Ihelol XT-S600 Smart Neck Massager User Manual
- June 6, 2024
- Ihelol
Table of Contents
Ihelol XT-S600 Smart Neck Massager
Precautions
Danger
This product Is prevented from using with any of the following medical electronic devices and venues:
-
Medical electronic devices such as ECG instruments;
-
Implantable medical electronic devices such as heart pacemakers;
-
Life-sustaining medical electronic devices such as ECMO;
Any of the said medical electronic devices when used with this product may cause malfunctioning or even maybe life-threatening, please know well and observe this rule; -
Gas stations, chemical plants, dusty workshops, gas and other flammable and explosive venues.
-
This product Is prohibited from use when charging.
Warning
- Never use this product with a high-frequency surgical instrument, otherwise, the electrode pad may be burnt, or this product may be damaged.
- Never use this product In the v1c1nity of short-wave or microwave therapy apparatus, otherwise, the function output may be unstable, and the normal funclion of therapy apparatus may be also compromised.
- Avoid making the electrode pad of this product contact with wounds and scar during a massage, otherwise, 1t may cause physical discomfort.
- Never place the electrode pad near the chest heart during a massage, otherwise, 1t would increase the hazard of heart fibrillation.
- It Is recommended to use this product for only a user. Whenever it Is used by more than one user, you must clean the electrode pad on this product before use, keep it clean and hygienic.
- To use this product, fit the electrode pad against the skin and make it evenly contact, otherwise, It may cause a slight.
- The electrode pad has a heating function when the same part Is heated for a long duration (over 15 m1ns), otherwise, It may cause low-temperature skin scald.
- Immediately stop 1t and seek medical attention whenever you endure physical discomfort
- This product Is provided with a lithium battery. Never drop, disassemble, modify, immerse in water or throw 1t into the fire, otherwise, It may cause damage or battery explosion.
- In any of the following circumstances, seek medical consultation before use, otherwise, 1t may cause an incident or physical discomfort.
- Any patient under medical treatment and suffering physical discomfort.
- Any patient diagnosed with heart disease and abnormalities of brain nerves.
- Any patient whose body temperature exceeds 37 .8’C (fever).
- Any patient diagnosed with infectious disease.
- Any patent with impaired skin perception or abnormal skin.
- Any patient who Is taking med1c1ne.
- Any patient with an abnormality of blood pressure.
- Any patient who cannot freely express his/her will.
- Any patient with perceptual disorder due to high peripheral circulation disorders such as diabetes.
- Any patient whose skin Is allergic to SS electrode pad.
- Never use this product, 1f
- Any patient diagnosed with malignant tumors
- Any patient diagnosed with bleeding tendency
- Pregnant or lying-in woman
- Any patient with acute suppuratIve 1nflammat1on
- Any patient implanted with cardiac pacemaker
- Any patient with metal implants
- Around the human heart
- Never use this product In any area around the heart, above the neck, on the head, In the mouth, in private parts or in any area with a skin disorder, and do not massage the soles of feet and palms at the same time. Moreover, never press on internal organs in use, otherwise, it may cause an incident or physical discomfort.
- Never use this product with other medical devices or use more than 2 units of this product at the same time, otherwise, It may cause an abnormality of function or physical discomfort.
- Never use this product for any purpose other than a massage as specified in this User Manual, otherwise, It may cause a 1nc1dent or malfunctioning.
- Never use any electrode pad with notch or damaged part, otherwise 1t may cause an incident and functioning.
- Never disassemble, repair or modify this product without prior approval, otherwise It may cause malfunctioning or electnc shock.
Cautions
-
Immediately stop,t whenever you feel physical discomfort.
-
Seek medical consultation and follow-up medical advice if your body or skin feels unusual discomfort.
-
If you intend to move the electrode pad or move,t to another person during massage, never use it with any medical electronic device inside or outside the body, otherwise ,t may malfunctioning of electronic components (e.g.:
e- watch wrong time). -
Turn off the power and unplug 11 it cannot run as normal, otherwise, II may have physical discomfort and, an incident.
-
Do not plug or unplug a USB cable with wet hands when charging, otherwise, it may cause an electric shock or incident.
-
Do not render it to any child or person who is unconscious of expression, otherwise, It may cause an incident or physical discomfort.
