Sentec Digital Monitoring System SDMS Instructions

June 6, 2024
Sentec

Sentec Digital Monitoring System (SDMS)
Quick Reference Guide
Software version
SMB SW-V08.03 and higher

Adult mode
`Adult´ if older than term birth + 12 months

Selection of Measurement Site and Sensor Attachment Accessory

Sentec Digital Monitoring System SDMS

Attachment Accessory
Earlobe:

Use Ear Clip for mature, intact skin.
All other sites:

Use MAR/e-MI for mature, intact skin or MAR/e-SF for sensitive, fragile skin.

WARNING: Refer to the Instruction Manual for the SDMS, Technical Manual for the SDM, or the respective Directions for Use of the sensor and/or disposables for warnings, cautions, and additional information such as instructions, routine checks, or maintenance recommendations. These are available on www.sentec.com/ifu

https://sentec.com/tv/v0/

Check SDM Settings, System Readiness, and Sensor Condition before Use

Ensure current SDM Settings/SDM Profile are appropriate for the patient, for the selected site, and the skin condition/skin tissue perfusion at the selected site.

Sentec Digital Monitoring System SDMS - Check SDM
Settings

When removing the sensor from the Docking Station, check the condition of its membrane and its integrity before applying it to the patient. Change the membrane if necessary.  o not use the sensor if any problems are noted.

Sensor Application Using a Multi-Site Attachment Ring

Sentec Digital Monitoring System SDMS - Site Attachment
Ring

  1. Clean the site and let it dry. If necessary, remove hair.
  2. Attach the Multi-Site Attachment Ring (MAR/e) to the measurement site. Verify that the skin under the adhesive is not wrinkled.
  3. Apply one small drop of Sentec Contact Gel to the skin area in the center of the ring. Avoid wetting the adhesive tape!
  4. Holding the sensor at its neck, approach the MAR/e and first insert the nose of the sensor into the ring. Click in the sensor by applying slight downward pressure on its neck. Rotate the sensor in the ring into the best position and press the sensor gently against the skin to spread the Contact Gel. Verify that air gaps between the skin and the sensor are eliminated and that the sensor can easily be rotated.
  5. Tape the cable to the skin and secure it with a Clothing Clip on the patient’s clothing or bed linen. For forehead/cheek placement, wrap the cable once around the ear first. Ensure that the sensor cable is loose enough not to be stretched during monitoring. Gently press on the sensor as a final application check.

Note: If more secure sensor attachment is required, e.g. in high humidity environments, for patients who perspire profusely and/or in challenging patient motion conditions, the Staysite™ Adhesive (model SA-MAR) can be used complementary with the Multi-Site Attachment Rings.

Patient Monitoring

After sensor application, verify that the SDM detects ‘Sensor-On-Patient’, initiate monitoring and the enabled parameters stabilize. If necessary, readjust sensor application or reposition the sensor.
Various preconfigured measurement screens are available.
Press the Display Button ( ) to cycle between the available screens. Press the Enter button ( ) to open a ‘Quick Access Menu’ permitting to set a baseline, to set an RHP reference, to mark ‘Operator Events’, or to perform a ‘PCO2 In- Vivo Correction’.
Note: SpO2 and PR usually stabilize within a few seconds. PCO2 typically increases to reach a stabilized value within 2 to 10 minutes.

Sentec Digital Monitoring System SDMS - Patient
Monitoring

Sensor Application Using an Ear Clip

Sentec Digital Monitoring System SDMS - Sensor
Application

  1. Clean the earlobe and let it dry. If necessary, remove hair.
  2. Attach the clip to the earlobe with the retainer ring on the backside of the earlobe. Verify that the skin under the retainer ring’s adhesive is not wrinkled and that the hole in the center of the retainer ring completely covers the skin.
  3. Take the sensor and apply one small drop of Contact Gel in the middle of the sensor surface. Ensure that the gel does not run off the sensor face. Pull the earlobe with the Ear Clip in a horizontal position. Snap the sensor into the clip.
  4. Then guide the earlobe back in a vertical position and – if this is not the case yet – rotate the sensor such that its cable points to the crown of the head.
  5. Verify the entire dark surface of the sensor is covered by the earlobe, that air gaps between the skin and the sensor are eliminated and that the sensor can easily be rotated.
  6. Wrap the sensor cable around the ear once, tape it to the cheek as shown in the picture, and secure it with a Clothing Clip on the patient’s clothing or bed linen. Ensure that the sensor cable is loose enough not to be stretched during monitoring. Gently squeeze the sensor and Ear Clip as a final application check.

Sensor Removal.

When monitoring is completed or monitoring time has elapsed, remove the sensor from the patient and clean/ inspect the skin. Clean the sensor. Inspect condition of membrane and integrity of sensor before inserting it into the Docking Station.

CAUTION: For site inspection and/or calibration, the sensor attachment accessory can remain on the same site for up to 24 hours and maybe reused for another sensor application. It is recommended to remove and discard the attachment accessory after 24 hours and to keep the measurement site free of adhesive for 8 to 12 hours.

IMPORTANT:

  • Clean sensor after use!
  • Check the condition of membrane and integrity of sensor prior/after use!
  • To maintain monitor readiness and minimize PCO2 drift potential, keep SDM switched on and store sensor in Docking Station in between monitoring!

Sensor Calibration

If a sensor calibration is mandatory, the message ‘Calibrate sensor’ is displayed and PCO2 values are replaced by ‘—’.

Sentec Digital Monitoring System SDMS - Calibrate
sensor

Calibration Intervals for Sentec TC Sensors can last up to 12 hours. Once elapsed, calibration is recommended and monitoring is possible for another 4 to 6 hours (PCO2 ‘questionable’). Thereafter, sensor calibration is mandatory.

Changing the Sensor Membrane

If the ‘Membrane Change Interval’ has elapsed, the SDM displays the message ‘Change sensor membrane‘ and marks PCO2 values as invalid (‘—’).
Use the QR code at left to navigate directly to our Membrane Change tutorial video.
https://sentec.com/tv/v0/

https://sentec.com/tv/v0/

Sentec Digital Monitoring System SDMS - Changing the Sensor
Membrane

Without being requested by the SDM, the sensor membrane must additionally be changed if it is damaged or missing, has a loose fit, or if there is trapped air or dry electrolyte under the membrane.
IMPORTANT: In default settings, the ‘Membrane Change Interval’ is 28 days.
It can be customized.

Online User Guide

Quickly access tutorial videos, FAQs, and manuals through our Online User Guide. https://www.sentec.com/onlineuserguide

Sentec Digital Monitoring System SDMS - QR 2

https://www.sentec.com/onlineuserguide

0123
Sentec AG, Ringstrasse 39, CH-4106 Therwil, Switzerland, www.sentec.co
HB-007713-d Date of release: 12/2020

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals