AMobile E-Guardian S409 Mask and Fever Screening Solution User Guide
- June 6, 2024
- AMobile
Table of Contents
E-Guardian S409
Mask & Fever Screening Solution
Quick Start Guide
Copyright Notice
©2020 All Rights Reserved.
This document contains proprietary information protected by copyright. All
rights are reserved. No part of this manual may be reproduced by any
mechanical, electronic, or other means in any form without prior written
permission of the manufacturer.
Disclaimer
The S409 Preliminary Fever Screening Solution is not intended to replace
professional medical devices. In the event that a potential fever temperature
is detected, the subject should be immediately rechecked with an approved
medical thermometer.
The information in this document is subject to change without prior notice in
order to improve the reliability, design, and function. It does not represent
a commitment on the part of the manufacturer.
Under no circumstances will the manufacturer be liable for any direct,
indirect, special, incidental, or consequential damages arising from the use
or inability to use the product or documentation, even if advised of the
possibility of such damages.
Federal Communications Commission (FCC) Statement
15.21
You are cautioned that changes or modifications not expressly approved by the
party responsible for compliance could void the user’s authority to operate
the equipment.
15.105(b)
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to part 15 of the FCC rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses, and can radiate
radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off
and on, the user is encouraged to try to correct the interference by one or
more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
- this device may not cause harmful interference and
- this device must accept any interference received, including interference
that may cause undesired operation of the device.
FCC RF Radiation Exposure Statement:
For body-worn operation, this phone has been tested and meets FCC RF exposure guidelines when used with an accessory that contains no metal and that positions the handset a minimum of 1.5 cm from the body. The use of other accessories may not ensure compliance with FCC RF exposure guidelines.
System Overview
The S409 mask and fever screening solution are designed for fast preliminary detection of potential fever symptoms. The S409 comprises the following main components:
- Thermal camera: An infrared thermography camera to detect potential fever. A temperature of >37.5°C is defined as potential fever with S409.
- Calibrator: To calibrate the thermography camera to enhance accuracy.
How the S409 Works
AI-Powered Human Recognition
The S409 is an AI-powered solution so it is able to recognize a human from the
thermal image and thus measure the forehead temperature and mask. An
individual wearing a mask can still be detected.
Real-time On-Screen Display
The S409 continuously measures the temperature and mask of people passing
through the S409. Screening results are displayed in real-time on the screen.
Audio Alert on Potential Fever
When potential fever is detected, the S409 sends an audio alert for further
examinations.
Installation Options
Designed for flexible installation, the S409 comes with an all-in-one
installation kit to support various installations.
A. VESA Mount
S409 can be installed on the VESA kit for the vehicle application.
B. Tripod Mount
S409 can be installed on the tripod with an universal 1/4″ threaded tripod
socket for installation on a tripod.
Package Contents
Package Contents
• S409 (including power cord and S409.)
| • Type-U power cord for the vehicle (Please note that this is for vehicle
version)
---|---
| • M8 to DC power cord for indoor (Please note that this is for indoor
version)
Getting Started
Before installation, please get yourself familiar with the components of the
S409.
Front View
Getting Started
Rear View
LED Indicator Status
Bottom View
Side View
Dimension
Installation Advice
Before proceeding to install the S409, please follow the guidelines below to
get the optimal performance of the system:
-
Install the S409 indoor with an ambient temperature of 15~30°C.
-
For the person to be detected, it’s suggested to stand at a distance of 1M~1.5M to the S409.
-
To avoid detection interference, prevent another heating source such as fluorescent or lamp from being constantly displayed in the S409’s field of view.
-
Avoid any covering on the forehead such as hair or hat to ensure the measurement.
-
The S409 is capable of detecting multiple human temperatures.
However, avoid overlaps between people to ensure the measurement. For optimal performance, it’s suggested that you allow only one person appears in the field of view at a time.
Installations
Installing the S409 – Vehicle
Please follow the steps below to install the S409 in the vehicle:
- Locate the VESA hole in the bottom of the S409 mounting plate, and install 4 M3*L9 screws on the S409.
- Tighten 4 M6*30 screws for two pairs of pipe clamps. The lower pipe clamp is about 125cm away from the car floor, and the distance between the two pipe clamps is about 10cm.
- Tighten 2 M6*10 screws for racket/device on the pipe to secure the position.
- Adjust the angle of S409 to 15 degrees to screen passengers in the vehicle. Then tighten the screws in the picture as left.
Cautions:
• DO NOT expose the S409 to direct sunshine, water, oil, dust, or corrosive
gas.
Mounting the S409 – Tripod Mount
- Locate the threaded hole in the bottom of the S409 mounting plate.
- Mount your tripod head to the S409 plate. Tighten the screw-head from the bottom of the tripod head to secure the assembly. Then mount it to your tripod.
- Adjust the angle and the height of the tripod to ensure the S409 and person to be detected are in the camera’s field of view.
- Connect the power cord of the provided power adapter to the power connector on S409. Then plug the power adapter into a power outlet. (use a coordinate plug adapter for your territory) The power of S409 will startup automatically.
(Please note that S409 will cause auto shut down after 10 seconds by itself if the power is unplugged.)
Cautions:
• DO NOT expose the S409 to direct sunshine, water, oil, dust, or corrosive
gas.
Power Connection
- The S409 comes with a Type-U (or M8 to DC) power cord for a power source.
- Attach the correct plug according to your location to the power adapter. Then plug the power adapter into a standard AC power outlet.
- When power is applied, the S409 will power on automatically.
Then the S409 will launch the Fever and Mask Screening APP automatically upon startup.
Using S409
When a person’s temperature is detected under 37.5°C, “PASS” will be displayed; otherwise, “OUT OF TEMP” will be displayed and a sound alert will be sent. If the person did not wear the mask, the screen would show “NO MASK” and a sound alert will be sent.
| When S409 detects the mask and the temperature is under 37.5, the monitor
will show a green “PASS” sign.
---|---
| If the temperature is over 37.5, the monitor will show a red “OUT of TEMP”
sign.
| If S409 does not detect the mask, it will show a yellow “NO MASK” sign.
Adjust the Volume
If users need to adjust the volume of the audio, follow the instruction below:
• Slide down the menu
Screen Mirroring
Wireless screen Mirroring is available for the S409. It’s a suitable vehicle
environment.
Please follow the instruction for the screen mirroring.
Then choose your devices. The screen on the S409 can be mirrored on your
display device.
****(Please note that screen mirroring may not be supported depending on your
model or brand of display devices.)
Specifications
Specifications Highlight
Measuring Distance Accuracy| 1M ~1.5M
Accuracy| ±0.5ºC
System Power| 5V/2A
Touch Screen| 5″ Capacitive touch screen
Control| Power button
O.S| Android 11.0
Thermal Imaging| Lepton 2.5
Array Format| 80 x 60
Calibration| Blackbody, ±0.3ºC
Dimension| 196.36 x 132.86 x 157.61 mm (W x H x D)
Weight| 920g
Working Temp.| 15 ~ 30 °C
Storage Temp.| -10 ~ 50 °C
Q & A
1. Q: What will the device do when subjects are not in masks| A: When
users are not in the mask and have fever simultaneously, the device will say
“Abnormal temperature”.
---|---
2. Q: Can the device scan a real-name system OR code?.| A: S409 needs to
connect the external blooth scanner for scanning the QR code.
3. Q: Can one device-detect more than two people at the same time in a
large field site.| A: It is recommended to set up a single gateway to detect
one person at a time when developing S409. S409 can detect up to four people
on a single screen at the same time.
4. Q: What if the alarm sounds
all the time and it is a
misjudgment?
| A: Please confirm if there is any round hot spot near to the subjects, and
remove it..
5. Q: What if the temperature of subjects is higher than the standard?
Shall they be prohibited from entering?
| A: When potential fever is detected, the S409 sends an audio alert. We
suggest you use the medical temperature scanner for further examinations.
6. Q: How far does the machine measure body temperature?| A: We
recommended distance is about 1-1.5M.
7. Q: The temperature can not be detected by s409 when it starts up.| A:
Please note that when S409 connects to the power, the temperature of the
calibrator is not warm at first. Wait for about 5 – 6 minutes for the
calibration furnace to reach the target calibration temperature.