Occupied Bandwidth] Limit: 200 kHz

Test Mode| Tested Frequency [MHz]| -26dB Bandwidth [kHz]| Occupied Bandwidth [kHz]
---|---|---|---



TM 1

| 502.0| 155.570| 107.393
511.1| 155.570| 108.392
521.5| 156.300| 108.392

[Test Plot of Occupied Bandwidth]

BUREAU-VERITAS-TKR-373MPPLUS-Wireless-Microphones-
\(6\)

Necessary Bandwidth

Regulation

§15.236(g) : Emissions within the band from one megahertz below to one megahertz above the carrier frequency shall comply with the emission mask in § 8.3 of ETSI EN 300 422-1 V1.4.2 (2011-08), Electromagnetic compatibility and Radio spectrum Matters (ERM); Wireless microphones in the 25 MHz to 3 GHz frequency range; Part 1: Technical characteristics and methods of measurement. Emissions outside of this band shall comply with the limits specified in section 8.4 of ETSI EN 300 422-1 V1.4.2 (2011-08).

Test Procedure

Necessary Bandwidth (BN) for Analogue Systems

  • The arrangement of test equipment as shown in figure B.1 shall be used. Note that the noise meter conforms to (quasi peak) without weighting filter (flat).
  • With the Low Frequency (LF) audio signal generator set to 500 Hz, the audio input level to the EUT shall be adjusted to 8 dB below the limiting threshold (-8 dB (lim)) as declared by the manufacturer.
  • The corresponding audio output level from the demodulator shall be measured and recorded.
  • The input impedance of the noise meter shall be sufficiently high to avoid more than 0,1 dB change in input level when the meter is switched between input and output.
  • The audio input level shall be increased by 20 dB, i.e. to +12 dB (lim), and the corresponding change in output level shall be measured.
  • It shall be checked that the audio output level has increased by ≤ 10 dB.
  • If this condition is not met, the initial audio input level shall be increased from -8 dB (lim) in 1 dB steps until the above condition is fulfilled, and the input level recorded in the test report. This level replaces the value derived from the  manufacturer’s declaration and is defined as -8 dB (I’m).
  • Measure the input level at the transmitter required to give +12 dB (lim).
  • The LF generator shall be replaced with the weighted noise source to ITU-R Recommendation BS.559-2 [1], band-limited to 15 kHz as described in IEC 60244 -13 [2], and the level shall be adjusted such that the measured input to the
  • transmitter corresponds to +12 dB (lim).
  • If the transmitter incorporates any ancillary coding or signalling channels (e.g. pilot-tones), these shall be enabled prior to any spectral measurements.
  • If the transmitter incorporates more than one audio input, e.g. stereo systems, the second and subsequent channels shall be simultaneously driven from the same noise source, attenuated to a level of -6 dB (lim).
  • The transmitter RF output spectrum shall be measured, using a spectrum analyser with the following settings:
    • centre frequency: fc: Transmitter (Tx) nominal frequency;
    • dispersion (Span): fc – 1 MHz to fc + 1 MHz;
    • Resolution BandWidth (RBW): 1 kHz;
    • Video BandWidth (VBW): 1 kHz;
    • detector: Peak hold.

Necessary Bandwidth (BN) for Digital Systems
NOTE 1: This parameter also includes the limits for spectral components within the out-of-band region. Principal

Spectrum Mask measuring method for digital transmitters:

  • Spectrum mask below 1 GHz, see figure 4, for the spectrum mask above 1 GHz, see figure 5.
    The transmitter shall be modulated with the test signals defined in clause 7.1.2. In any case the mask shall not be exceeded.
  1. Step 1: Measure the “Carrier Power” with the spectrum analyzer setup:

    • Center Frequency = fc
    • Span = Zero span
    • Detector = RMS
    • Trace Mode = Average
    • RBW&VBW = 5 x B
    • Sweep time ≥ 2 s
  2. Step 2: Measure the “Maximum Relative Level (dBc) at Specified Carrier Offsets” with the following spectrum analyzer setup:

    • Center Frequency = fc

    • Span ≥ 5 x B
      Detector = RMS
      Trace Mode = Peak Hold

    • RBW&VBW = 1 kHz

    • Sweep time ≥ 2 s

  3. Step 3: Measure the “transmitter wide band noise floor”:

    • The measurement of transmitter broad band noise floor shall be carried out according to clause 8.3.1.1.
    • Start Frequency = fc + 1,75B and fc – 1 MHz below 1 GHz, Start Frequency = fc + B and fc – 1 MHz above 1 GHz.
    • Stop Frequency = fc + 1 MHz and fc – 1,75 B below 1 GHz, Stop Frequency = fc + 1 MHz and fc -B above 1 GHz.
    • Detector = RMS
    • Trace Mode = Average
    • RBW&VBW = 1 kHz
    • Sweep time ≥ 2 s

NOTE 2: Two spectrum ranges are to be measured!

