Ingestible Adherence Tracker System
Product Information:
The ID-Cap System is a medical device developed by etectRx. It is
classified as an ingestible event marker (IEM) and has been cleared
by the United States Food and Drug Administration (FDA) for
marketing in the United States. The system consists of the ID-Cap
Reader, ID-Capsules, and the optional ID-Cap App. The device model
numbers are as follows: – ID-Cap Reader: Model Number ET71001.00 –
ID-Capsules: Model Numbers ET30001.51, ET30001.52 Product Usage
Instructions:
1. Before using the ID-Cap System, it is important to read and
fully understand the contents of the User Manual to ensure safe and
effective operation of the device. 2. The ID-Cap System is intended
for use by authorized users, including patients who have been
registered to use the system and clinicians or program
administrators who have been granted access to view ingestion event
data. 3. The system can be used with the ID-Cap App, which is
optional. Instructions for using the ID-Cap System with the app are
provided in the “Instructions for Use of the ID-Cap System with the
ID-Cap App” section of the User Manual. 4. In case of any issues or
difficulties, refer to the “Troubleshooting” section of the User
Manual for assistance. 5. For further assistance or inquiries,
refer to the “User Assistance” section or contact the manufacturer
directly. 6. Important notices regarding warranty and trademark
information can be found in the “Notices” section of the User
Manual. Note: The ID-Cap System is for prescription-use only.
ID-Cap System USER MANUAL
Pendant Reader
This User Manual provides important information about the ID-Cap System from etectRx. You must read and fully understand the contents of this User Manual before using the ID-Cap System to ensure safe and effective operation of the device.
Device Model Numbers
This User Manual pertains to the following device model numbers:
Model Number for the ID-Cap Reader
ET71001.00
Model Number for the ID-Capsule
ET30001.51, ET30001.52
1
TABLE OF CONTENTS Device Model
Numbers……………………………………………………………………………………………………………….. 1
Device Model Numbers 1……………………………………………………………………………………………………………2
GLOSSARY………………………………………………………………………………………………………………………………. 3
ID-CAP SYSTEM OVERVIEW ………………………………………………………………………………………………… 4
Introduction ………………………………………………………………………………………………………………………………………….4
Prescription-Use Only ……………………………………………………………………………………………………………………………4
Indications for Use ………………………………………………………………………………………………………………………………..4
Conditions of Use…………………………………………………………………………………………………………………………………..4 Device
Description- Pendant Reader ………………………………………………………………………………………………………4
The ID-Cap System…………………………………………………………………………………………………………………………4 ID-Capsule
(Ingestible Sensor Capsule) ……………………………………………………………………………………………5 ID-Cap Reader
………………………………………………………………………………………………………………………………5 Reader Display
………………………………………………………………………………………………………………………………6 The ID-Cap Reader Kit that
is available by prescription-only contains:…………………………………………………6 Mobile
Application (ID-Cap App) for Your Smartphone ……………………………………………………………………..6
Patient Users …………………………………………………………………………………………………………………………………………7 Locations
of Use…………………………………………………………………………………………………………………………………….7 When the Device
Should Not Be Used …………………………………………………………………………………………………….7 Expectations of
the Device ……………………………………………………………………………………………………………………..8 Data Captured by
the ID-Cap System…………………………………………………………………………………………………….9
IMPORTANT USE INFORMATION ………………………………………………………………………………………… 9
INSTRUCTIONS FOR USE OF THE ID-CAP SYSTEM WITH THE ID-CAP APP …………………. 11
Initial Setup of the ID-Cap Reader and its Charging Clip
……………………………………………………………………..11 Initial Setup of the ID-Cap App
……………………………………………………………………………………………………………11 Pair Your Reader and Phone Before
Use………………………………………………………………………………………………12 Routine Use of the ID-Cap System
with the ID-Cap App……………………………………………………………………….12 Setting up the
charger:…………………………………………………………………………………………………………………………14 To charge the
Reader:………………………………………………………………………………………………………………………….15 Device Return
……………………………………………………………………………………………………………………………………..15 Monitoring the Activity
of the Device……………………………………………………………………………………………………15 Conditions Affecting
Device Performance and Use………………………………………………………………………………..15 Maintenance
……………………………………………………………………………………………………………………………………….. 16 Cleaning the ID-Cap
Reader ………………………………………………………………………………………………………………..16 Storage and Handling
………………………………………………………………………………………………………………………….16
Storage and Operating Requirements ……………………………………………………………………………………………..16
Useful Life and Shelf Life ………………………………………………………………………………………………………………16
Disposal of Waste Products …………………………………………………………………………………………………………..16
Compliant Use …………………………………………………………………………………………………………………………………….16 Ordered by
Doctor or Licensed Practitioner ……………………………………………………………………………………16 Avoiding
Unsafe Use Conditions…………………………………………………………………………………………………….17
TROUBLESHOOTING …………………………………………………………………………………………………………… 18
USER ASSISTANCE ……………………………………………………………………………………………………………….. 19
MANUFACTURER …………………………………………………………………………………………………………………. 19
NOTICES ……………………………………………………………………………………………………………………………….. 20
Warranty ……………………………………………………………………………………………………………………………………………. 20 Trademark
………………………………………………………………………………………………………………………………………….20
2
Copyright ……………………………………………………………………………………………………………………………………………20 Quality
Assurance ……………………………………………………………………………………………………………………………….20 Device
Classification…………………………………………………………………………………………………………………………….21 Symbols
Reference Guide …………………………………………………………………………………………………………………….21 Environmental
Conditions of Use …………………………………………………………………………………………………………23 Biocompatibility
………………………………………………………………………………………………………………………………….23 Protection Against
Ingress of Solids and Liquids…………………………………………………………………………………..23 Regulatory
Information ……………………………………………………………………………………………………………………….24
Declaration of Conformity ……………………………………………………………………………………………………………24 Applied
Part ………………………………………………………………………………………………………………………………..24 CISPR Interference
Statement………………………………………………………………………………………………………24 FCC Interference Statement
…………………………………………………………………………………………………………24 FCC Compliance Identifiers
…………………………………………………………………………………………………………24 Electromagnetic
Compatibility………………………………………………………………………………………………………25 Information on the
Radio Subsystem …………………………………………………………………………………………….25
GLOSSARY
Term Authorized User Caution
etectRx ID-Cap App (optional)
ID-Cap Reader ID-Cap Reader Kit ID-Cap System ID-Capsule ID-Tag Warning
Definition Patient who has been registered to use the ID-Cap System or a
clinician or program administrator who has been granted access to view the
ingestion event data Important notification or instruction regarding a
hazardous situation which can cause material damage or lead to minor or
moderate injury
ATTENTION Hazardous situation which can cause material damage
or lead to minor or moderate injury
Manufacturer of the ID-Cap System Smartphone application for the patient that
may be used to report and display ingestion event data, transfer data to the
secure server for viewing by others as authorized, send notifications to
users, provide updates on the status of the Reader, and manually record an
ingestion event that has not been recorded by the ID-Cap System Wearable
reader which receives the signal from the ID-Tag and confirms the ingestion
event Kit comprised of the ID-Cap Reader, Reader accessories, and instruction
manuals supplied to the patient System comprised of a wearable ID-Cap Reader,
ID-Capsules, and a mobile application for the patient Capsule for oral
administration which contains the ID-Tag Ingestible sensor which transmits a
signal once ingested and is subsequently excreted Important notification or
instruction regarding a hazardous situation which can cause a serious or fatal
injury
WARNING Hazardous situation which can cause a
serious or fatal injury
3
ID-CAP SYSTEM OVERVIEW
Introduction
The ID-Cap System from etectRx is a medical device that has been cleared by
the United States Food and Drug Administration (FDA) for marketing in the
United States. The ID-Cap System is classified by the FDA as an ingestible
event marker (IEM).
