KEN-A-VISION T-22041 Vision Scope 2 Stereo Microscope Instruction Manual

June 17, 2024
Ken-A-Vision

KEN-A-VISION T-22041 Vision Scope 2 Stereo Microscope

KEN-A-VISION-T-22041-Vision-Scope-2-Stereo-Microscope-
PRODUCT

Specifications

  • Product Name: Vision Scope 2 Stereo Microscope
  • Model : T-22XXX Series T-26001
  • Manufacturer : Ken-A-Vision
  • Website: www.ken-a-vision.com

Product Usage Instructions

Storage and Handling

  1. Store the microscope in a dry and clean environment, away from direct sunlight, high temperatures, and violent vibration.
  2. Handle the microscope with care to avoid impact and abrupt movement during transportation.

Cleaning and Maintenance

  1. Keep all lenses clean by blowing off fine dust with a hand blower or wiping gently with a soft lens tissue.
  2. Clean fingerprints and oil marks on the lens surfaces with a tissue moistened with a 3:7 mixture of alcohol and ether.
  3. Avoid using organic solutions to clean other surfaces, especially plastic surfaces. If necessary, use a neutral detergent.
  4. Do not disassemble the microscope after assembly to prevent damage.
  5. After use, cover the microscope with the provided dust-cover and store it in a dry, clean environment to prevent rust.
  6. Periodically check and clean the microscope to ensure proper performance.

Setting up the Microscope

  1. Place the microscope base on a flat, firm surface and position the frosted stage plate on the base.

Focusing and Observation

  1. Check the coarse focus knob on either side of the main support to ensure it turns easily.
  2. To adjust the degree of tightness of the focusing arm, hold one focusing knob and turn the other to attain a suitable position. Clockwise direction is tight, counterclockwise is loose.
  3. Set the specimen in the center of the stage plate and clamp it if necessary.
  4. Plug the small round plug into the back of the microscope base and connect the opposite end to a power outlet.
  5. Turn the focusing knob and observe the specimen through the right eyepiece until the image is clear.
  6. Observe the specimen through the left eyepiece and adjust the diopter ring until the image is clear.

Adjusting Eyepieces and Light Intensity

  1. Adjust the two eyepieces along the direction indicated until they feel comfortable and a single image is seen.
  2. Use the left rheostat (dimmer knob) to adjust the light intensity of the bottom (transmitted) light.
  3. Use the right rheostat (dimmer knob) to adjust the light intensity of the upper (incidental) light.

Before use

Notice

  1. The microscope should be stored in a dry and clean environment. Do not expose the microscope to the direct sunlight. Avoid high temperatures and violent vibrations.
  2. This microscope is a precision instrument. Handle with care. Avoid impact and abrupt movement during transportation.
  3. To keep the image clear, clean any fingerprints or stains on the surfaces of the lens.
  4. Never turn the left and right focusing knob in the opposite direction at the same time. Microscope damage will occur.

Maintenance

  1. All lenses must be kept clean. Fine dust on the surface of the lens should be blown off with a hand blower or wiped off gently with soft lens tissue. Fingerprints and oil marks should be cleaned with a tissue moistened with a small amount of a 3:7 mixture of alcohol and ether.
  2. Never use organic solutions to clean the other surfaces (especially the plastic surfaces). If necessary, please use a neutral detergent.
  3. Once assembled, do not take the microscope apart as this may cause damage to the microscope.
  4. After use, cover the microscope with the provided dust cover and store it in a dry, clean environment to prevent rust.
  5. To ensure proper performance, please check and clean the microscope periodically.

Nomenclature

KEN-A-VISION-T-22041-Vision-Scope-2-Stereo-Microscope-FIG-
\(1\)

Assembly

KEN-A-VISION-T-22041-Vision-Scope-2-Stereo-Microscope-FIG-
\(2\)

Operation

  1. Position the glass stage
  2. Place the microscope base on a flat, firm surface and position the frosted stage plate on the base. (Figure 1)
  3. Adjust the degree of tightness of the focusing arm
  4. Check the coarse focus knob on either side of the main support. The coarse focus knob should turn easily.
  5. If you want to adjust the degree of tightness of the focusing arm, you can hold one of the focusing knobs and turn another one to attain a suitable position. The degree of tightness relies on the direction to be turned. The clockwise direction is tight; counterclockwise is loose. (Figure 2)
  6. Set the specimen slide
  7. Set the specimen in the center of the stage plate. If necessary, clamp the slide with the clips.
  8. To operate, plug the small round plug into the back of the base of the microscope and then plug the opposite end into the power outlet.
  9. Adjust focusing knobs and diopter
  10. Turn the focusing knob and observe the specimen through the right eyepiece until the image of the specimen is clear.
  11. Observe the specimen through the left eyepiece and adjust the diopter ring 1 until the image is clear. (Figure 3)
  12. Adjust the interpupillary distance
  13. Adjust the two eyepieces along the direction of the arrow shown in Figure 4 until the eyepiece feels comfortable and a single image is seen.
  14. Use the left rheostat (dimmer knob) to adjust the light intensity of the bottom (transmitted) light.
  15. Use the right rheostat (dimmer knob) to adjust the light intensity of upper (incidental) light.

