NorthEast Monitoring HE-LX Holter Analysis Software User Guide
- June 1, 2024
- NorthEast Monitoring
Table of Contents
- NorthEast Monitoring HE-LX Holter Analysis Software
- User Guide
- Introduction
- Patient Information
- Holter Analysis
- Review Methods
- Reports
- Preferences & Configurations
- Archive Patient Data
- Licensing and Utilities
- Calculation of Heart Rate
- Test Issues
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
NorthEast Monitoring HE-LX Holter Analysis Software
Specifications
- Product Name: NorthEast Monitoring, Inc. HE/LX Holter Analysis Software
- User Guide: National Country Language
- Version: 6.1A
- Part Number: NEMM028_REV_H_EN
- Copyright: 2024 NorthEast Monitoring, Inc.
- Operating System Compatibility: Microsoft Windows 10 and Windows 11
- Patents: US D889,662; US D905,253; US 6125296; US 6666182
- FDA Approval: U.S.A. FDA 510(k) Market Approval Number K930564
Introduction
Intended Use and Indications for Use
Intended Use: The Holter Analysis Software provides Holter analysis
(arrhythmia and trend data) to physicians on their personal computers offline.
It can analyze data from legacy recorders like DR180+, DR181, and DR300.
Operator Knowledge
Users must have extensive Holter knowledge to identify various rhythms,
artifact, ST segment changes, and pacemaker failures. Familiarity with Windows
operating systems is also required.
Reports
Details about generating and interpreting reports will be provided here.
User Guide
Microsoft Windows 10 and Windows 11 are registered trademarks of Microsoft
Corp.
LibreOffice is community-driven and developed software, and is a project of
the not-for-profit organization, The Document Foundation.
- The Holter Analysis Software has been developed by NorthEast Monitoring, Inc.
- NorthEast Monitoring is an FDA Registered Facility, number 1224919, that follows all FDA CGMP Manufacturing Practices.
- The Holter Analysis software is covered by one or more US patents: US D889, 662, US D905, 253, US 6125296, US 6666182.
The Holter Analysis Software has
U.S.A. FDA 510(k) Market Approval Number K930564
Introduction
Welcome to NorthEast Monitoring’s HE/LX Holter Analysis Software. Used in conjunction with a NorthEast Monitoring Holter Recorder, HE/LX Holter Analysis allows you to fully review all of the ECG recorded during the Holter test, including all normal, ventricular, supraventricular, and paced beats. You can quickly review and edit morphology types, significant arrhythmic events, strips saved for the printed report, data trends, and tables. You can also review and edit report information before it’s printed, and then print whatever pages are required to document each patient’s Holter test. In addition, the software automatically reads recording data from the recorder’s flashcard – including entries made using the Event button – and saves sample strips of event markers and diary entries. Archiving, Remote Reporting and Spectral Analysis are also included.
- Intended Use and Indications for Use
Intended Use:
The Holter Analysis Software is intended to provide Holter analysis
(arrhythmia and trend data) to physicians on their personal computers, on an
off-line basis.
Indications for Use:
- The Holter Analysis Software is to be used to analyze the data recorded by Holter recorders such as the NorthEast Monitoring, Inc. DR200 and DR400. Detection of Arrhythmias: The Holter Analysis Software assists in the evaluation of ECG recordings of cardiac rhythm when intermittent arrhythmias are suspected due to patient symptoms such as palpitations, transient ischemic attacks (TIAs), syncope (fainting), or other such symptoms as determined by the physician.
- Efficacy of Treatment: The Holter Analysis Software assists in documenting the effect of pharmacological treatment of known arrhythmias is effective by measuring the frequency and duration of the arrhythmia compared to the frequency and duration prior to treatment.
- Pacemaker Evaluation: The Holter Analysis Software assists in the evaluation of the function of implanted pacemakers to ensure that the pacemaker is functioning within prescribed limits.
- The Holter Analysis Software is to be used only on the order of a physician.
Legacy recorders
While they remain in the field, the Holter Analysis Software can be used to
analyze Holter data recorded on NorthEast Monitoring, Inc. legacy recorders
such as the NorthEast Monitoring, Inc. DR180+, DR181 and the DR300. The only
limitation is that as of version 6.1 of the software, 12-lead or oximetry are
no longer supported.
Specifications for HE/LX Analysis software
Operator Knowledge
Users must have extensive Holter knowledge to properly identify sinus and
paced rhythms, abnormal rhythms, supraventricular and ventricular arrhythmias,
artifact, ST segment changes, and pacemaker failures. In addition, all
instructions assume a working knowledge of computers and, specifically, a
Windows operating system.
Software and Performance Specifications
Arrhythmia Detection
- Ventricular Premature Beats (VPBs)
- Supraventricular Premature Beats (SVPBs)
Maximum time analyzed
- 14 days
Presentation of data
- Standard 25 mm/s for 3-lead recordings
- Calibration indication can be provided in reports
ST Segment slope
- Positive or negative slope indicated based on manual setting of cursers by trained operator
LAN Capability
- Allows multiple users to access a common database of patient data without conflict.
Pacemaker detection
- Displayed and annotated for the operator
Reports:
- Capability of labeling all arrhythmias with the operator’s input
- Modular with ability for operator to select modules to be included
- All reportable values and labels are editable by the operator
- Ability to remove and/or add sample ECG strips
The Software Package includes:
- Software HASP key
- USB flashcard reader
The Holter (HE/LX) Analysis software must be installed on your hard drive to run. The software can be found and downloaded from www.nemon.com under the Support > Downloads and Documents page. A license file, provided separately, and the HASP key must also be installed to run the software. A demo patient has been provided with your software. Otherwise, you will need a flash.dat file from a Holter recorder to begin analysis.
System requirements
The Holter Analysis Software is to be used to analyze ECG data recorded by a
NorthEast Monitoring Holter recorder. The Holter Analysis Software must be
installed on the hard drive of the computer. To run the Holter Analysis
Software, the computer must include:
- Microsoft Windows 10 or Windows 11 operating systems
- Processor with a speed of 1 GHz or faster
- At least 1 GB of memory, 2GB for Windows 10
- At least 10 GB of free space on your hard drive
- Monitor with a resolution of at least 1280 by 1024
- USB flashcard reader or a laptop PC card slot (For DR200 and DR300)
- Laser printer is recommended
Clinical benefits
The Holter software is used to analyze NorthEast Monitoring Holter recordings
that are used to assist in the diagnosis of heart issues that do not show up
on an electrocardiogram. Holter monitoring is not intended to replace real-
time telemetry monitoring for patients suspected of having life-threatening
arrhythmias.
Report Serious Incident
The user and/or patient must report any serious incident that has occurred in
relation to the Holter Monitor should be reported to North East Monitoring and
the applicable competent authority /government agency in the country where the
incident occurred.
Initializing a flashcard for DR200
If you purchased a flashcard from an outside vendor, the flashcard must first be initialized using the HE/LX Analysis software:
- First, insert the card into the card reader, and select File > Flashcard > Initialize.
- If a card is found, the Initialize Flashcard window opens.
- The drive selected is highlighted in blue. Ensure that the correct drive is selected for your card. If the drive has not been found, check to be sure that a card is in the slot and the reader is attached to the computer.
- Select the card format option of your choice.
- Press Erase.
Note: If you insert a card into the recorder and get a message that the Flashcard or memory card is missing, the card is not formatted or erased properly.
Patient Information
New Patient
Copy the Holter ECG file to an open directory
For DR400 recordings, plug recorder into PC, and for other recorders, insert
the flashcard into card reader. Select File > Open/New. Click on a directory
(row) that is blank and press the New button at bottom of window.
Enter Patient Information
While the file is copying, you can start to update the patient’s demographic
information, select the appropriate Type of analysis/report (configuration)
and enter any diary information into the Diary Symptoms window.
Copy different flashcard, if required
If the patient information displayed does not match the correct patient,
correct the file and press
“Copy different flashcard.” Once the data is correct, press “Start” to begin
analysis.
Patient Information
To open the Patient Information window for the current patient, select File >
Patient Information. While most of the Patient Information window is the same
as that of a new patient, you will see the addition of the Status button, and
the Re-analyze button replaces the Start button because the Holter signal has
already been analyzed.
If you choose to change the Type of Analysis/Report while editing patient
information for “current” patient you will be forced to redo analysis and all
edited ECG data will be lost. For more information about Types of
Analysis/Report, see Section 6 Configurations and Preferences. Diary window
You can manually enter diary entries here when provided on paper by the
patient. Once the recording is analyzed, you will also see events that were
created when a patient presses the diary or event button on the Holter
recorder.
Preview data on flashcard or DR400
Review the clerical data on a flashcard by inserting the card in reader and
select File > Preview Flashcard. The Preview window displays the Patient ID,
recorder number, date recorded and the start time on the flashcard. No data is
saved on desktop by previewing.
Receive Patients Remotely
You will need to have an account set up to receive patient files from LX
Remote. Contact Technical Support on how to get started.
In order to input files to HE/LX, go to File > Open/New and press Remote
Receive button at the bottom of the screen. You will now see the Remote
Receive window. By default, the FTP directory is usually set up as c:\nm\ftp.
You can change this by browsing and selecting the directory where your
remotely received files are saved on your desktop.
Copy a patient file onto a patient directory by:
• Select a blank patient directory on the right-side of the screen by clicking
on it
• Select the incoming patient on the left-side of the screen by clicking on it
• Both should now be highlighted
• Press the Copy button.
• Repeat these steps until you have copied over all of the patients. The FTP
Delete button should be used to delete FTP records after they have been copied
into the software for analysis
Archive
From the Patient Open window, you can go to the Archive screen which allows
you to save a copy of the patient recording and report to another location.
Delete
This will allow you to delete one or more patients. This data will be lost
forever unless saved or backed up elsewhere.
HIS Export
The software includes an interface to your hospital information system (HIS).
Once a patient is completely analyzed using HE/LX Analysis, the user is able
to save a copy of the patient data and final report in a location for your
hospital information system to retrieve.
The Export Process
The user creates the HIS files by going to the Patient List, clicking on a
patient, and then clicking on the HIS Export button. The HIS Export button
will only be available if the patient has a report to export. Three files are
created during this process:
- cmp – the existence of this file tells you that the other files are ready to be retrieved.
- pdf – the patient’s Holter report
- txt – a text file in an xlm format with the patient data.
The HISExportPath in the user’s h4w.ini file determines where these files are saved. By default, the files will appear in the user’s c:\HIS_Transfer directory. Each file name will begin with the following format, Pat-$IDNO-$DateRecorded-$CurrentTimeStamp.
Send Report Remote button
The NARP (NorthEast’s Automated Remote Process) must be running on your
computer to work, and the Remote user must be running LX Remote to receive the
report. When you select a patient with a report, this button will be enabled.
The Patient List
All of the patients saved in the software appear in the Patient List when you
select File > Open/New from the toolbar. From the Patient List you can choose
to open a patient’s record by clicking and highlighting it and pressing “OK”
or by double-clicking anywhere on the patient line.
From the Patient List, you also can create a new patient by clicking on an
empty directory (row) and clicking the New button at the bottom of the screen.
The New button is only available when a directory is empty.
On the Patient List, if you don’t know the meaning of a specific column, you can hover your cursor over the heading for that column and help will be displayed if it exists.
Customize the Patient List View
You are able to customize the Patient List for your specific needs in the
following ways:
- To change what columns appear on the list, right-click on the top row of the column headings. You will now see a list where you can check or uncheck items to include or exclude from the list.
- Once you have decided what columns you want to see, you can change the position of the column by clicking on it and dragging it to a new location.
- Grab the column divider to increase or decrease the size of any column.
- Sort on any column by double-clicking on the header for that column. An arrow will appear to show you which column is sorted.
Search Patients
The Search Patients box will search all visible columns for a match, and only
show patients who meet the search criteria.
Holter Analysis
Arrhythmia Analysis
Certain analysis and related documentation criteria are already set when you
click the Start button in the Patient Information window. They include all the
settings that appear in the five windows that are accessible using the
Settings button either in the Patient Information window or on the main Holter
toolbar. Those windows are What Strips to Auto Save, How Often Strips Auto
Save, Scanning Criteria and Spectral Analysis.
Scanning Criteria
The Scanning Criteria are used during Holter analysis to define some of the
arrhythmias labeled by the software, along with settings that control the
amount of information processed.
What Strips to Auto Save
All possible strip labels appear in this window. Each label can be turned off
or on to indicate whether sample strips of that type should be saved for the
final report. A check mark indicates that sample strips with that label will
be saved.
How Often Strips Auto Save
These settings control the distribution of strips that are saved for the
report.
Spectral Analysis
Spectral analysis settings for Pro users only.
Arrhythmia Re-Analysis
Re-analysis is required after changing most of the settings in the Scanning
Criteria window. The exceptions to this are the Lead Labels, Heart Rates and
Prematurities. If a setting requires re-analysis, the software asks you to
confirm whether it is appropriate to re-analyze. If you want analysis to
occur, click Yes. If not, click No and your edits will be discarded.
Because re-analysis is required after a change in most Scanning Criteria, be
sure to make any changes to the Settings before you work on the final report.
Any bin, template or beat editing, along with manually saved strips and typed
comments will be lost after re-analysis. Additionally, any changes to Saved
Strips, Tables and Report Summary for the final report should be done after
settings are updated, as if done before, they will be lost.
Update
An Update is required when What Strips to Auto Save, How Often Strips Auto
Save windows, and other items in Scanning Criteria (Heart Rates, Interval Size
and Prematurities) window are changed. An update does not go through a full
re-analysis, so bin, template or beat editing done before the update, will NOT
be lost. However, any changes to Saved Strips (except Alternatives) Tables and
Report Summary will be lost during an update.
Whether the Update button appears in the Review toolbar depends on the
“Automatically update tables” setting in the Preferences window. If the
software is set to automatically update, the button does not appear; if you
must update the data after making changes, the button appears.
ST segment Analysis
To review and/or update the location of the ST markers used during analysis,
Select Review > Calibration.
Pacemaker Analysis
Pacemaker activity is recorded without distorting the patient’s ECG, by
removing the effects of the pacemaker spike and replacing it with a pacemaker
marker. In order for pacemakers to be analyzed, the Pacemaker Criteria in the
Scanning Criteria window must be set properly.
Spectral Analysis
To perform HRV analysis, the software considers only normal-to-normal RR
intervals and performs the analysis based on the settings available in
Settings > Spectral Analysis. HRV Analysis can only be performed when the
Analysis time is 24 hours or less. For this reason, the HRV menu option from
the toolbar will be disabled when the Analysis duration is greater than 28
hours.
Review Methods
Patient ECG data can be reviewed on the monitor of your computer in different ways. You can review and edit:
- the templates established during analysis (Bin)
- the most significant events identified during analysis (Critical Events)
- strips saved for the final report (Saved Strips)
- on-screen full disclosure of all the ECG (Page)
- graphs showing the heart rate and RR interval data (Trends)
- tables compiled for the report (Tables) and
- Superimposition.
Color coding
Throughout the HE/LX Analysis software, the ECG is color-coded based on what
the system has labeled each beat as:
-
Green: Beats the software has identified as normal.
-
Yellow: Beats identified as supraventricular premature beats (SVPBs). They have a normal morphology, but fall early.
-
Red: Beats identified as ventricular premature beats (VPBs). They differ significantly from the normal; they are not necessarily premature.
-
Orange: Beats identified as part of an AF event. Either Atrial Fibrillation or Atrial Flutter (AF).
The following beats have a normal morphology, but an irregular HR. -
White: Beats identified as pauses, based on the definition in the Scanning Criteria window. Thewhite overrides any other color that the beat may also qualify for (e.g., red because it is a VPB).
-
Light blue (cyan): Signal that appears to be contaminated by artifact.
-
Cobalt blue: Beats identified as paced.
All data must be reviewed carefully to ensure that you agree with the beat labels the software has selected; if you do not agree, you can change them and their color will change appropriately.
Labels
In addition to the color-coded labels the software provides for each beat,
there are some labels only you can use to relabel beats. These are:
- Questionable (Unknown): Use this label to separate out beats you cannot identify and keep them from being included in another category. They are colored green, like normals.
- T-wave: Use this label if the software has identified a portion of the signal as a QRS, but it is not. This will remove the beat from the counts and will merge its RR interval with the preceding RR interval.
Measure ECG data
The data fields in the center of the toolbar – HR (2 RR), Time, and the ST
indicators for each channel – contain data calculated based on the two blue
calipers in the active strip. As you drag the blue calipers, those fields
change, reflecting the new caliper positions.
To measure a two-beat heart rate, place the calipers two RR intervals apart;
the measurement appears in the HR (2 RR) field. To measure ST in any of the
channels, position the left caliper in the isoelectric area of the PR interval
and the right caliper where you want to make the ST measurement; the
measurements for each channel appear in the appropriate fields.
To move the calipers keeping them the same distance apart, click the check box
next to Both and then drag the calipers. Click again to remove the mark and
move them separately.
To keep the calipers in the same locations as you move through different
screens of ECG, click the check box next to Lock; the calipers will stay in
the indicated locations unless you move them again. Click again to turn off.
To make ST measurements, place the left caliper in the isoelectric portion of
the PR interval, and place the right caliper where you want the ST segment
measurement to be made; the vertical distance between where the left caliper
intersects the ECG and where the right caliper intersects the ECG will appear
in the ST field for each channel (labeled ST 1, 2, 3).
Time-of-Day
All review windows are linked by time-of-day. In addition, the Page window is
linked to all other review windows through the right-hand button on the mouse.
From any other review window, a right-click will jump to the Page display,
retaining the current beat. After that, a right-click in Page will then take
you back to where you originally were, regardless of whether you change the
current beat in the Page window.
Bin Screens
The software determines what the patient’s normal QRS complex looks like and
establishes a template called “normal.” Each beat after that is compared to
the normal template; matches to that template are also called normal, while a
similar but slightly different morphology will establish a new template, also
called normal. A QRS complex that differs more significantly from the normal
template will establish a template called “ventricular.” A new template is
established for each different morphology identified by the software.
Subsequent matches to a template are labeled based on the template label, the
timing of the beat, and other criteria.
Moving from one level of the Bin window to the next can also be done by
double-clicking on the ECG in the bin, template, or beats display. Each
double-click changes the radio button position one button to the right.
After analysis, the templates that generally look alike are grouped together in “bins.” You can review these bins by morphology, that is, all normal bins or all ventricular bins. Within the Bin window, you can also review by template, displaying all the templates within each bin.
To change the morphology displayed, click on the arrow in the Morph field to
display your choices, and then click on the type you want displayed. To undo a
relabel, click the Undo button. It will restore the templates to their state
before the last relabel.
The template display contains up to 12 templates that matched the current bin.
If more than 12 templates fell into that bin, you can access additional pages
of templates by using the Page Down key, the scroll bar or the Scan button
Clicking the Beats radio button displays up to 24 of the beats that matched
the current template. Use the Page Down, the Scan button, or the scroll bar to
display additional matches to the template.
Critical Events Screen
Each type of critical event has a number associated with it – the number of
events of that type that were identified for this patient. Every event of each
type can be displayed, either one at a time (full-sized) or 12 at a time
(miniature).
A beat that appears in one category of Critical Events does not appear in all
other applicable categories. For example, if a VPB appears in Bigeminy, it
does not appear in VPB; if a paced beat appears in Sense Failure, it does not
appear in any other paced category. Therefore, do not rely on the counts in
Critical Events to provide comprehensive totals.
Histograms
The top portion of the Critical Events window presents a histogram showing the
distribution of the events within the type displayed – either an RR histogram
or a 24-hour histogram.
Relabel in Critical Events
All relabeling done in the Critical Events window is single-beat editing. To
relabel a beat within the Critical Events window, click on the event to select
it; this turns the relabel buttons from dim to colored. Click one of the
colored relabel buttons to relabel the selected beat.
Saved Strips Screen
The report includes full-size, 7.5-second, 25-mm/sec strips on a background
grid. Some strips are automatically saved based on the settings in the What
Strips to Auto Save window. You can also use the Keep button to manually save
strips while reviewing the Holter recording.
Review Saved Strips
To review the saved strips, click Saved Strips in the Holter menu. The Saved
Strips window displays a miniature version of the strips 12 at a time. Each is
labeled with its strip label and the time-of-day at which it occurred. Page
through them using the Page Up and Page Down keys, the up and down arrows of
the scroll bar, or the scroll button on your mouse.
All strips are three-channel unless oximetry data was collected for this
patient; if oximetry data is present, it appears in the area where channel 3
normally appears, and data fields of SpO2 data appear to the right of the
standard data fields.
ST event labels include the channel in which the ST segment change occurred.
Change the active strip
At any time, there is only one active strip, the strip outlined in blue. Four
fields above the strips refer specifically to the active strip. Those fields
include time-of-day, a strip number, HR (heart rate) and HR2 (the second heart
rate, that is, the heart rate of a run of VTAC or SVT in the strip). To change
the active strip, click on the one you want so that the outline surrounds it.
You can also change the active strip by clicking the List button in the
toolbar to open the List window. The List window lists each strip label and
corresponding heart rate in order of time-of-day. To display a particular
strip from the list, click on the appropriate entry on the list and click OK,
or double-click on the entry. To exit without changing the active strip, click
Cancel.
Edit a strip label
To change the label of the active strip, click on Edit in the toolbar. The
Edit window opens; it includes a field with the current label of the strip and
the heart rate of the ECG in the strip, along with the second heart rate, the
rate of either SVT or VTAC if it is present. (A second heart rate of 0
indicates that there is no run on the strip.)
Delete a strip
- To delete one strip in the Multiple strip display, click on it to make that strip the active one, then click Delete in the toolbar. To delete more than one strip, click on the first strip to make it the active strip; in addition to the blue highlight around the strip, there is also a yellow highlight around the time-of-day, indicating that the strip is selected. Click on any additional strips you want to delete, then click Delete in the toolbar. All of the selected strips (as indicated by the yellow highlight) are now deleted.
- When you delete a strip, its label becomes red; strips with red labels are not included in the printed report. To retrieve a deleted strip, click on it and then click Delete in the toolbar again.
- To delete all of the strips displayed, click the button labeled Del/Undel All. To retrieve all of the strips displayed, click the button again.
- To delete only one or more channels of a strip, click on a strip to make it active. Then, click on one of the check boxes labeled Channel 1, 2 and 3 to delete it.
Alternative strips
Some strips can be replaced by an alternative: maximum and minimum heart
rates, shortest and longest RR intervals, and fastest and longest runs of VTAC
and SVT and longest and fastest runs of AF. If you would prefer to select a
different alternative, click on the strip to make it active and the
Alternatives button appears.
Click the Alternatives button to make the Alternatives window opens,
displaying other choices for that label. To select a different strip, click on
the strip and then the Select new Alternative button.
Expand the active strip
To view a strip more closely, either click Single in the toolbar or double-
click on the strip. It then fills the Saved Strips window. Each beat is
labeled with either the heart rate (BPM) or the length (in milliseconds) of
the RR interval following the beat.
Saved Strips are re-compiled after every update or re-analysis, so be sure to
make changes to the automatically saved strips only after you have completed
all other editing. Any editing of automatically saved strips that occurs
before an update or re-analysis will be lost. Manually saved strips remain as
is.
Page Screen
The Page window allows you to review all of the ECG stored during the
recording, like an electronic full disclosure. The window is divided into two
displays: a miniature, single-channel presentation and an expanded three-
channel display with a background grid. Use the Full
screen/Expand button to toggle between the miniaturized ECG and a combination
screen formats. Single-channel page display
The single-channel page display contains a blue highlight box surrounding one
of the QRS complexes, the “current” beat. The time-of-day at that beat is
displayed in the time field in the upper left corner of the window.
The displayed ECG can be adjusted in these ways
- To adjust the ECG so that the highlighted beat appears in the center of the page, click Center.
- To move the highlight box to a different beat, click on the beat.
- To change the channel displayed, click on the Lead field and select a different channel from the list.
- To change the amplitude of the displayed signal, click on the Gain field and select a different size from the list.
- To change the amount of time displayed on each page, click on the Zoom field and select a different amount of time.
