GlucoRx Nexus Glucose Test Strips Instructions

June 6, 2024
GlucoRx

GlucoRx logo

312-0000000-XXX
Nexus Test Strips
Glucose Test Strips

Warnings

  • For in vitro diagnostic use (for use outside of the body only).
  • For single use only.
  • Healthcare professionals and other users testing multiple patients with this system should handle everything that has come into contact with human blood carefully to prevent transmitting infectious diseases, including sanitized objects.
  • Please read this sheet and your Blood Glucose Monitoring System Owner’s Manual before you use these test strips. Use only GlucoRx Nexus Test Strips with GlucoRx Nexus Blood Glucose Monitoring Systems to obtain accurate results, and be covered by the manufacturer’s warranty.
  • Results may be inaccurate when testing on patients with abnormally low blood pressure or those who are in shock.
  • For patients with impaired peripheral circulation, collection of capillary blood from the approved sample sites is not advised as the results might not be a true reflection of the physiological blood glucose level. It may apply under the following circumstances: severe dehydration as a result of diabetic ketoacidosis or due to stress hyperglycemia, hyperosmolar non-ketotic coma, shock, decompensated heart failure NYHA Class IV, or peripheral arterial occlusive disease.
  • Keep test strips and lancets away from small children. If swallowed, seek medical advice immediately.

Intended Use

GlucoRx Nexus test strips, when used together with GlucoRx Nexus Blood Glucose Monitoring Systems, allow your blood glucose levels to be measured by yourself at home or by healthcare professionals.
They use fresh whole blood samples from the finger, palm, upper arm, or forearm. This system is not intended for use in the diagnosis or screening of diabetes mellitus.
Professionals may use these test strips to test capillary and venous whole blood; home use is limited to capillary whole blood testing.

About Alternative Site Testing (AST)

IMPORTANT: There are limitations to performing AST. Please read your Nexus Blood Glucose Monitoring System Owner’s Manual and consult your doctor before you perform AST.
Alternative site testing (AST) is when individuals check their blood glucose levels using areas of the body other than the fingertip. The Nexus Test Strips allow AST to be performed on sites other than the fingertip. We strongly recommend that you perform AST ONLY at the following times:

  • During a pre-meal or fasting state (more than 2 hours since the last meal).
  • Two hours or more after taking insulin.
  • Two hours or more after exercise.

Do NOT use AST if:

  • You think your blood glucose is low (you may not notice if you are hypoglycaemic).
  • Your AST results are inconsistent with the way you feel.
  • You are testing for hyperglycemia.
  • Your routine blood glucose results often fluctuate.

To obtain a blood sample from the alternative sites, please rub the puncture site for approximately 20 seconds before following the procedures of “Testing Your Blood Glucose”.

Limitations

  • Xylose: Do not test blood glucose during or soon after a xylose absorption test. Xylose in the blood can give falsely elevated results.

  • Haematocrit: The Haematocrit level is limited to between 30% and 60% for TD-4277 Nexus and Nexus Blue, TD-4287 Nexus Mini, TD-4141 Nexus Mini Ultra, TD-4280 Nexus Voice, and TD-4283 GlucoRx GO; while the Haematocrit level is limited to between 25% and 70% for TD-4277 Nexus Pearl. Please ask your healthcare professional if you do not know your Haematocrit level.

  • Neonatal Use: These test strips must not be used for testing on newborns.
    These test strips are used for testing fresh capillary and venous whole blood.

  • Metabolites: Dopamine, L-Dopa, methyldopa, uric acid, ascorbic acid, and acetaminophen at normal blood concentration do not significantly affect blood glucose test results. The following table summarized the mentioned substance concentrations in excess of limitation with interference.

Substance| Limiting Concentration(mg/dL)| Therapeutic/Physiologic Concentration Range (or Upper Limit ) (mg/dL)
---|---|---
Dopamine| > 1.25| 0.03
L–Dopa| > 0.7| 0.02 – 0.28
Methyldopa| > 1.875| 0.1 – 0.5
Uric Acid| > 10| 2 – 8
Ascorbic Acid| > 5| 2
Acetaminophen| > 6.25| 0.45 – 3

  • There is no significant interference in the presence of galactose, maltose, or fructose observed in blood glucose tests.

  • Lipemic Effects: Blood triglycerides up to 33.9 mmol/L (3000 mg/dL) do not affect the results significantly, but may affect results at higher levels.

  • Altitude Effects:
    Altitudes up to 10,742 feet (3,275m) do not affect test results.