Test Setup

BUREAU-VERITAS-TKR-373MPPLUS-Wireless-Microphones-
\(7\)

Test ResultBUREAU-VERITAS-TKR-373MPPLUS-Wireless-Microphones-
\(8\)

[Test Plot of Necessary Bandwidth] BUREAU-VERITAS-TKR-373MPPLUS-
Wireless-Microphones- \(9\)

Frequency Stability
Regulation

§15.236(f) : The frequency tolerance of the carrier signal shall be maintained within ±0.005% of the operating frequency over a temperature variation of −20 degrees to +50 degrees C at normal supply voltage, and for a variation in the primary supply voltage from 85% to 115% of the rated supply voltage at a temperature of 20 degrees C. Battery operated equipment shall be tested using a new battery.

Test Procedure
Frequency stability with respect to ambient temperature

  • Supply the EUT with a nominal ac voltage or install a new or fully charged battery in the EUT. If possible, a dummy load shall be connected to the EUT because an antenna near the metallic walls of an environmental test chamber could affect the output frequency of the EUT. If the EUT is equipped with a permanently attached, adjustable-length antenna, then the EUT shall be placed in the center of the chamber with the antenna adjusted to the shortest length possible. Turn ON the EUT and tune it to one of the number of frequencies shown in 5.6.

  •  Couple the unlicensed wireless device output to the measuring instrument by connecting an antenna to the measuring instrument with a suitable length of coaxial cable and placing the measuring antenna near the EUT (e.g., 15 cm away), or by connecting a dummy load to the measuring instrument, through an attenuator if necessary.
    NOTE—An instrument that has an adequate level of accuracy as specified by the procuring or regulatory agency is the recommended measuring instrument.

  •  Adjust the location of the measurement antenna and the controls on the measurement instrument to obtain a suitable signal level (i.e., a level that will not overload the measurement instrument but is strong enough to allow measurement of the operating or fundamental frequency of the EUT).

  • Turn the EUT OFF and place it inside the environmental temperature chamber. For devices that have oscillator heaters, energize only the heater circuit.

  •  Set the temperature control on the chamber to the highest specified in the regulatory requirements for the type of device and allow the oscillator heater and the chamber temperature to stabilize.

  • While maintaining a constant temperature inside the environmental chamber, turn the EUT ON and record the operating frequency at startup, and at 2 minutes, 5 minutes, and 10 minutes after the EUT is energized. Four measurements in total are made.

  • Measure the frequency at each of frequencies specified in 5.6.

  • Switch OFF the EUT but do not switch OFF the oscillator heater.

  • Lower the chamber temperature by not more that 10 °C, and allow the temperature inside the chamber to stabilize.

  • Repeat step f) through step i) down to the lowest specified temperature.

Frequency stability when varying supply voltage
Unless otherwise specified, these tests shall be made at ambient room temperature (+15 °C to +25 °C). An antenna shall be connected to the antenna output terminals of the EUT if possible. If the EUT is equipped with or uses an adjustable-length antenna, then it shall be fully extended.

  •  Supply the EUT with nominal voltage or install a new or fully charged battery in the EUT. Turn ON the EUT and couple its output to a frequency counter or other frequency-measuring instrument.
    NOTE—An instrument that has an adequate level of accuracy as specified by the procuring or regulatory agency is the recommended measuring instrument.

  • Tune the EUT to one of the number of frequencies required in 5.6. Adjust the location of the measurement antenna and the controls on the measurement instrument to obtain a suitable signal level (i.e., a level that will not overload the measurement instrument but is strong enough to allow measurement of the operating or fundamental frequency of the EUT).

  • Measure the frequency at each of the frequencies specified in 5.6.

  • Repeat the above procedure at 85% and 115% of the nominal supply voltage as described in 5.13.

Test Setup

BUREAU-VERITAS-TKR-373MPPLUS-Wireless-Microphones-
\(10\)

Test Result
[Test Data of Frequency Stability]

Test Mode| Tested Frequency

[MHz]

| Temperature [ºC]| Voltage [VDC]| Measurement Frequency

[MHz]

| Frequency Error

[kHz]

| Limit [kHz]
---|---|---|---|---|---|---





















TM 1

| ****






502.0

| 50| ****




3.00

| 502.00060| 0.600| ****






±25.100

40| 502.00180| 1.800
30| 502.00300| 3.000
10| 502.00480| 4.800
0| 502.00519| 5.190
-10| 502.00519| 5.190
-20| 502.00500| 5.000


20

| 3.00| 502.00376| 3.760
2.55| 502.00203| 2.030
3.45| 502.00260| 2.600







511.1

| 50| ****




3.00

| 0.00040| 0.400| ****






±25.555

40| 0.00160| 1.600
30| 0.00300| 3.000
10| 0.00460| 4.600
0| 0.00500| 5.000
-10| 0.00519| 5.190
-20| 0.00460| 4.600


