FLIR IM75-2 Insulation Tester Digital Multi Meter User Manual

June 16, 2024
FLIR

IM75-2 Insulation Tester Digital Multi Meter

Product Information: FLIR MODEL IM75-2 Insulation Tester /

Digital MultiMeter

Specifications

  • Certifications: FCC Compliance, Industry Canada Compliance

  • Key Features: Insulation resistance measurements, Voltage
    measurements, Continuity test, Diode test, Capacitance
    measurements, and more

  • Meter Sections: Function Switch Positions, Function Buttons,
    Selector Pad, Rotary Switch

  • Display: Description, Icons and Indicators

  • Power: Auto Power Off feature

  • Measurement Modes: PI/DAR Test, Earth Bond Resistance
    Measurements (ZERO), VFD Mode, Frequency Mode, Pass/Fail Mode, Data
    Recording Mode (99-point)

  • Maintenance: Cleaning and Storage, Battery Replacement, Fuse
    Replacement

  • Disposal: Environmentally friendly disposal in accordance with
    regulations for electronic waste

Usage Instructions

1. Powering the Meter

To power on the FLIR MODEL IM75-2 Insulation Tester / Digital
MultiMeter, press the power button located on the device. The meter
will display the default setting upon startup.

1.1 Auto Power Off

The meter is equipped with an Auto Power Off feature that
automatically turns off the device after a period of inactivity. To
disable this feature, navigate to the setup utility and adjust the
settings accordingly.

2. Manual AC/DC Selection

The meter allows you to manually select AC or DC mode for
measurements. Use the function buttons or rotary switch to navigate
through the options and select the desired mode.

3. Auto Range

The Auto Range feature allows the meter to automatically select
the appropriate measurement range based on the detected signal.
This ensures accurate readings without the need for manual range
adjustments.

4. ‘Smart’ Out-of-Range Warnings

The FLIR MODEL IM75-2 provides ‘Smart’ Out-of-Range warnings
when the measured value exceeds the selected range. This helps
prevent incorrect readings and potential damage to the device.

5. Data Hold and Auto-Hold Modes

5.1 Data Hold Mode

The Data Hold mode allows you to freeze the displayed
measurement value on the screen. Press the Data Hold button to
activate this mode, and press it again to release the hold and
resume live measurements.

5.2 Auto-Hold mode

In Auto-Hold mode, the meter automatically holds the measurement
value on the screen when a stable reading is detected. This mode is
especially useful for capturing fluctuating measurements.

6. Voltage Measurements

Use the meter to measure voltage by selecting the appropriate
mode and connecting the test leads to the voltage source. Ensure
proper contact and follow safety guidelines while performing
voltage measurements.

7. Insulation Resistance Measurements

7.1 PI/DAR Test

The FLIR MODEL IM75-2 allows you to perform PI/DAR tests to
measure insulation resistance. Select the PI/DAR test mode, follow
the on-screen instructions, and connect the test leads to the
insulation to obtain accurate measurements.

8. Earth Bond Resistance Measurements (ZERO)

To measure earth bond resistance, select the ZERO mode on the
meter. Follow the on-screen instructions and connect the test leads
to the earth bond point. The meter will display the resistance
value.

9. Continuity Test

Perform continuity tests by selecting the continuity mode on the
meter. Connect the test leads to the circuit or component under
test. The meter will emit an audible beep if continuity is
detected.

10. Diode Test

Use the diode test mode to test the forward and reverse bias of
diodes. Connect the test leads to the diode and follow the
on-screen instructions. The meter will display the voltage drop
across the diode.

11. Capacitance Measurements

For capacitance measurements, select the capacitance mode on the
meter. Connect the test leads to the capacitor terminals and follow
the on-screen instructions. The meter will display the capacitance
value.

12. Extended Functionality

12.1 Selecting Modes

The FLIR MODEL IM75-2 offers various extended functionality
modes. Use the function buttons, selector pad, or rotary switch to
navigate through the available modes and select the desired
one.

12.2 VFD Mode (ACV only)

In VFD mode, the meter is optimized for measuring AC voltage
signals from variable frequency drives. Select this mode when
working with such signals for accurate readings.

12.3 Frequency Mode (ACV only)

The frequency mode allows you to measure the frequency of AC
voltage signals. Select this mode and connect the test leads to the
source to obtain frequency measurements.

12.4 Pass/Fail Mode (COMPARE)

The Pass/Fail mode enables you to compare measured values
against pre-set thresholds. Set the desired thresholds in the setup
utility and use this mode to quickly determine if the measurement
passes or fails.

12.5 Setup Utility

The setup utility provides access to various settings and
customization options for the meter. Navigate through the utility
using the function buttons, selector pad, or rotary switch, and
make the desired adjustments.

12.6 Silent Mode

The Silent mode allows you to mute the audible beep produced by
the meter during certain operations. Enable or disable this mode in
the setup utility as per your preference.

12.7 Data Recording Mode (99-point)

In Data Recording mode, the meter can record up to 99
measurement points. Activate this mode and perform measurements as
required. The recorded data can be accessed and reviewed later.

13. Keypad Lockout

The keypad lockout feature prevents accidental button presses
and changes to the settings. To activate or deactivate this
feature, navigate to the setup utility and adjust the keypad
lockout settings accordingly.

