Promega AM2100 GoTaq Enviro Wastewater SARS-CoV-2 Systems Instruction Manual

June 5, 2024
Promega

Promega AM2100 GoTaq Enviro Wastewater SARS-CoV-2 Systems

Promega-AM2100
-Go-taq-Enviro-Wastewater-SARS-CoV-2-Systems-product-img

Description

Early in the COVID-19 pandemic, scientific studies demonstrated that the genetic material of SARS-CoV-2—an enveloped RNA virus—could frequently be detected in the feces of infected individuals (1–3) and thereafter was also detected in wastewater (4). This finding mobilized the global water sector to investigate if wastewater-based epidemiology (WBE) using the genetic signal of SARS-CoV-2 could be used to track the spread of COVID-19 in communities. In the past, WBE was critical in identifying the community prevalence of poliovirus in support of the World Health Organization’s (WHO) program to eradicate poliomyelitis and was also used to investigate opioid use in communities (5). Detection of SARS-CoV-2 in wastewater has the potential to provide an integrated, community-level indication of the presence of COVID-19. The GoTaq® Enviro Wastewater SARS-CoV-2 Systems for wastewater samples are four-dye, three-target hydrolysis probe-based reverse transcription (RT)-qPCR kits that amplify SARS-CoV-2 genetic signatures. The systems include primer/probe sets prescribed by the U.S. Centers for Disease Control (CDC) that target the N1 or N2 regions of the SARS-CoV-2 nucleocapsid (N) gene, or a primer/probe set targeting the envelope (E) gene (6). The systems are designed for a one-step RT-qPCR. As an internal process control and to allow for normalization, the systems include primers to detect Pepper Mild Mottle Virus (PMMoV), an RNA virus commonly found in wastewater (7), and an internal amplification control (IAC). The GoTaq® Enviro-Master Mix, provided in this kit, uses proprietary enzymes and formulations that tolerate reverse transcriptase and PCR inhibitors such as humic acids, that can be present in nucleic acid samples purified from wastewater.

Product Components (Section 2) and Related Products (Section 8. C) include:

  • Target Genes: Three different primer/probe sets are supplied as 20X Primer/Probe/IAC mixtures for detecting either the N1 or the N2 fragment of the nucleocapsid (N) gene, or the envelope protein (E) gene of SARS-CoV-2. These amplicons are detected using Fluorescein- (FAM™-) labeled probes.
  • Controls: Each 20X Primer/Probe/IAC Mix contains two controls:
    • Process Control: Detection of the Pepper Mild Mottle Virus (PMMoV), an abundant RNA virus commonly found in wastewater. The primers and Quasar® 670 dye-labeled probe amplify and detect a 68bp region of the PMMoV genome and are detected in the Cy®5 channel.
    • Internal Amplification Control (IAC): The 20X Primer/Probe/IAC Mix contains primers, probes, and templates for an internal amplification control (IAC). The primers and the CAL Fluor® 560 dye-labeled probe amplify and detect a 285bp product from an RNA template that is included in every amplification reaction. IAC performance is used to detect RT-qPCR inhibitors (Taq polymerase and reverse transcriptase inhibitors) in the sample and are detected in the HEX™/JOE™/VIC® dye channel.
  • Passive Reference Dye: The 20X Primer/Probe/IAC Mix includes a reference dye (CXR, carboxy-X-rhodamine) that has similar spectral properties to ROX™ dye. The final concentration of the Passive Reference dye in RT-qPCR is 30nM.
  • DNA Polymerase and Reverse Transcriptase: GoTaq® Enviro qPCR Master Mix contains thermostable DNA polymerase and the GoScript™ Enzyme Mix contains reverse transcriptase. These mixes are designed to tolerate a diverse range of DNA polymerase and reverse transcriptase inhibitors, including those found in wastewater.
  • RNA Quantitation Standards: GoTaq® Enviro Wastewater SARS-CoV-2 Systems contain two in vitro transcribed RNA fragments: SARS-CoV-2 (N+E) RNA, 4 × 106 copies/µl, and PMMoV RNA, 4 × 106 copies/µl. These RNA fragments serve as stable quantitation standards that can be used to generate a standard curve.
  • Nuclease-Free Water: Can be used as a negative no-template control (NTC), for diluting the quantitation standards and for adjusting the setup volume for RT-qPCR reaction mixes.

Product Components and Storage Conditions

GoTaq® Enviro Wastewater SARS-CoV-2 System, N1/N2/E Not For Medical Diagnostic Use. The system contains sufficient reagents for 200 reactions at 20µl. Includes:

  • 1 × 100µl N1 & PMMoV Primer/Probe/IAC Mix, 20X
  • 1 × 100µl N2 & PMMoV Primer/Probe/IAC Mix, 20X
  • 1 × 100µl E & PMMoV Primer/Probe/IAC Mix, 20X
  • 2 × 1,000µl GoTaq® Enviro Master Mix, 2X
  • 1 × 100µl GoScript™ Enzyme Mix
  • 2 × 1.25ml Nuclease-Free Water
  • 1 × 100µl SARS-CoV-2 (N+E) RNA, 4 × 106 copies/µl
  • 1 × 100µl PMMoV RNA, 4 × 106 copies/µl

GoTaq® Enviro Wastewater SARS-CoV-2 System, N1 200 reactions AM2110

Not For Medical Diagnostic Use. The system contains sufficient reagents for 200 reactions at 20µl. Includes:

  • 2 × 100µl N1 & PMMoV Primer/Probe/IAC Mix, 20X
  • 2 × 1,000µl GoTaq® Enviro Master Mix, 2X
  • 1 × 100µl GoScript™ Enzyme Mix
  • 2 × 1.25ml Nuclease-Free Water
  • 1 × 100µl SARS-CoV-2 (N+E) RNA, 4 × 106 copies/µl
  • 1 × 100µl PMMoV RNA, 4 × 106 copies/µl

GoTaq® Enviro Wastewater SARS-CoV-2 System, N2

Not For Medical Diagnostic Use. The system contains sufficient reagents for 200 reactions at 20µl. Includes:

  • 2 × 100µl N2 & PMMoV Primer/Probe/IAC Mix, 20X
  • 2 × 1,000µl GoTaq® Enviro Master Mix, 2X
  • 1 × 100µl GoScript™ Enzyme Mix
  • 2 × 1.25ml Nuclease-Free Water
  • 1 × 100µl SARS-CoV-2 (N+E) RNA, 4 × 106 copies/µl
  • 1 × 100µl PMMoV RNA, 4 × 106 copies/µl

Product Components and Storage Conditions (continued)

GoTaq® Enviro Wastewater SARS-CoV-2 System, E

Not For Medical Diagnostic Use. The system contains sufficient reagents for 200 reactions at 20µl. Includes:

  • 2 × 100µl E & PMMoV Primer/Probe/IAC Mix, 20X
  • 2 × 1,000µl GoTaq® Enviro Master Mix, 2X
  • 1 × 100µl GoScript™ Enzyme Mix
  • 2 × 1.25ml Nuclease-Free Water
  • 1 × 100µl SARS-CoV-2 (N+E) RNA, 4 × 106 copies/µl
  • 1 × 100µl PMMoV RNA, 4 × 106 copies/µl

Storage Conditions: Store all components of the GoTaq® Enviro Wastewater SARS-CoV-2 Systems at –30°C to –10°C. Limit freeze-thaws to five cycles or less. Store the 20X Primer/Probe/IPC Mixes protected from light.

General Considerations

GoTaq® Enviro Wastewater SARS-CoV-2 Systems are sensitive; take precautions to minimize contamination. We recommend storing the reagents separately from RNA/TNA (total nucleic acid) samples. We also recommend using clean designated work areas and separate pipettes for pre-and post-amplification steps to minimize the potential for cross-contamination between RNA samples and to prevent carryover of nucleic acid from one run to the next. Wear a lab coat and protective eyewear. Wear gloves and change them often. Prevent contamination by using aerosol-resistant pipette tips. Always include a no- template control (NTC) reaction to detect contamination. We recommend performing NTC reactions in triplicate. Do not unseal reaction plates after amplification is complete. Unsealing the plates increases the risk of contaminating subsequent reactions with the amplified products.

