apogee INSTRUMENTS MQ-100 Quantum Meter Owner’s Manual
- June 6, 2024
- apogee INSTRUMENTS
Table of Contents
OWNER’S MANUAL
QUANTUM METER
Models MQ-100, MQ-200, and MQ-300 Series
Rev: 28-Oct-2020
CERTIFICATE OF COMPLIANCE
EU Declaration of Conformity
This declaration of conformity is issued under the sole responsibility of the
manufacturer:
Apogee Instruments, Inc.
721 W 1800 N
Logan, Utah 84321
USA
for the following product(s):
Models: MQ-100, MQ-200, MQ-301, MQ-303, MQ-306
Type: Quantum Meter
The object of the declaration described above is in conformity with the relevant Union harmonization legislation:
2014/30/EU
2011/65/EU
2015/863/EU| Electromagnetic Compatibility (EMC) Directive
Restriction of Hazardous Substances (RoHS 2) Directive
Amending Annex II to Directive 2011/65/EU (RoHS 3)
---|---
Standards referenced during compliance assessment:
EN 61326-1:2013
EN 50581:2012| Electrical equipment for measurement, control, and laboratory
use – EMC requirements
Technical documentation for the assessment of electrical and electronic
products with respect to the restriction of hazardous substances
---|---
Please be advised that based on the information available to us from our raw
material suppliers, the products manufactured by us do not contain, as
intentional additives, any of the restricted materials including lead (see
note below), mercury, cadmium, hexavalent chromium, polybrominated biphenyls
(PBB), polybrominated biphenyls (PBDE), bis(2-Ethylhexyl) phthalate (DEHP),
butyl benzyl phthalate (BBP), dibutyl phthalate (DBP), and biphenyls phthalate
(DIBP). However, please note that articles containing greater than 0.1% lead
concentration are RoHS 3 compliant using exemption 6c.
Further note that Apogee Instruments does not specifically run any analysis on
our raw materials or end products for the presence of these substances, but
rely on the information provided to us by our material suppliers.
Signed for and on behalf of:
Apogee Instruments, February 2021
Bruce Bugbee
President
Apogee Instruments, Inc.
INTRODUCTION
Radiation that drives photosynthesis is called photosynthetically active
radiation (PAR) and is typically defined as total radiation across a range of
400 to 700 nm. PAR is often expressed as photosynthetic photon flux density
(PPFD): photon flux in units of micromoles per square meter per second (µmol
m, equal to microEinsteins per square meter per second) summed from 400 to 700
nm (total number of photons from 400 to 700 nm). While Einsteins and
micromoles are equal (one Einstein = one mole of photons), the Einstein is not
an SI unit, so expressing PPFD as µmol m -2 -1 s -2 -1 s is preferred.
The acronym PPF is also widely used and refers to the photosynthetic photon
flux. The acronyms PPF and PPFD refer to the same parameter. The two terms
have co-evolved because there is not a universal definition of the term
“flux”. Some physicists define flux as per unit area per unit time. Others
define flux only as per unit time. We have used PPFD in this manual because we
feel that it is better to be more complete and possibly redundant.
Sensors that measure PPFD are often called quantum sensors due to the
quantized nature of radiation. A quantum refers to the minimum quantity of
radiation, one photon, involved in physical interactions (e.g., absorption by
photosynthetic pigments). In other words, one photon is a single quantum of
radiation.
Typical applications of quantum sensors include incoming PPFD measurement over
plant canopies in outdoor environments or in greenhouses and growth chambers,
and reflected or under-canopy (transmitted) PPFD measurement in the same
environments.
Apogee Instruments MQ series quantum meters consist of a handheld meter and a
dedicated quantum sensor that is integrated into the top of the meter housing
(MQ-100) or connected by cable to an anodized aluminum housing (MQ-200 and
MQ-300 series). Integrated and separate sensors consist of a cast acrylic
diffuser (filter), photodiode, and are potted solid with no internal air
space. MQ series quantum meters provide a real-time PPFD reading on the LCD
display and offer measurements for both sunlight and electric light
calibrations (menu selectable) that determine the radiation incident on a
planar surface (does not have to be horizontal), where the radiation emanates
from all angles of a hemisphere. MQ series quantum meters include manual and
automatic data logging features for making spot-check measurements or
calculating daily light integral (DLI).
SENSOR MODELS
Apogee MQ series quantum meters covered in this manual are self-contained and
come complete with a handheld meter and sensor.
