DRYDEN AQUA DA-GEN 150 Generator User Manual

July 5, 2024
DRYDEN AQUA

DRYDEN AQUA DA-GEN 150 Generator

Specifications:

  • Electrical consumption:
    • DA-GEN 24: 90 W
    • DA-GEN 45: 125 W
    • DA-GEN 90: 180 W
    • DA-GEN 150: 175 W
    • DA-GEN 240: 680 W
    • DA-GEN 360: 1000 W
    • DA-GEN 500: 1020 W
    • DA-GEN 750: 2880 W

Product Usage Instructions

System Installation:
Ensure proper electrical connections and use appropriate circuit breakers based on the product consumption specifications provided above.

Main Screen Overview:

  • Check the status of auxiliary relays, production intensity, pH setpoint, chlorine levels, water temperature, and other important parameters displayed on the main screen.
  • Monitor filtration status, alarms, and automatic measurements for pH, redox, free chlorine, and conductivity.

Electrical Connections:
Connect all sensors carefully to avoid irreparable damage to the device. Follow the instructions provided in the manual for proper sensor connections.

Control Panel Functions:

  • Control various functions such as ON/OFF switch, gas detector connection, pH dosing pump, additional chlorine pump (if available), heating control, and more.
  • Monitor the status of filtration relay and lights (manual/automatic).

Touch Screen Usage:

  • Access additional settings and controls through the touchscreen interface.
  • View important information such as current time, WiFi signal strength, and filtration mode.

DESCRIPTION

DA-Gen is a traditional salt electrolyze unit, working by splitting NaCL into Na and CL. By this, chlorine is added to the pool water as main disinfectant. Furthermore, free radicals are produced, which supports the chlorine in killing bacteria. The level of chlorine in the pool water is adjustable to the level the user decides and to the level demanded from the standards demanded from the authorities. The unit should be specified to fit the size of the pool, in order to be able to produce sufficient amount of chlorine.
The content of NaCL in the pool water has influence on the production capacity of the unit. Recommended level is between 2000 – 4000 TDS. The volume of the pool and the TDS will specify the size of the unit. By this, the unit can produce any level of chlorine demanded by user. The specification of the unit can be provided of the distributor of the unit. Hydrogen can be lead away from the unit by installing a pipe connection to outside air.

Control panel

  1. Connection cell
  2. Gas detector connection
  3. Main 230V connection
  4. ON/OFF switch
  5. 4 Amp fuse for panel and cell
  6. 4 Amp fuse relays

Cell

  1. Cell
  2. Gas detector connector
  3. Cell connector
  4. Cell housing
  5. Gas detector (internal)1

MAIN SCREEN

SYSTEM INSTALLATION

  1. External pump controller*
  2. Filter with AFM
  3. Filter pump
  4. Control Panel
  5. Free chlorine cell with rotameter Cell (always in vertical position if installed without paddle flow swit 6ch )  pH probe
  6. Redox probe and/or conductivity probe
  7. Paddle flow switch and temperature module APF® dosing pump
  8. pH dosing pump
  9. pH injection
  10. Prefilter
  11. Wifi module (see chapter 15)
  12. APF
  13. (not included)
  14. pH-Minus (not included)
  15. If Outdoor pool: ACO®  (not included)
  16. Besgo valve (not included)

Electrical consumption
Use of a 13 Amp time delay circuit breaker is recommended for private devices and a 16 A breaker for public devices. In case of sharing the power supply with other devices please consult a technician in order to dimension a correct installation.

ELECTRICAL CONNECTIONS

WATER PARAMETERS

The following values should be checked: On a daily basis:

  • pH*
  • Total, Free and combined chlorine ( = According to DS477)

On a weekly basis:
Alkalinity. Recommended Level: 100-200ppm

Salt level (NaCl) : According to the decided level in the pool
Recommended TDS level is between 2000 – 4000 ppm The volume of the pool and the TDS will specify the size of the unit to be used. It is recommended to check and adjust the TDS value once a week. The only chemical needed for operations of the unit is NaCl.

