InTemp CX450 Temp or Relative Humidity Data Logger Instruction Manual
- June 5, 2024
- InTemp
Table of Contents
- InTemp CX450 Temp or Relative Humidity Data Logger
- Specifications
- InTemp CX450 Temp/RH Logger
- Logger Components and Operation
- Getting Started
- Logger Alarms
- Minimum and Maximum Values
- Passkey Protection
- Downloading the Logger
- Logger Events
- Deploying and Protecting the Logger
- Battery Information
- Federal Communication Commission Interference Statement
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
InTemp CX450 Temp or Relative Humidity Data Logger
The InTemp CX450 logger measures ambient temperature and relative humidity (RH) for storage and transportation monitoring in the pharmaceutical, life science, and medical industries. This Bluetooth® Low Energy-enabled logger is designed for wireless communication with a mobile device. Using the InTemp app, you can easily configure the logger, download reports, and monitor tripped alarms. Or you can use InTempConnect® to configure and download the logger via the CX5000 Gateway. InTempVerify™ app is also available to easily download loggers and automatically upload reports to InTempConnect. Use the built-in LCD screen on the logger to view the current temperature and humidity, logging status and battery usage, and to check and clear maximum and minimum readings. Once data is uploaded to InTempConnect, you can track logger configurations and automatically upload logger data to build custom reports for further analysis.
Specifications
InTemp CX450 Temp/RH Logger
Included Items:
- Two AAA 1.5 V alkaline batteries
- Battery door and screw
- NIST Certificate of Calibration
Required Items:
- InTemp app
- Device with iOS or Android™ and Bluetooth
Logger Components and Operation
- Start Button: Press this button for 3 seconds to start the logger when it is configured to start “on button push.” You can also press this button for 1 second to bring the logger to the top of the list in the app. Press and hold both buttons on the logger simultaneously for 3 seconds to clear the minimum and maximum values on the LCD (see Minimum and Maximum Values) or for 10 seconds to reset a passkey (see Passkey Protection).
- Next or Mute Button: Press this button for 1 second to switch between the temperature and humidity readings on the LCD. You can also press this button for 1 second to mute a beeping alarm (see Logger Alarms). Press and hold both buttons on the logger simultaneously for 3 seconds to clear the minimum and maximum values (see Minimum and Maximum Values).
- Sensor Housing: Both the temperature and RH sensors are located within the housing that protrudes from the case.
- Audible Alarm Speaker: This is the speaker for the audible alarm that beeps when an alarm is tripped. See Logger Alarms.
- Alarm LED: This LED blinks every 5 seconds when an alarm is tripped. See Logger Alarms.
- Magnets: Use the four magnets on the back for mounting the logger.
- Vent: This is a membrane-covered vent for proper sensor operation. Do not remove the protective membrane.
- LCD: This screen shows the latest sensor readings and other status information. The LCD screen refreshes at the same rate as the logging interval. The example shows all symbols illuminated on the LCD screen followed by a table with descriptions of each symbol.
Note: If the logger has stopped logging because the memory is full, the LCD screen will remain on with “STOP” displayed until the logger is downloaded to your mobile device. Once the logger has been downloaded, the LCD will turn off automatically after 2 hours. The LCD will turn back on the next time the logger connects with your device.
Getting Started
InTempConnect is web-based software where you can monitor CX450 logger
configurations and view downloaded data online. Using the InTemp app, you can
configure the logger with your phone or tablet and then download reports,
which are saved in the app and automatically uploaded to InTempConnect. Or
anyone can download a logger using the InTempVerify app if the loggers are
enabled to be used with InTempVerify. See
www.intempconnect.com/help for details on
both the gateway and InTempVerify. If you do not need to access logged data
via the cloud based InTempConnect software, then you also have the option to
use the logger with the InTemp app only
Follow these steps to begin using the logger:
-
Set up an InTempConnect account. Follow all steps if you are a new administrator. If you already have an account and roles assigned, follow step c. If you are using the logger with the InTemp app only, skip to step 2.
-
Go to www.intempconnect.com and follow the prompts to set up an administrator account. You’ll receive an email to activate the account.
-
Log into www.intempconnect.com and add roles for the users you will be adding to the account. Click Settings and then Roles. Click Add Role, enter a description, select the privileges for the role and click Save.
