Health Life HL868AW Automatic Upper Arm Blood Pressure Monitor Instruction Manual
- June 15, 2024
- Health Life
Table of Contents
- HL868AW Automatic Upper Arm Blood Pressure Monitor
- Medical Disclaimer
- Intended Use
- About Blood Pressure
- Measurement Method
- Accuracy
- Precautions
- Precautions
- Device Overview
- Symbol Definitions
- Features
- Bluetooth Transmission
- Installing Batteries
- Using the AC/DC Adapter
- Applying the Cuff
- Measurement Procedure
- Memory Function
- Storage and Maintenance
- Troubleshooting
- Specifications
- Appendix
- Blood Pressure Diary
- Documents / Resources
Instruction Manual
Automatic
Upper Arm Blood Pressure Monitor
HL868AW Automatic Upper Arm Blood Pressure Monitor
Model No. HL868AW
Medical Disclaimer
This manual and product are not meant as a substitute for advice provided by
your doctor.
You are not to use the information contained herein, or this product for
diagnosing or treating a health problem or prescribing any medication. If you
have or suspect that you have a medical problem, promptly consult your
healthcare provider.
Intended Use
This device uses the oscillometric method to automatically measure systolic
and diastolic blood pressure as well as heart rate. All values can be read out
on one LCD panel. Measurement position is at human being’s upper arm. The
intended use of this over-the-counter device is for adults age 18 years and
older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm
to 43 cm) and for home use.
The device can accurately measure blood pressure in pregnant patients
including those with known or suspected in pre-eclampsia condition.
HL868AW detects the appearance of irregular heartbeats during
measurement; an indicated symbol will appear with measuring reading. And the
Risk Category Indicator will show the information with the readings on the
screen for the user tracking their blood pressure level.
Besides, the device features a built-in “Bluetooth Transmission” function,
which enables the device automatically transmits measuring results to paired
Bluetooth-enabled device. Also, users could simply synchronize the current
date and time, and check the battery status of blood pressure monitor by
means of Daily Check ® application software with the paired Bluetooth-enabled
device.
About Blood Pressure
-
What is blood pressure?
Blood pressure is the measurement of the force of blood pushing against the walls of the arteries. Arterial blood pressure is constantly fluctuating during the course of the cardiac cycle. The highest pressure in the cycle is called the systolic blood pressure, and represents the pressure in the artery when the heart is beating. The lowest pressure is the diastolic blood pressure, and represents the pressure in the artery when the heart is at rest. Both the systolic and the diastolic pressure are necessary for a physician to evaluate the status of a patient’s blood pressure.
Many factors such as physical activity, anxiety or the time of day, can influence your blood pressure. Blood pressure is typically low in the mornings and increases from the afternoon to the evening. It is on average lower in the summer and higher in the winter. -
Why is it useful to measure blood pressure at home?
Having one’s blood pressure measured by a doctor in a hospital or a clinic is often associated with a phenomenon called “White Coat Hypertension” where the patient becomes nervous or anxious, thus raising his blood pressure. There are also numerous other factors that might cause your blood pressure to be raised at a specific time of day. This is why medical practitioners recommend home monitoring as it is important to get readings of blood pressure during different times of the day to really get an idea of your real blood pressure.
Medical practitioners generally recommend the “Rule of 3”, where you are encouraged to take your blood pressure three times in a row (at 3 ~ 5 minute interval), three times a day for three days. After three days you can average all the results and this will give you an accurate idea of what your blood pressure really is.
A. AHA blood pressure classifications:
Standards for assessment of high or low blood pressure without regard to age,
have been established by the American Heart Association (AHA 2017), as shown
in the chart.
BLOOD PRESSURE CATEGORY| SYSTOLIC mm Hg (upper number)| | DIASTOLIC mmHg
(lower number)
---|---|---|---
NORMAL| LESS THAN 120| and| LESS THAN 80
ELEVATED| 120-129| and| LESS THAN 80
HIGH BLOOD PRESSURE (HYPERTENSION)STAGE1| 130-139| or| 80-89
HIGH BLOOD PRESSURE (HYPERTENSION)STAGE2| 140 OR HIGHER| or| 90 OR HIGHER
HYPERTENSIVE CRISIS (consult your doctor immediately)| HIGNER THAN 180|
And/or| HIGHER THAN 120
However, this chart is not exact for classification of blood pressure and
it’s intended to be used as a guide in understanding non-invasive blood
pressure measurements. Please consult with your physician for proper
diagnosis.
B. Variations in blood pressure:
Individual blood pressures vary greatly both on a daily and a seasonal basis.
These variations are even more pronounced in hypertensive patients. Normally
the blood pressure rises while at work and is at its lowest during sleeping
period.
(hypertension: means a person who has high blood pressure symptom.)
The graph below illustrates the variations in blood pressure over a whole day
with measurement taken every five minutes.
The thick line represents sleep. The rise in blood pressure at 4 PM (A in the
graph) and 12 AM (B in the graph) correspond to an attack of pain.
Measurement Method
HL868AW Automatic Upper Arm Blood Pressure Monitor measures blood pressure and
heart rate by oscillometric method, meaning the fluctuations in pressure are
measured. Once the cuff is wrapped around your upper arm, just turn on the
monitor and inflation automatically starts. The inflation of the cuff creates
pressure around the arteries inside upper arm.
Within the cuff is a gauge which senses the fluctuations (oscillations) in
pressure. The fluctuation measured represents the degree of intensity that
your arteries contracting with each heartbeat, and also a result of the
pressure that the cuff has placed on the upper arm. The monitor measures
these contractions and converts the information to a digital value. This is
the result displayed on the monitor screen.
Once the measurement is completed, the cuff will automatically deflate.
Accuracy
HL868AW Automatic Upper Arm Blood Pressure Monitor has been clinically tested
against a scientific device called mercury sphygmomanometer, considered the
gold standard in blood pressure measurement.
All HL868AW Automatic Upper Arm Blood Pressure Monitors have performed
equivalent to measurements taken with this scientific device and are within
the accuracy limits prescribed by the American National Standard for
Electronic or Automated Sphygmomanometers.
The SPHYGMOMANOMETER was clinically investigated according to the requirements
of ISO 81060-2:2018.
In case it is needed to have the device checked for calibration, please
consult the distributor. This is recommended to be considered every two years.
Precautions
- Read the Instruction Manual thoroughly before measuring and keep it at hand for your reference at any time.
- The device is designed for home use and not suitable for clinical use.
- The patient is an intended operator, who can operate the device by himself or herself, not necessarily by a physician or operator.
- This monitor is not intended for use in the MR environment.
❑ The device should not be used to either self-diagnose Hypertension or
exclude the diagnosis of Hypertension. If your blood pressure reading is out
of normal range, please consult your physician. Even your blood pressure
reading is within the “normal” range, the device cannot exclude the diagnosis
of Hypertension.
❑ Do not take a measurement in a too low (less than 41 ℉/5 ℃) or too high
(more than 104 ℉/40 ℃) temperature, nor in a place outside humidity ranges (15
% ~ 93 % R.H.) and atmospheric pressure ranges (700 ~ 1060 haps), or you may
get inaccurate readings.
❑ Wait 30 ~ 45 minutes before measurement if you’ve just consumed caffeinated
beverages or smoked cigarettes.
❑ Please rest for at least 5 ~ 10 minutes before taking a measurement.
❑ To allow your blood vessels to return to the condition prior to taking the
measurement, please wait at least 3 ~ 5 minutes between measurements. You may
need to adjust the wait time according to your personal physiological
situation.
❑ We recommend you using the same arm (preferably the left arm) and measuring
around the same time each day.
❑ Perform measurements in a quiet and relaxed environment at room temperature.
❑ Do not move or shake the device during a measurement. Please keep quiet and
do not talk during measurements.
❑ If you are pregnant, you should pay more attention to your blood pressure
changes, because during this time, it may change drastically.
❑ This monitor is clinically validated for use in pregnancy and preeclampsia.
When you detect unusual readings in pregnancy, you should measure again after
taking some rest. If the reading is still abnormal, consult your doctor or
gynecologist.
❑ This product is not suitable for:
◼ People with arrhythmias
◼ Undergoing intravenous injection on any limb
◼ Currently in a dialysis treatment
Precautions
❑ For those who have had a mastectomy or lymph node clearance, it is
recommended to take a measurement on the unaffected side.
❑ When used among medical electronic equipment on the same limb,
pressurization of the cuff may cause temporarily malfunction to other devices.
❑ If you have one of the circulatory problems such as arteriosclerosis,
diabetes, liver disease, kidney disease, severe hypertension, peripheral
circulation……, please consult your healthcare professional before using the
device.
❑ If the cuff is worn incorrectly, or the shape of the upper arm is special
(for example, the circumference of the upper arm differs largely from the
circumference of the forearm), excessive gap might occur between the arm cuff
and the arm, and it might lead to measurement errors or inaccuracies. If you
have any question about the condition of cuff wearing and/or measurement
result, please consult your healthcare professional.
❑ Blood pressure measurements taken with this device are equivalent to those
obtained by a trained observer using the cuff / stethoscope auscultation
method and are within the accuracy limits prescribed by the Standard of EN
1060-4.
❑ The applied part is cuff.
❑ Results are not intended for direct diagnosis. Please consult with a
physician if you have any questions or concerns about your results.
*Attention!
- Do not use the device on infants, children, or those who cannot express their own intention. To avoid accidental strangulation, keep this product away from children and do not drape tube around neck.
- The medical device should not be used adjacent to or stacked with other equipment. In case adjacent or stacked use is necessary. The medical device should be observed to verify normal operation in the configuration in which it will be used.
- Consider the electromagnetic compatibility of the device (ex. power disturbance, radio frequency interference etc.) Please use it indoor only.
- Over high frequency measurements may result in blood flow interference, which is likely to cause uncomfortable sensations, such as partial subcutaneous hemorrhage, or temporary numbness to your arm. In general, these symptoms should not last long. However, if you do not recover in time, please see your medical practitioners for help.
Device Overview
◆ Part Names and Product Components
Symbol Definitions
◆ Unit Display
SYMBOLS
| Definitions
---|---
Low Battery Symbol
| This symbol appears when the battery power is excessively low or the
polarity reverses.
We suggest you replace all batteries with new ones, and make sure the +/-
polarities are properly positioned.
Pulse Symbol
| Once pulse is detected, the symbol flashes with each pulse beat.
Our suggestion:
Please do not talk or move during measurements.
Irregular Heartbeat Detector
| This symbol appears when an irregular heartbeat was detected.
Our suggestion:
Repeat the measurement after resting for at least 5 minutes, and restart your
measurement while sitting down comfortably and quietly.
If symbols appear frequently, please contact your physician.
Excessive Body Motion Detector
| The symbol appears if body movement is detected during measurement,
especially the movement on the arm that the blood pressure monitor is worn on.
Notice: The measured blood pressure reading may not be accurate if the icon is
displayed.
Cuff Fitness detection Symbol
| The symbol appears if the cuff was wrapped incorrectly, which is too tight or too loose. This is the function aid in detecting if the cuff is wrapped properly.
Risk Category Indicator
| The arrowhead points out the specific Risk Category that your measurement reading fits in.
Bluetooth Symbol
| Under Bluetooth Data Transmission / Link Mode, LCD displays this symbol.
Average of Last 3 Measurements
| This symbol appears when LCD displays average value of last 3 readings.
AM/PM Averaging
| This symbol appears when LCD displays an average from the last 3 morning or last 3 evening measurements.
Features
◆ Risk Category Indicator (AHA 2017)
This device is equipped with Risk Category Indicator which classifies your
blood pressure measurements into five stages (Normal, Elevated, Hypertension
stage 1, Hypertension stage 2 and Hypertensive crisis) based on the blood
pressure
standards established by the American Heart Association (AHA). Besides, for
yours and your loved ones’ health, we further classify the five stages into
numeral ranges, which sorts out hypertension symptoms more clearly. Moreover,
to your convenience and readability, we use three corresponding colors to
represent your measuring result. Refer to below comparison chart for details:
BLOOD PRESSURE CATEGORY | SYSTOLIC mm Hg (upper number) | NDICATOR OLOR | ||
---|---|---|---|---|
NORMAL | <120 | and | <80 | |
ELEVATED | 120-129 | and | ||
HIGH BLOOD PRESSURE (HYPERTENSION) STAGE 1 | 130-139 80-89 | or | ||
HIGH BLOOD PRESSURE (HYPERTENSION) STAGE 2 | 140-180 | or | 90-120 | |
HYPERTENSIVE CRISIS (consult your doctor immediately) | >180 | and/or | >120 |
Source: AHA 2017
After measurement, LCD displays the systolic and diastolic pressure, heart
rate, date and time along with Risk Category Indicator bar. The higher the
blood pressure, the higher the bar. Compare the bar with the three colors at
the right of LCD display to know the classification of your blood pressure
based on American Heart Association standard (AHA 2017).
*Note!*
When a person’s systolic and diastolic pressures fall into different
categories, the higher category should apply.
e.g. systolic pressure 181 and diastolic pressure 99 Red category
(Hypertensive crisis)
e.g. systolic pressure 110 and diastolic pressure 95 Red category
(Hypertension stage 2)
Note !
The above table is not exact for classification of blood pressure and it’s
intended to be used as a guide in understanding non-invasive blood pressure
measurements.
Usually this is not a cause for concern; however we recommend you consult with
your physician for proper diagnosis or seek medical advice according to our
recommendation mentioned above. Please note that the device does not
appropriate to diagnose hypertension, and it is only for user reference on
blood pressure monitoring.
◆ Irregular Heartbeat Detector**
The symbol will appear on screen indicating a certain heartbeat irregularity
is detected during measurement.
The heartbeat rhythm that is more than or less than 25% from the average
rhythm is usually defined as an irregular heartbeat rhythm.
Talking, moving, shaking or having an irregular pulse during the measurement
can result in the appearance of this symbol.
Usually this is not a cause for concern, however if the symbol appears often,
we recommend you seek medical advice.
And please note that the device does not replace a cardiac examination, but
serves to detect pulse irregularities at an early stage.
*Note!
- The pulse display is not suitable for checking the frequency of heart pacemakers. If a certain pulse irregularity is detected during measurement often, we recommend you seek medical advice.
- As a safeguard, we recommend that if you have arrhythmias such as atrial or ventricular premature beats and atrial fibrillation or any other special conditions you should check with your physician before using your device.
- The IHB function is designed neither for use by people with arrhythmias nor for diagnosing or treating an arrhythmic problem. In order to filter the unstable status of user and avoid affecting the detection of heart rate from any movement, shaking or talking in the beginning of measurement, the method of averaging heartbeat intervals of subject device is calculated with the three proper heartbeat pulses detected in the beginning of measurement and that is different from a strict mathematical averaging of all recorded intervals.
- At least 3 beats with 25% or greater difference from the average heartbeat interval will generate the IHB icon on the screen.
◆ Bluetooth Transmission
HL868AW features a Bluetooth transmission function, which enables the device
to transmit measured results to paired Bluetooth-enabled device after
measurement automatically. When the connection is established, the device
would transmit systolic pressure, diastolic pressure, and pulse with time to
the Bluetooth-enabled device.
Before attempting to sync the device with your smart device, make sure
Bluetooth function is turned ON in both your smart device and the monitor, and
make sure your Bluetooth-enabled device has downloaded the App. See the “App
for Bluetooth” section for details.
◆ Transmit Readings
Bluetooth function can be activated only when Bluetooth is turned ON. To turn
on Bluetooth, please refer to Setting Year, Date and Time section.
-
There are 2 ways to activate Bluetooth Function.
a. Automatically Activate:
When a measurement is completed, the device activates Bluetooth Function automatically, and Bluetooth symbol will be flashed on the screen.
b. Manually Activate:
Under sleeping mode, user can press the button for 3 seconds to activate Bluetooth function, and Bluetooth symbol will be flashed on the screen. -
If HL868AW is connected successfully to your smart device, Bluetooth symbol will appear on the screen.
◆ Date/Time Synchronization
- The date/ time setting of HL868AW can be synchronized by Bluetooth enabled device (e.g. smartphone) which has downloaded and installed Daily Check ® application software.
