microlife BP B6 Connect Bluetooth Blood Pressure Monitor Instruction Manual
- June 6, 2024
- microlife
Table of Contents
- Preparation
- Before each measurement
- Microlife BP B6 Connect
- The appearance of the Atrial Fibrillation Indicator for Early Detection
- Using the Device for the First Time
- Checklist for Taking a Reliable Measurement
- Taking a Blood Pressure Measurement
- Data Memory
- Battery Indicator and Battery change
- Using a Mains Adapter
- Bluetooth® Function
- PC-Link Functions
- Error Messages
- Safety, Care, Accuracy Test, and Disposal
- Guarantee
- Technical Specifications
- Guarantee Card
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
BPB6 Connect
Bluetooth® Blood Pressure Monitor
Instruction Manual
Preparation
|
---|---
Sit on a back-supported chair and keep your legs uncrossed| Avoid thick or
close-fitting garments on the upper arm.
|
Place the artery mark on the cuff over your artery.| Fit the cuff closely, but
not too tight.
|
Position the cuff 1-2 cm above your elbow.| Keep your arm still and do not
speak during the measurement.
Before each measurement
Avoid eating, bathing, smoking or caffeine (approx. 30 min).
Avoid activity and relax for 5-10 min.
Measure before medication intake.
Microlife BP B6 Connect
1 ON/OFF button | 16 Systolic Value |
---|---|
2 Display | 17 Diastolic Value |
3 M-button (memory) | 18 Pulse Rate |
4 Time Button | 19 Battery Display |
5 User Button | 20 Cuff Fit Check |
-A: Suboptimal Cuff Fit
-B: Arm Movement Indicator «Err 2»
-C: Cuff Pressure Check «Err 3»
6 Cuff Socket
7 MAM Switch| 21 Cuff Signal Indicator «Err 1»
8 Traffic Light Indicator| 22 Irregular Heartbeat (IHB) Symbol
9 USB Port| 23 Atrial Fibrillation Indicator (AFIB)
10 Mains Adapter Socket| 24 User Indicator
11 Battery Compartment| 25 MAM Mode
12 Cuff| 26 Stored Value
13 Cuff Tube| 27 Clinical Blood Pressure Average «MyBP»
14 Cuff Connector| 28 Date/Time
Display| 29 Pulse Indicator
15 Active Bluetooth| 20 Average Indicator «MyCheck»
| Read the instructions carefully before using this device.
---|---
| Type BF applied part
| Keep dry
Intended use:
This oscillometric blood pressure monitor is intended for measuring non-
invasive blood pressure in people aged 12 years or older.
It is clinically validated in patients with hypertension, hypotension,
diabetes, pregnancy, pre-eclampsia, atherosclerosis, end-stage renal disease,
obesity and the elderly.
The device can detect an irregular pulse suggestive of Atrial Fibrillation
(AF). Please note that the device is not intended to diagnose AF. A diagnosis
of AF can only be confirmed by ECG. The patient is advised to see a physician.
Dear Customer,
This device was developed in collaboration with physicians and clinical tests
were carried out to prove its measurement accuracy to be of a very high
standard.*
Microlife AFIBsens is the world’s leading digital blood pressure measurement
technology for the detection of atrial fibrillation (AF) and arterial
hypertension. These are the two top risk factors of getting a stroke or heart
disease. It is important to detect AF and hypertension at an early stage, even
though you may not experience any symptoms. AF screening in general and thus
also with the Microlife AFIB algorithm is recommended for people 65 years and
older. The AFIB algorithm indicates that atrial fibrillation may be present.
For this reason, it is recommended that you visit your doctor when the device
gives an AFIB signal during your blood pressure measurement. The AFIB
algorithm of Microlife has been clinically investigated by several prominent
clinical investigators and showed that the device detects patients with AFIB
at a certainty of 97-100%. 1,2 If you have any questions, problems, or want to
order spare parts please contact your local Microlife-Customer Service. Your
dealer or pharmacy will be able to give you the address of the Microlife
dealer in your country. Alternatively, visit the internet at
www.microlife.com where you will find a wealth of
invaluable information on our products.