In the single-channel display, you can visually review pages of ECG by using the Page Up and Page Down keys, by clicking on the down arrow of the scroll bar, by using the scroll button on your mouse, or by clicking the Scan button. Turn the Scan button off by clicking it again. Control the speed of the scan by pressing + to make it faster and – to slow it down.
Expanded ECG display
The expanded display shows three channels of ECG in the bottom of the Page
window. To display expanded ECG, click the Expand button.
The expanded ECG in the bottom half of the window and the single-channel
display in the top half of the window are linked. The expanded strip is
centered on the highlight box in the top half. If you move the highlight box,
the ECG displayed in the bottom half changes appropriately.
Relabeling in the Page display
The type of relabeling performed is determined by the setting in the Mode
field:
- Single beat relabels just the current highlighted beat to whatever label you choose.
- All matches relabel the template to whatever label you choose.
To relabel in the Page window, click on the beat to be relabeled, select the
appropriate setting for the Mode field, and then click the appropriate relabel
button.
To relabel multiple single beats to the same label, click on the first beat,
then press the Shift key and click on each additional beat. A blue highlight
box surrounds each of the beats to be relabeled; click the appropriate relabel
button.
To relabel a string of beats to the same label, click on the first beat and then drag across to the last beat; the beats turn magenta. Then click the appropriate relabel button.
Whenever you use a relabel button, a message appears in the bottom strip of
the window indicating what label was given to the beat and how many beats were
relabeled. In addition, error messages appear there whenever you try to
relabel inappropriately.
Although the S and T labels are available in Page, only a single beat at a
time can be relabeled to S or T. If Mode is set to All matches and you use the
S or T relabel button, a single-beat edit will be performed
Insert a beat
If while you are reviewing the ECG in the Page window, you see that a
particular beat is included in the highlight box of the preceding beat, it
means that the beat was missed. This is usually because of very low amplitude,
but sometimes because of low slope. To force the system to count the beat, you
can use the Insert button in the expanded Page toolbar.
To insert a beat, first click near the beat so that it appears in the expanded
Page display, then drag or click the left-hand caliper to the location of the
missed QRS complex. Click the Insert button in the toolbar in the middle of
the window. The Insert window opens, with the time-of-day of the new beat
listed in the first field and a beat label in the Morphology field. Click on
the arrow in the Morphology field to display the list of label choices and
make your selection. Then click OK to insert that type of beat where the left-
hand caliper is.
Save strips for report
To save a strip, click on the beat you want at the center of the strip to move
the highlight box there, and then click the Keep button; the Keep window
opens. The Description field contains the current beat label. To relabel the
strip, either type the label in the Description field or select a label from
the scrolling list. Click OK to save.
The Keep window includes two heart rate fields: HR, which equals the heart
rate of the background rhythm of the strip, and HR 2, which is the rate of the
run (VTAC or SVT) on the strip, if there is one.
Both fields can be edited.
To save strips of an event longer than 7.5 seconds, drag the cursor across the
ECG to be saved (the selected ECG turns magenta) and then click Keep. In the
Keep window, you can enter the label of the first strip in the series and then
click the left button, which indicates how long a time period to be saved.
To save multiple strips, all with the same label, click on a beat at the
center of the ECG to be saved, then hold the Shift key down and click on
another beat. Then click Keep. In the Keep window, click the button labeled “n
strips” to save all the selected examples; click the button labeled “1 strip”
to save just the first.
Turning AF On/Off
If a patient is in intermittent atrial fibrillation or flutter, you can
disable the SVPB counts and/or relabel that region as AF. To do so, select the
ECG by dragging across it (it turns magenta), then click the AF On/Off button.
All the selected beats turn orange and be labeled as AF.
iTrends Windows
- On all of the trends, time-of-day appears on the horizontal axis. RR intervals are plotted so that the range within each minute appears as a vertical line; the top end of the line indicates the longest RR interval within that minute, and the bottom of the line indicates the shortest RR interval within it. The blue marker
- The blue vertical marker is located at the time-of-day of a particular 30-second segment. Click on either trend to move the marker to a different time-of-day. The time-of-day that appears in the large data box indicates the time-of-day at the marker. The data boxes at the top of the display indicate the data collected for the minute where the marker is located, including heart rate (HR), the shortest RR interval (RR Min), the longest RR interval (Max), the total number of VPBs and SVPBs, and the total number of VPBs and SVPBs that occurred during runs of VTAC and SVT. In addition, the ST data boxes include data collected for the 30-second interval where the marker is located, including the ST segment measurement for each channel in ST 1, 2 and 3, and the ST slope measurement for each channel in Slope 1, 2 and 3.
- Asterisks indicate that there is no data for that time period, usually because of artifact.
Relabel to Artifact
In the Trends window, to relabel a continuous period as artifact, click at the
time-of-day you want to start rejecting and then drag until the end time. The
time period turns magenta. Now click the Artifact button.
General Trends
The general trends present data in one-minute increments throughout the Holter
period, including the RR trend showing the range of RR interval measurements;
the heart rate trend showing the average heart rate; total VPB and VTAC
trends; and total SVPB and SVT trends.
ST Trends
The ST trends present the ST segment analysis data in 30-second increments
throughout the Holtered period for all three channels of data. The placement
of the ST calipers is automatic unless you re-set them in the Calibration
window.
The ST trends include three components for each channel:
- the patient’s baseline ST measurement, that is, the patient’s normal ST;
- the actual measurement made for each 30-second increment; and
- the slope indicator for each 30-second indicator.
The baseline measurement is trended as a blue line, the actual measurement is green, and the slope indicator is a vertical red line drawn from the actual measurement to the measured value at the slope caliper.
Tables Window
To review the tables compiled for a patient, click Tables in the Review
toolbar. The listing of what tables are available appears at the right of the
screen. The displayed table is highlighted in blue. To display a different
table instead, click on its name in the list.
Within the Interval Table Edit window, you can click in any editable data
field to type in your changes. Highlight an existing entry and type over it,
or click to the right of the entry and backspace to eliminate it and then type
your entry. To eliminate information within an interval, use the Zero All
buttons to zero out the day or enter a zero into each field you want to
remove.
Be sure to make any changes to the tables carefully. Incorrect information
entered in this window can cause inconsistencies in the printed report.
Additional features
Superimposition
Select Superimposition from the drop-down Review menu. Click on Scan to start
and stop the superimposition display. Control the speed of the scan by
repeatedly pressing + to make it faster and – to slow it down.
Calibration Adjustment
- Select Calibration from the drop-down Review menu. Three channels of calibration signal are displayed. The two horizontal lines for each channel should be lined up so that one is level with the top of the square wave and one is level with baseline. Drag the lines to move them.
- To increase the size of the signal for analysis, set the horizontal gain markers close together. To decrease the size of the signal for analysis, set the horizontal gain markers far apart.
- If you use the gain markers in this way, the signal is no longer calibrated and no ST measurements are correct.
- Whenever you make changes in the Calibration window, the signal must be re-analyzed when you exit. If you choose to not re-analyze, the changes are not saved.
Invert/Hide
To invert the signal in a channel or to hide it from view go to Review >
Invert/Hide. Click on the check box for each channel to be inverted or hidden;
click on it again to return the signal to normal.
Shorten analysis time
To shorten the analysis time, move the marker to the time-of-day at which you
would like to end analysis, then select Review > Shorten analysis time.
Reports
To access the report modules, select Reports from the Review toolbar. The
modules that are available for the current patient are listed in the right
half of the Reports window.
To include a module in the report, the check box next to the module name in
the Reports window must contain a check mark. Click on an empty box to add a
check mark, and click on a check mark to remove it. To turn all of the modules
on or off, click on the All On/Off check box under the report module list; to
change them all again, click on the All On/Off check box again.
Report heading
The Patient Information module includes a Report heading so that you can
customize the report for your facility. To enter text in a line, click on the
field and type your entry.
If your address comes up automatically, but you would like to change it for a
particular patient, you can either make your selection from the addresses you
have associated with different report configurations or you can edit it on
this screen.
Adding Logo to report
A reasonably sized log.jpg file can be put into the C:\nm\bin\logo\ directory.
Strip annotation
In the Reports window, set the Strip annotation field to indicate how you
would like the beats annotated. The beat labels consist of:
- N for normal
- S for SVPB
- V for VPB
- A for artifact
- P for paced (A, V, or AV)
- F for AF
- D for event marker
- ? for questionable/unknown
Report summary
The summary that prints on the front page of the report can take one of four different formats, although the content will vary depending on a couple of factors: if the patient has a pacemaker, paced data replaces ST data; if Afib has been turned On or Off, the Report summary includes the percentage of time that Afib was identified and excludes supraventricular counts for that period.
When you click on the Create Report button, the front page and list of reports
that you selected will be created in LibreOffice Writer. Using LibreOffice,
you can edit the report and enter Comments in the box labeled “COMMENTS”. Once
you are satisfied with your report, you should save it and create a PDF in
LibreOffice or by clicking on the Create PDF button.
If you enter COMMENTS into the box, those comments will be saved if you need
to recreate the report. Any other edits will be lost if you do not Save the
report, although the report will be deleted any time you reanalyze the
patient.
Status indicators
Use Status indicators to keep track of whether a patient’s Holter has been
edited, printed, and/or verified already. From the Patient Information screen,
you can also lock a patient’s record on the Status window. Click the check box
to add or remove a check mark.
Full disclosure
Full disclosure is a printout of all the ECG recorded during the Holter
monitoring period, in a miniaturized format. Each page is annotated with time-
of-day along the left margin. Update the Full Disclosure of the Report screen
to customize what and how the full disclosure appears.
The time per page check boxes control how much total ECG is printed per page.
If you choose to print two channels of ECG, the 30 min/page setting will print
both channels during a 15- minute time period, and the 60 min/page setting
will print both channels of a 30-minute time period.
Print
Print your report in Adobe after reviewing.
Preferences & Configurations
- Preferences window
To open the Preferences window, select File > Preferences.
Draw grid
You can choose whether there is a background grid behind the expanded strip
displayed in the Page window. A check mark indicates that a light grid will
appear. No check mark means that the grid will not appear.
Automatically update tables
A check mark should appear in the check box so that the software automatically
updates counts, tables and strip labels after you relabel a beat, template or
bin in any of the Review windows. If you choose to not automatically update,
you must manually run an update after making changes. Click on the blinking
red Update button to make this occur.
Save new Physician or Interpreting physician
This field allows the software to ask you whether to add a new physician or
interpreting physician name to the appropriate list when you close the Patient
Information window after typing a new name in either field.
Annotation
Indicate here whether the beats should be labeled on the screen with a beat-
by-beat heart rate calculation or RR interval length.
ST measurement
Set your preference for ST measurement here.
After analysis show
This field allows you to determine which window is displayed upon the
completion of analysis. Saved Strips / Single view / Default sort
Select the two default options that you prefer. The defaults will be used each
time you create a new patient, but can be overridden on the Saved Strips
screen.
Edit fields
Entries in those fields can be preset so that you can make a selection from a
list instead of typing the physician’s name or report summary phrases in for
each Holter test.
Physicians and Interpreting physicians
To add a name to the list in the Physician field of the Patient Information
window, go to File > Preferences > Physicians to add, edit or delete the
physician names.
Beat Label Edit
This screen allows you to customize the list of labels that appear on Strip
Labels, Saved Strips and the final report. Double-click on the No. of the
label that you want to edit on the column labeled “Editable custom label,”
edit and press Enter.
Note: Changes to labels must be made carefully because the meaning of the
label MUST NOT change. For example, when the system calls a beat ventricular,
it uses the VPB label when saving strips for the report; you can change the
text to read VE instead, but not SVPB or BBB, or your report will be
incorrect.
Indications and Medications Edit
This window is for customizing the indication and medication lists on the
Patient Information screen.
Running the Configurations program
Access the Configurator with the HE/LX Analysis program running via the File >
Preferences window.
Default Configurations
Several default configurations have been created for you to copy to build your
own customizable configurations. All but the Holter default configuration can
be deleted if you so choose. They can be brought back later by clicking the
“Restore” button.
Configuration window
Once set up by you, the main Configuration window opens with a listing of all
current configurations of your software. Each should have a unique name. To
make changes, you can either edit an existing configuration or create a new
one. You can also delete a configuration if you no longer need it.
To Create a Configuration
To create a new configuration, click on the name associated with a
configuration similar to the one you want to create, and then click the Copy
button. Now you can edit your new configuration to what you want. Be sure to
change the name to something that is meaningful to you.
Configuration folders
The window for a specific configuration consists of a pile of folders with
tabs. Each folder contains the controls for a particular window or portion of
the HE/LX Analysis software. Within the folders for a configuration, any entry
in any of the fields automatically appears whenever you create a new patient
using that configuration. To display the fields in a particular folder, click
on the tab for that folder.
Changes to Strip Labels must be made carefully because the meaning of the
label MUST NOT change. For example, when the system calls a beat ventricular,
it uses the VPB label when saving strips for the report; you can change the
text to read VE instead, but not SVPB or BBB, or your report will be
incorrect.
Changing or adding a configuration
For each configuration you create or edit, make changes in as many tabs as you
need to. When all of the tabs are updated to your liking, click the OK button
at the bottom of the window. Your new configuration will be saved and the
window closed; the main Configurator window then appears.
Cancel a configuration
To exit without saving the new configuration, click Cancel. The window closes
and the main Configurator window appears.
Exiting from the Configuration program
To exit from the Configuration program, click on the red Close button in the
upper right corner of the window.
Using a configuration (Type of Analysis / Report)
The configurations created or edited using the Configurator appear when you
start a new Holter test. When you select File > New to open the Patient
Information window for a new Holter test, a list of the Configuration
descriptions appears in the Type of analysis/report field; select your choice
from that list.
If you use an inappropriate configuration for a patient, after analysis, you
can either change the configuration for that patient and reanalyze, or you can
go to the patient’s Settings windows and make necessary changes. If the system
reanalyzes because of a new configuration or changed a setting, any editing
you did previously will be lost and need to be redone by you.
Archive Patient Data
HE/LX Analysis Patient Archive is accessible by going to the File > Open/New >
Patient List and clicking the Archive button at the bottom of the screen. The
Archive tool has two tabs – Archive Patients and Restore Patients.
Archiving a patient consists of two steps: 1) copy the patient to the Archive
folder and 2) delete the patient from the Patient List.
Once archived, if you need to go back to a patient who has been Archived and
deleted from the Patient List, you can go to the Archive folder and restore
the patient back to the Patient List.
Archive Type
From the toolbar, you can select an Archive Type. Archive Type “Full”, will
copy all patient data, including the entire Holter recording and report; and
Archive Type “Report only” will save an electronic version of the patient’s
Holter report, including ECG strips, but not the recording.
Only by doing a “Full” Archive, can the patient’s Holter data be restored and
reanalyzed.
Archive Patients Tab
The HE/LX Patient Archive Tab will initially open with the “Archive Patients”
tab selected. Once opened, all patients who can be archived will be
highlighted in blue. A patient is highlighted in blue when:
- They have been analyzed
- They have a report
- They have never been archived previously, or their information has been updated since they were last archived.
If you want to override the highlighted patients, you can hold the Shift-key
and click on any directory in order to manually select or deselect it. When
finished, click the “Archive Selected Patients” button at the bottom of the
screen and the selected patients will be copied to the Archive folder.
If a Patient is opened and viewed, but not updated, the system will perceive
this patient as being updated and show that the patient needs to be archived
again. To ensure that this does not become a problem for you, be sure to wait
to archive a patient until you are sure that they will most likely not be
viewed or reviewed again.
Delete Patient Records
Once archived, you will want to delete the patients from the Patient List in
order to make room for new patients. The Archived check box, labeled “A” on
the Patient List, will show you which patients have been archived and can be
deleted at this time.
Restore Patients Tab
To restore a patient that has been archived and deleted from the Patient List,
select the Restore Patients tab on the HE/LX Patient Archive Tool. When
selected, it will be blue. You will now see a searchable list of all archived
records on the right. Select the archived patient you want to restore on the
right, and an empty patient directory on the left. Then click the Restore
Selected Record at the bottom of the screen to copy the patient back to the
Active Patient List.
Once restored, the Patient record will be Locked. Only unlock the patient if
you want to update and create a new report.
Delete record after Restore
At this time, if you want to delete the Archived record of the patient you
just restored, you can go to the toolbar and select Archive > Delete Selected
Archive Record.
Report only Archive
If you archived patients using the “Report Only” option, you will not be able
to restore those patients as they no longer have the ECG file. Instead, you
can View the patient’s report by clicking on the View button at the bottom of
the screen.
Auto Archive
The Archive process consists of two actions, both of which can be run
automatically on a schedule or initiated by you by clicking the Auto Archive
button from the Archive Window. The two pieces to archiving are:
- Copy a patient from the Patient List to Archive after a defined number of days that the patient has a report created or modified
- Delete the patient from the Patient List after a defined number of days after the patient has been Archived and no subsequent changes to the patient have been made.
Scheduling Auto Archive
You can schedule the Archive program to run in the background. You can do this
by setting up a Scheduled Task from Microsoft’s Control Panel > Administrative
Tools. The command that should be entered where “c:nm” is the location where
the HE/LX Analysis is installed:
“c:\nm\java\bin\java -jar c:\nm\bin\archive.jar -auto”
Java must be installed for the Archive program to run. The Archive.ini file in
that directory can be modified to meet your specific business needs.
Archive.ini
An Archive.ini file is included in the installation in the bin directory. The
directory where the archived files are saved can be changed here.
Additionally, the number of days criteria used to Archive and/or Delete
patients can be found and edited in this file. They are:
- AutoArchiveDays – the number of days after a patient has a report created that the patient will be copied to the archive directory. Make -1 to turn this feature off.
- AutoDeleteDays – The number of days after a patient is archived that the record will be deleted from the Patient List. Make -1 to turn this feature off.
Licensing and Utilities
Licensing
All installations, regardless of license type, must include a license.ini file
in the bin directory
HASP Software Key and HASPFinder Utility
The HASP software key must be plugged into your computer or a visible network,
and a license.ini file must exist for the HE/LX Analysis software to run. The
HASPFinder utility will look up and show you if a NorthEast Monitoring HASP
key is visible to the computer. HASPFinder will also tell you if the HE/LX
Analysis has a license.ini file and if the license.ini file matches the HASP
key.
Settings Utility
The Settings window contains the following information that you can update:
- The name of your facility
- The primary user of the software
- Five lines for the name and address that appear automatically in the Report window
- The language that the software uses
- The number of patients stored in the software
Note: If you choose to reduce the number of patients in the Settings window, be sure that all patients are already backed up before you proceed. Additionally, you will need to use Internet Explorer to delete the empty folders from your patient directory so that they no longer appear in the software.
Go to and Restore Demo Patient buttons
If there is ever a time that you are unable to start the software because of a
problematic patient record, you can use the Go to Demo Patient button to open
the software. If for some reason the demo patient has been deleted, you can
first restore it using the Restore Demo Patient button.
Calculation of Heart Rate
Types of heart rates
A variety of heart rate calculations are made by HE/LX Analysis. They include:
- Current heart rate
- Minute-by-minute heart rate
- Beat-by-beat heart rate
- Mean heart rate in intervals
- Mean heart rate for Holtered period
- Second heart rate
- AF Heart rate
Current heart rate
This is a complex function that takes the current beat and the beats preceding
it into account. This weighted average follows these rules:
- If the differences between the adjacent beats of the preceding four RR intervals are no more than 12 percent of the average RR interval for the previous beat and the beats are all normal, then the new average RR interval is the simple average of the previous four RR intervals.
- If the previous four RR intervals were NOT bigeminy, VTAC or SVT AND the current RR interval is within 25 percent of the previous average AND the previous two beats were not ventricular AND the previous 10 beats were not supraventricular, then the new average RR interval is 1/8 of the current RR interval plus 7/8 of the previous average.
- If the previous four RR intervals were NOT bigeminy, VTAC or SVT AND the current RR interval is not within 25 percent of the previous average OR any of the previous two beats were ventricular OR any of the previous 10 beats were supraventricular, then the new average is 1/32 of the current RR interval plus 31/32 of the previous average.
- If the previous four RR intervals were bigeminy, VTAC or SVT, then the average RR interval is changed by 0.000087 seconds. It is increased if the current interval is longer than the previous interval, otherwise it is decreased.
Once the current average RR interval is determined, the current heart rate is
calculated as 60 divided by the current average RR interval, that is, current
HR = 60/(current RR interval).
The current heart rate is used as the heart rate that appears in the heart
rate data field for any displayed strip. This includes the heart rate
associated with any strip in the Selected Strips window and in the printed
report.
The current heart rate is also used to detect tachycardia and bradycardia. The
onset of either is determined to be when the current heart rate reaches the
tachycardia or bradycardia settings in the Scanning Criteria window.
The low and high heart rates reported in the Tables window and in the tables
of the printed report refer to the lowest and highest current heart rate
calculated during the interval.
Heart Rate calculation limits
The absolute limits are 20 and 300 (3 seconds to 0.2 seconds). If the
“additional dead time” is not set to zero, then this upper limit can be
lowered further. If the “additional dead time” is set to 0.1 or greater, then
the HR upper limit is = 30 / (additional dead time).
Minute-by-minute heart rate
The heart rate plotted in the Trends window is a minute-by-minute heart rate.
It is calculated as 60 times the number of beats processed in the minute
divided by the sum of all RR intervals of beats processed in the minute (in
seconds).
Beat-by-beat heart rate
The heart rate associated with each beat in expanded displays whenever the
Annotation field in the Preferences window is set to Heart Rate is the beat-
by-beat heart rate. It is calculated based on the RR interval following the
labeled beat. Beat-by-beat heart rate equals 60 divided by RR interval, that
is, HR = 60/ (RR).
Mean heart rate in intervals
In the tables (in Tables window and printed report), the mean heart rate
within each interval is calculated by dividing the number of beats in that
interval by the amount of time processed within the interval. Mean heart rate
for Holtered period In the Report Summary (in the Report Summary window and
printed report), the mean heart rate during the Holter test is the number of
beats counted divided by the amount of time processed.
Second heart rate
The second heart rate is the heart rate associated with a run of VTAC or SVT.
It is calculated as 120 divided by the sum of the current RR interval and the
previous RR interval. The second heart rate appears in strips with VTAC or SVT
in Selected Strips, the printed report, and the strip list, labeled HR2.
The second heart rate is used to determine where in the ventricular and
supraventricular run tables a run of VTAC or SVT appears. The heart rate
separating fast from slow runs is determined by the VTAC and SVT settings in
Scanning Criteria, but the rate of each event is considered to be the second
heart rate. The second heart rate is also used to determine which run is
identified as the fastest run of VTAC and SVT.
AF heart rate
The AF heart rate is used to determine the Fastest AF strip and is also the HR
that appears on any AF type strip created by the software. The AF heart rate
is calculated from the average of the RR intervals that are completely within
+- 3.75 seconds of the current beat, which is the same as the standard 7.5
second strip.
Heart rate strips
In the Critical Events window, there is a choice in the Type field called “HR
strips.” This displays all ECG from the Holter test divided into 7.5-second
strips. Each strip includes a time-of-day and a Strip HR. That Strip heart
rate is the total number of RR intervals (including partial ones, but
excluding artifact) within the strip divided by the sum of the RR intervals.
Defining dead-time
Dead-time is the amount of time (in seconds) after a detected QRS complex
during which the software will not look for another QRS complex. Generally,
this helps to prevent the mis-identification of tall T-waves as QRS complexes.
The operator can add more time to the tail end of the dead-time using the
Extra dead-time setting in the Scanning Criteria window. Because the recovery
time (i.e., the width of the T-wave) varies with heart rate, the dead-time
built into the software adjusts based on the current heart rate. At higher
rates, the dead-time decreases, and at lower rates, the dead-time increases.
Test Issues
The following is a list of issues that have been identified in this version of the software.
- When the operating system is set to a region that uses commas instead of periods to reflect decimal points, the Spectral Analysis will be unable to create 24-Hour Plot.
- Report: The page numbers on Saved Strip List do not adjust if a page is inserted or removed before the list.
- Scanning Criteria: Relabel events as artifact only works on reanalysis, but it should work on update too.
- Calibration will not work if the Artifact filter is turned off. The workaround is to keep the Artifact filter on in order to calibrate.
- Sometimes user prompted to close report, but there is no report. On rare occasion, you may need to restart HE/LX.
References