- When Bluetooth connection is established, the Bluetooth-enabled device can send commend with the date/ time information to BPM and the BPM’s date/ time will be updated.
Bluetooth Transmission
◆ App for Bluetooth
Download and install “DailyChek® ” app on your smart device from Google Play
or App store.
System requirement of the Bluetooth-enabled device
- Bluetooth 4.2 for Android 6.0 or above
- Bluetooth 4.2 for iOS 7.0 or above
NOTE:
- HL868AW is subject to and complies with electromagnetic compatibility (EMC)standard of IEC 60601-1-2, EN 301 489-1, EN 301 489-17, EN 300 328 and U.S. federal guidelines, Part 15 of the FCC (Federal Communications Commission) rules for devices with RF capability. These guidelines help ensure that your device will not affect the operation of other nearby devices. Additionally, other devices should not affect the use of your device.
- Other wireless devices that are in use nearby, such as a cell or mobile phone, or a wireless network, may prevent or delay the transmission of data from your device to paired Bluetooth-enabled device. Moving away from the source of the interference or turning off these devices to resolve the problem.
- Make sure HL868AW and paired Bluetooth-enabled device are within acceptable distance (no more than 10 meters) with each other. If not, put them closer.
- Be sure to select the correct user on the monitor before your blood pressure measurement begins.
- Bluetooth data transmission is not available under measurement.
Installing Batteries
When low battery symbol appears on the display, or the device has no reaction
toward operation, please change batteries.
Replace all worn-out batteries with new ones and do not mix new and used
batteries. All batteries used must be the same type. Do not mix alkaline,
standard (carbon-zinc) or rechargeable (cadmium) batteries either. Such action
may shorten the battery life or cause the device to malfunction.
Slide the battery cover and insert 4 AAA (1.5V, LR03) alkaline batteries into
the battery compartment as shown on the figure below. Make sure the polarities
“+” and “–” ends are coinciding with similar markings engraved on the battery
housing.
*Attention!
- Batteries are hazardous waste. Do not dispose of them together with the household garbage. Please discard worn-out batteries to the recycling site according to local regulations.
- Keep the battery away from children in case they choke on it.
- To prolong the battery life and prevent damage caused by leakage, remove the batteries from the device if the device is not to be used for a long period.
- Memories (if any) will not be deleted during battery replacement.
- Please replace all worn-out batteries with new ones when you are operating the Bluetooth transmission function and the low battery symbol appears on the display.
- After replacing the batteries, the date and time will be reset.
Using the AC/DC Adapter
This monitor is designed for operation with batteries or an AC/DC adapter.
Please use only a compatible AC/DC adapter with required voltage and current
as indicated in this manual.
*Note!
-
No batteries are needed when operating with an AC/DC adapter.
-
Please unload the batteries when operating with an AC/DC adapter for an extended period of time.
-
Leaving the batteries in the compartment for a long time may cause leakage, which may lead to damage of the unit.
-
Recommend Adapter specification.
Model: Fran Mar International, FRM06-S05-UU -
Rating:
Input: 100-240V, 50/60 Hz, 0.2A
Output: 5V, DC, 1.2A,
*Note!
When you use the blood pressure monitor with AC/DC adapter, do not position
the device to make it difficult to disconnect the adapter plug.
Applying the Cuff
❑ Wrap the cuff on a bare arm or over thin clothing. Thick clothing or a
rolled up sleeve will cause inaccurate blood pressure measurements.
❑ Press your brachial artery approximately 1 inch (2 ~ 3 cm) above the elbow
on the inside of your left arm to determine where your strongest pulse is.
❑ Slide the end of arm cuff furthest from the tube through the metal ring to a
loop. The smooth cloth should be on the inside of the cuff.
❑ If the cuff is located correctly, the velcro will be on the outside of the
cuff, metal ring will not touch your skin, and the arrow on the cuff should be
fallen within the Proper Fit Range (Figure 1). ❑ Put left arm through the
cuff loop. The tube should lie over the brachial artery on the inner part of
the arm.
The bottom edge of the cuff should be approximately 1 inch (2 ~ 3 cm) above
the inner elbow.
❑ Pull the end of the cuff so that it tightens evenly around your arm,
allowing room for 2 fingers to fit between the cuff and your arm (Figure 2).
❑ When the cuff is
positioned properly, press the velcro firmly against the pile side of the cuff
and make sure the cuff is not slipped during measurement.
❑ Sit on a chair comfortably, put your feet flat on the floor and lay your
forearm on the table, make sure your back and arm supported, legs uncrossed,
so that the cuff is at the same level as your heart(Figure 3). ❑ Relax your
arm and turn your arm upward.
❑ Make sure there are no kinks in the air tube.
Note!
- Fit the cuff snugly, leaving enough space for 1 inch (2 ~ 3 cm) between the inner elbow and the lower edge of the cuff, or the measurement may not be accurate.
- In case the cuff kept pumping up non-stop, unwrap the cuff at once.
- Do not wrap the cuff around any body part other than your arm.
- Do not use this device if your arm has any wound or injury, especially after surgery on the arm. Otherwise, it may cause infection at the surgical site. Please use the device after the wound has healed.
- If there is one of situations mentioned below, please dispose the device without reuse.
- If you have any infectious skin disease or the device is used by users with infectious skin disease, please do not continue using the device.
- Before using the device, user should check the appearance of cuff. If you notice blood or other soil on cuff, please do not use this device.
Measurement Procedure
◆ Switch on the monitor
A. Put in 4 AAA 1.5V (LR03) alkaline batteries.
B. All segments appear on the screen for 1.5 seconds.
C. The monitor will automatically turn to sleeping mode (all LCD segment are
cleared).
◆ Setting Year, Date and Time
-
Set Manually.
-
Under sleeping mode, to enter setting mode, press button for 3 seconds, then Month digit flashes.
Use button to select current month.
a. When above settings are done,
Press button to adjust current DATE.
Press button to select current date.
b. Continue to set current YEAR,
HOUR (1, 2……12PM, 1PM……,12) and MINUTE (00,01……,59) by following Step B.
c. Users can adjust MONTH-DATE-YEAR-HOUR-MINUTE in an orderly manner.
Press button to save the settings and switch to Bluetooth function setting. -
Using your Bluetooth smart device.
The date and time on your monitor can be automatically updated, when you connect it with your smart device.
Once the date and time have been successfully synced, future readings will automatically have the correct date and time.
◆ Turn the Bluetooth Function ON/OFF
Press button again to enter Bluetooth setting. The default setting of this
function is ON. Press button to turn the Bluetooth function ON or OFF. When
settings are done, press START/STOP button to confirm the entries.
The device will turn to sleeping mode and is ready to measure.
◆ Taking a Measurement
A. With the cuff wrapped around your arm, press START/STOP button to start
measurement. B. All display symbols appear on the screen for 1.5 seconds.
After all symbols disappear, the display will show “00”. The monitor is “Ready
to Measure” and will automatically inflate to the level that is right for
you.
C. As the cuff inflates, the monitor automatically determines your ideal
inflation level. This monitor detects your blood pressure and pulse rate
during inflation. The Heartbeat Symbol ( ) flashes at every heartbeat. Remain
still and do not move until the entire measurement process is completed. The
device will detect your pulse and determine the blood pressure.
*Note!
- If the cuff does not stop inflating, remove the cuff at once.
- To stop measurement, press START/STOP button. The cuff will deflate immediately after the button is pressed.
- The monitor will automatically switch to sleeping mode if no operation in 1 minute.
D. After the monitor has determined your blood pressure and heart rate, the
cuff automatically deflates. Your results including systolic pressure,
diastolic pressure, heart rate, corresponding risk category indicator,
irregular heartbeat detector and excessive body motion detector (if any) will
be displayed with date and time for 1 minute and saved to memory
automatically.
E. After measurement is completed, the device activates Bluetooth function
automatically and the Bluetooth symbol will flash on the screen.
If the monitor cannot be connected to paired Bluetooth-enabled device over 45
seconds, LCD will display error message “E4” and Bluetooth will be turned off.
F. Device automatically shuts off if there is no operation over 1 minute.
Memory Function
◆ Storing Data
After each measurement, the results including systolic and diastolic pressure,
heart rate and corresponding risk category indicator, irregular heartbeat
detector and excessive body motion detector (if any) with the time and date
will be automatically stored.
The monitor features 1 user memory capability. User holds the last 120
measurements, and automatically replacing the oldest data with a new one.
◆ Recalling Data
A. Press button to enter memory mode.
If there is no data stored before, nothing (except month, date, and time)
will be appeared on the display. If yes, the first reading will be the
average of last 3 measurements.
B. Press button again, LCD displays average of latest 3 AM measurements, and
then is average of latest 3 PM measurements. If fewer than three measurements
in AM or PM, the average will not be displayed.
C. Every new press of the button will recall a previous reading. The latest
reading will be recalled first. Press button repeatedly to scroll through the
previous readings stored in the memory.
D. To stop reading the memories, press START/STOP button to switch to
sleeping mode.
◆ Erasing data
A. Press button to enter memory mode.
B. Press and hold and button at the same time, all the data of the selected
user will be erased automatically. Note: Once the data is deleted, it
can NOT be restored.
Storage and Maintenance
◆ General Use
❑ Do not twist the cuff in any way.
❑ Do not press START/STOP button if the cuff is not wrapped around your
upper arm.
❑ Do not drop the product and avoid any strong impacts.
◆ Maintenance
❑ Do not attempt to disassemble or change any parts of the monitor, only
trained technicians are allowed to repair and disassemble the device,
including software upgrades, cuff, patches and maintenance, because a
substitution of a component different from supplied might result in
measurement error.
❑ If any suggestion or service is requested, please consult your service
station.
❑ Do not implement the maintenance procedures for equipment during
measurement.
To ensure that your device is in optimal use and to avoid damage, please refer
to the following instructions:
❑ Clean the device and the cuff with a soft dry cloth, or
❑ Use a dry cloth with water to clean the device (do not flush it directly, do
not soak it in water, and hold the device dry), or
❑ Do not use detergent or any strong chemicals to clean the device.
❑ Make sure the cuff is completely dry before using.
According to the different usage scenarios of the sphygmomanometer, there are
recommended methods to avoid infection as the following points:
❑ Only use it yourself (home use), it can be cleaned at ordinary times, and
wipe it once a month with a commercial available 75% alcohol cotton sheet (for
the cuff) for more than 30 seconds each time.
❑ If it is used for more than one person (home use), it can be cleaned at
ordinary times. It is disinfected once a week (for the cuff belt) with a
commercially available 75% alcohol cotton sheet, for more than 30 seconds
each time.
❑ After cleaning / disinfecting/ before using the device, please make sure
that there are no blood stains or soil on the LCD, the device and the cuff. If
there is any blood stains or soil, please dispose the device without reuse.
❑ If it is used in a complex environment (such as a hospital) or after
multiple people (non-family), please discard the old cuff and replace it with
a new one.
◆ Storage
❑ If the device is not to be used for a long time, please remove the batteries
from the device (leaking of battery acid can cause the device to malfunction).
❑ Always store the unit in the storage case after use. It is intended to be
transported or stored in a carrying case between uses.
❑ Do not place the device directly under sunlight, in high temperature, or in
humid or dusty places.
❑ Do not store the device in extremely low (less than –13℉/–25℃) and high
(more than 158℉/70℃) temperature, nor in a place its humidity exceeds 93% R.H.
Troubleshooting
SYMBOLS/SYMPTOMS | CONDITIONS/CAUSES | INDICATION/ CORRECTION |
---|---|---|
Unit does not turn on when START/STOP button is pushed. | Worn-out | |
batteries. | Replace them with 4 new AAA (LR03) alkaline batteries. | |
Battery polarities have been positioned incorrectly. | Re-insert the batteries |
in the correct positions.
Measuring Error Symbol appears when blood pressure value displayed is
excessively low or high.| Cuff has been placed incorrectly.| Wrap the cuff
properly so that it is positioned correctly.
Did you talk or move during measurement?| Measure again. Keep arm steady
during measurement.
Shaking of the arm with the cuff on
Measuring Error Symbol| Air circuit abnormality. Cuff tube may not be plugged into monitor correctly.| Check cuff connection. Measure again.
Measuring Error Symbol| Inflation pressure exceeding 300 mmHg.| Switch the unit off, then measure again.
Measuring Error Symbol| Can’t determine blood pressure measurement data.| Wrap
the cuff properly and keep steady. Measure again.
Cuff is worn improperly, or the shape of the upper arm is unusual (for
example, the circumference of the upper arm differs largely from the
circumference of the forearm), excessive gap might exist between the arm cuff
and the arm.| Wrap the cuff snugly so that it is positioned correctly.
If you have any question about the cuff wearing and/or measurement result,
please consult your healthcare professional.
Measuring Error Symbol| If the device cannot be connected to paired Bluetooth- enabled device over 45 seconds, LCD will display Error message “E4” and Bluetooth will be turned off.| Please press button for 3 seconds to start Bluetooth function.
Excessive Body Motion Detector Notice: The measured blood pressure reading
may not be accurate if the icon is displayed.| Body movement during
measurement, especially, the movement on the arm the blood pressure monitor is
worn on. e.g. Talking, moving or shaking of the arm with the cuff on while
measurement.| Measure again. Keep arm steady during measurement.
Cuff is worn improperly, or the shape of the upper arm is unusual (for
example, the circumference of the upper arm differs largely from the
circumference of the forearm), excessive gap might be exist between the arm
cuff and the arm.| Wrap the cuff properly and keep steady. Measure again.
If you have any question about the cuff wearing and/or measurement result,
please consult your healthcare professional.
Cuff Fitness Detection| The cuff was wrapped incorrectly (for example too
loosely or too tightly).| Please reference “Applying the Cuff ”section to wrap
the cuff correctly.
BPM cannot communicate with Bluetooth-enabled device| Paring has not been
completed.| Please re-pairing the BPM and Bluetooth -enabled device with each
other.
---|---|---
Bluetooth function is not turn on.| See the “the Bluetooth Transmission”
section to turn on Bluetooth function.
The distance between BPM and Bluetooth-enabled device is out of transmitting
range.| Please make sure the acceptable distance (≦10 meters) with each other.
Use an incompatible Bluetooth- enabled device.| Please refer to “Bluetooth
compatibility” and “RF Specification”
Use non-Bluetooth-enabled device.
Unexpected loss of electrical/mechanical integrity.| Re-insert the batteries
and try again.
Return the device to your local distributor or importer.
Note: If “EP” appears on the display, just return the device to your local
distributor or importer.
Limited Warranty
◆ Warranty For Two Years from the manufacturing date
Please note that this warranty does not cover damage caused by misuse or
abuse; accident; the attachment of any unauthorized accessory; alteration to
the product; improper installation;
unauthorized repairs or modifications; improper use of electrical/power
supply; loss of power; dropped product; malfunction or damage of an operating
part from failure to follow the manufacturer’s recommended maintenance;
transportation damage; theft; neglect; vandalism; or inappropriate; loss of
use during the period the product is at a repair facility or otherwise
awaiting parts or repair; or any other conditions whatsoever that are beyond
the control of importers or distributors.
Specifications
Model Number | HL868AW |
---|---|
Measurement Method | Oscillometric (Inflation) |
Rated Range of Cuff Pressure | 0 ~ 300 mmHg |
Rated Range of Determination | 40 ~ 280 mmHg |
Measurement Range of Heart Rate | 40 ~ 199 Beats/Minute |
Accuracy | Pressure: ± 3 mmHg Pulse: ± 5 % Max. |
Inflation | Automatic Inflation (Air Pump) |
Deflation | Automatic Air Release Control Valve |
Display | Liquid Crystal Display |
Memory | 120 Memory Total for 1 User |
Unit Dimensions | 4.5 x 5.61 x 1.8 inch (L x W x H) 114 x 143 x 46 mm (L x W x |
H)
Unit Weight (Cuff and batteries excluded)| 259.8 g ± 5 g (9.16 oz ± 0.17 oz)
Cuff Size| UC-01: Universal size cuff (9”~17”/ 23 ~ 43 cm)
Storage/ Transportation Environment| Temperature: -25 °C ~ 70 °C (-13 °F ~
158°F) Humidity: ≤ 93 % R.H.
Operation Environment| Temperature: 5 °C ~ 40 °C (41 °F ~ 104 °F) Humidity: 15
% ~ 93 % R.H. Atmospheric pressure: 700hPa ~ 1060hPa
Power Supply| 1. DC 6V, AAA “LR03” (1.5V) alkaline battery x 4
2. DC 5V 1.2A AC/DC adapter (Optional)
Battery Life| Approx. 250 measurements
Shelf life (battery)| 3 years (Temperature: 20 ± 2°C; Relative humidity: 65 ±
20%RH)
Product Life| 5 Years (4 times per day)
Sleeping Mode| Without any operation for 1 minute, device automatically shuts
off.
Accessories| 4 AAA (LR03) Alkaline Batteries, Arm Cuff with Tube, Instruction
Manual, Pouch
RF Type| Bluetooth 5.0 BLE
System requirement of the Bluetooth-enabled device| Bluetooth 4.2 for Android
6.0 or above Bluetooth 4.2 for iOS 7.0 or above
*The contents of this manual and the specifications of the device covered by this manual are subject to change for improvement without notice.
Note
Explanation of symbols:
**Symbol**
| Explanation| Health & Life Information
---|---|---
| Refer to instruction manual/booklet| –
| TYPE BF Applied Part| –
| To avoid inaccurate results caused by electromagnetic interference| Warning:
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30cm (12 inches)
to any part of the device, Otherwise, degradation of the performance of this
equipment could result.
| Waste of electrical and electronic equipment (WEEE)| Discard the used
product to the recycling collection point according to local regulations.
| **** Serial number| YYMMXXXXXX
| Ingress Protection Rating| First characteristic numeral- Degree of
protection against access to hazardous parts and against solid foreign objects
N1=2 (Protected against solid foreign objects of 12.5 mm Ø and greater) Second
characteristic numeral- Degree of protection against ingress of water N2=2
(Protected against vertically falling water drops when ENCLOSURE tilted up to
15°)
| Non-ionizing electromagnetic radiation| –
Device information:
– Internally powered equipment
– Not suitable for use in presence of flammable anesthetic mixture with air or
with Oxygen or nitrous oxide
– Continuous operation with short-time loading
Manufacturer: HEALTH & LIFE CO., LTD.
9F, No. 186, Jian Yi Road, Zhonghe District 23553, New Taipei City, Taiwan
www.healthandlife.com.tw
Distributor: Homeric’s USA, LLC
3000 N. Pontiac Trail, Commerce Township, MI 48390
Phone: 1-800-466-3342
Note
*Note !
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation.
This equipment generates uses and can radiate radio frequency energy and, if
not installed and used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment
does cause harmful interference to radio or television reception, which can be
determined by turning the equipment off and on, the user is encouraged to try
to correct the interference by one or more of the following measures:
The user is encouraged to try to correct the interference by one or more of
the following measures:
❑ Reorient or relocate the receiving antenna.
❑ Increase the separation between the equipment and the receiver.
❑ Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
❑ Consult the dealer or an experienced radio/TV technician for help.
FCC RF Radiation Exposure Statement:
- The equipment complies with FCC RF exposure limits set forth for an uncontrolled environment.
- The equipment must not be co-located or operating in conjunction with any other antenna or transmitter.
FCC Label Compliance Statement:
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received, including interference
that may cause undesired operation.
*Note !
To assure continued FCC compliance:
Changes or modifications not expressly approved by the manufacturer could void
the user’s authority to operate the equipment.
HL868AW essential performance per IEC 80601-2-30 additional essential
performance requirements:
◼ 201.12.1.102 Limits of the error of the manometer from environmental
conditions
Over the temperature range of 5 °C to 40 °C (41 °F~ 104 °F) and the relative
humidity range of 15 % to 93 %(non-condensing), the maximum error for the
measurement of the CUFF pressure at any point of the NOMINAL measurement
range shall be less than or equal to ± 3 mmHg (± 0.4 kPa) or 2 % of the
reading, whichever is greater.
◼ 201.12.1.107 Reproducibility of the blood pressure determination
The laboratory Reproducibility of the BLOOD PRESSURE
DETERMINATION of the AUTOMATED SPHYGMOMANOMETER shall be less than 3 mmHg (0.4
kPa).
Appendix
◆ Guidance and manufacturer’s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed
below, and should only be used in such environments:
Emissions test| Compliance| Electromagnetic environment –
guidance
---|---|---
RF emissions CISPR 11| Group 1| RF energy is used only to maintain device’s
operation. Therefore, its RF emissions are so low that it’s not likely to
cause any interference in nearby electronic equipment.
RF emissions CISPR 11| Class B| The device is suitable for use in all
establishments, including domestic establishments, and those directly
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuations/ flicker emissions IEC 61000-3-3| Complies
◆ Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed
below, and should only be used in such environments:
Immunity test| IEC 60601 test level| Compliance level|
Electromagnetic environment – guidance
---|---|---|---
Electrostatic discharge (ESD) IEC 61000-4-2| ± 8 kV contact discharge
± 15 kV air discharge| ± 8 kV contact discharge
± 15 kV air discharge| In the case of air discharge testing, the climatic
conditions shall be within the following ranges:
Ambient Temperature:15℃~35℃, Relative Humidity:30%~60%.
Power frequency (50 or 60 Hz) magnetic field IEC 61000-4-8| 30 A/m
50 or 60 Hz| 30 A/m 50 or 60 Hz| Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical commercial or
hospital environment.
Electrical fast transient/burst IEC 61000-4-4| ± 2 kV for power supply lines
± 1 kV for input/output lines| ± 2 kV for power supply lines
± 1 kV for input/output lines| Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5| AC Power port
±1 KV Line to Line| AC Power port
±1 KV Line to Line| Mains power quality should be that of a typical commercial
or hospital environment.
interruptions and voltage variations on power supply input lines
IEC 61000-4-11
| 0% UT; 0.5 cycle
At 0°,45°,90°,135°,180°,225
°,270°and 315°.
0 % UT; 1 cycles
70 % UT; 25/30 cycles 0 % UT; 250/300 cycle| 0% UT; 0.5 cycle
At 0°,45°,90°,135°,180°,225
°,270°and 315°.
0 % UT; 1 cycles
70 % UT; 25 cycles
0 % UT; 250 cycle| Mains power quality should be that of a typical commercial
or hospital environment. If the user of the device requires continued
operation during power mains interruptions, it is recommended that the device
be powered from an uninterruptible power supply or a battery.
◆ Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed
below, and should only be used in such environments:
Immunity test| IEC 60601 test level| Compliance level|
Electromagnetic environment – guidance
---|---|---|---
Conducted RF IEC 61000-4-6| 3V rms
At 0.15-80 MHz
6V rms
At ISM & Radio Amateur Freq.| 3V rms
At 0.15-80 MHz
6V rms
At ISM & Radio Amateur Freq.| Portable and mobile RF communications equipment
should be used no closer to any part of the device, including cables, than the
recommended separation distance calculated from the equation applicable to the
frequency of the transmitter.
Radiated RFIEC 61000-4-3
(Proximity fields from RF wireless communications
equipment IEC 61000-4-3)| 10 V/m at 80-2700 MHz
AM Modulation And 9-28V/m at
385-6000MHz,Pulse Mode and
other Modulation. The system shall
be tested as specified in
IEC60601-1-2 Table 9 for proximity
fields from RF wireless communications equipment using the test methods
specified in IEC 61000-4-3| 10 V/m at 80-2700 MHz
AM Modulation And 9-28V/m at
385-6000MHz,Pulse Mode and
other Modulation. The system shall
be tested as specified in IEC60601-1-2 Table 9 for proximity fields from RF
wireless communications equipment using the test methods specified in IEC
61000-4-3| Recommended separation distance Considering to reduce the
minimum separation distance, based on RISK MANAGEMENT, and using higher
IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation
distance. Minimum separation distances for higher IMMUNITY TEST LEVELS shall
be calculated using the following equation: E = 6/d where P is the
maximum power in W, d is the minimum separation distance in m, and E is
the IMMUNITY TEST LEVELS in V/m.
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, a should be less than the compliance level in
each frequency range.b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
a. Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to
verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the device.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
Test specifications for enclosure port immunity to RF wireless communications equipment.
Test frequency (MHz) | Modulation | IMMUNITY TEST LEVEL (V/m) |
---|---|---|
385 | Pulse modulation 18 Hz a) | 27 |
450 | FM ± 5 kHz deviation 1kHz sine b) | 28 |
710 | Pulse modulation 217 Hz a) | 9 |
745
780
810| Pulse modulation 18 Hz a)| 28
870
930
1720| Pulse modulation 217 Hz a)| 28
1845
1970
2450| Pulse modulation 217 Hz a)| 28
5240| Pulse modulation 217 Hz a)| 9
5500
5785
NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance
between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be
reduced to 1 m, The 1 m test distance is permitted by IEC 61000-4-3.
a). The carrier shall be modulated using a 50% duty cycle square wave signal.
b). AS an alternative to FM modulation, 50% pulse modulation at 18 Hz may be
used because while it does not represent actual modulation, it would be worst
case.
Blood Pressure Diary
Date:| Time:| | □Before
□After| Meal
---|---|---|---|---
Systolic / Diastolic:| | Pulse:|
Date:| Time:| | □Before
□After| Meal
Systolic / Diastolic:| | Pulse:|
Date:| Time:| | □Before
□After| Meal
Systolic / Diastolic:| | Pulse:|
Date:| Time:| | □Before
□After| Meal
Systolic / Diastolic:| | Pulse:|
Date:| Time:| | □Before
□After| Meal
Systolic / Diastolic:| | Pulse:|
Date:| Time:| | □Before
□After| Meal
Systolic / Diastolic:| | Pulse:|
Date:| Time:| | □Before
□After| Meal
Systolic / Diastolic:| | Pulse:|
Date:| Time:| | □Before
□After| Meal
Systolic / Diastolic:| | Pulse:|
Date:| | Time:| | □Before
□After| Meal
Systolic / Diastolic:| | Pulse:|
Date:| | Time:| | □Before
□After| Meal
Systolic / Diastolic:| | Pulse:|
P/N:323103780
VER:A001 20230419
Documents / Resources
|
Health Life HL868AW Automatic Upper Arm Blood Pressure
Monitor
[pdf] Instruction Manual
HL868AW Automatic Upper Arm Blood Pressure Monitor, HL868AW, Automatic Upper
Arm Blood Pressure Monitor, Arm Blood Pressure Monitor, Blood Pressure
Monitor, Pressure Monitor, Monitor
---|---