TROTEC BB20 Layer Thickness Measuring Device User Manual
- June 12, 2024
- TROTEC
Table of Contents
BB20
OPERATING MANUAL
LAYER THICKNESS
MEASURING DEVICE
**Notes regarding the instructions
**
Symbols
Warning of electrical voltage
This symbol indicates dangers to the life and health of persons due to
electrical voltage.
Warning
This signal word indicates a hazard with an average risk level which, if not
avoided, can result in serious injury or death.
Caution
This signal word indicates a hazard with a low risk level which, if not
avoided, can result in minor or moderate injury.
Note
This signal word indicates important information (e.g. material damage), but
does not indicate hazards.
Info
Information marked with this symbol helps you to carry out your tasks quickly
and safely.
Follow the manual
Information marked with this symbol indicates that the instructions must be
observed.
You can download the current version of the instructions and the EU
declaration of conformity via the following link:
BB20| https://hub.trotec.com/?id=43284
---|---
Safety
Read this manual carefully before starting or using the device. Always store
the manual in the immediate vicinity of the device or its site of use.
Warning
Read all safety warnings and all instructions.
Failure to follow the warnings and instructions may result in electric shock,
fire and/or serious injury.
Save all warnings and instructions for future reference.
- Do not use the device in potentially explosive rooms or areas and do not install it there.
- Do not use the device in aggressive atmosphere.
- Protect the device from permanent direct sunlight.
- Do not remove any safety signs, stickers or labels from the device. Keep all safety signs, stickers and labels in legible condition.
- Do not open the device.
- Do not insert rechargeable batteries into the remote control.
- Never charge batteries that cannot be recharged.
- Different types of batteries and new and used batteries must not be used together.
- Insert the batteries into the battery compartment according to the correct polarity.
- Remove discharged batteries from the device. Batteries contain materials hazardous to the environment. Dispose of the batteries according to the national regulations.
- Remove the batteries from the remote control if you will not be using the device for a longer period of time.
- Never short-circuit the supply terminal in the battery compartment!
- Do not swallow batteries! If a battery is swallowed, it can cause severe internal burns within 2 hours! These burns can lead to death!
- If you think batteries might have been swallowed or otherwise entered the body, seek medical attention immediately!
- Keep new and used batteries and an open battery compartment away from children.
- If the battery compartment does not close securely, stop using the product via the remote control.
- Observe the storage and operating conditions (see Technical data).
Intended use
Only use the device for coating thickness measurements.
Observe and comply with the technical data.
To use the device for its intended use, only use accessories and spare parts
which have been approved by Trotec.
Foreseeable misuse
Do not use the device in potentially explosive atmospheres, for measurements
in liquids or at live parts. Trotec accepts no liability for damages resulting
from improper use. In such a case, any warranty claims will be voided. Any
unauthorised modifications, alterations or structural changes to the device
are forbidden.
Personnel qualifications
People who use this device must:
- have read and understood the instructions, especially the Safety chapter.
Residual risks
Warning
Risk of suffocation!
Do not leave the packaging lying around. Children may use it as a dangerous
toy.
Warning
The device is not a toy and does not belong in the hands of children.
Warning
Dangers can occur at the device when it is used by untrained people in an
unprofessional or improper way!
Observe the personnel qualifications!
Caution
Keep a sufficient distance from heat sources.
Note
To prevent damages to the device, do not expose it to extreme temperatures,
extreme humidity or moisture.
Note
Do not use abrasive cleaners or solvents to clean the device.
Information about the device
Device description
The device BB20 is used to determine the coating thickness on ferromagnetic
and non-ferromagnetic metal surfaces. The measuring device functions according
to the magnetic induction principle (for coatings on ferromagnetic surfaces)
or the eddy- current principle (for coatings on non-ferromagnetic surfaces).
The BB20 is a compact and versatile device suitable for quick, precise and
non-destructive measurements of coating thicknesses.
Device depiction
No. | Designation |
---|---|
1 | USB interface |
2 | Blue button (back / cancel) |
3 | ▼ button |
4 | ZERO button |
5 | Sensor |
6 | Battery compartment with cover |
7 | Power button |
8 | ▲ button |
9 | Red button (select / confirm) |
10 | Display |
Display
No. | Designation |
---|---|
11 | Working mode display |
12 | Battery indication |
13 | PC access display |
14 | NFe/Fe display |
15 | Indication of the unit |
16 | Layer thickness display |
17 | Statistical data display: AVG, MAX, MIN, SDEV |
18 | Number of measured values |
Technical data
Parameter | Value |
---|---|
Model | BB20 |
Sensor | F (ferrous metals) |
Functional principle | Magnetic induction |
Measuring range | 0 to 1250 pm |
0 to 49.21 mils | 0 to 1250 pm |
0 to 49.21 mils
Possible metallic substrates (examples)| Iron, steel| Copper, aluminium, zinc,
bronze and others
Guaranteed tolerance (of the measured value)| 0 to 850 pm (±3 % ±1 pm)
850 to 1250 pm (±5 %)
0 to 33.46 mils (±3 % ±0.039 mils)
33.46 to 49.21 mils (±5 %)| 0 to 1000 pm (±3 % ±1.5 pm)
850 to 1250 pm (±5 %)
0 to 33.46 mils (±3 % ±0.059 mils)
33.46 to 78.7 mils (±5 %)
Accuracy| Oto 50 pm (0.1 pm)
50 to 850 pm (1 pm)
850 to 1250 pm (0.01 mm)
0 to 1.968 mils (0.001 mils)
1.968 to 33.46 mils (0.01 mils)
33.46 to 49.21 mils (0.1 mils)| 0 to 50 pm (0.1 pm)
50 to 850 pm (1 pm)
850 to 1250 pm (0.01 mm)
0 to 1.968 mils (0.001 mils)
1.968 to 33.46 mils (0.01 mils)
33.46 to 49.21 mils (0.1 mils)
Minimum bending radius| 1.5 mm| 3 mm
Diameter of the smallest measuring surface| 7 mm| 5 mm
Critical primary layer thickness| 0.5 pm| 0.3 pm
Ambient temperature| 0 to 40 °C (32 to 104 °F) at 20 to 90 % RH
Power supply| 2 batteries 1.5 V AAA
Dimensions (length x width x height)| 114 x 27 x 54 mm
Weight| 152 g
Scope of delivery
- 1 x Layer thickness measuring device BB20
- 2 x 1.5 V batteries AAA
- 1 x Transport case
- 1 x Steel and aluminium base plate with different film thicknesses for calibration
- 1 x USB cable
- 1 x Operating manual
- 1 x PC software
Transport and storage
Note
If you store or transport the device improperly, the device may be damaged.
Note the information regarding transport and storage of the device.
Transport
For transporting the device, use the transport case included in the scope of
delivery in order to protect the device from external influences.
Storage
When the device is not being used, observe the following storage conditions:
- dry and protected from frost and heat
- protected from dust and direct sunlight
- For storing the device, use the transport case included in the scope of delivery in order to protect the device from external influences.
- the storage temperature complies with the values specified in the Technical data
- Remove the batteries from the device.
Operation
Inserting the batteries
- Insert the supplied batteries before first use.
- Loosen the screw and open the battery compartment (6).
- Insert the new batteries in the battery compartment (6) with correct polarity.
- Place the cover back on the battery compartment (6) and reinsert the screw.
Switching the device on
Info
Note that moving from a cold area to a warm area can lead to condensation
forming on the device’s circuit board. This physical and unavoidable effect
can falsify the measurement. In this case, the display shows either no
measured values or they are incorrect. Wait a few minutes until the device has
become adjusted to the changed conditions before carrying out a measurement.
- Press the Power button (7).
