ADC Diagnostix Otoscope Coax Ophthalmoscope Dermascope Throat Illuminator Power Handles Instruction Manual
- June 10, 2024
- ADC
Table of Contents
- A SPECIAL THANK YOU
- INTENDED USE
- **** WARNINGS AND PRECAUTIONS
- SYMBOL DEFINITIONS
- BATTERY HANDLES AND INITIAL USE
- DIAGNOSTIX OTOSCOPE
- DIAGNOSTIX OPHTHALMOSCOPE
- DIAGNOSTIX DERMASCOPE
- DIAGNOSTIX THROAT ILLUMINATOR
- REPLACING THE LAMP
- INSTRUMENT HEAD COMPATIBILITY
- SPARE PARTS AND ACCESSORIES
- CLEANING AND DISINFECTION
- TECHNICAL SPECIFICATIONS
- MAINTENANCE
- ELECTROMAGNETIC COMPATIBILITY
- WARRANTY
- QUALITY STANDARDS
- HOW TO CONTACT US
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Diagnostix™
Otoscope
Coax Ophthalmoscope
Dermascope
Throat Illuminator
Power Handles
Instruction Manual
Questions?
Call ADC toll free:
1-800-232-2670
A SPECIAL THANK YOU
Congratulations on your purchase of an ADC® Diagnostix™ physical exam
instrument.
ADC professional diagnostic products are the instruments of choice where
accuracy and dependability are critical. This feature-rich instrument was
designed to simplify physical exams and non-invasive diagnostics. With proper
use and care these instruments will provide many years of dependable service.
Read this booklet thoroughly before using your new instrument.
INTENDED USE
These instruments are designed to facilitate examination of the eye, ear,
nose, throat, and skin.
If you have any questions call our toll-free number or visit our website.
Note: Only use ADC parts and accessories to ensure safe and functional
use of this product.
**** WARNINGS AND PRECAUTIONS
ADC Diagnostix instruments have been manufactured to the highest global
standards and are subjected to rigorous quality control. Read these
instructions for use carefully before putting the unit into operation and keep
them in a safe place. If you should have any questions, call our toll-free
number or visit our website. Our address can be found on the last page of this
booklet. Please note that all instruments described in these instructions for
use are only to be used by suitably trained personnel. The performance and
efficacy of these instruments are only guaranteed when genuine ADC parts and
accessories are used.
Warning: Do not use batteries, electrical cords, or replacement parts
other than those included with this product or supplied by the manufacturer.
Warning: Because prolonged intense light exposure can damage the retina,
the use of the device for ocular examination should not be unnecessarily
prolonged, and the brightness setting should not exceed what is needed to
provide clear visualization of the target structures. This device should be
used with filters that eliminate UV radiation (<400 nm) and, whenever
possible, filters that eliminate short-wavelength blue light (<420 nm).
The retinal exposure dose for a photochemical hazard is a product of the
radiance and the exposure time. If the value of radiance were reduced in half,
twice the time would be needed to reach the maximum exposure limit.
While no acute optical radiation hazards have been identified for direct or
indirect ophthalmoscopes, it is recommended that the intensity of light
directed into the patient’s eye be limited to the minimum level necessary for
diagnosis. Infants, aphakes, and persons with diseased eyes will be at greater
risk. The risk may also be increased if the person being examined has had any
exposure with the same instrument or any other ophthalmic instrument using a
visible light source during the previous 24 hours. This will apply
particularly if the eye has been exposed to retinal photography.
Warning: Otoscope MUST be used with included disposable specula.
Caution: Not made with natural rubber latex.
Attention: There may be a risk of ignition when the instrument is used in
the presence of flammable or combustible gases. Work in areas with good
ventilation.
Attention: Make sure to charge wall plug-in handles for at least 12 hours
before first use and for all subsequent charges to ensure optimal capacity and
battery life. Never attempt to take the instrument apart! There is a danger of
life-threatening electrical shock. Unplug the instrument before cleaning or
when disinfecting.
NOTE: To obtain the greatest performance from your instrument, it is
recommended that the instrument be used within a temperature range of 50°F to
104°F (10°C to 40°C), with a 10-95% relative humidity.
SYMBOL DEFINITIONS
The following symbols are associated with your diagnostic instrument.