Line quantum meters, MQ-300 series, provide spatially averaged PPFD
measurements. All sensors along the length of the line are connected in
parallel, and as a result, Apogee line quantum meters display PPFD values that
are averaged from the location of the individual sensors.
The sensor model number and serial number are located on a label on the backside of the handheld meter.
SPECIFICATIONS
| MQ-100| MQ-200| MQ-301|
MO-303/306
---|---|---|---|---
Calibration Uncertainty| ± 5 % (see calibration Traceability below)
Measurement| Less ess than 1 %
Long-term Drift (Non-stability)| Less than 2 % per year
Non-linearity| Less than 1 % (up to 3000 limot m-2 s-1)
Response Time| Less than 1 ms
Field of View| 180°
Spectral Range| 410 to 655 nm (wavelengths where the response is greater than
50 % of the maximum; see Spectral Response below)
Directional (Cosine) Response| ± 5 % at 75° zenith angle (see Cosine Response
below)
Temperature Response| 0.06 ± 0.06 % per C (see Temperature Response below)
Operating Environment| 0 to 50 C; less than 90 % non-condensing relative
humidity up to 30 C; less than 70 % non-condensing relative humidity from 30
to 50 C; separate sensors can be submerged in water up to depths of 30 m
Meter Dimensions| 126 mm length; 70 mm width; 24 mm height
Sensor Dimensions| Integrated w/Meter| 24 mm diameter;
33 mm height| 70 mm length; 15mm width; 15mm height| 50 mm length; 15mm width;
15mm height
Mass| 150 g| 180 g| 380 g| 300 g
Cable| 2 m of two-conductor, shielded, twisted-pair wire; TPR jacket (high
water resistance, high UV stability, flexibility in cold conditions)
Calibration Traceability
Apogee MQ series quantum meters are calibrated through side-by-side comparison
to the mean of four transfer standard quantum sensors under a reference lamp.
The reference quantum sensors are recalibrated with a 200 W quartz halogen
lamp traceable to the National Institute of Standards and Technology (NIST).
Spectral Response
Mean spectral response of six SQ-100 series quantum sensors ( error bars represent two standard deviations above and below mean ) compared to defined plant response to photons. Spectral response measurements were made at 10 nm increments across a wavelength range of 300 to 800 nm with a monochromator and an attached electric light source. Measured spectral data from each quantum sensor were normalized by the measured spectral response of the monochromator/electric light combination, which was measured with a spectroradiometer.
Temperature Response
Mean temperature response of eight SQ-100 series quantum sensors ( errors bars represent two standard deviations above and below mean ). Temperature response measurements were made at 10 C intervals across a temperature range of approximately 10 to 40 C in a temperature-controlled chamber under a fixed, broad-spectrum, electric lamp. At each temperature set point, a spectroradiometer was used to measure light intensity from the lamp and all quantum sensors were compared to the spectroradiometer. The spectroradiometer was mounted external to the temperature control chamber and remained at room temperature during the experiment.
Cosine Response
Directional (cosine) response is defined as the measurement error at a specific angle of radiation incidence. Error for Apogee SQ-100 series quantum sensors is approximately ± 2 % and ± 5 % at solar zenith angles of 45° and 75°, respectively.
Mean directional (cosine) response of six Apogee SQ-100 series quantum sensors. Directional response measurements were made on the rooftop of the Apogee building in Logan, Utah. The directional response was calculated as the relative difference of SQ-500 quantum sensors from the mean of replicate reference quantum sensors (LI-COR models LI-190 and LI190R, Kipp & Zonen model PQS 1). Data were also collected in the laboratory using a reference lamp and positioning the sensor at varying angles.
DEPLOYMENT AND INSTALLATION
Apogee MQ series quantum meters are designed for spot-check measurements and calculation of daily light integral (DLI; the total number of photons incident on a planar surface over the course of a day) through the built-in logging feature. To accurately measure the PFFD incidents on a horizontal surface, the sensor must be level. For this purpose, each MQ model comes with a different option for mounting the sensor to a horizontal plane.
The AL-210 leveling plate is recommended for use with the MQ-100.
The AL-100 leveling plate is recommended for use with the MQ-200. To facilitate mounting to a cross arm, the AL-120 mounting bracket is recommended.
MQ-300 series line quantum sensors are leveled using the built-in bubble level located in the handle of the sensor. In addition to leveling, all sensors should also be mounted such that obstructions (e.g., weather station tripod/tower or other instrumentation) do not shade the sensor.