14.1 General

METERLiNK is a wireless communication feature that allows the
FLIR MODEL IM75-2 to stream measurement data to compatible devices
or software. This enables real-time monitoring and data
logging.

14.2 Procedure

To utilize METERLiNK, ensure that the meter and the receiving
device or software both support this feature. Follow the specific
instructions provided by the manufacturer of the receiving device
or software to establish a connection and start streaming
measurement data.

Maintenance

1. Cleaning and Storage

Regularly clean the FLIR MODEL IM75-2 with a soft, dry cloth to
remove dirt and debris. Avoid using solvents or abrasive cleaners.
Store the meter in a cool, dry place when not in use to protect it
from damage.

2. Battery Replacement

To replace the battery, follow the instructions provided in the
user manual. Ensure that the replacement battery meets the
specified requirements and observe proper polarity during
installation.

3. Fuse Replacement

If the meter’s fuse needs replacement, consult the user manual
for detailed instructions. Only replace the fuse with the
appropriate type and rating to ensure safe and accurate
operation.

4. Disposal of Electronic Waste

When disposing of the FLIR MODEL IM75-2 or any electronic waste,
adhere to environmentally friendly practices and comply with
existing regulations for electronic waste disposal. Contact local
authorities or recycling centers for proper disposal methods.

FAQ

Q: Where can I access the latest manuals and notifications for

the FLIR MODEL IM75-2?

A: To access the latest manuals and notifications, visit the
Download tab at https://support.flir.com. Register
online to gain access to the download area, where you can find the
latest releases of manuals for this product and other FLIR
products.

Q: Does the FLIR MODEL IM75-2 have a warranty?

A: Yes, FLIR offers a limited lifetime warranty for the FLIR
MODEL IM75-2 Insulation Tester / Digital MultiMeter. For more
details on the warranty terms and conditions, refer to the warranty
section in the user manual.

USER MANUAL
FLIR MODEL IM75-2 Insulation Tester / Digital MultiMeter

Table of Contents
1. DISCLAIMERS
1.1 Copyright 1.2 Quality Assurance 1.3 Documentation 1.4 Disposal of Electronic Waste
2. SAFETY
2.1 FCC Compliance 2.2 Industry Canada Compliance
3. INTRODUCTION
3.1 Key Features
4. METER DESCRIPTION
4.1 Meter Sections 4.2 Function Switch Positions 4.3 Function Buttons, Selector Pad, and Rotary Switch
4.3.1 Selector Pad 4.3.2 Rotary Switch 4.4 Display Description 4.5 Display Icons and Indicators
5. OPERATION
5.1 Powering the Meter 5.1.1 Auto Power Off
5.2 Manual AC/DC Selection 5.3 Auto Range 5.4 `Smart’ Out-of-Range Warnings 5.5 Data Hold and Auto-Hold Modes
5.5.1 Data Hold Mode 5.5.2 Auto-Hold mode 5.6 Voltage Measurements 5.7 Insulation Resistance Measurements 5.7.1 PI/DAR Test 5.8 Earth Bond Resistance Measurements (ZERO )
FLIR IM75-2 USER MANUAL 2

4
4 4 4 4
5
6 7
8
8
9
9 10 10 11 11 12 13
14
14 14 14 14 14 14 14 14 15 15 15 16
IM75-2-UM_en-US Rev AA

5.9 Continuity Test

16

5.10 Diode Test

16

5.11 Capacitance Measurements

17

5.12 Extended Functionality

17

5.12.1 Selecting Modes

17

5.12.2 VFD Mode (ACV only)

18

5.12.3 Frequency Mode (ACV only)

18

5.12.4 Pass/Fail Mode (COMPARE)

18

5.12.5 Setup Utility

18

5.12.6 Silent Mode

19

5.12.7 Data Recording Mode (99-point)

19

5.13 Keypad Lockout

19

5.14 Streaming Measurement Data using METERLiNK

20

5.14.1 General

20

5.14.2 Procedure

20

6. MAINTENANCE

21

6.1 Cleaning and Storage

21

6.2 Battery Replacement

21

6.3 Fuse Replacement

21

6.4 Disposal of Electronic Waste

21

7. SPECIFICATIONS

22

7.1 General specifications

22

7.2 Electrical Range Specifications

23

7.3 Input Specifications

25

8. CUSTOMER SUPPORT

26

9. WARRANTY

27

9.1 FLIR Limited Lifetime Warranty

27

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 3

1. Disclaimers
1.1 Copyright
© 2023, FLIR Systems, Inc. All rights reserved worldwide. No parts of the software including source code may be reproduced, transmitted, transcribed, or translated into any language or computer language in any form or by any means, electronic, magnetic, optical, manual, or otherwise, without the prior written permission of FLIR Systems.
The documentation must not, in whole or part, be copied, photocopied, reproduced, translated, or transmitted to any electronic medium or machine- readable form without prior consent, in writing, from FLIR Systems.
Names and marks appearing on the products herein are either registered trademarks or trademarks of FLIR Systems and/or its subsidiaries. All other trademarks, trade names or company names referenced herein are used for identification only and are the property of their respective owners.
1.2 Quality Assurance
The Quality Management System under which these products are developed and manufactured has been certified in accordance with the ISO 9001 standard.
FLIR Systems is committed to a policy of continuous development; therefore, we reserve the right to make changes and improvements on any of the products without prior notice.
1.3 Documentation
To access the latest manuals and notifications, go to the Download tab at: https://support.flir.com. It only takes a few minutes to register online. In the download area you will also find the latest releases of manuals for our other products, as well as manuals for our historical and obsolete products.
1.4 Disposal of Electronic Waste
As with most electronic products, this equipment must be disposed of in an environmentally friendly way, and in accordance with existing regulations for electronic waste.
Please contact your FLIR Systems representative for more details.

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 4

2. Safety
Safety Notes
· Before operating the device, you must read, understand, and follow all instructions, dangers, warnings, cautions, and notes.
· FLIR Systems reserves the right to discontinue models, parts or accessories, and other items, or to change specifications at any time without prior notice.
· Remove the batteries if the device is not used for an extended period.
Warning Statements
· Do not operate the device if you do not have the correct knowledge. Formal qualifications and/or national legislation for the electrical inspections can apply. Incorrect operation of the device can cause damage, shock, injury or death to persons.
· Do not start the measuring procedure before you have set the function switch to the correct position. This can cause damage to the instrument and can cause injury to persons.
· Do not change to resistance when you measure the voltage. This can cause damage to the instrument and can cause injury to persons.
· You must disconnect the test leads from the circuit that you did a test on before you change the range. If you do not do this, damage to the instrument and injury to persons can occur.
· Do not replace the batteries before you remove the test leads. This can cause damage to the instrument and can cause injury to persons.
· Do not use the device if the test leads and/or the device show signs of damage. Injury to persons can occur.
· Be careful when you do the measurements if the voltages are more than 25 VAC rms or 35 VDC. There is a risk of shock from these voltages. Injury to persons can occur.
· Do not do diode, resistance, or continuity tests before you have removed the power from capacitors and other devices under test during a measurement. Injury to persons can occur.
· Do not use the device as a tool to identify live terminals. You must use the correct tools. Injury to persons can occur if you do not use the correct tools.
· Make sure that children cannot touch the device. The device contains dangerous objects and small parts that children can swallow. If a child swallows an object or a part, speak with a physician immediately. Injury to persons can occur.
· Do not let children play with the batteries and/or the packing material. These can be dangerous for children if they use them as toys.
· Do not touch expired or damaged batteries without gloves. Injury to persons can occur. · Do not cause a short-circuit of the batteries. This can cause damage to the instrument and can
cause injury to persons. · Do not put the batteries into a fire. Injury to persons can occur.

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 5

Cautions
Do not use the device for a procedure that it is not made for. This can cause damage to the protection.
This symbol, adjacent to another symbol or terminal, indicates that the user must refer to the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present.
Double insulation.
UL listing is not an indication or a verification of the accuracy of the meter
2.1 FCC Compliance
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
1. This device may not cause harmful interference. 2. This device must accept any interference received, including interference that may cause
undesired operation.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in an installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: 1. Reorient or relocate the receiving antenna. 2. Increase the separation between the equipment and receiver. 3. Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected. 4. Consult the dealer or an experienced radio/TV technician for help.

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 6

WARNING
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
2.2 Industry Canada Compliance
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 7

3. Introduction
Thank you for selecting the FLIR IM75-2 Insulation Test / Digital MultiMeter. This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.
3.1 Key Features
· 4000 count extra-large digital dual display with backlighting · Auto Range with smart over-range indication · Auto Hold and Standard Data Hold features · On-screen menu selection and navigator key drive · Variable-frequency drive mode (low-pass filter) · Front panel insulation resistance test voltage selector · COMPARE feature with front panel color coded PASS/FAIL LED indication · 0.1% DCV accuracy · 99-record data record/recall memory · High power built-in work lights · METERLiNK® app compatibility · Safety Category Rating: CAT IV-600V, CAT III-1000V

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 8

4. Meter Description
4.1 Meter Sections
1. Courtesy Work lights 2. LCD Display 3. Pass-Fail (green/red) LED 4. Rotary Function Switch 5. Function Buttons (detailed below) 6. Positive (+) probe input jack for Voltage,
Capacitance, Diode, & Continuity 7. COM (negative -) probe input jack 8. Positive (+) probe input jack for insulation
resistance and earth bond resistance

Fig 4-1 Front View

FLIR IM75-2 USER MANUAL

1. Courtesy Work Lights 2. Test Probe Holders 3. Tilt Stand 4. Battery and Fuse Access
Fig 4-2 Rear View
IM75-2-UM_en-US Rev AA 9

4.2 Function Switch Positions

OFF

The meter is switched OFF and in full power-saving mode.

The meter can measure voltage (V) through the probe inputs.

The meter can measure capacitance, diode polarity or continuity through the probe inputs. The type of measurement is chosen by the MODE button.