Symbol | Definition |
---|---|
Important Caution | |
Not made with natural rubber latex | |
Phthalate free | |
Consult instructions for use |
| Meets the general safety and performance requirements of Regulation (EU)
2017/745 of the European Union
| Medical device
| Authorized European Representative’s Information
| Manufacturer’s Information
Symbol| Definition
---|---
| Type BF applied part
| Do not dispose of this product as unsorted municipal waste. Prepare this
product for reuse or separate collection as specified by Directive 2002/96/EC
of the European Parlia- ment and the Council of the European Union on Waste
Elec- tronic and Electrical Equipment (WEEE). If this product is contaminated,
this directive does not apply.
|
Device double-earthed
| Keep Dry
| Do not use if the package is damaged
| Importer
| Catalog number
Symbol| Definition| | Symbol| Definition
---|---|---|---|---
| Exporter| | | Date of Manufacture
| Non-Sterile|
| Batch Code
| Unique Identifier| | | Follow instructions for use
BATTERY HANDLES AND INITIAL USE
5.1. Function
The ADC battery handles described in these instructions supply the instrument
heads with power (the lamps are contained in their respective instrument
heads).
5.2. Battery Handle Options
All the instrument heads described in these instructions fit on the following
battery handles and can therefore be individually combined.
For Otoscopes, Ophthalmoscopes, Dermascope, Throat Illuminator, Power
Handles
| |
---|---|---
Wall-Mounted Handle (with extension unit)| # 5660E| 3.5V, 230V or 120V
Standard Rechargeable Handle (requires desk charger)| # 5560| 3.5V
Plug-In Rechargeable Handle| # 5460| 3.5V, 230V or 120V
Note: These handles are compatible with ADC , Riester, and Welch Allyn 3.5V instrument heads.
- Welch Allyn is a registered trademark.
- Riester is a registered trademark
5.3. Inserting, Removing, and Charging Batteries
Screw off the battery cap on the lower part of the handle. Depending onwhich
handle you have purchased and for what voltage, insert the rechargeable
battery into the casing such that the positive end points toward the top of
the handle (Fig. 1). There is also an arrow next to the plus symbol on the
rechargeable battery that shows you the direction to insert into the handle.
Screw the battery cap onto the handle to secure.
Plug-in Rechargeable Handles
Prior to initial use, charge the plug-in handle for up to 24 hours in the
mains socket. To charge, remove top portion by unscrewing counter-clockwise to
reveal two-prong lug. Plug into electrical outlet.
Caution:
- If you do not plan to use the device for a long time, or if you travel with it, remove the rechargeable batteries from the handle.
- New batteries should be inserted once the light intensity of the instrument becomes weaker, even on a full charge.
- To obtain the best possible light output we recommend always fitting high-quality batteries.
- If you suspect that liquid or moisture could have entered the handle, do not charge under any circumstances. This could lead to a life-threatening electric shock, especially in the case of plug-in handles.
- To extend the service life of the battery, the battery should only be charged once the light intensity of the instrument has become weaker.
- Plug-in handle should be charged overnight (12 hours) to ensure uninterrupted power supply.
NOTE for #5460: 3.5V for charging in a 120V or 230V wall socket. When using the new 5460BAT, care must be taken to ensure that no insulation is affixed to the spring of the battery-handle cap. If the old 5460BAT is used, an insulation must be affixed to the spring to avoid short circuit.
Waste Disposal
Please note that rechargeable batteries must be disposed of as special waste.
You can obtain the relevant information from your local authority or from your
local environmental agency.
5.4. Fitting Instrument Heads
Fit the required instrument head on the receptacle on the upper part of
the handle such that the two recesses of the lower part of the instrument head
fit on the two rotruding guide studs on the battery handle. Press the
instrument head lightly on to the battery handle and twist the handle
clockwise until securely locked (Fig. 2). The head s removed by reversing the
process.
5.5 Locking Instrument Head
Instrument heads are equipped with a security feature. Use enclosed hex key
to lock head onto power handle. To secure, insert hex key into slot on head
and tighten by turning clockwise. Keep the hex key in a convenient place in
case the head must be removed for cleaning or service.
NOTE: Locking the head is recommended on wall-mounted systems only.
5.6 Adjusting the Light Intensity
With the ADC Adtronic™ electronic rheostat, it is possible to modulate the
light intensity. Smoothly control the applied power by holding the easy-grip
dial and turning either clockwise or counter-clockwise, varying the light
intensity stronger or weaker.
ATTENTION: At every switch-on of the battery handle the light intensity
is at 100%. An automatic safety switches the light off after 180 seconds.
Explanation of the symbol on the plug-in handle:
Caution: Observe the Operating Instructions!
DIAGNOSTIX OTOSCOPE
6.1. Purpose
The ADC Diagnostix Otoscope described in these instructions is produced for
illumination and examination of the auditory canal in combination with the ear
specula. ADC otoscopes are compatible with Riester and Welch Allyn specula.
6.2. Fitting and Removing Ear Specula
Screw the speculum clockwise until noticeable resistance is felt.
To remove the speculum, twist the speculum counter-clockwise.
6.3 Swivel Lens for Magnification
Standard Series: The swivel lens (3x max.) is fixed to the device and can be
swiveled 360°. (Fig. 3)
PMV Series: The focusing wheel enables you to adjust the focusing range in the
auditory canal (tympanic membrane). The adjustment wheel moves up and down to
focus the lens. (Fig. 4)
6.4 Insertion of External Instruments into the Ear
Standard Series only: If you wish to insert external instruments (e.g.,
tweezers) into the ear, you have to rotate the swivel lens located on the
otoscope head 180°.
6.5 Pneumatic Test
To perform the pneumatic test (examination of the eardrum), connect an
insufflator (sold separately, # 5122N). Once the tube for the insufflator is
attached to the connector port on the right side of the instrument head (Fig.
5), you can carefully insert the necessary volume of air into the ear canal.
6.6 Technical Data of the Lamp
Otoscope Hal/Xen 3.5V | 3.5V 720 mA | mean life span 15h |
---|---|---|
Otoscope LED 3.5V | 3.5V 300 mA | mean life span 10,000h |
DIAGNOSTIX OPHTHALMOSCOPE
7.1. Purpose
The ADC Diagnostix Ophthalmoscope described in these instructions is produced
for the examination of the eye..
2. Lens Wheel with Correction Lens
The correction lens can be adjusted on the lens wheel.
The following correction lenses are available:
Coax 3.5V Ophthalmoscope (Fig. 6)
Plus: 1-10, 12, 15, 20, 40
Minus: 1-10, 15, 20, 25, 30, 35
Coax Plus 3.5V Ophthalmoscope (Fig. 7)
Plus: 1-45 in single steps
Minus: 1-44 in single steps
The values can be read off in the illuminated field of view. Plus values are
displayed in green numbers, minus values with red numbers.
7.3. Apertures
The following apertures can be selected with the aperture hand wheel (Fig. 8).
Coax Ophthalmoscope
Half-moon, micro/small/large circular aperture, fixation star, and slit.
Coax Plus Ophthalmoscope
Half-moon, micro/small/large circular aperture, fixation star, slit, and grid.
Aperture | Function | Models |
---|
| Half Moon| For examinations with turbid lenses| Coax/Coax Plus
Micro Spot
Small Circle| Allows quick entry into small, undilated pupils
Excellent view of fundus through an undilated pupil
| Coax/Coax Plus
Coax/Coax Plus
Large Circle| For a dilated pupil and general examination| Coax/Coax Plus
Karo (Grid)| For topographic determination of retina changes| Coax Plus Only
Slit| To help determine levels of tumors and lesions| Coax/Coax Plus
Fixation star| Measuring eccentric fixation or locating lesions| Coax/Coax
Plus
7.4. Changing Filters
Using the filter wheel, the following filters can be switched for each
aperture.
Coax & Coax Plus Ophthalmoscope
Red-free filter, blue filter, and polarization filter.
Filter | Function |
---|---|
Red-free filter: | Contrast enhancing to assess fine vascular changes, e.g., |
retinal bleeding.
Polarizations filter:| For precise assessment of tissue colors and to minimize
retinal reflections.
Blue filter:| For improving recognition of vascular abnormalities or bleeding,
for fluorescence ophthalmology.
7.5. Focusing Device (Coax Plus)
Fast fine adjustment of the examination area to beobserved is achieved from
various distances by turning the focusing daisywheel. (Fig. 9)
7.6. Technical Data on the Lamp
Coax 3.5V ophthalmoscope: 3.5V / 690 mA / average service life 15h
Coax Plus 3.5V ophthalmoscope: 3.5V / 29 mA / average service life 100,000h
DIAGNOSTIX DERMASCOPE
8.1. Purpose
The Diagnostix Dermascope described in these instructions is produced for
early identification of changes of skin pigmentation (malignant melanomas).
8.2. Focusing
Focus the magnifying glass by rotating the eyepiece ring. (Fig. 10)
8.3. Contact Plates
Two contact plates are supplied:
- Without a scale.
- Including a scale of 0 – 10 mm for measuring melanotic skin changes, such as malignant melanoma.
8.4. Technical Data of Lamp
Dermascope LED: 3.5V 28 mA / mean life span 100,000h
DIAGNOSTIX THROAT ILLUMINATOR
9.1. Purpose
The throat illuminator described in these instructions is produced for
examination of the oral cavity and pharynx in combination with commercial
wooden and plastic blades.
9.2. Technical Data of Lamp
Illuminator Hal/Xen | 3.5V 720 mA | mean life span 15h |
---|---|---|
Illuminator LED | 3.5V 28 mA | mean life span 100,000h |
REPLACING THE LAMP
All Instrument Heads
Remove the instrument head from the battery handle. The lamp is located at the
base of the instrument head. Pull the lamp out of the instrument head with
thumb and forefinger or a suitable tool. Insert a new lamp.
*Use only ADC or Riester lamps.
Caution: The pin on the ophthalmoscope lamp must be inserted into the
guide groove on the instrument head.
INSTRUMENT HEAD COMPATIBILITY
All ADC 3.5V instrument heads are compatible with Riester and Welch Allyn power handles equipped with bayonet-style mount.
SPARE PARTS AND ACCESSORIES
For a complete list of our physical exam instrument spare parts and accessories, please visit our website at www.adctoday.com.
CLEANING AND DISINFECTION
Cleaning and disinfection of medical devices serves to protect the patient,
the user, and third parties, and to preserve the value of the medical devices.
Due to the product design and the materials used, no defined limit can be
specified for the maximum number of reprocessing cycles that can be carried
out. The life span of the medical devices is determined by their function and
by gentle handling of the devices. Defective products must undergo the
reprocessing procedure described before being returned for repair.
Cleaning and Disinfection
The instrument heads and handles can be cleaned externally with a moist cloth
until visually clean. Wipe disinfection as specified by the disinfectant
manufacturer. Only disinfectants with proven efficacy should be used, taking
into account the national requirements. After disinfection, wipe the
instrument down with a moist cloth to remove possible disinfectant residues.
The components that come into contact with the skin can be rubbed down with
alcohol or a suitable disinfectant.
Please Note!
-
Never immerse the instrument heads and handles in liquids!
Take care to ensure that no liquids get inside the casing! -
This item is not approved for automated reprocessing and sterilization.
Single-Use Ear Specula
For single Use Only
WARNING: Repeated use can cause infections.
TECHNICAL SPECIFICATIONS
Ambient Temperature: | 32°F to 104°F (0°C to 40°C) |
---|---|
Relative Humidity: | 30% to 70% non-condensing |
Transport and Storage Temperature: | 14°F to 131°F (-10°C to 55°C) |
Relative Humidity: | 10% to 95% non-condensing |
MAINTENANCE
These instruments and their accessories do not require any specific maintenance. Should an instrument have to be examined for any specific reason whatsoever, please return it to ADC.
ELECTROMAGNETIC COMPATIBILITY
Medical electrical equipment is subject to special precautionary measures with
regard to electromagnetic compatibility (EMC).
Portable and mobile high-frequency communication equipment can influence
medical electrical equipment. This ME device is intended for operation in an
electromagnetic environment as specified below. The user of the device should
ensure that it is operated in such an environment.
The ME device must not be used directly next to or arranged in a stack with
other devices. If the device has to be operated near to or in a stacked
arrangement with other devices, then the ME device should be monitored in
order to verify that it operates as intended in this arrangement. This ME
device is intended exclusively for use by professional medical staff. This
device can cause radio interference and can disrupt the operation of equipment
nearby. Suitable remedial measures, for instance realignment, rearrangement of
the ME device or, shielding, may become necessary.