  • Use only heparin for anticoagulation of fresh capillary or venous whole blood.
    The following compounds when determined to be in excess of their limitation and tested with the GlucoRx Nexus Blood Glucose Meter may produce elevated glucose results:
    Summary of substances and concentrations in excess of limitation with interference
    **** Substance| Limiting Concentration(mg/dL)|

Therapeutic/Physiologic Concentration Range (or Upper Limit ) (mg/dL)

---|---|---
Acetaminophen (Paracetamol)| > 6.25| 0.45 – 3
Ascorbic Acid| > 5| 2
Pralidoxime Iodide| > 5| ~ 10 (IV Dose 500 mg)
Uric Acid| > 10| 2 – 8

Storage and Handling

IMPORTANT: Do not use the test strips if they have expired.

  • Store the test strips in a cool, dry place between 2°C and 30°C (35.6°F and 86°F) and 10% to 85%. relative humidity.
  • Keep the test strips away from direct sunlight. Do not store the test strips in high humidity.
  • Store the test strips in their original vial ONLY. Do not transfer them to a new vial or any other containers.
  • Do not touch the test strips with wet hands.
  • Close the vial immediately after taking out a strip. Use each test strip immediately after taking it out of the vial.
  • Keep the vial closed at all times.
  • Do not bend, cut, or alter the test strips.

Strip Appearance

| 1. Absorbent Hole
Apply a drop of blood here. The blood will automatically be absorbed.
2. Confirmation Window
This shows if enough blood has been drawn into the absorbent hole of the strip.
3. Test Strip Handle
Hold this part to insert the test strip into the slot.
4. Contact Bars
Insert this end of the test strip into the meter. Push it in firmly until it will go no further.
The front side of the test strip should face upwards when inserting the test strip into the meter.
Test results may be wrong if the contact bar is not fully inserted into the test slot.
---|---

Testing Your Blood Glucose **

**

PLEASE WASH AND DRY YOUR HANDS BEFORE PERFORMING ANY TESTING.

| STEP 1
Insert the test strip fully into the slot of the meter until it will go no further.
When the strip is fully inserted, the meter will do several self-checks.
STEP 2
Collect a blood sample of about 0.5 μL with the test strip. A sufficient quantity of blood is required for the test to provide accurate results.
Touch the blood drop with the absorbent hole of the test strip, and wait until the confirmation window is fully covered. Do NOT apply a smeared blood sample. The meter will start counting down.
STEP 3
After a few seconds, the meter will display your blood glucose level. The last reading will be automatically saved in the meter. Turn it off by removing the test strip and throwing away the used test strip.
---|---

Please refer to your Owner’s Manual for more information.

The used lancet and test strip are potentially biohazardous. Please dispose of them carefully according to your local regulations.

Reading Your Result

Your blood glucose readings deliver plasma equivalent results and are displayed in millimoles of glucose per liter of blood (mmol/L). The measurement range of your meter is 1.1-33.3 mmol/L (20-600 mg/dL).

Reference values
Time of day Normal blood glucose range for people with diabetes
Fasting and before a meal 4-7 mmol/L (72-126 mg/dL)
2 hours after meals Less than 9 mmol/L (162 mg/dL)

Source: Diabetes UK. Blood glucose target. Balance: No.234, 2010 April. P.69
Please consult your doctor to determine a target range that works best for you.

Questionable or inconsistent results
If your test results are unusual or inconsistent with how you are feeling:

  • Make sure the confirmation window of the test strip is completely filled with blood.
  • Check the expiry date of the test strips.
  • Check the performance of your meter and test strip with the control solution.

PLEASE NOTE: Unusually high or low blood glucose levels may be symptoms of a serious medical condition.
If most of your results are unusually high or low, please contact your healthcare professional.

Quality Control Testing

Taidoc control solutions contain a known amount of glucose that can react with test strips. You can check the performance of the meter, test strip, and your technique by comparing the control solution results with the range printed on the label of your test strip vial. Checking regularly can ensure your test results are accurate. Please refer to your Owner’s Manual for complete testing instructions.
IMPORTANT: The reference range of the control solutions may vary with each new vial of test strips. Make sure you check the range on the label of your current vial.

Chemical Components of Sensor

Glucose dehydrogenase (E. coli) 8%
Electron shuttle 55%
Enzyme protector 8%
Non-reactive ingredients 29%

Additional Information for Healthcare Professionals

Always wear gloves and follow your facility’s biohazard control policy and procedures when performing
tests involving patient blood samples. Use fresh whole blood samples only. Professionals may use these
test strips to test capillary and venous whole blood.
Sample Size: 0.5 μL
Reaction Time: 5 seconds
System Measurement Range: 1.1 mmol/L to 33.3 mmol/L
Haematocrit Range: 30 % to 60% for TD-4277 Nexus and Nexus Blue, TD-4287 Nexus Mini, TD-4141 Nexus Mini Ultra, TD-4280 Nexus Voice and TD-4283 GlucoRx GO; 25% to 70% for TD-4277 Nexus Pearl

Accuracy

Diabetes experts have suggested that glucose meters should be within ±0.83 mmol/L of the reference method when the glucose concentration is lower than 5.5 mmol/L , and be within ±15% of the reference method when the glucose concentration is 5.5 mmol/L or higher.
The three tables below display how often the GlucoRx Nexus meters (TD-4277 Nexus and Nexus Blue, TD-4287 Nexus Mini, TD-4141 Nexus Mini Ultra, TD-4280 Nexus Voice, and TD-4283 GlucoRx GO and TD-4277 Nexus Pearl) achieve these targets respectively.
The charts are based on a study carried out on 160 patients (each patient was tested six times which had 960 test results) to see how well the GlucoRx Nexus meters (TD-4277 Nexus and Nexus Blue, TD-4287 Nexus Mini, TD-4141 Nexus Mini Ultra, TD-4280 Nexus Voice and TD-4283 GlucoRx GO and TD-4277 Nexus Pearl) performed compared to YSI-2300 reference method results.

TD-4277 Nexus and Nexus Blue
Table 1 Accuracy results for glucose concentration < 5.55 mmol/L

Within ±0.28 mmol/L Within ±0.55 mmol/L Within ±0.83 mmol/L *
55.1% (172/312) 88.8% (277/312) 99.4% (310/312)

Table 2 Accuracy results for glucose concentration ≥ 5.55 mmol/L

Within ±5 % Within ±10 % Within ±15 %*
51.4% (333/648) 85.2% (552/648) 97.5% (632/648)

Table 3 Accuracy results for glucose concentrations between 1.58 mmol/L to 30.22 mmol/L

Within ±0.83 mmol/L or ±15 %
98.1 (942/960)

TD-4287 Nexus Mini
Table 1 Accuracy results for glucose concentration < 5.55 mmol/L

Within ±0.28 mmol/L Within ±0.55 mmol/L Within ±0.83 mmol/L *
66.7% (208/312) 93.6% (292/312) 100% (312/312)

Table 2 Accuracy results for glucose concentration ≥ 5.55 mmol/L

Within ±5 % Within ±10 % Within ±15 %*
56.3% (365/648) 89.5% (580/648) 98.3% (637/648)

Table 3 Accuracy results for glucose concentrations between 1.35 mmol/L to 30.22 mmol/L

Within ±0.83 mmol/L or ±15 %
98.9% (949/960)

TD-4280 Nexus Voice

Table 1 Accuracy results for glucose concentration < 5.55 mmol/L

Within ±0.28 mmol/L Within ±0.55 mmol/L Within ±0.83 mmol/L *
59.3% (185/312) 93.9% (293/312) 100% (312/312)

Table 2 Accuracy results for glucose concentration ≥ 5.55 mmol/L

Within ±5 % Within ±10 % Within ±15 %*
58.0% (376/648) 87.7% (568/648) 97.5% (632/648)

Table 3 Accuracy results for glucose concentrations between 1.80 mmol/L to 30.08 mmol/L

Within ±0.83 mmol/L or ±15 %
98.3 (944/960)

TD-4283 GlucoRx GO
Table 1 Accuracy results for glucose concentration < 5.55 mmol/L

Within ±0.28 mmol/L Within ±0.55 mmol/L Within ± 0.83 mmol/L *
59.3% (185/312) 93.9% (293/312) 100% (312/312)

Table 2 Accuracy results for glucose concentration ≥ 5.55 mmol/L

Within ±5 % Within ±10 % Within ±15 %*
58.0% (376/648) 87.7% (568/648) 97.5% (632/648)

Table 3 Accuracy results for glucose concentrations between 1.80 mmol/L to 30.08 mmol/L

Within ±0.83 mmol/L or ±15 %
98.3% (944/960)

TD-4141 Nexus Mini Ultra
Table 1 Accuracy results for glucose concentration < 5.55 mmol/L

Within ±0.28 mmol/L Within ±0.55 mmol/L Within ± 0.83 mmol/L *
66.7% (256/384) 93.2% (358/384) 100% (384/384)

Table 2 Accuracy results for glucose concentration ≥ 5.55 mmol/L

Within ±5 % Within ±10 % Within ±15 %*
55.9% (322/576) 89.2% (514/576) 98.1% (565/576)

Table 3 Accuracy results for glucose concentrations between 1.42 mmol/L to 30.08 mmol/L

Within ±0.83 mmol/L or ±15 %
98.9% (949/960)

TD-4277 Nexus Pearl
Table 1 Accuracy results for glucose concentration < 5.55 mmol/L

Within ±0.28 mmol/L Within ±0.55 mmol/L Within ± 0.83 mmol/L *
50.6% (158/312) 84.0% (262/312) 98.4% (307/312)

Table 2 Accuracy results for glucose concentration ≥ 5.55 mmol/L

Within ±5 % Within ±10 % Within ±15 %*
54.2% (351/648) 88.6% (574/648) 98.8% (640/648)