Stay healthy — Microlife AG!
- This device uses the same measuring technology as the award-winning «BP 3BTO-A» model tested according to the British and Irish Hypertension Society (BIHS) protocol.
¹Kearley K, Selwood M, Van den Bruel A, Thompson M, Mant D, Hobbs FR et al.: Triage tests for identifying atrial fibrillation in primary care: a diagnostic accuracy study comparing single-lead ECG and modified BP monitors. BMJ Open 2014; 4:e004565.
²Wiesel J, Arbesfeld B, Schechter D: Comparison of the Microlife blood pressure monitor with the Omron blood pressure monitor for detecting atrial fibrillation. Am J Cardiol 2014; 114:1046-1048.
The appearance of the Atrial Fibrillation Indicator for Early Detection
(Active only in MAM mode)
This device is able to detect atrial fibrillation (AF). This symbol BM indicates that atrial fibrillation was detected during the measurement. Please refer to the next paragraph for information regarding the consultation with your doctor.
Information for the doctor on frequent appearance of the atrial fibrillation indicator This device is an oscillometric blood pressure monitor that also analyses pulse irregularity during measurement. The device is clinically tested. The AFIB symbol is displayed after the measurement if atrial fibrillation occurred during the measuring. If the AFIB symbol appears after having performed a full blood pressure measurement episode (triplicate measurements), the patient is advised to perform another measurement episode (triplicate measurements). If the AFIB symbol appears again, we recommend the patient seek medical advice. If the AFIB symbol appears on the screen of the blood pressure monitor, it indicates the possible presence of atrial fibrillation. The atrial fibrillation diagnosis however, must be made by a cardiologist based on ECG interpretation.
Keep the arm still during measuring to avoid false readings.
This device may not or wrongly detect atrial fibrillation in people with
pacemakers or defibrillators.
In the presence of atrial fibrillation, the diastolic blood pressure value may
not be accurate.
In the presence of atrial fibrillation using MAM-mode is recommended for more
reliable blood pressure measurement.
What is Atrial Fibrillation (AF)?
Normally, your heart contracts and relaxes to a regular beat. Certain cells in
your heart produce electrical signals that cause the heart to contract and
pump blood. Atrial fibrillation occurs when rapid, disorganized electrical
signals are present in the heart’s two upper chambers, called the atria;
causing them to contract irregularly (this is called fibrillation). Atrial
fibrillation is the most common form of heart arrhythmia. It often causes no
symptoms, yet it significantly increases your risk of stroke. You’ll need a
doctor to help you control the problem.
Who should be screened for Atrial Fibrillation?
AF screening is recommended for people over 65 years of age, since the chance
of having a stroke increases with age. AF screening is also recommended for
people from the age of 50 years who have high blood pressure (e.g. SYS higher
than 159 or DIA higher than 99) as well as those with diabetes, coronary heart
failure, or for those who have previously had a stroke.
In young people or in pregnancy AF screening is not recommended as it could
generate false results and unnecessary anxiety. In addition, young individuals
with AF have a low risk of getting a stroke as compared to elder people.
Risk factors you can control
Early diagnosis of AF followed by adequate treatment can significantly reduce
the risk of getting stroke. Knowing your blood pressure and knowing whether
you have AF is the first step in proactive stroke prevention. For more
information visit our website:
www.microlife.com/afib.
Using the Device for the First Time
Inserting the batteries
After you have unpacked your device, first insert the batteries. The battery
compartment AK is on the bottom of the device. Insert the batteries (4 x 1.5
V, size AA), thereby observing the indicated polarity.
Setting the date and time
- After the new batteries are fitted, the year number flashes in the display. You can set the year by pressing the M-button 3. To confirm and then set the month, press the time button 4.
- Press the M-button to set the month. Press the time button to confirm and then set the day.
- Follow the instructions above to set the day, hour, and minutes.
- Once you have set the minutes and pressed the time button, the date and time are set and the time is displayed.
- If you want to change the date and time, press and hold the time button down for approx. 3 seconds until the year number starts to flash. Now you can enter the new values as described above.
Selecting the correct cuff
Microlife offers different cuff sizes. Select the cuff size to match the
circumference of your upper arms (measured by close-fitting in the center of
the upper arm).
Cuff size | for the circumference of the upper arm |
---|---|
S | 17 – 22 cm |
M | 22 – 32 cm |
M – L | 22 – 42 cm |
L | 32 – 42 cm |
L – XL | 32 – 52 cm |
Only use Microlife cuffs.
- Contact your local Microlife Service if the enclosed cuff AL does not fit.
- Connect the cuff to the device by inserting the cuff connector into the cuff socket 6 as far as it will go.
If you buy a spare Microlife cuff, please remove the cuff connector 14 from the cuff tube 13 from the cuff supplied with the original device and insert this cuff connector into the tube of the spare cuff (valid for all cuff sizes).
Selecting the user
This device allows to the storage of the results for 2 individual users.
- Select the intended user (user 1 or user 2 BN) by pressing the user button 5.
Before each measurement, ensure that the correct user is selected.
Selecting standard or MAM mode
Before each measurement, select standard (single measurement) or MAM mode
(automatic triple measurement). In MAM mode, 3 measurements are automatically
taken in succession and the result is then automatically analyzed and
displayed. Because the blood pressure constantly fluctuates, a result obtained
in this way is more reliable than when a single measurement is performed.
-
To select MAM mode, slide the MAM switch 7 upwards to position «3» until the MAM-symbol BO appears on the display.
To change to standard mode (single measurement), slide the MAM switch downwards to position «1». -
The bottom, right-hand section of the display shows a 1, 2 or 3 to indicate which of the 3 measurements is currently being taken.
-
There is a break of 15 seconds between the measurements.
A count down indicates the remaining time. -
The individual results are not displayed. Your blood pressure will only be displayed after all 3 measurements are taken.
-
Do not remove the cuff between measurements.
-
If one of the individual measurements was questionable, a fourth one is automatically taken.
AF detection is only activated in MAM mode.
Checklist for Taking a Reliable Measurement
-
Avoid activity, eating or smoking immediately before the measurement.
-
Sit down on a back-supported chair and relax for 5 minutes.
Keep the feet flat on the floor and do not cross your legs. -
Always measure on the same arm (normally left). It is recommended that doctors perform double-arm measurements on a patient’s first visit in order to determine which arm to measure in the future. The arm with the higher blood pressure should be measured.
-
Remove close-fitting garments from the upper arm. To avoid constriction, shirt sleeves should not be rolled up – they do not interfere with the cuff if they are laid flat.
-
Always ensure that the correct cuff size is used (marking on the cuff).
• Fit the cuff closely, but not too tight.
• Make sure that the cuff is positioned 1-2 cm above the elbow.
• The artery mark on the cuff (ca. 3 cm long bar) must lie over the artery which runs down the inner side of the arm.
• Support your arm so it is relaxed.
• Ensure that the cuff is at the same height as your heart.
Taking a Blood Pressure Measurement
- Select standard (single measurement) or MAM mode (automatic triple measurement): see details in chapter «2.».
- Pressez le bouton ON/OFF 1 pour démarrer la mesure.
- The cuff will now pump up automatically. Relax, do not move, and do not tense your arm muscles until the measurement result is displayed. Breathe normally and do not talk.
- The cuff fit check BT on the display indicates that the cuff is perfectly placed. If the icon BT-A appears, the cuff is fitted suboptimally, but it is still ok to measure.
- When the correct pressure is reached, the pumping stops and the pressure falls gradually. If the required pressure was not reached, the device will automatically pump some more air into the cuff.
- During the measurement, the pulse indicator BS flashes in the display.
- The result, comprising the systolic AP and the diastolic AQ blood pressure and the pulse rate AR is displayed. Note also the explanations on further display symbols in this booklet.
- When the device has finished measuring, remove the cuff.
- Switch off the device. (The monitor does switch off automatically after approx. 1 min.).
AF detection is only activated in MAM mode.
You can stop the measurement at any time by pressing the ON/OFF button (e.g. if you feel uneasy or have an unpleasant pressure sensation).
This monitor is specially tested for use in pregnancy and pre-eclampsia. When you detect unusual high readings in pregnancy, you should measure after a short while again (eg. 1 hour). If the reading is still too high, consult your doctor or gynecologist. In pregnancy, the AFIB symbol can be ignored.
Manual inflation
In case of high systolic blood pressure (e.g. above 135 mmHg),
it can be an advantage to set the pressure individually. Press the ON/ OFF
button after the monitor has been pumped up to a level of approx. 30 mmHg
(shown on the display). Keep the button pressed until the pressure is about 40
mmHg above the expected systolic value – then release the button.
How not to store a reading
As soon as the reading is displayed press and hold the ON/OFF button 1 until
«M» BP is flashing. Confirm to delete the reading by pressing the time
button 4.
«CL» is displayed when the reading is deleted from the memory
successfully.
How do I evaluate my blood pressure?
The LED traffic light indicator on the left-hand side of display 8 indicates
within which range the measured blood pressure lies. The value is either
within the optimum (green), elevated (yellow), or high (red) range. The
classification corresponds to the following ranges defined by international
guidelines (ESH, ESC, JSH). Data in mmHg.
Range | Systolic | Diastolic | Recommendation |
---|---|---|---|
1. | blood pressure | ||
too high | > 135 | > 85 | Seek medical |
advice
2.| blood pressure
elevated| 130 – 134| 80 – 84| Self-check
3.| blood pressure
normal| < 130| < 80| Self-check
The higher value is the one that determines the evaluation.
For example, a blood pressure value of 140/80 mmHg or a value of 130/90
mmHg indicates «blood pressure too high».
Average Indicator «MyCheck»
This symbol 30 indicates after each measurement, if the most recent
measured value lies below, above or on the same level as your stored average
value (see also chapter «5. Data Memory»).
If the measured Systole or Diastole is more than 5mmHg higher than the stored
average, the arrow shows upwards.
If the measured Systole or Diastole is more than 5mmHg lower than the stored
average, the arrow shows downwards.
If the measured Systole and Diastole do not differ by more than 5mmHg from the
stored average, the arrow shows straight on.
If the measured systole and diastole differ in different directions from the
stored average, this is indicated first with the systole figure flashing,
together with the up or down arrow for two seconds. Thereafter, the diastole
figure flashes with the arrow pointing up or down for two seconds.
The appearance of the Irregular Heartbeat (IHB) Symbol
This symbol 22 indicates that an irregular heartbeat was detected.
In this case, the measured blood pressure may deviate from your actual blood
pressure values. It is recommended to repeat the measurement.
Information for the doctor in case of repeated appearance of the IHB
symbol:
This device is an oscillometric blood pressure monitor that also measures the
pulse during blood pressure measurement and indicates when the heart rate is
irregular.
Data Memory
This device automatically stores up to 99 measurement values for each of the 2
users.
Select either user 1 or 2 by pressing the user button 5.
Viewing the average of the last 28 days
Press the M-button 3 again. The display first shows « M » 26 and «
28A », which stands for the average measurement values of the last 28
days.
Blood pressure readings with suboptimal cuff fit 20 -A are not considered in the average value.
Viewing the clinical blood pressure average «MyBP»
Pressing the M-button 3 briefly when the device is switched off, allows
you to see the average clinically relevant blood pressure « MyBP». The
display first shows « M » 26 and « MyBP » 27. This average is
only shown when 12 clinically relevant measurement values in the last 28 days
have been performed.
Only measurements that were performed in the morning between 05:00-10:59 or in
the evening between 17:0022:59 are considered.
A maximum of 4 readings per day are considered (2 from the morning and 2 from
the evening).
Measurements performed in standard mode and in MAMmode are considered in the
average when they have been performed during the right time of the day.
Measurements performed in MAM-mode or single standard mode are both classed as
single measurements for working out the «MyBP average». Blood pressure
readings with suboptimal cuff fit BT-A are not considered in the average
value.
Viewing the stored single values
Pressing the M- button again, allows you to see the last performed
measurement. The display first shows « M » 26 and a value, e.g.
«M17». This means that there are 17 single values in the memory.
Pressing the M-button again displays the previous value. Pressing the M-button
repeatedly enables you to move from one stored value to another.
Pay attention that the maximum memory capacity of 99 memories is not exceeded.
When the 99 memory is full, the oldest value is automatically overwritten
with the 100th value. Values should be evaluated by a doctor before the
memory capacity is reached – otherwise data will be lost.
Clearing all values
Make sure the correct user is activated.
If you are sure that you want to permanently remove all stored values, hold
down the M-button (the device must have been switched off beforehand) until
«CL ALL» appears and then release the button. To permanently clear the memory,
press the time button while «CL ALL» is flashing. Individual values cannot be
cleared.
Cancel deletion: press ON/OFF button 1 while « CL ALL » is flashing.
Battery Indicator and Battery change
Low battery
When the batteries are approximately ¾ empty the battery symbol AS will flash
as soon as the device is switched on (partly filled battery displayed).
Although the device will continue to measure reliably, you should obtain
replacement batteries.
Flat battery – replacement
When the batteries are flat, the battery symbol AS will flash as soon as the
device is switched on (flat battery displayed). You cannot take any further
measurements and must replace the batteries.
- Open the battery compartment AK at the back of the device.
- Replace the batteries – ensure correct polarity as shown by the symbols in the compartment.
- To set a date and time, follow the procedure described in «Section 2.».
The memory retains all values although date and time must be reset – the year number, therefore, flashes automatically after the batteries are replaced.
Which batteries and which procedure?
Use 4 new, long-life 1.5 V, size AA alkaline batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the device is not going to be used for a prolonged period.
Using rechargeable batteries
You can also operate this device using rechargeable batteries.
Only use «NiMH» type reusable batteries.
Batteries must be removed and recharged when the flat battery symbol appears.
They should not remain inside the device as they may become damaged (total
discharge as a result of low use of the device, even when switched off).
Always remove the rechargeable batteries if you do not intend to use the
device for a week or more.
Batteries cannot be charged in the blood pressure monitor. Recharge batteries
in an external charger and observe the information regarding charging, care,
and durability.
Using a Mains Adapter
You can operate this device using the Microlife mains adapter (DC 6V, 600 mA).
Only use the Microlife mains adapter available as an original accessory
appropriate for your supply voltage.
Ensure that neither the mains adapter nor the cable are damaged.
- Plug the adapter cable into the mains adapter socket AT in the blood pressure monitor.
- Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is consumed.
Bluetooth® Function
This device can be used in conjunction with a smartphone running the
«Microlife Connected Health+» App.
Bluetooth® can be activated manually by pressing and holding the ON/OFF button
1 for about 4 seconds until the Bluetooth symbol begins flashing 15.
Bluetooth® has automatically activated 15 after a measurement has been
completed.
Bluetooth pairing and setup are required to use this device with the
«Microlife Connected Health+» App. It is recommended to manually activate
Bluetooth® and connect to the «Microlife Connected Health+» App to synchronize
the date and time settings between the device and the smartphone.
In case of difficulties connecting the device with the «Microlife Connected
Health+» App, try to reset the Bluetooth connection of this device to default
settings: Press and hold the ON/OFF button 1 for 8 seconds until «CL Pr» is
displayed. This clears the Bluetooth® pairing records on the device and resets
them to factory settings. Bluetooth® automatically activates after the reset.
For more detailed information visit
www.microlife.com/connect.
PC-Link Functions
This device can be used in conjunction with a personal computer (PC) running the Microlife Blood Pressure Analyzer+ (BPA+) software. The memory data can be transferred to the PC by connecting the monitor via a cable. If no download- voucher and cable are included download the BPA+ software from www.microlife.com/software and use a Micro-USB cable.
Error Messages
If an error occurs during the measurement, the measurement is interrupted and an error message, e.g. « Err 3 », is displayed.
Error | Description | Potential cause and remedy |
---|
« Err 1 »
21| Signal too
weak
| The pulse signals on the cuff are too weak. Re-position the cuff and repeat
the measurement.*
« Err 2 »
20-B| Error signal| During the measurement, error signals were detected
by the cuff, caused for instance by movement or muscle tension. Repeat the
measurement, keeping your arm still.
« Err 3 »
20-C
| Abnormal cuff
pressure| An adequate pressure cannot be generated in the cuff. A leak may
have occurred. Check that the cuff is correctly connected and is not too
loose. Replace the batteries if necessary. Repeat the measurement.
« Err 5 »| Abnormal
result| The measuring signals are inaccurate and no result can therefore be
displayed. Read through the checklist for performing reliable measurements and
then repeat the measurement.
« Err 6 »| MAM Mode| There were too many errors during the measurement in
MAM mode, making it impossible to obtain a final result. Read through the
checklist for performing reliable measurements and then repeat the
measurement.
«HI»| Pulse or cuff pressure too high| The pressure in the cuff is too
high (over 299 mmHg) OR the pulse is too high (over 200 beats per minute).
Relax for 5 minutes and repeat the measurement.
«LO»| Pulse too low| The pulse is too low (less than 40 beats per
minute). Repeat the measurement.
| Problem with Bluetooth®
connection| If any problem occurs with the Bluetooth connection, the
Bluetooth® icon AO blinks rapidly for approximately 10 seconds. To solve the
problem, please visit
www.microlife.com/connect.
- Please immediately consult your doctor, if this or any other problem occurs repeatedly.
Safety, Care, Accuracy Test, and Disposal
Safety and protection
-
Follow instructions for use. This document provides important product operation and safety information regarding this device.
Please read this document thoroughly before using the device and keep it for future reference. -
This device may only be used for the purposes described in these instructions. The manufacturer cannot be held liable for damage caused by the incorrect application.
-
This device comprises sensitive components and must be treated with caution. Observe the storage and operating conditions described in the «Technical Specifications» section.
-
Protect it from:
• water and moisture
• extreme temperatures
• impact and dropping
• contamination and dust
• direct sunlight
• heat and cold -
The cuffs are sensitive and must be handled with care.
-
Do not exchange or use any other kind of cuff or cuff connector for measuring with this device.
-
Only pump up the cuff once fitted.
-
Do not use this device close to strong electromagnetic fields such as mobile telephones or radio installations. Keep a minimum distance of 3.3 m from such devices when using this device.
-
Do not use this device if you think it is damaged or notice anything unusual.
-
Never open this device.
-
If the device is not going to be used for a prolonged period the batteries should be removed.
-
Read the additional safety information provided within the individual sections of this instruction manual.
-
If you suffer from a cardiac arrhythmia consult with your doctor before using the device. See also chapter «Appearance of the Irregular Heartbeat (IHB) Symbol» of this user manual.
-
The measurement results given by this device is not a diag It is not replacing the need for the consultation of a physician, especially if not matching the patient’s symptoms. Do not rely on the measurement result only, always consider other potentially occurring symptoms and the patient’s feedback. Calling a doctor or an ambulance is advised if needed.
-
Permanently high blood pressure values can damage your health and must be treated by your doctor!
-
Always discuss your values with your doctor and tell him/her if you have noticed anything unusual or feel unsure. Never rely on single blood pressure readings.
-
Under no circumstances should you alter the dosages of drugs or initiate treatment without consulting your doctor.
-
Deviations between measurements taken by your doctor or in the pharmacy and those taken at home are quite normal, as these situations are completely different.
-
The pulse display is not suitable for checking the frequency of heart pacemakers!
-
If you are pregnant , you should monitor your blood pressure regularly as it can change drastically during this time.
Ensure that children do not use this device unsupervised; some pads are small enough to be swallowed. Be aware of the risk of strangulation in case this device is supplied with cables or tubes.
Device care
Clean the device only with a soft, dry cloth.
Cleaning the cuff
Carefully remove spots on the cuff with a damp cloth and soapsuds.
WARNING: Under no circumstances may you wash the inner bladder!
Accuracy test
We recommend this device be tested for accuracy every 2 years or after
mechanical impact (e.g. being dropped). Please contact your local Microlife-
Service to arrange the test (see foreword).
Disposal
Batteries and electronic devices must be disposed of in accordance with the locally applicable regulations, not with domestic waste.
Guarantee
This device is covered by a 5-year guarantee from the date of purchase. During
this guarantee period, at our discretion, Microlife will repair or replace the
defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
- Transport costs and risks of transport.
- Damage caused by incorrect application or non-compliance with the instructions for use.
- Damage caused by leaking batteries.
- Damage caused by accident or misuse.
- Packaging/storage material and instructions for use.
- Regular checks and maintenance (calibration).
- Accessories and wearing parts: Batteries, power adapter (optional).
The cuff is covered by a functional guarantee (bladder tightness) for 2 years.
Should guarantee service to be required, please contact the dealer from where
the product was purchased or your local Microlife service. You may contact
your local Microlife service through our website:
www.microlife.com/support Compensation is
limited to the value of the product. The guarantee will be granted if the
complete product is returned with the original invoice. Repair or replacement
within the guarantee does not prolong or renew the guarantee period. The legal
claims and rights of consumers are not limited by this guarantee.
Technical Specifications
Operating conditions:| 10 – 40 °C / 50 – 104 °F 15 – 95 % relative
maximum humidity
---|---
Storage conditions:| -20 – +55 °C / -4 – +131 °F
15 – 95 % relative maximum humidity
Weight:| 415 g (including batteries)
Dimensions:| 157.5 x 105 x 61.5 mm
Measuring procedure:| oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V diastolic
Measurement range:| 20 – 280 mmHg – blood pressure
40 – 200 beats per minute – pulse
Cuff pressure
display range:
| 0 – 299 mmHg
Resolution:| 1 mmHg
Static accuracy:| pressure within ± 3 mmHg
Pulse accuracy:| ± 5 % of the readout value
Communication:| Bluetooth® 4.2
Compatibility:| iOS: iOS 10.0 or newer
Android: Android 5 or newer
Voltage source:| 4 x 1.5 V alkaline batteries; size AA Mains adapter DC
6V, 600 mA (optional)
Battery lifetime:| approx. 920 measurements (using new batteries)
IP Class:| IP20
Reference to
standards:| EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life:| Device: 5 years or 10000 measurements
Accessories: 2 years
This device complies with the requirements of the Medical Device Directive
93/42/EEC. Technical alterations reserved.
The Bluetooth® word mark and logos are registered trademarks owned by
Bluetooth SIG, Inc. and any use of such marks by Microlife Corp. is under
license. Other trademarks and trade names are those of their respective
owners.
Guarantee Card
Name of Purchaser………….
Serial Number…………………
Date of Purchase……………
Specialist Dealer……………..
Microlife AG
Espenstrasse 139
9443 Widnau / Switzerland
www.microlife.com
EC REP
Microlife UAB
P. Lukšio g. 32
08222 Vilnius / Lithuania
References
- The Most Iconic Photographs of All Time - LIFE
- Hypertension and Fever Management - Microlife AG
- Hypertension and Fever Management - Microlife AG
- Connect - Microlife AG
- Request Support Microlife BP Analyzer+ Software - Microlife AG
- Support - Microlife AG
- Technologies - Microlife AG
- Connect - Microlife AG
- Verenpainemittaus ja kuumemittaus - Microlife AG
- Microlife - Hipertenzijos ir karščiavimo kontrolė - Microlife AG
- Mājaslapa - Microlife AG
- Preventief je gezondheid managen - Microlife Nederland - Microlife AG
- Advies of hulp nodig? | Microlife - Microlife AG
- Технология AFIB - Microlife / Микролайф
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>