SYSTEM SETTINGS

PADDLE FLOW SWITCH

SUCTION LANCE (pH BOTTLE)

HYDROLYSIS

  1. Hydrolysis : Programming of hydrolysis functions

  2. Level :
    Hydrolysis – Desired disinfection production (Always 100%). Boost has no effect, leave as off.

  3. Mode : If the device has Free Chlorine and redox probes, choose the parameter that controls the cell’s chlorine generation.

MEASUREMENTS

  1. Measurements : Adjustment of setpoints and measuring probes.
  2. Setpoints for each measurement.
  3. Setpoints : according to public standards / DS 477

pH Calibration

  • 10.11 Calibration of pH probe: Recom-mended at least every 2-3 months during the usage season. Calibrate it always first with the Buffer (2pt).
  • 10.12 Calibration with buffers (buffer pH7 / pH10 / neutral): Follow the instructions that appear on the display.
  • 10.14 Manual calibration : Allows manual adjustment of the probes – only recommended to correct small deviations in the readings.
  • 10.15 Without removing the probe from the water, use the up/down arrows to adjust the reading so it matches your reference value (photometer or other measurement).

Temperature calibration

  • 10.22 Temperature calibration: To set the difference between the measured value of the probe and the actual temperature, use the up/down keys. Set to the actual temperature of the probe and press OK.

Free Chlorine calibration

  • 10.31 Calibration of the Free Chlorine probe: Recommended every month.
  • 10.32 Calibration with buffer (photometer DPD1): Follow the instructions in 7 steps that appear in the display.
  • 10.33 Step 1 of 7 – Calibrate Cl at 0 ppm (offset): Close the water flow through the probe and wait for 5 to 60 min until the reading is close to 0. Press OK
  • 10.34 Step 3 of 7 – Calibrate Cl: Set the water flow to the correct rate of 80-100 litres/hour. Wait for 1 to 10 min until there is a stable ppm reading. Press OK.
  • 10.35 Step 5 of 7 – Establish the real ppm values with the up/down keys according to your DPD1 (free chlorine) value. Press OK. 10.36 Step 7 of 7 – If this screen is not shown repeat the calibration process.
  • 10.38 Manual calibration: Open the water flow and set the flowmeter (rotameter) to the correct flowrate (50-100l/h). Wait until the current level is stable. Set the chlorine level with the up/down keys, manually (use a manual DPD1 test kit). Press OK when value is correct.

Redox Calibration
The redox value shows the oxidation/reduction potential and is used to determine the sterility of the water. Adjusting the ideal redox level (setpoint) is the last step in the system start up sequence.

  • 10.41 Calibration of the redox probe: Recommended every month.
  • 10.42 Calibration with buffer (buffer solution 465 mV): Follow the instructions that appear on the display
  • 10.44 Manual calibration : Not recommended!

Conductivity calibration

VARIABLE SPEED PUMP

FILTRATION

Manual mode

Only with an external pump controller unit

  1. Filtration:
    Configuration control of the filter pump. To set, select Filtration and confirm by pressing OK. The mode selection is done in Mode line with the plus/minus keys.

  2. Manual:
    Manually turns ON/OFF the filtration process. No timing or additional functions. The State (Status) line indicates whether the filtration pump is ON.

Automatic mode

Without an external pump controller unit

Automatic
In this mode the filtration is controlled by up to 3 timers.  We highly recommend to run your system in a 24/7 mode with a variable speed pump.

For example: During the night time (6:00 until 24:00 & 0:00 until 10:00) in low speed, during day time (10:00 until 6:00) in medium speed. To set the ON/OFF times select with the up/down keys in the timer line you want to change (1-3 ). The plus/minus keys open the selected start time field. Set the time with plus/minus keys. Scroll with the up key to the minute field and set it up with plus/minus keys. To confirm press OK and to cancel press return/escape.

Backwash: See chapter 13

Heating mode

  • Heating : This mode acts equally to the automatic mode, but besides it includes the option to work on a relay to control the temperature. The system works with a hysteresis of 1 degree (example: the setting temperature is 23° C, the system will activate itself when the temperature goes below 22° C and will not stop before it passes 23° C).
    Use the plus/minus keys to set the desired temperatures

    • Clima OFF: The heating only works within the set filtration periods.
    • Clima ON: Keeps the filtration working when the filtration period is finished if the water temperature is below the setting temperature.

