Femometer DMT-427 Thermometer for Adults Instruction Manual

June 14, 2024
Femometer

Femometer DMT-427 Thermometer for Adults Instruction Manual

Indications For Use : This digital thermometer provides a quick and highly accurate reading of an individual’s body temperature. The digital thermometer is intended to measure the human body’s temperature in regular mode orally, rectally or under the arm, and the device is reusable for clinical or home use on people of all ages.

Please Read Carefully Before Using: This appliance conforms to the following standards: EN 12470-3 Clinical thermometers-Part 3: Performance of compact electrical thermometers (non-predictive and predictive) with maximum device, ISO 80601-2-56 Medical electrical equipment-Part 2-56:Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement, EN 60601-1-11 Medical electrical equipment-Part 1-11: General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment and complies with the requirements of EN 60601-1-2(EMC), JEC/EN60601-1 (Safety) standards. And the manufacturer is ISO 13485 certified.

KET TO SYMBOLS

Femometer DMT-427 Thermometer for Adults - KET TO
SYMBOLS

GENERAL WARNINGS

  1. Read instructions thoroughly before using digital thermometer. 2) Choking Hazard: Thermometer cap and battery may be fatal if swallowed. Do not allow children to use this device without parental supervision. 3) Do not use thermometer in ear. Designed use is for oral, rectal, and armpit (axilla) readings only. 4) Do not place thermometer battery near extreme heat as it may explode. 5) Note: Use of the probe cover may result in a 0.1 °C(0.2°F) discrepancy from actual temperature. 6) Remove battery from the device when not in operation for a long time. 7) The use of temperature readings for self- diagnosis is dangerous. Consult your doctor for the interpretation of results. Self-diagnosis may lead to the worsening of existing disease conditions. 8) Do not attempt measurements when the thermometer is wet as inaccurate readings may result. 9) Do not bite the
    thermometer. Doing so may lead to breakage and/or injury. 10) Do not attempt to disassemble or repair the thermometer. Doing so may result in inaccurate readings. 11) After each use, disinfect the thermometer especially in case the device is used by more than one person. 12) Do not force the thermometer into the rectum. Stop insertion and abort the measurement when pain is present. Failure to do so may lead to injury. 13) Do not use thermometer orally after being used rectally. 14) For children who are two years old or younger, please do not use the devices orally. 15) If the unit has been stored at temperatures over 5°C ~40°C(4 I °F ~I 04°F), leave it in 5°C ~40°C(41 °F ~104°F) ambient temperature for about 15 minutes before using it.

SPECIFICATIONS

  • Type: Digital Thermometer (Not Predictive)

  • Measure Range: 32.0°C-42.9°C (90.0°F-109.9°F) (°C /°Fchosen by manufacturer )

  • Accuracy: ±0.1°C (±0.2°F) during 35.5°C~42.0°C (95.9°F~107.6°F) at 18°C~28°C (64.4°F~82.4°F) ambient operating range ±0.2°C (±0.4°F) for other measuring and ambient operating range

  • Operating mode: Direct Mode

  • Display: Liquid crystal display, 3 1/2 digits Memory: For storing the last measured value

  • Battery: One 1.5 V DC. button battery (size LR41or SR41, UCC 392)

  • Battery life: Approx. 200hours of continuous operation or 1 year with 3 measurements per day

  • Dimension: 12.4cm1.9cm1.1cm (L x W x H)

  • Weight: Approx. 11 grams including battery

  • Expected service life: Three years

  • Ambient operating range:

  • Temperature: 5°C~40°C (41°F~104°F) Relative humidity: 15%~95%RH Atmospheric Pressure : 700hPa ~ 1060hPa

  • Storage and transportation condition:
    Temperature: -20°C~55°C (-4°F~131°F) Relative humidity: 15%~95%RH Atmospheric Pressure : 700hPa ~ 1060hPa

  • Classification: Type BF

DIRECITIONS

  1. Press the On/Off Button next to LCD display. A tone will sound as the screen shows , followed by last recorded temperature. After showing the self-test temperature, the thermometer is now in the testing mode.

  2. Position thermometer in desired location (mouth, rectum, or armpit.)
    a) Oral Use: Place thermometer under tongue as indicated by “√” position shown in Figure. Close your mouth and breathe evenly through the nose to prevent the measurement from being influenced by inhaled/exhaled air. Normal temperature between 35.7°C and 37.3°C (96.3°F and 99.1°F).
    b) Rectal Use: Lubricate silver probe tip with petroleum jelly for easy insertion. Gently insert sensor approximately 1cm (less than 1/2 “) into rectum. Normal temperature between36.2°Cand 37.7°C (97.2°F and 99.9°F).
    c) Armpit Use: Wipe armpit dry. Place probe in armpit and keep arm pressed firmly at side.
    From a medical viewpoint, this method will always provide inaccurate readings, and should not be used if precise measurements are required.
    Normal temperature between 35.2°C and 36.7°C (95.4°F and 98.1°F).

  3. The degree sign flashes throughout the testing process. When flashing stops an alarm will beep for approximately 10 seconds. The measured reading will appear on the LCD simultaneously. The minimum measurement time until the signaling tone (beep) must be maintained without exception. The measurement continues even after the buzzer notification.
    *Note: Normally the buzzes are “Bi-Bi-Bi- Bi-”; Alarm beeps more rapidly when temperature reaches 37.8°C (100.0°F) or higher, and the buzzes are” Bi- Bi-Bi—– Bi-Bi-Bi——- Bi-Bi-Bi”

  4. To prolong battery life, press the On/Off Button to turn unit off after testing is complete. If no action is taken, the unit will automatically shut off after around 10 minutes.

INSERTING/REPLACING THE BATTERIES

  1. Pull the battery cover in the direction shown on the arrow. 2) Carefully extract the battery compartment to the right to ease the battery removal.
    Use a pointed object to remove the old batteries. 3) Place the new batteries with the positive pole (+) pointed upwards and the negative pole pointed downwards. 4) Gently push the battery compartment into its seat and re- assemble the cover. Should the display show the symbol” “, it means that they have to be replaced. 5) Remove the used batteries as described under points 1 -2 and dispose of them as special waste at the collection points for recycling. For further information concerning the disposal of exhausted batteries, refer to the store where the device containing the batteries was purchased, the Municipality or the local waste disposal service. Caution: For safety reasons, it is recommended never to allow children younger than 12 years to remove the batteries.

Femometer DMT-427 Thermometer for Adults -
TROUBLESHOOTING

MAINTENANCE

Clean the thermometer probe using a cloth moistened with alcohol or with any type of topic disinfectant before and/or after each use. Any other type of solution can damage the device. The thermometer body has to be cleaned using a dry cloth. Do not immerse the thermometer in water or other liquids. Never use chemical or abrasive products. The probe is the most fragile part of the thermometer. Pay utmost care during its cleaning. Do not use alcohol to clean the display. Preserve the thermometer in a dry and clean place, away from direct sunlight. Do not sterilize or boll the thermometer. Preserve the thermometer carefully In Its standard-supplied casing. Do not attempt to disassemble or repair the thermometer. Doing so may result In Inaccurate readings. It Is recommended the performance should be checked every 2 years.

DISPOSAL PROCEDURE

The symbol on the bottom of the device indicates the separated collection of electric and electronic equipment (Dir. 2012/19/Eu-WEEE). At the end of life of the device, do not dispose it as mixed solid municipal waste, but dispose it referring to a specific collection center located in your area or returning it to the distributor, when buying a new device of the same type to be used with the same functions, if
the appliance to be disposed of is less than 25 cm, it can be returned to a retail location that is over 40 without having to purchase a new similar device. This procedure of separated collection of electric and electronic devices is earned out forecasting a European environmental policy aiming at safeguarding, protecting and improving environment quality, as well as avoiding potential effects on human health due to the presence of hazardous substances in such equipment or to an improper use of the same or of parts of the same.

Caution! The wrong disposal of electric and electronic equipment may Involve sanctions. To correctly dispose of batteries (Dir. 2013/56/Eu) do not throw them Into household waste but rather dispose of them as special waste at specialised recycling centres. For more Information about disposal of used batteries, contact the store where you purchased the equipment containing the batteries, the municipality or the local waste disposal service.

LIMITED WARRANTY

The thermometer is guaranteed for one year from the date of purchase. If the thermometer does not function properly due to defective components or poor workmanship, we will repair or replace it free of charge. All components are covered by this warranty excluding the battery. The warranty does not cover damages to your thermometer due to improper handling. To obtain warranty service, an original or copy of the sales receipt from the original retailer is required.
Question ? E-mail support@femometer.com

Disposal of this product and used batteries should be carried out in accordance with the national regulations for the disposal of electronic products.

Made in China

JOYTECH HEALTHCARE CO.LTD.
No.365, Wuzhou Road, Yuhang Economic Development Zone,
Hangzhou city, 311100 Zhejiang,China

PRODUCT ILLUSTRATION

Femometer DMT-427 Thermometer for Adults - Figure 1

Electromagnetic Compatibility Information

The device satisfies the EMC requirements of the international standard !EC 60601-1-2. The requirements are satisfied under the conditions described in the table below. The device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use. Portable and mobile HF communications equipment can affect the device. Use of the unit in conjunction with non-approved accessories can affect the device negatively and alter the electromagnetic compatibility. The device should not be used directly adjacent to or between other electrical equipment.

Table 1

Table 2

Femometer DMT-427 Thermometer for Adults - Table 2

Table 3

Femometer DMT-427 Thermometer for Adults - Table 3

Table 4

Femometer DMT-427 Thermometer for Adults - Table 4

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Femometer User Manuals

Related Manuals