WOODPECKER UDS-P LED Ultrasonic Scaler Instruction Manual
- June 26, 2024
- WOODPECKER
Table of Contents
WOODPECKER UDS-P LED Ultrasonic Scaler
Product Usage Instructions
Setting Up:
- Place the ADentol E9uipmcnt on a stable, flat surface near a power outlet.
- Connect the power cord to the equipment and plug it into the power outlet.
- Ensure all components are securely attached and in place before operation.
Operating the Equipment:
To power on the ADentol E9uipmcnt, locate the power button on the control panel and press it. Use the provided user manual to familiarize yourself with the different functions and settings of the equipment.
Maintenance:
Regularly clean the equipment with a soft, damp cloth to remove any dust or debris. Avoid using harsh chemicals that may damage the surfaces.
FAQs
- Q: How do I troubleshoot if the equipment is not turning on?
- A: Check the power connection and ensure the power outlet is functioning. If the issue persists, contact customer support for assistance.
- Q: Can I use the ADentol E9uipmcnt for procedures other than dental applications?
- A: It is recommended to use the equipment only for its intended dental purposes to ensure safety and optimal performance.
FDA CE 0197
www.glwoodpecker.com
The installation and components of equipment
1.1 Instruction
Guilin Woodpecker Medical Instrument Co., Ltd. is a professional manufacturer in researching, developing and producing ultrasonic scalers. The product is mainly used for teeth cleaning and is also an indispensable equipment for teeth disease prevention and treatment. The product UDS-P LED ultrasonic scaler has scaling, perio and endo functions. It contains the following features:
- Optical handpiece, more convenient for clinical operation.
- Automatic frequency tracking ensures that the machine always works on the best frequency and performs more steadily.
- The handpiece is detachable and can be autoclaved to the high temperature of 135°C and the pressure of 0.22MPa.
- Digitally controlled, easy operation and more efficient for scaling.
These features make UDS-P LED become a new generation product in the world dental market.
1.2 Components
- The components of machine are listed in the packing list.
- Product performance and structure
Ultrasonic scaler is composed of electro circuit, water way and ultrasonic transducer.
- Scope of application
Ultrasonic scaler UDS-P LED is used for the dental calculus elimination and root canal treatment.
The main technical specifications
- Power supply input: 220-240V~ 50Hz/60Hz 150mA
- Main unit input: 24V~ 50Hz/60Hz 1.3A
- Output primary tip Vibration excursion: ≤100μm
- Output half-excursion force: <2N
- Output tip Vibration frequency: 28kHz±3kHz
- Output power: 3W to 20W
- Main unit fuse: T1.6AL 250V
- Power supply fuse: T0.5AL 250V
- Water pressure: 0.1bar to 5bar (0.01MPa to 0.5MPa)
- Weight of main unit: 0.68kg
- Weight of power supply: 1 kg
- Operating mode: Continuous operation
- Type of protection against electric shock: Class II
- Degree of protection against electric shock: Type BF applied part
- Degree of protection against harmful ingress of water: Ordinary equipment (IPX0), Protection degree against water (used on foot switch): IPX1
- Applied part of the equipment: handpiece and tip
- Degree of safety of application in the presence of a Flammable Anesthetic Mixture with air or with Oxygen or Nitrous Oxide: Equipment not suitable for being used in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide
1.4 Installation of the main components
Sketch map for installation and connection.
1.4.1. The components of the equipment are showed in picture1(a, b)
- Scaling indicator
- Perio indicator
- Endo indicator
- Option switch
- Increasing power switch
- Decreasing power switch
- Water control knob
- Power switch
- Power indicator
1.4.2. Sketch map for connection of foot switch, power supply and main unit are showed in picture 2
- Foot switch
- Plug of foot switch
- Foot switch socket
- Power cord plug
- Power supply
- Socket for power supply
- Input voltage 220~240V~
1.4.3. Sketch map for connection of water supply system are showed in picture3.
- Connect pure water
- Cap of connector
- Water pipe
- Water connector
1.4.4 Sketch map for connection of detachable handpiece are showed in picture4.
- LED lamp
- Nipple
- Light pipe
- Handpiece decorative ring
- Handpiece
- Index point
- The connector of cable
- Cable
1.4.5 Sketch map for how to install tip and endochuck with wrench are showed in picture5.
- Torque wrench
- Tip
- Handpiece
- Endochuck
- Endo wrench
- File
Product function and operation
2.1 Scaling function
2.1.1 Operation
- Open the packing box, make sure that all the parts and accessories are complete according to the packing list. Take the main unit out of the box and put it on a stable plane.
- Turn the water control switch to the max based on symbol as shown as 3.5.2 [note 1].
- Insert the plug of the foot switch to its socket (picture 2).
- Connect one end of the water pipe to the water entrance, and the other end to the pure water source (picture 3).
- Select a suitable scaling tip as you need, screw it on the handpiece tightly by the torque wrench (picture 5), then connect the handpiece and the connector of cable correctly.
- Insert the plug of the power supply to its socket, then get through to the power (picture 2).
- Switch on the main unit, then the scaling indicator and the first five lights of power regulator shine.
- Step on the foot switch, the tip begins to vibrate, and the LED lamp on the top of the handpiece shines. Release the foot switch, the LED lamp keep shining for 10 seconds.
- The normal frequency is extremely high. Under the normal working state of scaling tips, a light touch and a certain to-and-fro motion will eliminate the tartar without heating. Overexertion and long-time lingering are forbidden.
- Vibrating intensity: Adjust the vibration intensity as you need, generally turn the knob to the middle grade. According to patients
- During the clinical treatment, be sure not to make the end of tip touch the teeth vertically and not to make the tip overexert on the surface of the teeth in case of hurting the teeth and damaging the tip.
- After finishing operation, keep the machine working for 30 seconds on the water supply condition in order to clean the handpiece and the scaling tip.
- Unscrew the scaling tip and pull out handpiece, then sterilize them.
Notice: Don’t pull out the handpiece when the machine is working.
2.1.2 Instruction for main components of detachable handpiece (showed in picture 4)
- Nipple: The nipple can be removed. You can screw off the nipple and clean the pole with alcohol termly.
- Decorative ring: can be disassembled and cleaned with alcohol regularly, can be autoclaved under the high temperature and pressure.
- Handpiece: The main part of the whole handpiece, can be autoclaved under the high temperature and pressure.
- The connector of the cable: Connect the handpiece with the water source and power supply of the main unit.
- LED lamp, Light pipe: Clean them with purified water and sterilize them under the high temperature of 135°C and high pressure of 0.22Mpa.
Notice: Keep the connector dry.
2.1.3 Torque wrench instruction (showed in picture 5)
The torque wrench’s structure is designed in special way which can control the strength of the scaling tip’s installation properly and correctly. It also can guarantee the operator screw or unscrew the scaling tip effectively and keep their hands away from being scratched.
Operation
- Take the scaling tip into the torque wrench; operate as showed in picture 5.
- Tip installation: Hold the handpiece turn the tip toward direction as showed in picture 5 with the torque wrench. Turn two more circles when the tip stops, then the tip is installed.
- Tip uninstallation: Hold the handpiece, turn the wrench toward anti-clockwise direction.
- Sterilize it in sterilizer after each treatment.
- The torque wrench must be cooled naturally after sterilization to avoiding scalding when using next time.
- Keep the torque wrench in a cool, dry and ventilated place and keep it clean.
Precaution
The following sterilizing methods are forbidden.
- Braise in liquor;
- Dip in iodine, alcohol or glutaraldehyde;
- Torreify in oven or microwave oven.
Notice: we are not responsible for any damage of the torque wrench directly or indirectly made by any way in the above items.
2.2 Endo function
2.2.1 Usage process
- Fix endochuck to handpiece by endo wrench.
- Unscrew the screw cap on the endochuck.
- Put the ultrasonic file into the hole in the front of endochuck.
- Screw down the screw cap with endo wrench to tight up the ultrasonic file.
- Press option key, turn to endo function.
- When ultrasonic scaler turns into endo function, only the first lead light is on and the power is at first grade. Put the ultrasonic file into the patient’s root canal slowly, step on the foot switch, then make endo treatment. During the treatment, turn up the power gradually according to the needs.
2.2.2 Notice
- When fixing endochuck, it must be screwed down.
- The screw cap on the endochuck must be screwed down.
- Don’t press it too hard when the ultrasonic file is in the root canal.
- Don’t step on the foot switch until the ultrasonic file is in the root canal.
- The power range is supposed from 1st to 5th grade.
Sterilization and maintenance
3.1 Sterilization of detachable handpiece
3.1.1 Autoclaved under high temperature/pressure:
- 121°C/1bar (0.1MPa)
- 135°C/2.2bar (0.22MPa)
- Pull out the handpiece and unscrew the scaling tip and endochuck after each operation.
- Pack the handpiece with sterile gauze or sterile bag before sterilizing.
- Reuse handpiece after it cools naturally in case of scalding hand.
3.1.2 Notice
- Clear the cleaning liquid on the handpiece with compressed air before sterilization.
- Be sure that the scaling tip has been unscrewed from the handpiece and it cannot be sterilized with others.
- Please notice whether the outer of the handpiece is damaged during the treatment or sterilization, don’t smear any protective oil on the surface of handpiece.
- There are two waterproof “O” rings at the end of handpiece. Please lubricate them with dental lube frequently, as sterilization and repeated pulling and inserting will reduce their life-span. Change a new one once it is damaged or worn excessively.
The following sterilizing methods are forbidden:
- Put handpiece into any liquid for boiling.
- Dip handpiece in disinfectors such as iodine, alcohol and glutaraldehyde.
- Put handpiece into oven or microwave oven for baking.
Specifications
Specification | Details |
---|---|
Power supply input | 220-240V~ 50Hz/60Hz 150mA |
Main unit input | 24V~ 50Hz/60Hz 1.3A |
Output primary tip Vibration excursion | ≤100μm |
Output half-excursion force | <2N |
Output tip Vibration frequency | 28kHz±3kHz |
Output power | 3W to 20W |
Main unit fuse | T1.6AL 250V |
Power supply fuse | T0.5AL 250V |
Water pressure