Note: Mode only visible if the temperature probe is installed and/or heating is activated in the “Installer Menu”.

AUTOMATIC BACKWASH

Backwash Mode with Besgo Valve: The DA-GEN is configured for automatic backwash with Besgo. Use AUX 2!

  • Mode : Choose Auto
  • Start : Choose starting time
  • Interval : Set backwash time in seconds (Recommendation: min. 240 seconds with AFM®, min. 300 seconds with Sand)
  • Freq.: Choose frequency (at least weekly)
  • Test

LIGHTING

  • 14.1 Lighting
    14.2 Manual Mode (ON/OFF). You can activate the light also by pressing the short cut. You can set a timer, after that the light will go out.

  • 14.3 Automatic Mode: Switches lights according to timer settings. Additional you can switch the lights on by pressing the Shortcut button

  • 14.5 LED spotlight: In case of installallation of RGB LED lights in the pool, you can change the color of the lights in the pool. Select the length of time in seconds in Pulse length and then press Next Program option to apply the pulse. Refer to your LED spotlight manual to set its different colors.

WIFI SETTINGS

Installation Advice- Connecting the WIFI to the DA-GEN
Open the WIFI box and unscrew the cabels in the box. Put the cable through the cable hole at the DA-GEN controller box and plug it in. Connect the cable then again in the WIFI box and close it Do not remove the cables from the plug!

  • 15.1 Internet: Once the WiFi module is connected, restart the unit. The internet option will appear in the settings menu.
  • 15.2 WiFi: Select WiFi to scan the available networks accessible to the module. The search will be done automatically.
  • 15.3 Select the desired network accessible to the WiFi module.
  • 15.4 Enter the password in the pop-up keyboard. Scroll up and down with the up/down keys and left to right with the plus/minus keys. To select a letter press OK.
  • 15.5 Enter AP: If you do not find your Network in the automatic mode, then you can enter the network name manually. Check first if the network works on other devices.
  • 15.6 Status : Check the status of your connection.
  • 15.7 Test connection : Check if your connection has been successfully established. Once the WiFi module is connected to the network with both lights ON, enter in www.DA-GEN.com. Access the Register option and enter all the data requested.
  • 15.10 – 15.13 The system node ID that you will need for the registration progress is located under System settings > System info > Power module

Upon completion of the process, you will have total control of your pool, will be able change parameters such as setpoints, filtration hours and turn ON/OFF any auxiliary relays.

Attention : If the DA-GEN was once registered at vistapool.com it needs to be removed there by the manufacturer before you can register it at DA- GEN.com. Please contact your dealer.

COVER

Cover: If the DA-GEN is runned with a frequency controlled pump and if it is connected to the pool cover, the filtration speed will automaticly go to «medium» when the cover is opened. (Please check the filtration speed in Chapter E). Set the Reduction value to 0%!

How to install: If the cover is open, the contact has to be closed and vice versa

AUXILIARY RELAYS

  • 18.2 It is possible to control up to 4 extra auxiliary relays (water attractions, dosing pumps, etc). In default setting you will only see two available relays: APF® and ACO®: (AUX 1 and 3). AUX 2 is reserved for the Besgo Valve and AUX 4 for the heating and therefore not shown here. If you do not have a heating, you can deactivate it (Chapter C & D in the service manual) and you will get a additional free relay (AUX4).
  • 18.3 Manual mode (ON/OFF).
  • 18.4 Automatic mode : ON/OFF according to a timer that adjusts the start and end of the program. The timers can be configured with a frequency.
  • 18.5 Timer mode: Working time is programmed in minutes. Each time the Shortcut on the front panel is pressed, it will start up and run for the programmed time. This function is recommended for timing of air pumps of spas.
  • 18.7 Rename relays : It is posible to rename each auxiliary relay to suit the use you want to assign. By pressing the plus/minus keys, a pop-up keyboard will appear. Scroll up and down with the up/down keys and left to right with the plus/minus keys. To select a letter press the OK.

MAINTENANCE

Weekly checks

  • SALT CONCENTRATION : ~2000 – 4000 ppm TDS
  • CELL : Visual inspection to detect incrustations.

Cleaning the Cell
If necessary, carry out a monthly visual inspection. To clean the cell:

  • Stop the system and close the valves
  • Place the cell for no more than 10 min. in 3% hydrochloric acid or put it for 2 to 4 hrs in normal vinegar or take a high pressure cleaner.
  • Once the incrustations have softened remove with a hose to complete cleaning the cell.

DO NOT USE METALLIC OR SHARP OBJECTS TO REMOVE INCRUSTATIONS. Scratching the edges or surface of the cell will make it vulnerable to chemicals, deteriorate the cell and invalidate the guarantee.

General maintenance

  1. The pool must be vacuumed as usual and the skimmers emptied whenever necessary.
  2. FILTER BACKWASHING: At least once a week for 4 to 5 minutes.
  3. VERY IMPORTANT : Make sure the cell is off while cleaning the filter. If the system controls the filtration pump, use the option “backwash” of the programmed filtration mode. See chapter 13 – automatic backwash
  4. Check regularly the level of your pH and APF® bottle to prevent the dosing pump from running dry.
    pH / Redox / Conductivity – probes: The probes must be cleaned and recalibrated every 2 to 3 months. To clean the probe insert in electrode cleaner. After each clean the probes must be re-calibrated.

Attention : the probes should never dry out and must be kept wet if stored (when emptying the pool for winterizing, make sure to store the probe head in water).

TROUBLESHOOTING

Blank display

  • Check if ON/OFF switch is illuminated.
  • Check the connection wire between display and motherboard.
  • Check the 3.15 A fuse of the device – it could have tripped due to overload.
  • Check the power supply – 230V/50Hz.
  • If the problem persists contact TECHNICAL SERVICE

Hydrolysis does not reach the setpoint value

  • Low water temperature. Check the salt concentration (TDS) in water.
  • Check the cell status (it may be incrusted or calcified).
  • Clean the cell according to the instructions in section 19.
  • Check that the cell is not worn out (remember that the cell is guaranteed for 5,000 hours, approx. 2-3 years of summer usage).

Free chlorine level doesn’t reach the setpoint

  • Increase the filtration hours to 24 hours
  • Increase the hydrolysis level (to 100%).
  • Increase the salt concentraion (TDS) in the water.
  • In an outdoor pool: Add ACO® to the water.
  • Check if the reagents in test kit are in date.
  • Check if the temperature or number of users has risen.
  • If you want a higher chlorine level you have to increase the salt concentration. Attention: Higher risk of corrosion!

Hydrolysis display shows FLOW

  • Check gas and paddle flow detector cable.
  • Clean for incrustations of the paddle flow detector at the top of the cell housing.
  • Check to see if system is free of air (gas detector must always be submerged).

Excess of chlorine in the water

  • Lower hydrolysis cell intensity.
  • If your system includes automatic Redox control, check the Redox setpoint value. Reduce it by 50 to 100 mV.
  • If your system includes free chlorine measurment, adjust the setpoint value.
  • Check redox probe and calibrate it if necessary.
  • Check the free chlorine probe and calibrate it.

Cell incrusted in less than 1 month

  • Very hard waters with a high pH and total alkalinity: balance water adjusting pH and total alkalinity.
  • Check to ensure that the system automatically changes polarity approximately every 300 minutes.
  • Consult with our technical service to consider accelerating the polarity change (auto-cleaning) . WARNING: Accelerating the polarity change decreases the cell life (5,000 hours) proportionally. Don’t go below 200 minutes!
  • If the crust is not foaming when in contact with acid, it might be Struvite. In this case do not use anymore MgCl2, use only NaCl.

Alarm AL3 and pH dosing pump stopped

  • The maximum dosing time (standard 999 min.) is accomplished and the pH-Minus dosing pump stops in order to avoid the acidification of the water.
  • Delete the message and restart the metering. Do the following verifications in order to preclude errors on the device: Verify if the pH probe reading is correct (if not, calibrate the probe or substitute it with a new one); Verify if the acid/base reservoir is full and if the dosing pump is working correctly; Verify the variable speed of the dosing pump.