Guidelines and manufacturer’s declaration – electromagnetic emissions
The Diagnostix instrument is intended for operation in an electromagnetic
environment as specified below. The customer or the user of the Diagnostix
instrument should ensure that it is used in such an environment.
Emission Measurements| Compliance| Electromagnetic
Environment Guidelines
HF emissions according to CISPR 11| Group 1| The Diagnostix instrument employs
HF energy solely for an internal function. Its HF emission is therefore very
low and it is unlikely that neighboring electronic devices will be affected by
interference.
HF emissions according to CISPR 11| Class B| The Diagnostix instrument is
intended for use in all facilities, including living quarters and such as are
directly connected to a public power supply that also supplies buildings that
are used for residential purposes.
Harmonics emissions according to EC61000-3-2| Not applicable
Voltage fluctuation / flicker emissions according to IEC61000-3-3| Not
applicable
Guidelines and manufacturer’s declaration –
electromagnetic emissions
The Diagnostix instrument is intended for operation in an electromagnetic
environment as specified below. The customer or the user of the Diagnostix
instrument should ensure that it is used in such an environment.
Immunity Tests| IEC 60601 Test Level|
Compliance Level| Electromagnetic Environment Guidelines
Electrostatic discharge (ESD) according to IEC61000-4-2| ± 6 kV contact
discharge ± 8 kV air discharge| ± 6 kV contact discharge ± 8 kV air discharge|
Floors should be made of wood or concrete or be covered with ceramic tiles. If
the floor is covered with a synthetic material, the relative air humidity must
be at least 30%.
Fast transient electrical interference/bursts according to IEC61000-4-4| ± 2
kV for power lines ± 1 kV for input and output lines| Not applicable| The
quality of the supply voltage should correspond to that of a typical business
or hospital environment.
Surges IEC61000-4-5| ± 1 kV voltage phase-to-phase ± 2 kV voltage phase-to-
earth| Not applicable| The quality of the supply voltage should correspond to
that of a typical business or hospital environment.
Voltage dips, short-time interruptions and fluctuations in the supply voltage
according to IEC61000-4-11| <5% UT (>95 % drop
in UT) for 0.5 cycles
40% UT (60 % drop in UT) for 5 cycles
70 % UT (30 % drop in UT) for 25 cycles
|
Not applicable
| The quality of the supply voltage should correspond to that of a typical
business or hospital environment.
| <5% UT (>95 % drop
in UT) for 5 s
| |
Magnetic field at the mains frequency (50Hz) according to IEC61000-4-8| 3 A/m|
3 A/m| If image disturbances occur, the Diagnostix instrument may have to be
placed further away from the sources of mains-frequency magnetic fields, or
magnetic shielding may have to be installed: the mains-frequency magnetic
field should be measured at the intended set-up site in order to ensure that
it is small enough.
Note – UT is the alternating supply voltage
prior to application of the test level.
Guidelines and manufacturer’s declaration –
electromagnetic immunity
This Diagnostix instrument model is intended for operation in the electromagnetic environment specified below.
The customer or the user of this Diagnostix instrument should ensure that it is used in such an environment.
Immunity Tests| IEC 60601 Test Level|
Compliance Level| Electromagnetic Environment Guidelines
Conducted HF interference according to IEC61000-4-6 Radiated HF rerference
cording to C61000-4-3| 3 Vrms 150 kHz to 80MHz 3 V/m 80 MHz to 2.5GHz| Not
applicable 10 V/m 3 V/m| Portable and mobile radio equipment should not be
used within a distance from the Diagnostix instrument, including cables, that
is less than the recommended safety distance as calculated by the equation
that is appropriate for the transmission frequency.
| | | Recommended safety distance: d = 1.2uP d = 1.2uP 80 MHz to 1000 MHz d
= 2.3uP 1400 MHz to 2.5 GHz
| | | Where P is the nominal power of the transmitter in Watts (W) as
specified by the manufacturer of the transmitter, and d is the recommended
safety distance in meters (m).
| | | The field strength of stationary radio transmitters should be less
than the compliance level at all frequencies as verified by an on- site testa
| | | Interference is possible in the vicinity of equipment marked with
the following symbol
---|---|---|---
Note 1: At 80 MHz and 800 MHz, the higher value applies.
