MEDSOURCE MS-131002 Non-Contact Infrared Body Thermometer User Manual
- June 6, 2024
- MEDSOURCE
Table of Contents
- Foreword
- Safety Information
- Overview
- Equipment Symbols
- Technical Specifications
- Safety Classification of ME Equipment
- Operation
- Troubleshooting
- Replacing the Batteries
- Cleaning, Care, and Storage
- Disposal
- Calibration
- EMC Declaration
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
User Manual
Non-Contact Infrared Body Thermometer
Model MS-131002
Please read this manual before operating this device.
It contains important safety information.
HTD8813C
Foreword
This user manual is intended to provide the necessary information for the
proper operation of the model MS-131002 non-contact infrared body thermometer.
General knowledge of infrared thermometers and an understanding of the
features and functions of the thermometer are prerequisites for proper use.
The non-contact infrared body thermometer is a medical device designed to last
for three years.
Please read this manual before using the thermometer. If you do not fully
understand how to use the thermometer, please consult this manual.
Do not operate the
thermometer without completely reading and understanding these instructions.
Notice
Purchase or possession of this device does not carry any express or implied
license to use with replacement parts which would, alone or in combination
with this device, fall within the scope of one of the relating patents.
Safety Information
This device should only be used for the purposes described in these
instructions. The manufacturer cannot be held liable for the damage caused by
the incorrect application.
The non-contact thermometer is designed to minimize the possibility of hazards
from errors in the software program by following human factors design, risk
analysis, and software validation processes.
Warnings
Warnings are identified by the Warning symbol shown above.
- The non-contact infrared body thermometer is to be operated by consumers in the home setting and primary care setting as a screening tool. The manual includes directions for use, all precautionary information and specifications should be read before use.
- This thermometer is designed to measure human body temperature on the forehead. Do not use it for any other purpose.
- This thermometer is intended to be used as a screening tool in a home setting and a primary care setting.
- Do not use the thermometer if it malfunctions or has been damaged.
- If the ambient temperature changes too much, such as moving the thermometer from a cold room to a warm room, allow the thermometer to adjust to the new temperature. This may take up to 30 minutes. The operating temperature of the thermometer is 59 °F to 95 °F (15 °C to 35 °C). Remove the batteries if the thermometer is not likely to be used for a long time.
- The thermometer is not waterproof. Do not immerse it in water or other liquids. For cleaning and disinfection, please follow the instructions contained in the “Cleaning, Care and Storage” section.
- Do not touch the infrared sensor with your fingers.
- If a cold compress has been applied to the patient’s forehead, or other physical measures have been used to cool down the patient, avoid using the thermometer, as it may produce a low reading.
- When measuring body temperature using the forehead, select “body” mode. When measuring other objects, liquids, or foods, select “surface” mode.
- This product must be operated in a stable environment. If the ambient environment has changed, water may condense on the surface of the infrared sensor. In the event of condensation on the infrared sensor, see the “Cleaning, Care, and Storage” section for guidance.
- Do not use the thermometer near strong electrostatic or magnetic fields. These fields may affect the accuracy of the thermometer.
- When replacing the batteries, do not mix old and new batteries. This may damage the thermometer.
- The accuracy of the measurement may be elected if the forehead is covered by hair, perspiration, or clothing.
- This thermometer is intended for screening. If you have any doubt about the result, please measure the temperature using another method.
- The thermometer is calibrated during manufacturing. If used according to the instructions, periodic calibration is not required.
The thermometer should be kept within reach of children and pets. When not in
use, store the device in a dry place and protect it against moisture, heat,
lint, dust, and direct sunlight. Never place any heavy objects on the
thermometer or thermometer packaging.
Do not dispose of batteries in a fire.
Only use recommended batteries. Do not use rechargeable batteries.
This thermometer is not designed to replace diagnostic thermometers in
hospitals.
Do not drop, disassemble or modify the thermometer.
Do not use the thermometer if you think it is damaged or notice unusual
operation.
The thermometer is comprised of sensitive components and must be treated with
caution.
Observe the storage and operating conditions described in the “Technical
Specifications” section.
Do not perform service or maintenance while the thermometer is on.
When using the thermometer, do not touch the battery and patient
simultaneously.
Do not use the device if it is damaged/ degraded/ or if any of the
thermometers have become loose. The use of a damaged unit may cause improper
results or injury.
Health Canada Warning: The thermometer is not intended for use on children
under the age of two.
Based on current science and technology, the thermometer is not known to cause
any allergic reactions.
This equipment needs to be installed and put into service in accordance with
the information provided in the ACCOMPANYING DOCUMENTS.
Overview
Intended Use
The non-contact infrared body thermometer is designed to be used for
intermittent measurement and monitoring of human body temperature by consumers
in the home setting and primary care setting as a screening tool.
Description of Non-Contact Infrared Body Thermometer
Introduction and Operating Principle
This non-contact thermometer is a hand-held, reusable, battery-operated device
that measures human body temperature via the patient’s forehead.
The operating principle is based on infrared sensor technology. The infrared
sensor can output different signals when measuring different object
temperatures or in different ambient temperatures. An application-specific
integrated circuit then turns the signal from the infrared sensor into a
digital value for display on the LCD screen.
Description of Controls, Indicators, and Symbols
Figure 1: Overview
- Liquid Crystal Display (LCD)
- Battery Cover
- On/Scan Button
- Set Button
- Memo Button
- Mode Button
- Infrared Sensor
Figure 2: Display
- Surface Mode
- Body Mode
- Digital Display
- Fever Indicator
- Battery Indicator
- Buzzer On/O
- Memory Index
- Fahrenheit
- Celsius
- Memory Mode
Thermometer Applications
Thermometer Model Number | Thermometer | Adult | Pediatric |
---|---|---|---|
Ear | Forehead | Ear | Forehead |
HTD8813C/MS-131002 | Non-Contact Infrared Body Thermometer |
√
| |
√
Equipment Symbols
| Warning| | Compliance with WEEE Standard
---|---|---|---
| Non-Sterile Package| | DO NOT THROW AWAY Intended for
multiple uses
| Refer to Operating
Instructions| | Operating atmospheric pressure
| Operating Temperature| | Serial number
| Operating Humidity| | Recyclable
| Manufacturer| | This device complies with Part 15 of
FCC (Federal Communications
Commission) rules.
| Ingress Protection: The first 2 indicates protection from the ingress of
solid foreign objects 12.5 mm in diameter and greater. The second 2 indicates
protection from the ingress of vertically falling water drops when the
enclosure is tilted not greater than 15 degrees from vertical.
Technical Specifications
Specification | Value(s) |
---|---|
Measurement Units | Fahrenheit (°F) / Celsius (°C) |
Operating Mode
Direct mode
| Adjusted mode (Body mode)
(Surface mode)
Measuring Site| Forehead
Reference Body Site| Axillary
Rated Output Range| Body mode: 95.0 °F – 107.6 °F (35.0 °C – 42.0 °C)
Surface mode: 32.0 °F – 212.0 °F (0.0 °C – 100.0 °C)
Extended Output Range| Body mode:
Lower: 93.2 °F – 94.8 °F (34.0 °C – 34.9 °C)
Upper: 107.8 °f – 109.2 °F (42.1 C 42.9 °C)
Range| Body mode: 93.2 °F – 109.2 °F (34.0 °C – 42.9 °C)
Surface mode: 32.0 °F – 212.0 °F (0.0 °C – 100.0 °C)
---|---
Accuracy| Body mode:
93.2 °F – 94.8 °F ±0.5 °F (34.0 °C – 34.9 °C ±0.3 °C)
95.0 °F – 107.6 °F ±0.4 °F (35.0 °C – 42.0 °C ±0.2 °C)
107.8 °F – 109.2 °F ±0.5 °F (42.1 °C – 42.9 °C ±0.3 °C)
Surface mode: ±3.6 °F (±2.0 °C)
Temperature
Resolution| Fahrenheit: 0.1
Celsius: 0.1
Three Color Backlight (Color Alarm)| Green – Normal Temperature: 95.9°F – 99.2
°F (35.5 °C – 37.3 °C)
Yellow – Slight Fever: 99.3 °F – 100.4 °F (37.4 °C – 38.0 °C)
Red – High Fever: 100.5 °F – 109.2 °F (38.1 °C – 42.9 °C)
Notes: 1. Surface Mode always displays a green backlight.
2. In Body Mode, a reading of 93.2 °F – 95.8 °F(34.0 °C – 35.4 °C) displays a
green backlight.
Auto Power Off| ≤ 18 s
Measuring Time| ≤ 2 s
Measuring Distance| 0.4 – 2.0 in (1 – 5 cm)
Memory| 50 measurements
Power Supply Requirements
Batteries| 1.5V AAA Alkaline Battery x 2 (IEC Type LR03)
Adaptable Range| 2.6V – 3.6V
Environmental Conditions
Operating Conditions| Temperature: 59.0 °F – 95.0 °F (15.0 °C – 35.0 °C)
Relative Humidity: ≤ 85% Atmospheric Pressure: 70 – 106 KPa
Transport and Storage Conditions| Temperature: -4.0 °F – 131.0 °F (-20.0 °C –
55.0 °C)
Relative Humidity: ≤ 93% Atmospheric Pressure: 70 – 106 KPa
Physical Properties
Weight (without batteries)| 3.0 ounces (84 grams)
Size| Length: 5.4” (138mm)
Width: 3.7” (95mm)
Height: 1.6” (40mm)
Compliance
Basic Safety and Essential
Performance| EN 60601-1: 2006+A1:2013, EN 60601-1-2:2007
Type of Protection| Internally Powered Equipment (on battery power)
Degree of Protection| Non-Applied Part
Front Panel and Case Labeling| EN/ISO 15223-1:2012
---|---
Temperature| EN/ISO 80601-2-56:2012
Home Healthcare Environment| EN 60601-1-11:2010
|
---|---
Clinical Bias (Δcb)| -0.027 °C
Standard Deviation (σj)| 0.14 °C
Limits of Agreement (LA)| 0.26 °C
Clinical Repeatability (σr)| 0.07 °C
Safety Classification of ME Equipment
Protection against electric shock | Internally Powered ME Equipment |
---|---|
Applied Part | No Applied Parts |
Protection against harmful ingress of water or particulate matter | IP22 |
Mode of Operation | Continuous Operation |
Note: Not intended to be sterilized. Not for use in an OXYGEN RICH ENVIRONMENT .
Operation
2.1 Battery Installation
1)Pull the battery cover forward as indicated by the arrow.
2)Insert two AAA-size batteries. Ensure correct polarity as indicated inside
the battery cover.
3)Slide the battery cover back in until it snaps in place.
2.2 Before Using the Thermometer
Be sure to read and understand all warnings listed in the instructions
before use.
- If the ambient temperature changes too much, such as moving the thermometer from a cold room to a warm room, allow the thermometer to adjust to the new temperature. This may take up to 30 minutes.
- The ambient temperature around the patient should be stable. Keep away from large air flows such as fans or air-conditioning vents.
- Do not use the thermometer in bright sunlight.
- If a patient moves from a colder environment into a warmer test environment, they should remain in the test environment for at least 5 minutes prior to taking a measurement. his will ensure the patient is consistent with the ambient temperature.
2.3 Thermometer Self-Test
When the thermometer is off, press the On/Scan button to initiate a self-test.
The thermometer will briefly illuminate all segments of the display with green
backlighting and then complete a measurement.
2.4 Temperature Modes
The thermometer is capable of measuring body temperature and surface
temperature. By default, the thermometer will enter Body Mode when turned on.
To toggle between body temperature and surface temperature, press the Mode
Button when the device is on.
2.5 Body Temperature
Taking a Body Measurement
- Align the thermometer with the middle of the forehead to measure body temperature (between and above the eyebrows).
- Ensure the distance between the thermometer and the patient’s forehead is between 0.4” and 2.0” (1cm – 5 cm). See Figure 3.
Figure 3
- Press and release the On/Scan button.
- The measurement will be displayed within two seconds.
- The thermometer will produce an audible indication when the measurement is complete if audio is enabled.
- Wait at least one second before taking the next measurement.
- If a continuous series of measurements are being performed, wait at least 30 seconds after every 5th measurement.
Body Measurement Results
- If the measurement is below 93.2 °F (34.0 °C), the display will show “Lo”, a green backlight, and beep three times if audio is enabled.
- If the measurement is between 93.3 °F and 99.2 °F (34.1 °C and 37.3 °C) the display will show the temperature, a green backlight, a happy face icon, and beep once if audio is enabled. This indicates normal body temperature.
- If the measurement is between 99.3 °F and 100.5 °F (37.4 °C and 38.0 °C), the display will show the temperature, a yellow backlight, a sad face icon and emit a series of beeps if audio is enabled. This indicates a slight fever
- If the measurement is between 100.6 °F and 109.2 °F (38.1 °C and 42.9 °C), the display will show the temperature, a red backlight, a sad face icon and emit a series of beeps if audio is enabled. This indicates a high fever.
- If the measurement is greater than 109.2 °F (42.9 °C), the display will show “Hi”, a green backlight, and beep three times if audio is enabled.
2.6 Surface Temperature
Taking a Surface Measurement
- Ensure the distance between the thermometer and the surface to be measured is between 0.4” and 2.0” (1 cm and 5 cm).
- Press and release the On/Scan button.
- The measurement will be displayed within one second.
- The thermometer will produce a tone when the measurement is complete if audio is enabled. Surface Measurement Results
- If the temperature is less than 32.0 °F (0.0 °C), the display will show “Lo”, a green backlight, and beep three times if audio is enabled.
- If the temperature is between 32.0 °F and 212.0 °F (0.0 °C and 100.0 °C), the display will show the temperature, a green backlight, and beep once if audio is enabled.
- If the temperature is greater than 212.0 °F (100.0 °C), the display will show “Hi”, a green backlight, and beep three times if audio is enabled.
2.7 Memory Mode
The thermometer will automatically store the previous 50 temperature measurements. Temperatures that are registered High or Low are out of range and are not stored.
-
To enter Memory Mode, press the Memo Button while the unit is off or after the completion of a temperature reading.
-
Upon entering Memory Mode, the most recent temperature measurement will be displayed.
-
Press the Memo Button to display the next temperature measurement.
-
Each stored measurement will display the following information:
o The memory index value (1-50, newest to oldest).
o The temperature.
o The units (°F or °C).
o Whether body or surface temperature.
o For body temperature, a happy or sad face icon -
Every stored measurement will be displayed with a green backlight.
-
To clear the stored values, press and hold the Memory Button until “Clr” is displayed.
-
Empty memory cells will be displayed with “ – – – -“ as the temperature value.
-
2.8 Parameter Mode
The thermometer has four parameters that can be tailored to certain populations or environmental conditions.
Entering Parameter Mode -
Ensure the thermometer is on.
-
Press and hold the Set Button until “F1” and then “Unit” is displayed.
Temperature Units (F1)
The thermometer can be set to display the temperature in Fahrenheit or Celsius. -
Press either the Memo Button or the Mode Button to toggle between Fahrenheit and Celsius.
-
When the desired units have been selected, press the Set Button to move to the next parameter.
Fever Threshold (F2)
The thermometer allows the fever threshold to be modified. This is the value at
which the thermometer will indicate a slight fever. The default value is 100.5
°F.
-
The threshold value will be displayed as it is modified.
-
Press the Memo Button to increase the threshold by approximately 0.2 °F (0.1 °C).
-
Press the Mode Button to decrease the threshold by approximately 0.2 °F (0.1 °C).
-
When the desired threshold has been selected, press the Set Button to move to the next parameter.
Audio (F3)
The thermometer can be used with audio enabled or disabled. -
Press either the Memo Button or the Mode Button to toggle between audio enabled and audio disabled.
-
When the desired audio setting has been selected, press the Set Button to move to the final parameter.
Note: The audio can also be enabled and disabled via Set Button when not in Parameter Mode. -
Temperature offset (F4)
The thermometer allows a surface temperature of the set to be entered. This will increase or decrease the displayed surface temperature, based on the set value. The default setting value is zero. The set range is ± 9.0 °F (± 5.0 °C). -
The offset value is displayed as it is modified.
-
Press the Memo Button to increase the offset by approximately 0.2 °F (0.1 °C).
-
Press the Mode Button to decrease the offset by approximately 0.2 °F (0.1 °C).
-
When the desired temperature oset has been selected, press the Set Button to exit Parameter Mode.
2.9 Notes
- If you experience problems with this thermometer, such as configuration, maintenance, or use, please contact the SERVICE PERSONNEL. Do not attempt to open or repair the device yourself.
- Please report to us any unexpected operations or events.
- The patient is an intended operator. They can determine the state of the batteries and charge the batteries. The patient can maintain the device and its accessories according to the user manual under normal circumstances.
Troubleshooting
Message
| Situation|
Solution
---|---|---
****| Body: The temperature measured is not within the typical human body
temperature range of 93.2 °F to 109.2 °F (34.0 °C to 42.9 °C).
Surface: The temperature measurement is not within the measurable range of
32.0 °F – 212.0 °F (0.0 °C – 100.0 °C).| Ensure the correct distance and
location are being used to measure the temperature.
Do not attempt to measure temperatures outside the allowable range.
| Improper Distance| Ensure the proper distance is used: 0.4” – 2.0” (1
cm – 5 cm).
Incorrect test position.| Ensure the correct position is used.
The subject’s hair, antipyretic stickers, or perspiration is affecting the
measurement.| Remove any items that may interfere with the measurement.
F4 temperature offset is set incorrectly.| Adjust the temperature offset
value.
| The device temperature exceeds the allowable operating temperature
range.| Move to a location within the operating temperature range and allow
the thermometer to adjust to the new temperature.
| The screen flickers and then turns off.| Replace the batteries. If the
error persists, the device may be damaged.
| The battery capacity is too low. Temperature measurement is not
allowed.| Replace batteries.
| The ambient temperature has changed too quickly.| Allow the thermometer
to adjust to the new temperature.
****| Blank Screen
a) The power is off
b) Improper battery installation
c) The batteries are depleted.
d) Device is damaged| a) Press the On/Scan button.
b) Check battery polarity.
c) Replace the batteries
d) Contact retailer or service center.
Replacing the Batteries
Caution: The thermometer does not operate with dead or low batteries and
does not allow connection to an external power supply.
1)Before replacing the batteries, ensure the device is off.
2)Pull the battery cover forward as indicated by the arrow.
3) Remove the old batteries.
4)Insert two AAA-size batteries. Ensure correct polarity as indicated inside
the battery cover.
5)Slide the battery cover back in until it snaps in place.
6)Dispose of the batteries in accordance with local regulations.
If the device does not function after replacing the batteries:
- Check for proper battery polarity.
- Try a fresh set of batteries.
- The unit may have locked up if the batteries were removed while the unit was powered. In this case, remove the batteries, wait 30 seconds, and then reinsert the batteries.
Warning
Do not recharge, disassemble, or dispose of in a fire.
- The typical service life for a set of new batteries is 2000 measurements with an operation time of 18 seconds per measurement.
- Only use the recommended batteries.
- Do not attempt to recharge non-rechargeable batteries.
- Do not dispose of batteries in a fire.
- Remove the batteries if the thermometer is not to be used for a long period of time.
Cleaning, Care, and Storage
- It is important to protect the infrared lens from dirt and damage; it is very delicate.
- Use a clean, soft cloth to clean the surface of the device and LCD. Do not use solvents or immerse the device in water or other liquids.
- Always keep the thermometer within the storage temperature range of -4.0 °F to 131.0 °F (-20.0 ° to 55.0 °C) and humidity range of ≤ 93% (non-condensing) as specified.
- It is recommended to store the thermometer in a dry location free from dust.
- Do not expose the thermometer to direct sunlight, high temperature or humidity, or any other extreme environmental conditions as it may affect the proper operation of the device.
- If the ambient temperature changes too much, such as moving the thermometer from a cold room to a warm room, allow the thermometer to adjust to the new temperature. This may take up to 30 minutes. Condensation may have formed on the infrared sensor.
Disposal
- Batteries should be disposed of in accordance with local regulations.
- Do not dispose of the thermometer in the unsorted municipal waste stream. Enquire about options
for environmentally friendly and appropriate disposal. Take local regulations into account.
Calibration
The thermometer is initially calibrated at the time of manufacture. If this thermometer is used according to the instructions, periodic re-adjustment is not required. If at any time you question the accuracy of the temperature measurements, please contact service personnel.
EMC Declaration
- This equipment needs to be installed and put into service in accordance with the information provided in the ACCOMPANYING DOCUMENTS; This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment.
- Do not use a mobile phone or other devices that emit electromagnetic fields near the unit. This may result in an incorrect operation of the unit.
- Caution: This unit has been thoroughly tested and inspected to assure proper performance and operation.
- Caution: This machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Non-Contact Infrared Body Thermometer is intended for use in the
electromagnetic environments specified below. The Customers or users of the
Non-Contact Infrared Body Thermometer should ensure that it is used in such an
environment.
Immunity Test| IEC 60601 Test Level| Compliance Level|
Electrostatic Discharge (ESD)
IEC 61000-4-2| ±6 kV contact
±8 kV air| ±6 kV contact
±8 kV air| Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at least 30%.
Electrical Fast
Transient/Burst IEC 61000-4-4| ±2 kV for power supply lines
±1 kV for input/output lines| Not Applicable| The main power quality should be
that of a typical commercial or hospital environment.
Surge IEC 61000-4-5| ±1 kV line(s) to line(s)
±2 kV for line(s) to earth| Not Applicable| The main power quality should be
that of a typical commercial or hospital environment.
Voltage dips, short interruptions, and voltage variations on power supply
input lines IEC 61000-4-11| <5% UT (>95% dip in UT) for 0.5 cycle 40% UT (60%
dip in UT) for 5 cycles 70% UT (30% dip in UT) for 25 cycles <5% UT (>95% dip
in UT) for 5 secs| Not Applicable| The main power quality should be that of a
typical commercial or hospital environment. If the user of the Non-Contact
Thermometer requires continued operation during a mains power interruption, it
is recommended that the device be powered from an uninterruptable power supply
or battery.
Power Frequency (50 Hz/60 Hz)
Magnetic Field IEC 61000-4-8| 3 A/m| 3 A/m| Power frequency magnetic fields
should be at levels characteristic of a typical location in a typical
commercial or hospital environment.
Note: UT is the AC mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration – Electromagnetic Emission
The Non-Contact Infrared Body Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Non-Contact Infrared Thermometer should ensure that it is used in such an
environment.
Emission Test| Compliance| Electromagnetic Environment – Guidance
RF Emissions CISPR 11| Group 1| The Non-Contact Infrared Body Thermometer uses
RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic
equipment.
RF Emission CISPR 11| Class B
Harmonic Emissions IEC 6100-3-2| Not Applicable| The Non-Contact Infrared Body
Thermometer is suitable for use in all establishments, other than domestic and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Non-Contact Infrared Body Thermometer is intended for use in the
electromagnetic environments specified below. The customer or user of the Non-
Contact Infrared Body Thermometer should ensure that it is used in such an
environment.
Immunity Test| IEC 60601 Test Level| Compliance Level| Electromagnetic
Environment – Guidance
Conducted RF IEC 61000-4-6
Radiated RF IRC61000-4-3
| 3VRMS 150kHz to 80 MHz
3 V/M 80MHz to 2.5 GHz
| Not Applicable
3 V/M
| Portable and mobile RF communications equipment should be used no closer to any part of the Non-Contact Infrared Thermometer, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended Separation Distance d = 1.2 √p (150 kHz – 80 MHz)
d = 1.2 √p (80 MHz – 800 MHz)
d = 1.2 √p (800 MHz – 2.5 GHz)
Where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters.
Field strengths from fixed RF transmitters, as determined by electromagnetic site surveysª should be less than the compliance level in each frequency rangeª. Interference may occur in the vicinity of equipment marked with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
a. Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM, and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to
verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the device.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
Recommended Separation Distances between
Portable and Mobile RF Communications Equipment and the Non-Contact Infrared Thermometer
The Non-Contact Infrared Thermometer is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or
user of the device can help prevent electromagnetic interference by
maintaining distance between portable and mobile RF communications equipment
(transmitters) and the device as recommended below, according to the maximum
output power of the communications equipment.
Separation distance according to the frequency of the
transmitter (m)
Rated Maximum Output Power of Transmitter (W)| 150 kHz – 80 MHz
d = 1.2√p
| 80 MHz – 800 MHz
d = 1.2√p
| 800 MHz – 2.5 GHz
d = 2.3√p
0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by the absorption and reflection from structures, objects, and people.
MANUFACTURED FOR :
MEDSOURCE
4201 NOREX DRIVE
CHASKA, MN 5531 8
800-876-8264
info@medsourcelabs.com
www.medsourcelabs.com
HAIDA TECHNOLOGY CO., LTD.
Add:4F BaShiDa High-Tech Park,XiangDong Industrial Area,DaLing Shan
Town,DongGuan City,GuangDong,China
Made in China
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>