Keysight ECal Modules – Security Features and Volatility Guide
85091 ECal Electronic Calibration Module
This manual provides documentation for the following models:
RF Two-Port: 85091/2/3/6/8/9C/D, 85094D, and N7550/1/2A
RF Four-Port: N4431A/B and N4431D
Microwave Two-Port: N4690A/B/C/D,
N4691A/B/D, N4692/3/4A/D, N4696A/B/
D, and N7553/4/5A
Microwave Four-Port: N4432/3A and N4432/3D
Note: The individual modules may not have the model number listed, but
will instead have a part number based upon it; such as 85092-60007,
N4691-60004, N4433-60003, etcetera.
This is the Security Features and Volatility Guide for the ECal Modules.
Notices
© Keysight Technologies, Inc.
2014-2023
No part of this manual may be reproduced in any form or by any means
(including electronic storage and retrieval or translation into a foreign
language) without prior agreement and written consent from Keysight
Technologies, Inc. as governed by United States and international copyright
laws.
Document Part Number
N4693-90016
Edition
Edition 2, August 2023
Supersedes: February 2023
Printed in USA
Published by:
Keysight Technologies
1400 Fountain grove Parkway
Santa Rosa, CA 95403
Keysight Technologies FieldFox Handheld Analzers ECal Modules
Security Features and Volatility Documentation
Contacting Keysight Sales and Service Offices
Assistance with test and measurements needs and information on finding a local
Keysight office is available on the internet at
http://www.keysight.com/find/assist.
If you do not have access to the internet, please contact your field engineer.
NOTE
In any correspondence or telephone conversation, refer to the product by its
model number and full serial number. With this information, the
Keysight representative can determine whether your unit is still within its
warranty period.
Products Covered by this Document
NOTE
The individual modules may not have the model number listed, but will instead
have a part number based upon it; such as 85092-60007, N4691-60004,
N4433-60003, etcetera.
Table 1
Products Covered by this Document
Product Family Name | Product Names | Model Names |
---|---|---|
Electronic Calibration Module | RF Two-Port | 85091C/D, 85092C/D, 85093C/D, |
85094D, 85096C/D,
85098C/D, 85099C/D
N7550A, N7551A, N7552A
| RF Four-Port| N4431A, N4431B N4431D
| Microwave Two-Port| N4690A, N4690B, N4690C, N4691A, N4691B, N4692A,
N4693A, N4694A, N4696A, N4696B
N4690D, N4691D, N4692D, N4693D, N4694D, N4696D
N7553A, N7554A, N7555A
| Microwave Four-Port| N4432A, N4433A N4432D, N4433D
Contacting Keysight Sales and Service Offices
Products Covered by this Document
Introduction
This document describes instrument security features and the steps to
declassify an instrument through memory sanitization or removal. For
additional information please go to http://www.Keysight.com/find/ad and
security instrument tab. Keysight ECal Modules Security Features and
Volatility Documentation
Contacting Keysight Sales and Service Offices Instrument Memory
Instrument Memory
This section contains information on the types of memory available in your instrument. It explains the size of memory, how it is used, its location, volatility, and the sanitization procedure.
Contacting Keysight Sales and Service Offices
Memory Sanitization Procedures
Memory Sanitization Procedures
NOTE
IMPORTANT! N755xA ECal models require ’ECal Module Security Wipe for the
PNA’ revision 3.50 or later.
Memory Clearing, Sanitization and/or Removal Procedures
This section explains how to clear, sanitize, and remove memory from your
instrument for all memory that can be written to during normal operation.
For Memory Clearing and Sanitization:
NOTE
IMPORTANT! The memory clearing and sanitization process only removes any
user data stored in the ECal memory (i.e., it does NOT remove the factory
calibration).
To clear or sanitize the ECal memory, two ECal Data Wipe Utility has been
developed which will destroy all user data per US DoD 5220.22M. To download
the latest version, click on the URL below.
http://na.support.Keysight.com/pna/apps/applications.htm
Select ECal Module Security Wipe Stand-Alone for PC and run this program on
any windows PC or ENA with firmware revision 9.2 or later. If the ECal is used
with an PNA, then select the ’ECal Module Security Wipe for the PNA’ which
require the PNA to have firmware revision A.03.50 or later. Follow the
instructions on the ECalWipeReadMe document that is contained within the
utility.
For Memory Removal:
This memory can NOT be removed without damaging the ECal.
Learn more, refer to “Procedure for Declassifying a Faulty Instrument” on page
9.
Procedure for Declassifying a Faulty Instrument
Definitions:
Clearing – The process of eradicating the data on media before reusing the
media so that the data can no longer be retrieved using the standard
interfaces on the instrument. Clearing is typically used when the instrument
is to remain in an environment with an acceptable level of protection.
Sanitization – The process of removing or eradicating stored data so that
the data cannot be recovered using any known technology. Instrument
sanitization is typically required when an instrument is moved from a secure
to a non-secure environment such as when it is returned to the factory for
calibration. (The instrument is declassified.) Keysight memory sanitization
procedures are designed for customers who need to meet the requirements
specified by the US Defense Security Service (DSS). These requirements are
outlined in the “Clearing and Sanitization Matrix” issued by the Cognizant
Security Agency (CSA) and referenced in National Industrial Security Program
Operating Manual (NISPOM) DoD 5220.22M ISL 01L-1 section 8301.
Security erase – A term that is used to refer to either the clearing or
sanitization features of Keysight instruments.
Instrument declassification – A term that refers to procedures that must be
undertaken before an instrument can be removed from a secure environment such
as is the case when the instrument is returned for calibration.
Declassification procedures will include memory sanitization and or memory
removal. Keysight declassification procedures are designed to meet the
requirements specified by the DSS NISPOM security document (DoD 5220.22M
chapter 8).
This information is subject to change without notice.
© Keysight Technologies, Inc. 2014-2023
Edition 2, August 2023
N4693-90016
www.keysight.com
Documents / Resources
|
KEYSIGHT 85091 ECal Electronic Calibration
Module
[pdf] User Manual
85091 ECal Electronic Calibration Module, 85091, ECal Electronic Calibration
Module, Electronic Calibration Module, Calibration Module, Module
---|---
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>