Prescription-Use Only
CAUTION: Federal law restricts this device to sale by or on the order of a
licensed practitioner.
Indications for Use
The ID-Cap System consists of a wearable reader for ambulatory recording of
events signaled by swallowing the IDCapsule which contains the ID-Tag, an
ingestible sensor. The ID-Cap System is intended to log, track, and trend
intake times and enables unattended data collection for clinical applications.
The ID-Cap System may be used in any instance where quantifiable analysis of
ingestion events, including events signaled by the co-incidence with or
coingestion with the ID-Capsule, is desirable.
Conditions of Use
The ID-Cap System records an event signaled by swallowing the ID-Capsule. ID-
Capsules may be taken on a specific schedule to mark ingestion events. When
used to mark events signaled by the co-incidence with or coingestion with the
ID-Capsule, the ID-Capsule should be taken at the same time as the specific
event of interest or at the same time as a dose of medication.
The ID-Cap System should be used with the patient’s use of our mobile
application. The ID-Cap App collects the ingestion event and Reader status
messages from the Reader and forwards them via the mobile network to the IDCap
secure server, allowing authorized users to view ingestion events as they
occur. The ID-Cap App reports and displays information about ingestion events
and ingestion event history, sends notifications to the user, and provides
updates on the status of the ID-Cap Reader. It can also allow the patient to
manually record an ingestion event that has not been recorded by the ID-Cap
System when applicable. If the ID-Cap App and phone are not used with the ID-
Cap System, the Reader may collect and store information that can be
downloaded at a later time.
Device Description- Pendant Reader
The ID-Cap System
Your ID-Cap System consists of a wearable ID-Cap Reader, ID-Capsules, and a
mobile application for recording of events signaled by swallowing the ID-
Capsule which contains an ingestible sensor. Your ID-Cap Reader is worn like a
watch and collects information from the ingested ID-Capsules that are taken as
directed by your doctor. Your doctor may instruct you to take the ID-Capsule
on a specific schedule, at the same time as a specific event, or with a
specific medication. The ingestible sensor in the ID-Capsule communicates with
your ID-Cap Reader after it reaches your stomach. This allows the system to
measure the pattern, regularity, and schedule of your ingestion events. With a
simple Bluetooth connection to your mobile device, your ID-Cap Reader sends
information to the ID-Cap App to help you keep track of your ingestion events.
With your permission, this information will be shared with your healthcare
provider to help them improve your health outcomes.
4
ID-Capsule (Ingestible Sensor Capsule)
ID-Capsules are available by prescription only for use with the ID-Cap Reader.
ID-Capsules are taken by mouth. The ID-Capsule is a capsule shell that
contains an ingestible sensor called the ID-Tag. The ID-Tag is made of
ingredients that do not easily react with other chemicals and are found in the
food chain. The ingestible sensor communicates with your ID-Cap Reader after
it reaches your stomach. It is powered by your stomach fluid. The ID-Tag has
been designed to be safe for human consumption. Clinical studies have
demonstrated the safety of ID-Capsules. Animal tests have demonstrated that
the ID-Tag is biocompatible and non-toxic. Key ingredients of the ingestible
ID-Tag that are in human contact include magnesium, silver, silver chloride,
polyimide, and epoxy. Additional information about ID-Capsules, including the
specific ingredients of the capsule shells, may be requested from the
manufacturer. The ID-Tag passes through your digestive system intact and is
excreted in the feces.
WARNING DO NOT take an ID-Capsule if you are allergic to any of its
ingredients. Inform your doctor of any allergies and consult with your doctor
before using.
ID-Cap Reader
The ID-Cap Reader is available by prescription only for use with ID-Capsules.
The ID-Cap Reader is a lanyardworn reader which you wear around your neck. The
ID-Cap Reader detects ingested ID-Capsules. The Reader does not require skin
contact to detect the signal and does not require an adhesive patch. Animal
tests and clinical studies have demonstrated that the Reader is biocompatible
and non-irritating. The Reader must be worn with the display screen facing
away from your body. Your ID-Cap Reader may be worn during most activities,
including exercise, but should be removed prior to any activities that would
expose the Reader to water, including showering, bathing, and swimming.
ATTENTION CAUTION: DO NOT submerge the ID-Cap Reader in or expose it to
excessive amounts of water. Your ID-Cap Reader is designed to be weather
resistant, but it may not work properly when exposed to water or other
liquids. The Reader contains a lithium polymer rechargeable battery. It is
charged using a charging clip that is supplied in the Reader Kit. The Reader
connects with authorized computing devices, such as tablets or smartphones,
using Bluetooth technology.
5
Reader Display
What the Reader Display indicates:
· If the display screen is black, the Reader is powered OFF. Once powered ON,
DO NOT turn the Reader off, even when not in use. The Reader conserves power
when not in use.
· The reader can be turned on by pressing and holding the button for 3
seconds. · When the Reader is positioned properly on the supplied charging
clip, the Reader display will turn on. · Please check the battery which will
display on your home screen. · An ingestion can be detected with just one bar
of battery. · The pill icon will display for approximately one hour when a
pill is detected, and then revert back to time of
day.
ID-Cap Reader Kit
The ID-Cap Reader Kit that is available by prescription-only contains: · ID-
Cap Pendant Reader with pre-attached Lanyard · USB-C Charger Clip · AC Adapter
· Quick Start Guide
Mobile Application (ID-Cap App) for Your Smartphone
The ID-Cap System may be used with a mobile application, the ID-Cap App, which
can be loaded on your smartphone. The ID-Cap App: · Allows the user to view
and track ingestion events · Transfers data to the secure server for viewing
by your healthcare provider(s) and others as authorized · Sends notifications
to the user, such as acknowledgements of reported ingestion events, low
battery alerts, and
reminder notifications · Provides updates on the status of the Reader and
System · Can allow the user to manually record an ingestion event that has not
been recorded by the ID-Cap System or if
the ID-Cap System was not used (if applicable). The ID-Cap App collects the
ingestion event and Reader status messages and forwards them via the mobile
network to the ID-Cap secure server, allowing authorized users to view
ingestion events as they occur. Your healthcare provider may view your
ingestion data on a web-based system called the ID-Cap Dashboard. The ID-Cap
App on
6
your phone reports and displays information about your ingestion events, sends
notifications to you, and provides updates on the status of your Reader and
the System. If you are receiving reminder notifications for use of the IDCap
System, you have the option to stop or modify them in your phone App settings.
The App also allows you to manually record an ingestion event that has not
been recorded by the ID-Cap System, if applicable. Refer to your doctor or
study team for information about the functionality of the App, instructions
for use, and minimum requirements for your smartphone to run the App.
Patient Users
Federal law restricts this device by prescription only of a licensed
practitioner. The ID-Cap System is intended to be used by patients for
ambulatory recording of events signaled by swallowing the ID-Capsule.
ATTENTION Your healthcare provider should evaluate you, as well as the risks
and benefits of the device, to ensure that the decision to use the ID-Cap
System is appropriate. The patient (or patient’s caregiver) must be able to
read, understand, and follow the instructions provided in the User Manual and
Quick Start Guide. The patient (or patient’s caregiver) must be able to set
up, charge, and use the IDCap Reader and follow instructions regarding how and
when to take ID-Capsules to mark ingestion events. Use of the ID-Cap System is
reserved for users who can follow ID-Capsule ingestion instructions without
interfering with other health behaviors, such as properly taking prescribed
medication.
Locations of Use
The ID-Cap System has been designed for ambulatory use, enabling unattended
data collection for clinical applications.
ATTENTION CAUTION: The ID-Cap System is not intended for use on aircraft.
See Technical Information in Appendix A of this User Manual for additional
restrictions on locations of use due to electromagnetic interference. The AC
adapter for the charging clip may not be compatible with power sources other
than 120V, 60Hz (U.S. Standard). If being used outside of the United States,
the user may need an adapter or a converter to convert to the proper voltage.
When the Device Should Not Be Used
Contraindications are conditions under which the device should not be used
because the risk of use clearly outweighs any possible benefit.
WARNING DO NOT use the ID-Cap System in patients or situations in which it is
contraindicated as summarized below.
7
The ID-Cap System should not be used in the following situations: · DO NOT use
the ID-Cap System if you are pregnant or breastfeeding. Use in pregnant or
breastfeeding women
has not been studied. · Avoid exposure to known sources of electromagnetic
interference (EMI) with medical devices, such as
magnetic resonance imaging (MRI) systems, diathermy, lithotripsy,
electrocautery, Radio Frequency Identification (RFID) devices, and
electromagnetic security systems, such as metal detectors. Note that the
presence of RFID devices may not be obvious. · DO NOT use the ID-Cap System if
you have a cardiac pacemaker or other implanted electronic medical device. The
ID-Cap System has not been studied in individuals with implanted electronic
medical devices. · DO NOT use the ID-Cap System if you are unable to swallow
oral medications. · DO NOT use the ID-Cap System if you are allergic to any of
the ingredients in the ingestible ID-Capsule. · DO NOT use the ID-Cap System
if you have a significant medical condition which may affect capsule passage
through your digestive system.
Risks and Benefits
The ID-Cap System records events signaled by swallowing ID-Capsules. The
System logs, tracks, and trends intake times for clinical use. This allows the
ID-Cap System to measure and report the pattern, regularity, and schedule of
your ingestion events. This information may help you or your healthcare
provider better understand your health behaviors and improve your health
outcomes.
The risks of the ID-Cap System must be evaluated in light of its benefits and
limitations. The risks of the ID-Cap System include:
· Risk of incomplete, inaccurate, or misinterpreted ingestion event data that
may impact treatment decisions · Minor, self-limiting adverse events, such as
minor digestive problems and the sensation of the capsule in
the esophagus · Risks associated with failure to eliminate the ID-Tag in the
feces · Risk of breach of confidentiality, privacy, or security involving
ingestion data · Risk of allergic reactions (including life-threatening
reactions) when taking any ingestible product
Expectations of the Device
When the ID-Capsule is ingested, the capsule will soften and disintegrate.
During this process, the ID-Tag will power up and begin sending messages. The
ID-Cap Reader detects the messages from the ingested ID-Capsule. The average
time for the Reader to detect an ID-Capsule ingestion is about 6.5 minutes
after the ingestion. The ID-Tag will continue to send messages repeatedly
until it runs out of power in approximately 20-30 minutes. When it is worn as
instructed, the Reader will continue to record these messages from the ID-Tag
and store this information in its memory. The ID-Tag passes through your
digestive system intact and is excreted in the feces.
The ID-Cap System is used effectively with the patient’s use of the mobile application. When the Reader is paired with the patient’s smartphone using the ID-Cap App, the user may receive a notification of the ingestion event on his or her smartphone about 6.5 minutes on average after taking the ID-Capsule. It can also allow the patient to manually record an ingestion event that has not been recorded by the ID-Cap System if applicable. The ID-Cap App collects the ingestion event and Reader status messages and forwards them via your phone’s mobile network to the ID-Cap monitoring database. This enables authorized users, such as your doctor, to view ingestion events as they occur when the phone is paired with the Reader. No personal or protected health information (PHI) is sent from the ID-Tag, ID-Cap Reader, or ID-Cap App
Parameter Time to Detection on ID-Cap Reader from ID-Capsule Ingestion Average Duration of ID-Tag Detections by ID-Cap Reader Time to Phone Notification from ID-Capsule Ingestion Time to Server Notification from ID-Capsule Ingestion
Results (mean) 6.4 minutes 27.9 minutes 6.5 minutes 6.5 minutes
8
Data Captured by the ID-Cap System
The following data are collected and reported using the ID-Cap System: · Date
and time of ingestion events detected by the Reader (signaled by swallowing
the ID-Capsule while wearing the Reader) · Date and time of ingestion events
manually recorded by the patient using the ID-Cap App · Reader status messages
The Reader status messages communicate information about whether your Reader
is on, its battery level, its operational state (that is, whether or not it is
ready for use in detecting ingestions of the ID-Capsule), and its movement.
The movement of the device is measured by an accelerometer within the device.
No other data is collected by the accelerometer other than a change in the
position of the Reader. The Reader status messages also collect some
information about your use of the Reader. The information that is captured
includes whether the Reader is powered ON or OFF, the Reader battery level,
whether the Reader is on or off the charging pad, and Reader movement.
When paired with the phone, the Reader stays in regular communication with the
ID-Cap App and secure server by sending Reader status messages. When not
paired with the App, these messages are stored on the Reader for future
download.
IMPORTANT: No personal or protected health information (or “PHI”) is sent from
the ingestible sensor, ID-Cap Reader, or ID-Cap App. There is not a Global
Positioning System (GPS) in the Reader, and the Reader does not collect any
location information. The ID-Cap App also does not capture or track any
location information.
IMPORTANT USE INFORMATION
The ID-Cap System records an event signaled by swallowing the ID-Capsule. ID-
Capsules may be taken on a specific schedule to mark ingestion events. When
used to mark events signaled by the co-incidence with or coingestion with the
ID-Capsule, the ID-Capsule should be taken at the same time as the specific
event of interest or at the same time as a dose of medication. Your doctor
will prescribe an ingestion schedule and provide instructions for taking the
ID-Capsules. The ID-Cap System logs, tracks and trends your ID-Capsule
ingestion times.
Please keep in mind the following while using the ID-Cap System: · The ID-Cap
System simply records and tracks your ingestions of ID-Capsules when you are
wearing the ID-
Cap Reader. · Talk with your doctor or pharmacist if you have questions about
the ID-Cap System. · With your consent, your healthcare provider may be
monitoring your ingestion events using the ID-Cap System
and may contact you regarding the information received. · The ID-Cap System
may occasionally fail to detect or record an ingestion. If this occurs or if
you forget to
use the Reader, DO NOT take an additional ID-Capsule. There is no need to
repeat the ingestion. Wait until your next scheduled ingestion. If an
ingestion is not detected or not reported in the App, you can manually record
the ingestion event in the ID-Cap App, if applicable.
Always take your medication as prescribed regardless of the status and use of
the ID-Cap SystGENERAL WARNINGS AND PRECAUTIONS
WARNINGS and PRECAUTIONS Ensure that you read and fully understand the
warnings and precautions outlined here. If you have any questions about how to
safely and effectively use the ID-Cap System,
ask your doctor or pharmacist.
General System
· DO NOT use if you are pregnant or breastfeeding.
9
· DO NOT use as the sole basis for healthcare decisions or treatment changes
related to ingestion events. Detection accuracy is less than 100%. Patients
instructed by a doctor to take the ID-Capsule to mark an event signaled by the
co-incidence with or co-ingestion with the ID-Capsule may selectively adhere
to one or the other. Therefore, ingestion event data should be interpreted
with caution. Patients should discuss ingestion events with their doctor prior
to making changes in their treatment or health behaviors.
· Avoid exposure to known sources of electromagnetic interference (EMI) with
medical devices, such as magnetic resonance imaging (MRI) systems, diathermy,
lithotripsy, electrocautery, Radio Frequency Identification (RFID) devices,
and electromagnetic security systems, such as metal detectors. Note that the
presence of RFID devices may not be obvious.
· The ID-Cap System has not been studied in individuals who have implanted
electronic medical devices. · Keep components out of reach of children. · DO
NOT use the ID-Cap System while on aircraft.
ID-Cap Reader
· Keep the Reader powered ON and charged at all times, so it is ready for use.
The Reader may not function properly and may not record ingestion events if it
does not have a sufficient charge.
· Do not turn the Reader off at any time as it may impact the proper
functioning of the System. · Do not attempt to replace the battery or open the
Reader case. The Reader contains no user-serviceable parts. · The ID-Cap
Reader is designed to be weather resistant, but DO NOT submerge in or expose
to excessive
amounts of water or other liquids. · If the ID-Cap Reader or any of the
charging accessories becomes damaged during normal use, discontinue use
of the Reader, and contact your healthcare provider for a replacement.
ID-Capsule
· DO NOT ingest if you are unable to take oral medications. · DO NOT tamper
with, open, or expose to any liquid before ingestion. · DO NOT take more than
five ID-Capsules per day with a minimum of 90 minutes between ID-Capsule
ingestions to ensure that each ID-Tag is appropriately identified, and each
ingestion event is detected. · Take with a sufficient amount of water. DO NOT
chew the ID-Capsules. · DO NOT take if you are allergic to any of the
ingredients in the ingestible ID-Capsule.
10
INSTRUCTIONS FOR USE OF THE ID-CAP SYSTEM WITH THE ID-CAP APP Initial Setup of
the ID-Cap Reader and its Charging Clip
1. Begin by removing all contents from the ID-Cap Reader kit. Save the Reader
kit packaging.
2. Fully charge your reader with the included charging clip.
Reader Display · The reader display will show a lightning bolt over the
battery icon when charging. · IMPORTANT: Always keep the Reader powered ON. DO
NOT turn the Reader OFF. The Reader
conserves power when not in use. 3. Download the “IDCap App” on your
smartphone.
Initial Setup of the ID-Cap App
PROCEED with these steps if you are using your smartphone (and the ID-Cap App)
to receive and communicate ingestion event information. This will allow
authorized users to view and track your ingestion events when they occur and
as they are recorded by the ID-Cap Reader. To download and install “IDCap App”
on your smartphone, complete the following steps:
1. Download “IDCap App” from the App Store or Google Play. Search for “IDCap
App” without any deviations in spelling or spacing and select the correct app
(“IDCap App”) or scan the QR code below with your smartphone to download the
application.
11
2. Open the App and follow the
prompts to enter your 6-
digit Reader ID and 12-digit App activation code. All codes needed to activate the App can be
found on the GOLD sticker on the side of the Reader Kit box. The Reader Kit box is the
cardboard box in which the Reader and accessories were packaged. The codes found that you
will need to enter when prompted include the Reader ID, App activation code, and Bluetooth
pairing code (which will be entered after you activate the App).
3. If you cannot locate these codes, contact the individual who dispensed the Reader to you.
4. Accept the terms of use and agreement regarding the privacy policy to proceed.
5. If prompted, you must allow permissions and notifications for App activation.
Pair Your Reader and Phone Before Use
With a simple Bluetooth connection to your phone, your ID-Cap Reader sends
information to the ID-Cap App to help you keep track of your ingestion events.
With your permission, this information will be shared with your healthcare
providers to help them improve your health outcomes.
Before pairing your Reader with your phone, make sure you have completed the setup steps listed on the previous page to power up the Reader and activate the App.
Pair the Reader with your phone by completing the following steps:
1. Ensure the Reader and your phone are within 10 feet of each other.
2. Ensure Bluetooth on your phone is turned ON (i.e., enabled).
3. Ensure Location Services on your phone are turned ON. 4. If the ID-Cap App
is not running, open it. 5. When prompted, enter the Bluetooth pairing code
(032669) in the ID-Cap
App to complete the pairing, the Bluetooth icon on the Reader will stop
The Bluetooth icon on the reader display will blink until it is
connected to your smartphone.
blinking.
6. Your phone will automatically pair with the Reader.
7. When successfully paired the home screen of the ID-Cap App will indicate that the Reader is
connected and will display the Reader battery level as shown to the right
Routine Use of the ID-Cap System with the ID-Cap App
Use your Reader every time you ingest an ID-Capsule to record the ingestion
event. Before first use, make sure you follow the instructions for setting up
your Reader and ID-Cap App.
12
FOR EACH INGESTION EVENT, complete the six steps outlined below to record the ID-Capsule ingestion on the Reader and the ID-Cap App. Each of these steps is described in more detail in the text below this summary graphic.
1. Confirm Reader is ON, paired, and has at least one bar of battery.
2. Place reader around your neck.
3. Ingest ID-Capsule as prescribed by your doctor. Take one ID-Capsule at the
scheduled time or at the same time as the
ingestion event to be tracked.
4. Look for the pill detection message on the reader indicating a confirmed
ingestion. This may take several minutes. Look for the ingestion event record
in the App. The pill icon will display for one hour and then revert back to
time
of day.
5. You may remove the Reader after the pill detection message is received or
after 30 minutes if no
message is received.
6. Place Reader on the supplied charging clip. Leave Reader powered ON. Reader will not fully charge unless it is ON.
If you forget to put on the Reader before you ingest the ID-Capsule, you may still be able to detect the ingestion. If this occurs, put the Reader on as soon as you remember. The ingestible sensor sends a signal for an average of nearly 30 minutes after it is swallowed, so there is still an opportunity for the Reader to detect it.
WARNING If you have taken an ID-Capsule and have not received confirmation of the ingestion on the Reader and/or App, DO NOT take an additional ID-Capsule until the time your next dose is scheduled. There is no need to repeat the ingestion if it was not recorded properly. If an ingestion is not detected or not reported in the App, you may be able to manually record the ingestion event in the ID-Cap App if applicable.
Manually Recording an Ingestion Event in the ID-Cap App (if applicable, your
Doctor will inform and train you if this is available to you)
The System may occasionally fail to detect or record an ingestion event. If
this occurs or if you forget to wear your
13
Reader when taking an ID-Capsule, DO NOT take an additional ID-Capsule. There is no need to repeat the ingestion. You can manually record the ingestion event in the App if applicable.
To manually record an ingestion event in the ID-Cap App, follow these steps:
1. Select the + sign in the top right corner of the App Home page. 2. Enter
the date and time that the ID-Capsule was taken.
3. Remember that the ingestion event must be recorded within two days of the
ingestion in the ID-Cap App and cannot be deleted or edited once it is
manually entered.
The type of ingestion event recorded in the App can be identified by the icon displayed at the end of each ingestion event listing on the Home, History, and Calendar screens. Manually recorded ingestion events are displayed with the icon of the person and check mark. Reader-detected ingestion events are displayed with the icon of the Reader.
Symbol for manually entered ingestion event
Symbol for Reader-detected ingestion event
Charging the ID-Cap Reader
The Reader contains a rechargeable battery. Keep the Reader charged so it is
ready for use. It is recommended to leave the Reader on the Reader charging
clip when not in use as shown below.
Only use the supplied charging clip, USB cable, and AC adapter to charge the
Reader. Using a charger other than the one provided with your Reader may
affect the charging of the Reader. When you are not using the Reader to detect
an ingestion, it is recommended to keep it on the charging clip. The Reader
kit contains the following three items for charging the Reader: a charging
clip, an AC adapter or wall plug, and a USB cable.
Setting up the charger:
14
· Connect the charging pad to the wall plug using the supplied USB cable.
· Plug the charging pad into a standard electrical outlet at a convenient
location.
To charge the Reader:
· Attach the Reader on the charging clip as shown above. · After the Reader
reaches full charge, the Charging Complete screen will be displayed on the
reader screen.
· The Reader battery has a life of approximately 24 hours on a full charge.
The Reader battery level is displayed on the Home screen of the Wrist reader.
The battery level must be at least 1 bar to detect an ingestion.
Do not attempt to replace the battery or open the Reader case. The Reader
contains no user-serviceable parts.
If the reader charge level drops below a level where it can detect a pill, the
critical battery level screen (see below) will be displayed momentarily when
the button is pressed. This screen will be displayed until the reader is
charged.
Device Return
After your last ID-Capsule ingestion, return the ID-Cap Reader, charging clip,
and accessories to your healthcare provider or to etectRx by packaging it in
the Reader Kit box and returning it as instructed.
Monitoring the Activity of the Device
Reader status information and ingestion event data are sent to the
manufacturer and authorized users who have been granted access to your
ingestion event data. This information includes movement of the device as
measured by an accelerometer within the device and battery charging status, as
well as additional information about the Reader and its operation. There is
not a Global Positioning System (GPS) in the Reader, and the Reader does not
collect any location information. The ID-Cap App does not capture or track any
location information.
As with all wireless communications, an ingestion event message will
occasionally not be received by the ID-Cap Reader. If this occurs more often
than just occasionally, contact your healthcare provider or etectRx. If you
are experiencing problems with your ID-Cap System, please refer to the
Troubleshooting Section of this User Manual or contact your doctor.
Conditions Affecting Device Performance and Use
The following conditions may affect device performance, the integrity of the
data collected by the System, and/or user interactions with the ID-Cap System:
· Failure to follow instructions for use · Technical issues affecting any of
the following:
o ID-Cap Reader (e.g., low battery) o ID-Tag, ID-Capsule, or messaging between
the ID-Tag and Reader o Smartphone and/or ID-Cap App when using the device
with the ID-Cap App · Connectivity between the Reader and smartphone ·
Distractions in the user environment that could interrupt use
15
Maintenance
The ID-Cap Reader contains no user-serviceable parts. For maintenance of the
device, please contact the manufacturer, etectRx. ID-Cap Reader should be
returned to etectRx for cleaning and recertification before assignment to the
next patient.
Cleaning the ID-Cap Reader
The ID-Cap Reader should be cleaned periodically as needed. To clean the
Reader, DO NOT place under water. Use a clean cloth dampened lightly with
water and household soap or a mild detergent to wipe the exterior of the
Reader case. Immediately dry it off with a clean, dry cloth.
Storage and Handling
Storage and Operating Requirements
The ID-Cap System is intended for storage and operation in a room-temperature
environment. Store ID-Capsules in original sealed containers. ID-Capsules may
be sensitive to moisture and humidity.
Please refer to the storage and handling information in Appendix A for
environmental conditions required for proper operation, storage, and transport
of ID-Cap Readers and ID-Capsules.
Useful Life and Shelf Life
The ID-Cap Reader has a shelf life of 18 months from the date of manufacture.
The ID-Cap Reader has a service life of 12 months from initial patient use and
must be returned to the manufacturer at the conclusion of the scheduled
ingestions. The ID-Cap Reader contains no user-serviceable parts.
The ID-Capsules have a shelf life of three months from the date of manufacture
when stored in their original packaging. Refer to the use date on the ID-
Capsule container. The following signs may indicate a failing device:
· Failure to mark ingestion events · Discoloration or deformity of the ID-
Capsule shell · Discoloration or deformity of the Reader or its materials ·
Display screen not showing
Disposal of Waste Products
Do not dispose of the ID-Cap Reader, any Reader accessories, or any ID-
Capsules. This medical device may contain substances that could be harmful to
the environment or human health if improperly handled at the end of the
product’s life. To avoid release of such substances into the environment and
to reduce the use of natural resources, all devices, both used and unused,
should not be disposed with household waste.
Compliant Use
Ordered by Doctor or Licensed Practitioner
Federal law (USA) restricts this device to be provided by prescription only on
the order of a licensed practitioner.
ATTENTION CAUTION: Your healthcare provider should evaluate you, as well as
the risks and benefits of the device, to ensure that the decision to use the
ID-Cap System is appropriate.
16
Avoiding Unsafe Use Conditions
The ID-Cap System is not intended for use in the diagnosis or treatment of any
medical conditions. The ID-Cap System has not been tested or approved for use
in the presence of strong magnetic or electrical fields.
The Reader contains no user-serviceable parts. Do not attempt to replace the battery or open the Reader case.
Your ID-Cap Reader may be worn during most activities, including exercise, but should be removed prior to any activities that would expose the Reader to water, including showering, bathing, and swimming.
Important Safety Instructions for Compliant Use
Electromagnetic Compatibility
The ID-Cap System is compliant with electromagnetic compatibility standards
specified in IEC 60601-1, Medical electrical equipment – Part 1: General
requirements for basic safety and essential performance.
ATTENTION
CAUTION: Use caution when operating the ID-Cap System near other electrical
equipment. Check the instructions or ask your doctor to make sure the devices
are compatible.
There are no known significant risks of reciprocal interference posed by the ID-Cap System. When using the ID-Cap System near other electrical equipment, consult the manufacturer’s instructions for Electromagnetic Compatibility of all devices to ensure potential electromagnetic or other interference is avoided or minimized.
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the ID-Cap Reader, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result. Use of accessories, transducers, and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
Electrical Connection
ATTENTION
CAUTION: The charging clip should be plugged into a power outlet using the
supplied AC adapter. Plugging the charging pad into the USB port of another
electronic device may prevent it from charging the Reader or may increase the
time required to charge the Reader.
Charging
ATTENTION
CAUTION: The reader must be turned ON to achieve full charge. If the reader is
turned off it will charge only partially until it is powered ON.
Environment
ATTENTION
CAUTION: Do not submerge the Reader in water. The Reader is designed to
operate between 41-104 degrees Fahrenheit.
Malfunction
The ID-Cap System is designed to have basic user interaction. Refer to the
Troubleshooting Section in this User Manual if you experience a problem. If
the device is not working properly, contact your doctor, or the manufacturer.
17
TROUBLESHOOTING
If you experience difficulties or device malfunction, refer to the information below or contact your doctor. To report an injury from the use of the device contact your doctor or study team.
ID-Capsule
Problem You experience clinical worsening or new clinical symptoms during use
of the ID-Cap System.
Solution Seek medical attention.
Problem
Solution
You are confused about how or when Talk with your doctor
to take your ID-Capsules.
Problem The ID-Capsule gets wet or damaged.
Solution Do not take the wet or damaged ID-Capsule. Swallow a replacement capsule if provided.
ID-Cap Reader
Problem The ID-Cap Reader will not power on.
Solution Option 1: Press and hold the power button for three seconds.
Option 2: Ensure that the device charger is set up and connected correctly and
place the Reader on the charging clip in the correct orientation. After
charging for 30 minutes, press and hold the power button for three seconds to
turn the Reader on.
Option 3: Contact your doctor or etectRx for a replacement device if the reader does not turn on.
Problem The display screen on the ID-Cap Reader is not working.
Solution Contact your doctor or etectRx for a replacement device.
Problem The ID-Cap Reader is shut off or its battery fully discharges.
Solution Charge the reader if needed. Attempt to power on the Reader by holding the power button for three seconds.
If your Reader is being used with the ID-Cap App on your smartphone, pair the Reader with the smartphone before dosing.
NOTE: Always keep the Reader powered on and sufficiently charged.
Mobile Application
Problem An ingestion event appears on the reader but is not displayed in the
IDCap App
Solution Option 1: In your home screen on your ID-Cap App, check your ID-Cap Reader battery status for sufficient power and ensure an active connection between the ID-Cap Reader and the smartphone. If no connection is indicated, close the App completely and re-start the App.
18
Option 2: Restart your phone: press and hold the power button to turn the phone off and press and hold the button to turn it back on. Re-start the ID- Cap App.
Option 3: If the app home screen indicates that the device is not paired, check that Bluetooth on the phone is turned on. If Bluetooth is off, turn it on. If Bluetooth is on but not connected, shut down app, go to settings and forget the reader device, restart app, and enter pairing code when prompted.