⇒ The display will be switched on and the device ready for operation.
Menu
Press the Red button (9) to open the main menu.
Press the ▼(3) or▲ button (8) to select the desired menu item.
In order to confirm the selected menu item, press the Red button (9) again.
Press the Blue button (2) to return to the previous menu item.
Setting the measuring mode
To adjust the measuring mode, please proceed as follows:
- Press the Red button (9) to open the main menu.
- In the menu, select the menu item Options and confirm by pressing the Red (9) button.
- Select the Measure mode menu item, then, select the desired mode. Confirm with the Red button (9).
Single mode:
After every measurement, an acoustic signal will be emitted. All measurements
are automatically saved.
Continuous mode:
The sensor must not be lifted between the individual measurements. No acoustic
signal will be emitted after the measurement. All measurements are
automatically saved.
Setting the working mode
To set the working mode, please proceed as follows:
- Press the Red button (9) to open the main menu.
- In the menu, select the menu item Options and confirm by pressing the Red (9) button.
- Select the Working mode menu item, then, select the desired mode. Confirm with the Red button (9).
The device offers two different working modes:
DIRECT:
This mode is suitable for quick and easy measurements.
Measurements are cached during a measurement series.
Once the device is switched off or changes into another working mode, the
measurement results will be deleted. The statistical analysis programme is
able to assess 80 measurements. Once the memory is full, the oldest
measurements will be overwritten.
GROUP:
The GROUP mode comprises group 1 to 4. In each group, 80 individual and 5
statistical measurements can be saved. Calibration and limit values can be
adjusted and saved individually.
Once the memory is full, current measurements will no longer be saved.
Measurements can be carried out as usual.
Setting the sensor mode
To adjust the sensor mode, please proceed as follows:
- Press the Red button (9) to open the main menu.
- In the menu, select the menu item Options and confirm by pressing the Red (9) button.
- Select the Used probe menu item, then, select the desired mode. Confirm with the Red button (9).
-
AUTO:
The sensor automatically selects a measuring mode. -
Fe:
The sensor operates according to the magnetic induction principle if it is placed on a magnetic surface. -
Not Fe:
The sensor operates according to the eddy-current principle if it is placed on non-magnetic metals.
Units
You can choose between metric (μm and mm) and imperial (mils) units.
If the measured value exceeds 850 μm, the device switches to mm.
To change the unit, proceed as follows:
- Press the Red button (9) to open the main menu.
- In the menu, select the menu item Options and confirm by pressing the Red (9) button.
- Select the Unit setting menu item, then, select the desired unit. Confirm with the Red button (9).
Reset
Resetting deletes the entire data in all memories. In order to carry out a
reset, follow these steps:
-
Switch the device off.
-
Simultaneously press the Zero button (4) and the Power button (7).
⇒ sure to reset will be displayed. -
Press the Red button (9) to confirm or the Blue button (2) to cancel the process.
The device automatically switches back on.
Background illumination
In the menu, you can control the background illumination using Options and
Backlight, during a measurement, use the Blue button (2).
Indication of statistics
In the menu item LCD Statistic, you can choose between the following
indications:
- Average
- Maximum
- Minimum
- Standard deviation (Sdev.)
The selected indication appears in the bottom right corner of the display in
the statistical data indication (17).
Statistics of the measured values appear on the bottom left corner in the
display in the measured values indication (18).
To view all statistical values of the respective groups, go to the menu and
select Statistic view.
Measured value indication
To view all measured values of the respective groups, go to the menu and
select Measure view.
Automatic switch-off
The device switches off after 3 minutes of inactivity. To deactivate automatic
switch-off, go to the Options section in the menu and select the Auto poweroff
function. Select Disable and confirm by pressing the Red (9) button.
Device calibration
Carry out a calibration before each measuring operation.
You can perform the calibration e.g. at an untreated or uncoated spot of the
object to be tested or use the set included in the scope of delivery.
Calibration mode
-
Press the Red button (9) to open the main menu.
-
Select the Calibration menu item, then select Enable.
⇒ You return automatically back to Calibration. -
Press the Blue button (2) to return back to the start screen.
Calibration is carried out here.
⇒ The following messages appear on the display:
• cal zero: There is no one-point or two-point calibration.
• cal 1 to 2: There is a one-point or two-point calibration.
• zero y: There is a zero-point calibration.
Info
Please bear in mind that the device at hand is a precision measuring device
that can determine coating thicknesses of no more than a few micrometres (1 µm
is equivalent to one thousandth of a millimetre). The surface condition of
most measuring objects is hardly ever perfectly even and homogeneous, even
though it might appear differently to the naked eye. Observed under the
microscope, even the smoothest surface looks much like a mountain and valley
landscape. The teensiest of scratches, cavities or contaminations can
therefore already have a negative effect on the expected measurement result,
seeing as they will also be a part of the measurement to a greater or lesser
extent. It does, however, not affect the accuracy of the device. Even after
the calibration unexpected measurement deviations of a few micrometres always
have to be regarded in this context.
Hence it is important to handle the supplied calibration accessories with care
to avoid scratches and dirt on their respective surface as far as possible.
Zero point calibration
An uncoated sample is required.
-
Place the sensor onto an uncoated sample. The calibration sample is to be identical with the actual sample in terms of material composition, shape and surface condition.
⇒ The following message will be displayed: <x.x μm> -
Lift the sensor again and keep it away from the sample (at least 10 cm).
-
Press and hold the Zero button (4) for approx. 2 seconds.
⇒ The following result appears on the display:
<0.0 μm>
CAL1
ZeroY
⇒ The calibration is terminated.
One-point calibration
This calibration is suited for high-precision measurements, e.g. for thin
coatings.
A calibration film, a coated and an uncoated sample are required.
-
Activate the calibration mode as described before.
-
Carry out a zero-point calibration.
-
Place the calibration film on an uncoated sample.
-
Place the sensor on the sample.
⇒ The measurement is effected. -
Lift the sensor again.
-
Adjust the required film thickness by using the Down/ Right (3) or the Up/Left (8) button. The film thickness should be at least the size of the estimated coating thickness.
-
Repeat this step several times.
-
Now carry out the measurement on a coated sample.
-
Press the Zero button (4) to adopt the current calibration.
-
To exit the current calibration, press the Blue button (2).
Two-point calibration
This calibration is especially suited for measurements on rough surfaces or
for high-precision measurements.
****You need two films of different film thickness. The thicker film should
be approx. 1.5 times thicker than the thinner film.
- At first, carry out a zero-point calibration.
- Using the first film, carry out a one-point calibration.
- Repeat this step with the second film.
- Place the sensor onto the coating that is to be measured.
After a brief acoustic signal is emitted, you can lift the sensor again.
⇒ The measured value appears on the display.
Calibration on radiated surfaces
Values of layer thickness measurements carried out on radiated surfaces are
usually too high. The average thickness can be determined as follows:
Method A
- Carry out the steps for one-point and two-point calibration as described above. Use a smooth sample with the same surface curvature and material type as the measuring object that is to be used later on.
- In order to determine the average value Xo, carry out approx. 10 measurements on the uncoated sample.
- Afterwards, carry out approx. 10 measurements on the coated sample to determine the average value Xm.
⇒ The difference between both average values is the average coating thickness Xeff. Take the standard deviation of both values Xm and Xo into account.
Xeff = (Xm – Xo) ± S
Method B
- Carry out a zero calibration with approx. 10 measurements on an uncoated sample.
- Afterwards, carry out a film calibration on the uncoated sample.
- Repeat this calibration with films of different thicknesses (max. thickness 50 μm). Together, these films should be as thick as the assumed layer thickness.
⇒ The layer thickness can be read from the display.
Method C
- Carry out the steps for two-point calibration with two films as described above.
- Use several 50 μm films to approach the actual surface as precisely as possible.
⇒ The layer thickness can be read from the display.
General information on measurements
If the calibration has been carried out correctly, the measured value lies
within the guaranteed measuring tolerance. Incorrect measured values can be
deleted in the menu. The last value results from the statistical calculation
and the guaranteed degrees of tolerance of the measured value.
Strong electrical or electromagnetic fields can influence the measured values.
The measuring tip is to be cleared of contamination prior to each calibration.
Wait for at least 4 seconds after each measurement before proceeding to the
next measurement as the device does not respond to measurement sequences
carried out too quickly.
Deleting functions
In order to delete measured values, select Delete in the menu, then select the
measured values you would like to delete.
-
Current Data
Deletes the last measured value. -
All Data
All data can be deleted during the respective working mode. -
Group Data
Comprises Delete all data. In addition, the limit values and one-point and two-point calibration values will be deleted.
Limit value functions
Limit values can be entered using the Limit function. It is possible to enter
limit values before, during or after carrying out measurement series.
- Press the Red button (9) to open the main menu.
- Select the Limit menu item, then select Limit setting.
⇒ Using the▼ (3) and ▲ (8) buttons, you can determine the upper limit value in the High limit section and the lower limit value in the Low limit section.
Each measured value beyond the fixed tolerance limit appears on the display as
follows:
H: Measured value is above upper limit value.
L: Measured value is below lower limit value.
Statistical measurements
The device is able to calculate statistics based on 80 measurements. A total
of 400 measured values can be saved.
No measured values are saved in direct mode, however, calculating statistics
is possible.
If you change between the different working modes or if the device is switched
off, statistics are deleted from the direct mode.
The following statistical values are calculated:
- NO.: Number of measured values in working mode
- AVG: Average
- Sdev.: Standard deviation
- MAX: Maximum value
- MIN: Minimum value
Memory space full
If the memory space is full in group mode, statistics will no longer be
updated. However, measurements can still be carried out. These measurements
will not be part of the statistics. In single measurement mode, FULL appears
on the display. If the memory space is full in direct mode, old measurements
are replaced by new ones and statistics will be updated.
Menu structure
Main menu | Submenu 1 | Submenu 2 |
---|
Statistic view
(Statistic view)| Average view
Minimum view
Maximum view
Number view
Standard deviation
view (Sdev. view)| –
Options| Measuring mode| Single mode
Continuous mode
Working mode
(Working mode)| Direct
Group 1
Group 2
Group 3
Group 4
Used probe
(Used probe)| AUTO
Fe
Not Fe
Units
(Unit setting)| pm
mils
mm
Background
illumination
(Backlight)| On
Off
LCD Statistic
LCD Statistic| Average
Maximum
Minimum
Standard deviation
Automatic switch-
off (Auto poweroft)| Activate
Deactivate
Limit values (Limit)| Limit value settings
(Limit settings)| Upper limit value
Lower limit value
Delete limit value
(Delete limit)| –
Delete| Current data| –
All data
Group data
Measured value indication (Measure view)| –| –
Calibration| Activate (Enable)| –
Deactivate (Disable)
Delete zero N
Delete zero F
PC software
Using the software Coating Thickness Tester stored measurement data can be
called up and saved via USB cable. The software is available for download at
www.trotec.de.
Info
The supplied free software is designed for useful basic functionalities.
Trotec assumes no liability with regard to this free software and also
provides no support on that score. Trotec accepts no liability concerning the
use of this free software and is under no obligation to make adjustments or to
further develop updates or upgrades.