Note 2: These guidelines may not apply in all situations. The
propagation of electromagnetic waves is influenced by reflection and
absorption by buildings, objects and people.
a. The field strength of stationary transmitters, such as base stations
of wireless telephones and mobile field radio services, amateur radio
stations, AM and FM radio and television transmitters cannot be precisely
determined theoretically in advance. In order to determine the electromagnetic
environment due to stationary HF transmitters, an investigation of the
location is advisable. If the field strength determined at the location of the
Diagnostix instrument exceeds the compliance level indicated above, then the
Diagnostix. instrument must be monitored with regard to its normal operation
at each place where it is used. If unusual performance characteristics are
observed, additional measures such as re-alignment of the Diagnostix
instrument or its removal to another place may be necessary.
b. In the frequency range of 150 kHz to 80 MHz, the field strength should
be smaller than 3 V/m.
Recommended safety distances between portable and
mobile HF communication devices and the Diagnotix
® **L**
This Diagnostix instrument is intended for operation in an electromagnetic
environment in which the radiated HF interference is monitored. The customer
or user of this Diagnostix instrument can help prevent electromagnetic
interference by observing minimum distances between portable and mobile HF
communication equipment (transmitters) and this Diagnostix instrument as
recommended below, depending on the maximum output power of the communication
equipment.
Nominal power of the transmitter W| Safety distance
that applies to the transmitter frequency m
150 kHz to 80 MHz
Not applicable
| 80 MHz to 1000 MHz d = 1.2 u P| 1400 MHz to
2.5GHz d = 2.3 u P
0.01| | 0.12| 0.23
0.01| | 0.38| 0.73
1| | 1.2| 2.3
10| | 3.8| 7.3
100| | 12| 23
For transmitters whose nominal power is not indicated in the table above, the
distance can be determined using the equation belonging to the respective
column, where P is the nominal power of the transmitter in Watts (W) as
specified by the manufacturer of the transmitter.
Note 1: At 80 MHz and 1400 MHz, the distance for the higher frequency
range applies.
Note 2: These guidelines may not apply in all situations. The propagation
of electromagnetic waves is influenced by reflection and absorption by
buildings, objects and people.
WARRANTY
This Diagnostix instrument is warranted for two years on instruments and
lifetime on LED lamps, from date of purchase. The warranty does not apply to
damage caused by improper handling, accidents, improper use, or alterations
made to the instrument by third parties. The warranty is only valid after the
product is registered online at https://www.adctoday.com/register.
For our European Customers
On request, send to ADC by e-mail
(info@adctoday.com), we can send to you this
manual on paper form within 7 calendar days at no additional cost to the user.
Our website, https://www.adctoday.com, where these instructions for use are
available fulfills the requirements of personal data protection, according to
Directive 95/46/EC and GDPR on the protection of individuals with regard to
the processing of personal data and on the free movement of such data. Any
serious incident that has occurred in relation to this medical device should
be reported to ADC and the competent authority of the Member State in which
the user and/or patient is established.
For Australian Consumers
Our goods come with guarantees that cannot be excluded under the Australian
Consumer Law. You are entitled to a replacement or refund for a major failure
and ompensation for any other reasonable foreseeable loss or damage. You are
also entitled to have the goods repaired or replaced if the goods fail to be
of acceptable quality and the failure does not amount to a major failure.
QUALITY STANDARDS
Device standard
This device is manufactured to meet the European and United States standards
for: ISO 62471, ISO 15004, ISO 10942, ISO 15004-1, and ISO 15004-2
Electromagnetic compatibility
Device fulfills the stipulations of the International standard IEC60601-1-2
HOW TO CONTACT US
To register your product and obtain further detailed user information about
our products and services visit us at:
www.adctoday.com
and follow the links.
For questions, comments, or suggestions call us toll free at:
1-800-232-2670 ADC
55 Commerce Drive
Hauppauge, NY 11788
U.S.A.
SC Cattus SRL
Str. Baneasa Nr. 8,
Targu-Mures, Jud. Mures
Romania, EU
IB p/n 93-5400-00 rev 5 (10/31/22)
Made in Germany
Inspected and Packaged in the U.S.A.
tel: 631-273-9600
toll free: 1-800-232-2670
fax: 631-273-9659
www.adctoday.com
email: info@adctoday.com
Printed in U.S.A.