Option 4: If the ingestion event does not populate after completing Options 1-3, report the ingestion event in the ID-Cap App as a manual entry (if applicable)
Problem An ingestion event is not detected by the reader after waiting 30 minutes
Solution If the ingestion event does not appear on the reader, please report the ingestion event in the ID-Cap App as a manual ingestion event entry (if applicable)
USER ASSISTANCE
If you need assistance with setting up or using the ID-Cap System or to report
any problems operating the device, please contact your doctor. To request
technical information or assistance with maintenance of the device, or to
report unexpected events or quality issues, please contact the manufacturer,
etectRx. Email: support@etectrx.com
MANUFACTURER
etectRx, Inc. 747 SW 2nd Ave, Suite 365T, IMB 24, Gainesville, FL, 32601 Ph:
352-443-5713 www.etectrx.com US Patent
No. 9743880 | Made in USA
19
NOTICES Warranty
etectRx makes no warranty for any data or information that is collected
erroneously by the device, or misuse or malfunction as a result of abuse,
accidents, alteration, misuse, neglect, or failure to maintain the device as
instructed. Users of the ingestible sensor and/ or ID-Cap Reader who
experience clinical worsening or new clinical symptoms should seek medical
attention. Healthcare providers should exercise their clinical judgment in
interpreting and using any data from the ID-Cap System for clinical decision-
making.
Trademark
All ID-Cap System product names appearing in this User Manual are trademarks
owned by etectRx unless otherwise noted. Nothing contained herein shall be
construed as conferring by implication, estoppel, or otherwise, any license or
right under any patent or trademark of etectRx or any third party.
Copyright
Copyright © 2019 etectRx, Inc, Gainesville, FL, United States. All rights
reserved. etectRx hereby authorizes users to copy this User Manual for the
express purpose of safely and effectively using the ID-Cap System and only for
non-commercial use, provided any copy of these materials which you make shall
retain all copyright and other proprietary notices contained herein. This User
Manual cannot otherwise be copied, distributed, displayed, modified, posted,
or transmitted or used for any public or commercial purpose without prior
permission of etectRx. Nothing contained herein shall be construed as
conferring any license or right under any copyright of etectRx.
Quality Assurance
etectRx has established and follows a quality system to help ensure that our
products consistently meet applicable requirements and specifications. ID-Cap
Readers, ID-Tags, and ID-Capsules are manufactured and assembled in validated
production environments. Current good manufacturing practices (CGMPs) are
followed throughout the process to assure that our products will meet high
quality standards. Consistency, uniformity, and conformance to specifications
are maintained at etectRx through process monitoring and control checks that
occur throughout production processes.
20
ETLB7018.01 OCT 2022
APPENDIX A: TECHNICAL INFORMATION
Device Classification
CAUTION: Federal (United States) law restricts this device to sale by or on
the order of a licensed practitioner.
The ID-Cap System is categorized as a Class II medical device by the United States Food and Drug Administration and identified as an ingestible event marker under 21 CFR 880.6305. An ingestible event marker is a prescription device used to record time-stamped, patient-logged ingestion events. The ingestible component links wirelessly through radio frequency communication to an external recorder which records the date and time of ingestion. The patient is the intended operator of the ID-Cap System.
Symbols Reference Guide
Symbol
Meaning
Caution Consult Accompanying Documents – Hazardous situation which can cause
material damage or lead to minor or moderate injury
Warning Consult Accompanying Documents – Hazardous situation which can cause a serious or fatal injury
Model Number
Manufacturer
Date of Manufacture Serial Number Class II Equipment
Keep Dry
CAUTION: United States federal law restricts this device to sale by or on the
order of a licensed practitioner.
Not for General Waste
Lot Number Read Instructions Before Use
21
Symbol
Product Does Not Contain Latex
Meaning
Independent Lab Tested by F-Squared Labs
Use by Date
Temperature Range Relative Humidity Range Atmospheric Pressure Range Do Not Use If Package Is Damaged
Non-Sterile Product Shock Protection Type BF Applied Part Single Use Only MRI Unsafe Do Not Wear During Magnetic Resonance Imaging (MRI) Procedures
Emits Radio Waves
Ingress Protection Rating of IP53
The ID-Cap Reader has an Ingress Protection rating of IP53. This means that
the enclosure is protected from limited dust ingress, and it has been rated
for protection from water spray less than 60 degrees from vertical. Stand-By
This symbol identifies the button by means of which the ID-Cap Reader is
switched on in order to bring it into the stand-by condition.
22
Environmental Conditions of Use
The ID-Cap System has been designed for ambulatory use, enabling unattended
data collection for clinical applications. The ID-Cap System is intended for
storage and operation in a room-temperature environment.
Temperature
Relative Humidity
Atmospheric Pressure
ID-Cap Reader
Condition
Operating Storage Transport
Temperature
41°F 104°F -13°F 158°F -13°F 158°F
Relative Humidity (non-condensing)
15% 90% 15% 90% 15% 90%
Atmospheric Pressure
700 1060 hPa 700 1060 hPa 700 1060 hPa
If the internal temperature of the ID-Cap Reader exceeds 132°F, the device automatically shuts down and cannot be used until it cools. During this time, an overheat message is displayed until the ID-Cap Reader cools to below 113°F.
ID-Capsules
Condition Storage Transport
Temperature 59°F – 86°F 59°F – 86°F
Relative Humidity 35% 65% 35% 65%
Store ID-Capsules in original sealed containers at the recommended temperature and humidity. ID-Capsules may be sensitive to temperature changes. Properly stored and sealed containers will provide optimum performance. Avoid storing ID-Capsules under or near sources of water.
Biocompatibility
Biocompatibility and toxicity requirements have been met for the biological
evaluation of the ID-Cap System and its component materials that come into
contact with the human body. The following tests have been performed under the
International Standard ISO-10993 to demonstrate that the ID-Cap Reader, ID-
Capsule, and ID-Tag are biocompatible and non-toxic:
· Cytotoxicity, sensitization, primary buccal irritation, intracutaneous
irritation, pyrogenicity, intramuscular implantation, acute systemic toxicity,
and subacute systemic toxicity tests for the ingestible ID-Tag
· Cytotoxicity, sensitization, intracutaneous irritation, and pyrogenicity
tests for the ingestible ID-Capsule · Chemical characterization and
degradation study of the ID-Capsule and toxicological risk assessment ·
Cytotoxicity, sensitization, and primary skin irritation tests for the
wearable ID-Cap Reader
Protection Against Ingress of Solids and Liquids
The ID-Cap Reader has an Ingress Protection rating of IP53. This means that
the enclosure is protected from limited dust ingress, and it has been rated
for protection from water spray less than 60 degrees from vertical.
ATTENTION
CAUTION: For continued safety, if the enclosure becomes damaged, do not use
the ID-Cap Reader and contact the manufacturer.
23
Regulatory Information
Declaration of Conformity
etectRx declares that the ID-Cap System is compliant with the following standards:
Product Safety
AAMI/ANSI/ES 60601-1: 2005; (R) 2012 +A1, +A2
Home Use Safety
AAMI/IEC 60601-1-11: 2015
Electromagnetic Compatibility (EMC)/ Electromagnetic Immunity (EMI)
AAMI/IEC 60601-1-2: 2014
Wireless Coexistence
ANSI C63.27: 2017
Battery
IEC 62133: 2012
Biocompatibility/Toxicity
ISO 10993-1: 2009; ISO 10993-5: 2009; ISO 10993-6:
2016; ISO 10993-10: 2010; ISO 10993-11: 2017; ISO
10993-12: 2012; USP 40-NF35: 2017
Applied Part
The ID-Cap Reader may touch the patient during the normal use of the device
and is treated as an Applied Part. The ID-Cap Reader meets all requirements of
Applied Parts.
CISPR Interference Statement
Medical Electric Equipment needs special precautions regarding Electromagnetic
Compatibility (EMC) and needs to be installed and put into service according
to the EMC information provided in this section of the User Manual. Portable
and mobile RF communications equipment can affect Medical Electrical
Equipment. The ID-Cap System may be interfered with by other equipment, even
if that equipment complies with CISPR emission requirements.
FCC Interference Statement
This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
This equipment has been tested and found to comply within the limits of a Class B digital device pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to radio or television
reception (which can be determined by turning the equipment on and off), the
user is encouraged to try to correct the interference by using one or more of
the following measures:
· Reorient or relocate the receiving antenna. · Increase the separation
between the equipment and the receiver. · Connect the equipment into an outlet
on a circuit different from that to which the receiver is connected. · Consult
the dealer or an experienced radio/TV technician for help.
ATTENTION
CAUTION: Changes or modifications not expressly approved by etectRx could void
your authority to operate the equipment.
FCC Compliance Identifiers
FCC ID for the ID-Cap Reader: 2AL2U-020002 FCC ID for the ID-Tag: 2AL2U-
ET2000150
24
ISED Interference Statement
ENGLISH: This device complies with Industry Canada license-exempt RSS
standard(s). Operation is subject to the following two conditions: (1) this
device may not cause interference, and (2) this device must accept any
interference, including interference that may cause undesired operation of the
device.
FRENCH: Le présent appareil est conforme aux CNR d’Industrie Canada
applicables aux appareils radio exempts de licence. L’exploitation est
autorisée aux deux conditions suivantes : (1) l’appareil ne doit pas produire
de brouillage, et (2) l’utilisateur de l’appareil doit accepter tout
brouillage radioélectrique subi, même si le brouillage est susceptible d’en
compromettre le fonctionnement.
Electromagnetic Compatibility
The ID-Cap System has been evaluated and deemed compliant with the
requirements in EN60601-1-2 Class B for Electromagnetic Compatibility (EMC).
Medical Electrical Equipment needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information provided
in this User Manual. Portable and mobile RF communications equipment can
affect Medical Electrical Equipment. The ID-Cap Reader should not be used
adjacent to or stacked with other electromagnetic equipment. This may impact
the performance of the IDCap Reader or the other equipment. If adjacent or
stacked use with other electromagnetic equipment is necessary, verify that the
ID-Cap Reader operation is normal in the configuration in which it will be
used. The device will function normally when subject to the immunity testing
described below.
Information on the Radio Subsystem
The ID-Tag is a low power transmitter operating in the frequency range of 286
to 320 MHz, with an average RF output power of <1 mW. The modulation is ASK
(Amplitude Shift Keying).
The ID-Cap Reader incorporates a BluetoothTM radio subsystem which is
compliant with the Bluetooth Low Energy standard.
The following information is provided to satisfy the requirements of EN/IEC
60601-1-2:
The Bluetooth radio transmits and receives in the 2.4 GHz frequency band on
channels which are equally spaced at 2 MHz intervals between 2402 MHz and 2480
MHz.
The effective receive bandwidth is 1 MHz.
The transmit modulation is frequency-hopping using GFSK (Gaussian Frequency
Shift Keying) with a bandwidthbit period product BT-0.5. The Modulation index
is between 0.28 and 0.35.
The effective isotropic radiated power is 1.95 dBm (P=1.55 mW).
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The ID-Cap System is intended for use in the electromagnetic environment specified below. The customer or the user of ID-Cap Reader should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic Environment Guidance
RF Emissions CISPR 11
Group 1
The ID-Cap Reader uses RF energy for its internal functions and BLE. Its RF emissions are low and are not likely to cause any interference in nearby electronic equipment.
RF Emissions CISPR 11
Class B
Harmonic emissions IEC 61000-3-2 Voltage fluctuations/ flicker emissions
Not Applicable Not Applicable
The ID-Cap System is suitable for use in all establishments, including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes.
IEC 6100-3-3
25
Guidance and Manufacturer’s Declaration Electromagnetic Immunity
The ID-Cap System is intended for use in the electromagnetic environment
specified below. The customer or the user of the ID-Cap System should assure
that it is used in such an environment.
Immunity Test
Electrostatic discharge (ESD) IEC 61000-4-2
IEC 60601 Test Level
+/- 8 kV contact +/15 kV air
Compliance Level
+/- 8 kV contact +/- 15 kV air
Electromagnetic Environment Guidance
The relative humidity should be at least 5%.
Electrical fast transient / burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input
lines IEC 61000-4-11
+/- 2 kV for power supply lines
+/- 1 kV for input/output lines +/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth <5 % UT (>95 % dip in UT) for 0.5 cycle
40 % UT (60 % dip in UT) for 5 cycles
Not Applicable
Not Applicable
Not Applicable
70 % UT (30 % dip in UT) for 25 cycles
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
<5 % UT (>95 % dip in UT) for 5 s 30 A/m
30 A/m
Power frequency magnetic fields from common appliances in the home are not expected to affect the device.
Conducted RF IEC 61000-4-6
3 Vrms
Not
150 kHz to 80 MHz Applicable
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. Keep the ID- Cap System away from sources of high levels of power line magnetic fields (in excess of 30 A/m) to reduce the likelihood of interference.
Radiated RF IEC 61000-4-3
10 V/m
10 V/m
80 MHz to 2.7 GHz
The ID-Cap System is suitable for the electromagnetic environment of typical homes and commercial or hospital settings.
NOTE UT is the a.c. mains voltage prior to application of the test level.
26
References
- etectRx - Dosing Documentation for Clinical Trials
- etectRx - Dosing Documentation for Clinical Trials
Read User Manual Online (PDF format)