20

| 3.00| 0.00347| 3.470
2.55| 0.00347| 3.470
3.45| 0.00260| 2.600







521.5

| 50| ****




3.00

| 0.00040| 0.400| ****






±26.075

40| 0.00160| 1.600
30| 0.00320| 3.200
10| 0.00440| 4.400
0| 0.00519| 5.190
-10| 0.00539| 5.390
-20| 0.00519| 5.190


20

| 3.00| 0.00260| 2.600
2.55| 0.00203| 2.030
3.45| 0.00318| 3.180

Radiated Spurious Emission

Regulation
§15.236(g) : Emissions within the band from one megahertz below to one megahertz above the carrier frequency shall comply with the emission mask in § 8.3 of ETSI EN 300 422-1 V1.4.2 (2011-08), Electromagnetic compatibility and Radio spectrum Matters (ERM); Wireless microphones in the 25 MHz to 3 GHz frequency range; Part 1: Technical characteristics and methods of measurement. Emissions outside of this band shall comply with the limits specified in section 8.4 of ETSI EN 300 422-1 V1.4.2 (2011-08).
Test Procedure
On a test site, the sample shall be placed at the specified height on a non- conducting support. The transmitter shall be operated at the power as specified under clause 8.2, delivered to the antenna (see clause 5.1.1).
Radiation of any spurious components shall be detected by the test antenna and receiver, over the frequency range specified below, excluding the 250 % (out of band region) band of frequencies centred on the channel on which the transmitter is intended to operate.
NOTE: The 250 % (out of band region) exclusion is covered by measurements carried out in clauses 8.3.1 and 8.3.2.

  • The measuring receiver, as defined in table 4, shall be tuned over the frequency range 25 MHz to 4 GHz for equipment operating on frequencies below 1 GHz or in the frequency range of 25 MHz to 12,75 GHz for equipment operating on frequencies above 1 GHz.
  • At each frequency at which a component is detected, the sample shall be rotated to obtain maximum response and the effective radiated power of that component determined by a substitution measurement.
  • The measurement shall be repeated with the test antenna in the orthogonal polarization plane.
  • If the transmitter allows for standby operation, the tests shall be repeated with the transmitter in standby mode.

Test Setup

Test Result of Radiated Spurious Emission
Remarks

  1. Result (dBm) = S/A Reading Value(dBm) + Total Factor(dB/m) + DCCF(dB)
  2. Total Factor(dB/m) = T.F (dB/m) = Antenna Factor(dB/m) + Cable Loss(dB) – Pre-Amplifier Gain(dB)
  3. Margin(dB) = Result (dBm) – Limit (dBm)
  4. Measurement Distance = 3 m
  5.  DCCF = Duty Cycle Correction Factor.
  6. No other spurious and harmonic emissions were found greater than listed emissions on above table

Radiated Emissions BUREAU-VERITAS-TKR-373MPPLUS-Wireless-Microphones-
\(12\)

AC Conducted Emissions (150 kHz to 30 MHz)
Regulation
§15.207(a) : Except as shown in paragraphs (b) and (c) of this section, for an intentional radiator that is designed to be connected to the public utility (AC) power line, the radio frequency voltage that is conducted back onto the AC power line on any frequency or frequencies, within the band 150 kHz to 30 MHz, shall not exceed the limits in the following table, as measured using a 50 μH/50 ohms line impedance stabilization network (LISN). Compliance with the provisions of this paragraph shall be based on the measurement of the radio frequency voltage between each power line and ground at the power terminal. The lower limit applies at the boundary between the frequency ranges.

  • Decreases with the logarithm of the frequency.

Test Procedure

  •  The EUT was placed 0.4 meters from the conducting wall of the shielded room with EUT being connected to the power mains through a line impedance stabilization network (LISN). Other support units were connected to the power mains through another LISN. The two LISNs provide 50 ohm / 50 μH of coupling impedance for the measuring instrument.
  • Both lines of the power mains connected to the EUT were checked for maximum conducted interference.
  •  The frequency range from 150 kHz to 30 MHz was searched. Emission levels under (Limit – 20 dB) was not recorded.

Remark : The resolution bandwidth and video bandwidth of test receiver is 9 kHz for quasi-peak detection (QP) and average detection (AV) at frequency 0.15 MHz – 30 MHz.

Test Setup

Note: 1.Support units were connected to second LISN.

Test Result
– N/A
The device was only powered by battery when operating.

  • Appendix – Information of the Testing Laboratories
  • We, Bureau Veritas Consumer Products Services Korea. Our laboratories are FCC recognized accredited test firms and accredited and approved according to ISO/IEC 17025.
  • Test Firm Name : BV CPS ADT Korea Ltd.
  • Address : Innoplex No.2 106, Sinwon-ro 306, Yeongtong-gu, Suwon-si, Gyeonggi-do, 16675 KOREA

FCC

Designation Number : KR0158
Test Firm Registration Number : 666061

USED

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals