EQUITEST ITALIANO HT10A Continuity Tester User Manual
- June 9, 2024
- EQUITEST
Table of Contents
EQUITEST ITALIANO HT10A Continuity Tester
PRECAUTIONS AND SAFETY MEASURES
The accessory has been designed in compliance with IEC/EN61557 and IEC/EN61010-1 guidelines relevant to electronic measuring devices. For your safety and to prevent damaging the accessory, please strictly follow the procedures described in this manual and read carefully all notes preceded by symbol . Before and after carrying out measurements, observe the following instructions:
- Do not carry out any measurement in humid environments.
- Do not carry out any measurements in case gas, explosive materials or flammables are present, or in dusty environments.
- Avoid any contact with the circuit under test if no measurement is in progress.
- Avoid any contact with exposed metal parts, with unused measuring probes, circuits, etc.
- Do not carry out any measurement in case you find anomalies in the accessory such as deformation, breaks, substance leaks, absence of display on the screen, etc.
- Pay special attention when measuring voltages higher than 25V AC, since a risk of electrical shock exists.
In this manual, and on the accessory, the following symbols are used
- Warning: observe the instructions given in this manual; improper use could damage the accessory or its components.
- High voltage danger: electrical shock hazard
- Accessory with double insulation
- AC Voltage or Current
- Ground reference
PRELIMINARY INSTRUCTION
CAUTION
- The accessory can be used for RESISTANCE measurements on installations with CAT III 300V to ground
- Do not use the accessory on loads with technical specifications different by the same described in § 7. We invite you to follow the normal safety rules aimed at protecting against dangerous voltages and protecting the accessory against incorrect use
- Check that the rated operating voltage value indicated on the back label should be consistent with the power supply voltage of the system under test
- Do not use the accessory in circuits with voltages and currents higher than the rated ones
- Do not carry out any measurement in case you find anomalies in the accessory such as deformation, breaks, substance leaks, absence of display on the screen, etc.
DURING USE
Please carefully read the following recommendations and instructions:
CAUTION
The Failure to observe the warnings and/or instructions may damage the
accessory and/or its components or be a source of danger for the operator
- Before changing function, disconnect the test leads from the circuit under test
- When the accessory is connected to the circuit under test, never touch any terminal, even if unused
- Avoid measuring resistance if external voltages are present. Even if the accessory is protected, excessive voltage could cause damage
DEFINITION OF MEASUREMENT (OVERVOLTAGE) CATEGORY
Standard “IEC/EN61010-1: Safety requirements for electrical equipment for
measurement, control and laboratory use, Part 1: General requirements” defines
what measurement category, commonly called overvoltage category, is. § 6.7.4:
Measured circuits, reads: circuits are divided into the following measurement
categories:
-
Measurement category IV is for measurements performed at the source of a low-voltage installation.
Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units. -
Measurement category III is for measurements performed on installations inside buildings.
Examples are measurements on distribution boards, circuit breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in fixed installation, and equipment for industrial use and some other equipment, for example, stationary motors with permanent connection to fixed installation. -
Measurement category II is for measurements performed on circuits directly connected to the low-voltage installation.
Examples are measurements on household appliances, portable tools and similar equipment. -
Measurement category I is for measurements performed on circuits not directly connected to MAINS.
Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS-derived circuits. In the latter case, transient stresses are variable; for that reason, the standard requires that the transient withstand capability of the equipment is made known to the user
GENERAL DESCRIPTION
The EQUITEST accessory is designed for carrying out the continuity measurement of protective and equipotential conductors with test current> 10A AC in combination with the following models of test instruments (MASTER unit) of the HT family. EQUITEST can be used in combination with the following HT safety testers:
*Model ()** | Connection | FW version |
---|---|---|
COMBI519 |
Optical with cable C2050
|
2.01 (or higher)
COMBI521
EASYTEST
SUPERCOMBIS
(*) The list of available models can be changed without notice. In case of doubt contact the after-sales service
The accessory uses the 4-terminal measurement method that allows self-
compensation of the resistance of the test cables and therefore their
extension without the need to
perform any preliminary calibration
The followed features are available:
- Continuity of protective and equipotential conductors with current> 10A AC and no-load voltage <24V AC
- Power supplies available: 110V ± 10% or 230 / 240V ± 10%
- LED indications for measurement execution control
- Connection via optical/USB cable to MASTER units
- Detection of Manual and Automatic operating modes
- WiFi connection to mobile devices via the HTAnalysis APP
- Input protection fuse
PREPARATION FOR USE
INITIAL CHECKS
Before shipping, the accessory has been checked from an electric as well as a
mechanical point of view. All possible precautions have been taken so that the
accessory is delivered free of damage. However, a thorough check of the
accessory is recommended to detect any damage suffered during transport. In
case anomalies are found, immediately contact the forwarding agent. Also check
whether the packaging contains all parts indicated in § 7.4. In case of
discrepancy, please contact the Dealer. Should the accessory be returned,
please follow the instructions given in § 8.
CAUTION
Should the accessory be used differently from what is specified by the
manufacturer, the protection provided may be impaired
POWER SUPPLY
The accessory is powered by 230VAC, 50/60Hz (default condition) or 110VAC,
50/60Hz via the supplied plug cable.
CAUTION
Check that the rated operating voltage value indicated on the back label
should be consistent with the power supply voltage of the system under test
STORAGE
To guarantee accurate measurement, after a long storage time, wait for the
accessory to come back to normal condition (see § 7.3).
NOMENCLATURE
DESCRIPTION OF THE ACCESSORY
- Input for connection to MASTER instrument via C2050 cable
- WiFi LED
- POWER LED
- STATUS LED
- Indication of LED status
- Input terminal C1
- Input terminal C2
- Input terminal P1
- Input terminal P2
- Plug cable for connection to mains supply
DESCRIPTION OF MEASURING LEADS
STATUS DESCRIPTION OF LED INDICATORS
LED | Description |
---|
POWER
| This LED identifies the power on status of the accessory. The following situations are possible:
Ø Off à accessory not powered
Ø Green à accessory powered with correct voltage
Ø Red à accessory powered with incorrect voltage
WiFi
| This LED identifies the connection status of the internal WiFi module. The following situations are possible:
Ø Off à no WiFi connection
Ø Green steady à WiFi connection active
Ø Green flashing à accessory waiting for WiFi connection
Ø Red à internal WiFi module malfunction (contact the after-sales service)
STATUS
| This LED identifies the type of measurement performed in connection with the MASTER unit (see relevant user manual). The following situations are possible:
Ø Green à operation in Manual mode
Ø Orange à operation in Autotest mode
Ø Red à there are connection errors (contact the after-sales service)
OPERATING INSTRUCTIONS
CONTINUITY MEASUREMENT OF PROTECTIVE CONDUCTORS WITH 10A
CAUTION
- The power supply voltage of the accessory must be: 230/240V ± 10% (207V ÷ 265V) or 110V ± 10% (100V ÷ 120V). Do not use external power supply voltages external from the above ranges to avoid damaging the instrument
- Check that the rated operating voltage value indicated on the back label should be consistent with the power supply voltage of the system under test
- Connect the accessory to the power supply using the supplied network cable (see Fig. 1 – part 10). The POWER LED is steady green to identify the correct power supply voltage, the STATUS LED is steady green to identify the operation of the accessory in Manual mode (see user manual of the MASTER instrument) and the WiFi LED is on in green flashing color to identify that the accessory is trying to make (if available) a WiFi connection with an external mobile device via the HTAnalysis APP (refer to the HTAnalysis instructions for managing the operation)
- Connect the accessory to the MASTER instrument (see Fig. 3) by using the C2050 cable supplied (see Fig. 1 – part 1)
- Check the correct detection of the accessory on the MASTER instrument (see relevant user manual)
- Connect the two black cables of the C7000 cable set to the input sockets P1 and P2 and the two red cables to the input sockets C1 and C2 (see Fig. 3)
- Connect the alligator clips of the C7000 cable set to the ends of the object under test (see Fig. 3)
- Press the GO/STOP key on the MASTER instrument to activate the test in Manual mode. The accessory provides the test current >10A AC and the measurement result is shown on the display of the MASTER instrument (see related user manual)
- By selecting the Automatic mode on the MASTER instrument (measurement activated without using the GO/STOP key) check that the STATUS LED should be steady orange and connect the alligator clips to the ends of the object under test (see Fig. 3) to automatically activate the test
- To save and manage the result, refer to the user manual of the MASTER instrument
MAINTENANCE
GENERAL
During use and storage, respect the recommendations listed in this manual to
avoid possible damage or dangers during use. Do not use the accessory in
environments with high humidity levels or high temperatures.
CAUTION
Only skilled and trained technicians should perform maintenance operations.
Before carrying out maintenance operations, disconnect all cables from the
input terminals.
CLEANING THE ACCESSORY
Use a soft and dry cloth to clean the accessory. Never use wet cloths,
solvents, water, etc.
END OF LIFE
CAUTION : the symbol on the accessory indicates that the appliance and its accessories must be collected separately and correctly disposed of.
TECHNICAL SPECIFICATIONS
TECHNICAL CHARACTERISTICS
Accuracy calculated as ± [%reading + (num. dgt) * resolution] at 23°C ± 5°C,
<80%RH
Continuity of protection conductors
Range [ W ] | Resolution [ W ] | Accuracy |
---|---|---|
0.001 ¸ 1.999 | 0.001 | ±(1.0%rdg + 2dgt) |
- Test current: 10A AC (max test cable resistance 0.7)
- Open voltage: 4< Vo < 24V AC
- Measurement time: 10 periods (@ 50Hz), 12 periods (@60Hz)
- Measurement method: 4-wires
- Overheating : after at least 20 consecutive tests
Test current
Range [A] | Resolution [A] | Accuracy |
---|---|---|
0.00 ¸ 19.99 | 0.01 | ±(1.0%rdg + 2dgt) |
REFERENCE GUIDELINES
- Safety: IEC/EN61010-1, IEC/ EN61010-2-030, IEC/EN61557-1
- EMC: IEC/EN61326-1, IEC/EN61326-2-2
- RED: ETSI EN300328, ETSI EN303446-1, ETSI EN301489-17
- EMC environment of use: portable, Class B, Group 1
- Insulation: double insulation
- Pollution level: 2
- Measurements: IEC/EN61439-1, IEC/EN60204-1, IEC/EN62305-3
- Measurement category: CAT III 300V to ground and between inputs
GENERAL CHARACTERISTICS
Power supply
- Main’s supply: 230V/240V ±10%, 50/60Hz or 110V ±10%, 50/60Hz
- Power consumption: max 70W (@230V, 300mA) (@110V, 600mA)
- Fuse protection: Time-Leg 250V/1A (5x20mm)
CAUTION
Check that the rated operating voltage value indicated on the back label
should be consistent with the power supply voltage of the system under test
External communication
- MASTER instrument interface: optical cable C2050
- Mobile devices interface: WiFi (via APP HTAnalysis)
- Internal status indications: two colors LEDs
Mechanical characteristics
- Dimensions (L x W x H): 210 x 115 x 60mm (8 x 5 x 2in)
- Weight (with integrated cable): approx. 1kg (32ounces)
- Mechanical protection: IP40
Environmental conditions
- Working temperature: 0°C 40°C (32°F 104°F)
- Working humidity: <80%RH
- Storage temperature: -10°C 60°C (14°F 140°F)
- Storage humidity: <80%RH
- Max operating altitude: 2000m (6562ft)
This accessory compliance with the requirements of Low Voltage Directive
2014/35/EU (LVD) and of EMC Directive 2014/35/EU
This accessory satisfies the requirements of European Directive 2011/65/EU
(RoHS) and 2012/19/EU (WEEE)
ACCESSORIES
SeeSee enclosed packing list the enclosed packing list
ASSISTANCE
WARRANTY CONDITIONS
This accessory is warranted against any material or manufacturing defect, in
compliance with the general sales conditions. During the warranty period,
defective parts may be replaced. However, the manufacturer reserves the right
to repair or replace the product. Should the accessory be returned to the
After-sales Service or to a Dealer, transport will be at the Customer’s
charge. However, shipment will be agreed in advance. A report will always be
enclosed to a shipment, stating the reasons for the product’s return. Only use
original packaging for shipment. Any damage due to the use of non-original
packaging material will be charged to the Customer. The manufacturer declines
any responsibility for injury to people or damage to property.
The warranty shall not apply in the following cases:
- Repair and/or replacement of accessories and battery (not covered by warranty).
- Repairs that may become necessary as a consequence of an incorrect use of the accessory or due to its use together with non-compatible appliances.
- Repairs that may become necessary as a consequence of improper packaging.
- Repairs which may become necessary as a consequence of interventions performed by unauthorized personnel.
- Modifications to the accessory performed without the manufacturer’s explicit authorization.
- Use not provided for in the accessory’s specifications or in the instruction manual.
The content of this manual cannot be reproduced in any form without the manufacturer’s authorization.
Our products are patented, and our trademarks are registered. The manufacturer reserves the right to make changes in the specifications and prices if required by improvements in technology.
ASSISTANCE
If the accessory does not operate properly, before contacting the After-sales
Service, please check the conditions. Should the accessory still operate
improperly, check that the product is operated according to the instructions
given in this manual. Should the accessory be returned to the After-sales
Service or to a Dealer, transport will be at the Customer’s charge. However,
shipment will be agreed in advance. A report will always be enclosed to a
shipment, stating the reasons for the product’s return. Only use original
packaging for shipment. Any damage due to the use of non-original packaging
material will be charged to the Customer.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>