ZERO Earth Bond Resistance testing mode. 50V, 100V, 250V, 500V, 1KV: Insulation resistance test voltage selections

4.3 Function Buttons, Selector Pad, and Rotary Switch

MODE HOLD

· Use this button to change the operating mode of the meter (ex: AC vs. DC) · Also selects Insulation Resistance values to compare for PASS/FAIL testing
Press to HOLD readings on the display during normal DMM measuring modes (see Section 5.5 Data Hold and Auto-Hold Modes). Press and hold for Keypad lockout mode (see Section 5.13 Keypad Lockout). In INSULATION TESTING MODE this button will serve to enable the test lock function for continuous insulation resistance testing. See Section 5.7 Insulation Resistance Measurements for further information.
Use this Selector Pad to start/stop Insulation Resistance Tests, to enable extended functionality modes, and to navigate mode menu options. Further details provided in Section 4.3.1 Selector Pad

The Rotary Switch is used to select a measurement function. Details are provided in Section 4.3.2 Rotary Switch

Press this button to exit an extended functionality mode.
· Press this button to enable/disable the display backlight. · Press and hold the button for 2 seconds to enable/disable the work light.
Press to enable/disable METERLiNK communication, see section 5.14 Streaming measurement data using METERLiNK.

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 10

4.3.1 Selector Pad
There are five (5) `soft’ function buttons arranged in a square, as shown below. The functions change depending on the sub-function chosen at the time. Generally, the arrow buttons are used for navigation and the TEST-OK button is used for selecting an object or starting a test. Detailed information below:

FIMig745-3SeSleelcetcotor rPPaadd
Pressing the TEST OK button selects a menu-driven option. The TEST OK button is also used to enter the sub-menus for selection screens.
NOTE: When in INSULATION TESTING MODE pressing and holding TEST-OK outputs a test voltage (until the button is manually released). In Test Lock mode this button (used in conjunction with the HOLD/LOCK button) is used to Start and Stop continuous Insulation Resistance tests. See Section 5. Insulation Resistance Measurements for further details. Do not confuse the Insulation Resistance test lock feature with the Keypad Lockout feature (section 5.13).
The LEFT () button scrolls to previous choices in the current level of the menu. When reaching the start of a group of selections, further presses of the LEFT button wraps to the end of the list. The RIGHT () button scrolls to next choices in the current level of the menu. When reaching the end of a group of selections, further presses of the RIGHT button wraps to the start of the list. The UP () button scrolls to previous selectable options in the current level of the menu. When reaching the start of a group of selections, further presses of the UP button are ignored. The DOWN () button scrolls to previous selectable options in the current level of the menu. When reaching the end of a group of selections, further presses of the DOWN button are ignored.
4.3.2 Rotary Switch
There are nine (9) dedicated Rotary Switch positions:
· OFF: The meter is switched OFF.
· AC/DC VOLTAGE: The meter can measure voltage through the probe inputs. The type of measurement (AC/DC) is chosen by the MODE button. The default measurement is AC voltage.

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 11

· CAPACITANCE/DIODE/CONTINUITY: The meter can measure capacitance, diode polarity or continuity through the probe inputs. The type of measurement is chosen by the MODE button.
· Zero : The meter can measure earth-bond resistance, through the probe inputs. The measurement unit is Ohms ().
· 50V: The meter can source a 50VDC TEST Voltage and measure Insulation Resistance through the Zero /Insulation and COM probe inputs.
· 100V: The meter can source a 100VDC TEST Voltage and measure Insulation Resistance through the Zero /Insulation and COM probe inputs.
· 250V: The meter can source a 250VDC TEST Voltage and measure Insulation Resistance through the Zero /Insulation and COM probe inputs.
· 500V: The meter can source 500VDC TEST Voltage and measure Insulation Resistance through the Zero /Insulation and COM probe inputs.
· 1000V: The meter can source 1000VDC TEST Voltage and measure Insulation Resistance through the Zero /Insulation and COM probe inputs.
4.4 Display Description

Fig 4-4 Display
1. Main display area 2. Secondary display area 3. Bar graph (matches the reading on the main display) 4. Modes of operation (detailed later in this guide) 5. Units of measure for primary and secondary displays 6. Alerts and status icons (detailed below)

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 12

4.5 Display Icons and Indicators

TEST H

Indicates that Bluetooth communication is active Indicates an Insulation Test in progress and test voltage is present Indicates that the measured voltage is greater than 30 V (AC or DC) Indicates that the meter is in Hold mode (solid H) or auto hold (flashing H) Indicates that the meter is in locked mode

APO

Memory icon shown with the active memory location (1-99) Indicates the battery voltage status Indicates that the auto power off function is enabled Indicates that the meter is measuring AC voltage Indicates that the meter is measuring DC voltage Analog bar graph VFD (variable frequency drive) mode icon Frequency mode icon

Earth Bond Resistance icon

INSUL COMP
PI DAR

Insulation Resistance icon Compare function icon Polarization Index icon Dielectric Absorption Ratio icon Settings mode icon Silent mode icon

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 13

5. Operation
Note: Before operating the device, you must read, understand, and follow all instructions, dangers, warnings, cautions, and notes.
Note: When the meter is not in use, the function switch should be set to the OFF position.
Note: When connecting the probe leads to the device under test, connect the COM (negative) lead before connecting the positive lead. When removing the probe leads, remove the positive lead before removing the COM (negative) lead.
5.1 Powering the Meter
1. Set the function switch to any position to switch on the meter. 2. If the battery indicator shows that the battery voltage is low or if the meter does not power on, replace the battery. See section 6.2 Battery Replacement.
5.1.1 Auto Power Off
The meter enters sleep mode after 30 minutes of inactivity. The meter beeps three times 20 seconds before powering off. Press any button or turn the Rotary Switch to prevent the meter from powering off. The auto power off time- out is then reset. To disable auto power off (APO) press the MODE button while turning on the meter. To change the APO time, read section 5.12.6 Setup Utility
5.2 Manual AC/DC Selection
When measuring voltage, press the MODE button to select AC or DC voltage.
5.3 Auto Range
Auto-Range is the only mode of operation for Capacitance; ranges are automatically selected by the meter. Insulation Resistance tests allow for voltage test selection using the rotary switch.
5.4 Smart’ Out-of-Range Warnings If the input is over/under the full-scale range, the display will show >x or <x, where x is the high or low end of the full-scale range. 5.5 Data Hold and Auto-Hold Modes These two modes are discussed in detail below. In the Settings menu, if Auto- Hold is set to OFF, the meter will operate in the standard Hold mode. If Auto- Hold is set to ON in the settings menu, the meter will operate in the Auto- Hold mode. 5.5.1 Data Hold Mode In Data Hold mode, press the HOLD button to capture the measurement that is shown on the main display. The captured reading will be displayed in the secondary, upper display area (the main display will then function normally). TheH’ display icon will appear solid on the LCD in the Hold mode. To exit the Hold mode, press the HOLD button again.
5.5.2 Auto-Hold mode
If Auto-Hold mode is switched ON in the settings menu, press the HOLD button to activate Auto-Hold mode (the `H’ display icon will appear flashing). In Auto-Hold mode readings are

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 14

captured in the say way as in standard Hold mode. The difference is that the captured reading, displayed on the smaller, secondary display digits, will update whenever a reading 50 digits higher or lower than the captured reading is measured. To exit the Auto-Hold mode press the HOLD button.
5.6 Voltage Measurements
1. Set the function switch to the position. 2. Insert the black probe lead into the COM terminal and red probe lead into the V terminal. 3. Use the MODE button to select AC or DC voltage measurements.
· The indicator will be displayed for AC measurements. · The indicator will be displayed for DC measurements. 4. Connect the probe leads in parallel to the part under test 5. Read the voltage value on the display
5.7 Insulation Resistance Measurements
1. Use the rotary switch to select the insulation voltage (50V, 100V, 250V, 500V, or 1kV). 2. Insert the black probe lead into the COM terminal and red probe lead into the Insulation
terminal. 3. Select INSUL to enable the Insulation Resistance Function mode as described in section 5.12.1
Selecting modes. 4. During the Test, the insulation voltage will appear in the secondary (upper) reading area. The
resistance will be shown in the main reading area. 5. Press TEST OK to start test. 6. To lock the test for continuous testing, press the HOLD/LOCK button (lock icon will appear) and
then press TEST OK to test continuously. 7. The TEST icon on the LCD will be shown while the voltage is outputted. 8. Press TEST OK again to stop the test
NOTE: The 99-point Data Record memory function cannot be activated in the usual way while using the Insulation Resistance function (since the TEST OK button is used for testing in this case and not for saving readings). The last reading, however, is held on the display and can then be saved in the normal way as described in the Extended Functionality section below under Section 5.12.7 Data Record Mode.
5.7.1 PI/DAR Test
Note: PI (Polarization Index); DAR (Dielectric Absorption Ratio) 1. Select the PI or DAR test as described in section 5.12.1 Selecting modes. 2. Press the TEST OK button to test. 3. Press the MODE button to show the test time at the upper reading area. (LCD will show
CLOCK icon and the elapsed time). PI is a ten-minute test and DAR is a one- minute test. 4. The test will stop automatically, and the result (ratio) will be shown on the main display
area. The PASS/FAIL LED will indicate the test result (Green for PASS, and Red for FAIL).

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 15

5.8 Earth Bond Resistance Measurements (ZERO )

1. Move the function dial to the Earth-bond resistance position (ZERO ); the Earth-bond

resistance icon

will appear.

2. Insert the black probe lead into the COM terminal and red probe lead into the ZERO/Ohm/ Insulation terminal.
3. Before testing, short the ends of the probes and press the MODE button. The ZERO icon will appear on the display and the offset value will be stored. If the test leads are not shorted or if there is residual resistance, the display will show the > 2.00 ohms (this indication could be due to faulty leads, improper test lead insertion, or meter problem. Tests should not be attempted until root cause of the high resistance is determined.
4. Connect the test leads to the Device under test. 5. Press TEST OK to run a test.

6. To lock the test ON, press the HOLD/LOCK button (the LOCK icon will appear) and then

press TEST OK to test continuously. 7. Press TEST OK again to stop the test. 8. Note that the Earth Bond Resistance Measurement mode can also be used to measure
standard resistance to 40kohms

NOTE: The 99-point Data Record memory function cannot be activated in the usual way while using the Earth Bond Resistance function (since the TEST OK button is used for testing in this case and not for saving readings). The last reading, however, is held on the display and can then be saved as described in the Extended Functionality section below under Section 5.12.7 Data Recording Mode.
5.9 Continuity Test
Warning: Do not do diode, resistance or continuity tests before removing the power from capacitors and other devices under test during a measurement. Injury to persons can occur.

1. Set the function switch to the position. 2. Insert the black probe lead into the COM terminal and red probe lead into the V terminal.
3. Use the MODE button to select continuity measurement. The indicator will be displayed. 4. Insert the black probe lead into the COM (negative) terminal and the red probe lead into the
positive terminal. 5. Touch the tips of the probe across the circuit or component under test. 6. If the resistance is 30 ± 5 (nominal) or less, the meter will beep.

Note: This threshold is user selectable in the SET UP menu under the Cntin (continuity) setting: Range: 10-50 ; Increment: 1; Default: 30 .
5.10 Diode Test
Warning: Do not do diode, resistance, or continuity tests before you have removed the power from capacitors and other devices under test during a measurement. Injury to persons can occur.
1. Set the function switch to the position.

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 16

2. Use the MODE button to select the diode test function. The indicator will be displayed. 3. Insert the black probe lead into the COM (negative) terminal and the red probe lead into the
positive V terminal. 4. Touch the tips of the probes across the diode or semiconductor junction under test. Make a
note of the value on the display. 5. Reverse the polarity of the probes, by interchanging the probe test locations. 6. Touch the tips of the probe across the diode or semiconductor junction under test. Make a
note of the new value on the display. 7. The diode or semiconductor junction can be evaluated as follows:
If one of the readings displays a value (Typically 0.4.00 to 0.900 V) and the other reading displays a value (-0.400 to -0.900 V), the component is good.
If both readings display OL, the component is either open or shorted. Use Continuity mode to determine if it is open or shorted.
5.11 Capacitance Measurements
Warning: Do not do capacitance tests before removing the power to the capacitor or other devices under test during a measurement. Injury to persons can occur.
1. Set the function switch to the position. 2. Use the MODE button to select the capacitance measurement. The F (Farad) unit will be
displayed. 3. Insert the black probe lead into the COM (negative) terminal and the red probe lead into the
positive terminal. 4. Touch the tips of the probe across the part under test. 5. Read the capacitance value on the display.
Note: For very large capacitance values, it may take several seconds for the measurement to settle and the final reading to stabilize.
5.12 Extended Functionality
A variety of modes are available that provide extended functionality as detailed below.
5.12.1 Selecting Modes
The mode icons applicable for the selected measurement type are displayed in the lower part of the display. When a mode is enabled, the icon is framed.

Fig 5-1 Mode Icons
Use the left/right arrow buttons to navigate to the desired mode icon. The currently selected icon will flash. 1. Press the TEST OK button to enable the selected (flashing) mode.

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 17

2. Use the up/down arrow buttons to step through the mode options (if any). Refer to the section related to the specific mode for detailed instructions.
3. Press the TEST OK button to disable the selected (flashing) mode.

5.12.2 VFD Mode (ACV only)
In VFD (variable-frequency drive) mode, high-frequency noise is eliminated from the voltage
measurement using a low-pass filter. VFD mode is available when measuring ACV.

1. Select

and then press the TEST OK button to enable the selected (flashing) mode.

5.12.3 Frequency Mode (ACV only)
In Frequency mode, the frequency is displayed in the main display and the period is displayed in the secondary display. Frequency mode is available when measuring AC voltage.
1. Select and enable Frequency mode as described in section 5.12.1, Selecting modes.

5.12.4 Pass/Fail Mode (COMPARE)
1. Select the desired insulation voltage value using the rotary switch. 2. Select COMP and enable the Pass/Fail Function mode as described in section 5.12.1
Selecting modes. 3. Use the MODE button to select the resistance value to compare. Resistance value segments
and a COMPARE icon are shown on the LCD. See the test voltages and corresponding resistance value options in the table below. 4. Press TEST to start the process. 5. Observe the green light / red light for PASS / FAIL indication.

Test Voltages (V) 50 100 250 500 1000

Resistance Options (k) 100 /200 /500

Resistance Options (M) 1/2/5/10/20/50 1/2/5/10/20/50/100 1/2/5/10/20/50/100/200 1/2/5/10/20/50/100/200/500 1/2/5/10/20/50/100/200/500

5.12.5 Setup Utility
In Setup, define the settings for various meter options: · Auto power off (indicated by the text APO): A mode where the time period after which the meter enters sleep mode can be set. The range is 1 to 30 minutes, or OFF. The factory default is 10 minutes.
· Auto backlight off (indicated by the text b.Lit): A mode where the time period after which the backlight turns off can be set. The range is 1 to 30 minutes, or OFF. The factory default is 5 minutes.
· Continuity threshold (indicated by the text Cntin): A mode where the threshold for continuity tests can be set.
· Auto hold (indicated by the text A.Hold): Select Auto hold mode ON (Auto hold mode active) or OFF (Data hold mode active). For more information, see section 5.5 Data hold mode and Auto hold mode.

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 18

1. Select and enable Setup mode as described in section 5.12.1 Selecting modes. 2. Use the up/down arrows button to cycle through the mode options APO, b.Lit, Cntin,
A.Hold, and RESET shown on the secondary display. 3. Press the TEST OK button to activate the displayed option.
· APO: Use the left/right arrow buttons to change the auto power off time. · b.Lit: Use the left/right arrow buttons to change the auto backlight off time. · A.Hold: Use the left/right arrow buttons to configure auto hold/data hold. On indicates that the auto hold mode is active. Off indicates that the data hold mode is active. · Cntin: Use the left/right arrow buttons to change the continuity threshold. · RESET: Press the TEST OK button to reset the settings to the factory default.

5.12.6 Silent Mode
In Silent mode, the alert beeper is disabled. Silent mode does not affect the continuity beeper. Select and enable Silent mode as described in section 5.12.1 Selecting modes.

5.12.7 Data Recording Mode (99-point)
The meter has 99 memory locations for the storage of measurement data.

Select and enable MEM mode as described in section 5.12.1 Selecting modes.

Use the arrow keys to cycle through the mode options: SAVE, LOAD, and CLEAR shown on the

secondary display.

SAVE: The data on the main display is saved to a memory location indicated by the `88′ in the

upper display area.

LOAD: The data stored in the memory location (identified by the `88′ indicator) is displayed.

Use arrow keys to scroll the memory locations. Use the

button to exit the LOAD mode.

CLEAR: The data in all memory locations is erased.

5.13 Keypad Lockout
In Lockout mode the meter ignores all button presses except HOLD. The auto power off function, see section 5.1.1 Auto Power off, is disabled in Lockout mode. Press and hold the HOLD/LOCK button for 3 seconds to enter/exit the lockout mode. In lockout mode, the lock indicator is displayed. Do not confuse Lockout mode with the test lock feature used for continuous Insulation Resistance testing (see Section 5.7 Insulation Resistance Measurements).

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 19

5.14 Streaming Measurement Data using METERLiNK
5.14.1 General
You can stream measurement data to a mobile device using the METERLiNK app. The app is available from the Apple Store® (iOS) or from Google PlayTM (Android).
5.14.2 Procedure
1. Download the METERLiNK app to your mobile device and open the app. 2. Power the meter and select the desired measurement mode.
3. Press the meter’s Bluetooth button to enable communications. 4. Follow the instructions and prompts on the METERLiNK app screens.

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 20

6. Maintenance
6.1 Cleaning and Storage
Clean the meter with a damp cloth and mild detergent; do not use abrasives or solvents.
If the meter is not to be used for an extended period, remove the batteries and store them separately.
6.2 Battery Replacement
The Battery symbol flashes with no `bars’ when the batteries have reached a critical level of 7.0V. The symbol stays active and visible while the LCD is powered.
The meter displays readings within specifications while the low battery indicator is on. When this is no longer possible, the display must be blanked. The meter powers off before it displays an out of tolerance voltage.
1. To avoid electrical shock, disconnect the meter if connected to a circuit, remove the probe from the terminals, and set the function switch to the OFF position before attempting to replace the batteries.
2. Unscrew and remove the battery compartment cover. 3. Replace the six standard AAA batteries, observing correct polarity. 4. Secure the battery compartment cover.
Never dispose of used batteries or rechargeable batteries in household waste. As consumers, users are legally required to take used batteries to appropriate collection sites, the retail store where the batteries were purchased, or wherever batteries are sold.
6.3 Fuse Replacement
The fuse accessed via the battery compartment cover. The fuse is rated 440mA/1000V, ceramic fast blow with a minimum Interrupt Rating of 10kA.
6.4 Disposal of Electronic Waste
As with most electronic products, this equipment must be disposed of in an environmentally friendly way, and in accordance with existing regulations for electronic waste. Please contact your FLIR Systems representative for more details.

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 21

7. Specifications

7.1 General specifications

Display
Controls
Backlight Work Lights Measurement Rate Display Rate Input Impedance AC Voltage Bandwidth Power Supply Battery Life Considerations
Auto Power Off (APO) Measurement Types Over-current Protection (fuse) Polarization Index and Dielectric Absorption Ratio Tests

4,000-count with bar graph
9-position rotary switch
4-way Selector Pad with center OK/TEST button
LED: Pass (Green), Fail (Red)
(5) dedicated function buttons: Mode, Cancel, Hold/Lock, Bluetooth, Backlight
White LED
White LED array
2 samples per second, nominal
2 times per second
10M VDC/VAC < 100pF
50Hz to 500Hz
6 x `AAA’ (LR03) batteries
The meter can perform 1200 Earth-Bond resistance measurements with new alkaline batteries. These are standard tests of 1 with a duty cycle of 5 seconds ON and 25 seconds OFF. Insulation Resistance: The meter can perform 300 Insulation Resistance tests with new alkaline batteries. These are standard tests of 1M with a duty cycle of 5 seconds ON and 25 seconds OFF. This does not include Bluetooth or Backlight use.
User programmable OFF to 30 minutes in 1-minute increments, with audible pre- alert (20 seconds)
True RMS AC Volts, DC Volts, Earth Bond Resistance, Insulation Resistance, Capacitance, Diode, and Continuity
440mA/1000V, ceramic fast blow fuse with a minimum Interrupt Rating of 10kA PI (Polarization Index) = R10-min/R1-min DAR (Dielectric Absorption Rations) = R1-min/R30-sec Where R10-min: The insulation resistance measured at the 10minute point after pressing the TEST button. R1-min: The insulation resistance measured at the 1-minute point after pressing the TEST button. R30-sec: The insulation resistance measured at the 30 second point after pressing the TEST button.

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 22

Other Indications Internal Memory Operating Temperature Storage Temperature
Operating Humidity
Storage Humidity Indoor use only Altitude Dimensions Weight Agency Approvals IP Rating Drop test rating

Progressive Low Battery, Over Range (99) storage locations 32 to 122°F (0 to 50°C) -4 to 140°F (-20 to 60°C) Max 80% up to 95°F (35°C) decreasing linearly to 60% at 113°F (45°C) 80% maximum
6561 ft. (2000 m) 2.4 x 3.9 x 8.2 in. (60.7 x 98.1 x 207.1 mm) 1.33 lbs. (605 g), including batteries FCC Class B, CE, UL, RCM, UKCA IP54 9.8 ft. (3 m)

7.2 Electrical Range Specifications

Note: Accuracy is stated at 65°F to 83°F (18°C to 28°C) and less than 75% RH)

Function

Range

Resolution Accuracy of Reading

DC Voltage

1000.0V 0.1V

± (0.1% + 5 digits)

AC Voltage

1000.0V 0.1V

50Hz to 60Hz ± (1.5% +5 digits); 61Hz to 500Hz ± (2.0% + 5 digits)

VFD AC Voltage 1000.0V 0.1V

50 to 60Hz ± (1.5% +5 digits); 61 to 400Hz ± (5.0% + 5 digits)

Notes: 1. Measurement start-voltage: <50V for VAC (add 20 digits for accuracy)

2. VFD/LPF Cutoff: 800Hz (-3dB point)

3. Input Impedance 10M (less than 100pF)

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 23

Function
Earth-Bond Resistance

Range 40.00

Resolution Accuracy of Reading

0.01

± (1.5% + 5 digits) *

400.0 4000 40.00k

0.1 1 0.01K

± (1.5% + 3 digits)

Notes:

1. * 1.00 add 3 digits. 2. Open Circuit Test Voltage: >4.0V, <8V 3. Short Circuit Current 200.0mA 4. Live Circuit Detection if Greater than or equal to 2V @AC/DC Inputs test inhibited.
The top LCD will display the actual live voltage sensed.

Function Continuity

Range 400.0

Resolution Accuracy of Reading

0.1

± (0.5% + 2 digits)

Diode

2.000V

0.001V ± (1.5% + 2 digits)

Notes: Continuity: Built-in buzzer sounds when measured resistance threshold is met (adjustable from 10 to 50, default is 30). Continuity MAX Test Current: 1mA Continuity MAX Open Circuit Voltage >3.0V Diode MAX Test Current: 1mA Diode MAX Open Circuit Voltage: 3.0V Overload Protection: 1000V AC rms or DC

Function

Range

Resolution

Capacitance

1000 µF 10.00mF

1µF 0.01mF

Notes: 1. Overload Protection: 1000V AC rms or DC

Accuracy of Reading ± (1.2% + 2 digits) ± (1.2% + 20 digits)

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 24

Function

Range 400.0Hz

Resolution Accuracy of Reading

0.1Hz

± 5 digits

Frequency (ACV)

4.000kHz 40.00kHz

0.001kHz ± 5 digits 0.01kHz ± 5 digits

Function
Insulation Resistance

Resistance Range

Resolution

4.000M 40.00 M

0.001M 0.01M

400.0M 4000 M

0.1M 1M

4.1G to 20.0G 0.1G

Test voltages

Min. Resistance

50V

50k

100V

100k

250V

250k

500V

500k

1000V

1M

Accuracy of Reading
± (1.5% + 5 digits)
± (3.0% + 5 digits)
± (10.0% + 3 digits)
Max. Resistance 50.0M 100.0M 250.0M 500.0M 20G

Notes:

1. Test Current = 1 mA 2. Test Voltage Accuracy: +0% to +20%; Short Circuit Test Current: 1mA (nominal) 3. Auto Discharge Function: Discharge time <1 second for Cap. <1µF 4. Max. Capacitive Load: Operable with up to 1µF load. 5. Live Circuit Detection: If >2V AC/DC detected test inhibited.
The top LCD will display the actual live voltage sensed.

7.3 Input Specifications
Function AC Voltage, DC Voltage Insulation Test

Maximum Input 1000V DC/AC No LIVE Voltage IN – Protected

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 25

8. Customer Support
For technical assistance, repair and returns, document downloads, and general information, visit the support site, link below: https://support.flir.com

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 26

9. Warranty
9.1 FLIR Limited Lifetime Warranty This product is protected by FLIR’s Limited Lifetime Warranty.
Visit https://support.flir.com/prodreg to register your product and read the Limited Lifetime Warranty document.

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 27

Website https://www.flir.com
Customer Support https://support.flir.com

Disclaimer
Specifications subject to change without further notice. Models and accessories subject to regional market considerations. License procedures may apply. Products described herein may be subject to US Export Regulations. Please refer to exportquestions@flir.com with any questions.
There is no export-controlled information contained in this user manual.

Publication Identification No.: Release version: Release Date: Language:

IM75-2_en-US AA 2023 September en-US

Copyright © 2023, Teledyne FLIR Commercial Systems, Inc. All rights reserved worldwide.

FLIR IM75-2 USER MANUAL

IM75-2-UM_en-US Rev AA 28

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>