Table 3 Accuracy results for glucose concentrations between 1.55 mmol/L to 30.56 mmol/L

Within ±0.83 mmol/L or ±15%
98.6% (947/960)

*Acceptance criteria in EN ISO 15197: 2015, 95% of all differences in glucose values (i.e., YSI-2300 reference values minus GlucoRx Nexus’s glucose values) should be within ±0.83 mmol/L (15 mg/dL) for glucose concentration < 5.5 mmol/L (100 mg/dL), and within ±15% for glucose concentration ≥ 5.5mmol/L
(100 mg/dL).
Note: When Nexus Test Strips results are compared to the reference values, difference values below 5.5 mmol/L are expressed in mmol/L, while those above 5.5 mmol/L are in percent.

Consensus Error Grid (CEG)

The analysis of the Consensus Error Grid showed 100% of test results were within zone A and zone B compared to the Yellow Spring 2300 reference method. (100% of test results were within zone A, while 0% was within zone B, with no results in zones C, D, or E).
Figure 1~4 Error grid analysis of the Consensus Error Grid (mmol/L)

| |
---|---|---

Table 1 Definitions of the Error Grid Zones (CEG)

Risk level (CEG zone) Percentage Risk to the diabetic patient
100% A No effect on clinical action.
0% B Altered clinical action – little or no effect on clinical outcome.
0% C Altered clinical action – likely to affect clinical outcome.
0% D Altered clinical action – could have significant medical risk.
0% E Altered clinical action – could have dangerous consequences.
User performance

TD-4277 Nexus and Nexus Blue
A study evaluating glucose values from fingertip capillary blood samples obtained by 154 laypersons showed the following results:
100% within ±0.83 mmol/L of the medical laboratory values at glucose concentrations below 5.55 mmol/L, and 97.2% within ±15% of the medical laboratory values at glucose concentrations at or above 5.55 mmol/L.

TD-4287 Nexus Mini
A study evaluating glucose values from fingertip capillary blood samples obtained by 154 laypersons showed the following results:
100% within ±0.83 mmol/L of the medical laboratory values at glucose concentrations below 5.55 mmol/L, and 96.3% within ±15% of the medical laboratory values at glucose concentrations at or above 5.55 mmol/L.

TD-4280 Nexus Voice
A study evaluating glucose values from fingertip capillary blood samples obtained by 155 laypersons showed the following results:
100% within ±0.83 mmol/L of the medical laboratory values at glucose concentrations below 5.55 mmol/L, and 97.2% within ±15% of the medical laboratory values at glucose concentrations at or above 5.55 mmol/L.

TD-4283 GlucoRx GO
A study evaluating glucose values from fingertip capillary blood samples obtained by 155 laypersons showed the following results:
100% within ±0.83 mmol/L of the medical laboratory values at glucose concentrations below 5.55 mmol/L, and 98.1% within ±15% of the medical laboratory values at glucose concentrations at or above 5.55 mmol/L.

TD-4141 Nexus Mini Ultra
A study evaluating glucose values from fingertip capillary blood samples obtained by 154 laypersons showed the following results:
100% within ±0.83 mmol/L of the medical laboratory values at glucose concentrations below 5.55 mmol/L, and 96.3% within ±15% of the medical laboratory values at glucose concentrations at or above 5.55 mmol/L.

TD-4277 Nexus Pearl
A study evaluating glucose values from fingertip capillary blood samples obtained by 160 laypersons showed the following results:
100% within ±0.83 mmol/L of the medical laboratory values at glucose concentrations below 5.55 mmol/L, and 97.2% within ±15% of the medical laboratory values at glucose concentrations at or above 5.55 mmol/L.

Precision

The CV (%) is less than 5% both in intermediate precision and repeatability precision.

Symbol Information

Symbol

|

Referent

---|---
| In vitro diagnostic medical device
| Consult instructions for use
| Temperature limit
| Use-by date
| Batch code
| Caution
| Manufacturer
| Authorized Representative in the European Community
| Do not re-use

0123

| CE mark
| Humidity limitation
| Catalog number

Distributed by GlucoRx Ltd.
Unit 1C Henley Business Park,
Pirbright Road, Guildford, Surrey,
GU3 2DX, UK
Website: www.glucorx.co.uk
Customer Careline: 01483 755133

TaiDoc Technology Corporation
B1-7F, No. 127, Wugong 2nd Rd., Wugu Dist.,
24888 New Taipei City, Taiwan
www.taidoc.com
MedNet EC-REP GmbH
Borkstraβe 10, 48163 Mϋnster, Germany

Use Only With GlucoRx Nexus Blood Glucose Monitoring Systems.

For self-testing.
ver 11.0 2020/05

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals