WatchGas PDM Family Multi Dock User Manual

June 6, 2024
WatchGas

WatchGas PDM Family Multi Dock

Warning

  •  Any unauthorized attempt to repair or modify the product, or any other cause of damage beyond the range of the intended use, including damage by fire, lightning, or another hazard, voids the liability of the manufacturer.
  • Do not use if the device appears to be damaged.
  • Use the device in a clean environment, which has no hazardous gases.
  • Do not use the expired gas cylinders.
  • Ensure to use the certified gas cylinders.
  • Ensure gas types and concentrations match with manuals and devices.
  • Severe vibration or shock to the device may cause a sudden reading change and cause the device to malfunction.
  • Do not allow water and liquids to reach this device.

READ THE OPERATION MANUAL BEFORE USE

  •  Before the operation, read the manual carefully and follow all the instructions.
  •  IR link only works when it is placed at the bay 1 on the docking station.
  •  The device only operates when the power LED turns green.
  •  Before upgrading firmware, connect the plug to avoid any data loss and malfunction by unexpected power off. And the file name is fixed as below.

Overview

The WatchGas PDM Docking Station keeps the monitor correctly configured and tested and fully functional without the actual presence of a certified service technician. Bump testing portable monitors is becoming more and more prevalent in the market. Ensuring sensors are responding to a known concentration of gas remains the best way to verify the monitor has not become impeded. WatchGas PDM docking station provides bump testing, event management and calibration from a simple-to-use multi-unit station and gives fleet or safety managers the ability to update configurations on large fleets. It also maintains the all- important event logs that demonstrate users are working within company requirements.

Gas usage-controlled Lowers the cost of bump testing
Manages calibration Manages calibration and stores records of the event
Flexible Can monitor a wide array of test parameters and detector

settings, providing excellent traceability
All data is stored in a USB stick| Simple to transfer the data
Rechargeable battery or power| Ideal for in-vehicle or multi-site use
Holds up to 4 PDM’s at once| Faster testing of larger fleets

PDM Dock Configuration EASY TO MAINTAIN

Changeable alarm levels| Ensures compliance with changing regulations or company procedure
---|---
WatchGas PDM Docking Station| Keeps monitors correctly configured without service or technician team
‘Bump test due’ indication| Alerts the user and team members that the monitor is not compliant
Compact and lightweight| Non-intrusive
---|---
Single button operation| Delivers simplicity for users
Reduces training time
Wide operating temperature range| Effective in a wide variety of harsh enviroments: 41 to + 104˚F ( 5 to + 40˚C)
---|---
IP Rating| IP66 and IP67

Note: Users can change settings, upgrade firmware and download log events through the Docking Station IR-Link**WatchGas-PDM-Family-Multi-Dock-
FIG-2

  •  Docking PC Programs areas above image.
  •  To interact with Docking Station, ensure to install the Docking Station software and have the IR Link cable** plugged into the computer. Place the top side of the IR Link face down on detector bay #1.
    •  IR-Link only operates on bay #1 and power should be on. ** IR-Link is available separately.WatchGas-PDM-Family-Multi-Dock-FIG-3

DOCKING INFO

Serial S/N: Factory Serial S/N
Version: Docking station firmware version
Auto Power-Off (Max 3600 seconds): Auto Power-Off
Maximum Bump Time: Default: 30 / Min: 20 / Max: 120 seconds
Location: Configure a location of the docking station installed.
Gas Expiration Date: Set up the expiration date of a gas cylinder

manually.
Gas Bottle Lot #:| Set up a bottle serial number manually.

ALLOW CHECK
Users can enable/disable the following options by checking each box. Calibration/Bump test/Hibernation/Configuration(Detector)/Firmware Upgrade Read: Retrieve the stored settings Write: Update the adjusted settings

option (SET BY DOCK)

Alarms O 2 H 2 S CO
LOW 1%vol 5 ppm 25 ppm
HIGH 23% vol 5 ppm 25 ppm

User ID configuration Calibration Interval (0~365 days): Set up an interval of the calibration alert Bump Station Interval (0~24 hours): Set up an interval of the self-test.

Log Configuration

EVENT LOGGER PDM (FOR THE LAST 30 ALARMS)

 Specification

Size 47 x 41 x 22cm (18.50 x 16.14 x 8.66 in.)

Weight

(without gas cylinder)

| 8.2kg
Operating Temperature| 5 to + 40˚C (41 to + 104˚F)
Warranty| Full 2 years
Battery Type| Rechargeable Lithium Ion
Battery Life| 1,000 Bump Tests
LEDS| 5 red / green / orange LEDs

( 1 for each unit, 1 for power )

Memory| USB 8GB Standard
Log Capacity| Approximately Up to 5 million tests (8GB, removable USB memory)
Tests Performed| Bump Test and Calibration
Information Stored| Bump/Cal Logs, Individual Monitor Event Logs, Firmware and Unit Configurations
**Gases Available*| CO, H₂S, O₂, NH3, SO2 en H2
Unit Compatibility| Works with WatchGas PDM
Calibration Gas| Compatible with 36cm cylinders
Charging Adaptor**| 12V, 3A

Limited Warranty

WatchGas warrants this product to be free of defects in workmanship and materials under normal use and service for two years from the date of purchase from the manufacturer or from the product’s authorized reseller.
The manufacturer is not liable (under this warranty) if its testing and examination disclose that the alleged defect in the product does not exist or was caused by the purchaser’s (or any third party’s) misuse, neglect, or improper installation, testing, or calibrations. Any unauthorized attempt to repair or modify the product, or any other cause of damage beyond the range of the intended use, including damage by fire, lightning, water damage, or another hazard, voids the liability of the manufacturer. In the event that a product should fail to perform up to manufacturer specifications during the applicable warranty period, please contact the product’s authorized reseller or WatchGas service center at +31 (0)85 01 87 709 to repair/return information.

  • WatchGas B.V. Sextantstraat 61
  • 2901 ZZ Capelle aan den IJssel
  • The Netherlands
  • [email protected] www.watchgas.eu
  • All rights reserved. No part of this publication may be reproduced, distributed, or transmitted in any form or by any means, including photocopying, recording, or other electronic or mechanical methods, without the prior written permission of the publisher, except in the case of brief quotations embodied in critical reviews and certain other noncommercial uses permitted by copyright law. For permission requests, contact WatchGas B.V.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

WatchGas User Manuals

Related Manuals