Rust on metallic components in the pool

  • Metallic elements lack standardised earth connection. Contact an electrician to solve the problem.
  • Rusted components are not stainless steel (minimum 316/V4A/1.4571).
  • The salt concentration (TDS) is too high.
  • Attention Stainless Steel parts must be cleaned regularly

Polarity 1 reaches maximum intensity, but polarity 2 (auto clean) does not reach maximum intensity

  • If the salt concentration is correct : The cell is reaching its end of life. As of this moment check the intensity every 15-20 days.
  • When polarity 2 does not reach intensity of Polarity 1, we recommend substituting the cell for a new one if it happens during the summer period. If it happens during winter, change the cell before the next summer period.

Dosing pump is not working properly

  • Check fuse on the right side of the dosing pump
  • Check (and change) the dosing speed
  • Check electrical connections
  • Check tubes and fittings for leaks
  • Check if injection valve is blocked
  • Check if suction lance/suction weight is blocked
  • Check if error message «TANK» appears. If yes replace bottle, if not check the polarity of the suction lance or replace the suction lance

IMPORTANT NOTES

  • WARNING
    Keep chemical levels in pool as instructed in this manual.

  • CLEANING FILTER
    Very Important: Make sure the cell is off while cleaning/backwashing the filter. If the system controls the filtration pump, use the option “filter cleaning” of the programmed filtration mode. See section 5 – Filtration / Filter Cleaning of the General Installation Guide.

  • VERY IMPORTANT
    Remember that the system needs some time to adapt to your pool (up to 14 days)!

  • SECURITY
    To avoid accidents, children should not handle this product unless supervised by an adult. Children should be supervised at all times when in or near a spa, pool or jacuzzi.

  • HANDLING AND DOSING DANGEROUS CHEMICALS
    Chemicals should be handled with extreme caution. When preparing acid, always add acid to water, never add water to acid, because very dangerous gases may be produced.

ELECTRICAL CONNECTIONS

SERVICE MENU

  1. The 7 available relays can be hooked up to various predefined external devices being controlled by the unit.
  2. The predefined functions are:*
    • pH : Acid pH-pump.
    • Filter : Filtration pump.
    • Light : Pool lights.
    • AUX 1: APF®
    • AUX 2: Besgo Valve
    • AUX 3: ACO®
    • AUX 4: Heat pump or other heating device.

Recommended relay settings.

Note : “NO” will deactivate the predefined parameters and leave the relay available.

SERVICE SETTINGS

  1. Parameters related to external devices
  2. Setting the polarity times. In the case of high alkalinity, the times in Hydro Pol 1 + 2 should be reduced.

TYPE OF PUMP

  • E.2 With the plus/minus keys, select the pump type connected to the system (the default is a standard pump type). The configuration allows the control of two different variable speed pumps (Variable Speed A or Variable Speed B).
    In case of a variable speed pump (A or B), establish the speed when the cover is closed, when the pool heating is connected and/or it controls a backwash filter (Besgo).

  • E.3 Variable Speed Pump A (Hayward® or similar): During the filtration periods, the corresponding relay closes. The filtration pump opens and closes contacts depending on the speed:

    • Common + 1 – Slow speed
    • Common + 1 + 2 – Medium speed
    • Common + 1 + 2 + 3 – Fast speed
  • Variable Speed Pump A B (Speck® or similar): During the filtration periods, the corresponding relay closes. It’s necessary to connect a wire from the filtration relay to the common. The filtration pump opens and closes contacts depending on the speed:

    • Common + 1 – Slow speed
    • Common + 2 – Medium speed
    • Common + 3 – Fast speed

Consult the wiring-schemata in the appendix!

DOSING PUMPS

  • F.2 There are 2 modes for the pH dosing pump:

  • F.3 Normal : Delay – Delay time between detection of incorrect value and the start of dosing.

  • F.4 Repetitive: With the two timers you can programm the ON and OFF time of the dosing pump
    Scaling: Proportional reduction of the dosing time (Pump On time) when getting close to the setpoint. Example 100%: The pump time is reduced proportionally if the pH is closer than 1 to the setpoint. Example 50%: The pump time is reduced proportionally if the pH is closer than 0.5 to the setpoint.

  • F.6 Other pumps : With the two timers you can programm the ON and OFF time of the dosing pump

  • F.8 It corresponds to the behavior of the system after AL3 activation:

    • Ignore – AL3 is not shown in the display.
    • Inform – After the selected interval, the AL3 alarm is displayed.
    • Force stop – After the selected interval, the AL3 alarm is displayed on the display and the dosing pump stops. To reset the alarm and the dosing pump, press .

EXTRA SETTINGS

  • G.2 Gas (0) – The FL1 alarm is only activated by cell’s gas sensor (external flow switch annulled).
    • Siempre ON (1) – The FL1 alarm is never activated (invalidates cell’s gas sensor and external flow switch);
    • Paddle (2) – The FL1 alarm is activated by external flow switch (gas sensor annulled).
    • Paddle or gas (3) – When both cell’s gas sensor and external flow switch are connected, and either of them detects lack of flow, The FL1 alarm is activated. To connect the external flow switch use the FL1 terminal
    • Paddle + Gas (4) – When both cell’s gas sensor and external flow switch are connected, and both of them detects lack of flow, The FL1 alarm is activated. To connect the external flow switch use the FL1 terminal
    • Paddle delay – Delay before FL1 is activated
    • Relay control through flow detection – Manage the FL1 alarm deactivation in case of lack of flow. Recommended option for floculant dosification or similar.

COUNTERS

H.2 Reset counters : There are two levels of working hours counters which log the working hours of the components and devices.

In this service menu the installer can reset the working hour counters on the first level (for example when a new cell is installed). The second level of the working hour counters can only be accessed by the factory.

CONNECTION

I.2 Node address : Used for the configuration of more than 2 user interfaces. For normal operation of the system, keep the value to 1 for this parameter

WEB IP

K.2 Server control and connection port in case there is WIFI Module connected to the system. For the proper functioning of the system, do not change the default values unless you receive a notice from your provider

EU DECLARATION OF CONFORMITY
This declaration of conformity is issued under the sole responsibility of the manufacturer, it is carried out based on tests carried out under the tutelage of the regulatory agency designated by the manufacturer.

  • Product Description: Saline Chlorinator
  • Product reference: OXONG – OX1NG – OX2NG – OX3NG
    • – Declaration of Conformity L.V.D – Low voltage according to LVD Directive 2014/35/EU
    • Harmonized standard : EN 60335-1:2012 + AC:2014 + A11:2014 + A13:2017 + A1:2019 + A14:2019 + A2:2019 + A15:2021; EN 62233:2008 + AC:2008
    • Test number: SAFEKRIPS220601
    • – Declaration of conformity C.E.M – Electromagnetic compatibility in accordance with the ECM Directive 2014/30/EU
    • Harmonized standard : EN 55014-1:2021, EN 55014-2:2021, EN IEC 61000-3-2:2019 + A1:2021, EN 61000-3-3:2013/A1:2019, IEC 61000-3-3:2013/A2:2021, IEC 61000-3-11:2019
    • Test number : EMCOKRIPS220601, EMFIKRIPS220601
    • RoHS Declaration of Conformity under the Directive on the Restriction of the Use of Certain Hazardous
    • Substances in Electrical and Electronic Equipment (RoHS) 2011/65/EU and Amendment (EU) 2015/863
    • Harmonized standard: EN IEC 63000:2018, UNE-EN IEC 62474:2019, EN 62321 * (8 STD determination)
    • Declaration of conformity Ecodesign according to Directive 2009/125/EC

We certify the conformity of the products with the corresponding standards, listed in this declaration, provided that their installation and use comply with the prescribed standards. Place, date: Yuncos, 28/11/2023

Authorized signature: Carlos Martínez Ortega Technology Knowledge Manager, Europe

FAQ

Q: What should I do if the device shows a flow alarm?
A: If the device indicates a flow alarm (FL1 or FL2), check for any blockages in the flow path, ensure proper water flow, and reset the alarm if necessary.

Q: How do I reset alarms related to dosing time exceeded?
A: To reset alarms related to maximum dosing time exceeded (AL3), follow the instructions provided in the manual for alarm reset procedures.

Q: What is the recommended electrical setup for private vs. public installations?
A: For private installations, use a 13 Amp time delay circuit breaker, while for public installations, a 16 A breaker is recommended. Consult a technician for shared power supply setups.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals