velleman DVM811 Digital Multimeter User Manual

June 6, 2024
Velleman

DVM811


USER MANUAL

Introduction

To all residents of the European Union
Important environmental information about this product
This symbol on the device or the package indicates that disposal of the device after its lifecycle could harm the environment. Do not dispose of the unit (or batteries) as unsorted municipal waste; it should be taken to a specialized company for recycling. This device should be returned to your distributor or to a local recycling service. Respect the local environmental rules.
If in doubt, contact your local waste disposal authorities.

Thank you for choosing Velleman! Please read the manual thoroughly before bringing this device into service. If the device was damaged in transit, do not install or use it and contact your dealer.

Used Symbols

| AC (Alternating Current)
---|---

| DC (Direct Current)

| Both AC and DC


| Risk of Electric shock. A potentially hazardous voltage is possible.
**| Caution: risk of danger, refer to the user manual for safety information.
Warning : a hazardous condition or action that may result in injury or death
Caution : condition or action that may result in damage to the meter or equipment under test

**| Double insulation (class 2-protection)


| Earth

| Fuse
****| Capacitor


| Diode


| Continuity

General Guidelines

Refer to the Velleman® Service and Quality Warranty on the last pages of this manual.

| This symbol indicates: Read instructions
Not reading the instructions and manual can lead to damage, injury, or death.
---|---
| This symbol indicates: Danger
A hazardous condition or action that may result in injury or death
| This symbol indicates: Risk of danger/damage
Risk of a hazardous condition or action that may result in damage, injury, or death
| This symbol indicates: Attention; important information
Ignoring this information can lead to hazardous situations.
| WARNING: To avoid electrical shock always disconnect the test leads prior to opening the housing. To prevent fire hazards, only use fuses with the same ratings as specified in this manual.
Remark: refer to the warning on the battery compartment
| Avoid cold, heat, and large temperature fluctuations. When the unit is moved from a cold to a warm location, leave it switched off until it has reached room temperature. This is to avoid condensation and measuring errors.
|

Protect this device from shocks and abuse. Avoid brute force when operating.

| Pollution degree 2-device. For indoor use only. Keep this device away from rain, moisture, splashing, and dripping liquids. Not for industrial use. Refer to §8 Pollution degree.
| Keep the device away from children and unauthorized users.
| Risk of electric shock during operation. Be very careful when measuring live circuits.
| There are no user-serviceable parts inside the device.
Refer to an authorized dealer for service and/or spare parts.
| This is an installation category CAT III measuring instrument. Refer to §7 Overvoltage/installation category.
---|---
| Read this addendum and the manual thoroughly. Familiarise yourself with the functions of the device before actually using it.
| All modifications of the device are forbidden for safety reasons. Damage caused by user modifications to the device is not covered by the warranty.
| Only use the device for its intended purpose. Using the device in an unauthorized way will void the warranty. Damage caused by disregard of certain guidelines in this manual is not covered by the warranty and the dealer will not accept responsibility for any ensuing defects or problems.

Maintenance

There are no user-serviceable parts inside the device. Refer to an authorized dealer for service and/or spare parts.
Before performing any maintenance activities, disconnect the test leads from the jacks.
For instructions on replacing the battery or fuse, refer to §11 Battery and fuse replacement.
Do not apply abrasives or solvents to the meter. Use a damp cloth and mild detergent for cleaning purposes.

During Use

Risk of electric shock during operation. Be very careful when measuring live circuits.

  • Never exceed the limit value for protection. This limit value is listed separately in the specifications for each range of measurement.
  • Do not touch unused terminals when the meter is linked to a circuit that is being tested.
  • Never use the meter with CAT II installations when measuring voltages that might exceed the safety margin of 500 V above earth ground. Never use the meter with CAT III installations when measuring voltages that might exceed the safety margin of 300 V above earth ground.
  • Set the range selector at its highest position if the intensity of the charge to be measured is unknown beforehand.
  • Disconnect the test leads from the tested circuit before rotating the range selector in order to change functions.
  • When carrying out measurements on a TV set or switching power circuits, always remember that the meter may be damaged by any high amplitude voltage pulses at test points.
  • Always be careful when working with voltages above 60 VDC or 30 VAC rms. Keep your fingers behind the probe barriers at all times during measurement.
  • Never perform resistance, diode, or continuity measurements on live circuits. Make sure all capacitors in the circuit are depleted.

General Description

Refer to the illustration on page 2 of this manual:

  1.  Display 3 ½ digits, LCD: 32 x 14 mm
  2. Selection button
  3. Fixed test leads

Overvoltage/Installation Category

DMMs are categorized depending on the risk and severity of transient overvoltage that might occur at the point of the test. Transients are short- lived bursts of energy induced in a system, e.g. caused by a lightning strike on a power line.
The existing categories according to EN 61010-1 are:

CAT I

| A CAT I-rated meter is suitable for measurements on protected electronic circuits that are not directly connected to mains power, e.g. electronics circuits, control signals…
---|---

CAT II

| A CAT II-rated meter is suitable for measurements in CAT I- environments and mono-phase appliances that are connected to the mains by means of a plug and circuits in a normal domestic environment, provided that the circuit is at least 10 m apart from a CAT III- or 20 m apart from a CAT IV-environment. E.g. household appliances, portable tools…

CAT III

| A CAT III-rated meter is suitable for measurements in CAT I- and CAT II- environments, as well as for measurements on (fixed) mono- or poly-phased appliances which are at least 10 m apart from a CAT IV environment, and for measurements in or on distribution level equipment (fuse boxes, lighting circuits, electric ovens).

CAT IV

| A CAT IV-rated meter is suitable for measuring in CAT I-, CAT II- and CAT III-environments as well as on the primary supply level.
Note that for all measurements on equipment for which the supply cables run outdoors (either overhead or underground) a CAT IV meter must be used.

Warning:
This device was designed in accordance with EN 61010-1 installation category CAT II 500 V and CAT III 300 V. This implies that certain restrictions in use applications that are related to voltages and voltage peaks which can occur within the environment of use. Refer to the table above.
This device is only suitable for measurements up to 500 V in CAT II and up to 300 V in CAT III

Pollution Degree

IEC 61010-1 specifies different types of pollution environments, for which different protective measures are necessary to ensure safety. Harsher environments require more protection, and the protection against the pollution which is to be found in a certain environment depends mainly on the insulation and the enclosure properties. The pollution degree rating of the DVM indicates in which environment the device may be used.

Pollution degree 1| No pollution or only dry, nonconductive pollution occurs. Pollution has no influence. (only to be found in hermetically sealed enclosures)
---|---
Pollution degree 2| Only non-conductive pollution occurs. Occasionally, temporary conductivity caused by condensation is to be expected. (home and office environments fall under this category)
Pollution degree 3| Conductive pollution occurs, or dry non-conductive pollution occurs that becomes conductive due to condensation that is to be expected. (industrial environments and environments exposed to outside air – but not in contact with precipitation)
Pollution degree 4| The pollution generates persistent conductivity caused by conductive dust or by rain or snow. (exposed outdoor environments and environments where high humidity levels or high concentrations of fine particles occur)

Warning: This device was designed in accordance with EN 61010-1 pollution degree 2. This implies that certain restrictions in use applications are related to the pollution which can occur within the environment of use. Refer to the table above.
This device is only suitable for measurements in Pollution degree class 2 environments.

Specifications

This device is not calibrated when purchased!
Regulations concerning the environment of use:
Use this meter only for measurements in CAT I, CAT II, and CAT III environments (see §7).
Use this meter only in a pollution degree 2 environment (see §8).
Ideal working conditions include:
temperature: 0 °C to 40 °C (32 °F to 104 °F)
relative humidity: max. 80 %
altitude: max. 2000 m (6560 ft)
voltage ………… 500 V
fuse protection
F0.5 A/500 V, 5 x 20 mm
power supply …………1 x CR2032 (incl.)
display ……………. LCD, 1999 counts
display dimensions ……..61 x 32 mm
over-range ………….yes
continuity buzzer ………… yes
diode test ……….. yes
low-battery indication ………… yes
ranging mode ………….manual
data hold ……………..no
backlight ……………no
auto power-off ………..no
dimensions ……………… 97 x 51 x 27 mm
weight (with battery) ……..100 g
storage environment
temperature ……………-20 °C to 60 °C
humidity ……………< 90 % RH
test lead probe ………. CAT II 500 V / CAT III 300 V, 10 A; L = 70 cm

9.1 DC VOLTAGE


| Do not measure circuits that may contain voltages > 500 VDC or > 500 VAC
---|---
range| resolution| accuracy
200 mV| 0.1 mV| ****

± (1.0 % rdg + 2 digits)

2 V| 1 mV
20 V| 10 mV
200 V| 100 mV
500 V| 1 V| ± (1.2 % rdg + 2 digits)

Overload protection: 500 V DC or AC rms
Impedance: 1 MΩ
9.2 AC VOLTAGE

****| Do not measure circuits that may contain voltages

500 VDC or > 500 VAC

---|---
range| resolution| accuracy
200 V| 100 mV| ± (1.5 % rdg + 3 digits)
500 V| 1 V

Average sensing calibrated to RMS of sine wave
Frequency range: 40-500 Hz
Overload protection: 500 V DC or AC rms
Impedance: 450 kΩ
9.3 DC CURRENT


| Do not measure circuits that may contain voltages

500 VDC or > 500 VAC

---|---
range| resolution| accuracy
2000 µA| 1 µA| ± (1.5 % rdg + 2 digits)
20 mA| 10 µA
200 mA| 100 µA| ± (2.0 % rdg + 2 digits)

Overload protection: F0.5 A/500 V fuse
9.4 RESISTANCE


| Do not conduct resistance measurements on live circuits
---|---
range| resolution| accuracy
200 Ω| 0.1 Ω| ± (1.0 % rdg + 5 digits)
2000 Ω| 1 Ω| ****

± (1.0 % rdg + 3 digits)

20 kΩ| 10 Ω
200 kΩ| 100 Ω
2000 kΩ| 1 kΩ| ± (1.5 % rdg + 3 digits)

Overload protection: F0.5 A/500 V fuse
9.5 DIODE AND CONTINUITY


| Do not conduct diode or continuity measurements on live circuits
---|---
range| description| test condition


| display reads the approximate forward voltage of the diode| ****

forward DC current ± 10 µA reversed DC voltage ± 1.8 V


| built-in buzzer sounds if resistance < 50 Ω| open-circuit voltage ± 1.8 V

Overload protection: F0.5 A/500 V fuse
9.6 BATTERY TEST

range load current resolution accuracy
1.5 V 50 mA 10 mV ± (5.0 % rdg + 5 digits)
9 V 5 mA 10 mV

Operation

10.1 VOLTAGE MEASUREMENT

Do not measure circuits that may contain voltages > 500 VDC or > 500 VAC
Use extreme caution when measuring voltages higher than 60 VDC or 30 VAC rms.
Always place your fingers behind the protective edges of the test probes while measuring!

  1. Set the rotary switch in the desired V= or V~ position.
  2. Connect the test leads to the source being measured.
  3. Read the voltage value and the polarity of the red test lead on the LCD display.

Notes

  • If the range is not known beforehand, set the selector switch to a high range and lower gradually.
  • An over-range is indicated by 1 or -1. Set to a higher range.

10.2 DC CURRENT MEASUREMENT
Do not measure circuits that may contain voltages > 500 VDC or > 500 VAC
Use extreme caution when measuring voltages higher than 60 VDC or 30 VAC rms.
Always place your fingers behind the protective edges of the test probes while measuring!

  1. Set the rotary switch in the desired mA= position.
  2. Open the circuit in which the current is to be measured and connect the test leads to the circuit IN SERIES.
  3. Read the current value and the polarity of the red lead connection on the LCD display.

Notes

  • If the range is not known beforehand, set the selector switch to a high range and lower gradually.
  • An over-range is indicated by 1 or -1. Set to a higher range.
  • The maximum input current is 200 mA.

10.3 RESISTANCE MEASUREMENT
****Do not conduct resistance measurements on live circuits. Make sure all capacitors in the circuit are depleted.

  1. Set the rotary switch in the desired Ω position.
  2. Connect the test leads to the resistor to be measured and read the LCD display.

Notes

  • If the resistance being measured is connected to a circuit, turn off the power and discharge all capacitors before applying the test probes.
  • An over-range is indicated by 1 or -1. Set to a higher range.

10.4 DIODE AND CONTINUITY TEST

****Do not conduct diode or continuity measurements on live circuits. Make sure all capacitors in the circuit are depleted.

  1. Set the rotary switch in the (diode) or (continuity) position.
  2. For measuring the diode, connect the red test lead to the anode of the diode to be tested and the black test leads to the cathode of the diode. The approx. forward voltage drop of the diode will be displayed. For measuring continuity, connect the test leads to two points of the circuit to be tested. If continuity exists, the built-in buzzer will sound.

10.5 BATTERY TEST

  1. Set the rotary switch in the 1.5V or 9V position.
  2. Connect the test leads to the battery to be tested and read the LCD display.

Battery and Fuse Replacement

WARNING: To avoid electrical shock always disconnect the test leads prior to opening the housing. To prevent fire hazards, only use fuses with the same ratings as specified in this manual. Remark: refer to the warning on the battery compartment
There are no user-serviceable parts inside the device. Refer to an authorized dealer for service and/or spare parts.
****Disconnect the test leads from the test points and remove the test leads from the measuring terminals before replacing the batteries or fuses.

  • When” ” is displayed, the battery should be replaced.
  • Fuses rarely need replacement and blown fuses almost always result from human error.

To replace the battery:

  • Switch of the meter.
  • Remove the screw at the back of the case and gently open the housing.
  • Remove the old battery and insert a new one.
  • Close the housing and fasten the screw.

Battery: 1x CR2032, make sure to respect the polarity
Fuses: F0.5 A/500 V, 5 x 20 mm
Make sure the meter is closed tight and put the protective edge back in place before using the meter.

Troubleshooting

If the device acts abnormal while measuring, this means that the internal fuse is defective.
Keep in mind that a low battery level could lead to incorrect measurements. Replace the battery on a regular basis. (tip: the reduced luminosity of the backlight/LCD display indicates a low battery level)
Use this device with original accessories only. Velleman nv cannot be held responsible in the event of damage or injury resulting from (incorrect) use of this device. For more info concerning this product and the latest version of this manual, please visit our websitewww.velleman.eu. The information in this manual is subject to change without prior notice.

© COPYRIGHT NOTICE
The copyright to this manual is owned by Velleman nv. All worldwide rights reserved. No part of this manual may be copied, reproduced, translated, or reduced to any electronic medium or otherwise without the prior written consent of the copyright holder.

Velleman® Service and Quality Warranty Since its foundation in 1972, Velleman® acquired extensive experience in the electronics world and currently distributes its products in over 85 countries. All our products fulfill strict quality requirements and legal stipulations in the EU. In order to ensure the quality, our products regularly go through an extra quality check, both by an internal quality department and by specialized external organizations. If, all precautionary measures notwithstanding, problems should occur, please make an appeal to our warranty (see guarantee conditions).
General Warranty Conditions Concerning Consumer Products (for EU):

  • All consumer products are subject to a 24-month warranty on production flaws and defective material as from the original date of purchase.

  • Velleman® can decide to replace an article with an equivalent article or to refund the retail value totally or partially when the complaint is valid and a free repair or replacement of the article is impossible, or if the expenses are out of proportion. You will be delivered a replacing article or a refund at the value of 100% of the purchase price in case of a flaw that occurred in the first year after the date of purchase and delivery, or a replacing article at 50% of the purchase price or a refund at the value of
    50% of the retail value in case of a flaw occurred in the second year after the date of purchase and delivery.

  • Not covered by warranty:

– all direct or indirect damage caused after delivery to the article (e.g. by oxidation, shocks, falls, dust, dirt, humidity…), and by the article, as well as its contents (e.g. data loss), compensation for loss of profits;
– consumable goods, parts, or accessories that are subject to an aging process during normal use, such as batteries (rechargeable, non-rechargeable, built-in or replaceable), lamps, rubber parts, drive belts… (unlimited list);
– flaws resulting from fire, water damage, lightning, accident, natural disaster, etc.…;
– flaws caused deliberately, negligently, or resulting from improper handling, negligent maintenance, abusive use or use contrary to the manufacturer’s instructions;
– damage caused by a commercial, professional or collective use of the article (the warranty validity will be reduced to six (6) months when the article is used professionally);
– damage resulting from inappropriate packing and shipping of the article;
– all damage caused by modification, repair, or alteration performed by a third party without written permission by Velleman®.

  • Articles to be repaired must be delivered to your Velleman® dealer, solidly packed (preferably in the original packaging), and be completed with the original receipt of purchase and a clear flaw description.
  • Hint: In order to save on cost and time, please re-read the manual and check if the flaw is caused by obvious causes prior to presenting the article for repair. Note that returning a non-defective article can also involve handling costs.
  • Repairs occurring after warranty expiration is subject to shipping costs.

The above conditions are without prejudice to all commercial warranties. The above enumeration is subject to modification according to the article (see article’s manual).

Made in PRC
Imported by Velleman nv
Legen Heirweg 33, 9890 Gavere, Belgium
www.velleman.eu

Documents / Resources

| velleman DVM811 Digital Multimeter [pdf] User Manual
DVM811, Digital Multimeter, DVM811 Digital Multimeter, Multimeter
---|---

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals