ebro TLC 1598 Precision Thermometer User Manual
- June 5, 2024
- ebro
Table of Contents
Precision thermometer
TLC 1598
Please read this manual carefully before operating your new general purpose
thermometer.
The operating manual introduces you with clear and simple instructions to the
handling of the thermometer.
Information that are useful and important for understanding the function, are
highlighted in the instructions text.
On behalf of a safe handling of the thermometer, follow the safety notes
marked with the character.
Safety notes
Caution when measuring high temperatures! Do not touch the temperature sensor with bare hands immediately after measuring! Risk of burn injury! As a basic principle, do not put the thermometer in the hot zone during measuring. Risk of explosion of the lithium battery!
Do not use the thermometer in explosion endangered areas!
Description
The general purpose thermometer is a handy measuring instrument operated with
a lithium battery, used to measure the temperature of air, liquids and plastic
material (meat, fruit, etc.) The electronics are controlled by a
microprocessor.
This ensures a high measuring precision and linearity over the complete
measuring range.
The thermometer is equipped with a stainless steel tilting sensor needle.
1………….. display (LCD)
2………….. button ON/OFF
3………….. tilting sensor needle
Extent of supply
Check contents of the packing for completeness and integrity. If you discover any damage or have any reason for a complaint, please contact you distributor or:
Xylem Analytics Germany Sales GmbH & Co. KG
ebro Am Achalaich 11 82362 Weilheim Germany
Phone: +49.(0)841.954.78.0
Fax: +49.(0)841.954.78.80
Internet: www.ebro.com E-Mail:
ebro@xylem.com
Parts list:
- thermometer
- calibration certificate
- for accessories see appendix I
Operation
Switching on the thermometer
To switch the thermometer on, press the “ON/OFF” button for at least one
second.
The device will perform a system test during which all segments are visible in
the display.
After this the measuring channel is tested. The display shows “38.8”.
Afterwards the first measured value will appear in °C.
If an error message is displayed, refer to chapter 4 “Troubleshooting” on page
10.
Temperature measuring
For conserving battery power, the thermometer will switch off automatically after 2 hours of operation.
Tilt the sensor needle out and place it at the desired position.
Use the probe only as directed, in order to avoid damages and faulty
measurements.
When the measurement is finished, switch the device off by pressing “ON/OFF”.
Troubleshooting
| Possible cause| Remedy
---|---|---
Measuring range exceeded Probe defective| Consider measuring rang e
Call service
Below measuring range
Short circuit of the probe| Consider measuring range
Call service
Calibration services
In order to guarantee a high measuring precision, the thermometer must be calibrated every year. Therefore we offer you a calibration service.
- Fill in the enclosed service card.
- After one year we will recall your thermometer for calibration.
Send us your devices and we will send them after calibration back.
Appendix
Accessories
Description | Identifier |
---|---|
Artificial leather case | EB 5 L |
Plastic box, blue | EB 1598 |
Technical data
Measuring range | 50 to +200 °C |
---|---|
Sensor | Pt 1000 |
Operating temperature | 0 to +50 °C |
Storage temperature | -10 to +60 °C |
Resolution | 0,1 °C |
Measuring precision | ±0,2 °C ±1 Digit |
Display test | 38.8 ± 0.1 |
Thermal time constant T99:
moved water| approx. 8 s
Battery| Lithium 3.6 V/ 0.4 Ah
Life time of battery| typically 4 years
Power off| automatically after2 hours
Dimensions| (L x W x H)
Housing| 158 x 44 x 20 mm
Sensor needle| 106 mm, Ø 3 mm
Housing material| ABS
Safety class| IP 54
Weight| approx. 70 g
Service-Adresse / Service Address / Adress du Service
Xylem Analytics Germany Sales GmbH & Co. KG
ebro Am Achalaich 11 82362 Weilheim Germany
Phone: +49.(0)841.954.78.0
Fax: +49.(0)841.954.78.80
Internet: www.ebro.com E-Mail:
ebro@xylem.com
…………………………………………………………………………………….
Hersteller
Xylem Analytics Germany GmbH
Am Achalaich 11
82362 Weilheim
Germany
References