DAVID SCOTT 1405 Siemens Koordiant U Angio Table Pad Instruction Manual
- June 17, 2024
- DAVID SCOTT
Table of Contents
DAVID SCOTT 1405 Siemens Koordiant U Angio Table Pad
Specifications
- Product: David Scott Company pads and straps
- Cleaning Protocols: Follow the below instructions
Cleaning Instructions
- Inspect pad before and after use to make sure there are no compromising defects.
- Use disinfectant solution listed below to avoid discoloration:
- – H202 (Hydrogen Peroxide)
- – PAA (Peracetic Acid)
- Follow manufacturer’s instructions for concentrations and use of cleaning products.
- Ensure no excess fluids remain on pads during and after cleaning.
- If recommended cleaning products are not available, use a neutral disinfectant or disinfectant with sodium hypochlorite solution (.1% or 1000 parts per million available chlorine).
Special Considerations
For special considerations such as large volume blood spills, contamination,
or exposure to high-risk tissue and fluids, pads may be cleaned with an
intermediate to high-level distinction dilution
rate of 10% or 1:10 dilution rate of household bleach as recommended by CFC
and AORN guidelines. Please note that this is not recommended for routine use
and requires a damp cloth rinse post-cleaning.
Instructions for use – Cleaning instruction for David Scott
Company Mattresses, Pads, and Straps
Cleaning Instructions
To ensure prolonged use of David Scott Company pads and straps please follow the below cleaning protocols.
- Inspect pad before and after use to make sure there are no compromising defects
- Use of disinfectant solution OTHER than those listed below may cause discoloration or deterioration of pad surface:
- Coverage® Plus Germicidal Surface Wipes Disinfecting/Deodorizing/Cleaning Wipes
- Quaternary Ammonium Compound (Quats) With Ethanol Solvent
- Quaternary Ammonium Compound (Quats) with Isopropyl Alcohol solvent
- Quaternary Ammonium Compound (Quats) with IPA + 2-Butoxyethanol solvent
- Quaternary Ammonium Compound (Quats) + Biguanide
- H202 (Hydrogen Peroxide)
- Neutral Cleaners
- The Use of H202 + PAA (Hydrogen Peroxide + Peracetic Acid) is strongly discouraged
- Allows Follow manufacturer’s instructions for concentrations and use of cleaning products
- Ensure no excess fluids remain on pads during and after cleaning
- If recommended cleaning products are not available in your area, use a neutral disinfectant or disinfectant with sodium hypochlorite solution (.1% or 1000 parts per million available chlorine)
For special considerations such as large volume blood spills, contamination or exposure to high-risk tissue and fluids, pads may be cleaned with an intermediate to high level distinction dilution rate of 10% or 1:10 dilution rate of household bleach as recommended by CFC and AORN guidelines. Please note this is not recommended for routine use and requires a damp cloth rinse post cleaning
Additional Precautions
- Dilute all cleaners/disinfectants and germicides in accordance with manufacturer’s instructions
- Allow adequate drying time before placing pad into service
- Some cleaning agents can leave a residue, it is recommended to rinse and wipe clean with a slightly damp cloth
- Avoid applying strong adhesive tape directly to mattress surface