Installation requirements
Ensure that the following minimum requirements for installing the PC software
are fulfilled:
-
Supported operating systems:
– Windows 7
– Windows 8
– Windows 8.1
– Windows 10 -
Hardware requirements:
– Standard USB interface
– Min. 7 MB of free hard disk space
– Recommended resolution: 1280*1024 with 16 bit
– .NET Framework 2.0 or higher
Installing the PC software
- Insert the data medium with the software into the drive or download the current software from the Service area of Trotec download centre.
- Double-click the installation file Setup.exe.
- Follow the instructions of the installation wizard.
Starting the PC software
- Connect the device to the PC using the supplied USB cable.
- Start the PC software.
⇒ The software connects to the device.
⇒ Data stored on the device can now simply be visualised in form of a chart or graphic.
Saving measured values (export)
You can export a selected group as csv or txt file and save it on your PC.
There the table looks as it does in the software.
- Open the desired file.
- You select either Save .TXT or Save .CSV.
Errors and faults
The device has been checked for proper functioning several times during
production. If malfunctions occur nonetheless, check the device according to
the following list.
The device does not switch on:
- Check the charging status of the batteries. Change the batteries, if required.
- Check that the batteries are properly positioned. Check the polarity is correct.
- Never carry out an electrical check yourself; instead, contact your Trotec customer service.
Table of faults
The following error codes can be displayed:
Error code | Cause of error |
---|---|
Err1, Err2, Err3 | Faulty contact between sensor and mainboard |
Err1 | Faulty signal of eddy-current sensor |
Err2 | Faulty signal of magnetic induction sensor |
Err3 | Faulty signals of both sensors |
Err4, Err5, Err6 | No meaning, reserved for later versions of the device |
Err7 | Faulty measurement. There might be a hardware problem. |
Maintenance and repair
Battery change
A battery change is required, when the battery indication (12) lights up or
the device can no longer be switched on. See chapter Operation
Cleaning
Clean the device with a soft, damp and lint-free cloth. Make sure that no
moisture enters the housing. Do not use any sprays, solvents, alcohol-based
cleaning agents or abrasive cleaners, but only clean water to moisten the
cloth.
Repair
Do not modify the device or install any spare parts. For repairs or device
testing, contact the manufacturer.
Disposal
Always dispose of packing materials in an environmentally friendly manner and
in accordance with the applicable local disposal regulations.
The icon with the crossed-out waste bin on waste electrical or electronic
equipment is taken from Directive 2012/19/EU. It states that this device must
not be disposed of with the household waste at the end of its life. You will
find collection points for free return of waste electrical and electronic
equipment in your vicinity. The addresses can be obtained from your
municipality or local administration. You can also find out about other return
options that apply for many EU countries on the website
https://hub.trotec.com/?id=45090. Otherwise, please contact an official
recycling centre for electronic and electrical equipment authorised for your
country.
The separate collection of waste electrical and electronic equipment aims to
enable the re-use, recycling and other forms of recovery of waste equipment as
well as to prevent negative effects for the environment and human health
caused by the disposal of hazardous substances potentially contained in the
equipment.
In the European Union, batteries and accumulators must not be treated as
domestic waste, but must be disposed of professionally in accordance with
Directive 2006/66/EC of the European Parliament and of the Council of 6
September 2006 on batteries and accumulators. Please dispose of batteries and
accumulators according to the relevant legal requirements.
Only for United Kingdom
According to Waste Electrical and Electronic Equipment
Regulations 2013 (SI 2013/3113) (as amended) and the Waste
Batteries and Accumulators Regulations 2009 (SI 2009/890) (as amended),
devices that are no longer usable must be collected separately and disposed of
in an environmentally friendly manner.
Trotec GmbH
Grebbener Str. 7
D-52525 Heinsberg
+49 2452 962-400
+49 2452 962-200
info@trotec.com
www.trotec.com
Documents / Resources
|
TROTEC BB20 Layer Thickness Measuring
Device
[pdf] User Manual
BB20, BB20 Layer Thickness Measuring Device, Layer Thickness Measuring Device,
Thickness Measuring Device, Measuring Device
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>