premier 9065 Series Dental Surgical Instruments Instruction Manual

June 1, 2024
Premier

premier 9065 Series Dental Surgical Instruments

premier 9065 Series Dental Surgical Instruments

PURPOSE

The following instructions are for reusable surgical instruments, unless indicated otherwise by specific product labelling. These reusable devices are supplied non-sterile and must be cleaned and sterilized before first use and after each subsequent use. These instructions are intended for use only by persons with the required specialist knowledge and training.

SCOPE OF APPLICATION

*These instructions apply to the below listed items here, distributed by Premier:

Instrument Description Applicable SKUS
Rubber Dam Clamp Forceps 9065131
Forceps – Mandibular 9065008, 9065011, 9065018, 9065021, 9065031, 9065035,

9065039, 9065051, 9065052, 9065053, 9065058, 9065064, 9065077, 9065079.
Forceps – Maxillary| 9065009, 9065010, 9065015, 9065016, 9065024, 9065025, 9065027, 9065028, 9065030, 9065032, 9065034, 9065054, 9065057, 9065078.
Root Forceps| 9065019, 9065043, 9065044, 9065045, 9065063
Rongeurs| 9065070, 9065071, 9065073, 9065076

WARNINGS

Follow instructions and warnings as issued by manufacturers of any decontaminants, disinfectants and cleaning agents used. Wherever possible avoid use of mineral acids and harsh, abrasive agents.

  • Do not use chemical sterilant with caustic ingredients such as: surgical scrub solutions, povidone-iodine solutions, bleach, peroxide solutions, Virox 3, Sporox and Cidex PA. Do not use garment or surface disinfectants.
  • Do not autoclave with substandard stainless steel surgical instruments as this may cause a reaction and lead to rust or decolorization.
  • No part of the process shall exceed 140º C.
  • When reprocessing medical devices, handle with care, wearing protective clothing and face visors or goggles where appropriate.

LIMITATIONS ON REPROCESSING

  • Do not flash auto clave.
  • Devices may continue to be used if inspected and determined to be in a safe and good condition for its intended use.

INSTRUCTIONS AFTER POINT OF USE

  • If possible, soiled devices should be placed in a holding solution (combined disinfectant / enzyme  solution)immediately after use and prior to cleaning.

PREPARATION BEFORE CLEANING

  • Reprocess all devices as soon as it is reasonably practical to do so.
  • Where necessary, disassemble devices that require disassembly for adequate cleaning and reprocessing.

CLEANING/DISINFECTION

Manual cleaning/disinfection:
Some devices cannot be effectively cleaned by a washer-disinfector, those items should be manually cleaned as follows and then cleaned in a washer- disinfector as below:

  • Using a sink dedicated for instrument cleaning (not used for hand washing), rinse excess soil from the device (water temp<35°C).
  • Keeping the device submerged, with a brush, apply CE marked cleaning solution to all surfaces. Pay particular attention to underside and any jointed part of the device. Always brush away from the body. Do not use a wire brush.
  • Rinse the device thoroughly with clean water, so that the water reaches all parts of the device, then carefully hand dry with lint-free gauze or use a drying cabinet.

Automated cleaning/disinfection:

  • Use only CE marked and validated washer-disinfector machines and cleaning agents, following the manufacturers’ instructions for use, warnings and recommended cycles.
  • Load devices carefully, ensure any delicate parts of the device are not liable to damage during the loading procedure.

Surgical instruments have been validated to the following automatic washer- disinfector cycle:
Pre-wash 28 ºC for 4 minutes, Wash at 50 ºC for 8 minutes, Rinse at 50ºC for 4 minutes.
Thermal Disinfection at 90ºC – 95ºC for 1 minute, Drying for 25 minutes.
Validated detergents: Instro-Klenz, Enzy care, Olympic Spray dry 1000 and Spray clean 2000.

INSPECTION, MAINTENANCE, AND TESTING

  • Visually inspect and functionally check all devices. Dispose damaged devices or if contaminants cannot be removed. If any soil is still visible, return the device for repeat decontamination.
  • If any devices have been disassembled, reassemble in accordance with any product specific instructions for use and test.

PACKAGING

All devices should to be packed following local protocol in accordance with BS standards.

STERILIZATION

Use CE marked and validated vacuum autoclave operating at 134–137ºC 2.25 bar for a minimum holding time of 3 minutes always following the instructions of the machine manufacturer.

  • When sterilizing multiple instruments in one autoclave cycle, ensure that the sterilizer manufacturer’s stated maximum load is not exceeded.
  • Ensure all devices are dry before sterilization. If the devices cannot be dried prior to sterilization, then use distilled/de-ionized water in the final-rinse stage of cleaning.
  • Surgical instruments have been validated to the following sterilizer cycle:
    Sterilizing temperature 134 – 137ºC, Sterilizing time 3 minutes, Normal drying time 5 minutes, Extended Drying time 15 minutes.

STORAGE

Ensure all devices are dry before storage, and stored in dry, clean conditions at an ambient room temperature.

ADDITIONAL INFORMATION

  • Other forms of cleaning (alkaline and neutral) and sterilization (Cidex OPA, Ethylene oxide up to 65ºC for anesthetic devices) are permitted. However, always follow the reprocessing instructions provided by Premier and always consult with us if in any doubt over the suitability of any process used.
  • Follow cleaning and sterilizing guidelines as per HTM 01-01.
  • These instructions have been validated by the manufacturer of the medical device as being capable of preparing a medical device for reuse. It remains the responsibility of the processor to ensure that the processing, as actually performed using equipment, materials and personnel in the processing facility, achieves the desired result. This requires verification and/or validation and routine monitoring of the process.

For   Your Information – HTM (Health Technical Memorandum) series provides guidance on various aspects of healthcare facility management, and HTM 01-01 focuses on the decontamination of reusable medical devices.

Customer Support

©2023 Premier Dental® Products Company | All Rights Reserved V-PRO®, Virox 3, Sporox, Cidex PA, Instro-Klenz, Enzy care, Olympic Spray dry 1000 and Spray clean 2000 are not registered trademarks of Premier Dental Products Company

SymbolDistributed by: Premier® Dental Products Company 1710 Romano Drive, Plymouth Meeting, PA 19462, USA
Phone: 888-670-6100 / 610-239-6000
Fax: 610-239-6171
premierdentalco.com
info@premierdentalco.com
Logo

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Premier User Manuals

Related Manuals