About this guide

This guide provides instructions for mounting the snap-on centering discs on a flexible single lead probe.

Safety messages

WARNING

Failure to follow safe installation and servicing guidelines could result in death or serious injury.

Ensure only qualified personnel perform the installation.

Required equipment
  • Tape measure
  • Cutting pliers
  • Hand file
Product recycling/disposal

Recycling of equipment and packaging should be taken into consideration and disposed of in accordance with local and national legislation or regulations.

Snap-on centering disc kits

Part number Description
03300-1658-0001 Kit, 2- to 4-in. snap-on centering disc, PEEK, 1 pc
03300-1658-0002 Kit, 2- to 4-in. snap-on centering disc, PEEK, 3 pcs
03300-1658-0003 Kit, 2- to 4-in. snap-on centering disc, PEEK, 5 pcs
Centering disc size recommendation

Figure 2-1: Top View

Centering disc size recommendation

Table 2-1: Centering Disc Dimensions

Disc size Actual disc diameter
2-in. 1.50 in. (38 mm)
3-in. 2.36 in. (60 mm)
4-in. 3.15 in. (80 mm)

Table 2-2: Centering Disc Size Recommendation

Pipe size Pipe schedule
5s, 10s, 20s 30, 40s, 40

XXS
2-in./DN50| 2-in.| 2-in.| 2-in.| N/A(1)| 2-in.| 2-in.
2½-in./DN65| 3-in.| 3-in.| 2-in.| 2-in.| 2-in.| 2-in.
3-in./DN80| 4-in.| 3-in.| 3-in.| 3-in.| 3-in.| 2-in.
3½-in./DN90| 4-in.| 4-in.| 4-in.| N/A(1)| N/A(1)| 3-in.
4-in./DN100| 4-in.| 4-in.| 4-in.| 4-in.| 4-in.| 4-in.

Mounting considerations

Figure 3-1: Minimum and Recommended Distance

Minimum and Recommended Distance

A. Minimum: 3.3 ft. (1 m) Recommended: 9.8 ft. (3 m)

Table 3-1: Maximum Number of Centering Discs

Model Water (DC = 80) Oil (DC = 2)
3300 20 5
3308 16 5
5300 20 11

Note: Use as few centering discs as possible.

Operating temperature limits

  • -67 to 392 °F (-55 to 200 °C)

Effect on measurement performance

See each product’s Product Data Sheet for more information.

Mount the snap-on centering discs

Procedure

  1. If required, cut the centering discs at the grooves.
    CAUTION
    Be careful of sharp edges. Wear protective gloves!
    Mount the snap-on centering discs

  2. Remove any sharp edges with a hand file.

  3. Snap on the centering discs on the probe.

Support

For more information: www.emerson.com

©2021 Emerson. All rights reserved.
Emerson Terms and Conditions of Sale are available upon request. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Emerson User Manuals

Related Manuals