WLC A1-C-2 Precision Scale
PRECAUTIONS
Prior to installation, use or maintenance activities, carefully read this user
manual and follow the provided guidelines.
| Prior to the first use, carefully read this user manual. Use the weighing
device only as intended.
---|---
| Place weighed loads in the centre of the weighing pan.
| Load the weighing pan with loads of gross weight which does not exceed the
maximum capacity.
| Mind not to leave heavy loads on the weighing pan for longer periods
of time.
| Protect the indicator against considerable temperature variation, solar and
UV radiation, substances causing chemical reactions.
| The C32 scale must not be operated in hazardous areas endangered with
explosion of gases, and in dusty environments.
| In case of damage, immediately unplug the device from the mains.
| Scales to be decommissioned must be decommissioned in accordance with valid
legal regulations.
| Do not let battery discharge in case of prolonged storage of the
device in low temperature.
| A worn out battery can be replaced only by the manufacturer or by the
authorized service.
| Accumulators do not belong to regular household waste. The European
legislation requires discharged accumulators to be collected and disposed
separately from other communal waste with the aim of being recycled. Symbols
on batteries identify harmful compounds: Pb = lead, Cd =
cadmium, Hg = mercury. Dear user, you are obliged to dispose of the worn out
batteries as regulated.
| If the scale is to be operated in conditions that are difficult due to
electrostatics (e.g. printing house, packing centre, etc.), you must connect
it to the earth wire. To enable this, the device features functional earthing
terminal, marked with symbol.
INTENDED USE
WLC/A precision scales enable fast and accurate mass measurements under laboratory and industrial conditions. The devices are equipped with an internal battery which allows their operation in places where there is no access to the mains. The WLC series features a stainless steel weighing pan, and a backlit LCD guaranteeing clear weighing result presentation. WLC/A precision scales are equipped with the following interfaces: 2 x RS232, USB type A, and USB type B, wireless communication (option). The interfaces enable cooperation between the scale and peripheral devices (e.g. printer, computer, USB flash drive).
| This equipment has been tested and found to comply with the limits for
a Class A digital device, pursuant to part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment.
This equipment generates, uses and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual, may cause
harmful interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case
the user will be required to correct the interference at his own
expense.
---|---
| This equipment has been tested and found to comply with the limits for a Class A according to standards, listed in declaration of conformity, required by European regulations in order to use CE marking. This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments. Class A products may also be utilized in residential/domestic environments but may cause interference and require the user to take adequate corrective measures.
WARRANTY CONDITIONS
- RADWAG feels obliged to repair or exchange all elements that appear to be faulty by production or by construction.
- Defining defects of unclear origin and means of their elimination can only be realized with assistance of manufacturer and user representatives.
- RADWAG does not bear any responsibility for damage or losses resulting from unauthorized or inadequate performing of production or service processes.
- The warranty does not cover:
- mechanical damage caused by product exploitation other than intended, damage of thermal and chemical origin, damage caused by lightning, overvoltage in the power network or other random event,
- inappropriate cleaning habits.
- Loss of warranty takes place if:
- a repair is carried out outside RADWAG authorized service point,
- service claims intrusion into mechanical or electronic construction by unauthorized people,
- the device does not bear company security stickers.
- Warranty conditions outline the warranty period for rechargeable batteries attached to the device for 12 months.
- For detailed warranty conditions read the warranty certificate.
- Contact with the central authorized service: +48 (48) 386 63 30.
MAINTENANCE ACTIVITIES
In order to ensure safety in the course of cleaning, it is necessary to disconnect the device from the mains. With this condition met, uninstall the weighing pan and other detachable components.
Cleaning ABS Components
To clean dry surfaces and avoid smudging, use clean non-colouring cloths made
of cellulose or cotton. You can use a solution of water and detergent (soap,
dishwashing detergent, glass cleaner). Gently rub the cleaned surface and let
it dry. Repeat cleaning process if needed. In the case of hard to remove
contamination, e.g. residues of adhesive, rubber, resin, polyurethane foam
etc., you can use a special cleaning agents based on a mixture of aliphatic
hydrocarbons that do not dissolve plastics. Before using the cleanser for all
surfaces we recommend carrying out tests. Do not use cleansers containing
abrasive substances.
Cleaning Glass Components
Select dissolvent depending on a dirt. Never soak the glass panes in alkaline
solutions since they interact with glass and may cause damage. Do not use
cleansers containing abrasive substances.
For organic dirt use acetone first, next use water or detergent. For other
than organic dirt use diluted acid solutions (soluble salts of hydrochloric or
nitric acid) or base solutions (ammonium or sodium base). To remove ACIDS use
protophilic solvent (sodium carbonate), to remove BASE use protogenic solvent
(mineral acid of various concentration). In case of heavy contamination use
brush and detergent, nevertheless avoid detergents containing large and hard
molecules which could potentially scratch glass panes. At the end of the
cleaning process rinse the pane using distilled water. Use soft brush with
wooden or plastic handle exclusively in order to avoid risk of scratches. Do
not use wire brush. Rinsing is a necessary cleaning process stage allowing to
remove remaining soap, detergents and other cleansers from the panes prior
their reinstallation. After preliminary cleaning process stage, rinse the pane
using running water first, distilled next. Avoid drying the panes either using
paper towel or forced air circulation since some fibres, grains or
contamination of other type could permeate into the panes thus causing
weighing errors. We do not recommend using driers when drying measuring glass
tools. It is a frequent treatment to leave glass components on a rack to dry.
Cleaning Stainless Steel Components
Avoid using cleansers containing any corrosive chemicals, e.g. bleach
(including chlorine). Do not use cleansers containing abrasive substances.
Always remove the dirt using microfiber cloth to avoid damage of protective
coating. In case of a daily maintenance:
- Remove the dirt using cloth dipped in warm water.
- For best results, add a little bit of dishwashing detergent.
Cleaning Powder-Coated Components
For preliminary cleaning process stage you need running water or wet sponge
featuring large holes, this will help you to remove loose, heavy dirt. Do not
use cleansers containing abrasive substances. Next, using cloth and cleanser-
water solution (soap, dishwashing liquid) gently rub the cleaned surface.
Avoid using cleanser without water since it may result with damage of the
cleaned surface, please mind that large amount of water mixed with cleanser is
a must.
Cleaning Aluminium Components
While cleaning aluminium components use products acid by nature, e.g. spirit
vinegar, lemon. Do not use cleansers containing abrasive substances. Avoid
using hard brush, this may cause scratches.
It is recommended to use microfibre cloth. While polishing the surface use
circular movements. Use clean, dry cloth to make the surface shine.
SERVICE AND REPAIR
In case of any problems with correct operation of the scale, contact the closest manufacturer’s service point. In case of defects, deliver the faulty product to the manufacturer’s service point. If the product cannot be delivered to the manufacturer’s service point, call the service and report the defect. Repair scope and method will be set up.
RECYCLING
HY10 scales must be recycled, they are not to be treated as a regular
household waste. Scales to be decommissioned must be decommissioned in
accordance with valid legal regulations.
MECHANICAL DESIGN
Dimensions
Connectors Arrangement
- power supply socket 12 VDC
- RS232 (1) connector
- RS232 (2) connector
- USB A “host”
- USB B “device”
RS232 Connectors
Inputs / Outputs
Standard C32 scale is equipped with 4 optoisolated inputs and 4 semiconductor
outputs (solid-state relays). The signals are sent via DSUB15/F connector.
Technical Specifications
Output parameters
Output quantity| 4
Output type| Solid-state relay
Cable cross-section| 0.14 – 0.5 mm2
Maximum output current| 0.5 A DC
Maximum output voltage| 30 VDC
Input parameters
Input quantity| 4
Input type| Optoisolated
Cable cross-section| 0.14 – 0.5 mm2
Input voltage range| 5¸24VDC
I/O Schematic Diagrams
Operation panel
Keys:
INSTALLATION
Unpacking and Installation
- Unpack the device and place it on a flat and even surface. Keep it far away from any sources of heat.
- Install the weighing pan, follow the below figures:
Under-Pan Weighing
The WLC/A1, WLC/A2 scales enable weighing loads below scale (so called under-
pan weighing). This is an alternative for loads of non-standard dimensions
and shapes, and for loads generating magnetic field.
Preparing the scale for the under-pan weighing:
1. Unpack the scale, assembly it following section 7.1 of this user manual.|
–
---|---
2. Put the scale right side down.
| ****
3. Remove the hole plug.
| ****
4. Mount the hook and put the scale bottom side down.
|
Levelling
To level the weighing instrument turn its feet. Keep turning the feet until
the air bubble takes central position:
Connecting the Scale to the Mains
The weighing device can be connected to the mains only with a power supply
that comes standard with the particular model. Nominal voltage of the power
supply (specified on the power supply data plate) has to be compatible with
the mains nominal voltage.
Procedure:
- Connect the power supply to the mains. Plug it to the power supply socket (back side of the scale housing).
- Display test proceeds (all symbols are backlit for a moment), program name and number is displayed first, ZERO indication with reading unit next (displayed reading unit is conditioned by scale type).
- in case the weighing result is not zero after indication stabilisation, press
Warm-Up Time
- For correct scale operation, the workroom temperature must range between +15C + 30C.
- It takes 30 minutes for the device to warm up.
- During the thermal stabilization, the indications on the screen can change.
- If you want to carry out adjustment operation, make sure that your device is thermally stable.
- Temperature and humidity variations during operation may cause indication errors. To remove the indication errors perform user adjustment.
Battery Charge Status
The scale of standard design is equipped with an internal battery. Battery
state is signalled by display pictogram, the pictogram is displayed in the top
bar
pictogram display mode | Meaning |
---|---|
No pictogram | Battery charged. Regular scale operation. |
Pictogram displayed continuously | Too low battery charge (the scale is about |
to shut down). Charge the battery immediately.
Blinking pictogram, blink frequency: ca. 1 s| Battery charge in progress. The
device is connected to the power supply charging the battery.
Blinking pictogram, blink frequency: ca. 0.5 s| Battery error. Battery is
damaged.
Battery Charge Status Check
Depending on the battery state, a respective status is displayed on the
screen for 2s:
80% | Battery power supply. Battery power given in %. |
---|---|
CHArGE | Battery charge in progress. The device is connected to the power |
supply charging the battery.
-Err5-| Battery error. Battery is damaged.
OPERATING THE MENU
In order to navigate the program menu use the operation panel.
Return to the Weighing Mode
Introduced menu modifications are automatically saved to scale memory upon
return to the home screen. To return to the home screen press repeatedly.
WEIGHING
Load the weighing pan. Read the result when stability marker is displayed. To assure long-term operation and correct mass measurements, follow the rules presented below:
DIAGRAMS OF CONNECTION CABLES
TECHNICAL SPECIFICATIONS
For technical specifications of respective scales go to RADWAG website www.radwag.pl.
TROUBLESHOOTING
Problem | Cause | Solution |
---|
Scale start-up fail.
| Power supply disconnected.| Connect the power supply to the scale.
Battery discharged.| Connect the power supply to the mains, charge the
battery.
No battery (not installed or installed incorrectly).| Check if the battery is
installed correctly (polarization).
The scale switches off automatically.
| <7.4.t1> parameter set to value enforcing scale shut- down after particular time
interval.
| ****
Go to <P7.Othr> menu, set <7.4.t1> parameter to ‘nonE’ value.
During the start-up, message ‘LH’ is displayed.| Weighing pan loaded during the start-up.| Unload the weighing pan. Zero indication is displayed.
Communication with the computer not established.
| Incorrect computer port set in parameter <5.1.1.Prt>.| Enter < P5.ducE /
5.1.PC> submenu and set correct <5.1.1.Prt> parameter value.
Incorrect transmission parameters for the selected computer port.|
Enter <P4.Conn> menu and set correct transmission parameters for the selected
computer port.
Incorrect printout frequency for continuous transmission.| Enter < P5.ducE /
5.1.PC> submenu and set correct <5.1.3.Int> parameter value.
No printout on a scale-connected printer.
| Incorrect printer port set in
<5.2.1.Prt> parameter.
| Enter < P5.ducE / 5.2.Prtr> submenu and set correct <5.2.1.Prt> parameter
value.
Incorrect transmission parameters for the selected printer port.|
Enter <P4.Conn> menu and set correct transmission parameters for the selected
printer port.
No variable declared in weighing printout project.| Enter <P6.Prnt /
6.2.GLP> submenu and declare variables that are to be printed.
Communication with the additional display not established..
| Incorrect additional display port set in <5.3.1.Prt> parameter.| Enter <
P5.ducE / 5.3.AdSP> submenu and set correct <5.3.1.Prt> parameter value.
Incorrect transmission parameters for the selected computer port.|
Enter <P4.Conn> menu and set correct transmission parameters for the selected
additional display port.
Displayed mass unit does not comply with the scale data plate.
| Changed scale start unit in
<9.1.UnSt> parameter.
| Enter <P9.Unit / 9.1.UnSt> submenu and set unit complying with the scale
data plate.
Changed custom unit in
<9.2.Unin> parameter.
| Enter <P9.Unit / 9.2.Unin> submenu and set unit complying with the scale data plate.
ERROR MESSAGES
– E r r 2 – | Value beyond zero range. |
---|---|
– E r r 3 – | Value beyond tare range. |
– E r r 4 – | Adjustment weight or start mass out of range (±1% for |
adjustment weight, ±10 for start mass).
– E r r 5 –| Battery error. Battery is damaged.
– E r r 8 –| Time of the following operations exceeded: taring, zeroing,
start mass determining, adjustment process.
– n u l l –| Zero value from converter.
– F U L L –| Weighing range exceeded.
– L H –| Start mass error, indication out of range (±10% of start mass).
– H i –| Display range of total mass on scale display exceeded in
‘Totalizing’ mode.
– u L o –| Too low battery charge. The scale is about to shut down.
– E r r L o –
| – Determined mass of single part in ‘Parts counting’ mode too small.
– Value of ‘Min’ threshold is greater than value of ‘Max’ threshold in ‘+/- control’ mode.
– E r r H i –
| – Entered value of single part greater than maximum capacity in ‘Parts counting’ working mode.
– Entered value of ‘Max’ threshold greater than maximum capacity in ‘+/- control’ mode.
– Entered reference mass greater than maximum capacity in ‘Percent weighing’ mode.