NeuroNexus AVIH Unboxing and Implant Instruction Manual
- June 15, 2024
- NeuroNexus
Table of Contents
NeuroNexus AVIH Unboxing and Implant
Product Usage Instructions
Unboxing
- Slowly lift the package and array up out of the holder and box, being mindful of the probe.
Attach Package IST-2mm Rod
- Take the provided IST-2mm rod and thread it into the IST-Holder while holding the sides of the package.
- If not using AVH/AVIH-IST probe inserter mount:
- Attach the IST-2mm rod to a stereotaxic arm.
- Once secure, lower the array down for implantation.
Attach IST-AVH/AVIH
- Place IST-AVH/AVIH on the back side of the IST Holder.
- Thread the two pieces together with two M2x4mm pan head Phillips screws, being mindful of the probe to prevent damage.
Mounting Probe to IST-AVH/AVIH
- Following the PEG procedure, slide the probe onto the probe platform area.
- Hold the probe in place until PEG hardens, securing the probe to the IST-AVH/AVIH.
- Attach the IST-2mm rod to the stereotaxic frame. The probe is now ready for insertion, being mindful of the shanks of the probe to prevent damage.
Remove IST-AVH/AVIH
- Release the probe from the AVH/AVIH-IST probe insertion mount following the PEG procedure.
- Remove the two backside M2 pan head screws holding the IST-AVH/AVIH to the IST holder.
- Carefully remove the IST-AVH/AVIH, ensuring that the probe is completely detached and being mindful of the probe to prevent damage.
Remove IST-Holder
- If you need to remove the package from the stereotaxic frame and secure it to the skull with dental cement, you can remove the IST-holder.
- Remove the two M2 screws holding the front plate on.
- Carefully remove the front plate, making sure the package doesn’t fall out and ensuring a good handle on the front plate.
FAQ
- Q: Where can I find general surgical guidelines and craniotomy instructions?
- A: You can find general surgical guidelines and craniotomy instructions here.
- Q: What should I do if I need a replacement?
- A: Please reach out to your account manager to inquire about a replacement.
Introduction
Read this manual carefully before planning your surgery.
This manual applies the following packages:
AVH64, AVIH64,AVH128, AVIH128, AVH256 and AVIH256
- Examine the probe before opening the box.
- Make sure the probe hasn’t broken free from the gel-film, or the package hasn’t freed itself from the shipping holder.
- Note: Any damage one might see that has occurred before the box has been opened.
- If there is visual damage, please document it with photos.
- Then reach out to your account manager to see about replacement.
General surgical guidelines and craniotomy instructions can be found here: https://www.neuronexus.com/files/technicalsupportdocuments/General- Surgery-Guidlines.pdf
AVH/AVIH Box Overview
Installation Instructions
Remove AVH/AVIH Package From Shipping Box
- A. Carefully remove any tape outside the box. NOTE: It is wise to discard the tape to prevent it accidentally sticking to the polyimide cable.
- B. Open the box slowly. Before removing the probe, remove the film over the green gel-film (highlighted red) holding the probe in place. Do not remove the probe until the film has been removed.
- C. Remove the M2 screw holding the package in the L-shipping bracket. Do not leave the screw in the bottom of the box as this could lead to damaging the probe. Before removing the package from the box, using a flat tweezer, carefully pry the probe so it is free from the gel-film. Be mindful of the probe so it doesn’t crash into the side of the box.
- D. Grip the sides of the package (see blue arrows) and with flat tweezers hold the cable just above the probe (see blue arrow).
- E. Slowly lift the package and array up out of the holder and box. Be always mindful of the probe.
Attach IST-AVH/AVIH
- A. Place IST-AVH/AVIH on back side of IST Holder
- B. Thread the two pieces together with two M2x4mm pan head Phillips screws. NOTE: Be always mindful of the probe at all times, lessening the chance of it getting damaged.
Mounting Probe to IST-AVH/AVIH
- A. Following the PEG procedure, slide the probe on to the probe platform area.
- B. There, one will hold in place the probe until PEG hardens securing the probe to the IST-AVH/AVIH.
- C. When probe is securely in place, one can attach the IST-2mm rod to stereotaxic frame. Now probe is ready for insertion. NOTE: Be always mindful of the shanks of the probe at all times, lessening the chance of it getting damaged.
Remove IST-AVH/AVIH
- A. Release the probe from the AVH/AVIH-IST probe insertion mount, by following the PEG procedure.
- B. Remove the two backside M2 pan head screw holding the IST-AVH/AVIH to IST holder.
- C. Once screws are moved, carefully remove the IST-AVH/AVIH. Make sure the probe is completely detached.
- NOTE: Be always mindful of the probe at all times, lessening the chance of it getting damaged.
Remove IST-Holder
- A. If one must remove the package from the stereotaxic frame and would like to secure the package to the skull with dental cement, the IST-holder can be removed.
- B. Remove the two M2 screws holding the front plate on.
- C. Once those are removed then the front plate can be removed. Make sure that package doesn’t fall out. Also make sure you have a good handle on the front plate as you don’t want to drop it and have it crash into your implanted probe.
- D. Holding the package, press on the IST-Holder carefully so that it pops out and away from the package. Now, the package can be secured to the skull using whatever approved in-house method.