-
If you feel physical discomfort because of product manufacturing, immediately stop it and unplug it before moving it, otherwise,t may cause an intense electric shock and physical discomfort.
-
Do not use this product,n high humidity areas such as a bathroom, or use,t while taking a shower, otherwise,t may cause an intense electric shock and product damage.
-
Do not use,t when asleep, otherwise, may cause product malfunctioning and the electrode pad may lead to physical discomfort when moved to other parts.
-
Do not use this product when dnv, ng, otherwise,t may cause physical discomfort and incident due to intense skin irritation.
-
Do not affix the electrode on the wounded area,
-
otherwise,t may cause skin inflammation.
-
Do not make the metal part of the belt, necklace, and other, terns contact the electrode when massaging, otherwise,t may be subject to intense electric shock.
-
Do not apply the same area for a long duration (over 30 mins), otherwise, it may cause muscle fatigue on the area where massaged and physical discomfort.
-
Wear this product property on left and right side of the neck in a symmetrical way, otherwise,t may cause an incident or physical discomfort.
-
When unplugging a USB cable, do not attempt to pull and drag the wire, but instead gnp and pull out the wire plug, otherwise ,t may cause electric shock or product damage.
-
The electrode pad will heat up, any user whos not sent, veto heat, or is susceptible to heat allergies should seek medical consultation before use.
-
The product surface may heat up 11 operated for a long duration, but this ,s a normal phenomenon.
-
Remove the battery of remote controller 11 left unused (over a penod of 3 months).
-
If the battery of the remote controller,s about to run out, please promptly replace new battery of the same model: CR2032.
-
Never load the battery of the remote controller,n reverse electrodes when replacing a new battery.
-
After the battery of the remote controller is out of service, dispose of it by following local regulations.
Recommendation for Use
- Make sure that the buttons can function well before use.
- If the battery gets lost during InitIal use or if the product Is left unused for a long period, charge the product by following the 1nstruct1ons as contained,n the User Manual, until you can make sure that ,t can run as normal.
- Dispose of,t according to local regulations on environmental protection, otherwise, It may cause environmental pollution.
- This product Is intended for user at 16 or above.
- Never use or store it,n:
- The area exposed to the sun
- Hot and humid area
- Flooded area
- Dusty area
- Near fireworks
- Area susceptible to vibration and shock 7 Area under electromagnetic interference 8 Petrol or gas station
- LPG station
- Flammable and explosive chemical plant
Technical Data
Electrical schematic diagram
Product Components
Remote Controller
Procedures
- Moisten the neck skin with a wet towel before use, wear the product on neck with the electrodes fit to the neck skin, set it In proper place, turn on the product and adJust 1t to an appropnate level and mode of massage.
- Press and hold for two seconds, turn on the product (until you can hear the buzzer “beep” <8? and the indicator becomes hi), go to the default mode and set at the default level (0). Press and hold for two seconds when 1I is turned on, the buzzer will “beep” twice and the product Is turned off. (Press and hold the off button, do not attempt to turn 1t on until the battery indicator stops flashing).
- “+” button on the main unit Is used to turn up the level of massage, every press Is followed by one buzzer “beep” , which means one level up. There are 15 levels numbered from Oto 15. At a level of 15, press “+” button, then the buzzer will “beep” twice, but the level stays the same at 15.
- •-• button on the main unit is used to tum down the level of massage, every press Is followed by one buzzer “beep”, which means one level down. There are 15 levels numbered from 15 to 1. At a level of 1, press •-• button, then the buzzer will “beep” twice, but the level stays the same at 1.
- “M” button on the main unit is the mode button, press it to switch mode. There are 3 optional modes, every press Is followed by one buzzer “beep”, which means one mode up. Three modes can be switched cyclically (Note: M1, M2, M3). Every mode switch Is followed by a return to the default 0.
- “M” button on the remote controller Is the mode button, press II to switch mode. There are 3 optional modes, every press Is followed by one buzzer “beep” , which means one mode up. Three modes can be switched cyclically (Note: M1, M2, M3). Every mode switch Is followed by a return to the default 0.
- “+” button on the remote controller is the level up button, every press is followed by one buzzer “beep”, which means one level up. There are 15 levels numbered from Oto 15. At a level of 15, press “+” button, then the buzzer will “beep” twice, but the level stays the same at 15.
- •-· button on the remote controller Is the level up button, every press Is followed by one buzzer “beep”, which means one level up. There are 15 levels numbered from 15 to 1. At a level of 1, press “+” button, then the buzzer will “beep” twice, but the level stays the same at 1.
- “the Ill ” buzzer on the will remote “beep” controller once, Is then the heating main on the unit button, starts press this heating, button, two electrode pas will heat up at a temperature of 37-43’C.
- 0. “.144-” on the remote controller Is the healing off button, press this button, the buzzer will “beep” once, then the main unit stops heating.
- The timing run set for the main unit Is 15 minutes, which means that the unit will automatically stop running 15 minutes after turning on.
- When the unit is turned on, the product will automatically shut down within 20 seconds 11 you don’ t wear II properly.
How to Charge Battery
- When the battery level Is low, the unit will shut down 7 seconds after “beeping” , you may plug one end of the M1cro-USB cable to the unit, with the other end plugged to the power adapter (power output: 5V1 A/2A) for charging. When charging, the red indicator, green 1nd1cator, and blue 1nd1cator will flash alternately; white indicator becomes lit when fully charged.
- To replace the battery, insert a key of suitable width into the battery knob slot, turn II clockwise to the counter posItIon, open the cover and replace battery, make sure that the cathode Is loaded upward. (Battery mode: CR2032)
Tips
- The product will automatically shut down after running for 15 minutes, so 11 Is recommended to run for a duration shorter than 30 minutes, use 1t on demand every week.
- This product should be stored In a well-ventilated and dry area after use.
- Never attempt to use 11 when placed under any other heavy obJect.
- Never attempt to pull the rubber plug at USB jack, failing which 1t may break and fall off.
Daily maintenance
- To clean and maintain the product, tum off the main unit, clean electrode pads with a wet tissue. When 1t is very dirty, wipe and d1s1nfect electrodes
- with a slight amount of 75% medical alcohol. For routine care, wipe clean by applying neutral detergent with a soft cloth,
- never use any corrosive detergent or soak or rinse it In the water. Prevent water or other liquids from getting into the main unit when cleaning, for fear of short electric shock In use.
Troubleshooting
- Refer to the following tips and alternatives if something goes wrong with this product.
- Unplug it 1f the product Is faulty or becomes unserviceable.
- Never attempt to disassemble or repair the product without prior approval, failing which 1t may cause product damage or invalidate the warranty term. In this case, we will be excused from any consequence.
Fault Descr ptIon| Alterna t ive Solution
---|---
The product cannot start
| • The internal rechargeable battery Is low.
• The remote controller battery runs out of power.
• Make sure if the main unit On/Off button is pressed down.
The prowct C3IV101 be recllarged or the chargr,g oocator is oot cisplayed|
Check 11 USB cable is properly connected to the main unit and cellphone power
adapter.
Fau I Description Alternative Solution
• Check if the electrode pads directly contact the skin? The product can Adjust the position until II IS closely fit to the skin. display as normal, • Check 11 there is any hair or clothes, accessories but user Is between electrode pads and skin? Move or remove unconscious of objects between them.
massage • Check 11 the level of massage is set to 0? Increase the
level of massage.
• Check 1f the electrode pads directly contact the neck skin? AdJust the position until 1t is closely fit to the skin.
• Check 1f the massage duration Is too long? It Is recommended lo massage 15 minutes a cycle, two cycles a day.
• Check if the level of massage Is too high? Level down.
With sting feeling • Check 1f the neck skin Is allergic to electrode pads? Anyone With severe skin allergy should receive
desensitization treatment first.
• Check n the climate is freezing and skin is dry? Moisten the skin with a wet towel before use.
• Check if the skin surface Is wounded? Never attempt to place any electrode on the wounded area for massage.
• Check if the electrode pads directly contact the neck skin? AdJust the pos1t1on until 1t IS closely Ill to the skin and turn to an appropriate level of massage.
Slight irritation • Check 1f the electrode Is stained or oily? Clean the electrode surface.
• Check 11 the battery level Is low? If so, recharge the battery.
• Check if the level of massage Is too low? Level up.
Instructions for Recycling of Products and Parts
Suggestive Instructions for ecycling
For the purpose of environmental protection, you as a consumer Is responsible for separating this product or any of parts (when scrapped} from domestic refuses, and transfer ,t to a competent recycling site who will resort, disassemble, recycle and renew this product or any of parts (when scrapped} according to applicable national regulations. Description and Content of Hazardous Substances in Product.
| | Haza, Im s Sullstances
---|---|---
Part Name| Pll| He| Crl| Cr(VI| P9R| PRnF-
Metal parts| 0| 0| 0| 0| 0| 0
PCB| X| 0| 0| 0| 0| 0
Plastic parts| O| 0| 0| 0| 0| 0
€)
EFUPMark
| “EFUP” of this product is up to 10 years, we define this penod according to the characteristics of product design. “EFUP” is only valid when the product ,s used under normal conditions as described in the User Manual.
Note:
- 1. Other parts and their material that comprise this product, but are not listed above are considered non-hazardous substances.
2. All parts and their homogeneous materials which contain hazardous substances as listed above are consistent with the strict requirement of China RoHS on restrictive use of hazardous substances, you can rest assured to useIt.
Precautions for Disposal
&
Danger
| The battery loaded In this product Is a spec,al rechargeable battery. Never attempt to use It for other products, and do not charge the rechargeable battery that has been removed.
• Do not throw battery 1n fire for fear of overheating
• Do not pierce w,th a nail, impact by gravity, disassemble or modify
structure
• Do not connect the cathode and anode with a metal piece
• Do not store 1t wrth necklace or hairpin
• Do not charge, use and place rt by furnace, or exposed under the sun, failing to which It may cause heating, fire and break off.
• Do not dismantle the massager and replace the ,nside rechargeable hth,um 10n battery by a non- rechargeable battery.
Any of the above circumstances may cause heating, fire, cracking or explosion.
---|---
&
Warn i ng
| Keep the removed rechargeable battery out of children’ s reach for fear of aoodental swallowing.
• Seek 1mmed1ate medical attentionIn case of accidental swallowing
• If the battery fluid leaks, never touch the fluid with bare hands and dispose rt by the following way.
• If the battery fluid accidentally splashes into your eyes, you may lose sight. Do not attempt to rub your eyes with hands, nnse w,th clean water and seek medical consultation. If your body or clothesIs stained with battery fluid, 11 may cause skin imtat,on or wound, flush with clean water.� you feel unwell, seek ,mmed,ate medical attention.
• Disassemble the product only when disposed, failing to which II may cause fire, electric shock or wound; � you intend to repair the product (other than battery replacement), consult the store from which you bought this product.
Give support for the recycling of rechargeable batteries (lithium-ion batteries}. Battery model: 103048-1500mAh Battery polarity: Inside the battery compartment, red line indicates the anode, and black line indicates the cathode.
The lithium-ion battery load in this product,s a valuable resource that can be recycled and renewable. Remove the battery when It Is going to scrap, send It to a renewable and disposal site for rechargeable batteries.
Product Warranty Card
Warranty Term and Condition:
- This warranty card is only valid within the territory of Mainland China (other than Taiwan, Macau, Hong Kong).
- 2.0seemoon electrical appliance bought by consumer for family needs will be guaranteed for one year from the date of purchase provided that it Is used as normal.
- If you request free repair service for a malfunctioning product under normal use conditions within the warranty term, connect to the purchase channel on presentation of this warranty card and payment slip.
- In any of the following circumstances, this product will be provided paid repair instead of free repair:
- If It is damaged due to improper use, maintenance, or care;
- lf no “Three Guarantees” certificate and valid payment slip Is presented;
- lf the model as indicated on the “Three Guarantees” certificate Is different from the product model or there Is some content altered;
- lf this product Is damaged due to the Force Majeure event.
- Keep the warranty card in good order and no renewal Is made upon loss.
- Any user as holder of a valid warranty card will be entitled to other nights to the extent permitted by applicable national provisions during the period of repair request, except for the product repair as contained In this warranty card.
User Warranty Card
Repair Service Record
FCC Statement
This device complies with part 15 of the FCC rules. Operation is subject to
the following two conditions: (1) this device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation. Changes or
modifications not expressly approved by the party responsible for compliance
could void the user’s authority to operate the equipment.
NOTE : This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates uses and
can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
The device has been evaluated to meet general RF exposure requirement. The device can be used in portable exposure conditions without restriction.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>