-
Click Settings and then Users to add users to your account. Click Add User and enter the email address and first and last name of the user. Select the roles for the user and click Save. New users will receive an email to activate their user accounts.
-
Set up the logger. Insert two AAA batteries in the logger, observing polarity. Insert the battery door in the back of the logger making sure it is flush with the rest of the logger case. Use the included screw and a Phillips-head screwdriver to screw the battery door into place.
-
Download the InTemp app and log in.
-
Download InTemp to a phone or tablet from the App Store® or Google Play™.
-
Open the app and enable Bluetooth in the device settings if prompted.
-
InTempConnect users: Log in with your InTempConnect account email and password from the InTempConnect User screen.
InTemp app only users: Swipe left to the Standalone User screen and tap Create Account. Fill in the fields to create an account and then log in from the Standalone User screen. -
Set up a logger profile.
InTempConnect users (requires privileges): -
In InTempConnect, select Loggers > Logger Profiles.
-
Click Add Logger Profile.
-
Type a profile name.
-
If you plan to use the logger with the InTempVerify app, then select InTempVerify enabled.
-
Select CX450 for the Logger Family.
-
Select a Logging Interval, Start option, and any alarms.
Tap Save to add the new profile. Note: Set up any optional trip information fields before continuing (see www.intempconnect.com/help for details). -
InTemp app only users:
-
In the app, tap the Settings icon and tap CX450 Logger.
-
Tap the plus in the upper right corner.
-
Type a profile name.
-
Tap Logging Interval. Select a logging interval and tap Save.
-
Tap Start and select the option for when you want logging to begin.
-
Set up alarms if desired.
-
Tap Save to add the new profile.
-
Configure the logger.
-
Tap the Devices icon in the app. Find the logger in the list and tap it to connect to it.
-
Once connected, tap Configure. Swipe left and right to select a logger profile. Type a name or label for the logger. Tap Start to load the selected profile to the logger. InTempConnect users: If the profile does not yet appear in the app, close and then reopen it, and repeat steps a and b. In addition, if trip information fields were set up, you will be prompted to enter additional information. Tap Start in the upper right corner when done.
Note: You can also use the CX5000 Gateway to automatically configure loggers. See www.intempconnect.com/help for details.
If you are having trouble connecting:
* Make sure the logger is within range of your mobile device. The range for successful wireless communication is approximately 30.5 m (100 ft) with full line-of-sight.
* If your device can connect to the logger intermittently or loses its connection, move closer to the logger, within sight if possible.
* Change the orientation of your phone or tablet to ensure the antenna in your device is pointed toward the logger. Obstacles between the antenna in the device and the logger may result in intermittent connections
* If the logger appears in the list, but you cannot connect to it, close the app, power down the mobile device, and then turn it back on. This forces the previous Bluetooth connection to close.
- Deploy and start the logger. Place the logger in the location where you will be monitoring the temperature and humidity. Logging will begin based on the settings in the profile selected.
Once logging begins, press the Next button on the logger to switch between the latest temperature and humidity readings, as well as the minimum and maximum values for both (see Minimum and Maximum Values for more details). See Downloading the Logger for details on how to access the logged data.
Logger Alarms
You can set up an alarm to trip on the logger when a temperature and/or
humidity reading falls above or below a specified value or when the logger
battery is low. Specific alarm settings are set in the logger profile that you
create in either InTempConnect or in the app.
When a temperature or humidity alarm trips:
- The logger LED will blink every 5 seconds.
- The alarm icon will appear on the LCD and in the app.
- If audible alarms are enabled in the logger profile, the logger will beep as follows every 15 seconds:
- One quick beep for temperature alarms.
- Two quick beeps for RH alarms.
- One long beep if multiple alarms were tripped at the same time.
- An Alarm Tripped event is logged.
When a low battery alarm trips:
- The battery icon on the LCD will flash.
- The logger will quickly beep three times every 15 seconds.
- A Low Battery event is logged.
To mute a beeping alarm, press the Mute button on the logger. Once muted, you
cannot turn the beeping back on. Note that if a temperature and/or humidity
alarm is occurring at the same time as a low battery alarm, pressing the mute
button will silence all alarms.
Download the logger to view details about the tripped alarm and to clear
temperature anda humidity alarm indicators in the app and on the LCD. Replace
the batteries in the logger to clear a battery alarm. Note: Download the
logger before replacing the batteries to ensure no data is lost.
Minimum and Maximum Values
The logger LCD displays the minimum and maximum temperature and humidity
readings for the entire logging period. These values reset automatically when
the logger is downloaded and restarted or stopped and reconfigured.
You can also clear these values as needed while the logger is logging by
pressing both buttons on the logger simultaneously for 3 seconds until HOLD
disappears on the LCD. Dashes (–) will then appear on the LCD for the minimum
and maximum values until the next logging interval. The values will then
continue to be updated for the remaining logging period or until they are
cleared again. Note: This clears the data on the screen only. Actual logger
and reporting data will not be cleared with this reset.
Passkey Protection
The logger is protected by an encrypted passkey automatically generated by the InTemp app for InTempConnect users and optionally available if you are using the InTemp app only. The passkey uses a proprietary encryption algorithm that changes with every connection.
InTempConnect Users
Only InTempConnect users belonging to the same InTempConnect account can
connect to a logger once it is configured. When an InTempConnect user first
configures a logger, it is locked with an encrypted passkey that is
automatically generated by the InTemp app. After the logger is configured,
only active users associated with that account will be able to connect to it.
If a user belongs to a different account, that user will not be able to
connect to the logger with the InTemp app, which will display an invalid
passkey message. Administrators or users with the required privileges can also
view the passkey from the device configuration page in InTempConnect and share
them if needed. See
www.intempconnect.com/help for more
details. Note: This does not apply to InTempVerify. if the logger was
configured with a logger profile in which InTempVerify was enabled, then
anyone can download the logger with the InTempVerify app.
InTemp App Only Users
If you are using the InTemp app only (not logging in as an InTempConnect
user), you can create an encrypted passkey for the logger that will be
required if another phone or tablet attempts to connect to it. This is
recommended to ensure that a deployed logger is not mistakenly stopped or
purposely altered by others.
To set a passkey:
- Tap the Devices icon and connect to the logger.
- Tap Set Logger Passkey.
- Type a passkey up to 10 characters.
- Tap Save.
- Tap Disconnect.
Only the phone or tablet used to set the passkey can then connect to the logger without entering a passkey; all other mobile devices will be required to enter the passkey. For example, if you set the passkey for the logger with your tablet and then try to connect to the device later with your phone, you will be required to enter the passkey on the phone but not with your tablet. Similarly, if others attempt to connect to the logger with different devices, then they would also be required to enter the passkey. To reset a passkey, simultaneously press both the top and bottom button on the logger for 10 seconds, or connect to the logger in the app, tap Set Logger Passkey, and select Reset Passkey to Factory Default.
Downloading the Logger
You can download the logger to a phone or tablet and generate reports that
include temperature and humidity readings, events, user activity, alarm
information, and more. Reports can be shared immediately upon download or
accessed later in the InTemp app.
InTempConnect users: Privileges are required to download, preview, and share
reports in the InTemp app. Report data is automatically uploaded to
InTempConnect when you download the logger. Log into InTempConnect to build
custom reports
(requires privileges).
Note: InTempConnect users can also download CX loggers automatically on a
regular basis using the CX5000 Gateway. Or, if the logger was configured with
a logger profile in which InTempVerify was enabled, then anyone can download
the logger with the InTempVerify app. For details on the gateway and
InTempVerify, see www.intempconnect/help.
To download the logger with the InTemp app:
- Tap the Devices icon and connect to the logger.
- Tap Download.
- Choose a download option:
- Download & Continue. The logger will continue logging once the download is complete.
- Download & Restart. The logger will start a new data set using the same profile once the download is complete.
- Download & Stop. The logger will stop logging once the download is complete.
To download multiple loggers with the InTemp app:
- Tap Devices, then Bulk Download.
- The screen changes to Bulk Download mode. This changes the way the screen behaves when you tap on a logger tile. Tap on a tile to select it for bulk download. You can select up to 20 loggers. Text at the bottom of the page updates to indicate how many loggers are selected.
- Tap Download X Loggers to begin the download.
- Download & Continue. The loggers continue logging once the download is complete.
- Download & Restart (CX400, CX450, CX503, CX603, and CX703 models only). The logger starts a new configuration using the same profile once the download is complete. You will be prompted to enter trip information again (if applicable). Note that if the logger profile is set up to start with a button push, you must push the button on the logger for logging to restart.
- Download & Stop. The logger stops logging once the download is complete. The downloads begin and run one after another. The screen displays the download queue.
- Click Cancel to cancel the downloads and return to the Devices screen, not in Bulk Download mode.
- The screen displays Done when all the loggers are downloaded.
A report of the download is generated and is also uploaded to InTempConnect if
you are logged into the InTemp app with your InTempConnect user credentials.
If you are using the bulk download feature, one report per logger is
generated.
In the app, tap Settings to change the default report type and report sharing
options. The report is also available in Secure PDF and XLSX formats for
sharing later. Tap the Reports icon to access previously downloaded reports.
See www.intempconnect.com/help for
details on working with reports in both the InTemp app and InTempConnect.
Logger Events
The logger records the following events to track logger operation and status. These events are listed in reports downloaded from the logger.
Deploying and Protecting the Logger
Use the four magnets on the back of the logger case to mount it to a magnetic
surface.
The logger is designed for indoor use and can be permanently damaged by
corrosion if it gets wet. Protect it from direct water contact. If the logger
gets wet, remove the battery immediately and dry the circuit board.
Note: Static electricity may cause the logger to stop logging. The logger has
been tested to 8 KV but avoid electrostatic discharge by grounding yourself to
protect the logger. For more information, search for “static discharge” on
onsetcomp.com.
Battery Information
The logger requires two user-replaceable AAA 1.5 V alkaline or optional lithium batteries for operation at the extreme ends of the logger operating range. Expected battery life varies based on the ambient temperature where the logger is deployed, the frequency of connecting to the phone or tablet and downloading reports, the duration of audible alarms, and battery performance. New batteries typically last 1 year with logging intervals 1 minute or greater. Deployments in extremely cold or hot temperatures or a logging interval faster than 1 minute can impact battery life. Estimates are not guaranteed due to uncertainties in initial battery conditions and operating environment.
Note: Make sure the batteries installed have flat negative terminals. There should be no indent in the bottom of the batteries. Batteries with indents in the negative terminals may become loose and prevent proper operation.
To install or replace the batteries:
- Download the logger before replacing the batteries to ensure no data is lost.
- If the battery door is already installed on the back of the logger, use a Phillips-head screwdriver to remove it.
- Remove any old batteries.
- Insert two new batteries observing polarity.
- Screw the battery door into place.
WARNING: Do not cut open, incinerate, heat above 85°C (185°F), or recharge lithium batteries. The batteries may explode if the logger is exposed to extreme heat or conditions that could damage or destroy the battery case. Do not dispose of the logger or batteries in fire. Do not expose the contents of the batteries to water. Dispose of the batteries according to local regulations for lithium batteries.
Federal Communication Commission Interference Statement
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one of the following measures:
- Reorient or relocate the receiving antenna
- Increase the separation between the equipment and receiver
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected
- Consult the dealer or an experienced radio/TV technician for help
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
FCC Caution: Any changes or modifications not expressly approved by the party
responsible for compliance could void the user’s authority to operate this
equipment.
Industry Canada Statements
This device complies with Industry Canada license-exempt RSS standard(s).
Operation is subject to the following two conditions: (1) this device may not
cause interference, and (2) this device must accept any interference,
including interference that may cause undesired operation of the device.
To comply with FCC and Industry Canada RF radiation exposure limits for
general population, the logger must be installed to provide a separation
distance of at least 20cm from all persons and must not be co-located or
operating in conjunction with any other antenna or transmitter.
1-508-759-9500 (U.S. and International)
1-800-LOGGERS (564-4377) (U.S. only)
www.onsetcomp.com/intemp/contact/support
© 2020–2022 Onset Computer Corporation. All rights reserved. Onset, InTemp,
InTempConnect, and InTempVerify are trademarks or registered trademarks of
Onset Computer Corporation. App Store is a service mark of Apple Inc. Google
Play is a trademark of Google Inc. Bluetooth is a registered trademark of
Bluetooth SIG, Inc. Bluetooth and Bluetooth Smart are registered trademarks of
Bluetooth SIG, Inc. All other trademarks are the property of their respective
companies.
Patent #: 8,860,569 25186-F
References
- Onset HOBO and InTemp Data Loggers
- InTempConnect
- intempconnect.com/help
- Onset HOBO and InTemp Data Loggers
- InTemp Technical Support | Onset's HOBO and InTemp Data Loggers