Installation and user’s manual
LFR30B
INFRARED FOREHEAD THERMOMETER
Foreword
Please read the instruction manual carefully before using the infrared forehead thermometer for the first time, because the correct temperature measurement can be made only when the infrared thermometer is used in a correct way. The instruction manual presents the specific steps for measuring forehead temperature with the thermometer and effective tips on how to reliably measure the forehead temperature. Please keep the instruction manual properly.
Chapter 1 Safety Instructions
1.1 Warning
-
Keep the thermometer out of reach by children under 12 years old alone.
-
Do not use the thermometer for any other purpose than intended.
-
The thermometer is not waterproof, do not immerse it in water or other liquids of any kind.
-
Do not keep the thermometer in an extreme environment.
-
Please keep a distance of 0-5cm between the thermometer sensor and the eye and brow when measuring.
-
If the infrared thermometer is kept in a place where the temperature is lower or higher than that of the place where it is used, please put it in the room where it is to be used 30 minutes in advance.
-
If the thermometer itself or its scanning device is damaged, do not continue to use it, do not repair the damaged thermometer without authorization, and do not insert any sharp objects into the scanning area or other openings of the thermometer.
-
The thermometer measurement does not substitute for diagnosis by physicians.
-
This infrared forehead thermometer does not apply to premature or underage infants.
-
Do not allow children to take their own temperature without supervision
-
Do not remove the thermometer until you hear the beep.
-
Please try to take the temperature in the same place; otherwise, you may get different results.
-
Do not hold the sensor when measuring to avoid the error code due to temperature instability.
1.2 Note: -
Please follow some instructions offered in Cleaning and Maintenance to clean the thermometer.
-
Remove the battery when the thermometer will not be used for a long time.
-
The thermometer contains high-quality precision parts; do not drop the thermometer or expose it to shock or vibration; do not twist the thermometer or its sensor.
-
Seek medical treatment in time in case of dysphoria, vomiting, diarrhea, dehydration, appetite, or behavioral pattern changes for unknown reasons.
-
Please note the storage and use conditions in the section of “Product Specifications”.
-
Protect the thermometer sensor from dirt and dust.
1.3 Description of graphic symbols
Symbol| Description| Symbol|
Description
---|---|---|---
**| Warning**| | Upward
| Avoid rain| | Avoid sunlight
| Type BF application part| | Fragile
| Sequence number| | Production date
| Temperature limit| | Humidity limit
| Disposal instructions for electronic devices| | Consult the instruction
manual
| Validity| | Atmospheric pressure
| Manufacturer
1.4 Environmental protection
The company designs and manufactures products for body temperature measurement
according to the safety and environmental protection requirements. The
equipment will not cause any harm to people or the environment if any outer
cover of the product is not taken apart or the equipment is always used in a
correct way. When materials that are potentially hazardous to the environment
must be used, as permitted by laws and regulations, they must be handled in
the right way.
Warning:
Do not dispose of the waste generated by the body temperature measurement
products along with the industrial or domestic waste. According to the local
and national environmental regulations, the waste generated by ultrasonic
equipment should be disposed of in the right way, and the equipment should be
discarded when it reaches its service life. Reusable materials can be recycled
by qualified waste firms to reduce environmental pollution. Please consult our
service organization for matters concerned or dispose of properly according to
the local waste collection methods.
Chapter 2 Product Description
2.1 Scope of application
The product applies to showing the body temperature of the subject by
measuring the thermal radiation on the forehead.
2.2Contraindication
N/A
2.3 Structural composition
The infrared forehead thermometer is composed of a sensor, a button, a display
screen, a battery cover and a body.
2.4 Product photo
2.5 Display screen
1. Object mode symbol | 5. Unit symbol |
---|---|
2. Temperature display value | 6. Mnemonic symbol |
3. Sound switch symbol | 7. Low battery symbol |
4. Body mode symbol | 8. Decimal point |
2.6 Key parts
NO | Component name | Specification and model |
---|---|---|
1 | Sensor | HMS-K1C1 |
2 | MCU | HY11P54 |
2.7 Packing list
Names of articles | Quantity |
---|---|
Instruction manual, including warranty card and certificate of conformity | 1 |
2 batteries, AAA 1.5V | 1 |
Main engine | 1 |
※ Product packaging should contain the items described above. In case of any shortage, please contact Shenzhen LEPU Intelligent Medical Equipment Co., Ltd. or the agent distributor in time.
Chapter 3 Why the Infrared Forehead Thermometer is Required?
3.1 Quick
With the innovative infrared technology, it can quickly measure the body
temperature in non-contact mode.
3.2 Accurate and reliable
By measuring the heat energy emitted from the forehead and calculating the
body temperature accordingly, accurate readings can be obtained as long as it
is held within a range of 5cm when measuring.
3.3 Simple and easy to use The infrared forehead thermometer is about
inductive measurement. It can easily measure body temperature, even for
sleeping children.Compared with the rectal thermometer, a non-contact frontal
thermometer can reduce the discomfort of children, and it is simpler and more
practical than other thermometers in use.
3.4 Safe and hygienic
Contactless measurement can prevent the spread of bacteria;
It is absolutely safe for children and adults;
Chapter 4 Product Installation and Use
4.1Check
Please check the packing case carefully before unpacking. In case of any
damage is found, please contact the carrier immediately. Open the package
correctly, take out the infrared forehead thermometer and other components
from the case with care, and check them one by one against the packing list.
When the equipment is moved a different environment, the difference in
temperature or humidity may lead to condensation to it, in which case no use
is allowed before condensation disappears.
4.2Install or replace batteries
The first step after unpacking is to install the battery. The battery holder
is on the back of the infrared forehead thermometer. The battery installation
steps are as follows: (1) Open the battery cover. (2) Insert the battery and
keep +, – poles of the battery in line with +, – poles of the battery holder.
(3) Close the battery cover.
Note: Please use 2 AAA alkaline batteries. Please use batteries if the same type, do not mix used and new batteries. Remove the battery when the thermometer will not be used for a long time. The symbol “ ” appearing on the screen indicates that the battery runs out; please replace the battery before measuring.
4.3 Starting up
Press the “ON/OFF” button, then the backlight of the display comes on, the LCD
is shown in full screen and displays the latest group of memory values, the
equipment enters the waiting state for measurement; at this time, the
backlight goes out, the LED on the forehead keeps flashing, and the equipment
gets ready for measurement. If there is no operation for about 60 seconds, the
equipment will shut down automatically.
4.4 Placement
Place the thermometer between the eyebrows, at a distance within 5cm from the
center of the forehead; In non-contact mode, the blue pilot light will point
to the area you are aiming at. If the eyebrow area is covered by hair, sweat
or dirt, please clean it in advance to improve the accuracy of the reading.
Keep the thermometer and forehead still when
measuring, as movement will negatively affect the temperature reading.
4.5 Mode switch
Toggle the slide switch to select the temperature measurement/calibration mode
4.6 Sound switch settings
In shutdown mode, press and hold the button “Memory”, it first displays “—-
M”, which flashes along with the symbol “M”; press and hold the memory button
for more than 4 seconds, it displays “—- C + human icon + sound icon”; if the
sound is on currently, a “bi” will be heard when entering this mode.
Note:
The equipment will shut down automatically if there is no button press for 60
seconds. It defaults to the current setting after rebooting, while it will
save previous settings in case of power failure.
4.7 Body temperature measurement
Toggle the slide switch to the body mode, press the button “ON/OFF
measurement” to turn on the forehead thermometer, align the thermometer sensor
to the position between eyebrows within 5cm from the forehead, and press the
button “ON/OFF measurement”, at this time, the distance focusing light comes
on, quickly adjust the appropriate distance (preferably when the focusing
light spot is within the pea grain size), when a “bi” is heard about 1S later
(no “bi” heard if the sound is disabled), it means that the body temperature
has been measured, with the result displayed on the LCD screen.
4.8 Object temperature measurement
Toggle the slide switch to the calibration mode press the button “ON/OFF
measurement” to turn on the thermometer, align the thermometer sensor to the
object to be measured, and press the button “ON/OFF measurement”, at this
time, the distance focusing light comes on, quickly adjust the appropriate
distance (preferably when the focusing light spot is within the pea grain
size), when a “bi” is heard about 1S later (no “bi” heard if the sound is
disabled), it means that the surface temperature has been measured, with the
result displayed on the LCD screen.
Note:
When the measured temperature is <40.0℃, the LCD backlight is green; when the
measured temperature is ≥40.0℃, the LCD backlight is green red and gives a
warning with the sound of “Bi”, “Bi”, “Bi”.
4.9 Memory query
Press the button “Memory” to start up the thermometer, the screen displays “—-
M” and the symbol “M” flashes. Press the button “Memory” again to display the
number of memory groups + M icon, and about 1 second later, show the display
memory value + the symbol “M” flashes.
4.10 Memory deletion
In shutdown mode, press and hold the button “Memory”, the sound setting
appears 4 seconds later at first, and keep holding to clear the memory value 8
seconds later, at this time, it shows “CLR + M”, which flashes with the sound
of “Bi-Bi-Bi”. After clearing, the equipment will shut down automatically.
4.11 Low battery indicator
When the battery voltage is below 2.60V, only the low-voltage symbol will
appear after starting up, and it cannot be used for measurement before the
replacement of the battery.
Chapter 5 Special Instructions for Safe Use
You should know the normal body temperature of individuals when they are
healthy, which will help you to accurately judge whether they have a fever. To
get the normal body temperature, please take more measurements when they are
healthy. The normal temperature of children can be as high as 37.7°C or as low
as 36.1°C. Please confirm it with a standard electronic thermometer.
The human body can regulate the temperature to keep the normal body
temperature within a certain fluctuation range, up to 1°C within a day.
Besides, the internal temperature of the human body, i.e. the body core
temperature, is different from the surface temperature of the skin, so we
cannot simply define what temperature is “normal”, as the body temperature is
always in connection with to the measurement site. The level of body
temperature is also affected by ambient temperature, age, sleep time, hormonal
readiness and physical activity.
Note: Avoid taking the temperature until after 30 minutes has lapsed after 30 minutes in the room (the subject to be measured and the infrared forehead thermometer should be at the same ambient temperature for at least 30 minutes). Keep the infrared forehead thermometer and forehead still when measuring, do not move the thermometer before the last beep is heard. Do not take the baby’s temperature immediately after breastfeeding.
Wait a few minutes before taking the temperature after waking up. Do not eat,
drink or do any other physical activity before or during the temperature
measurement. If there is a hat on head, please take it off and wait for 10
minutes before taking the temperature. Please clean the dirt or hair on
forehead before taking the temperature. Front bangs may cause readings to
rise. Wait for 10 minutes before taking the temperature after cleaning the
forehead. Please take the temperature in strict accordance with the
instructions. Temperature readings may be affected by improper placement.
For the following circumstances, it is recommended to measure the same site
three times and take the highest one as the final.1) Newborn babies less than
100 days old. 2) Children under three years old, with low immunity and greatly
affected by fever on health. 3) The user learns to use the infrared forehead
thermometer for the first time, who has little knowledge about the operation
and fails to get stable readings.
If patients intend to take body temperature by their own, we recommend contact measurement.
Chapter 6 Calibration
The infrared forehead thermometer has been calibrated before delivery from the
factory. If you have any doubt about its accuracy, please contact the after-
sales service.
We recommend a technical inspection for the measurement every two years, and
it is required to comply with the applicable national regulations of the local
place. Technical inspection for the measurement may be carried out by the
government agencies in charge or by authorized fee-paying maintenance
services.
Chapter 7 Troubleshooting
The infrared forehead thermometer has been calibrated before delivery from the
factory. If you have any doubt about its accuracy, please contact the after-
sales service.
We recommend a technical inspection for the measurement every two years, and
it is required to comply with the applicable national regulations of the local
place. Technical inspection for the measurement may be carried out by the
government agencies in charge or by authorized fee-paying maintenance
services.
7.1 Measured temperature too high
Error message on the screen
| Meaning of show|
Possible causes and solutions
---|---|---
| The measured body temperature is too high (above43.0°C)| The result exceeds
the measurement range limit
| The measured object temperature is too high (above 100°C)| The result
exceeds the measurement range limit
7.2 Measured temperature too low
Error message on the screen
| Meaning of show|
Possible causes and solutions
---|---|---
| Measured body temperature is too low (below 32.0°C)| The result exceeds the
measurement range limit
| Measured object temperature is too low (below 0°C)| The result exceeds the
measurement range limit
7.3 Ambient temperature out of range
Error message on the screen
| Meaning of show|
Possible causes and solutions
---|---|---
| The ambient temperature is lower than 5°C| Move to another warm room
| The ambient temperature is higher than 40°C| Move to another cold room
7.4 Other error messages
Error message on the screen
| Meaning of show|
Possible causes and solutions
---|---|---
| Eeprom error| Please contact the after-sales service.
| Hardware error| Please contact the after-sales service.
Chapter 8 Cleaning and Maintenance
Clean with a dry soft cloth. If the engine is particularly dirty, wipe with a wet and wrung out cloth and then dry it with a dry cloth. Disinfect the product only when necessary; disinfection is recommended carried out by wiping the product surface twice with disinfectant. Do not soak or use gas for disinfection. Use of medical alcohol is recommended.
If necessary, clean the infrared forehead thermometer first when the repeated measurement of the body temperature is required. Clean the thermometer shell and sensor with an alcohol swab or cotton ball dipped in alcohol (70%), and be careful to avoid liquid entering the thermometer. No use of corrosive detergent, diluent or benzene solvent is allowed. Do not immerse the thermometer in water or other cleaning solutions of any kind.
Do not expose the equipment to high temperature, high humidity, dust, or direct sunlight.
Remove the battery when the thermometer will not be used for a long time.
Chapter 9 Warranty Service
The warranty period of the product is 1 year from the date of sale (see the
package for the manufacturing date).
Any damage caused by improper use, battery leakage, failure to operate in
accordance with the requirements, or transfer of the thermometer to other
users is out of warranty.
We will not provide free warranty service for the failure caused by the user’s reasons as follows:
Repair service beyond warranty will be charged accordingly.
(1) Failure caused by unauthorized disassembly or refitting of the product.
(2) Failure caused by dropping during use or handling.
(3) Failure due to lack of proper maintenance.
(4) Failure to operate in accordance with the correct instructions in the
operating manual
Failure is caused by failing to operate correctly as instructed by the
operation manual, etc.
After-sale service unit: Shenzhen LEPU Intelligent
Medical Equipment Co., Ltd.
Address of after-sale service unit: BLD 9, Baiwangxin High-Tech Industrial
Park, Songbai Road, Xili
Street, Nanshan District, Shenzhen
After-sale service phone: 400-830-9392
Chapter 10 Product Specification
10.1 Product safety specification
Parameter|
Specification
---|---
Shock protection type| Equipment supplied with internal power
Shock protection grade| Type BF
Operating mode| Continuous
Movement level| Transportable equipment
EMC group| Class B, Group 1
10.2 Product environment specification
Environment
|
Specification
---|---
Temperature range| Operating temperature: 15°C~40°C
Storage temperature: -20°C~+55°C
Humidity range| Working humidity: ≤ 85%, non-condensing
Storage humidity: <95%, noncondensing
Ambient air pressure range| Working air pressure: 70 kPa ~106 kPa
Storage air pressure: 70 kPa ~106 kPa
10.3 Product hardware specification
Parameter|
Specification
---|---
Product name| Infrared Forehead Thermometer
Product model| LFR30B
Size| About 168.5(L) mm x 36(W)mm x 48(H) mm
Weight| About 95g (including battery)
Display screen| Segmented LCD display screen
Power supply| d.c.3V (2 batteries, AAA)
Data Storage| Up to 99 sets of measurements can be stored
Product service life| 5 years
10.4 Product measurement specification
Parameter|
Specification
---|---
Measuring position| Forehead temperature (between eyebrows)
Measurement range| 32.0°C~42.9°C
Resolving power| 0.1°C
Unit| °C
Accuracy| ±0.3℃(34℃~42.9℃) ±0.4℃(32℃~33.9℃)
Chapter 11 Product Registration Information
Medical Device Registration Certificate No.:GDXZZ20192070111
Product Technical Requirement No.: GDXZZ20192070111
Production License No.: GDSYJXSCX No. 20173019
Type Approval Certificate for Measuring Instruments: CPA2019T129-44
Registrant/manufacturer: Shenzhen LEPU Intelligent Medical Equipment Co., Ltd.
Address of registrant/manufacturer: North Floor 3, BLD 9, Baiwangxin High-Tech
Industrial Park, Songbai Road, Xili Street, Nanshan District, Shenzhen
Manufacturing Date: See the package
Chapter 12 Guide of EMC
Guidance and manufacturer’s declaration – electromagnetic emissions
The product is intended for use in the electromagnetic environment specified
below. The customer or the user should assure that it is used in such an
environment.
Emissions Test| Compliance| Electromagnetic Environment – Guidance
RF emissions GB 4824| Group 1| The product uses RF energy only for its
internal function. Therefore, the emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Class B
Harmonic Emissions GB 17625.1| Not applicable| The product is suitable for use
in all establishments, including domestic establishments and those directly
connected to the public low voltage power supply network that supplies
buildings used for domestic purposes.
Voltage fluctuations/flicker emissions GB 17625.2| Not applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
This product is intended for use in the electromagnetic environment specified below. The customer or the user should assure that it is used in such an environment.
Immunity Test
| IEC 60601 Testing electrical level| Compliance level|
Electromagnetic Environment guidance
Electrostatic discharge GB/T 17626.2| ±6 kV contact ±8 kV air| ±6 kV contact
±8 kV air
Not applicable| Floors are recommended to be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the relative humidity should be
at least 30%.
Electrical fast transient pulse trainGB/T 17626.4| ±2kV for power supply lines
±1kV for input/output lines| Not applicable
Surge GB/T 17626.5t| ±1 kV differential mode
±2 kV common mode| Not applicable| Not applicable
Voltage dips, short interruptions and voltage variations on power supply input
lines GB/T 17626.11| <5 % UT, for 0.5 cycle (>95% dip in UT); 40 % UT for 5
cycles (60% dip in UT); 70 % UT for 25 cycles (30% dip in UT); <5% UT for 5s
(>95% dip in UT)| Not applicable| Not applicable
Power frequency magnetic field (50/60Hz) GB/T 17626.8t| 3A/m| 3A/m| Power
frequency magnetic fields should be at levels characteristic of a typical
location in a typical commercial or hospital environment.
Note: UT is the AC. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The product is intended for use in the electromagnetic environment specified
below. The customer or the user should assure that it is used in such an
environment.
Immunity Test
| IEC 60601 test level| Compliance level|
Electromagnetic Environment guidance
---|---|---|---
Conducted RF GB/T 17626.6| 3 V (effective value)
150 kHz~80 MHz| Not applicable| Portable and mobile RF communications
equipment shall not be closer to any part of the product than the recommended
separation distance in use, including the cable. calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
RF radiation GB/T 17626.3| 3 V/m 80 MHz~2.5 GHz| 3V/m| d=1.2 Q
d= 1.2 Q80MHz~800MHz
d=2.3 Q 800MHz~2.5GHzWhere:
P – Maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer;
d – the recommended operation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey c should be less than the compliance level in each
frequency ranged.
Interference may occur in the vicinity of equipment marked with the following
symbol.
---|---|---|---
Note 1: At 80MHz and 800MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
a. Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcasts cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field intensity of the product is higher than the applied coincident
RF level above, the product shall be observed to verify its normal operation.
If abnormal performance is observed, additional measures may be necessary,
such as re-orienting or relocating the product.
b. Over the frequency range 150KHz to 80MHz, field strengths should be less
than 3/Vm.
The product is intended for use in the electromagnetic environment controlled
by RF radiation disturbance. The customer or the user of this product may
prevent electromagnetic interference through the minimum distance between the
portable and mobile RF communications equipment (transmitter) and the product
recommended below according to the maximum output power of the communications
equipment.
Maximum output power rating of the transmitter/W| 150 kHz~80 MHz d= 1.2 Q|
80MHz~800MHz d= 1.2 Q| 800MHz~2.5GHz d= 2.3 Q
0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
The maximum output power rating of the transmitter and the recommended
separation distance d in meters (m) not listed in above table may be
determined by the equation in the corresponding transmitter frequency column,
where, P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Certificate of Compliance
Product name: Infrared Thermometer
Product model: LFR30B
Manufacturing date:
Checker:
Registrant/manufacturer: Shenzhen LEPU
Intelligent Medical Equipment Co., Ltd.
Address: North Floor 3, BLD 9, Baiwangxin
High-Tech Industrial Park, Songbai Road,
Xili Street, Nanshan District, Shenzhen
Warranty Card
Product model: Body number:
Date of purchase: Customer name:
Store name:Tel:
Address:
Seal of store
Notice on warranty:
- High-quality after-sales service can protect your legitimate rights and interests.
- For your convenience in accepting maintenance services, please keep the warranty card properly.
- The warranty period of the product is one year.
- The following circumstances are beyond warranty:
a. No warranty card or expiration of the warranty period.
b. Failure caused by unauthorized disassembly or refitting of the product.
c. Failure caused by dropping during use or handling.
d. Failure due to lack of proper maintenance.
e. Failure to operate in accordance with the correct instructions in the operating manual
Failure is caused by failing to operate correctly as instructed by the operation manual, etc. Shenzhen LEPU Intelligent Medical Equipment Co., Ltd.