Materials to Be Supplied by Users

  • sterile aerosol-resistant barrier pipette tips
  • pipettes dedicated to pre-amplification work
  • 1.5ml tubes to prepare the reaction mixes
  • 0.5ml low-bind tubes (e.g., Eppendorf Cat.# 022431005) to prepare the standard dilutions
  • qPCR plates or strip tubes with caps
  • qPCR thermocycler (FAM™, HEX™, Cy®5 channels; ROX™ channel if reference dye required; see Section 8. B)

System Usage
The GoTaq® Enviro Wastewater SARS-CoV-2 Systems are RT-qPCR kits designed for detecting SARS-CoV-2 genetic signals from wastewater samples that have been preprocessed. Processing includes viral concentration and purification of nucleic acid. The purified nucleic acid is then used for RT-qPCR. Viral concentration and purification can be achieved by using one of the following kits:

  • Wizard® Enviro TNA Kit (Cat.# A2991)
  • Maxwell® RSC Enviro TNA Kit (Cat.# AS1831)

Alternative viral concentration methods can also be used, such as PEG 8000/NaCl precipitation, membrane filtration, centrifugal ultrafiltration, skimmed milk flocculation, or other methods. Nucleic acid purification can be performed on the concentrated viral material using manual or automated systems.

GoTaq® Enviro Wastewater SARS-CoV-2 Systems Protocol
Note: To avoid contamination of samples with external sources of PCR templates, perform all steps with aerosol-resistant pipette tips.

Preparing Standard Curve Dilutions for SARS-CoV-2 (N+E) and PMMoV RNAs Promega-AM2100-Go-taq-Enviro-Wastewater-SARS-CoV-2-Systems-
fig-1

  1. Thaw the SARS-CoV-2 (N+E) RNA, 4 × 106 copies/μl, and PMMoV RNA, 4 × 106 copies/μl, avoiding long exposures to ambient temperature.
  2. Combine and dilute the SARS-CoV-2 (N+E) and PMMoV RNAs (4 × 106 copies/µl) 100-fold by adding 2µl of each RNA to 196µl of Nuclease-Free Water, for a final concentration of 4 × 104 copies/µl.
  3. Perform subsequent serial tenfold dilutions in low-binding 0.5ml tubes. For example, combine 5µl of RNA with 45μl of Nuclease-Free Water to obtain the following standard curve dilutions (4 × 104–4 copies/µl; see Table 1 and Figure 1). Vortex each dilution for 3–5 seconds prior to removing an aliquot for the next dilution. Change pipette tips between dilutions.

Table 1. Standard Curve Dilutions for SARS-CoV-2 (N+E) and PMMoV RNAs


Tube (Figure 1)

| SARS-CoV-2 (N+E)

and PMMoV RNA (copies/µl)

| Copies/Well

(5μl sample/ 20µl reaction)

| ****

Copies/Well (4μl sample/ 10µl reaction)

---|---|---|---
A| 4 × 104| 2 × 105| 1.6 × 105
B| 4 × 103| 2 × 104| 1.6 × 104
C| 4 × 102| 2 × 103| 1.6 × 103
D| 40| 2 × 102| 1.6 × 102
E| 4| 20| 16

Preparing Standard Curve Dilutions for SARS-CoV-2 (N+E) and PMMoV RNAs (continued)

Dilution scheme for combined SARS-CoV-2 (N+E) and PMMoV RNA standards. 4.B. Preparing the RT-qPCR Reaction Mix (20µl Reaction Volume)

We recommend preparing three technical RT-qPCR replicates for increased statistical power.

  1. Vortex the GoTaq® Enviro-Master Mix briefly before use to ensure homogeneity. Centrifuge briefly to collect contents at the bottom of the tube.
  2. Determine the number of reaction wells needed. This should include reactions for combined SARS-CoV-2 (N+E) and PMMoV RNA standards and negative control reactions. Add 1 or 2 reactions to this number to compensate for pipetting error.
  3. While this approach consumes a small additional amount of each reagent, it ensures that enough RT-qPCR amplification mix will be available for all samples. It also ensures that each reaction contains the same RT-qPCR amplification mix.

Table 2. Reaction Mix Worksheet for 20μl Reaction Volume


RT-qPCR Reaction Mix

| Volume per Reaction (X)| Number of Reactions (n)| Final Volume (X × n)
---|---|---|---
GoTaq ® ** Enviro Master Mix, 2X| 10µl| |
GoScript
™** Enzyme Mix| 0.4µl| |
Primer/Probe/IAC Mix, 20X| 1µl| |
Nuclease-Free Water| 3.6µl| |

  1. Assemble the reaction mix by combining the GoTaq® Enviro-Master Mix, GoScript™ Enzyme Mix, Primer/Probe/IAC Mix 20X, and Nuclease-Free Water calculated in Step 2.
  2. Pipette 15µl of RT-qPCR reaction mix into wells of 96-well qPCR plates.
  3. Add 5µl of extracted nucleic acid, combined SARS-CoV-2 (N+E) and PMMoV RNA standards or Nuclease-Free Water for NTC. The final reaction volume should be 20µl.
  4. Seal and centrifuge the plate at approximately 300 × g for 1 minute to ensure all liquid is collected at the bottom of the plate wells. Protect from extended light exposure and elevated temperatures before cycling. The samples are now ready for thermal cycling.

Preparing the RT-qPCR Reaction Mix (10µl Reaction Volume) The systems are also amenable to 10µl reaction volumes.

  1. Vortex GoTaq® Enviro-Master Mix briefly before use to ensure homogeneity. Centrifuge briefly to collect contents at bottom of the tube.
  2. Determine the number of reaction wells needed. This should include reactions for the combined SARS-CoV-2 (N+E) and PMMoV RNA standards and negative control reactions. Add 1 or 2 reactions to this number to compensate for pipetting error. While this approach consumes a small additional amount of each reagent, it ensures that enough RT-qPCR amplification mix will be available for all samples. It also ensures that each reaction contains the same RT-qPCR amplification mix.

Table 3. Reaction Mix Worksheet for 10μl Reaction Volume


RT-qPCR Reaction Mix

| Volume per Reaction (X)| Number of Reactions (n)| Final Volume (X × n)
---|---|---|---
GoTaq ® ** Enviro Master Mix, 2X| 5µl| |
GoScript
™** Enzyme Mix| 0.2µl| |
Primer/Probe/IAC Mix, 20X| 0.5µl| |
Nuclease-Free Water| 0.3µl| |

  1.  Assemble the reaction mix by combining the GoTaq® Enviro-Master Mix, GoScript™ Enzyme Mix, Primer/Probe/IAC Mix 20X, and Nuclease-Free Water calculated in Step 2.
  2. Pipette 6µl of RT-qPCR reaction mix into wells of 96-well qPCR plates.
  3. Add 4µl of extracted nucleic acid, combined SARS-CoV-2 (N+E), and PMMoV RNA standards or Nuclease-Free Water for NTC. The final reaction volume should be 10µl.
  4. Seal and centrifuge the plate at approximately 300 × g for 1 minute to ensure all liquid is collected at the bottom of the plate wells. Protect from extended light exposure and elevated temperatures before cycling. The samples are now ready for thermal cycling.

Prepare the reaction mix by combining GoScript™ Enzyme Mix, GoTaq® Enviro- Master Mix, primers, and probe. Assemble the reaction. Perform RT-qPCR using standard or FAST cycling conditions on a real-time PCR instrument. Analyze amplification and standard curve data.Promega-AM2100-Go-taq-Enviro-
Wastewater-SARS-CoV-2-Systems-fig-2

An overview of the GoTaq® Enviro Wastewater SARS-CoV-2 Systems protocol.

Thermal Cycling

The PCR cycling parameters and instrument settings are shown here are provided as guidelines and can be modified as necessary for optimal results.

Standard Cycling Conditions

Step Temperature (°C) Time Number of Cycles
Reverse transcription 45 15 minutes 1
RT inactivation/GoTaq® activation 95 2 minutes 1
Denaturation 95 15 seconds 40
Annealing/extension 62 60 seconds

FAST Cycling Conditions

Step Temperature (°C) Time Number of Cycles
Reverse transcription 45 15 minutes 1
RT inactivation/GoTaq® activation 95 2 minutes 1
Denaturation 95 3 seconds 40
Annealing/extension 62 30 seconds

Collect data from the following fluorescence channels at the end of each 62°C annealing/extension step. Performing >40 PCR cycles is not recommended as it may generate nonspecifically

Fluorophores Target
FAM N1, N2, E (SARS-CoV-2)
CAL Fluor®560/HEX/JOE/VIC Internal Amplification Control
ROX/CXR Reference Dye
Quasar®670/Cy5 PMMoV

Dispose of PCR plates as biohazardous waste per your institutional guidelines. To avoid DNA contamination of your lab space and subsequent samples, do not open the PCR plates after completing amplification and data collection.

Thermal Cycling

| SARS-CoV-2 (N+E) and

PMMoV RNA Standards or NTC

| ****


Nucleic Acid Purified from Wastewater

---|---|---
| 1| 2| 3| 4| 5| 6| 7| 8| 9| 10| 11| 12
A| 2 × 105| 2 × 105| 2 × 105| Sample

1

| Sample

1

| Sample

1

| Sample

9

| Sample

9

| Sample

9

| Sample

17

| Sample

17

| Sample

17

B| 2 × 104| 2 × 104| 2 × 104| Sample

2

| Sample

2

| Sample

2

| Sample

10

| Sample

10

| Sample

10

| Sample

18

| Sample

18

| Sample

18

C| 2 × 103| 2 × 103| 2 × 103| Sample

3

| Sample

3

| Sample

3

| Sample

11

| Sample

11

| Sample

11

| Sample

19

| Sample

19

| Sample

19

D| 2 × 102| 2 × 102| 2 × 102| Sample

4

| Sample

4

| Sample

4

| Sample

12

| Sample

12

| Sample

12

| Sample

20

| Sample

20

| Sample

20

E| 20| 20| 20| Sample

5

| Sample

5

| Sample

5

| Sample

13

| Sample

13

| Sample

13

| Sample

21

| Sample

21

| Sample

21

F| NTC| NTC| NTC| Sample

6

| Sample

6

| Sample

6

| Sample

14

| Sample

14

| Sample

14

| Sample

22

| Sample

22

| Sample

22

G| | | | Sample

7

| Sample

7

| Sample

7

| Sample

15

| Sample

15

| Sample

15

| Sample

23

| Sample

23

| Sample

23

H| | | | Sample

8

| Sample

8

| Sample

8

| Sample

16

| Sample

16

| Sample

16

| Sample

24

| Sample

24

| Sample

24

Example plate layout for one target (N1, N2, or E), plus standard RNAs and NTCs

Note: For the 10µl reaction size refer to Table 1, Section 4. A for accurate copy numbers per reaction in the standard curve wells.

Data Analysis and Interpretation

Evaluate qPCR Assay Standard Curves (FAM (SARS-CoV-2)/Cy5 (PMMoV))

Common qPCR analysis software packages apply a linear regression to the standard dilution series data and calculate the best fit of the standard curve using y = MX + b, where x = Log10 concentration; y = Cq/Ct; m = slope. r2 measures goodness of fit to the regressed line and m is a measure of efficiency, where m= –3.3 indicates 100% PCR efficiency (i.e., the amplification product is doubled at each cycle). The y-intercept (b in the equation) is they value Cq at x = 0. For example, b corresponds to the Cq value for a sample with a concentration of 1 copy/reaction (Log10(1) = 0). In general, the standard curve for each PCR target has an average slope (m) in the range of –3.0 to –3.7 (corresponding to a qPCR efficiency of 86%<E<115%) and an r2 value >0.970. We recommend monitoring y-intercept values for any significant changes from run to run.

Analyze Internal Amplification Control Signal (HEX/CAL Fluor 560/VIC/JOE)

The IAC can be used to evaluate the overall performance of the SARS-CoV-2 RT- qPCR amplification reaction and to detect DNA polymerase and/or reverse transcriptase inhibition. The probe used is a dual-labeled probe (CAL Fluor® 560/BHQ1). Depending on the qPCR instrument used, HEX, JOE, VIC and CAL Fluor® 560 channels can be used to record the amplification signal. Depending on the qPCR instrument and analysis software used, the IAC Ct should fall in the range of 20–30 for the NTC reactions.
If the IAC Ct in a sample well is shifted significantly (Ct ≥ 2) compared to NTC well, PCR inhibitors are present in the experimental sample, and results should be considered qualitative and not quantitative. Repeat the purification or clean-up of nucleic acid if necessary. If a sample yields no detectable amplification for SARS-CoV-2 but exhibits IAC amplification (Ct = 20–30) and PMMoV amplification (Ct = 20–40), SARS-CoV-2 is not detectable with this system. If the IAC fails to amplify or the IAC Ct is shifted > 3 Ct compared to NTC wells, no conclusions can be made about the absence of SARS-CoV-2 genetic material in a sample. Results can be considered invalid. See Table 4 for examples. IAC can fail to amplify if the assay is set up incorrectly.

Analyze Process Control Signal, PMMoV (Cy5/Quasar® 670)
Wastewater samples typically exhibit PMMoV fluorescence growth curves that cross the threshold at <40 cycles. If a sample yields no detectable amplification for SARS-CoV-2 but exhibits IAC amplification (Ct = 20–30), and PMMoV amplification (Ct = 20–40), SARS-CoV-2 is not detectable with this system. Failure to detect PMMoV in wastewater samples may indicate:

  • improper extraction of nucleic acid from samples resulting in loss of RNA, RNA degradation or both
  • inhibition of reverse transcriptase, DNA polymerase, or both by inhibitors in the sample
  • absence of sufficient nucleic acid due to poor collection or pasteurization of sample
  • improper assay setup and execution
  • reagent or equipment malfunction

If the PMMoV reaction (Cy®5 channel) is negative, IAC is positive and SARS- CoV-2 N1 or N2 or E are positive, the result can be considered valid because PMMoV negativity may reflect a low PMMoV viral load. If all SARS-CoV-2 markers, PMMoV (process control), an internal amplification control (IAC) are negative for the specimen, the results are invalid. If the residual sample is available, repeat the extraction procedure and retest. If all markers remain negative after retesting, report the results as invalid. A new specimen should be collected if possible.

D. No-Template Control (NTC)
For an NTC, use nuclease-free Water in the RT-qPCR instead of a nucleic acid- containing sample or RNA standards. NTC samples should produce amplification curves for the IAC in the HEX™ channel. Sample contamination is indicated if FAM™ or Cy®5 NTC reaction channels exhibit a fluorescence curve with Ct value indicating a copy number greater than the limit of quantification (LOQ). LOQ for the assay is 20 copies per reaction for the SARS-CoV-2 genetic signal (in the FAM channel: N1, N2, and E, respectively).

Limit of Detection (LOD) and Limit of Quantification (LOQ)

LOD is defined as the lowest amount of analyte in a sample that can be detected with a 95% probability. The LOD for these assays is 8 copies of nucleic acid per reaction.
LOQ is defined as the lowest amount of analyte in a sample that can be quantitatively determined with less than a 25% coefficient of variance (8–10).
In Nuclease-Free Water (reference matrix) spiked with SARS-CoV-2 RNA, the LOQ of the assay is 20 copies per reaction. If SARS-CoV-2 genetic signal amplification appears after the signal for the LOQ, the amount of SARS-CoV-2 in the sample cannot be determined with certainty.

Interpreting Wastewater Signal Profiles

Table 4. Interpreting Wastewater Results.

| ****

SARS-CoV-2

| | | PMMoV

Process Control (Cy5)

| ****

Internal Amplification Control (HEX)

|
---|---|---|---|---|---|---
N1 (FAM)| N2 (FAM)| E (FAM)| NTC (FAM)| Result
Any one or more is positive| –| +/–| +/–| SARS-CoV-2

Detected

–| –| –| –| +| +| SARS-CoV-2

Not Detected

Any one or more is positive| +| +/–| +/–| Invalid (False Positive)
–| –| –| –| +/–| –| Invalid (False Negative)

Calculating Viral Nucleic Acid
The following formula can be applied to quantitate the amount of SARS-CoV-2 nucleic acid in a sample:

Viral genome (copies/liter) = Copies in RT-qPCR × 1,000

The volume of nucleic acid extract used in RT-qPCR (ml)*

| × Concentration factor

*If 5µl of the nucleic acid extract is used in RT-qPCR, the value in ml is 0.005.

Concentration factor = Wastewater sample volume used (ml)

The volume of nucleic acid extracted (ml)

Normalization with PMMoV

Quantitation of PMMoV viral genome copies can be performed using the same approach as for SARS-CoV-2 using the PMMoV RNA Quant Standard.
Changes in SARS-CoV-2 levels can be analyzed relative to the PMMoV levels by using this formula:

Relative SARS-CoV-2 signal = SARS-CoV-2 signal (copies/liter)

PMMoV signal (copies/liter)

Specificity Testing

Wastewater TNA isolates contain abundant nucleic acid originating from various bacterial and viral species. The GoTaq® Enviro Wastewater SARS-CoV-2 systems were carefully designed to amplify only the designated SARS-CoV-2 genomic target.

Table 5. Specificity of the GoTaq ® **Enviro Wastewater SARS- CoV-2 Systems When Tested Against Respiratory Pathogens.**

Pathogen N1 N2 E
SARS-CoV-2 + + +
OC43
229E
HKU1
NL63
Influenza A
RSV
L. pneumophila
P. aeruginosa

Appendix

References

  1. Wang, D. et al. (2020) Clinical characteristics of 138 hospitalized patients with 2019 novel coronavirus-infected pneumonia in Wuhan, China. JAMA 323(11), 1061–9.

  2. Xiao, F. et al. (2020) Evidence for gastrointestinal infection of SARS-CoV-2. Gastroenterology 158(6), 1831–3.

  3. Xu, Y. et al. (2020) Characteristics of pediatric SARS-CoV-2 infection and potential evidence for persistent fecal viral shedding. Nat. Med. 26(4), 502–5.

  4. Ahmed, W. et al. (2020) First confirmed detection of SARS-CoV-2 in untreated wastewater in Australia: A proof of concept for the wastewater surveillance of COVID-19 in the community. Sci. Total Environ. 728, 138764.

  5. Asghar, H. et al. (2014) Environmental surveillance for polioviruses in the Global Polio Eradication Initiative. J. Infect. Dis. 210, Suppl 1: S294–303.

  6. Corman, V.M. et al. (2020) Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR.
    Euro. Surveill. 25(3), 2000045.

  7. Symonds, E.M., Rosario, K. and Breitbart, M. (2019) Pepper mild mottle virus: Agricultural menace turned effective tool for microbial water quality monitoring and assessing (waste)water treatment technologies. PLoS Pathog. 15(4), e1007639.

  8. Forootan, A. et al. (2017) Methods to determine the limit of detection and limit of quantification in quantitative real-time PCR (qPCR). Biomol. Detect. Quantity. 12, 1–6.

  9. Kralik, P. and Ricchi, M. (2017) A basic guide to real-time PCR in microbial diagnostics: Definitions, parameters, and everything. Front. Microbiol. 8, 108.

  10. Bustin, S.A. et al. (2009) The MIQE guidelines: minimum information for publication of quantitative real-time PCR experiments. Clin. Chem. 55(4) 611–22.

  11. Mondal, S. et al. (2021) A direct capture method for purification and detection of viral nucleic acid enables epidemiological surveillance of SARS-CoV-2. Sci. Total Environ. 7(795), 148834.

qPCR Instruments and Reference Dye Requirements Instruments that do not require supplemental reference dye:

  • Bio-Rad CFX96 Real-Time PCR Detection System
  • Bio-Rad/MJ Research Chromo4™ Real-Time Detector
  • Bio-Rad iCycler iQ® and iQ®5 Real-Time PCR Detection Systems
  • Roche LightCycler® 480 Real-Time PCR System
  • MyGo Pro IT-IS

Instruments that require low levels (30nM) of reference dye:

  • Applied Biosystems® 7500 and 7500 FAST Real-Time PCR System
  • Applied Biosystems® QuantStudio® Real-Time PCR Systems
  • Applied Biosystems® ViiA® 7 Real-Time PCR System
  • Stratagene/Agilent Mx3000P® and Mx3005P® Real-Time PCR Systems
  • Stratagene/Agilent Mx4000® Multiplex Quantitative PCR System Instruments that require high levels (500nM) of reference dye:
  • Applied Biosystems® StepOne™ and StepOnePlus™ Real-Time PCR Systems
  • Applied Biosystems® 7300 and 7900HT Real-Time PCR System

8.C. Related Products

Amplification Systems and Accessories Product

|

Size

|

Cat.#

---|---|---
SARS-CoV-2 (N+E) dsDNA Quant Standard| 100µl| AM2060
PMMoV RNA Quant Standard| 100µl| AM2070
SARS-CoV-2 (N+E) RNA Quant Standard| 100µl| AM2050
GoTaq® Enviro qPCR System| 200 reactions| AM2000
| 1,000 reactions| AM2001
GoTaq® Enviro RT-qPCR System
| 200 reactions| AM2010
| 1,000 reactions| AM2011
IPC qPCR Inhibition Control Assay, CAL Fluor® 560| 100 reactions| AM2030
IAC RT-qPCR Inhibition Control Assay, CAL Fluor® 560
| 100 reactions| AM2040
GoScript™ Reverse Transcriptase| 100 reactions| A5003
| 500 reactions| A5004
RNasin® Plus RNase Inhibitor| 2,500u| N2611
| 10,000u| N2615
Set of dATP, dCTP, dGTP, dUTP| 10µmol each| U1335
| 40µmol each| U1245
RQ1 RNase-Free DNase| 1,000u| M6101
MgCl2| 1.5ml| A3511
Nuclease-Free Water| 50ml| P1193
*For Research Use Only. Not for use in diagnostic procedures. Not For Medical Diagnostics Use.| |
Manual Nucleic Acid Purification Systems and Reagents| |
Product| Size| Cat.#
Wizard® Enviro TNA Kit| 25 preps| A2991
Vac-Man® 96 Vacuum Manifold| 1 each| A2291
Eluator™ Vacuum Elution Device| 4 each| A1071
PEG 8000, Molecular Biology Grade| 500g| V3011
Sodium Chloride, Molecular Biology Grade| 1kg| H5273
Automated RNA Purification

Product

|

Size

|

Cat.#

---|---|---
Maxwell® RSC Enviro TNA Kit| 48 preps| AS1831
Maxwell® RSC PureFood GMO and Authentication Kit| 48 preps| AS1600

Summary of Changes

The following change was made to the 4/22 revision of this document :

  1. In Section 8.B, the Applied Biosystems® 7300 and 7900HT Real-Time PCR System was removed from the list of “Instruments that require low levels (30nM) of reference dye”.© 2021–2022 Promega Corporation.

All Rights Reserved

GoTaq, Maxwell, Vac-Man, and Wizard have registered trademarks of Promega Corporation. Evaluator and NoScript are trademarks of Promega Corporation. Applied Biosystems is a registered trademark of Applied Biosystems LLC. CalFluor and Quasar are registered trademarks of Biosearch Technologies. Cy is a registered trademark of GE Healthcare Biosciences. FAM is a trademark of Applera Corporation. iCycler iQ is a registered trademark and Chromo4 is a trademark of Bio-Rad Laboratories, Inc. LightCycler is a registered trademark of Roche. Mastercycler is a registered trademark of Eppendorf-Netheler-Hinz GmbH. Mx3000P, Mx3005P, and Mx4000 are registered trademarks of Agilent Technologies, Inc. QuantStudio, VIC, and ViiA are registered trademarks and HEX and ROX are trademarks of Thermo Fisher Scientific, Inc. Products may be covered by pending or issued patents or may have certain limitations. Please visit our Web site for more information. All prices and specifications are subject to change without prior notice. Product claims are subject to change. Please contact Promega Technical Services or access the Promega online catalog for the most up-to-date information on Promega products.