OPERATION AND MEASUREMENT
MQ series quantum meters are designed with a user-friendly interface allowing quick and easy measurements.
To power the meter, slide the included battery (CR2320) into the battery
holder, after removing the battery door from the meter’s back panel. The
positive side (designated by a “+” sign) should be facing out from the meter
circuit board.
Press the power button to activate the LCD display. After two minutes of non-
activity, the meter will revert to sleep mode and the display will shut off to
conserve battery life.
Press the mode button to access the main menu, where the appropriate
calibration (sunlight or electric light) and manual or automatic logging are
selected, and where the meter can be reset.
Press the sample button to log a reading while taking manual measurements.
Press the up button to make selections in the main menu. This button is also
used to view and scroll through the logged measurements on the LCD display.
Press the down button to make selections in the main menu. This button is also
used to view and scroll through the logged measurements on the LCD display.
The LCD display consists of the total number of logged measurements in the upper right-hand corner, the real-time PPFD value in the center, and the selected menu options along the bottom.
Calibration: To choose between sunlight and electric light calibration, push the mode button once and use the up/down buttons to make the appropriate selection (SUN or ELEC). Once the desired mode is blinking, press the mode button three more times to exit the menu.
Logging: To choose between manual or automatic logging, push the mode button once and use the up/down buttons to make the appropriate selection (SMPL or LOG). Once the desired mode is blinking, press the mode button two more times to exit the menu. When in SMPL mode press the sample button to record up to 99 manual measurements (a counter in the upper right-hand corner of the LCD display indicates the total number of saved measurements). When in LOG mode the meter will power on/off to make a measurement every 30 seconds. Every 30 minutes the meter will average the sixty 30 second measurements and record the averaged value to memory. The meter can store up to 99 averages and will start to overwrite the oldest measurement once there are 99 measurements. For every 48 averaged measurements (making a 24 hour period), the meter will also store an integrated daily total in moles per meter squared per day (mol m -2 -1 d ).
Reset: To reset the meter, in either SMPL or LOG mode, push the mode button three times (RUN should be blinking), then while pressing the down button, press the mode button once. This will erase all of the saved measurements in memory, but only for the selected mode. That is, performing a reset when in SMPL mode will only erase the manual measurements, and performing a reset when in LOG mode will only erase the automatic measurements.
Review/Download Data: Each of the logged measurements in either SMPL or
LOG mode can be reviewed on the LCD display by pressing the up/down buttons.
To exit and return to the real-time readings, press the sample button. Note
that the integrated daily total values are not accessible through the LCD and
can only be viewed by downloading to a computer.
Downloading the stored measurements will require the AC-100 communication
cable and software (sold separately). The meter outputs data using the UART
protocol and requires the AC-100 to convert from UART to USB, so standard USB
cables will not work. Set up instructions and software can be downloaded from
the Apogee website (http://www.apogeeinstruments.com/ac-100-communcation-
cable/).
Spectral Error
The combination of diffuser transmittance, interference filter transmittance,
and photodetector sensitivity yields the spectral response of a quantum
sensor. A perfect photodetector/filter/diffuser combination would exactly
match the defined plant photosynthetic response to photons (equal weighting to
all photons between 400 and 700 nm, no weighting of photons outside this
range), but this is challenging in practice. A mismatch between the defined
plant photosynthetic response and sensor spectral response results in a
spectral error when the sensor is used to measure radiation from sources with
a different spectrum than the radiation source used to calibrate the sensor
(Federer and Tanner, 1966; Ross and Sulev, 2000).
Spectral errors for PPFD measurements made under common radiation sources for
growing plants were calculated for Apogee SQ-100/300 and SQ-500 series quantum
sensors using the method of Federer and Tanner (1966). This method requires
PPFD weighting factors (defined plant photosynthetic response), measured
sensor spectral response (shown in Spectral Response section on page 7), and
radiation source spectral outputs (measured with a spectroradiometer). Note,
this method calculates spectral error only and does not consider calibration,
directional (cosine), temperature, and stability/drift errors. Spectral error
data (listed in the table below) indicate errors less than 5 % for sunlight in
different conditions (clear, cloudy, reflected from plant canopies,
transmitted below plant canopies) and common broad-spectrum electric lamps
(cool white fluorescent, metal halide, high-pressure sodium), but larger
errors for different mixtures of light-emitting diodes (LEDs) for the SQ-100
series sensors. Spectral errors for the SQ-500 series sensors are smaller than
those for SQ-100 series sensors because the spectral response of SQ-500 series
sensors is a closer match to the defined plant photosynthetic response.
Quantum sensors are the most common instrument for measuring PPFD because they
are about an order of magnitude lower cost the spectroradiometers, but
spectral errors must be considered. The spectral errors in the table below can
be used as correction factors for individual radiation sources.
Spectral Errors for PPFD Measurements with Apogee SQ-100 and SQ-500 Series Quantum Sensors
Radiation Source (Error Calculated Relative to Sun, Clear Sky)|
SQ-100/300 Series
PPFD Error [%]| SQ-500 Series
PPFD Error [%]
---|---|---
Sun (Clear Sky)| 0| 0
Sun (Cloudy Sky)| 0.2| 0.1
Reflected from Grass Canopy| 3.8| -0.3
Transmitted below Wheat Canopy| 4.5| 0.1
Cool White Fluorescent (T5)| 0| 0.1
Metal Halide| -2.8| 0.9
Ceramic Metal Halide| -16.1| 0.3
High-Pressure Sodium| 0.2| 0.1
Blue LED (448 nm peak, 20 nm full-width half-maximum)| -10.5| -0.7
Green LED (524 nm peak, 30 nm full-width half-maximum)| 8.8| 3.2
Red LED (635 nm peak, 20 nm full-width half-maximum)| 2.6| 0.8
Red LED (667 nm peak, 20 nm full-width half-maximum)| -62.1| 2.8
Red, Blue LED Mixture (80 % Red, 20 % Blue)| -72.8| -3.9
Red, Blue, White LED Mixture (60 % Red, 25 % White, 15 % Blue)| -35.5| -2
Cool White LED| -3.3| 0.5
Warm White LED| -8.9| 0.2
Federer, C.A., and C.B. Tanner, 1966. Sensors for measuring light available
for photosynthesis. Ecology 47:654-657.
Ross, J., and M. Sulev, 2000. Sources of errors in measurements of PAR.
Agricultural and Forest Meteorology 100:103-125.
Spectral Errors for PPFD and YPFD Measurements with Apogee MQ Series Quantum Meters
Radiation Source (Error Calculated Relative to Sun, Clear Sky)| PPFD
Error [%]| YPFD Error [%]
---|---|---
Sun (Clear Sky)| 0| 0
Sun (Cloudy Sky)| 1.4| 1.6
Reflected from Grass Canopy| 5.7| -6.3
Reflected from Deciduous Canopy| 4.9| -7
Reflected from Conifer Canopy| 5.5| -6.8
Transmitted below Grass Canopy| 6.4| -4.5
Transmitted below Deciduous Canopy| 6.8| -5.4
Transmitted below Conifer Canopy| 5.3| 2.6
Radiation Source (Error Calculated Relative to Cool White
Fluorescent, T5)| |
Cool White Fluorescent (T5)| 0| 0
Cool White Fluorescent (T8)| -0.3| -1.2
Cool White Fluorescent (T12)| -1.4| -2
Compact Fluorescent| -0.5| -5.3
Metal Halide| -3.7| -3.7
Ceramic Metal Halide| -6| -6.4
High-Pressure Sodium| 0.8| -7.2
Blue LED (448 nm peak, 20 nm full-width half-maximum)| -12.7| 8
Green LED (524 nm peak, 30 nm full-width half-maximum)| 8| 26.2
Red LED (635 nm peak, 20 nm full-width half-maximum)| 4.8| -6.2
Red, Blue LED Mixture (85 % Red, 15 % Blue)| 2.4| -4.4
Red, Green, Blue LED Mixture (72 % Red, 16 % Green, 12 % Blue)| 3.4| 0.2
Cool White Fluorescent LED| -4.6| -0.6
Neutral White Fluorescent LED| -6.7| -5.2
Warm White Fluorescent LED| -10.9| -13
Quantum sensors can be a very practical means of measuring PPFD and YPFD from multiple radiation sources, but spectral errors must be considered. The spectral errors in the table above can be used as correction factors for individual radiation sources.
Yield Photon Flux Density (YPFD) Measurements Photosynthesis in plants does not respond equally to all photons. Relative quantum yield (plant photosynthetic efficiency) is dependent on wavelength (green line in the figure below) (McCree, 1972a; Inada, 1976). This is due to the combination of spectral absorptivity of plant leaves (absorptivity is higher for blue and red photons than green photons) and absorption by non-photosynthetic pigments. As a result, photons in the wavelength range of approximately 600-630 nm are the most efficient.
Defined plant response to photons (black line, weighting factors used to
calculate PPFD), measured plant response to photons (green line, weighting
factors used to calculate YPFD), and SQ-100 series and SQ-300 series quantum
sensor response to photons (sensor spectral response).
One potential definition of PAR is weighting photon flux density in units of
µmol m -2 -1 s at each wavelength between 300 and 800 nm by measuring relative
quantum yield and summing the result. This is defined as yield photon flux
density (YPFD, units of µmol m -2 -1 s ) (Sager et al., 1988). There are
uncertainties and challenges associated with this definition of PAR.
Measurements used to generate the relative quantum yield data were made on
single leaves under low radiation levels and at short time scales (McCree,
1972a; Inada, 1976). Whole plants and plant canopies typically have multiple
leaf layers and are generally grown in the field or greenhouse over the course
of an entire growing season. Thus, actual conditions plants are subject to are
likely different than those the single leaves were in when measurements were
made by McCree (1972a) and Inada (1976). In addition, the relative quantum
yield shown in the figure above is the mean from twenty-two species grown in
the field (McCree, 1972a). Mean relative quantum yield for the same species
grown in growth chambers was similar, but there were differences, particularly
at shorter wavelengths (less than 450 nm). There was also some variability
between species (McCree, 1972a; Inada, 1976).
McCree (1972b) found that equally weighting all photons between 400 and 700 nm
and summing the result, defined as photosynthetic photon flux density (PPFD,
in units of µmol m -2 -1 s ), was well correlated to photosynthesis, and very
similar to the correlation between YPFD and photosynthesis. As a matter of
practicality, PPFD is a simpler definition of PAR. At the same time as
McCree’s work, others had proposed PPFD as an accurate measure of PAR and
built sensors that approximated the PPFD weighting factors (Biggs et al.,
1971; Federer and Tanner, 1966). The correlation between PPFD and YPFD
measurements for several radiation sources is very high (figure below), as an
approximation, YPFD = 0.9PPFD. As a result, almost universally PAR is defined
as PPFD rather than YPFD, although YPFD has been used in some studies. The
only radiation sources are shown (figure below) that don’t fall on the
regression line are the high-pressure sodium (HPS) lamp, reflection from a
plant canopy, and transmission below a plant canopy. A large fraction of
radiation from HPS lamps is in the red range of wavelengths where the YPFD
weighting factors (measured relative quantum yield) are at or near one. The
factor for converting PPFD to YPFD for HPS lamps is 0.95, rather than 0.90.
The factor for converting PPFD to YPFD for reflected and transmitted photons
is 1.00.
Correlation between photosynthetic photon flux density (PPFD) and yield photon flux density (YPFD) for multiple different radiation sources. YPFD is approximately 90 % of PPFD. Measurements were made with a spectroradiometer (Apogee Instruments model PS-200) and weighting factors shown in the previous figure were used to calculate PPFD and YPFD.
Biggs, W., A.R. Edison, J.D. Eastin, K.W. Brown, J.W. Maranville, and M.D.
Clegg, 1971. Photosynthesis light sensor and meter. Ecology 52:125-131.
Federer, C.A., and C.B. Tanner, 1966. Sensors for measuring light available
for photosynthesis. Ecology 47:654-657.
Inada, K., 1976. Action spectra for photosynthesis in higher plants. Plant and
Cell Physiology 17:355-365.
McCree, K.J., 1972a. The action spectrum, absorptance, and quantum yield of
photosynthesis in crop plants. Agricultural Meteorology 9:191-216.
Underwater Measurements and Immersion Effect
McCree, K.J., 1972b. Test of current definitions of photosynthetically active
radiation against leaf photosynthesis data. Agricultural Meteorology
10:443-453.
When a quantum sensor that was calibrated in the air is used to make
underwater measurements, the sensor reads Sager, J.C., W.O. Smith, J.L.
Edwards, and K.L. Cyr, 1988. Photosynthetic efficiency and phytochrome
photoequilibria determination using spectral data. Transactions of the ASAE
31:1882-1889.
Immersion Effect
Correction Factor When a radiation sensor is submerged in water, more of the
incident radiation is backscattered out of the diffuser than when the sensor
is in the air (Smith, 1969; Tyler and Smith, 1970). This phenomenon is caused
by the difference in the refractive index for air (1.00) and water (1.33) and
is called the immersion effect. Without correction for the immersion effect,
radiation sensors calibrated in the air can only provide relative values
underwater (Smith, 1969; Tyler and Smith, 1970). Immersion effect correction
factors can be derived by making measurements in air and at multiple water
depths at a constant distance from a lamp in a controlled laboratory setting.
Apogee SQ-100 series and SQ-300 series quantum sensors have an immersion
effect correction factor of 1.08. This correction factor should be multiplied
by PPFD measurements made underwater to yield accurate PPFD.
Further information on underwater measurements and the immersion effect can be
found on the Apogee webpage (http://www.apogeeinstruments.com/underwater-par-
measurements/).
Smith, R.C., 1969. An underwater spectral irradiance collector. Journal of
Marine Research 27:341-351.
Tyler, J.E., and R.C. Smith, 1970. Measurements of Spectral Irradiance
Underwater. Gordon and Breach, New York, New York. 103 pages
MAINTENANCE AND RECALIBRATION
Blocking of the optical path between the target and detector can cause low readings. Occasionally, accumulated materials on the diffuser of the upward- looking sensor can block the optical path in three common ways:
- Moisture or debris on the diffuser.
- Dust during periods of low rainfall.
- Salt deposit accumulation from the evaporation of sea spray or sprinkler irrigation water.
Apogee Instruments’ upward-looking sensors have a domed diffuser and housing
for improved self-cleaning from rainfall, but active cleaning may be
necessary. Dust or organic deposits are best removed using water, or window
cleaner, and a soft cloth or cotton swab. Salt deposits should be dissolved
with vinegar and removed with a cloth or cotton swab. Salt deposits cannot be
removed with solvents such as alcohol or acetone. Use only gentle pressure
when cleaning the diffuser with a cotton swab or soft cloth to avoid
scratching the outer surface. The solvent should be allowed to do the
cleaning, not mechanical force. Never use abrasive material or cleaner on the
diffuser.
Although Apogee sensors are very stable, nominal calibration drift is normal
for all research-grade sensors. To ensure maximum accuracy, recalibration
every two years is recommended. Longer time periods between recalibration may
be warranted depending on tolerances. See the Apogee webpage for details
regarding the return of sensors for recalibration
(http://www.apogeeinstruments.com/tech-support-recalibration-repairs/).
To determine if your sensor needs recalibration, the Clear Sky Calculator
(www.clearskycalculator.com) website and/or smartphone app can be used to
indicate the total shortwave radiation incident on a horizontal surface at any
time of day at any location in the world. It is most accurate when used near
solar noon in spring and summer months, where accuracy over multiple clear and
unpolluted days is estimated to be ± 4 % in all climates and locations around
the world. For best accuracy, the sky must be completely clear, as reflected
radiation from clouds causes incoming radiation to increase above the value
predicted by the clear sky calculator. Measured values of total shortwave
radiation can exceed values predicted by the Clear Sky Calculator due to
reflection from thin, high clouds and edges of clouds, which enhances incoming
shortwave radiation. The influence of high clouds typically shows up as spikes
above clear sky values, not a constant offset greater than clear sky values.
To determine recalibration needs, input site conditions into the calculator
and compare total shortwave radiation measurements to calculated values for a
clear sky. If sensor shortwave radiation measurements over multiple days near
solar noon are consistently different than calculated values (by more than 6
%), the sensor should be cleaned and re-leveled. If measurements are still
different after a second test, email
calibration@apogeeinstruments.com
to discuss test results and the possible return of sensor(s).
Steps to Replace a Handheld Meter Battery
-
Use a Phillips head screwdriver to remove the screw from the battery cover.
-
Remove the battery cover by slightly lifting and sliding the outer edge of the cover away from the meter.
-
Use your thumb to slide the battery out of the battery holder.
a. If the battery is difficult to move, turn the meter on its side so that the opening for the batter is facing downward and tap the meter downward against an open palm to dislodge the battery enough that it can be removed as described in step 3. -
To place the battery back in, simply slide it back into the battery holder with the flat side of the battery facing up.
NOTE: The battery cradle can be damaged by using an incorrectly sized battery. If the battery cradle is damaged, the circuit board will need to be replaced. To avoid this costly problem, use only a CR2320 battery.
Homepage of the Clear Sky Calculator. Two calculators are available: one for quantum sensors (PPFD) and one for pyranometers (total shortwave radiation).