Configuration chart

T-22XXX Series Configuration

Product Number   T-22001 T-22011 T-22021 T-22041 T-22051 T-22061
Magnifications   10x & 30x 15x & 45x 20x & 60x 20x & 40x 30x & 60x 40x

& 80x


Eyepieces

| 2 – 10x| О|  |  | О|  |
2 – 15x|  | О|  |  | О|
2 – 20x|  |  | О|  |  | О
Heads /

Objectives

| 1x & 3x| О| О| О|  |  |
2x & 4x|  |  |  | О| О| О

Note : The items marked “О” are included; others are optional.

T-26001 Series Configuration

Zoom

Magnifications

|  | ****

7x – 45x

| ****

10.5x – 67.5x

| ****

14x – 90x

---|---|---|---|---


Eyepieces

| 2 – 10x| О|  |
2 – 15x|  | О|
2 – 20x|  |  | О
Objective Zoom

Range

| ****

0.7x – 4.5x

| ****

О

| ****

О

| ****

О

Note : The items marked “О” are included; others are optional.

Technical specifications

Note : The items marked “О” are included; others are optional.

T-22XXX Series optical specifications


Objective Mag.

| ****

Working Distance (mm)

| Eyepiece| Eyepieces     (optional)
---|---|---|---
SC6EP10| SC6EP15| SC6EP20


Mag.

| Objective

field

| ****

Mag.

| Objective

field

| ****

Mag.

| Objective

field

1X| ****



100

| 10X| 20| 15X| 15| 20X| 11
2X| 20X| 10| 30X| 7.5| 40X| 5


3X

| ****

30X

| ****

6.7

| ****

45X

| ****

5

| T-22XXX

SERIESX

| ****

3.3


4X

| ****

40X

| ****

5

| T-22XXX

SERIESX

| ****

3.75

| ****

80X

| ****

2.5

Auxiliary objectives for T-22XXX Series


Auxiliary objectives

| ****

Magnification

| Working distance(mm)
---|---|---
SC6OB5| 0.5X| 165
SC6OB15| 1.5X| 45
SC6OB20| 2X| 30

  1. Working distance is fixed regardless of the magnification factor.
  2. Total magnification = Objective mag. X Auxiliary mag. X Auxiliary mag.
  3. Diameter of field of view (mm) = Eyepiece field
    • Objective magnification X Auxiliary objective magnification
  4. Photo adaptor mag. = Objective mag. X Auxiliary objective mag. X Photo eyepiece mag.

Configuration specifications of T-22XXX Series

Model| ****

T-22001

| ****

T-22041

---|---|---








Head

| Objective

magnification

| 1X、3X| 2X、4X
Working distance| 100mm
Observation angle| 45°
Interpupillary

distance adjustment

| Linkage between left and right eyepiece tube range of single

adjustment:     54-75mm


Diopter adjustment

| Range of single adjustment :±5D
 |
 |
Mount with

auxiliary objectives

| Screw hole:M48*0.75
Objective| Field of view| φ20mm




Main body

| Mount the head| Mount the head in the bracket hole (diameter: φ76mm)


Focusing device

| Focus adjustment by turning the focusing knobs.

Range of single adjustment: 10 mm


Glass stage

| Diameter :φ60mm
Clips| Put it on the base from top

Troubleshooting

Trouble Cause Remedy



1.Double images

| Interpupillary distance

correct

| is| not| Adjust  the  interpupillary

distance

Diopter      adjustment

correct

| is| not| ****

Adjust the diopter

Left and right eyepieces are

different magnifications.

| Mount  eyepieces  of  the

same magnification


2.Dirt appears in the field of view

| Dirt on the specimen| ****

Clean the specimen

Dirt on the surface of eyepiece| Clean the surface


3.Image is not clear

| Dirt  on  the  surface

objective

| of| the| ****

Clean the objective


4.Image is not clear while adjusting the focus

| Diopter      adjustment

correct

| is| not| ****

Adjust the diopter

Focus is not correct| Adjust the focus


5.The focusing knob does not turn     smoothly

| ****


The focusing knob is too tight

| ****

Loosen it to a suitable position


6.The image is obscure because of the head is slipping down by during observation

| ****



The focusing knob is too loose

| ****



Tighten it to a suitable position




8.Eyes feel tired easily

| Diopter      adjustment

correct

| is| not| ****

Adjust the diopter

Brightness of light is not correct| Adjust the brightness
The  bulb

correctly

| was  not| inserted| ****

Insert it correctly

Ken-A-VViision Mfg. Co., Inc 5615 Raytown Roa add, Kansas City, MO 64133 USA Phone : (816) 35353-4787 Fax : 816-358-5072 www.ken-a-vision.com

https://manual-hub.com/

FAQ’S

What are the available product numbers for Vision Scope 2 Stereo Microscope?

The available product numbers are T-22001, T-22011, T-22021, T-22041, and T-22051.

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals