KEN-A-VISION T-26001 Vision Scope 2 Stereo Microscope Instruction Manual

June 17, 2024
Ken-A-Vision

KEN-A-VISION T-26001 Vision Scope 2 Stereo Microscope

KEN-A-VISION-T-26001-Vision-Scope-2-Stereo-Microscope-
product

Product Information

Specifications

  • Product Name: Vision Scope 2 Stereo Microscope
  • Model: T-22XXX Series T-26001
  • Manufacturer: Ken-A-Vision
  • Website: www.ken-a-vision.com

Frequently Asked Questions

  • What should I do if there are fingerprints or stains on the lens surfaces?
    • Clean the lens surfaces with a tissue moistened with a 3:7 mixture of alcohol and ether.
  • Can I use organic solutions to clean the plastic surfaces of the microscope?
    • No, organic solutions should not be used on plastic surfaces. If necessary, use a neutral detergent.
  • Should I disassemble the microscope after use?
    • No, it is not recommended to disassemble the microscope as it may cause damage.
  • How should I store the microscope after use?
    • Cover the microscope with the provided dust-cover and store it in a dry, clean environment to prevent rust.
  • How often should I check and clean the microscope?
    • The microscope should be checked and cleaned periodically to ensure proper performance.

Before Use

Notice

  1. Microscope should be stored in a dry and clean environment. Do not expose the microscope to the direct sunlight. Avoid high temperatures and violent vibration.
  2. This microscope is a precision instrument. Handle with care. Avoid impact and abrupt movement during transportation.
  3. To keep the image clear, clean any fingerprints or stains on the surfaces of the lens.
  4. Never turn the left and right focusing knob in the opposite direction at the same time. Microscope damage will occur.

Maintenance

  1. All lenses must be kept clean. Fine dust on the surface of the lens should be blown off with hand blower or wiped off gently with a soft lens tissue. Fingerprints and oil marks should be cleaned with a tissue moistened with a small amount of a 3:7 mixture of alcohol and ether.
  2. Never use organic solution to clean the other surfaces (especially the plastic surfaces). If necessary, please use a neutral detergent.
  3. Once assembled, do not take the microscope apart as this may cause damage to the microscope.
  4. After use, cover the microscope with the provided dust-cover and store in a dry, clean environment to prevent rust.
  5. To ensure proper performance, please check and clean the microscope periodically.

Nomenclature

KEN-A-VISION-T-26001-Vision-Scope-2-Stereo-Microscope-
fig-1

Assembly

KEN-A-VISION-T-26001-Vision-Scope-2-Stereo-Microscope-
fig-2

Operation

Position the glass stage

  1. Place the microscope base on a flat, firm surface and position the frosted stage plate on the base. (Figure 1)

Adjust the degree of tightness of the focusing arm

  1. Check the coarse focus knob on either side of the main support. The coarse focus knob should turn easily.
  2. If you want to adjust the degree of tightness of the focusing arm, you can hold one of the focusing knobs and turn another one to attain a suitable position. The degree of tightness relies on the direction to be turned. The clockwise direction is tight; counterclockwise is loose. (Figure 2)

Set the specimen slide

  1. Set the specimen in the center of the stage plate. If necessary, clamp the slide with the clips.
  2. To operate, plug the small round plug into the back of the base of the microscope and then plug the opposite end into the power outlet.

Adjust focusing knobs and diopter

  1. Turn the focusing knob and observe the specimen through the right eyepiece until the image of the specimen is clear.
  2. Observe the specimen through the left eyepiece and adjust the diopter ring 1 until the image is clear. (Figure 3)

Adjust the interpupillary distance

  1. Adjust the two eyepieces along the direction of the arrow shown in Figure 4 until eyepiece feels comfortable and a single image is seen.
  2. Use the left rheostat (dimmer knob) to adjust light intensity of the bottom (transmitted) light.
  3. Use the right rheostat (dimmer knob) to adjust light intensity of upper (incidental) light.

Configuration Chart

T-22XXX Series Configuration

Product Number   T-22001 T-22011 T-22021 T-22041 T-22051 T-22061
Magnifications   10x & 30x 15x & 45x 20x & 60x 20x & 40x 30x & 60x 40x

& 80x


Eyepieces

| 2 – 10x| О|  |  | О|  |
2 – 15x|  | О|  |  | О|
2 – 20x|  |  | О|  |  | О
Heads /

Objectives

| 1x & 3x| О| О| О|  |  |
2x & 4x|  |  |  | О| О| О

Note: The items marked “О” are included; others are optional.

T-26001 Series Configuration

Zoom Magnifications   7x – 45x 10.5x – 67.5x 14x – 90x

Eyepieces

| 2 – 10x| О|  |
2 – 15x|  | О|
2 – 20x|  |  | О
Objective Zoom

Range

| ****

0.7x – 4.5x

| ****

О

| ****

О

| ****

О

Note: The items marked “О” are included; others are optional.

Technical specifications

T-22XXX Series optical specifications


Objective Mag.

| ****

Working Distance (mm)

| Eyepiece| Eyepieces     (optional)
---|---|---|---
SC6EP10| SC6EP15| SC6EP20


Mag.

| Objective

field

| ****

Mag.

| Objective

field

| ****

Mag.

| Objective

field

1X| ****



100

| 10X| 20| 15X| 15| 20X| 11
2X| 20X| 10| 30X| 7.5| 40X| 5


3X

| ****

30X

| ****

6.7

| ****

45X

| ****

5

| T-22XXX

SERIESX

| ****

3.3


4X

| ****

40X

| ****

5

| T-22XXX

SERIESX

| ****

3.75

| ****

80X

| ****

2.5

Auxiliary objectives for T-22XXX Series

Auxiliary objectives Magnification Working distance(mm)
SC6OB5 0.5X 165
SC6OB15 1.5X 45
SC6OB20 2X 30
  • Working distance is fixed regardless of the magnification factor.
  • Total magnification = Objective mag. X Auxiliary mag. X Auxiliary mag.
  • Diameter of field of view (mm) =KEN-A-VISION-T-26001-Vision-Scope-2-Stereo-Microscope-fig-7
  • Photo adaptor mag. = Objective mag. X Auxiliary objective mag. X Photo eyepiece mag.

Configuration specifications of T-22XXX Series

Model| ****

T-22001

| ****

T-22041

---|---|---








Head

| Objective

magnification

| 1X、3X| 2X、4X
Working distance| 100mm
Observation angle| 45°
Interpupillary

distance adjustment

| The linkage between left and right eyepiece tube range of single adjustment: 54-75mm


Diopter adjustment

| Range of single adjustment:±5D
 |
 |
Mount with

auxiliary objectives

| Screw hole: M48*0.75
Objective| Field of view| φ20mm




Main body

| Mount the head| Mount the head in the bracket hole (diameter: φ76mm)


Focusing device

| Focus adjustment by turning the focusing knobs.

Range of single adjustment: 10 mm


Glass stage

| Diameter:φ60mm
Clips| Put it on the base from top

Troubleshooting

Trouble Cause Remedy



1.Double images

| Interpupillary distance

correct

| is| not| Adjust the interpupillary

distance

Diopter      adjustment

correct

| is| not| ****

Adjust the diopter

Left and right eyepieces are

different magnifications.

| Mount eyepieces of the

same magnification


2.Dirt appears in the field of view

| Dirt on the specimen| ****

Clean the specimen

Dirt on the surface of eyepiece| Clean the surface


3.Image is not clear

| Dirt on the surface

objective

| of| the| ****

Clean the objective


4.Image is not clear while adjusting the focus

| Diopter      adjustment

correct

| is| not| ****

Adjust the diopter

Focus is not correct| Adjust the focus


5.The focusing knob does not turn     smoothly

| ****


The focusing knob is too tight

| ****

Loosen it to a suitable position


6.The image is obscure because of the head  is slipping down by during observation

| ****



The focusing knob is too loose

| ****



Tighten it to a suitable position




8.Eyes feel tired easily

| Diopter      adjustment

correct

| is| not| ****

Adjust the diopter

Brightness of light is not correct| Adjust the brightness
The bulb

correctly

| was not| inserted| ****

Insert it correctly

Contact