Clear Sky Calculator for quantum sensors. Site data are input in blue cells in the middle of the page and an estimate of PPFD is returned on the right-hand side of the page.
TROUBLESHOOTING AND CUSTOMER SUPPORT
Verify Functionality
Pressing the power button should activate the LCD and provide a real-time PPFD
reading. Direct the sensor head toward a light source and verify the PPFD
reading responses. Increase and decrease the distance from the sensor to the
light source to verify that the reading changes proportionally (decreasing
PPFD with increasing distance and increasing PPFD with decreasing distance).
Blocking all radiation from the sensor should force the PPFD reading to zero.
Battery Life
When the meter is maintained properly the coin cell battery (CR2320) should
last for many months, even after continuous use. The low battery indicator
will appear in the upper left-hand corner of the LCD display when the battery
voltage drops below 2.8 V DC. The meter will still function correctly for some
time, but once the battery is drained the pushbuttons will no longer respond
and any logged measurements will be lost.
Pressing the power button to turn off the meter will actually put it in sleep
mode, where there is still a slight amount of current draw. This is necessary
to maintain the logged measurements in memory. Therefore, it is recommended to
remove the battery when storing the meter for many months at a time, in order
to preserve battery life.
Low-Battery Error after Battery Replacement
A master reset will usually correct this error, please see the master reset
section for details and cautions. If a master reset does not remove the low
battery indicator, please double-check that the voltage of your new battery is
above 2.8 V, this is the threshold for the indicator to turn on.
Master Reset
If a meter ever becomes non-responsive or experiences anomalies, such as a low
battery indicator even after replacing the old battery, a master reset can be
performed that may correct the problem. Note that a master reset will erase
all logged measurements from memory.
Step 1: Press the power button so that the LCD display is activated.
Step 2: Slide the battery out of the holder, which will cause the LCD
display to fade out.
Step 3: After a few seconds, slide the battery back into the holder.
The LCD display will flash all segments and then show a revision number (e.g.
“R1.0”). This indicates the master reset was performed and the display should
return to normal.
Error Codes and Fixes
Error codes will appear in place of the real-time reading on the LCD display
and will continue to flash until the problem is corrected. Contact Apogee if
the following fixes do not rectify the problem.
Err 1: battery voltage out of range. Fix: replace the CR2320 battery
and perform a master reset.
Err 2: sensor voltage out of range. Fix: perform a master reset.
Err 3: not calibrated. Fix: perform a master reset.
Err 4: CPU voltage below minimum. Fix: replace the CR2320 battery and
perform a master reset.
Modifying Cable Length
Although it is possible to splice additional cable to the separate sensor of
the appropriate MQ model, note that the cable wires are soldered directly into
the circuit board of the meter. Care should be taken to remove the back panel
of the meter in order to access the board and splice on the additional cable,
otherwise, two splices would need to be made between the meter and sensor
head. See Apogee webpage for further details on how to extend sensor cable
length: (http://www.apogeeinstruments.com/how-to-make-a-weatherproof-cable-
splice/).
Unit Conversion Charts Apogee MQ series quantum sensors are calibrated to
measure PPFD in units of µmol m -2 -1 s. Units other than photon flux density
(e.g., energy-flux density, illuminance) may be required for certain
applications. It is possible to convert the PPFD value from a quantum sensor
to other units, but it requires the spectral output of the radiation source of
interest. Conversion factors for common radiation sources can be found on the
Unit Conversions page in the Support Center on the Apogee website
(http://www.apogeeinstruments.com/unit-conversions/). A spreadsheet to
convert PPFD to energy flux density or illuminance is also provided on the
Unit Conversions page in the Support Center on the Apogee website
(http://www.apogeeinstruments.com/content/PPFD-to-
IlluminanceCalculator.xls).
RETURN AND WARRANTY POLICY
RETURN POLICY
Apogee Instruments will accept returns within 30 days of purchase as long as
the product is in new condition (to be determined by Apogee). Returns are
subject to a 10 % restocking fee.
WARRANTY POLICY
What is Covered
All products manufactured by Apogee Instruments are warranted to be free from
defects in materials and craftsmanship for a period of four (4) years from the
date of shipment from our factory. To be considered for warranty coverage an
item must be evaluated by Apogee.
Products not manufactured by Apogee (spectroradiometers, chlorophyll content
meters, EE08-SS probes) are covered for a period of one (1) year.
What is Not Covered
The customer is responsible for all costs associated with the removal,
reinstallation, and shipping of suspected warranty items to our factory.
The warranty does not cover equipment that has been damaged due to the
following conditions:
- Improper installation or abuse.
- Operation of the instrument outside of its specified operating range.
- Natural occurrences such as lightning, fire, etc.
- Unauthorized modification.
- Improper or unauthorized repair.
Please note that nominal accuracy drift is normal over time. Routine recalibration of sensors/meters is considered part of proper maintenance and is not covered under warranty.
Who is Covered
This warranty covers the original purchaser of the product or other party who
may own it during the warranty period.
What Apogee Will Do
At no charge Apogee will:
- Either repair or replace (at our discretion) the item under warranty.
- Ship the item back to the customer by the carrier of our choice.
Different or expedited shipping methods will be at the customer’s expense.
How To Return An Item
-
Please do not send any products back to Apogee Instruments until you have received a Return Merchandise Authorization (RMA) number from our technical support department by submitting an online RMA form at www.apogeeinstruments.com/tech-support-recalibration-repairs/. We will use your RMA number for tracking the service item. Call 435-245-8012 or email techsupport@apogeeinstruments.com with questions.
-
For warranty evaluations, send all RMA sensors and meters back in the following condition: Clean the sensor’s exterior and cord. Do not modify the sensors or wires, including splicing, cutting wire leads, etc. If a connector has been attached to the cable end, please include the mating connector – otherwise, the sensor connector will be removed in order to complete the repair/recalibration. Note: When sending back sensors for routine calibration that have Apogee’s standard stainless-steel connectors, you only need to send the sensor with the 30 cm section of cable and one-half of the connector. We have mating connectors at our factory that can be used for calibrating the sensor.
-
Please write the RMA number on the outside of the shipping container.
-
Return the item with freight pre-paid and fully insured to our factory address shown below. We are not responsible for any costs associated with the transportation of products across international borders.
Apogee Instruments, Inc.
721 West 1800 North Logan, UT
84321, USA -
Upon receipt, Apogee Instruments will determine the cause of failure. If the product is found to be defective in terms of operation to the published specifications due to a failure of product materials or craftsmanship, Apogee Instruments will repair or replace the items free of charge. If it is determined that your product is not covered under warranty, you will be informed and given an estimated repair/replacement cost.
PRODUCTS BEYOND THE WARRANTY PERIOD
For issues with sensors beyond the warranty period, please contact Apogee at
techsupport@apogeeinstruments.com
to discuss repair or replacement options.
OTHER TERMS
The available remedy of defects under this warranty is for the repair or
replacement of the original product, and Apogee Instruments is not responsible
for any direct, indirect, incidental, or consequential damages, including but
not limited to loss of income, loss of revenue, loss of profit, loss of data,
loss of wages, loss of time, loss of sales, accruement of debts or expenses,
injury to personal property, or injury to any person or any other type of
damage or loss.
This limited warranty and any disputes arising out of or in connection with
this limited warranty (“Disputes”) shall be governed by the laws of the State
of Utah, USA, excluding conflicts of law principles and excluding the
Convention for the International Sale of Goods. The courts located in the
State of Utah, USA, shall have exclusive jurisdiction over any Disputes.
This limited warranty gives you specific legal rights, and you may also have
other rights, which vary from state to state and jurisdiction to jurisdiction,
and which shall not be affected by this limited warranty. This warranty
extends only to you and cannot be transferred or assigned. If any provision of
this limited warranty is unlawful, void or unenforceable, that provision shall
be deemed severable and shall not affect any remaining provisions. In case of
any inconsistency between the English and other versions of this limited
warranty, the English version shall prevail.
This warranty cannot be changed, assumed, or amended by any other person or
agreement
APOGEE INSTRUMENTS, INC. | 721 WEST 1800 NORTH, LOGAN, UTAH 84321, USA
TEL: 435-792-4700 | FAX:
435-787-8268 | WEB:
APOGEEINSTRUMENTS.COM
Copyright © 2021 Apogee Instruments, Inc.
References
- 📧calibration@apogeeinstruments.com
- 📧techsupport@apogeeinstruments.com
- AC-100: Communication Cable - Apogee Instruments, Inc.
- How to Make a Weatherproof Cable Splice
- Recalibration and Repair | Apogee Instruments
- Underwater PAR Measurements | Apogee Instruments
- Unit Conversions | Apogee Instruments
- Clear Sky Calculator | Apogee Instruments Inc.
- Clear Sky Calculator | Apogee Instruments Inc.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>