UNI-T USB Datalogger User Manual
- June 6, 2024
- UNI-T
Table of Contents
UNI-T USB Datalogger
Introduction
The USB datalogger (Hereinafter referred to as “logger”) is a low power consumption, high-accuracy temperature & humidity device. It is designed with IP65 dust/water protection, large data storage capacity, and features such as auto storage, USB transfer, real-time display, data export, etc. It is widely used in food processing, cold-chain transportation, warehouse storage, and other applications.
Accessories
User manual………………………………………..1 piece
Battery…………………………………………………1 piece
Holder ………………………………………………..1 piece
Screw…………………………………………………..2 pieces
Safety information
- Check if the logger is damaged before use.
- Replace the battery when the logger displays ” “.
- If the logger is found abnormal, please stop using and contact your seller.
- Do not use the logger near explosive gas, volatile gas, corrosive gas, vapor and powder.
- Do not charge the battery.
- 3.0V CR2032 battery is recommended.
- Install the battery according to its polarity.
- Take out the battery if the logger is not used for a long time.
Structure (Figure 1)
No. | Description |
---|
1
| USB cover
2
| Indicator (Green light: logging, red light: alarm)
3
| Display screen
4
| Start/select
5
| Stop/switch humidity and temperature
6
| Holder
Display (Figure 2)
No | Description |
---|
1
| Start
2
| Stop
3
| Minimum value
4
| Marking
5
| Mean kinetic temperature
6
| Number of sets
7
| Temperature unit
8
| Low battery
9
| Humidity unit
10
| Temperature & humidity display area
11
| Time display area
12
| Set a fixed time/delay
13
| Alarm due to abnormal logging
14
| Not alarm
15
| Set lower value
16
| Set upper value
17
| Maximum value
Setting
USB communication
- Download the instruction and PC software according to the attached file, then, install the software step by step.
- Insert the logger into USB port of PC, the main interface of logger will display “USB”. After the computer identifies the USB, open the software to set parameters and analyze the data.
- Open the computer software to browse and analyze data. As for how to use the software, users can click the help option on the operation interface to find “software manual”.
Parameter configuration
Model| The computer identifies the logger model automatically.|
---|---|---
Unit| °C or °F
ID| Users can set the ID, the range is 0~255
SN| Factory number
Description| Users can add descriptions in numbers or letters.
Alarm delay| The logger will not alarm when the temperature is beyond the
upper/lower value, but it will alarm when the exceeding time is longer than
the delay time.
Single alarm| The logger alarms if the continuous alarm time is greater than
the delay time. The logger will not alarm if the reading resumes normal within
the delay time.
accumulated alarm| The logger alarms if the accumulated alarm time is greater
than the delay time.
Temperature and humidity Adjusting| Adjust the logger error by changing the
calibration value.
Sampling interval| 10 seconds to 24 hours.
Sampling delay| Start logging after the delay time. 0 to 240 minutes.
Start logging| Press the button to start, start immediately through the
software, start at a fixed time.
Operations
Starting the logger
There are three starting modes:
- Press the button to start the logger.
- Start logging through the software.
- Start logging at preset fixed time.
- Mode 1: Long press the start button for 3 seconds in main interface to start logging. This start mode supports start delay, if delay time is set, the logger will start logging after a delayed time.
- Mode 2: Start logging through the software: On PC software, when parameter setting is completed, the logger will start logging after user unplugs the logger from the computer.
- Mode 3: Start the logger at preset fixed time: On PC software, when parameter setting is completed, the logger will start logging at preset time after user unplugs the logger from the computer. Mode 1 now is disabled.
Stopping the logger
When the logger reaches up to the maximum capacity, it will stop logging, the “Press the button to stop” mode supports the “Stop logging through the computer software” mode.
There are two stop modes:
- Press the button to stop.
- Stop logging through the software.
- Mode 1: In main interface, long press stop button for 3 seconds to stop the logger, this function will be disabled if “Stop logging through the software” is set.
- Mode 2: After connecting the logger to the computer, click the stop icon on the main interface of the computer to stop logging.
Function Interface 1
Short press stop button to switch between temperature and humidity in the main interface. Users can short press stop button at any time to go back to the main interface. If no button is pressed for 5 seconds, the logger will enter the power-saving mode.
Marking
When the device is in logging state, long press start button for 3 seconds to mark the current data for future reference, the mark icon and current value will flash 3 times, total number of mark value is 10.
Function Interface 2
In the main interface, press the start button and stop button together for 3 seconds to enter the Function Interface 2, short press start button to view: Y/M/D, device ID, maximum numbers of remaining storage groups, numbers of marking groups.
Alarm State
When the logger is operating, Alarm disabled: Green LED flashes every 10 seconds and main interface displays “”. Alarm enabled: Red LED flashes every 10 seconds and main interface displays “×”. No LED lights when the logger is in stopping state.
Viewing data
Users can view the data in stop or operating state.
- View the data in stop state: Connect the logger to the PC, if the LED flashes at this time, the PDF report is being generated, do not unplug the logger at this time. After the PDF report is generated, users can click the PDF file to view and export the data from the computer software.
- View the data in operating state: Connect the logger to the PC, the logger will generate a PDF report for all previous data, at the same time, the logger will continue logging data and it can only generate a PDF report with new data the next time.
Maintenance
Battery replacement (Figure 4)
Replace the battery with the following steps when the logger displays ” “.
- Rotate the battery cover counter-clockwise.
- Install CR2032 battery and waterproof rubber ring.
- Install the cover in arrow direction and rotate it clockwise.
Cleaning the logger
Wipe the logger with soft cloth or sponge dipped with a little water, detergent, soapy water. Do not clean the logger with water directly to avoid damage to the circuit board.
Specification
Function | UT330T | UT330TH |
---|---|---|
Range | Accuracy | Accuracy |
Temperature | -30.0℃ -20.1℃ | ±0.8℃ |
-20.0℃ 40.0℃ | ±0.4℃ | |
40.1℃ 70.0℃ | ±0.8℃ | |
Humidity | 0 99.9%RH | / |
Resolution | Temperature: 0.1°C; Humidity: 0.1%RH | |
Logging capacity | 64000 sets | |
Logging interval | 10s 24h | |
Unit/alarm setting | The default unit is °C. Alarm types include single and |
accumulated alarm, the default type is single alarm. Alarm type can be changed
through the PC soft.
Start mode| Press the button to start the logger or start the logger through
the software (Immediately/delay/at fixed time).
Logging delay| 0min 240min, it defaults at 0 and can be changed through the PC
software.
Device ID| 0 255, it defaults at 0 and can be changed through the PC software.
Alarm delay| 0s 10h, it defaults at 0 and can be changed through the PC
software.
Screen off time| 5s
Battery type| CR2032
Data export| View the real-time data and export the data (PDF, XLS) through
the computer
Working time| 140 days at an test interval of 15min.
Working temperature & humidity| -30°C ~ 70°C, ≤99%, non-condensable
Storage temperature| -50°C~70°C
Protection degree| IP65