Programmable Digital Voltmeter
Please read through the manual
before installment and operation
Chapter 1. General Introduction
This series digital display meters are a new generation of programmable
intelligent voltage meters, which are mainly used in the real-time measurement
and indication on AC or DC current in an electric circuit .With features of
high precision, good stability, calibration-free long-termoperation and
onsite parameter setting on a panel, it is an idea substitute of traditional
dial instrument or ordinary digital instrument.
This series of meter features a modular design by utilizing large-scale
integrated circuit, high-performance single microprocessor, digital signal
processing technology and SMT techniques. The calibrating potentiometer inside
the meter is cancelled and soft calibration is adopted instead. The modular
and uniform design mode substantially reduces the production costs of the
meter , which makes it highly cost-effective.
Chapter 2. Technical parameters
-
Measuring range
1 1 Digital AC Ammeter Direct measurement AC0~600V Accessory device AC 0~9999V(PT */ 100V)
1 2 Digital DC Ammeter Direct measurement DC 0~600V -
Accuracy rating ±0.5 % FS±1 digit
-
Sampling rate: about 1 times/s
-
Overload consistent 1 2times instantaneous 2times/30s
-
Measuring display mode: RMS measurement, four-digit LED nixietube display
-
Display resolution: Max. display resolution of Voltage 0.1V
-
Input circuit consumption: <0.5VA
-
Auxiliary power supply: AC 220V±10%50/60Hz or AC/DC 85~265V
-
Auxiliary supply consumption:< 3VA
-
Overflow indication: Displaying character”HHHH”or”LLLL”
-
Operational environment: free of corrosive gas with temperature of -10~50°C , and humidity 85%RH.
Chapter 3. Programming and usage
-
Panel description
-
Key function
Set key: Press this key 2s to enter the programmable mode Under the programmable mode it is used to enter the next menu
Shift key: Under the programmable mode it is used to left shift the cursor one digit, and quit the programmable mode and return the measuring value display interface.
Down key: under the programmable mode, it is used for digression of parameter value or enter the next menu;
Up key: under the programmable mode, it is used for progressive increase of parameter value or enter the previous menu; -
How to operate
The instrument displays program version code and enters measuring valve display statue automatically 2s later when applying power . Enter programming status by pressing down SET 2S. The operational flowchart is as following:
Note: Password to enter the menu is 0000(user can modify the password CODE ) Note: Under the programmable mode select the menu and press the SET key to modify the parameter Use the key to input or modify the parameter. -
Program parameter specification
NO-| Parameter code| Parameter name| Setting range| Parameter default| Description
---|---|---|---|---|---
1| | Programmable mode
PROG| /| /| Prompt to enter the programmable mode
2| | Programmable password
CODE| /| 0000| Password to enter the programmable mode
3| | Multiply power of PT
PT| 1-100| 1| Set multiplying power of voltage transformer “PT”(Primary value of voltage transformer/secondary value)
4| | Setting programmable password
CODE| 0-9999| 0000| Modify programming password
Chapter 4. Installment and connection
-
Shape and cutout hole dimension (unit: mm)
Shape| Panel dimension| Case dimension| Cutout hole dimension
---|---|---|---
W| H| W| H| D| W| H
120x120square| 120| 120| 110| 110| 80| 112| 112
96x96square| 96| 96| 90| 90| 80| 92| 92
80x80square| 80| 80| 74| 74| 80| 76| 76
72x72square| 72| 72| 66| 66| 80| 68| 68
48x48square| 48| 48| 44| 44| 70| 45| 45 -
Method of installation
Choose the corresponding hole cutout dimension from the table above , make a hole in the installation screen, insert the instruments into the hole, place the two clamping pieces into the clamping holder and push and tighten them by hand. -
Terminal arrangement and function declaration of instrument(please accord to the one of instrument case) 3. 1Auxiliary power supply (POWER): The voltage range of operation is AC 220V,50/60Hz. It is suggested to install a fuse of 1A beside of the live wire when using the AC power supply to prevent damaging the instrument. In the areas with poor power quality, the surge suppressor and quick pulse group suppressor should be installed in the power supply circuit.
Electrical quantity signal input (I input ):HI represents “+”port of DC input signal, live wire terminal of AC voltage input signal and inlet wire terminal of AC current input signal. Input voltage should not be higher than the maximum value (AC 600V, or you should consider of using PT and installing fuse of 1A on voltage input port. While the current is higher than AC10A, you should consider of using CT)
3.2 Typical connection
Chapter 5.Cautions
- Please confirm if the power supply, input signal and each terminal wiring of the meter are correct and reliable before applying the power.
- The instrument must be preheated for 15 minutes to guarantee the precision of measurement.
- The instrument should not be rapped, knocked and vibrate excessively and its using environment should meet the technical requirements.
- The meter has been calibrated according to the measuring range required by the customer upon order. The user should check once again if the measuring range of the meter is fit with the specifications of the transformer and set the measuring range again if not.
Chapter 6. Packing and Storage
The instrument and accessories with packing should keep storage conditions cool and dry and free of wet and corrosive gas with temperature not more than 70°Cand not less than -40°C,and relative humidity85%RH.
Appendix 1 : Example of programming
Example 1: The factory default parameter is AC 0-600V(PT=1); If the
voltage transformer is5KV/100V should modify the PT multiplying power as
50(5000/100)
Example 2: The factory default parameter is AC 0-600V(PT=1); If the
voltage transformer is8KV/100V should modify the PT multiplying power as
80(8000/100)
Example 3: The factory default programmable password is 0000 modify the
password as 1234,
Return upon error password